K Number
K243490

Validate with FDA (Live)

Date Cleared
2025-06-06

(206 days)

Product Code
Regulation Number
866.3365
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LIAISON PLEX® Gram-Positive Blood Culture Assay (BCP), performed using the automated, sample-to-result LIAISON PLEX® System, is a qualitative multiplexed in vitro diagnostic test for the simultaneous detection and identification of selected gram-positive pathogens and/or selected genetic determinants associated with antimicrobial resistance in positive blood culture bottles. BCP is performed directly on blood culture media using blood culture bottles identified as positive by a continuous monitoring blood culture system and which contain gram-positive bacteria as determined by Gram stain.

The BCP Assay detects and identifies the following:

Gram Positive Resistance Markers:

  • mecA/mecC
  • vanA
  • vanB

Genera and Species:

  • Bacillus spp.
  • Enterococcus faecalis
  • Enterococcus faecium
  • Listeria spp.
  • Staphylococcus spp.
  • Staphylococcus aureus
  • Staphylococcus epidermidis
  • Staphylococcus lugdunensis
  • Streptococcus spp.
  • Streptococcus agalactiae
  • Streptococcus anginosus group
  • Streptococcus pneumoniae
  • Streptococcus pyogenes

Negative results for antimicrobial resistance genes do not indicate bacterial susceptibility as there are multiple mechanisms that can contribute to resistance.

The LIAISON PLEX® BCP Assay contains targets for the detection of genetic determinants associated with resistance to methicillin (mecA/C) and vancomycin (vanA and vanB) to aid in the identification of potentially antimicrobial-resistant organisms in positive blood culture samples. In mixed growth, the LIAISON PLEX BCP Assay does not specifically attribute vanA/vanB-mediated vancomycin resistance to either E. faecalis or E. faecium, or mecA/mecC-mediated methicillin resistance to either Staphylococcus spp., S. aureus, S. epidermidis or S. lugdunensis.

The antimicrobial resistance gene or marker detected may or may not be associated with the agent responsible for disease. Negative results for these select antimicrobial resistance gene and marker assays do not indicate susceptibility, as multiple mechanisms of methicillin and vancomycin resistance exist.

The LIAISON PLEX® BCP Assay is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial bloodstream infections (BSI). The LIAISON PLEX® BCP Assay is not intended to monitor these infections. Sub-culturing of positive blood cultures is necessary to recover organisms for antimicrobial susceptibility testing (AST), for identification of organisms not detected by the LIAISON PLEX BCP Assay, to detect mixed infections that may not be detected by the LIAISON PLEX BCP Assay, for association of antimicrobial resistance genes to a specific organism, or for epidemiological typing.

Device Description

The LIAISON PLEX® Gram-Positive Blood Culture Assay (BCP Assay) is an automated test for the detection and identification of nucleic acid from gram-positive bacteria in a positive blood culture media sample. The BCP Assay is performed directly on blood culture media using blood culture bottles identified as positive by a continuous monitoring blood culture system, and which contain gram-positive bacteria, as determined by a Gram stain.

The LIAISON PLEX® System is a fully automated, bench-top "sample-to-answer" device that performs sample preparation and microarray-based hybridization for the detection of target-specific nucleic acids. The test reagents are supplied as a single, disposable test cartridge. Target amplification is not performed as part of the BCP Assay workflow, as it is a non-amplified, direct detection test performed on the LIAISON PLEX® System.

AI/ML Overview

The provided text describes the acceptance criteria and the study that proves the LIAISON PLEX® Gram-Positive Blood Culture Assay meets those criteria. However, it's important to note that this document is a 510(k) Clearance Letter and 510(k) Summary, which focuses on demonstrating substantial equivalence to a predicate device for regulatory clearance. It is not an academic paper detailing a clinical trial with specific acceptance criteria that would typically be associated with AI/ML diagnostic performance metrics like those in a multi-reader multi-case (MRMC) study or the establishment of ground truth by multiple experts.

The study described here is for an in vitro diagnostic test for the detection of nucleic acid sequences, not for an AI/ML-based diagnostic imaging device. Therefore, many of the requested points (e.g., number of experts for ground truth, adjudication method, MRMC study, human reader improvement with AI assistance) are not applicable to this type of device and study.

I will interpret the provided information in the context of the device described (a multiplex nucleic acid assay) and its verification and validation.


Acceptance Criteria and Device Performance

The acceptance criteria for this device are implicitly derived from the performance goals demonstrated in the analytical and clinical studies, aiming for high sensitivity and specificity in detecting specified microorganisms and resistance markers. The reported performance is presented in several tables throughout the document.

Table 1: Acceptance Criteria (Implicit) and Reported Device Performance

Performance Metric CategorySpecific Metric (Implicit Acceptance Criteria)Reported Device PerformanceReference Table(s) in Document
Analytical Performance
Precision/ReproducibilityHigh site-to-site reproducibility across multiple operators and sites (e.g., >95% agreement)99.7% overall reproducibility (95% CI: 99.3% - 99.9%) across 3 external sites and 1 internal site.Table 1 (Site-to-Site Reproducibility)
High within-laboratory precision/repeatability (e.g., >95% agreement)100% within-laboratory precision/repeatability (95% CI: 97.9% - 100%)Table 2 (Within-Laboratory Precision/Repeatability)
High lot-to-lot reproducibility (e.g., >95% agreement)100% lot-to-lot reproducibility (95% CI: 92.1% - 100%)Table 3 (Lot-to-Lot Reproducibility)
Growth & Detection100% detection of target organisms at ring positivity and 8 hours post-ring positivity100% positivity for target positive samplesTable 5 (Growth & Detection Results)
Analytical ReactivityHigh inclusivity for diverse strains of target organisms (e.g., 100% detection)Generally 100% detection across tested strains (some minor exceptions noted for specific strains/targets, e.g. S. mutans, E. faecium FP)Table 6 (Microorganism Markers Inclusivity), Table 7 (Resistance Markers Inclusivity)
Analytical SpecificityLow cross-reactivity with off-panel organisms (e.g., no detection)97 of 103 off-panel species showed no cross-reactivity; 5 showed cross-reactivity with one reportable target; 1 showed positivity in 1/6 replicates.Table 9 (Cross Reactivity Summary)
InterferenceNo significant interference from co-infections, microbial interference, or interfering substances100% positivity for on-panel targets in presence of co-infections/off-panel microbes. 0% for negative samples. 100% target detection with all interfering substances.Table 12 (Competitive Inhibition), Table 13 (Microbial Interference), Table 14 (Interfering Substances)
Carry-Over/Cross ContaminationNo carry-over or cross-contamination100% agreement between expected and observed results, indicating no cross-contamination.Table 15 (Carry-Over/Cross Contamination)
Clinical Performance
Sensitivity (PPA)High PPA for all target organisms and resistance markersRanges from 88.9% (Streptococcus anginosus group prospective) to 100% for most targets (Combined Data).Table 20 (Clinical Performance - Combined Specimens), Table 22 (Contrived Data Set)
Specificity (NPA)High NPA for all target organisms and resistance markersRanges from 95.7% (mecA/mecC prospective) to 100% for most targets (Combined Data).Table 20 (Clinical Performance - Combined Specimens), Table 22 (Contrived Data Set)

2. Sample Sizes and Data Provenance

  • Test Set (Clinical Study):
    • Prospective Arm (Arm 1): 562 unique specimens enrolled, 509 included in analysis.
    • Pre-selected Arm (Arm 2): 163 pre-selected left-over specimens, 162 included in analysis.
    • Contrived Arm (Arm 3): 225 contrived specimens.
    • Total specimens analyzed for clinical performance: 509 (prospective) + 162 (pre-selected) + 225 (contrived) = 896 unique samples.
  • Data Provenance:
    • Prospective Arm: Collected between April 2024 and August 2024 from four geographically diverse clinical sites within the United States. Data is prospective.
    • Pre-selected Arm: Sourced from ten vendors in the United States and one site in Italy. Data are retrospective (left-over, de-identified specimens).
    • Contrived Arm: Specimens were prepared, blinded, randomized, and tested at three external testing sites and one internal testing site between June 2024 to September 2024. These are contrived (synthetic) samples.

3. Number of Experts and Qualifications for Ground Truth

  • Not Applicable in the traditional sense for this IVD device. The ground truth for this nucleic acid assay is established through reference methods such as culture with automated microbiological/biochemical identification (VITEK 2), PCR followed by bi-directional sequencing (BDS), or a combination thereof. This is a laboratory-based diagnostic test, not an imaging device requiring human expert interpretation for ground truth.

4. Adjudication Method for the Test Set

  • Not Applicable in the traditional sense. The "adjudication" for the gold standard (reference method) involves a hierarchical algorithm (Table 19). If initial VITEK 2 results were insufficient or for specific targets like Bacillus spp. and resistance markers (mecA/mecC, vanA, vanB), PCR followed by bi-directional sequencing was used. This is a technical validation process against established laboratory methods rather than a consensus among clinical experts interpreting an output.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Not done. This type of study (MRMC) is typically used for diagnostic imaging devices where human readers interpret images with and without AI assistance. This device is a molecular diagnostic assay that provides direct results, and therefore, an MRMC study is not relevant.

6. Standalone (i.e., algorithm only without human-in-the-loop performance)

  • Yes, in essence. The LIAISON PLEX® BCP Assay is an automated, sample-to-result system. Its performance (sensitivity, specificity) is evaluated independently against a reference method. While trained operators are involved in loading samples and running the system, the diagnostic result itself is generated by the "system" (instrument + assay + software algorithm) without human diagnostic interpretation of the raw signal data.

7. The Type of Ground Truth Used

  • Laboratory Reference Methods:
    • Culture followed by Automated microbiological/biochemical identification using VITEK 2 for most bacterial species.
    • PCR followed by bi-directional sequencing (BDS) for Bacillus spp. and resistance markers (mecA/mecC, vanA, vanB). This method was also used to confirm discordant results or for targets with low prevalence in initial testing.
    • In some cases, MALDI-ToF assay (Matrix-Assisted Laser Desorption/Ionization-Time Of Flight) was also mentioned as a Standard of Care method in the footnotes for discordant result explanations.

8. The Sample Size for the Training Set

  • Not explicitly stated in terms of a "training set" for an AI/ML model. This device is a direct detection assay based on hybridization, not a machine learning model that undergoes a training phase with a distinct dataset. Therefore, the concept of a "training set" in the context of data used to train an AI model is not applicable to this traditional IVD. The development process would involve extensive analytical characterization, probe design, and optimization using various microbial strains and clinical samples, but this is distinct from "training data" for an AI algorithm.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable for this type of device. As explained above, this is not an AI/ML device that requires a "training set" with ground truth established in the AI/ML context. The assay's design and optimization would rely on established microbiological methods and genetic sequencing to define target sequences and ensure specificity and inclusivity.

U.S. Food & Drug Administration 510(k) Clearance Letter

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

June 6, 2025

Luminex Corporation
Rocio Rueda
Senior Regulatory Affairs Associate
4088 Commercial Avenue
Northbrook, Illinois 60062

Re: K243490
Trade/Device Name: LIAISON PLEX Gram-Positive Blood Culture Assay
Regulation Number: 21 CFR 866.3365
Regulation Name: Multiplex Nucleic Acid Assay For Identification Of Microorganisms And Resistance Markers From Positive Blood Cultures
Regulatory Class: Class II
Product Code: PAM
Dated: November 7, 2024
Received: November 12, 2024

Dear Rocio Rueda:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K243490 - Rocio Rueda Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K243490 - Rocio Rueda Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Noel J. Gerald -S

Noel J. Gerald, Ph.D.
Deputy Division Director
Division of Microbiology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 2

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K243490

Device Name: LIAISON PLEX® Gram-Positive Blood Culture Assay

Indications for Use (Describe)

The LIAISON PLEX® Gram-Positive Blood Culture Assay (BCP), performed using the automated, sample-to-result LIAISON PLEX® System, is a qualitative multiplexed in vitro diagnostic test for the simultaneous detection and identification of selected gram-positive pathogens and/or selected genetic determinants associated with antimicrobial resistance in positive blood culture bottles. BCP is performed directly on blood culture media using blood culture bottles identified as positive by a continuous monitoring blood culture system and which contain gram-positive bacteria as determined by Gram stain.

The BCP Assay detects and identifies the following:

Gram Positive Resistance Markers(a):

  • mecA/mecC
  • vanA
  • vanB

Genera and Species:

  • Bacillus spp.
  • Enterococcus faecalis
  • Enterococcus faecium
  • Listeria spp.
  • Staphylococcus spp.
  • Staphylococcus aureus
  • Staphylococcus epidermidis
  • Staphylococcus lugdunensis
  • Streptococcus spp.
  • Streptococcus agalactiae
  • Streptococcus anginosus group
  • Streptococcus pneumoniae
  • Streptococcus pyogenes

(a) Negative results for antimicrobial resistance genes do not indicate bacterial susceptibility as there are multiple mechanisms that can contribute to resistance.

The LIAISON PLEX® BCP Assay contains targets for the detection of genetic determinants associated with resistance to methicillin (mecA/C) and vancomycin (vanA and vanB) to aid in the identification of potentially antimicrobial-resistant organisms in positive blood culture samples. In mixed growth, the LIAISON PLEX BCP Assay does not specifically attribute vanA/vanB-mediated vancomycin resistance to either E. faecalis or E. faecium, or mecA/mecC-mediated methicillin resistance to either Staphylococcus spp., S. aureus, S. epidermidis or S. lugdunensis.

The antimicrobial resistance gene or marker detected may or may not be associated with the agent responsible for disease. Negative results for these select antimicrobial resistance gene and marker assays do not indicate susceptibility, as multiple mechanisms of methicillin and vancomycin resistance exist.

The LIAISON PLEX® BCP Assay is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial bloodstream infections (BSI). The LIAISON PLEX® BCP Assay is not intended to monitor

Page 5

FORM FDA 3881 (8/23) Page 2 of 2

these infections. Sub-culturing of positive blood cultures is necessary to recover organisms for antimicrobial susceptibility testing (AST), for identification of organisms not detected by the LIAISON PLEX BCP Assay, to detect mixed infections that may not be detected by the LIAISON PLEX BCP Assay, for association of antimicrobial resistance genes to a specific organism, or for epidemiological typing.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 6

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 1 of 43

510(k) Summary

This Summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

Preparation date: 04 April 2025

A. 510(k) Number:

K243490

B. Purpose for Submission:

Traditional 510(k), New Device

C. Measurand:

Nucleic acid sequences for the following organisms: Bacillus spp., Enterococcus faecalis, Enterococcus faecium, Listeria spp., Staphylococcus spp., Staphylococcus aureus, Staphylococcus epidermidis, Staphylococcus lugdunensis, Streptococcus spp., Streptococcus agalactiae, Streptococcus anginosus group, Streptococcus pneumoniae, Streptococcus pyogenes

Nucleic acid sequences for the following resistance markers: mecA/mecC, vanA, vanB

D. Type of Test:

Qualitative Multiplexed Direct Detection Hybridization Assay

E. Applicant:

Rocio Rueda, Luminex Corporation
4088 Commercial Avenue
Northbrook, IL 60062
(847) 400-9000

F. Proprietary and Established Names:

LIAISON PLEX® Gram-Positive Blood Culture Assay

G. Regulatory Information:

Product CodeClassificationRegulation SectionPanel
PAMII21 CFR 866.3365 – Multiplex Nucleic Acid Assay for Identification of Microorganisms and Resistance Markers from Positive Blood Cultures83 (Microbiology)

Page 7

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 2 of 43

H. Intended Use:

1. Intended use(s):

The LIAISON PLEX® Gram-Positive Blood Culture (BCP) Assay, performed using the automated, sample-to-result LIAISON PLEX® System, is a qualitative multiplexed in vitro diagnostic test for the simultaneous detection and identification of selected gram-positive pathogens and/or selected genetic determinants associated with antimicrobial resistance in positive blood culture bottles. The LIAISON PLEX BCP Assay is performed directly on blood culture media using blood culture bottles identified as positive by a continuous monitoring blood culture system and which contain gram-positive bacteria as determined by Gram stain.

The LIAISON PLEX BCP Assay detects and identifies the following:

Gram-Positive Resistance Markers^aGenera and Species
mecA/mecCBacillus spp.
vanAEnterococcus faecalis
vanBEnterococcus faecium
Listeria spp.
Staphylococcus spp.
Staphylococcus aureus
Staphylococcus epidermidis
Staphylococcus lugdunensis
Streptococcus spp.
Streptococcus agalactiae
Streptococcus anginosus group
Streptococcus pneumoniae
Streptococcus pyogenes

^a Negative results for antimicrobial resistance genes do not indicate bacterial susceptibility as there are multiple mechanisms that can contribute to resistance.

The LIAISON PLEX® BCP contains targets for the detection of genetic determinants associated with resistance to methicillin (mecA/mecC) and vancomycin (vanA and vanB) to aid in the identification of potentially antimicrobial-resistant organisms in positive blood culture samples. In mixed growth, the LIAISON PLEX® BCP Assay does not specifically attribute vanA/vanB-mediated vancomycin resistance to either E. faecalis or E. faecium, or mecA/mecC-mediated methicillin resistance to either Staphylococcus spp., S. aureus, S. epidermidis or S. lugdunensis.

The antimicrobial resistance gene or marker detected may or may not be associated with the agent

Page 8

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 3 of 43

responsible for disease. Negative results for these select antimicrobial resistance gene and marker assays do not indicate susceptibility, as multiple mechanisms of methicillin and vancomycin resistance exist.

The LIAISON PLEX BCP Assay is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial bloodstream infections (BSI). The LIAISON PLEX BCP Assay is not intended to monitor treatment of these infections. Sub-culturing of positive blood cultures is necessary to recover organisms for antimicrobial susceptibility testing (AST), for identification of organisms not detected by LIAISON PLEX BCP Assay, to detect mixed infections that may not be detected by the LIAISON PLEX BCP Assay, for association of antimicrobial resistance genes to a specific organism, or for epidemiological typing.

Page 9

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 4 of 43

2. Indication(s) for use:

Same as intended use.

3. Special conditions for use statement(s):

For prescription use only.
For in vitro diagnostic use only.

4. Special instrument requirements:

For use with LIAISON PLEX® Systems.

I. Device Description:

The LIAISON PLEX® Gram-Positive Blood Culture Assay (BCP Assay) is an automated test for the detection and identification of nucleic acid from gram-positive bacteria in a positive blood culture media sample. The BCP Assay is performed directly on blood culture media using blood culture bottles identified as positive by a continuous monitoring blood culture system, and which contain gram-positive bacteria, as determined by a Gram stain.

The LIAISON PLEX® System is a fully automated, bench-top "sample-to-answer" device that performs sample preparation and microarray-based hybridization for the detection of target-specific nucleic acids. The test reagents are supplied as a single, disposable test cartridge. Target amplification is not performed as part of the BCP Assay workflow, as it is a non-amplified, direct detection test performed on the LIAISON PLEX® System.

J. Substantial Equivalence Information:

1. Predicate device name(s):

VERIGENE Blood Culture Gram-Positive (BC-GP) Nucleic Acid Test

2. Predicate 510(k) number(s):

K122514

3. Comparison with predicate:

The following tables compare the LIAISON PLEX® Gram-Positive Blood Culture Assay to the VERIGENE Blood Culture Gram-Positive (BC-GP) Nucleic Acid Test.

Comparison to Predicate DevicePredicate Device: VERIGENE Blood Culture Gram-Positive (BC-GP) Nucleic Acid Test, K122514Candidate Device: LIAISON PLEX® Gram-Positive Blood Culture Assay
Product CodePAMPAM
Regulation Number21 CFR 866.336521 CFR 866.3365

Page 10

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 5 of 43

Comparison to Predicate DevicePredicate Device: VERIGENE Blood Culture Gram-Positive (BC-GP) Nucleic Acid Test, K122514Candidate Device: LIAISON PLEX® Gram-Positive Blood Culture Assay
Organism DetectedOrganisms: Staphylococcus spp., Staphylococcus aureus, Staphylococcus epidermidis, Staphylococcus lugdunensis, Streptococcus spp., Streptococcus pneumoniae, Streptococcus pyogenes, Streptococcus agalactiae, Streptococcus anginosus group, Enterococcus faecalis, Enterococcus faecium, Listeria spp.Resistance Markers: mecA, vanA, vanBOrganisms: Bacillus spp., Enterococcus faecalis, Enterococcus faecium, Listeria spp., Staphylococcus spp., Staphylococcus aureus, Staphylococcus epidermidis, Staphylococcus lugdunensis, Streptococcus spp., Streptococcus agalactiae, Streptococcus anginosus group, Streptococcus pneumoniae, Streptococcus pyogenesResistant markers: mecA/mecC, vanA, vanB
MeasurandNucleic acid from Organisms detectedSame
Intended UseThe Verigene® Gram-Positive Blood Culture Nucleic Acid Test (BC-GP), performed using the sample-to-result Verigene System, is a qualitative multiplexed in vitro diagnostic test for the simultaneous detection and identification of potentially pathogenic gram-positive bacteria which may cause bloodstream infection (BSI). BC-GP is performed directly on blood culture bottles identified as positive by a continuous monitoring blood culture system and which contain gram-positive bacteria.BC-GP detects and identifies the following:Bacterial Genera and SpeciesStaphylococcus spp.Staphylococcus aureusStaphylococcus epidermidisStaphylococcus lugdunensisStreptococcus spp.Streptococcus pneumoniaeStreptococcus pyogenesStreptococcus agalactiaeStreptococcus anginosus groupEnterococcus faecalisEnterococcus faeciumListeria spp.Resistance Markers¹mecAvanAvanB¹In mixed growth, BC-GPThe LIAISON PLEX® Gram-Positive Blood Culture Assay (BCP), performed using the automated, sample-to-result LIAISON PLEX® System, is a qualitative multiplexed in vitro diagnostic test for the simultaneous detection and identification of selected gram-positive pathogens and/or selected genetic determinants associated with antimicrobial resistance in positive blood culture bottles. BCP is performed directly on blood culture media using blood culture bottles identified as positive by a continuous monitoring blood culture system and which contain gram-positive bacteria as determined by Gram stain.The BCP Assay detects and identifies the following:Genera and SpeciesBacillus spp.Enterococcus faecalisEnterococcus faeciumListeria spp.Staphylococcus spp.Staphylococcus aureusStaphylococcus epidermidisStaphylococcus lugdunensisStreptococcus spp.Streptococcus agalactiaeStreptococcus anginosus groupStreptococcus pneumoniaeStreptococcus pyogenes

Page 11

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 6 of 43

Comparison to Predicate DevicePredicate Device: VERIGENE Blood Culture Gram-Positive (BC-GP) Nucleic Acid Test, K122514Candidate Device: LIAISON PLEX® Gram-Positive Blood Culture Assay
does not specifically attribute van-mediated vancomycin resistance to either E. faecalis or E. faecium, or mecA-mediated methicillin resistance to either S. aureus or S. epidermidis.BC-GP is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial bloodstream infections; however, it is not used to monitor these infections. Sub-culturing of positive blood cultures is necessary to recover organisms for antimicrobial susceptibility testing (AST), for identification of organisms not detected by BC-GP, differentiation of mixed growth, for association of antimicrobial resistance marker genes to a specific organism, or for epidemiological typing.Resistance Markers^amecA/mecCvanAvanB^a Negative results for antimicrobial resistance genes do not indicate bacterial susceptibility as there are multiple mechanisms that can contribute to resistance.The LIAISON PLEX BCP Assay contains targets for the detection of genetic determinants associated with resistance to methicillin (mecA/mecC) and vancomycin (vanA and vanB) to aid in the identification of potentially antimicrobial-resistant organisms in positive blood culture samples. In mixed growth, BCP does not specifically attribute vanA/vanB-mediated vancomycin resistance to either E. faecalis or E. faecium, or mecA/mecC-mediated methicillin resistance to either Staphylococcus spp., S. aureus, S. epidermidis or S. lugdunensis.The antimicrobial resistance gene or marker detected may or may not be associated with the agent responsible for disease. Negative results for these select antimicrobial resistance gene and marker assays do not indicate susceptibility, as multiple mechanisms of methicillin and vancomycin resistance exist.The LIAISON PLEX® BCP Assay is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial bloodstream infections (BSI). The LIAISON PLEX® BCP Assay is not intended to monitor these infections. Sub-culturing of positive blood cultures is necessary to recover organisms for antimicrobial susceptibility testing (AST), for identification of organisms not detected by LIAISON PLEX BCP Assay, to detect mixed infections that may not be detected by the LIAISON PLEX BCP Assay, for association of antimicrobial resistance genes to a specific organism, or for epidemiological typing.

Page 12

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 7 of 43

Comparison to Predicate DevicePredicate Device: VERIGENE Blood Culture Gram-Positive (BC-GP) Nucleic Acid Test, K122514Candidate Device: LIAISON PLEX® Gram-Positive Blood Culture Assay
does not specifically attribute van-mediated vancomycin resistance to either E. faecalis or E. faecium, or mecA-mediated methicillin resistance to either S. aureus or S. epidermidis.BC-GP is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial bloodstream infections; however, it is not used to monitor these infections. Sub-culturing of positive blood cultures is necessary to recover organisms for antimicrobial susceptibility testing (AST), for identification of organisms not detected by BC-GP, differentiation of mixed growth, for association of antimicrobial resistance marker genes to a specific organism, or for epidemiological typing.Resistance Markers^amecA/mecCvanAvanB^a Negative results for antimicrobial resistance genes do not indicate bacterial susceptibility as there are multiple mechanisms that can contribute to resistance.The LIAISON PLEX BCP Assay contains targets for the detection of genetic determinants associated with resistance to methicillin (mecA/mecC) and vancomycin (vanA and vanB) to aid in the identification of potentially antimicrobial-resistant organisms in positive blood culture samples. In mixed growth, BCP does not specifically attribute vanA/vanB-mediated vancomycin resistance to either E. faecalis or E. faecium, or mecA/mecC-mediated methicillin resistance to either Staphylococcus spp., S. aureus, S. epidermidis or S. lugdunensis.The antimicrobial resistance gene or marker detected may or may not be associated with the agent responsible for disease. Negative results for these select antimicrobial resistance gene and marker assays do not indicate susceptibility, as multiple mechanisms of methicillin and vancomycin resistance exist.The LIAISON PLEX® BCP Assay is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial bloodstream infections (BSI). The LIAISON PLEX® BCP Assay is not intended to monitor these infections. Sub-culturing of positive blood cultures is necessary to recover organisms for antimicrobial susceptibility testing (AST), for identification of organisms not detected by LIAISON PLEX BCP Assay, to detect mixed infections that may not be detected by the LIAISON PLEX BCP Assay, for association of antimicrobial resistance genes to a specific organism, or for epidemiological typing.
Automated System (Sample to Answer)AutomatedSame
InstrumentationVERIGENELIAISON PLEX®
Sample TypesPositive Blood CultureSame

Page 13

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 8 of 43

K. Standards/Guidance Documents Referenced:

  • Class II Special Controls Guideline: Multiplex Nucleic Acid Assay for Identification of Microorganisms and Resistance Markers from Positive Blood Cultures (May 2015)
  • Electronic Submission Template for Medical Device 510(k) Submissions - Guidance for Industry and Food and Drug Administration Staff (October 2, 2023).
  • Content of Premarket Submissions for Device Software Functions - Guidance for Industry and Food and Drug Administration Staff (June 14, 2023).
  • Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions - Guidance for Industry and Food and Drug Administration Staff (September 23, 2023).
  • Statistical Guidance on Reporting Results from Studies Evaluating Diagnostic Tests - Guidance for Industry and FDA Staff (March 13, 2007).
  • CLSI. User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline - Second Edition. CLSI document EP12-A2. Wayne, PA: Clinical and Laboratory Standards Institute; 2008.
  • CLSI. Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline. CLSI document EP25-A. Wayne, PA: Clinical and Laboratory Standards Institute; 2009.
  • ISO 14971:2019 Medical devices - Application of risk management to medical devices
  • IEC 62366-1:2015 +A1:2020 Medical devices- Part 1: Application of usability engineering to medical devices
  • ISO 62304:2006 Medical device software - Software life-cycle processes
  • ISO 15223-1:2021: Medical Devices- Symbols to be used with medical device labels, labeling and information to be supplied - Part 1: General requirements
  • IEC 61010-1 Ed. 3.1 2017-01: Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements
  • EN 61010-2-101:2002/IEC 61010-2-101:2015: Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment.
  • IEC 60601-1-2:2014 (Edition 4.0): Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
  • ISO 13485:2016/EN ISO13485:2016; Medical devices-Quality Management System - Requirements for regulatory purposes
  • ISO 20916:2019; In vitro diagnostic medical devices. Clinical performance studies using specimens from human subjects. Good study practice
  • EN ISO18113-1:2011; In vitro diagnostic medical devices- Information supplied by the manufacturer (labeling). Terms, definition and general requirements

Page 14

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 9 of 43

  • EN ISO 18113-2:2011; In vitro diagnostic medical devices- Information supplied by the manufacturer (labeling) – Part 2: In vitro diagnostic reagents for professional use
  • EN ISO 18113-3:2011; In vitro diagnostic medical devices- Information supplied by the manufacturer (labeling) – Part 3: In vitro diagnostic instruments for professional use
  • EN ISO 23640:2015; In vitro diagnostic medical devices- Evaluation of stability of in vitro diagnostic reagents
  • IEC 61326-1:2012; Electrical equipment for measurement control and laboratory use - EMC requirements - Part 1: General requirements
  • EN 61326-2-6:2006/IEC 61326-2-6:2012; Electrical equipment for measurement control and laboratory use - EMC requirements - Part 2-6: Particular requirements - In vitro diagnostic (IVD) medical equipment

L. Test Principle:

The LIAISON PLEX® Gram-Positive Blood Culture Assay (BCP Assay) is performed directly on blood culture media using blood culture bottles identified as positive by a continuous monitoring blood culture system, and which contain gram-positive bacteria, as determined by a Gram stain.

The system consists of an instrument and a single-use, disposable test cartridge. The user loads an aliquot of the sample into the sample port of the LIAISON PLEX® Gram-Positive Blood Culture Assay Cartridge. Next, the user sets up the sample order on the LIAISON PLEX® System by first entering the sample information or scanning the barcode ID located on the sample tube, then scanning the barcode ID located on the test cartridge. Last, the user inserts the test cartridge into the processing module to initiate the test. The LIAISON PLEX® System identifies the assay being run and automatically initiates the proper testing protocol to process the sample, analyze the data, and generate test results.

The LIAISON PLEX® System automates the BCP Assay sample analysis through the following steps:

a) Sample Preparation: Nucleic acid extraction via mechanical and chemical cell lysis and magnetic bead-based nucleic acid isolation; PCR amplification is not used in this assay. By performing direct detection, the assay specifically detects the viable bacteria and minimizes interference from trace nucleic acids or non-viable bacteria that may be present at much lower levels and potentially lead to false positive results.

b) Hybridization: Extracted nucleic acid hybridize to target-specific capture DNA on a microarray format, and target-specific mediator and gold nanoparticle probe hybridize to captured nucleic acids;

c) Signal Analysis: Gold nanoparticle probes bound specifically to target-containing spots in the microarray are silver-enhanced, and light scatter from the spots is measured and further analyzed to determine the presence (Detected) or absence (Not Detected) of a target.

Page 15

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 10 of 43

M. Performance Characteristics:

1. Analytical performance:

a. Precision/Reproducibility:

Site-to-site Reproducibility

Site-to-site reproducibility of the LIAISON PLEX® BCP Assay was evaluated by testing LIAISON PLEX® BCP Assay cartridges across five non-consecutive days with at least two operators each at multiple sites; three external sites and one internal site. The reproducibility panel, blinded to operators, consisted of a total of 16 blood culture samples. 7 samples each contained representative on-panel organisms individually cultured at ring positivity, and eight hours after ring positivity, one sample was contrived with an off-panel organism (Escherichia coli), and the last sample was a negative blood culture matrix (NBM) specimen. The blinded reproducibility panel was tested in triplicate by each operator on each testing day for five non-consecutive testing days. The call agreement and 95% confidence intervals are presented in Table 1. Overall, results of site-to-site reproducibility evaluation demonstrated 99.7% reproducibility of the LIAISON PLEX® BCP Assay with a 95% CI of 99.3% to 99.9%.

Table 1. LIAISON PLEX® Gram-Positive Blood Culture Assay Site-to-Site Reproducibility Results Summary

Organism IDReportable TargetsSample TypeAgreement (%) with Expected results95% CI
Site1Site 2
Staphylococcus aureusStaphylococcus spp. S. aureusPanel 1 – RP100% (30/30)100% (30/30)
Panel 1 – RP+8 Hours100% (30/30)96.7% (29/30)^a
Enterococcus faecalisEnterococcus faecalisPanel 2 – RP100% (30/30)100% (30/30)
Panel 2 – RP+8 Hours100% (30/30)100% (30/30)
Bacillus subtilisBacillus spp.Panel 3 – RP100% (30/30)100% (30/30)
Panel 3 – RP+8 Hours100% (30/30)100% (30/30)
Listeria monocytogenesListeria spp.Panel 4 – RP100% (30/30)100% (30/30)
Panel 4 – RP+8 Hours100% (30/30)100% (30/30)
Staphylococcus epidermidisStaphylococcus spp. Staphylococcus epidermidisPanel 5 – RP100% (30/30)100% (30/30)
Panel 5 – RP+8 Hours100% (30/30)100% (30/30)
Streptococcus pneumoniaeStreptococcus spp. Streptococcus pneumoniaePanel 6 – RP100% (30/30)100% (30/30)
Panel 6 – RP+8 Hours100% (30/30)100% (30/30)
Streptococcus pyogenesStreptococcus spp. Streptococcus pyogenesPanel 7 – RP100% (30/30)100% (30/30)

Page 16

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 11 of 43

Organism IDReportable TargetsSample TypeAgreement (%) with Expected results95% CI
Site1Site 2
Panel 7 – RP+8 Hours100% (30/30)100% (30/30)
Escherichia coliNoneContrived Negative100% (60/60)100% (60/60)
No TargetNoneNegative Blood Matrix100% (30/30)100% (30/30)
Resistance markers
Staphylococcus aureusmecA/CPanel 1 – RP100% (30/30)100% (30/30)
Panel 1 – RP+8 Hours100% (30/30)96.7% (29/30)
Enterococcus faecalisvanBPanel 2 – RP100% (30/30)100% (30/30)
Panel 2 – RP+8 Hours100% (30/30)100% (30/30)
Staphylococcus epidermidismecA/mecCPanel 5 – RP100% (30/30)100% (30/30)
Panel 5 – RP+8 Hours100% (30/30)100% (30/30)

^a False positive S. lugdunensis, S. anginosus group, and Streptococcus spp. result in 1/30 replicates. Reportable targets were fully detected.
^b False positive E. faecium and vanA result in 1/30 replicates.
^c False positive S. lugdunensis result in 1/30 replicates. Reportable targets were fully detected.
^d False positive Staphylococcus spp. result in 1/30 replicates. Reportable targets were fully detected.

Precision/Repeatability

Within laboratory (operator-to-operator) precision/repeatability of the LIAISON PLEX® BCP Assay was evaluated based on the results generated by two operators testing samples (two on-panel organisms (Staphylococcus aureus and Enterococcus faecalis) at ring positivity and at 8 hours after ring positivity, contrived negative and negative blood culture samples) at the internal site (Site 1). Within laboratory precision/repeatability of the LIAISON PLEX® BCP Assay was 100%, and results are summarized in Table 2.

Table 2. LIAISON PLEX® Gram-Positive Blood Culture Assay Within-Laboratory Precision/Repeatability Results Summary

Organism IDReportable TargetsSample TypeAgreement (%)Overall 95% CI
Staphylococcus aureusStaphylococcus spp | S. aureusPanel 1 – RP100% (30/30)88.6% - 100%
Panel 1 – RP+8 Hours100% (30/30)88.6% - 100%
Enterococcus faecalisEnterococcus faecalisPanel 2 – RP100% (30/30)88.6% - 100%
Panel 2 – RP+8 Hours100% (30/30)88.6% - 100%
Escherichia coliNoneContrived Negative100% (30/30)88.6% - 100%
No TargetNoneNegative Blood Matrix100% (30/30)88.6% - 100%
Resistance Markers
Staphylococcus aureusmecA/CPanel 1 – RP100% (30/30)88.6% - 100%

Page 17

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 12 of 43

Organism IDReportable TargetsSample TypeAgreement (%)Overall 95% CI
Panel 1 – RP+8 Hours100% (30/30)88.6% - 100%
Enterococcus faecalisvanBPanel 2 – RP100% (30/30)88.6% - 100%
Panel 2 – RP+8 Hours100% (30/30)88.6% - 100%
Overall Agreement (All Targets/Sample Types)100% (180/180)97.9% - 100%

Page 18

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 13 of 43

Lot-to-lot Reproducibility

Lot-to-lot reproducibility of the LIAISON PLEX® BCP Assay was evaluated by one operator over a minimum of five non-consecutive days across three lots of LIAISON PLEX® BCP Assay cartridges and four LIAISON PLEX® systems. The same reproducibility panel of six blood culture samples; two representative on-panel organisms (Staphylococcus aureus and Enterococcus faecalis) individually cultured at ring positivity, and eight hours after ring positivity, one negative sample contrived with an off-panel organism (Escherichia coli), and one negative blood culture matrix (NBM) sample, were prepared and tested for the evaluation of the lot-to-lot reproducibility. Each sample type was tested in triplicate per cartridge lot on a given testing day. Lot-to-lot reproducibility of the LIAISON PLEX® BCP Assay cartridges was 100%, and results are summarized in Table 3.

Table 3: LIAISON PLEX® Gram-Positive Blood Culture (BCP) Assay Lot-to-Lot Reproducibility Results

Organism IDReportable TargetsSample TypeReagent LotResults (%) agreementResults (95% CI)
Staphylococcus aureusStaphylococcus spp. S. aureusPanel 1 – RPLot 1100% (15/15)
Lot 2100% (15/15)79.6% - 100%
Lot 3100% (15/15)
Overall100% (45/45)92.1% – 100%
Panel 1 – RP+8 HoursLot 1100% (15/15)
Lot 2100% (15/15)79.6% - 100%
Lot 3100% (15/15)
Overall100% (45/45)92.1% – 100%
Enterococcus faecalisEnterococcus faecalisPanel 2 – RPLot 1100% (15/15)79.6% - 100%
Lot 2100% (15/15)
Lot 3100% (15/15)
Overall100% (45/45)92.1% – 100%
Panel 2 – RP+8 HoursLot 1100% (15/15)79.6% - 100%
Lot 2100% (15/15)
Lot 3100% (15/15)
Overall100% (45/45)92.1% – 100%
Escherichia coliNoneContrived NegativeLot 1100% (15/15)79.6% - 100%
Lot 2100% (15/15)
Lot 3100% (15/15)
Overall100% (45/45)92.1% – 100%
No TargetNoneNegative Blood MatrixLot 1100% (15/15)79.6% - 100%
Lot 2100% (15/15)
Lot 3100% (15/15)
Overall100% (45/45)92.1% – 100%
Resistance Marker
Staphylococcus aureusmecA/CPanel 1 – RPLot 1100% (15/15)
Lot 2100% (15/15)79.6% - 100%
Lot 3100% (15/15)
Overall100% (45/45)92.1% – 100%
Panel 1 – RP+8 HoursLot 1100% (15/15)
Lot 2100% (15/15)79.6% - 100%
Lot 3100% (15/15)

Page 19

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 14 of 43

Organism IDReportable TargetsSample TypeReagent LotResults (%) agreementResults (95% CI)
Overall100% (45/45)92.1% – 100%
Enterococcus faecalisvanBPanel 2 – RPLot 1100% (15/15)79.6% - 100%
Lot 2100% (15/15)
Lot 3100% (15/15)
Overall100% (45/45)92.1% – 100%
Panel 2 – RP+8 HoursLot 1100% (15/15)79.6% - 100%
Lot 2100% (15/15)
Lot 3100% (15/15)
Overall100% (45/45)92.1% – 100%

b. Linearity/assay reportable range:

Not applicable. The LIAISON® PLEX Gram-Positive Blood Culture Assay is a qualitative assay.

Page 20

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 15 of 43

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Controls

Several controls are built into the assay and system to ensure identification of processing errors and to establish validity of test results.

Internal Controls

Each LIAISON PLEX® Gram-Positive Blood Culture Assay cartridge includes internal controls to ensure performance of sample preparation and detection. The internal extraction control is present in the lysis tube when the sample is added. Sample preparation is initiated and the extraction control assesses extraction, nucleic acid recovery, and detection. Finally, addition of a post-extraction hybridization control serves as an indicator of successful hybridization. Internal control results are reported as Pass or Fail on the printed reports (see Table 4 for detailed explanations of each control result). Internal controls must generate a signal above threshold in each internal reaction for the system to report a valid test result.

Table 4. Interpretation of Controls on the LIAISON PLEX® Gram-Positive Blood Culture Assay Report

Internal Control ResultExplanationSuggested Action
PassTest was completed and internal controls were successful, indicating that valid results were generated.Review and report results
FailOne or more internal control failed.Repeat test with a new cartridge
External Controls

Positive and negative external controls should be tested with each new lot or shipment of reagents, or monthly, (whichever occurs first), or in accordance with updated local, regional, state, and/or federal guidelines. Verified negative blood matrix can be used as the negative control. Previously characterized positive samples or verified negative blood matrix spiked with well characterized organisms may be used as the external positive control. External controls should be used in accordance with laboratory protocols and in accordance with local, state, and federal accrediting organizations, as applicable.

Stability
Specimen Stability

Performance of the LIAISON PLEX® Gram-Positive Blood Culture (BCP) Assay was assessed using specimens tested in a fresh state (at bottle/ring positive and at bottle/ring positive + 8 hours) and after exposure to various storage conditions. Conditions tested included refrigerated storage (2° to 8°C) and room temperature storage (15°C to 30°C) to span the typical "fresh specimen" storage conditions across multiple time points. Positive specimens containing five target organisms representing a total of 9 reportable gram-positive targets

Page 21

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 16 of 43

and select resistance markers were tested as well as negative blood matrix control specimen containing no target organisms. 100% detection of all reportable targets was observed across all timepoints. The negative sample demonstrated 0% positivity across all time-points and storage conditions tested. The results demonstrated that specimens may be stored under the following temperature conditions without impacting the performance of the LIAISON PLEX® BCP cartridge:

  • Up to 72 hours (3 days) at Refrigerated (2-8°C) or Room Temperature (15-30°C) storage conditions.
Device Stability

A shelf-life study was conducted to evaluate the real-time stability of the LIAISON PLEX® BCP Assay at the recommended storage conditions of room temperature (15°C – 30°C). Real-time stability was assessed using three Positive Control panels which interrogate representative targets in the assay, and one Negative Control which consisted of negative blood matrix. Results of real-time stability for all tested lots demonstrated the LIAISON PLEX® BCP Assay is stable for at least 1 month when stored at 15°C – 30°C, and up to 3 months when stored at 15°C – 30°C for two of the three cartridge lots evaluated. Testing for the third lot is ongoing. Shelf-life will be extended based upon results of on-going stability testing.

An open box stability study was performed to evaluate the stability of the LIAISON PLEX® BCP Assay cartridges at room temperature once removed from their foil pouch. Testing was performed shortly after kits were manufactured and will be repeated at the end of the product shelf-life. Non-aged cartridges were tested at 0 hours (T0), 3 hours (T1), and 9 hours (T2) after removal from their pouch. Each time point included testing of three Positive Control panels which interrogate seven reportable targets in the assay, and one Negative Control which consisted of negative blood matrix. One unexpected result (false positive) for Staphylococcus lugdunensis and Streptococcus spp. occurred for a negative target. Expected results were observed at all other timepoints and conditions. Results of open-box stability indicate the cartridges are stable for up to nine hours after cartridges are removed from their foil pouches and stored at room temperature.

Fresh vs. Frozen Specimen Stability

A fresh vs. frozen specimen stability study was performed to evaluate the performance of the LIAISON PLEX® Gram-Positive Blood Culture (BCP) assay in specimens that have been prepared "fresh" (at two growth durations referred to as Bottle/Ring Positive and Bottle/Ring Positive + 8 hours) and subjected to a defined number of freeze/thaw (F/T) cycles as well as those experiencing a prolonged storage in frozen conditions. The study was performed using six representative organisms detected by the LIAISON PLEX® BCP Assay cultured individually in blood culture bottles. Performance testing using negative blood matrix served as a control test during the study. Material was tested under 5 different conditions – Material was tested at multiple freeze thaw cycles: 1st Freeze-thaw, 2nd Freeze-thaw, and 3rd Freeze-thaw, and was frozen for a minimum of 8 hours in between each freeze-thaw cycle. In the prolonged testing, material was frozen at time points of 15 days and 30 days. A total of 701 replicates were included in this study. The results

Page 22

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 17 of 43

demonstrated 100% positivity for target positive samples at both growth durations and all freeze-thaw conditions. The negative blood matrix demonstrated 0% positivity of all reportable targets for the assay at all freeze-thaw conditions.

d. Growth and Detection Study

The growth and detection study was performed to evaluate detection of each organism at the time of culture positivity and 8 hours after culture positivity for the organisms listed in Table 5. Three bottles for each organism were grown to ring positive or 8 hours after ring positive and three replicates were tested for each bottle, for a total of 9 replicates tested per organism. The results, presented in Table 5, demonstrated 100% positivity for target positive samples and 0% positivity when tested with negative blood.

Table 5. LIAISON PLEX® Gram-Positive Blood Culture Assay Growth & Detection Results Summary

Organism TestedExpected TargetsRing Positivity8 Hours After Ring Positivity
Per Bottle (CFU/mL)Positive Agreement/ Total (% Detected)
Bacillus subtilis ATCC 19659Bacillus spp.1.29E+089/9 (100%)
6.10E+08
3.60E+08
Enterococcus faecium ATCC 700221Enterococcus faecium | vanA2.37E+089/9 (100%)
3.01E+08
2.24E+08
Enterococcus faecalis ATCC 51575Enterococcus faecalis | vanB6.87E+089/9 (100%)
9.27E+08
9.83E+08
Listeria monocytogenes ATCC 15313Listeria spp.7.57E+089/9 (100%)
9.63E+08
7.73E+08
Staphylococcus aureus ATCC BAA-2312Staphylococcus spp.| Staphylococcus aureus | mecA/mecC1.52E+079/9 (100%)
2.34E+07
1.88E+07
Staphylococcus epidermidis ATCC 35984Staphylococcus spp. |Staphylococcus epidermidis | mecA/mecC1.55E+089/9 (100%)
4.40E+08
2.84E+08
Staphylococcus lugdunensis ATCC 49576Staphylococcus spp.| Staphylococcus lugdunensis5.67E+089/9 (100%)
7.43E+08
8.90E+08
Streptococcus agalactiae ATCC 12386Streptococcus spp.| Streptococcus1.21E+099/9 (100%)
1.32E+09

Page 23

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 18 of 43

Organism TestedExpected TargetsRing Positivity8 Hours After Ring Positivity
Per Bottle (CFU/mL)Positive Agreement/ Total (% Detected)
agalactiae1.12E+09
Streptococcus anginosus ATCC 33397Streptococcus spp.| Streptococcus anginosus group7.57E+089/9 (100%)
4.97E+08
3.63E+08
Streptococcus pneumoniae ATCC 49619Streptococcus spp.| Streptococcus pneumoniae8.30E+089/9 (100%)
8.40E+08
9.80E+08
Streptococcus pyogenes ATCC 700294Streptococcus spp.| Streptococcus pyogenes3.43E+089/9 (100%)
3.40E+08
4.70E+08
Organism TestedExpected TargetsRing Negative
Negative BloodNone0/3 (0%)

^a False Positive Bacillus spp. result in one replicate. Three additional replicates were tested and no additional FPs were observed.

e. Analytical Reactivity (Inclusivity)

The analytical reactivity (inclusivity) of the LIAISON PLEX® BCP Assay was evaluated by using a collection of 184 isolates, representing the genetic diversity of the analytes included as part of the LIAISON PLEX® BCP Assay. Table 6 contains the summary of the organisms tested and the genus and species detected by the LIAISON PLEX® BCP Assay. Table 7 contains the summary for the antimicrobial resistance marker reportable targets also detected by the assay.

Table 6: LIAISON PLEX® Gram-Positive Blood Culture (BCP) Inclusivity Summary (Microorganism Markers)

Reportable Target (Genus)Reportable Target (Species)Organism# of strains% Detected
Bacillus spp.N/ABacillus cereus4100%
Bacillus licheniformis2100%
Bacillus subtilis3100%
Bacillus thuringiensis2100%
N/AEnterococcus faecalisEnterococcus faecalis9100%
Enterococcus faeciumEnterococcus faecium9100%^a
Listeria spp.N/AListeria grayi2100%
Listeria innocua2100%
Listeria ivanovii2100%
Listeria monocytogenes6100%
Listeria seeligeri2100%

Page 24

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 19 of 43

Reportable Target (Genus)Reportable Target (Species)Organism# of strains% Detected
Listeria welshimeri2100%
Staphylococcus spp.S. aureusStaphylococcus aureus43100%
S. epidermidisStaphylococcus epidermidis8100%
S. lugdunensisStaphylococcus lugdunensis5100%
N/AStaphylococcus argenteus2100%^b
Staphylococcus auricularis2100%
Staphylococcus capitis2100%
Staphylococcus caprae1100%
Staphylococcus cohnii2100%
Staphylococcus haemolyticus2100%
Staphylococcus hominis3100%
Staphylococcus intermedius2100%
Staphylococcus muscae1100%^c
Staphylococcus pasteuri1100%
Staphylococcus saccharolyticus3100%
Staphylococcus saprophyticus2100%
Staphylococcus schleiferi1100%
Mammaliicoccus (Staphylococcus) sciuri2100%
Staphylococcus simulans2100%
Staphylococcus warneri2100%
Staphylococcus xylosus1100%
Streptococcus spp.S. agalactiaeStreptococcus agalactiae5100%
S. anginosus GroupStreptococcus anginosus2100%
Streptococcus constellatus3100%
Streptococcus intermedius2100%
S. pneumoniaeStreptococcus pneumoniae5100%
S. pyogenesStreptococcus pyogenes5100%
N/AStreptococcus bovis3100%
Streptococcus dysgalactiae2100%
Streptococcus equi2100%
Streptococcus equinus2100%
Streptococcus gallolyticus3100%
Streptococcus gordonii2100%
Streptococcus infantarius subsp. coli1100%
Streptococcus infantis2100%
Streptococcus mitis2100%
Streptococcus mutans288.9%^d
Streptococcus oralis2100%
Streptococcus parasanguinis2100%

Page 25

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 20 of 43

Reportable Target (Genus)Reportable Target (Species)Organism# of strains% Detected
Streptococcus peroris1100%
Streptococcus pseudopneumoniae1100%
Streptococcus salivarius1100%
Streptococcus sanguinis2100%

^a One replicate from one strain resulted in a false positive (FP) S. lugdunensis result, three additional replicates were run for that strain. No additional FPs were observed.
^b Staphylococcus argenteus cross reacts with S. aureus.
^c Staphylococcus muscae cross reacts with Listeria spp.
^d One strain of Streptococcus mutans, ATCC 25175, was not detected as Streptococcus spp. in one of three initial replicates tested. Three additional replicates were run, giving an overall detection rate for Streptococcus spp. of 5/6 for that strain. S. mutans strain 31383 was fully detected in all replicates.

Table 7: LIAISON PLEX® Gram-Positive Blood Culture (BCP) Inclusivity Summary – Antimicrobial Resistance Markers

Reportable Target (Resistance Marker)Organism# of strains% Detected
mecA/CStaphylococcus argenteus1100%
mecA/CStaphylococcus aureus35100%
Staphylococcus epidermidis5100%
Staphylococcus hominis1100%
Mammaliicoccus (Staphylococcus) sciuri2100%
vanAEnterococcus faecalis3100%
Enterococcus faecium3100%^a
vanBEnterococcus faecalis3100%
Enterococcus faecium3100%

^a One replicate from one strain resulted in a false positive (FP) S. lugdunensis result, and three additional replicates were run for that strain. No additional FPs were observed.

Predicted (in silico) Reactivity (Inclusivity) Results

For all targets, in silico inclusivity analysis was performed using sequences available in the GenBank and WGS (whole genome shotgun) databases from March to April 2024. Alignments of the signal fragment for each target were generated using MAFFT (version 7.490). For all targets, the inclusivity analysis involved assessing the percent homology of each oligo sequence to its binding region on each target sequence retrieved from the public databases. The predicted inclusivity based on sequence homology is the percentage of sequences with at least 90% oligo identity. In determining oligo identity, the highest percent identity of each oligo in the same component was assessed and then the lowest percent identity of the two components (capture and mediator probes) is used to characterize the inclusivity of the sequence. In silico analysis results are summarized below in Table 8.

Page 26

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 21 of 43

Table 8 – In silico Analysis Results

Reportable TargetInclusive Organism/TargetTotal # Sequences in Alignment# Sequences with Percent Oligo Identity ≥ 90%Predicted Inclusivity Percentage (%)
Bacillus spp.^aBacillus cereus group67226718100
Bacillus subtilis group43204318100
Enterococcus faecalisEnterococcus faecalis746745100
Enterococcus faeciumEnterococcus faecium555555100
Listeria spp.^bListeria monocytogenes88128812100
other Listeria species89987697
Staphylococcus aureusStaphylococcus aureus23072307100
Staphylococcus epidermidisStaphylococcus epidermidis250250100
Staphylococcus lugdunensisStaphylococcus lugdunensis162162100
Staphylococcus spp.^cStaphylococcus spp.1335513354100
Streptococcus agalactiaeStreptococcus agalactiae454454100
Streptococcus anginosus GroupStreptococcus anginosus356356100
Streptococcus constellatus7070100
Streptococcus intermedius127127100
unspecified species from Streptococcus anginosus group77100
Streptococcus pneumoniaeStreptococcus pneumoniae693693100
Streptococcus pyogenesStreptococcus pyogenes732732100
Streptococcus spp.^dStreptococcus spp.2358923574100
mecA/mecC^emecA2201215898
mecC3232100
vanA^fvanA57055197
vanB^fvanB215215100

^a Includes sequences for 26 species of B. cereus group and 15 species of B. subtilis group. Oligo designs may also detect Bacillus species that are not part of these two groups.
^b Includes sequences for 28 Listeria species.
^c Includes sequences for 44 Staphylococcus species.
^d Includes sequences for 108 Streptococcus species.
^e Includes sequences for Staphylococcus species only.
^f Includes sequences for E. faecalis and E. faecium only.

f. Analytical Specificity (Exclusivity)

Cross-Reactivity

The analytical specificity study was performed to evaluate cross-reactivity of the LIAISON PLEX® BCP Assay with "off-panel" organisms, including both phylogenetically related to the on-panel organisms and organisms likely to be present in typical blood culture samples. Table 9 contains the summary for the genus and species reportable targets. Out

Page 27

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 22 of 43

of the 103 off-panel species tested, 97 resulted in no cross-reactivity, 5 resulted in cross-reactivity with one reportable target, and 1 organism resulted in positivity for 1 out of 6 replicates tested. Organisms with cross-reactivity and the BCP target that was detected are listed in Table 10.

Table 9: LIAISON PLEX® BCP Assay Analytical Specificity (Cross Reactivity) Summary

OrganismPositivityOrganismPositivity
Gram Positive
Abitrophia defectiva0%Enterococcus flavescens0%
Aerococcus viridans0%Enterococcus gallinarum, vanC0%
Arcanobacterium bernardiae0%Enterococcus hirae0%
Arcanobacterium haemolyticum0%Enterococcus mundtii0%
Corynebacterium amycolatum0%Enterococcus raffinosus0%
Corynebacterium diphtheriae0%Erysipelothrix rhusiopathiae0%
Corynebacterium flavescens0%Kocuria kristinae0%
Corynebacterium genitalium0%Kytococcus sedentarius, methicillin resistant0%
Corynebacterium glutamicum0%Lactobacillus acidophilus0%
Corynebacterium jeikeium0%Lactobacillus crispatus0%
Corynebacterium pseudodiptherticum0%Lactobacillus rhamnosus0%
Corynebacterium renale0%Leuconostoc carnosum100%
Corynebacterium striatum0%Leuconostoc mesenteroides0%
Corynebacterium urealyticum0%Pediococcus acidilactici0%
Cutibacterium acnes0%Pediococcus pentosaceus0%
Cutibacterium avidum0%Peptostreptococcus anaerobius0%
Propionibacterium freudenreichii0%Planococcus citreus0%
Enterococcus avium100%Planococcus kocurii0%
Enterococcus casseliflavis, VRE, vanC0%Rothia dentocariosa0%
Enterocococcus dispar0%Rothia (Stomatococcus) mucilaginosa0%
Enterococcus durans0%
Gram Negative
Acinetobacter calcoaceticus0%Haemophilus influenzae0%
Acinetobacter pittii0%Herbaspirillum huttiense0%
Aggregatibacter aphrophilus100%Kingella kingae0%
Bacteroides fragilis0%Klebsiella oxytoca0%
Brevundimonas diminuta0%Klebsiella pneumoniae0%
Burkholderia cepacia0%Klebsiella variicola0%
Capnocytophaga ochracea0%Leclercia adecarboxylata0%
Cardiobacterium hominis0%Moraxella catarrhalis0%
Cedecea lapagei0%Morganella morganii0%
Citrobacter amalonaticus0%Neisseria lactamica0%

Page 28

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 23 of 43

OrganismPositivityOrganismPositivity
Citrobacter freundii0%Neisseria meningitidis0%
Comamonas testosteroni0%Neisseria mucosa0%
Delftia acidovorans0%Neisseria sicca0%
Eikenella corrodens0%Parabacteroides distasonis0%
Elizabethkingia meningoseptica0%Pasteurella aerogenes0%
Enteric Group 1370%Prevotella bivia0%
Enterobacter aerogenes0%Proteus mirabilis0%
Enterobacter cloacae0%Proteus penneri100%
Enterobacter hormaechei0%Proteus vulgaris100%
Escherichia albertii0%Pseudomonas aeruginosa0%
Escherichia coli0%Pseudomonas alcaligenes0%
Escherichia fergusonii0%Raoultella ornithinolytica0%
Escherichia hermannii0%Salmonella enterica0%
Fusobacterium necrophorum0%Serratia marcescens0%
Fusobacterium nucleatum0%Stenotrophomonas maltophilia0%
Haemophilus haemolyticus0%
Yeast/Fungi
Aspergillus fumigatus0%Cryptococcus neoformans0%
Candida albicans0%Debaryomyces hansenii (Candida famata)0%
Nakaseomyces glabratus (Candida glabrata)0%Kluyveromyces lactis0%
Pichia kudriavzevii (Candida krusei)0%Saccharomyces cerevisiae0%
Candida parapsilosis16.7%^aSchizosaccharomyces pombe0%
Candida tropicalis0%

^a Initial testing results showed 1 of 3 replicates detected as S. lugdunensis. An additional 3 replicates were tested (for a total of 6 replicates) and no additional false positives were seen.

Table 10: LIAISON PLEX® BCP Assay Analytical Specificity (Cross Reactivity) Summary

Cross-Reactive OrganismBCP Target Detected
Enterococcus aviumEnterococcus faecium
Leuconostoc carnosumStreptococcus spp.
Aggregatibacter aphrophilusStreptococcus spp.
Proteus penneriStreptococcus spp.
Proteus vulgarisStreptococcus spp.

Page 29

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 24 of 43

In Silico Cross-Reactivity

For in silico exclusivity assessment of the oligo sequences incorporated in the assay designs against on-panel and off-panel organisms listed in Table 11, based on analysis of sequences available in the GenBank nt database as of July 27, 2024, the following potential cross-reactivity is predicted:

  • Some strains of Enterococcus durans and 3 strains of Streptococcus species (AY123726.1, FJ577604.1, MK608388.1) are predicted to produce false positive Enterococcus faecalis results.
  • One strain of Enterococcus durans (KT877992.1), 1 strain of Enterococcus faecalis (CP092577.1) and some strains Enterococcus avium are predicted to produce false positive Enterococcus faecium results.
  • Staphylococcus aureus oligo designs are predicted to detect some strains of Staphylococcus argenteus and one strain of Staphylococcus schweitzeri (LR134304.1).
  • Staphylococcus spp. oligo designs are predicted to detect some strains of Macrococcoides caseolyticum.
  • Some strains of Streptococcus species (S. lactarius, S. milleri, S. oralis, S. periodonticum, S. pneumoniae, S. rubneri, S. vaginalis) are predicted to produce false positive Streptococcus anginosus Group results.
  • Streptococcus pneumoniae oligo designs are predicted to detect a few strains of various Streptococcus species (S. downii, S. mitis, S. oralis).
  • Some strains of Aggregatibacter aphrophilus, unspecified Bacillus species, Enterococcus species (E. durans, E. gallinarum), Glaesserella parasuis, Haemophilus species (H. ducreyi, H. influenzae, H. parahaemolyticus, H. parainfluenzae), Lactococcus species (L. garvieae, L. lactis), Leuconostoc carnosum, Moellerella wisconsensis, Morganella morganii, Pasteurella multocida, Proteus species (P. appendicitidis, P. columbae, P. faecis, P. penneri, P. terrae, P. vulgaris), Providencia species (P. alcalifaciens, P. hangzhouensis, P. heimbachae, P. huaxiensis, P. manganoxydans, P. rettgeri, P. rustigianii, P. stuartii, P. vermicola, P. zhijiangensis), Serratia species (S. marcescens, S. symbiotica) and Xenorhabdus species (X. bovienii, X. hominickii, X. nematophila) are predicted to produce false positive Streptococcus spp. results.
  • Listeria spp. oligo designs are predicted to detect some strains of Staphylococcus muscae.

Table 11: Potential Cross-Reactive Organisms assessed in the In Silico Exclusivity Analysis

On-Panel OrganismsOff-Panel Organisms
Gram-Positive BacteriaGram-Negative BacteriaResistance MarkersYeasts / Viruses / Parasites
Bacillus spp.Abiotrophia defectivaAcinetobacter spp.AmpCAspergillus flavus
Enterococcus faecalisActinomyces israeliiActinobacillus hominisCMYAspergillus fumigatus
Enterococcus faeciumActinomyces naeslundiiActinobacillus ureaeCTX-MAspergillus niger
Listeria spp.Actinomyces odontolyticusAeromonas caviaeIMPAspergillus terreus
Staphylococcus aureusAerococcus sanguinicolaAeromonas hydrophilaKPCBlastomyces dermatitidis
Staphylococcus epidermidisAerococcus urinaeAeromonas sobriaMCRCandida albicans
Staphylococcus lugdunensisAerococcus viridansAggregatibacter actinomycetemcomitansNDMCandida auris
Staphylococcus spp.Arcanobacterium bernardiaeAggregatibacter aphrophilusompK36Candida dubliniensis
Streptococcus agalactiaeArcanobacterium haemolyticumBacteroides caccaeOXACandida famata
Streptococcus anginosus groupArthrobacter psychrolactophilusBacteroides fragilisRAHNCandida glabrata

Page 30

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 25 of 43

On-Panel OrganismsOff-Panel Organisms
Gram-Positive BacteriaGram-Negative BacteriaResistance MarkersYeasts / Viruses / Parasites
Streptococcus pneumoniaeBrochothrix thermosphactaBacteroides ovatusSHVCandida guilliermondii
Streptococcus pyogenesCarnobacterium divergensBacteroides thetaiotaomicronSMECandida haemulonii
Streptococcus spp.Carnobacterium maltaromaticumBacteroides uniformisSPMCandida inconspicua
mecACellulomonas turbataBacteroides vulgatusTEMCandida kefyr
mecCCellulosimicrobium cellulansBacteroides xylanisolvensvanCCandida krusei
vanAClostridioides difficileBordetella bronchisepticavanDCandida lipolytica
vanBClostridium bifermentansBordetella parapertussisvanMCandida lusitaniae
Clostridium clostridioformeBordetella pertussisVIMCandida metapsilosis
Clostridium perfringensBrevundimonas diminutaCandida multis-gemmis
Clostridium ramosum (Thomasclavelia ramosa)Brevundimonas vesicularisCandida nivariensis
Clostridium septicumBurkholderia cepaciaCandida norvegensis
Clostridium tertiumBurkholderia malleiCandida orthopsilosis
Clostridium tetaniBurkholderia multivoransCandida parapsilosis
Corynebacterium spp.Burkholderia pseudomalleiCandida sojae
Cutibacterium acnesCampylobacter hominisCandida tropicalis
Cutibacterium avidumCapnocytophaga ochraceaCandida duobushaemulonii
Cutibacterium granulosumCardiobacterium hominisCandida viswanathii
Enterococcus aviumChlamydia trachomatisCoccidioides immitis
Enterococcus casseliflavusChlamydophila pneumoniaeCoccidioides posadasii
Enterococcus cecorumChromobacterium violaceumCryptococcus amylolentus
Enterococcus disparCitrobacter spp.Cryptococcus gattii
Enterococcus duransComamonas testosteroniCryptococcus neoformans
Enterococcus flavescensDelftia acidovoransCryptococcus uniguttulatus
Enterococcus gallinarumEikenella corrodensCutaneotrichosporon curvatum
Enterococcus hiraeElizabethkingia meningosepticaCyberlindnera fabianii
Enterococcus mundtiiEnterobacter spp.Geotrichum capitatum (Magnusiomyces capitatus)
Enterococcus raffinosusEnterobacteriaceaeHistoplasma capsulatum
Erysipelothrix rhusiopathiaeEscherichia coliKluyveromyces lactis
Finegoldia magnaFusobacterium necrophorumKodamaea ohmeri
Gemella haemolysansFusobacterium nucleatumLodderomyces elongisporus
Gemella morbillorumHaemophilus influenzaeMagnusiomyces capitatus
Globicatella spp.Haemophilus aegyptiusMillerozyma farinosa
Granulicatella adiacensHaemophilus ducreyiNaganishia albida
Granulicatella elegansHaemophilus haemolyticusPapiliotrema laurentii
Kocuria kristinaeHaemophilus parahaemolyticusPenicillium chrysogenum
Kocuria rhizophilaHaemophilus parainfluenzaeRhodotorula mucilaginosa
Kytococcus sedentariusHaemophilus parasuisSaccharomyces cerevisiae
Lactobacillus acidophilus
Lactobacillus crispatus

Page 31

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 26 of 43

On-Panel OrganismsOff-Panel Organisms
Gram-Positive BacteriaGram-Negative BacteriaResistance MarkersYeasts / Viruses / Parasites
Lactobacillus rhamnosusHaemophilus quentiniSchizosaccharomyces pombe
Lactococcus garvieaeHaemophilus sputorumTalaromyces marneffei
Lactococcus lactisHerbaspirillum huttienseTrichosporon asahii
Leuconostoc carnosumKingella denitrificansWickerhamomyces anomalus
Leuconostoc citreumKingella kingaeBK Virus
Leuconostoc mesenteroidesKingella negevensisChikungunya Virus
Macrococcus caseolyticusKingella oralisCytomegalovirus
Micrococcus luteusKlebsiella oxytocaDengue Virus
Mycobacterium avium complex (MAC)Klebsiella pneumoniaeEnterovirus
Mycobacterium fortuitumKlebsiella variicolaEpstein Barr Virus
Mycobacterium mucogenicumLegionella pneumophilaHepatitis A virus
Mycoplasma hominisLeptospira interrogansHepatitis B virus
Mycoplasma pneumoniaeMoraxella catarrhalisHepatitis C virus
Nocardia farcinicaMoraxella osloensisHuman alphaherpesvirus 1
Parvimonas micraMorganellaceaeHuman alphaherpesvirus 2
Pediococcus acidilacticiMycobacterium tuberculosisHuman betaherpesvirus 6
Pediococcus pentosaceusNeisseria gonorrhoeaeHuman betaherpesvirus 7
Peptostreptococcus anaerobiusNeisseria lactamicaHuman Immunodeficiency Virus
Planococcus citreusNeisseria meningitidisJC Virus
Planococcus kocuriiNeisseria mucosaMeasles Virus
Propionibacterium freudenreichiiNeisseria siccaMumps Virus
Propionibacterium propionicum (Arachnia propionica)Parabacteroides distasonisParvovirus B19
Rhodococcus equiParabacteroides merdaeRubella Virus
Rothia dentocariosaPasteurella aerogenesVaricella Zoster Virus
Rothia mucilaginosaPasteurella canisWest Nile Virus
Sarcina ventriculiPasteurella multocidaZika Virus
Solibacillus silvestrisPasteurella stomatisPlasmodium falciparum
Ureaplasma parvumPrevotella biviaTrypanosoma cruzi
Ureaplasma urealyticumPrevotella buccae
Vagococcus fluvialisPrevotella denticola
Weissella paramesenteroidesPrevotella melaninogenica
Prevotella oralis
Proteus spp.
Pseudomonas spp.
Psychrobacter cryohalolentis
Psychrobacter immobilis
Ralstonia mannitolilytica
Ralstonia pickettii

Page 32

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 27 of 43

On-Panel OrganismsOff-Panel Organisms
Gram-Positive BacteriaGram-Negative BacteriaResistance MarkersYeasts / Viruses / Parasites
Salmonella spp.
Serratia spp.
Shigella spp.
Stenotrophomonas acidaminiphila
Stenotrophomonas maltophilia
Stenotrophomonas nitritireducens
Stenotrophomonas rhizophila
Treponema pallidum
Veillonella parvula
Vibrio alginolyticus
Vibrio parahaemolyticus
Vibrio vulnificus

g. Interference

Competitive Inhibition / Co-Infection and Microbial Interference

The competitive inhibition and microbial interference study was executed to evaluate the performance of the LIAISON PLEX® BCP Assay in the presence of clinically significant levels of potential co-infections of on-panel organisms and potentially interfering (off-panel) microbes. To evaluate potential co-infections, varying ratios of two on-panel target organisms including a low titer (concentration at bottle/ring positive) and a high titer specimen (concentration at bottle/ring positive +8 hours) were tested in combination, as listed in Table 12. The on-panel target organisms were chosen to be representative of clinically relevant poly-microbial infections in blood culture specimens. To assess microbial interference, three off-panel microbes were combined in pairs with each of three representative on-panel target organisms, as listed in Table 13. These specific organism sets were chosen to mimic potentially interfering micro-organisms that are commonly found in positive blood culture samples but are not designed to be detected by the LIAISON PLEX® BCP Assay.

Results are shown in Tables 12 and Table 13. On-panel targets were detected with 100% positivity for all combinations tested, both in the presence of competitive on-panel targets and in the presence of competitive off-panel pathogens.

Page 33

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 28 of 43

Table 12: LIAISON PLEX® BCP Assay Competitive Inhibition Summary

On-Panel High Titer TargetPositivityOn-Panel Low Titer TargetPositivity
Enterococcus faecalis100%Staphylococcus epidermidis100%
100%Enterococcus faecium100%
Enterococcus faecium100%Staphylococcus epidermidis100%
100%Enterococcus faecalis100%
Staphylococcus aureus100%Staphylococcus epidermidis100%
100%Streptococcus agalactiae100%
Staphylococcus epidermidis100%Enterococcus faecalis100%
100%Enterococcus faecium100%
100%Staphylococcus aureus100%
100%Staphylococcus lugdunensis100%
100%Streptococcus pneumoniae100%
Staphylococcus lugdunensis100%Staphylococcus epidermidis100%
Streptococcus agalactiae100%Staphylococcus aureus100%
Streptococcus pneumoniae100%Staphylococcus epidermidis100%

Table 13: LIAISON PLEX® BCP Assay Microbial Interference Summary

On-Panel Low Titer TargetPositivityOff-Panel High Titer TargetPositivity
Staphylococcus epidermidis100%Escherichia coli0%
100%Klebsiella pneumoniae0%
100%Proteus mirabilis0%
Enterococcus faecium100%Escherichia coli0%
100%Klebsiella pneumoniae0%
100%Proteus mirabilis0%
N/AN/AEscherichia coli0%
N/AN/AKlebsiella pneumoniae0%
N/AN/AProteus mirabilis0%
Interfering Substances

The interfering substances study was performed to evaluate the performance of the LIAISON PLEX® BCP Assay in the presence of non-microbial (endogenous and exogenous) interfering substances which may be present in blood culture specimens. Six representative "on-panel" organisms were individually tested to assess effectiveness of target detection in the presence of six typically occurring interfering substances. Each interfering substance was tested across five replicates for each organism. In addition, five replicates of a negative control was tested alongside the positive specimens to assess impact of the same interfering agents in specimens containing no target. A positive control (specimen without interfering substances) for all four targets was also tested to assess for detection capabilities.

Page 34

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 29 of 43

As seen in Table 14, 100% target detection was observed for all six targets without interferent and in the presence of all six interfering substances. 0% target detection was observed with the negative sample with and without interfering substances.

Table 14: LIAISON PLEX® BCP Assay Interfering Substances Summary

OrganismInterfering Substance & Tested Concentration
Unconjugated BilirubinConjugated BilirubinHemoglobinIntralipidγ-globulinSodium polyanetholsulfonateNo interferent
20 mg/dL20 mg/dL14 g/L3000 mg/dL6 g/dL0.25% w/v
% Positivity
Staphylococcus aureus100%100%100%100%100%100%100%
Staphylococcus epidermidis100%100%100%100%100%100%100%
Streptococcus pneumoniae100%100%100%100%100%100%100%
Streptococcus agalactiae100%100%100%100%100%100%100%
Enterococcus faecalis100%100%100%100%100%100%100%
Enterococcus faecium100%100%100%100%100%100%100%
Negative Blood Matrix0%0%0%0%0%0%0%
Carry-Over/Cross Contamination

This study was performed to evaluate the risk of carry-over and cross contamination occurring during normal use of the device when highly concentrated positive specimens are processed alongside negative specimens. Two operators tested 30 high concentration positive samples consisting of Staphylococcus aureus at a final concentration of 1.90E+08 CFU/mL, and 30 negative samples consisting of Negative Blood Matrix. Staphylococcus aureus was selected as it is a representative organism with three expected positive target detections on the LIAISON PLEX® BCP Assay.

Testing was performed on two LIAISON PLEX® instruments containing 6 modules over the course of four days. The samples were loaded into cartridges alternating between positive and negative samples (checkerboard fashion), six specimens at a time using sample prep trays. The results, presented in Table 15, demonstrate 100% agreement between expected and observed results, indicating that no cross contamination occurred within runs and no carry-over was observed across runs.

Page 35

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 30 of 43

Table 15. Overall Results

Overall Percent Agreement between Expected and Observed Results100%
Staphylococcus aureus Positivity in High Positive Control Replicates100%
Staphylococcus aureus Positivity in Negative Control Replicates0%

h. Assay Cut-off

The specific assay parameters for the LIAISON PLEX® BCP Assay are considered confidential and proprietary.

2. Comparison Studies:

a. Method comparison with predicate device:

Refer to Section 3 Clinical Performance.

b. Matrix Comparison: Testing of Blood Culture Bottle Types / Matrix Equivalency

The performance of LIAISON PLEX® Gram-Positive Blood Culture (BCP) Assay was evaluated across 12 additional types of commercially available blood culture media bottles using target organisms representative of all "on-panel" BCP targets with resistance markers, representative off-panel (gram-negative) targets, and negative blood matrix (NBM). The results, presented in Table 16, demonstrate that all 12 bottle (media) types are compatible with the LIAISON PLEX® Gram-Positive Blood Culture (BCP) Assay. BACT/ALERT® FA Plus blood culture bottle type is established in the clinical evaluation and for all other analytical studies, such as, growth and detection and analytical reactivity/ inclusivity verification testing and was not included in testing performed for the media equivalency study. Table 17 presents the range of CFU/mL by organism across all bottle types tested in the study.

Table 16: LIAISON PLEX® Gram-Positive Blood Culture (BCP) Assay Media Equivalency/ Universal Blood Culture Bottle Result Summary

Manufacturer SystemBlood Culture Bottle ManufacturerBlood Culture Bottle Type^cUnique BottlesSeeded OrganismsNegative Blood Matrix^d
Gram-positive BacteriaGram-negative Bacteria
Expected Results
bioMérieux BACT/ALERT® 3D SystembioMérieux BACT/ALERT®Aerobic BACT/ALERT® SA20/2010/101/1
Anaerobic BACT/ALERT®20/2010/101/1

Page 36

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 31 of 43

Manufacturer SystemBlood Culture Bottle ManufacturerBlood Culture Bottle Type^cUnique BottlesSeeded OrganismsNegative Blood Matrix^d
Gram-positive BacteriaGram-negative Bacteria
SN
BACT/ALERT® FN Plus20/2010/101/1
Pediatric BACT/ALERT® PF Plus58/5810/101/1
Becton Dickinson 9050 / FX40Becton Dickinson^aBACTEC™ Aerobic
BACTEC™ Standard20/20^b10/101/1
BACTEC™ Plus20/2010/101/1
Anaerobic
BACTEC™ Standard58/58^b10/101/1
BACTEC™ Plus58/5810/101/1
Pediatric BACTEC™ Peds Plus58/5810/101/1
Lytic Anaerobic BACTEC™ Lytic20/2010/101/1
N/AThermo Scientific^aVersaTREK™ Aerobic REDOX™ 1 EZ Draw™56/5610/101/1
Anaerobic REDOX™ 2 EZ Draw™56/5610/101/1
Bottles with Expected Results /Total Number Blood Bottles464/464^b (100%)120/120 (100%)12/12 (100%)

^a For gram-negative blood culture preparation, BACTEC™ and VersaTREK™ systems were not available and corresponding media bottles were placed in a standard laboratory incubator with a shaker for growth. For gram-positive blood culture growth, only VersaTREK™ bottles needed to be placed in a standard laboratory incubator with a shaker in lieu of VersaTREK™ system; BACT/ALERT® and BACTEC™ blood bottles were grown using corresponding automated blood culture systems up to bottle ring positivity.
^b One organism, (E. faecalis strain 51575) required use of a ring positivity + 8 bottle to confirm the bottle matrix was not inhibitory.
^c bioMérieux BACT/ALERT® FA Plus blood culture bottle type (PN: 410851) was not tested in this study protocol as the effectiveness of this is bottle type was established through blood culture growth required for other analytical studies including growth and detection and analytical reactivity/inclusivity verification.
^d Each unique bottle of Negative Blood Matrix was tested in replicates of three.

Table 17: Range of CFU/mL by Organism Across All Bottle Types Tested

Gram-Positive OrganismsStrain IDExpected ResultBottle Titer Range (CFU/mL)
Bacillus cereusATCC 10702Bacillus spp.6.10E+07^b - 4.40E+08
Bacillus subtilisATCC 19659Bacillus spp.4.90E+05 - 3.73E+08
Enterococcus faecalisATCC 51575Enterococcus faecalis | vanB8.63E+07 - 3.50E+09^a
Enterococcus faecalisClinical Isolate CLCS VRE-1Enterococcus faecalis | vanA9.20E+06 - 6.50E+08
Enterococcus faeciumATCC 700221Enterococcus faecium | vanA3.17E+07 - 9.77E+08
Enterococcus faeciumATCC 51858Enterococcus faecium | vanB3.83E+06 - 5.50E+08

Page 37

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 32 of 43

Gram-Positive OrganismsStrain IDExpected ResultBottle Titer Range (CFU/mL)
Listeria ivanoviiATCC 700402Listeria spp.1.50E+08 - 1.40E+09
Listeria monocytogenesATCC 15313Listeria spp.6.60E+07 - 1.88E+09
Staphylococcus aureusATCC BAA-2312Staphylococcus spp. | S. aureus | mecA/C8.00E+06 - 1.29E+09
Staphylococcus aureusCDC AR-0227Staphylococcus spp. |S. aureus | mecA/C1.43E+07 - 3.50E+09
Staphylococcus epidermidisATCC 35984Staphylococcus spp. | S. epidermidis |mecA/C6.93E+06 - 1.43E+09
Staphylococcus lugdunensisATCC 49576Staphylococcus spp. | S. lugdunensis2.20E+07 - 3.46E+09
Streptococcus agalactiaeATCC 12386Streptococcus spp. | S. agalactiae2.70E+07 - 1.71E+09
Streptococcus anginosusATCC 33397Streptococcus spp. | S. anginosus Group1.82E+07 - 4.37E+09
Streptococcus constellatusATCC 27823Streptococcus spp. | S. anginosus Group4.00E+07 - 2.53E+09
Streptococcus pneumoniaeATCC 49619Streptococcus spp. | S. pneumoniae2.70E+06 - 2.88E+09
Streptococcus pyogenesATCC 700294Streptococcus spp. | S. pyogenes2.80E+06 - 2.07E+09
Corynebacterium diphtheriaeATCC 27010No target detected2.80E+06 - 1.31E+09
Corynebacterium striatumATCC 43735No target detected4.30E+07 - 1.17E+09
Cutibacterium acnesATCC 6919No target detected2.83E+07 - 5.90E+09
Gram-negativeStrain IDExpected ResultBottle Titer Range (CFU/mL)
Acinetobacter baumanniiIHMA 128307No target detected3.90E+07 - 9.10E+08
Escherichia coliNCTC 13846No target detected6.70E+08 - 2.11E+09
Haemophilus influenzaeATCC 9007No target detected2.30E+08 - 2.93E+09
Klebsiella oxytocaIHMA 683079No target detected6.23E+07 - 1.56E+09
Klebsiella pneumoniaeIHMA 629630No target detected2.17E+08 - 1.19E+09
Klebsiella variicolaClinical Isolate V0512No target detected2.60E+08 - 1.68E+09
Pseudomonas aeruginosaIHMA 576602No target detected2.10E+07 - 9.70E+08
Proteus mirabilisATCC 12453No target detected1.10E+08 - 9.03E+08

Page 38

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 33 of 43

Gram-Positive OrganismsStrain IDExpected ResultBottle Titer Range (CFU/mL)
Serratia marcescensIHMA 1642209No target detected7.57E+08 - 2.87E+09
Citrobacter freundiiIHMA 549813No target detected3.93E+08 - 2.06E+09

^a Two bottles tested at ring positive with titers of 8.63E+07 and 1.46E+08 were not fully detected. Bottles grown to ring positivity + 8 hours were used to confirm that the results were not due to bottle matrix.
^b No viable counts were achieved for B. cereus in two of the BD BACTEC™ Standard Anaerobic media bottles.

3. Clinical Performance:

A multi-site clinical study established the diagnostic accuracy of the LIAISON PLEX® Gram-Positive Blood Culture (BCP) Assay for the detection and identification of pathogenic gram-positive organisms in positive blood culture. The clinical performance of the LIAISON PLEX® BCP Assay was evaluated using clinical specimens prospectively collected between April 2024 and August 2024 from four geographically diverse clinical sites within the United States. The clinical study utilized remnant, de-identified blood culture specimens collected from patients exhibiting clinical signs and symptoms of bloodstream infection, evidenced by positive identification by a continuous monitoring blood culture system.

A total of 562 unique prospectively collected specimens that met the pre-determined inclusion criteria were enrolled in the study. Clinical runs and re-runs using the LIAISON PLEX® BCP Assay were tested on the LIAISON PLEX® System by trained operators at four clinical sites. For targets that exhibited low prevalence rates in the prospective study, the prospective specimen set was supplemented with 163 pre-selected left-over, de-identified specimens sourced from ten vendors in the United States and one site in Italy. The pre-selected specimens were identified by Standard of Care (SoC) testing and confirmed as positive by VITEK 2 and/or PCR/BDS according to the reference method algorithm prior to enrollment in the study. To minimize bias, pre-selected specimens were tested across four external and one internal site in a randomized, blinded manner along with negative specimens.

Out of the 562 specimens enrolled in the prospective arm of the study, 53 prospective specimens were excluded from the analysis (one (1) duplicate patient enrollment, one (1) due to sample handling errors and 51 with incomplete reference testing results due to lack of clinical isolates, mixed growth, insufficient growth, or no growth). The remaining 509 specimens were included in the analysis.

For targets that exhibited low prevalence rates in the prospective study, the prospective specimen set was supplemented with 163 pre-selected (Arm 2) and 225 contrived specimens (Arm 3). Of these, one pre-selected specimen was excluded (sample handling error), leaving 162 pre-selected and 225 contrived specimens included in the analysis.

Table 18 provides a summary of the general demographic information for the 509 prospectively collected and 162 pre-selected specimens that were included in the study analysis.

Page 39

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 34 of 43

Table 18: LIAISON PLEX® BCP Assay Summary of the General Demographic Information

Prospective (N=509)Preselected (N=162)
#Specimens(%)#Specimens(%)
Gender All Sites
Male295 (58.0%)87 (53.7%)
Female214 (42.0%)71 (43.8%)
Gender Unknown0 (0.0%)4 (2.5%)
Total509 (100.0%)162 (100.0%)
Age (years)
0-115 (2.9%)4 (2.5%)
>1-50 (0.0%)1 (0.6%)
>5-2111 (2.2%)5 (3.1%)
>21-65266 (52.3%)83 (51.2%)
>65213 (41.8%)57 (35.2%)
Age Unknown4 (0.8%)12 (7.4%)
Total509 (100.0%)162 (100.0%)
Subject Status
Emergency Room88 (17.3%)0 (0.0%)
Hospitalized391 (76.8%)4 (2.5%)
Outpatient2 (0.4%)0 (0.0%)
Status Unknown28 (5.5%)158 (97.5%)
Total509 (100.0%)162 (100.0%)
Blood Culture Bottle Type
BacT/ALERT FA Plus148 (29.1%)48 (29.6%)
BacT/ALERT FN Plus93 (18.3%)31 (19.1%)
BacT/ALERT PF Plus4 (0.8%)2 (1.2%)
BacT/ALERT SA Standard Aerobic27 (5.3%)0 (0.0%)
BacT/ALERT SN Standard Anaerobic21 (4.1%)0 (0.0%)
BD BACTEC Lytic Anaerobic55 (10.8%)2 (1.2%)
BD BACTEC Peds Plus8 (1.6%)0 (0.0%)
BD BACTEC Plus Aerobic135 (26.5%)2 (1.2%)
BD BACTEC Standard Aerobic18 (3.5%)42 (25.9%)
BD BACTEC Standard Anaerobic0 (0.0%)7 (4.3%)
Bottle Type Unknown0 (0.0%)28 (17.3%)
Bottle Type Total509 (100.0%)162 (100.0%)

The clinical performance of the LIAISON PLEX® BCP Assay was compared to culture followed by automated microbiological/biochemical identification using VITEK 2, PCR followed by BDS, or a combination according to the algorithm described in Table 19 below.

Page 40

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 35 of 43

Table 19: LIAISON PLEX® BCP Assay Summary of Comparator Method

LIAISON PLEX® BCP Assay TargetComparator Method
Enterococcus faecalisCulture followed by Automated microbiological/biochemical identification using VITEK 2
Enterococcus faecium
Listeria spp.
Staphylococcus spp.
Staphylococcus aureus
Staphylococcus epidermidis
Staphylococcus lugdunensis
Streptococcus spp.
Streptococcus agalactiae
Streptococcus anginosus group
Streptococcus pneumoniae
Streptococcus pyogenes
Bacillus spp.PCR followed by bi-directional sequencing
mecA/mecC
vanA
vanB

Of the 671 clinical specimens collectively included in the prospective and pre-selected study analysis (Arms 1 and 2 combined), 646 samples (96.3%) generated valid LIAISON PLEX BCP Assay results (i.e., Detected or Not Detected) on the first attempt. There were 25 specimens (3.7%) with an invalid result on the initial run. Of the 25 specimens retested, 24 specimens generated valid BCP results after a single retest for a final testing success rate of 99.9% (670/671).

Prospective and Pre-selected Clinical Evaluation

Clinical Performance (Sensitivity, Specificity, and 95% confidence interval) of the LIAISON PLEX® BCP Assay compared to the reference method is summarized in Table 20 for combined prospective and pre-selected specimens.

Table 20. LIAISON PLEX® Gram-Positive Blood Culture Assay Performance with Prospective and Pre-selected Specimens - Arm 1 and Arm 2

Pathogen TargetSensitivity/PPASpecificity/NPA
TP / (TP+FN)Sensitivity /PPA (%)95% CITN / (TN+FP)Specificity/NPA (%)95% CI
Bacteria
Bacillus spp.
Prospective1/1100%20.7% - 100%506/50799.8%98.9% - 100%
Pre-selected17/1989.5%68.6% - 97.1%127/127100%97.1% - 100%
Combined18/2090%69.9% - 97.2%633/634¹99.8%99.1% - 100%

Page 41

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 36 of 43

Pathogen TargetSensitivity/PPASpecificity/NPA
TP / (TP+FN)Sensitivity /PPA (%)95% CITN / (TN+FP)Specificity/NPA (%)95% CI
Enterococcus faecalis
Prospective38/38100%90.8% - 100%468/468100%99.2% - 100%
Pre-selected2/2100%34.2% - 100%125/125100%97% - 100%
Combined40/40100%91.2% - 100%593/593100%99.4% - 100%
Enterococcus faecium
Prospective16/16100%80.6% - 100%489/49099.8%98.9% - 100%
Pre-selected20/20100%83.9% - 100%106/10799.1%94.9% - 99.8%
Combined36/36100%90.4% - 100%595/597²99.7%98.8% - 99.9%
Listeria spp.
Prospective0/0NA%NA506/506100%99.2% - 100%
Pre-selected5/5100%56.6% - 100%122/122100%96.9% - 100%
Combined5/5100%56.6% - 100%628/628100%99.4% - 100%
Staphylococcus spp.
Prospective316/32298.1%96% - 99.1%182/18498.9%96.1% - 99.7%
Pre-selected20/20100%83.9% - 100%107/107100%96.5% - 100%
Combined336/342³98.2%96.2% - 99.2%289/291⁴99.3%97.5% - 99.8%
Staphylococcus aureus
Prospective160/16199.4%96.6% - 99.9%344/34599.7%98.4% - 99.9%
Pre-selected0/0NA %NA127/127100%97.1% - 100%
Combined160/16199.4%96.6% - 99.9%471/472⁵99.8%98.8% - 100%
Staphylococcus epidermidis
Prospective93/9696.9%91.2% - 98.9%405/41098.8%97.2% - 99.5%
Pre-selected0/0NANA126/12799.2%95.7% - 99.9%
Combined93/96⁶96.9%91.2% - 98.9%531/537⁷98.9%97.6% - 99.5%
Staphylococcus lugdunensis
Prospective6/6100%61% - 100%500/500100%99.2% - 100%
Pre-selected20/20100%83.9% - 100%107/107100%96.5% - 100%
Combined26/26100%87.1% - 100%607/607100%99.4% - 100%
Streptococcus spp.
Prospective97/9899%94.4% - 99.8%406/40899.5%98.2% - 99.9%
Pre-selected78/8097.5%91.3% - 99.3%47/47100%92.4% - 100%
Combined175/17898.3%95.2% - 99.4%453/455⁸99.6%98.4% - 99.9%
Streptococcus agalactiae
Prospective21/21100%84.5% - 100%485/485100%99.2% - 100%
Pre-selected17/17100%81.6% - 100%110/110100%96.6% - 100%
Combined38/38100%90.8% - 100%595/595100%99.4% - 100%

Page 42

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 37 of 43

Pathogen TargetSensitivity/PPASpecificity/NPA
TP / (TP+FN)Sensitivity /PPA (%)95% CITN / (TN+FP)Specificity/NPA (%)95% CI
Streptococcus anginosus group
Prospective8/988.9%56.5% - 98%496/49799.8%98.9% - 100%
Pre-selected25/2696.2%81.1% - 99.3%101/101100%96.3% - 100%
Combined33/3594.3%81.4% - 98.4%597/598⁹99.8%99.1% - 100%
Streptococcus pneumoniae
Prospective11/11100%74.1% - 100%494/49599.8%98.9% - 100%
Pre-selected21/21100%84.5% - 100%106/106100%96.5% - 100%
Combined32/32100%89.3% - 100%600/60199.8%99.1% - 100%
Streptococcus pyogenes
Prospective21/21100%84.5% - 100%485/485100%99.2% - 100%
Pre-selected16/16100%80.6% - 100%111/111100%96.7% - 100%
Combined37/37100%90.6% - 100%596/596100%99.4% - 100%
Resistance Marker Genes¹⁰
mecA/mecC
Prospective155/15798.7%95.5% - 99.6%154/16195.7%91.3% - 97.9%
Pre-selected3/3100%43.9% - 100%18/18100%82.4% - 100%
Combined158/16098.8%95.6% - 99.7%172/17996.1%92.1% - 98.1%
vanA
Prospective8/8100%67.6% - 100%47/47100%92.4% - 100%
Pre-selected22/22100%85.1% - 100%1/1100%20.7% - 100%
Combined30/30100%88.6% - 100%48/48100%92.6% - 100%
vanB
Prospective1/1100%20.7% - 100%54/54100%93.4% - 100%
Pre-selected0/0NANA23/23100%85.7% - 100%

¹The one Bacillus spp. FP was positive by Standard of Care MALDI-ToF assay.
²Out of two Enterococcus faecium FPs, one was positive by Standard of Care molecular and biochemical assays.
³Out of six Staphylococcus spp. FNs, one was negative by BDS and one was negative by Standard of Care MALDI-ToF assay.
⁴Out of two Staphylococcus spp. FPs, one was positive by BDS and one was positive by Standard of Care MALDI-ToF and molecular assays
⁵The one Staphylococcus aureus FP was positive by BDS and Standard of Care molecular assay.
⁶Out of three Staphylococcus epidermidis FNs, one was negative by BDS.
⁷Out of six Staphylococcus epidermidis FPs, three were positive by both BDS and Standard of Care MALDI-ToF assay, one was positive by both BDS and Standard of Care molecular assay and one was positive by Standard of Care MALDI-ToF assay and not tested by BDS.
⁸Out of two Streptococcus spp. FPs, one was positive by BDS and one was positive by Standard of Care molecular assay.
⁹The one Streptococcus anginosus group FP was positive by both BDS and Standard of Care molecular assay.
¹⁰The LIAISON PLEX BCP assay will report the presence or absence of resistance markers only if an applicable organism is also detected, therefore the total number of evaluable samples for each resistance marker is dependent on the number of applicable organisms enrolled.

Page 43

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 38 of 43

Pathogen TargetSensitivity/PPASpecificity/NPA
TP / (TP+FN)Sensitivity /PPA (%)95% CITN / (TN+FP)Specificity/NPA (%)95% CI
Combined1/1100%20.7% - 100%77/77100%95.3% - 100%

Contrived Specimen Testing

For low-prevalence targets that did not reach the minimum number of positives from the combined prospective and pre-selected populations, a total of 225 specimens were contrived and tested as part of Arm 3. To minimize bias, contrived specimens were blinded, randomized, and tested along with negative clinical specimens at three external testing sites and one internal testing site between June 2024 to September 2024. Results from contrived specimens were analyzed separately from the prospective and pre-selected data sets.

Out of the 225 specimens included in the contrived study analysis, 211 specimens (93.8%) generated valid LIAISON PLEX BCP Assay results (i.e., Detected or Not Detected) on the first attempt. There were 14 specimens with an invalid result on the initial run. Of the 14 specimens retested, all 14 specimens generated a valid result after a single retest for a final testing success rate of 100% (225/225).

A summary of the contrived specimen set is provided in Table 21.

Table 21: Summary of Contrived Specimens

OrganismResistance Marker(s)Strain# of Independent Specimens Tested
Bacillus amyloliquefaciensNAATCC 2335010
Bacillus atrophaeusNAATCC 645510
Bacillus cereusNAATCC 1177810
Bacillus licheniformisNAATCC 1458010
Bacillus thuringiensisNAATCC 1079210
Bacillus spp. Total50
Enterococcus faecalisvanBATCC 70080210
ATCC 5129910
6418826210
Enterococcus faecalis Total30
Enterococcus faeciumvanBATCC 5185810
JMI CS-71210
Enterococcus faecium Total20
Listeria grayiNAATCC 2540110
Listeria innocuaNAATCC 3309010
Listeria ivanoviiNAATCC 1911910
Listeria monocytogenesNAATCC 191148
Listeria welshimeriNAATCC 3589711
Listeria spp. Total49
Staphylococcus lugdunensisNAATCC 8449746250
Staphylococcus lugdunensis Total50

Page 44

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 39 of 43

Clinical Performance (Sensitivity, Specificity, and 95% confidence interval) of the LIAISON PLEX® BCP Assay compared to the reference method is summarized in Table 22 for contrived specimens.

Table 22: Sensitivity/PPA and Specificity/NPA of Contrived Data Set – Arm 3

Pathogen TargetSensitivity/PPASpecificity/NPA
AnalyteTP / (TP+FN)Sensitivity/PPA (%)95% CITN /(TN+FP)Specificity/NPA (%)95% CI
Bacteria
Bacillus spp.50/50100%92.9% - 100%175/175100%97.9% - 100%
Enterococcus faecalis30/30100%88.6% - 100%195/195100%98.1% - 100%
Enterococcus faecium20/20100%83.9% - 100%205/205100%98.2% - 100%
Listeria spp.49/49100%92.7% - 100%176/176100%97.9% - 100%
Staphylococcus spp.50/50100%92.9% - 100%175/175100%97.9% - 100%
Staphylococcus aureus0/0NANA225/225100%98.3% - 100%
Staphylococcus epidermidis0/0NANA225/225100%98.3% - 100%
Staphylococcus lugdunensis50/50100%92.9% - 100%175/175100%97.9% - 100%
Streptococcus spp.0/0NANA225/225100%98.3% - 100%
Streptococcus agalactiae0/0NANA225/225100%98.3% - 100%
Streptococcus anginosus group0/0NANA225/225100%98.3% - 100%
Streptococcus pneumoniae0/0NANA225/225100%98.3% - 100%
Streptococcus pyogenes0/0NANA225/225100%98.3% - 100%
Resistance Marker Genes¹
mecA/mecC0/0NANA50/50100%92.9% - 100%
vanA0/0NANA50/50100%92.9% - 100%
vanB50/50100%92.9% - 100%0/0NANA

¹The LIAISON PLEX BCP Assay will report the presence or absence of resistance markers only if an applicable organism is also detected. Therefore the total number of evaluable samples for each resistance marker is dependent on the number of applicable organisms enrolled.

LIAISON PLEX BCP Assay Performance for Genus-Level Targets by Species

The LIAISON PLEX BCP Assay reports genus-level results for Bacillus spp., Staphylococcus spp. and Streptococcus spp. Sensitivity/PPA for each of these target organisms stratified at the species level is presented in Table 23 below. Specimens with reference results of low discrimination between two species of the same genus that were not further resolved are listed as "spp. (unknown)".

Page 45

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 40 of 43

Table 23: Sensitivity/PPA of Genus-level Targets, Stratified by Species

OrganismProspective (Arm 1)Pre-selected (Arm 2)Contrived (Arm 3)
Sensitivity/PPA95% CISensitivity/PPA95% CISensitivity/PPA95% CI
Bacillus spp.100% (1/1)20.7% - 100%89.5% (17/19)68.6% - 97.1%100% (50/50)92.9% - 100%
Bacillus amyloliquefaciensNANANANA100% (10/10)72.2% - 100%
Bacillus atrophaeusNANANANA100% (10/10)72.2% - 100%
Bacillus cereusNANANANA100% (10/10)72.2% - 100%
Bacillus licheniformisNANANANA100% (10/10)72.2% - 100%
Bacillus thuringiensisNANANANA100% (10/10)72.2% - 100%
Listeria spp.NANA100% (5/5)56.6% - 100%100% (49/49)92.7% - 100%
Listeria grayiNANANANA100% (10/10)72.2% - 100%
Listeria innocuaNANANANA100% (10/10)72.2% - 100%
Listeria ivanoviiNANANANA100% (10/10)72.2% - 100%
Listeria monocytogenesNANA100% (5/5)56.6% - 100%100% (8/8)67.6% - 100%
Listeria welshimeriNANANANA100% (11/11)74.1% - 100%
Staphylococcus spp.98.1% (316/322)96% - 99.1%100% (20/20)83.9% - 100%100% (50/50)92.9% - 100%
Staphylococcus aureus99.4% (160/161)96.6% - 99.9%NANANANA
Staphylococcus epidermidis96.9% (93/96)91.2% - 98.9%NANANANA
Staphylococcus lugdunensis100% (6/6)61% - 100%100% (20/20)83.9% - 100%100% (50/50)92.9% - 100%
Staphylococcus arlettae100% (1/1)20.7% - 100%NANANANA
Staphylococcus auricularis66.7% (2/3)20.8% - 93.9%NANANANA
Staphylococcus capitis100% (9/9)70.1% - 100%NANANANA
Staphylococcus caprae100% (1/1)20.7% - 100%NANANANA
Staphylococcus haemolyticus100% (6/6)61% - 100%NANANANA
Staphylococcus hominis96.3% (31/32)84.3% - 99.4%NANANANA

Page 46

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 41 of 43

OrganismProspective (Arm 1)Pre-selected (Arm 2)Contrived (Arm 3)
Sensitivity/PPA95% CISensitivity/PPA95% CISensitivity/PPA95% CI
Staphylococcus pseudintermedius100% (1/1)20.7% - 100%NANANANA
Staphylococcus saccharolyticus0% (0/1)0% - 79.3%NANANANA
Staphylococcus saprophyticus100% (1/1)20.7% - 100%NANANANA
Staphylococcus simulans100% (1/1)20.7% - 100%NANANANA
Staphylococcus vitulinus0% (0/1)0% - 79.3%NANANANA
Staphylococcus warneri100% (3/3)43.9% - 100%NANANANA
Staphylococcus spp. Unknown87.5% (7/8)52.9% - 97.8%NANANANA
Streptococcus spp.99% (97/98)94.4% - 99.8%97.5% (78/80)¹91.3% - 99.3%NANA
Streptococcus agalactiae100% (21/21)84.5% - 100%100% (17/17)81.6% - 100%NANA
Streptococcus anginosus88.9% (8/9)56.5% - 98%96.2% (25/26)81.1% - 99.3%NANA
Streptococcus pneumoniae100% (11/11)74.1% - 100%100% (21/21)84.5% - 100%NANA
Streptococcus pyogenes100% (21/21)84.5% - 100%100% (16/16)80.6% - 100%NANA
Streptococcus dysgalactiae100% (15/15)79.6% - 100%NANANANA
Streptococcus gallolyticus100% (1/1)20.7% - 100%NANANANA
Streptococcus gordonii100% (4/4)51% - 100%NANANANA
Streptococcus parasanguinis100% (1/1)20.7% - 100%NANANANA
Streptococcus salivarius100% (2/2)34.2% - 100%NANANANA
Streptococcus sanguinis100% (2/2)34.2% - 100%NANANANA
Streptococcus spp. Unknown100% (11/11)74.1% - 100%NANANANA

¹For one specimen the assay returned a true positive call for Streptococcus anginosus without detecting Streptococcus spp. As a result, there are two total FNs for the Streptococcus spp. target, but only one listed when stratified by species.

The clinical performance for each resistance marker stratified by eligible organism (identified by the reference method for clinical specimens or the inoculated organism for contrived specimens) is listed in Tables 24-26.

Page 47

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 42 of 43

Table 24: Performance of mecA/mecC, Stratified by Organism

OrganismProspective (Arm 1)Pre-selected (Arm 2)Contrived (Arm 3)
Sensitivity/PPA95% CISensitivity/PPA95% CISensitivity/PPA95% CI
Staphylococcus aureus100% (64/64)94.3% - 100%NANANANA
Staphylococcus epidermidis97% (65/67)89.8% - 99.2%NANANANA
Staphylococcus lugdunensis100% (4/4)51% - 100%100% (3/3)43.8% - 100%NANA
Staphylococcus capitis100% (2/2)34.2% - 100%NANANANA
Staphylococcus haemolyticus100% (6/6)61.0% - 100%NANANANA
Staphylococcus hominis100% (12/12)75.7% – 100%NANANANA
Spp. unknown100% (2/2)34.2% - 100%NANANANA
Overall98.70% (155/157)95.4% - 99.6%100% (3/3)43.8% - 100%NANA

Table 25: Performance of vanA, Stratified by Organism

OrganismProspective (Arm 1)Pre-selected (Arm 2)Contrived (Arm 3)
Sensitivity/PPA95% CISensitivity/PPA95% CISensitivity/PPA95% CI
Enterococcus faecalisNANA100% (2/2)34.2% - 100%NANA
Enterococcus faecium100% (8/8)67.6% - 100%100% (20/20)83.9% - 100%NANA
Overall100% (8/8)67.6% - 100%100% (22/22)85.1% - 100%NANA

Table 26: Performance of vanB, Stratified by Organism

OrganismProspective (Arm 1)Pre-selected (Arm 2)Contrived (Arm 3)
Sensitivity/PPA95% CISensitivity/PPA95% CISensitivity/PPA95% CI
Enterococcus faecalisNANANANA100% (30/30)88.6% - 100%
Enterococcus faecium100% (1/1)20.7% - 100%NANA100% (20/20)83.9% - 100%
Overall100% (1/1)20.7% - 100%NANA100% (50/50)92.9% - 100%

Page 48

LIAISON PLEX® Gram-Positive Blood Culture Assay Traditional 510(k)

Confidential & Restricted 510(k) Summary Page 43 of 43

N. Proposed Labeling:

The labeling provided in the submission satisfies the requirements of 21 CFR 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

§ 866.3365 Multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures.

(a)
Identification. A multiplex nucleic acid assay for identification of microorganisms and resistance markers from positive blood cultures is a qualitative in vitro device intended to simultaneously detect and identify microorganism nucleic acids from blood cultures that test positive by Gram stain or other microbiological stains. The device detects specific nucleic acid sequences for microorganism identification as well as for antimicrobial resistance. This device aids in the diagnosis of bloodstream infections when used in conjunction with other clinical and laboratory findings. However, the device does not replace traditional methods for culture and susceptibility testing.(b)
Classification. Class II (special controls). The special control for this device is FDA's guideline document entitled “Class II Special Controls Guideline: Multiplex Nucleic Acid Assay for Identification of Microorganisms and Resistance Markers from Positive Blood Cultures.” For availability of the guideline document, see § 866.1(e).