K Number
K243667

Validate with FDA (Live)

Device Name
Sonic DL
Date Cleared
2025-06-05

(190 days)

Product Code
Regulation Number
892.1000
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Sonic DL is a Deep Learning based reconstruction technique that is available for use on GE HealthCare 1.5T, 3.0T, and 7.0T MR systems. Sonic DL reconstructs MR images from highly under-sampled data, and thereby enables highly accelerated acquisitions. Sonic DL is intended for imaging patients of all ages. Sonic DL is not limited by anatomy and can be used for 2D cardiac cine imaging and 3D Cartesian imaging using fast spin echo and gradient echo sequences. Depending on the region of interest, contrast agents may be used.

Device Description

Sonic DL is a software feature intended for use with GE HealthCare MR systems. It includes a deep learning based reconstruction algorithm that enables highly accelerated acquisitions by reconstructing MR images from highly under-sampled data. Sonic DL is an optional feature that is integrated into the MR system software and activated through purchasable software option keys.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for Sonic DL, based on the provided FDA 510(k) clearance letter:

Acceptance Criteria and Device Performance for Sonic DL

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly list specific quantitative "acceptance criteria" against which a single "reported device performance" is measured in a pass/fail manner for all aspects. Instead, it presents various performance metrics and qualitative assessments that demonstrate the device's acceptable performance compared to existing standards and its stated claims. For the sake of clarity, I've summarized the implied acceptance thresholds or comparative findings from the quantitative studies and the qualitative assessments.

Metric/CriterionAcceptance Criteria (Implied/Comparative)Reported Device Performance (Sonic DL)
Non-Clinical Testing (Sonic DL 3D)
Peak-Signal-to-Noise (PSNR)Equal to or above 30dBEqual to or above 30dB at all acceleration factors (up to 12)
Structural Similarity Index Measure (SSIM)Equal to or above 0.8Equal to or above 0.8 at all acceleration factors (up to 12)
ResolutionPreservation of resolution grid structure and resolutionPreserved resolution grid structure and resolution
Medium/High Contrast DetectabilityRetained compared to conventional methodsRetained at all accelerations; comparable or better than conventional methods
Low Contrast DetectabilityNon-inferior to more modestly accelerated conventional reconstruction methods at recommended acceleration ratesMaintained at lower acceleration factors; non-inferior at recommended rates (e.g., 8x Sonic DL 3D ~ 4x parallel imaging; 12x Sonic DL 3D ~ 8x parallel imaging)
Model Stability (Hallucination)Low risk of hallucination; dataset integrity preservedLow risk of hallucination; dataset integrity preserved across all cases
Clinical Testing (Sonic DL 3D) - Quantitative Post Processing
Volumetric Measurements (Brain Tissues) - Relative MAE 95% CILess than 5% for most regions (brain tissues)Less than 5% for most regions
Volumetric Measurements (HOS) - Relative MAE 95% CILess than 3% for Hippocampal Occupancy Score (HOS)Less than 3% for HOS
Intra-class Correlation Coefficient (ICC)Exceeded 0.75 across all comparisonsExceeded 0.75 across all comparisons
Clinical Testing (Sonic DL 3D) - Clinical Evaluation Studies (Likert-score)
Diagnostic QualityImages are of diagnostic qualitySonic DL 3D images are of diagnostic quality (across all anatomies, field strengths, and acceleration factors investigated)
Pathology RetentionPathology seen in comparator images can be accurately retainedPathology seen in ARC + HyperSense images can be accurately retained
Decline with AccelerationRetain diagnostic quality overall despite declineScores gradually declined with increasing acceleration factors yet retained diagnostic quality overall
Clinical Claims
Scan Time ReductionSubstantial reduction in scan timeYields substantial reduction in scan time
Diagnostic Image QualityPreservation of diagnostic image qualityPreserves diagnostic image quality
Acceleration FactorsUp to 12xProvides acceleration factors up to 12

2. Sample Size Used for the Test Set and Data Provenance

  • Quantitative Post Processing Test Set:
    • Sample Size: 15 fully-sampled datasets.
    • Data Provenance: Retrospective, acquired at GE HealthCare in Waukesha, USA, from 1.5T, 3.0T, and 7.0T scanners.
  • Clinical Evaluation Studies (Likert-score based):
    • Study 1 (Brain, Spine, Extremities):
      • Number of image series evaluated: 120 de-identified cases.
      • Number of unique subjects: 54 subjects (48 patients, 6 healthy volunteers).
      • Age range: 11-80 years.
      • Gender: 26 Male, 28 Female.
      • Pathology: Mixture of small, large, focal, diffuse, hyper- and hypo-intense lesions.
      • Contrast: Used in a subset as clinically indicated.
      • Data Provenance: Retrospective and prospective (implied by "obtained from clinical sites and from healthy volunteers scanned at GE HealthCare facilities"). Data collected from 7 sites (4 in United States, 3 outside of United States).
    • Study 2 (Brain):
      • Number of additional cases: 120 cases.
      • Number of unique subjects: From 30 fully-sampled acquisitions.
      • Data Provenance: Retrospective, collected internally at GE HealthCare, 1.5T, 3.0T, and 7.0T field strengths.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Quantitative Post Processing: The "ground truth" here is the fully sampled data and the quantitative measurements derived from it. No human experts are explicitly mentioned for establishing this computational 'ground truth'.
  • Clinical Evaluation Studies:
    • Study 1: 3 radiologists. Their specific qualifications (e.g., years of experience, subspecialty) are not provided in the document.
    • Study 2: 3 radiologists. Their specific qualifications are not provided in the document.

4. Adjudication Method for the Test Set

The document does not explicitly state a formal adjudication method (like 2+1 or 3+1). For the clinical evaluation studies, it mentions that "three radiologists were asked to evaluate the diagnostic quality of images" and "radiologists were also asked to comment on the presence of any pathology." This suggests individual assessments were either aggregated, or findings were considered concordant if a majority agreed, but a specific arbitration or adjudication process for disagreements is not detailed.


5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

  • Was an MRMC study done? Yes, the two Likert-score based clinical studies involved multiple readers (3 radiologists) evaluating multiple cases (120 de-identified cases in Study 1 and 120 additional cases in Study 2) for comparative effectiveness against ARC + HyperSense.
  • Effect size of human readers improvement with AI vs. without AI assistance: The document states that "Sonic DL 3D images are of diagnostic quality while yielding a substantial reduction in the scan time compared to ARC + HyperSense images." It also noted that "pathology seen in the ARC + HyperSense images can be accurately retained in Sonic DL 3D images." However, it does not quantify the effect size of how much human readers improve with AI assistance (Sonic DL) versus without it. Instead, the studies aim to demonstrate non-inferiority or comparable diagnostic quality despite acceleration. There's no performance gain stated explicitly for the human reader in terms of diagnostic accuracy or confidence when using Sonic DL images compared to conventional images; rather, the benefit is in maintaining diagnostic quality with faster acquisition.

6. Standalone (Algorithm Only) Performance Study

  • Was a standalone study done? Yes, extensive non-clinical testing was performed as a standalone assessment of the algorithmic performance. This included evaluations using:
    • Digital reference objects (DROs) and MR scans of physical ACR phantoms to measure PSNR, RMSE, SSIM, resolution, and low contrast detectability.
    • A task-based study using a convolutional neural network ideal observer (CNN-IO) to quantify low contrast detectability.
    • Reconstruction of in vivo datasets with unseen data inserted to assess model stability and hallucination risk.
      These studies directly evaluated the algorithm's output metrics and behavior independently of human interpretation in a clinical workflow, making them standalone performance assessments.

7. Type of Ground Truth Used

  • Non-Clinical Testing:
    • Quantitative Metrics (PSNR, RMSE, SSIM, Resolution, Contrast Detectability): Fully sampled data was used as the reference "ground truth" against which the under-sampled and reconstructed Sonic DL 3D images were compared.
    • Model Stability (Hallucination): The "ground truth" was the original in vivo datasets before inserting previously unseen data, allowing for evaluation of whether the algorithm introduced artifacts or hallucinations.
  • Quantitative Post Processing (Clinical Testing):
    • Fully sampled data sets were used as the reference for comparison of volumetric measurements with Sonic DL 3D and ARC + HyperSense images.
  • Clinical Evaluation Studies (Likert-score based):
    • The implied "ground truth" was the diagnostic quality and presence/absence of pathology as assessed by the conventional ARC + HyperSense images, which were considered the clinical standard for comparison. The radiologists were essentially comparing Sonic DL images against the standard of care images without a separate, absolute ground truth like pathology for every lesion.

8. Sample Size for the Training Set

The document does not specify the sample size used for training the Sonic DL 3D deep learning model. It only mentions that Sonic DL is a "Deep Learning based reconstruction technique" and includes a "deep learning convolutional neural network."


9. How the Ground Truth for the Training Set Was Established

The document does not describe how the ground truth for the training set was established. It is standard practice for supervised deep learning models like Sonic DL to be trained on pairs of under-sampled and corresponding fully-sampled or high-quality (e.g., conventionally reconstructed) images, where the high-quality image serves as the 'ground truth' for the network to learn to reconstruct from the under-sampled data. However, the specifics of this process (e.g., data types, annotation, expert involvement) are not mentioned in this document.

FDA 510(k) Clearance Letter - Sonic DL

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

June 5, 2025

GE Medical Systems, LLC
Emily Fang
Regulatory Affairs Leader
3200 N. Grandview Blvd
Waukesha, Wisconsin 53188

Re: K243667
Trade/Device Name: Sonic DL
Regulation Number: 21 CFR 892.1000
Regulation Name: Magnetic Resonance Diagnostic Device
Regulatory Class: Class II
Product Code: LNH
Dated: May 6, 2025
Received: May 6, 2025

Dear Emily Fang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

Page 2

K243667 - Emily Fang Page 2

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

Page 3

K243667 - Emily Fang Page 3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Daniel M. Krainak, Ph.D.
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K243667

Device Name
Sonic DL

Indications for Use (Describe)

Sonic DL is a Deep Learning based reconstruction technique that is available for use on GE HealthCare 1.5T, 3.0T, and 7.0T MR systems. Sonic DL reconstructs MR images from highly under-sampled data, and thereby enables highly accelerated acquisitions. Sonic DL is intended for imaging patients of all ages. Sonic DL is not limited by anatomy and can be used for 2D cardiac cine imaging and 3D Cartesian imaging using fast spin echo and gradient echo sequences. Depending on the region of interest, contrast agents may be used.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

Sonic DL
Traditional 510(k) Premarket Notification
Page 1 of 6

K243667

510(k) Summary

In accordance with 21 CFR 807.92, the following summary of information is provided:

Date: November 25, 2024

Submitter: GE Medical Systems, LLC
3200 N. Grandview Blvd.
Waukesha, WI 53188

Primary Contact: Emily Fang
Regulatory Affairs Leader
609-285-9878
Emily.Fang@gehealthcare.com

Secondary Contact: Glen Sabin
Regulatory Affairs Director
262-894-4968
Glen.Sabin@gehealthcare.com

Device Trade Name: Sonic DL
Common/Usual Name: MR System
Classification Name: Magnetic Resonance Diagnostic Device
Regulation Number: 21 CFR 892.1000
Product Code: LNH
Predicate Device(s): Sonic DL (K223523)

Device Description

Sonic DL is a software feature intended for use with GE HealthCare MR systems. It includes a deep learning based reconstruction algorithm that enables highly accelerated acquisitions by reconstructing MR images from highly under-sampled data. Sonic DL is an optional feature that is integrated into the MR system software and activated through purchasable software option keys.

Sonic DL CineSonic DL 3D
Pulse Sequence CompatibilityFIESTA CineFast Spin Echo (3D FSE)Gradient Echo (3D GRE)
Anatomic CoverageCardiacAll anatomies
Field Strength Compatibility1.5T, 3.0T1.5T, 3.0T, 7.0T
Contrast CompatibilityYesYes
Maximum Acceleration12Up to 12

Page 6

Sonic DL
Traditional 510(k) Premarket Notification
Page 2 of 6

Indications for Use

Sonic DL is a Deep Learning based reconstruction technique that is available for use on GE HealthCare 1.5T, 3.0T, and 7.0T MR systems. Sonic DL reconstructs MR images from highly under-sampled data, and thereby enables highly accelerated acquisitions. Sonic DL is intended for imaging patients of all ages. Sonic DL is not limited by anatomy and can be used for 2D cardiac cine imaging and 3D Cartesian imaging using fast spin echo and gradient echo sequences. Depending on the region of interest, contrast agents may be used.

Comparison of Technological Characteristics

The proposed Sonic DL feature that is the subject of this 510(k) is a modification to the earlier version of the feature described in the predicate device submission, K223523. The predicate device used deep learning convolutional neural networks to reconstruct MR images from highly under-sampled 2D cardiac cine acquisitions (also known as "Sonic DL Cine"). The proposed Sonic DL feature has been extended to include a new deep learning convolutional neural network for use with 3D Cartesian acquisitions ("Sonic DL 3D"). The proposed Sonic DL includes both Sonic DL Cine, unchanged from the predicate device, and the new Sonic DL 3D.

Summary of Nonclinical Testing

The Sonic DL 3D model has been evaluated using data generated from digital reference objects (DROs) and MR scans of physical ACR phantoms. Comparative image quality metrics (Peak-Signal-to-Noise (PSNR), Root-Mean-Square Error (RMSE), and Structural Similarity Index Measure (SSIM)), resolution, and low contrast detectability were measured at varied acceleration factors, contrasts, and noise levels. Measurements were taken using fully sampled data as reference and compared against a conventional GE HealthCare acquisition scheme and reconstruction algorithm using ARC + HyperSense.

The nonclinical testing indicated that Sonic DL 3D at all acceleration factors (up to 12) demonstrated PSNR and SSIM equal to or above 30dB and 0.8 respectively, preserved resolution grid structure and resolution. Medium or high contrast detectability was retained at all accelerations, and low contrast detectability was retained at lower acceleration factors but decreased as acceleration factor increased. In comparison to conventional reconstruction methods, Sonic DL 3D exhibited comparable or better performance in preserving small structures and low contrast detectability.

A task-based study using a convolutional neural network ideal observer (CNN-IO) was conducted to further quantify the low contrast detectability of Sonic DL 3D. Compared to parallel imaging and zero-filled reconstructions, Sonic DL 3D had superior performance in every test condition. Detectability with Sonic DL 3D decreased as the acceleration rate increased, but at a more gradual rate than parallel imaging or zero-filling. Low contrast detectability with Sonic DL at the recommended acceleration rates (tabulated below) was non-inferior to more modestly accelerated conventional reconstruction methods. AUC analysis indicates that Sonic DL 3D at 8x has comparable low contrast detectability to parallel imaging at 4x, and Sonic DL 3D at 12x is comparable in detectability to parallel imaging at 8x. While average low contrast detectability is reduced at higher acceleration rates, the use of rates above those recommended may be acceptable in certain clinical scenarios.

The stability of the Sonic DL 3D model was shown by reconstructing in vivo datasets including different anatomies, contrasts, and field strengths, with previously unseen data inserted into the image. Images were reconstructed at different acceleration factors with varied intensities and locations of the inserted

Page 7

Sonic DL
Traditional 510(k) Premarket Notification
Page 3 of 6

data. Results showed that dataset integrity was preserved across all cases, and the model exhibited a low risk of hallucination.

Overall, Sonic DL 3D was shown to accelerate 3D MR acquisitions up to 12-fold while outperforming conventional parallel imaging and compressed sensing at similar acceleration. The nonclinical testing did not identify any adverse impacts to image quality or other concerns related to safety and performance at the recommended acceleration rates; extreme acceleration rates were found to have decreased low contrast detectability.

Summary of Clinical Testing

Quantitative Post Processing

Images were evaluated to confirm that the use of Sonic DL 3D does not adversely impact downstream quantification. Volumetric measurements of key brain tissues were performed to assess whether Sonic DL 3D images have similar quantitative analysis results compared to ARC + HyperSense images. Sonic DL 3D and ARC + HyperSense images were retrospectively generated from 15 fully-sampled data sets acquired at GE HealthCare in Waukesha, USA from 1.5T, 3.0T, and 7.0T scanners. Data was also generated with and without AIR Recon DL to evaluate the compatibility of Sonic DL 3D and AIR Recon DL in the reconstruction pipeline.

Across all sampling methods and quantitative measurements, it was found that the 95% confidence interval for relative mean absolute error (MAE) was less than 5% for most regions and 3% for the Hippocampal Occupancy Score (HOS). The 95% confidence interval for the intra-class correlation coefficient exceeded 0.75 across all comparisons, aligning with acceptance criteria established in published literature. This analysis highlights the robustness of Sonic DL 3D, and that Sonic DL 3D and AIR Recon DL are mutually compatible at various acceleration factors.

Bland-Altman analysis confirmed that Sonic DL would not adversely impact quantification compared to either the fully sampled acquisitions or the clinical standard ARC + HyperSense reconstructions. The observed biases were small relative to the magnitude of the measured metrics, indicating minimal differences between methods.

Whole Brain Volume (mL)Mean Diff95% CICerebral White Matter Volume (mL)Mean Diff95% CI
Fully sampled vs ARC 2x1x1.21.1[-4.4, 6.5]Fully sampled vs ARC 2x1x1.2-1.5[-4.4, 1.5]
Fully sampled vs Sonic DL 41.4[-2.4, 5.3]Fully sampled vs Sonic DL 4-2.0[-5.2, 1.3]
Fully sampled vs Sonic DL 126.8[-4.9, 18.6]Fully sampled vs Sonic DL 12-2.3[-10.1, 5.5]
Cortical Gray Matter Volume (mL)Mean Diff95% CIVentricle Total Volume (mL)Mean Diff95% CI
Fully sampled vs ARC 2x1x1.23.0[1.2, 4.9]Fully sampled vs ARC 2x1x1.2-0.1[-0.3, 0.2]
Fully sampled vs Sonic DL 43.2[0.7, 5.6]Fully sampled vs Sonic DL 4-0.1[-0.4, 0.1]
Fully sampled vs Sonic DL 129.1[2.1, 16.0]Fully sampled vs Sonic DL 12-0.2[-0.8, 0.4]
HOC Average OccupancyMean Diff95% CI
Fully sampled vs ARC 2x1x1.20.004[-0.007, 0.014]
Fully sampled vs Sonic DL 40.007[-0.009, 0.023]
Fully sampled vs Sonic DL 120.014[-0.018, 0.046]

Page 8

Sonic DL
Traditional 510(k) Premarket Notification
Page 4 of 6

Clinical Evaluation Studies

Additionally, two Likert-score based clinical studies were performed using Sonic DL 3D images acquired from clinical sites and from healthy volunteers scanned at GE HealthCare facilities. In one study, three radiologists were asked to evaluate the diagnostic quality of images acquired and reconstructed with Sonic DL 3D and with a comparator, ARC + HyperSense, for 120 de-identified cases of brain, spine, and extremities as summarized below:

Number of image series evaluated:120
Number of unique subjects:54
Patients:48
Healthy Volunteers:6
Sites contributing data:7
United States:4
Outside of United States:3
Gender of subjects
Male:26
Female:28
Age range of subjects:11-80 years
Pathology:Subjects from clinical sites included examples of various pathologies representing a mixture of small, large, focal, diffuse, hyper- and hypo-intense lesions
Contrast:Contrast agents were used in a subset of the data, as clinically indicated
Equipment Used:GE HealthCare 1.5T, 3.0T, and 7.0T MR Systems

The results showed that Sonic DL 3D images are of diagnostic quality while yielding a substantial reduction in the scan time compared to ARC + HyperSense images.

Radiologists were also asked to comment on the presence of any pathology in the images. Analysis of these comments showed that images acquired and reconstructed with Sonic DL 3D agreed with those obtained with ARC + HyperSense across various anatomical subgroups and field strengths. These results demonstrate that pathology seen in the ARC + HyperSense images can be accurately retained in Sonic DL 3D images.

In a second study, three radiologists evaluated clinically relevant anatomic structures in Sonic DL 3D brain images. The study included 120 additional cases retrospectively generated from 30 fully-sampled acquisitions obtained at 1.5T, 3.0T, and 7.0T field strengths. Each case was generated from fully sampled data that was either (1) under-sampled to simulate Sonic DL 3D accelerations of 4 to 10 then reconstructed with Sonic DL 3D or (2) under-sampled to simulate an ARC acceleration factor of 2x1 and a HyperSense factor of 1.2. All cases were collected internally at GE HealthCare.

Despite the higher acceleration factors used, the Sonic DL 3D images received diagnostic scores in all anatomic structures and at all acceleration factors investigated. The scores gradually declined with increasing acceleration factors yet retained diagnostic quality overall. The results of this study confirm that Sonic DL 3D provides diagnostic quality images at the range of acceleration factors studied.

Page 9

Sonic DL
Traditional 510(k) Premarket Notification
Page 5 of 6

Assessment of Representative Data

In addition to the clinical evaluations described above, supplementary testing was done to further support broad use of Sonic DL 3D. Representative datasets with different anatomies, contrast weightings, with/without contrast agent, orientations, with/without AIR Recon DL, and acceleration factors up to 12 were assessed for Sonic DL 3D and compared to standard clinically appropriate protocols using ARC + HyperSense. Breast and liver data were collected from academic/clinical partners while the remaining images were acquired on a volunteer cohort internal to GE HealthCare.

Assessment of the representative data confirmed that Sonic DL 3D performs well across varied imaging conditions, retaining relevant anatomical details without structural losses or concerning artifacts.

Clinical Publications

The following peer-reviewed studies provide further quantitative and qualitative evidence of the ability of Sonic DL to improve acquisition speed and/or resolution across various clinical applications in the brain, joints, breast, prostate, and abdomen:

[1] Shintaku T, Ide S, Nagaya H, Ishimoto Y, Watanabe K, Oyu K, et al. Improved assessment of juxtacortical lesions in multiple sclerosis using highly-accelerated high-resolution double inversion recovery MR imaging with deep learning-based reconstruction. Magn Reson Med Sci. 2025.

[2] Ishimoto Y, Shintaku T, Ide S, Watanabe K, Oyu K, Kasai S, et al. Improved assessment of microbleeds by highly-accelerated 3D SWAN imaging with deep learning-based reconstruction. ISMRM 2025.

[3] Carretero L, Nunes B, Zhu X, Sanchez-Lacalle E, Sundaran D, Dholakia J, et al. 2-Minute 3D FSE Knee MRI with 10-fold accelerated Sonic DL – Rapid morphometric and qualitative assessment of Cartilage and Meniscus. ISMRM 2025.

[4] Milshteyn E, Pohl M, Cochran RL, Gaddipati A, Venkatachari A, Kolupar T, et al. Evaluation of Accelerated 2D and 3D Acquisition Strategies for T2-Weighted MRI of the Prostate. ISMRM 2025.

[5] Watanabe K, Kasai S, Umemura Y, Tatsuo S, Oyu K, Nozaki A, et al. 0.9mm isotropic 1min MPRAGE using highly-accelerated Deep learning Reconstruction for Brain Structural Analysis. ISMRM 2024.

[6] Sato K, Tanaka S, Murayama R, Takayama Y, Nozaki A, Zhu X, et al. Fast hepatobiliary phase gadoxetate-enhanced imaging under breath-holding utilizing DL reconstruction (Sonic DL): preliminary experience. ISMRM 2025.

[7] Wang P, Lu C, Keen K, Wilmes LJ, Chou SH, Chung M, et al. Shortening 3D T2-Weighted Breast MRI scan time using deep learning based reconstruction: A phantom and patient reader study. ISMRM 2025.

[8] Sato K, Tanaka S, Murayama R, Takayama Y, Nozaki A, Zhu X, et al. Quadruple arterial phase dynamic EOB imaging using a novel DL reconstruction visualizing aortic wax and wane phenomenon: preliminary results. ISMRM 2025.

Recommended Acceleration Ranges

The following table summarizes the maximum acceleration factors tested and the recommended acceleration ranges for each anatomical region. These values were established through comprehensive performance evaluations including model observer testing, quantitative volumetric analysis, and clinical image quality assessments.

Page 10

Sonic DL
Traditional 510(k) Premarket Notification
Page 6 of 6

AnatomyMaximum Acceleration TestedRecommended Default Acceleration Range
1.5T3.0T7.0T
Brain124 – 66 – 88 – 10
Spine124 – 66 – 8-
Musculoskeletal124 – 68 – 108 – 10
Abdomen124 – 68 – 10-
Pelvis124 – 66 – 8-
Breast104 – 66 – 8-

Conclusion Drawn from Performance Testing

The nonclinical and clinical testing demonstrated that Sonic DL 3D satisfies the product claims of reducing scan time while preserving diagnostic image quality, providing acceleration factors up to 12, and can enable the acquisition of higher resolution images with similar scan times.

Note that at very high acceleration factors, there may be a slight degradation in image quality affecting detectability for small low contrast lesions. Clinicians should use their best judgement in making these tradeoffs based on the specifics of the patient and clinical scenario, as is customary in clinical standard of care.

The proposed Sonic DL feature has been developed under GE HealthCare's quality management system and is at least as safe and effective as the earlier version of Sonic DL that is the legally marketed predicate device. For both the proposed Sonic DL feature and the predicate device, the primary question of safety and effectiveness is that of image quality. Performance data that were collected demonstrate the proposed Sonic DL feature provides an adequate level of image quality appropriate for diagnostic use. The performance testing did not identify any new hazards, adverse effects, safety concerns, or performance concerns that are significantly different from those associated with MR imaging in general.

Therefore, GE HealthCare believes that Sonic DL is substantially equivalent to the predicate device and is safe and effective for its intended use.

N/A