K Number
K243470

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-12-03

(25 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Fix2Lock is intended use for fixation of soft tissue to bone, using suture, in the following procedure; Orthopedic surgery for shoulders, knees, foot/ankle, hand/wrist and elbow.

Device Description

This product is used for orthopedic surgery, which soft tissue fix such as ligaments, tendons, and capsules to bone. The implant part is made of PEEK (polyether ether ketone, ASTM F2026) with non-absorbable suture and consists of Driver Handle and Driver Shaft for anchor insertion. This product is sterilized product and single use only.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device called "Fix2Lock (PEEK Self Punching)". This submission aims to demonstrate substantial equivalence to a legally marketed predicate device.

The information provided does not pertain to an AI/ML device or its performance criteria, as indicated by the nature of the acceptance criteria and the listed tests. The device is a physical orthopedic implant. Therefore, I cannot extract information related to AI/ML specific criteria such as sensitivity/specificity, AUC, reader studies, ground truth establishment for AI, or training/test set sample sizes for an AI model.

The "acceptance criteria" discussed in the document are primarily related to physical and material properties, sterilization, packaging, and shelf life of the bone fixation fastener. The "study" proving these criteria refers to bench tests conducted to verify the device meets all design specifications and is substantially equivalent to the predicate device.

Here's a summary of the information that can be extracted based on the provided text, using the closest applicable interpretations:

1. A table of acceptance criteria and the reported device performance

Since this is not an AI/ML device, the "acceptance criteria" are compliance with relevant standards for medical devices and meeting design specifications. The "reported device performance" is a statement of compliance with these standards and equivalency to the predicate.

Acceptance Criteria CategorySpecific Criteria (Relevant Standards)Reported Device Performance
Material- ASTM F2026: 2017 (PEEK polymers for surgical implant applications) - ASTM F2848: 2021 (Medical-Grade Ultra-High Molecular Weight Polyethylene Yarns and Surgical Implants)The subject device utilizes PEEK material, and bench tests demonstrated compliance with these standards.
Mechanical Performance- ASTM F543: 2017 (Metallic medical bone screws)Bench tests demonstrated compliance with this standard.
Sterilization- ISO 11135:2014 (Ethylene oxide sterilization methods) - ISO 11138-1:2017, ISO 11138-2:2017 (Biological indicators for sterilization) - ISO 11140-1:2014 (Chemical indicators)Bench tests demonstrated compliance with these standards, ensuring the device is sterile via EtO sterilization.
Microbiological- ISO 11737-1:2018 [AMD1:2021] (Determination of microorganism population) - ISO 11737-2:2019 (Tests of sterility)Bench tests demonstrated compliance with these standards.
Packaging & Shelf Life- ISO 11607-1:2019, ISO 11607-2:2019 (Packaging for terminally sterilized medical devices) - ASTM F1980:2016 (Accelerated Aging of Sterile Barrier Systems) - ASTM F88/F88M-21 (Seal strength) - ASTM F1929:2015 (Detecting Seal Leaks)Bench tests demonstrated compliance with these standards, confirming packaging integrity and a 3-year shelf life.
Bacterial Endotoxin- USP <85> (Bacterial Endotoxin Test) - USP <161> (Medical Devices-Bacterial Endotoxin and Pyrogen Tests) - ISO 11737-3:2023 (Bacterial endotoxin testing)Bench tests demonstrated compliance with these standards.
Substantial EquivalenceDevice must be substantially equivalent to the predicate device in terms of intended use, technological characteristics, and safety and effectiveness.The subject device (Fix2Lock PEEK Self Punching) is determined to be Substantially Equivalent (SE) to the predicate device (K231322) in respect of safety and effectiveness.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document mentions "bench tests" were conducted. It does not specify sample sizes for these tests, nor does it provide details on data provenance beyond the product's manufacturer (Osteonic Co., Ltd., South Korea) and the standards used. These are typical engineering tests, not clinical studies.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This is not an AI/ML device involving expert "ground truth" for diagnostic or analytical tasks. The "ground truth" here is compliance with engineering and material standards, typically verified by laboratory testing and technical experts/engineers according to standard protocols.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This concept applies to expert review processes, typically in clinical studies or AI ground truth establishment. The "adjudication" for this device would be the interpretation of bench test results against established standard specifications.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-assisted device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device is established by compliance with internationally recognized engineering and medical device standards (e.g., ISO, ASTM, USP) for materials, mechanical performance, sterilization, packaging, and biological safety.

8. The sample size for the training set

Not applicable. This is not an AI/ML device.

9. How the ground truth for the training set was established

Not applicable. This is not an AI/ML device.

{0}------------------------------------------------

December 3, 2024

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Osteonic Co., Ltd. % Lee Sanglok Manager Wise Company Inc. #1107, #1108, 204 Gasandigital 1-ro, Geumcheon-gu Seoul. 08502 Korea, South

Re: K243470

Trade/Device Name: Fix2Lock (PEEK Self Punching) Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: November 8, 2024 Received: November 8, 2024

Dear Lee Sanglok:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

CHRISTOPHER FERREIRA -S

Christopher Ferreira, MS Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K243470

Device Name

Fix2Lock (PEEK Self Punching)

Indications for Use (Describe)

Fix2Lock is intended use for fixation of soft tissue to bone, using suture, in the following procedure; Orthopedic surgery for shoulders, knees, foot/ankle, hand/wrist and elbow.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Date of Submission: 2024.11.08

01. Applicant

OSTEONIC Co., Ltd.

405Ho, 505-2Ho, 505-3Ho, 508Ho, 603Ho, 902Ho, 1004Ho, 1005Ho, 1202Ho, 1206Ho, 38 Digital-ro 29-gil, Guro-gu, Seoul, Korea, 102Ho, 103Ho, 24, Digital-ro 27-gil, Guro-gu, Seoul, Korea, 1012Ho, 1013Ho, 272, Digital-ro, Guro-gu, Seoul, Korea TEL: +82-2-6902-8442 FAX: +82-2-6902-8401 Email: dhkim@osteonic.com

02. Submission Correspondent

Sanglok, Lee Wise COMPANY Inc. 1107Ho, 1108Ho, 204 Gasan digital 1-ro, Geumcheon-gu, Seoul, Korea TEL: +82 70 8812 3619 / +82 2 831 3615 FAX: +82 50 4031 3619 Email: info@wisecompany.org

03. Subject Device Identification

Trade Name: Fix2Lock (PEEK Self Punching) Common Name: fastener, fixation, non-biodegradable, soft tissue Classification Name: Smooth or threaded metallic bone fixation fastener Classification Product Code: MBI Panel: Orthopedic Regulation Number: 21 CFR 888.3040 Device Class: II

04. Predicate Device

510(k) Number: K231322 Device Name: Fix2Lock (PEEK Self Punching) Manufacturer: Osteonic Co., Ltd.

05. Reference Device

510(k) Number: K233893 Device Name: AlphaVent Knotless SP PEEK Anchor Manufacturer: Stryker Endoscopy

06. Device Description

This product is used for orthopedic surgery, which soft tissue fix such as ligaments, tendons, and capsules to bone. The implant part is made of PEEK (polyether ether ketone, ASTM F2026) with non-absorbable suture and consists of Driver Handle and Driver Shaft for anchor insertion. This product is sterilized product and single use only.

07. Indication for use

Fix2Lock is intended use for fixation of soft tissue to bone, using suture, in the following procedure; Orthopedic surgery for shoulders, knees, foot/ankle, hand/wrist and elbow.

{5}------------------------------------------------

08. Non-Clinical Test Conclusion

Bench tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device complies with the following standards:

▪ Material

  • ASTM F2026: 2017 Standard specification for polyetheretherketone(PEEK) polymers for surgical implant applications - ASTM F2848: 2021 Standard Specification for Medical-Grade Ultra-High Molecular Weight Polyethylene Yarns and Surgical Implants

· Mechanical performance

  • ASTM F543: 2017 Standard specification and test methods for metallic medical bone screws

▪ Sterilization, shelf life and packaging for sterile product

  • ISO 11135:2014, Sterilization of health-care products Ethylene oxide Requirements for the development validation and routine control of a sterilization and routine control of a sterilization process for medical devices
  • ISO 11138-1 Third edition 2017-03, Sterilization of health care products Biological indicators Part 1: General requirements
  • ISO 11138-2:2017, Sterilization of health care products Biological indicators Part 2: Biological indicators for ethylene oxide sterilization processes
  • ISO 11140-1:2014, Sterilization of health care products -- Chemical indicators -- Part 1: General requirements
  • ISO 11737-1 Third edition 2018-01 [Including AMD1:2021], Sterilization of health care products Microbiological methods - Part 1: Determination of a population of microorganisms on product [Including Amendment 1 (2021)]
  • ISO 11737-2 Third edition 2019-12, Sterilization of medical devices Microbiological methods Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process
  • ISO 11607-1 Second edition 2019-02, Packaging for terminally sterilized medical devices part 1: requirements for materials, sterile barrier systems and packaging system
  • ISO 11607-2 Second edition 2019-02, Packaging for terminally sterilized medical devices Part 2: Validation requirements for forming, sealing and assembly processes
  • ASTM F1980:2016 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
  • ASTM F88/F88M-21 Standard test method for seal strength of flexible barrier materials.
  • ASTM F1929:2015 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
  • Bacterial Endotoxin
  • USP <85> Bacterial Endotoxin Test
  • USP <161> Medical Devices-Bacterial Endotoxin and Pyrogen Tests
  • ISO 11737-3:2023, Sterilization of health care products Microbiological methods Part 3: Bacterial endotoxin testing

{6}------------------------------------------------

09. Comparison of Technological Similarities and Differences

Table 1: Substantial Equivalence Comparison

Modified(Subject) DeviceUnmodified(Predicate) DeviceREFERENCE Device
ProductNameFix2Lock (PEEK Self Punching),Fix2Lock (PEEK Self Punching),K231322AlphaVent Knotless SP PEEKAnchor, K233893EquivalenceDiscussion
Product codeMBIMBIMBISame
RegulatoryclassClass IIClass IIClass IISame
RegulationNumber21 CFR 888.304021 CFR 888.304021 CFR 888.3040Same
Intended useFix2Lock is intended use forfixation of soft tissue to bone, usingsuture, in the following procedure;Orthopedic surgery for shoulders,knees, foot/ankle, hand/wrist andelbow.Fix2Lock is intended use forfixation of soft tissue to bone, usingsuture, in the following procedure;Orthopedic surgery for shoulders,knees, foot/ankle, hand/wrist andelbow.The AlphaVent Knotless SP PEEKAnchor is intended to be used forsoft-tissue to bone fixation in theshoulder, foot and ankle, knee, handand wrist, elbow, and hip inskeletally mature pediatric and adultpatients. It is indicated for use in thefollowing procedures: Shoulder:Rotator Cuff Repair, BankartRepair, SLAP Lesion Repair, BicepsTenodesis, Acromio-clavicularSeparation Repair, Deltoid Repair,Capsular Shift or CapsulolabralReconstruction Foot/Ankle: LateralStabilization, Medial Stabilization,Achilles Tendon Repair, HalluxValgus Reconstruction, Mid-footReconstruction,MetatarsalLigament Repair/Tendon RepairKnee: Anterior Cruciate LigamentRepair, Medial Collateral LigamentRepair, Lateral Collateral LigamentRepair, Patellar Tendon Repair,Posterior Oblique Ligament Repair,Iliotibial Band Tenodesis,Quadriceps Tendon Repair,Secondary or Adjunct Fixation forACL/PCL Reconstruction or Repair,Meniscal Root Repair, MPFLRepair/Reconstruction Hand/Wrist:Scapholunate LigamentReconstruction, Ulnar or RadialCollateral Ligament ReconstructionElbow: Biceps TendonReattachment, Ulnar or RadialCollateral Reconstruction, LateralEpicondylitis Repair Hip: CapsularRepair, Acetabular Labral Repair,Gluteus Medius Repair, ProximalHamstring RepairSame
OperatingPrinciplesThis product is used for orthopedicsurgery, which soft tissue fix suchas ligaments, tendons, and capsulesto bone.This product is used for orthopedicsurgery, which soft tissue fix suchas ligaments, tendons, and capsulesto bone.This product is used for orthopedicsurgery, which soft tissue fix such asligaments, tendons, and capsules tobone.Same
MaterialPEEK anchor and non-absorbablesuturesPEEK anchor and non-absorbablesutures, needlePEEK anchor and non-absorbablesuturesSimilar
StructureThis product consists of animplanted anchor, non-absorbablesuture, driver shaft and handleThis product consists of animplanted anchor, non-absorbablesuture, driver shaft and handleThis product consists of animplanted anchor, non-absorbablesuture, driver shaft and handleSame
ProductSize3.75mm ~ 6.5mm2.3mm ~ 5.5mm4.75mm, 5.5mm, 6.5mmSimilar
SterilizationSterile (EtO sterilization)Sterile (EtO sterilization)Sterile (EtO sterilization)Same
Single Use/ReuseSingle useSingle useSingle useSame
Packaging1 EA / BOX1 EA / BOX-Same
Shelf life3Years3Years-Same

{7}------------------------------------------------

The Product Size of subject device is similar with the predicate device and the safety was evaluated as the performance bench test.

Reference device had been added to demonstrate the equivalency with the marketed device. This is to resolve that the product size (6.5mm) is larger than the product size range of unmodified device.

10. Substantially Equivalent Conclusion

Based on above, the subject device, Fix2Lock (PEEK Self Punching), is determined to be Substantially Equivalent (SE) to the Unmodified(Predicate) device, Fix2Lock (PEEK Self Punching) (K231322) ,in respect of safety and effectiveness.

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.