K Number
K243514

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-04-03

(141 days)

Product Code
Regulation Number
882.4560
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Salvo® Robotic Navigation Instruments are indicated for use during the preparation and placement of Salvo® Spine System screws in open or minimally invasive procedures. The Salvo® Robotic Navigation Instruments are specifically designed for use with Globus Medical's ExcelsiusGPS® Robotic Navigation Platform, which is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of an instrument holder or guide tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous procedures, provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy.

Device Description

The Salvo® Robotic Navigation Instruments are surgical instruments that are designed to be used with Globus Medical's ExcelsiusGPS® Robotic Navigation Platform. These instruments are intended to facilitate the placement of screws during spinal surgery. The instruments are reusable and offered as non-sterile, to be cleaned and steam sterilized by the end user.

AI/ML Overview

The provided FDA 510(k) clearance letter describes the Salvo® Robotic Navigation Instruments and asserts their substantial equivalence to predicate devices. However, the document does not contain the detailed information required to fill out all requested sections about acceptance criteria, specific study results, sample sizes, expert qualifications, or ground truth details. The letter primarily focuses on the regulatory submission process and the conclusion of substantial equivalence based on dimensional analysis and geometric comparison and compatibility validation testing, rather than a comprehensive performance study that would typically include the requested metrics for AI-powered devices.

Specifically, this device appears to be surgical instruments used with a robotic navigation platform, not an AI or software algorithm itself that would typically have performance metrics like sensitivity, specificity, or AUC. The "dimensional analysis and geometric comparison" and "validation testing to demonstrate compatibility" are more aligned with traditional medical device testing for form, fit, function, and safety, rather than evaluating the diagnostic or predictive performance of an AI system.

Therefore, many sections below will state "Not applicable" or "Not specified in the provided document" because the input document does not detail these aspects as would be expected for an AI/ML-driven device study.


Acceptance Criteria and Device Performance Study Details

Given the nature of the device (robotic navigation instruments, not an AI diagnostic/predictive tool), the acceptance criteria and study detailed in the provided document focus on physical characteristics, compatibility, and safety rather than AI performance metrics.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria (As implied by the document)Reported Device Performance (Summary from document)
Accuracy PerformanceDemonstrated accuracy relative to predicate devices through dimensional analysis and geometric comparison. (Implied: instruments operate within acceptable tolerances for navigation)Evaluated through dimensional analysis and geometric comparison to predicate devices, establishing safety and effectiveness for accuracy.
Platform CompatibilityCompatibility with Globus Medical's ExcelsiusGPS® Robotic Navigation Platform, including its built-in safety features.Validation testing demonstrated compatibility of the subject device with the built-in safety feature of the ExcelsiusGPS end effector.
Material & Design EquivalenceEquivalent material of manufacture and device design to predicate devices.Similar device design and dimensions, materials of manufacture, and principles of operation to predicates.
Sterilization ProcessNon-sterile and reusable; amenable to steam sterilization by the end user (identical to predicates).Non-sterile, reusable instruments to be steam sterilized by the end user.
Intended Use EquivalenceSame intended use as the primary predicate device.Same intended use as the primary predicate: preparation and placement of screws in spinal procedures with the ExcelsiusGPS platform.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not applicable/Not specified. The document refers to "dimensional analysis and geometric comparison" and "validation testing," which implies testing of instrument prototypes or production units, not a patient or image dataset.
  • Data Provenance: Not applicable/Not specified.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Not applicable/Not specified. The evaluation involved engineering analyses (dimensional, geometric, compatibility) of the instruments themselves, not interpretation of clinical data requiring expert consensus.

4. Adjudication Method for the Test Set

  • Not applicable. As no clinical images or patient data were interpreted, no adjudication method was mentioned or implied.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance.

  • No. An MRMC study was not done. The device is a surgical instrument, not an AI-assisted diagnostic or interpretative system for human readers. "Clinical testing is not applicable" is explicitly stated.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done.

  • Not applicable. This is not an algorithm or AI system.

7. The Type of Ground Truth Used

  • Engineering specifications and predicate device characteristics. The "ground truth" for the tests appears to be established engineering tolerances, mechanical specifications, and the established characteristics of the predicate devices for comparison. For compatibility, the ground truth would be the functional requirements of the ExcelsiusGPS® system's safety features.

8. The Sample Size for the Training Set

  • Not applicable. This device is not an AI/ML algorithm that requires a training set. The "design and dimensions" of the instruments would be engineered and iterated upon, not "trained" in the machine learning sense.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. As there is no training set for an AI/ML algorithm, this question is not relevant.

FDA 510(k) Clearance Letter - Salvo® Robotic Navigation Instruments

Page 1

April 3, 2025

Spine Wave, Inc.
Ronald Smith
Executive Vice President, Quality, Regulatory & Clinical Affairs
Three Enterprise Drive
Suite 210
Shelton, Connecticut 06484

Re: K243514
Trade/Device Name: Salvo® Robotic Navigation Instruments
Regulation Number: 21 CFR 882.4560
Regulation Name: Stereotaxic Instrument
Regulatory Class: Class II
Product Code: OLO
Dated: February 28, 2025
Received: February 28, 2025

Dear Ronald Smith:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K243514 - Ronald Smith Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K243514 - Ronald Smith Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali -S

Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative, Repair
and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K243514

Device Name
Salvo® Robotic Navigation Instruments

Indications for Use (Describe)

The Salvo® Robotic Navigation Instruments are indicated for use during the preparation and placement of Salvo® Spine System screws in open or minimally invasive procedures. The Salvo® Robotic Navigation Instruments are specifically designed for use with Globus Medical's ExcelsiusGPS® Robotic Navigation Platform, which is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of an instrument holder or guide tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous procedures, provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary

510(k) #: K243514
Prepared on: 2025-03-31

Contact Details

21 CFR 807.92(a)(1)

FieldValue
Applicant NameSpine Wave, Inc.
Applicant AddressThree Enterprise Drive Suite 210 Shelton CT 06484 United States
Applicant Contact Telephone203-712-1846
Applicant ContactMr. Ronald Smith
Applicant Contact Emailrsmith@spinewave.com

Device Name

21 CFR 807.92(a)(2)

FieldValue
Device Trade NameSalvo® Robotic Navigation Instruments
Common NameStereotaxic instrument
Classification NameOrthopedic Stereotaxic Instrument
Regulation Number882.4560
Product Code(s)OLO

Legally Marketed Predicate Devices

21 CFR 807.92(a)(3)

Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K220862E-GPS Navigated InstrumentsOLO
K171651ExcelsiusGPS®OLO

Device Description Summary

21 CFR 807.92(a)(4)

The Salvo® Robotic Navigation Instruments are surgical instruments that are designed to be used with Globus Medical's ExcelsiusGPS® Robotic Navigation Platform. These instruments are intended to facilitate the placement of screws during spinal surgery. The instruments are reusable and offered as non-sterile, to be cleaned and steam sterilized by the end user.

Intended Use/Indications for Use

21 CFR 807.92(a)(5)

The Salvo® Robotic Navigation Instruments are indicated for use during the preparation and placement of Salvo® Spine System screws in open or minimally invasive procedures. The Salvo® Robotic Navigation Instruments are specifically designed for use with Globus Medical's ExcelsiusGPS® Robotic Navigation Platform, which is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of an instrument holder or guide tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous procedures, provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy.

Indications for Use Comparison

21 CFR 807.92(a)(5)

The subject Salvo® Robotic Navigation Instruments has the same intended use as the primary predicate. Specifically, the devices are used during the preparation and placement of screws in open or minimally invasive spinal procedures with Globus Medical's ExcelsiusGPS® Robotic Navigation Platform. Therefore, the subject Salvo® Robotic Navigation Instruments is substantially equivalent to

Page 6

its predicate in intended use, and no new issues of safety and effectiveness are raised.

Technological Comparison

21 CFR 807.92(a)(6)

The subject Salvo® Robotic Navigation Instruments have equivalent technological characteristics to that of the predicate devices and the minor differences do not raise any new issues of safety and effectiveness. Specifically, the following characteristics are similar between the subject and predicates: device design and dimensions, materials of manufacture, and principles of operation. Identical to the predicates, the subject Salvo® Robotic Navigation Instruments are also non-sterile, reusable instruments to be steam sterilized by the end user.

Non-Clinical and/or Clinical Tests Summary & Conclusions

21 CFR 807.92(b)

The Salvo® Robotic Navigation Instruments have been evaluated through a dimensional analysis and geometric comparison to the predicate devices to establish the safety and effectiveness for accuracy performance. Validation testing was also conducted to demonstrate compatibility of the subject device with the built-in safety feature of the ExcelsiusGPS end effector.

Clinical testing is not applicable.

The overall technology characteristics and engineering analysis lead to the conclusion that the Salvo® Robotic Navigation Instruments are substantially equivalent to the predicate device.

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).