K Number
K243512

Validate with FDA (Live)

Date Cleared
2025-06-10

(210 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

FUJIFILM Endoscope Models EG-840T and EG-840TP is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

FUJIFILM Endoscope Model EG-840N is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum. This product can be inserted orally or nasally.

Never use this product for any other purposes.

Device Description

The insertion portion of the device has a mechanism (hereinafter "the bending portion") which bends the tip from right to left and up and down, and a flexible tube (hereinafter "the flexible portion") consists of the bending portion and operating portion with a knob which controls the bending portion. The forceps channel which runs through the operating portion to the tip is arranged inside the insertion portion for inserting the surgical instrument. The insertion portion of the endoscopes comes into contact with the mucosal membrane.

The tip of the insertion portion is called the "Distal end" which contains the Imaging section, Distal cap, Objective lens, Air/water nozzle, Water jet nozzle (Except EG-840N), Instrument channel outlet, and Light guide. The bending portion is controlled by knobs on the control portion/operation section to angulate the distal end to certain angles.

The Flexible portion refers to the long insertion area between the Bending portion and the Control portion (a part of Non-insertion portion). This portion contains light guides (glass fiber bundles), air/water channels, a forceps/suction channel, a CrOS image sensor, and cabling. The glass fiber bundles allow light to travel through the endoscope to illuminate the body cavity, thereby providing enough light to the CMOS image sensor to capture an image and display the image on a monitor. The forceps channel is used to introduce biopsy forceps and other endoscopic accessories, as well as providing suction.

The control portion/operating section provides a grip to grasp the endoscopes and contains mechanical parts to operate the endoscopes. This section includes a Forceps inlet, which allows endoscope accessories to be introduced. The Scope connector connects the endoscopes to the light source.

AI/ML Overview

The provided FDA 510(k) clearance letter pertains to endoscopes, which are hardware devices, not AI/ML software. Therefore, the information requested regarding acceptance criteria, study details, and data provenance for an AI/ML device is not explicitly present in the provided text.

The document focuses on demonstrating substantial equivalence to a predicate device through various non-clinical tests (electrical safety, biocompatibility, endoscope-specific testing, software validation, and reprocessing validation) rather than a clinical effectiveness study involving human readers or AI performance metrics.

However, I can extract information about the device's performance specifications that were tested to prove its general functionality and safety, which can be interpreted as fulfilling certain "acceptance criteria" for a physical medical device.

Here's a breakdown of the available information based on your request, with the caveat that it does not directly address AI/ML performance:


1. A table of acceptance criteria and the reported device performance

The document lists various performance specifications that the device met. It does not provide specific numerical acceptance criteria alongside reported performance values in a table format. Instead, it states that "The subject device met performance specifications in the following additional testing." This implies that the device did meet predefined internal thresholds for these parameters.

Acceptance Criterion (Performance Specification Tested)Reported Device Performance
Field of viewMet performance specifications
Diameter of forceps channelMet performance specifications
Uneven illuminationMet performance specifications
Bending capabilityMet performance specifications
Viewing directionMet performance specifications
Color reproducibilityMet performance specifications
Rate of suctionMet performance specifications
ResolutionMet performance specifications
Air volumeMet performance specifications
Working lengthMet performance specifications
LG outputMet performance specifications
Water volumeMet performance specifications
Electrical safetyMet specified standards (ANSI/AAMI ES 60601-1:2012, IEC 60601-1-2:2020, IEC60601-1-6:2020 and IEC 60601-2-18:2009)
BiocompatibilityMet specified standards (ISO 10993-1:2018, ISO 10993-5:2009, and ISO 10993-10:2010), in accordance with FDA guidance.
Endoscope specific testingMet specified standards (ISO 8600-1:2015, ISO 8600-3:2019, and ISO 8600-4:2014)
Software specific testing (Validation)Met specified standards (IEC 62304:2015), in accordance with FDA guidance.
Cleaning, disinfection, and sterilization instructions validationMet FDA guidance, demonstrated substantial equivalence in performance to predicate device after reprocessing.

2. Sample size used for the test set and the data provenance

  • Sample Size: Not specified for any of the tests. The document mentions "testing" or "validation activities" without detailing the number of devices or trials involved.
  • Data Provenance: Not specified. As this is a 510(k) for a physical medical device (endoscope), the testing would typically be conducted in a laboratory setting by the manufacturer, rather than involving patient data in the context of imaging performance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. The tests performed are engineering and safety validations. There is no mention of human expert involvement for establishing "ground truth" in the context of image interpretation or diagnosis, as this is not an AI/ML diagnostic tool.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is a concept typically used in clinical studies involving interpretation by multiple readers, which is not described here.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-assisted device. The document describes a traditional endoscope.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. The device itself is an endoscope, not an algorithm. Software validation was conducted (IEC 62304:2015), but this relates to the software controlling the endoscope's functions, not a standalone diagnostic AI.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the listed performance specifications (field of view, resolution, etc.), the "ground truth" would be established by objective physical measurements using calibrated equipment and engineering standards. For biocompatibility, it's based on biological response to materials, and for reprocessing, it's based on sterility and decontamination efficacy. There is no "ground truth" in the diagnostic sense as there would be for an AI/ML algorithm.

8. The sample size for the training set

Not applicable. This is not an AI/ML device that requires a training set.

9. How the ground truth for the training set was established

Not applicable.

FDA 510(k) Clearance Letter - FUJIFILM Endoscope Models

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

June 10, 2025

FUJIFILM Corporation
℅ Chaitrali Kulkarni
Sr. Regulatory Affairs Specialist
Fujifilm Healthcare Americas Corporation
81 Hartwell Ave. Suite 300
Lexington, Massachusetts 02421

Re: K243512
Trade/Device Name: FUJIFILM Endoscope Model EG-840N; FUJIFILM Endoscope Model EG-840T; FUJIFILM Endoscope Model EG-840TP
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: FDS
Dated: May 9, 2025
Received: May 9, 2025

Dear Chaitrali Kulkarni:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K243512 - Chaitrali Kulkarni
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K243512 - Chaitrali Kulkarni
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shanil P. Haugen -S

Shanil P. Haugen, Ph.D.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K243512

Device Name:
FUJIFILM Endoscope Model EG-840N;
FUJIFILM Endoscope Model EG-840T;
FUJIFILM Endoscope Model EG-840TP

Indications for Use (Describe)

FUJIFILM Endoscope Models EG-840T and EG-840TP is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

FUJIFILM Endoscope Model EG-840N is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum. This product can be inserted orally or nasally.

Never use this product for any other purposes.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number"

FORM FDA 3881 (8/23)
Page 1 of 1

Page 5

510(k) Summary

510(k)#: K243512
Prepared on: 2025-06-05

Contact Details

21 CFR 807.92(a)(1)

FieldValue
Applicant NameFUJIFILM Corporation
Applicant Address798 Miyanodai Kaisei-Machii Ashigara Kami-Gun KANAGAWA 258-8538 Japan
Applicant Contact Telephone704-517-4886
Applicant ContactMs. Chaitrali Kulkarni
Applicant Contact Emailhcusregulatoryaffairs@fujifilm.com
Correspondent NameFujifilm Healthcare Americas Corporation
Correspondent Address81 Hartwell Ave. Suite 300 Lexington MA 02421 United States
Correspondent Contact Telephone704-517-4886
Correspondent ContactMs. Chaitrali Kulkarni
Correspondent Contact Emailhcusregulatoryaffairs@fujifilm.com

Device Name

21 CFR 807.92(a)(2)

FieldValue
Device Trade NameFUJIFILM Endoscope Model EG-840N; FUJIFILM Endoscope Model EG-840T; FUJIFILM Endoscope Model EG-840TP
Common NameEndoscope and accessories
Classification NameGastroscope And Accessories, Flexible/Rigid
Regulation Number876.1500
Product Code(s)FDS

Legally Marketed Predicate Devices

21 CFR 807.92(a)(3)

Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K240142FUJIFILM Endoscope Model EG-840N; FUJIFILM Endoscope Model EG-840TFDS

Device Description Summary

21 CFR 807.92(a)(4)

A. Intended Use/indications for Use

FUJIFILM Endoscope Model EG-840T, EG-840TP is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

FUJIFILM Endoscope Model EG-840N is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum. This product can be inserted orally or nasally. Never use this product for any other purposes.

Page 6

K243512
Page 2 of 3

B. Technological Characteristics

The insertion portion of the device has a mechanism (hereinafter "the bending portion") which bends the tip from right to left and up and down, and a flexible tube (hereinafter "the flexible portion") consists of the bending portion and operating portion with a knob which controls the bending portion. The forceps channel which runs through the operating portion to the tip is arranged inside the insertion portion for inserting the surgical instrument. The insertion portion of the endoscopes comes into contact with the mucosal membrane.

The tip of the insertion portion is called the "Distal end" which contains the Imaging section, Distal cap, Objective lens, Air/water nozzle, Water jet nozzle (Except EG-840N), Instrument channel outlet, and Light guide. The bending portion is controlled by knobs on the control portion/operation section to angulate the distal end to certain angles.

The Flexible portion refers to the long insertion area between the Bending portion and the Control portion (a part of Non-insertion portion). This portion contains light guides (glass fiber bundles), air/water channels, a forceps/suction channel, a CrOS image sensor, and cabling. The glass fiber bundles allow light to travel through the endoscope to illuminate the body cavity, thereby providing enough light to the CMOS image sensor to capture an image and display the image on a monitor. The forceps channel is used to introduce biopsy forceps and other endoscopic accessories, as well as providing suction.

The control portion/operating section provides a grip to grasp the endoscopes and contains mechanical parts to operate the endoscopes. This section includes a Forceps inlet, which allows endoscope accessories to be introduced. The Scope connector connects the endoscopes to the light source.

C. Principles of Operation

The Endoscope Model EG-840N, EG-840T, and EG-840TP function on the same principles of operation as the predicate device. EG-840N, EG-840T, and EG-840TP are endoscopes for observation, diagnosis, and treatment of the upper digestive tract. After connection to the light source and video processor, the devices guide the light through the fiber bundles situated inside the insertion portion of the device. The light that emits from the distal end of the insertion portion reflects from the target region and forms an image on the CMOS image sensor through a group of object lenses placed at the distal end of the devices. An electric signal from the image sensor is transmitted to the video processor connected to the device. The video processor converts the electric signal into a video signal and displays an image on a monitor.

Intended Use/Indications for Use

21 CFR 807.92(a)(5)

FUJIFILM Endoscope Models EG-840T and EG-840TP is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

FUJIFILM Endoscope Model EG-840N is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum. This product can be inserted orally or nasally. Never use this product for any other purposes.

Indications for Use Comparison

21 CFR 807.92(a)(5)

The indication for use for the proposed device and the predicate devices are same.

Technological Comparison

21 CFR 807.92(a)(6)

The subject devices FUJIFILM Endoscope Model EG-840N, EG-840T and EG-840TP are substantially equivalent to the predicate devices based on the same intended use, indications for use, technological characteristics and materials other than the delayed reprocessing instructions and software version.

The differences in the delayed reprocessing instructions raise no new issues of safety or effectiveness. Cleaning, Disinfection, and Sterilization (hereinafter "CDS") validation testing data demonstrated that the subject devices are substantially equivalent in performance to the predicate devices.

The difference in software version has been validated through Software & Cybersecurity evaluation testing.

Thus, the subject devices FUJIFILM Endoscope Model EG-840N, EG-840T and EG-840TP are substantially equivalent to the predicate device, FUJIFILM Endoscope Model EG-840N, EG-840T and EG-840TP (K240142).

Page 7

K243512
Page 3 of 3

Non-Clinical and/or Clinical Tests Summary & Conclusions

21 CFR 807.92(b)

Electrical safety of the subject device was evaluated using following standards: ANSI/AAMI ES 60601-1:2012, IEC 60601-1-2:2020, IEC60601-1-6:2020 and IEC 60601-2-18:2009.

Biocompatibility of the subject device was evaluated using the following consensus standards: ISO 10993-1:2018, ISO 10993-5:2009, and ISO 10993-10:2010. Biocompatibility testing was conducted in accordance with the FDA guidance, Use of International Standards ISO 10993-1, "Biological evaluation of medical devices - Part 1 : Evaluation and testing within a risk management process", issued June 16,2016.

Endoscope specific testing was conducted using the following consensus standards: ISO 8600-1:2015, ISO 8600-3:2019, and ISO 8600-4:2014.

Software specific testing was conducted using the following consensus standards: IEC 62304:2015. The software validation activities were performed in accordance with the FDA Guidance, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, issued on May 11, 2005.

Validation of the cleaning, disinfection, and sterilization instructions was performed in accordance with FDA's guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling," published March 17,2015.

The subject device met performance specifications in the following additional testing:

  • Field of view
  • Diameter of forceps channel
  • Uneven illumination
  • Bending capability
  • Viewing direction
  • Color reproducibility
  • Rate of suction
  • Resolution
  • Air volume
  • Working length
  • LG output
  • Water volume

The subject devices FUJIFILM Endoscope Model EG-840N, EG-840T and EG-840TP are substantially equivalent to the predicate devices based on the same intended use, indications for use, technological characteristics and materials other than the delayed reprocessing instructions and software version.

The differences in the delayed reprocessing instructions raise no new issues of safety or effectiveness. CDS validation testing data demonstrated that the subject devices are substantially equivalent in performance to the predicate devices. The difference in software version has been validated through Software & Cybersecurity evaluation testing.

Thus, the subject devices FUJIFILM Endoscope Model EG-840N, EG-840T and EG-840TP are substantially equivalent to the predicate device, FUJIFILM Endoscope Model EG-840N, EG-840T and EG-840TP (K240142).

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.