K Number
K243443

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-03-19

(133 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Digital X-ray detector, EXPD-N Series, is designed for use in digital imaging solutions for general radiographic diagnosis of human anatomy. This device is intended for use in all general diagnostic procedures, replacing film or screen-based radiographic systems for both adult and paediatric patients. It is not intended for use in mammography.

Device Description

In comparison to existing devices, the new detectors incorporate a Flexible a-Si in the TFT material within the panel. The primary difference from the conventional glass a-Si panel is that the electronic circuits, such as silicon, are deposited on a plastic substrate instead of a glass substrate during the manufacturing of the TFT panel. Since only the material of the substrate on which the silicon is deposited changes, the overall image performance remains unaffected. Another difference is the pixel pitch. While existing products feature only a pixel pitch of 140μm, the new models include an option with a pixel pitch of 100um. The resolution of an X-ray detector has a significant impact on MTF (Modulation Transfer Function) and sensitivity.

AI/ML Overview

This medical device submission is for an X-ray detector, not an AI/ML device. Therefore, the typical acceptance criteria and study requirements for AI/ML devices, such as those related to multi-reader multi-case studies, standalone performance, and ground truth establishment with expert consensus or pathology, are not applicable here.

The submission focuses on establishing substantial equivalence to a predicate device based on technical characteristics and physical performance, confirming it is suitable for general radiographic diagnosis.

Here's a breakdown of the provided information, tailored to the context of a non-AI X-ray detector:

1. Table of Acceptance Criteria and the Reported Device Performance

The acceptance criteria are implicitly defined by demonstrating substantial equivalence to the predicate device in terms of technical characteristics and performance metrics relevant to X-ray image quality. The table below compares the subject device's performance to the predicate device, highlighting where performance is similar or improved.

ItemAcceptance Criteria (Implied by Predicate Device K193017 Performance)Subject Device (EXPD-N Series) Reported Performance
Intended UseGeneral radiographic diagnosis, replaces film/screen-based systems, adult & pediatric, Not for mammography.General radiographic diagnosis, replaces film/screen-based systems, adult & pediatric, Not for mammography. (Same)
Anatomical SitesGeneral RadiographyGeneral Radiography (Same)
Dimensions (mm)EVS 3643W/WG/WP: 460(W) x 386(L) x 15(H)EVS 4343W/WG/WP: 460(W) x 460(L) x 15(H)EXPD 3643N/NP/NU/N1/U1: 460(W) x 386(L) x 15.5(H)EXPD 4343N/NP/NU/N1/U1: 460(W) x 460(L) x 15.5(H) (Slight difference in thickness, otherwise similar)
Pixel Pitch140 μm140 μm (for N/NP/NU models)100 μm (for N1/U1 models) (Improved resolution option added)
Image Size (pixels)EVS 3643W/WG/WP: 2,560 x 3,072EVS 4343W/WG/WP: 3,072 x 3,072EXPD 4343N/NP/NU: 3,072 × 3,072 (Same)EXPD 3643N/NP/NU: 2,560 × 3,072 (Same)EXPD 4343N1/U1: 4,302 × 4,302 (Improved with 100μm pixel pitch)EXPD 3643N1/U1: 3,534 × 4,302 (Improved with 100μm pixel pitch)
Active Area (mm)EVS 3643W/WG/WP: 430 x 358EVS 4343W/WG/WP: 430 x 430EXPD 4343N/NP/NU: 430.2mm × 430.2mm (Similar)EXPD 3643N/NP/NU: 353.4mm × 430.2mm (Similar)EXPD 4343N1/U1: 430.08mm × 430.08mm (Similar, adapted for 100μm pixel pitch)EXPD 3643N1/U1: 358.4mm × 430.08mm (Similar, adapted for 100μm pixel pitch)
TFT Materiala-Si, IGZOa-Si, Flexible a-Si, IGZO (New Flexible a-Si material introduced, otherwise similar)
Cycle Time< 8.5 sec.< 8.5 sec. (Same)
Power Supply100240V, 50/60 Hz100240V, 50/60 Hz (Same)
Dynamic Range16 bits16 bits (Same)
MTFEVS 4343W: 50.0% at 2.0 lp/mmEVS 3643W: 42.5% at 2.0 lp/mmetc.EXPD 4343N/NP/NU/3643N/NP/NU: 35% at 2.0 lp/mm (Comparable, some variations)EXPD 4343N1/U1/3643N1/U1: 40% at 2.0 lp/mm (Improved with 100μm pixel pitch)
DQEEVS 4343W: 52.8% at 1.0 lp/mmEVS 3643W: 53.3% at 1.0 lp/mmetc.EXPD 4343N/NP/NU/3643N/NP/NU: 60% at 0.5 lp/mm (Comparable, different spatial frequency reported)EXPD 4343N1/U1/3643N1/U1: 60% at 0.5 lp/mm (Comparable, different spatial frequency reported)
Resolution3.5 lp/mmEXPD 4343N/NP/NU/3643N/NP/NU: 3.5 lp/mm (Same)EXPD 4343N1/U1/3643N1/U1: 5 lp/mm (Improved with 100μm pixel pitch)
Communication MethodWire, Wireless IEEE 802.11a/g/n (2.4 GHz / 5 GHz)Wire, Wireless IEEE 802.11n/ac (2.4 GHz / 5 GHz) (Updated wireless standard)
Software (Image Processing program)EConsole 1 (K152172)EConsole 2 (K240243) (Newer software version)

Study Proving Acceptance Criteria:

The study proving the device meets the acceptance criteria is a non-clinical performance testing and expert review study.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size for Test Set: Not explicitly stated as a number of images. The document refers to "clinical images acquired from various anatomical sites."
  • Data Provenance: The document does not explicitly state the country of origin for the clinical images. It indicates the data was retrospective, as it refers to "clinical images acquired."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

  • Number of Experts: Not explicitly stated, but referred to as "qualified expert clinicians."
  • Qualifications of Experts: "qualified expert clinicians." (Specific specialties or years of experience are not provided.)

4. Adjudication Method for the Test Set:

  • Adjudication Method: "clinical images acquired from various anatomical sites were reviewed by qualified expert clinicians to assess imaging performance." No specific adjudication method like 2+1 or 3+1 is mentioned. It implies a qualitative assessment by experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, an MRMC comparative effectiveness study was not done. This device is an X-ray detector, not an AI-powered diagnostic tool, so a study on AI assistance for human readers is not relevant to this submission.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • No, a standalone (algorithm only) performance study was not done. This device is an X-ray detector; its performance is evaluated as a physical imaging device, not a standalone algorithm. The "performance" refers to image quality metrics like MTF, DQE, and resolution.

7. The Type of Ground Truth Used:

  • Type of Ground Truth: The ground truth for image quality assessment was established through expert qualitative assessment of clinical images. The stated goal was to demonstrate that "the device produces accurate X-ray images suitable for diagnostic purposes."

8. The Sample Size for the Training Set:

  • Sample Size for Training Set: Not applicable. This is a hardware device (X-ray detector) and its associated image processing software (EConsole 2). While the software itself might have been developed using some data, the submission does not detail a "training set" in the context of an AI/ML model for diagnostic interpretation. The submission indicates physical performance testing and expert review of acquired images.

9. How the Ground Truth for the Training Set Was Established:

  • How Ground Truth for Training Set Was Established: Not applicable, as this is not an AI/ML device with a defined "training set" requiring ground truth for model development. The focus is on the physical performance characteristics of the detector and the suitability of its output images for diagnostic purposes.

{0}------------------------------------------------

March 19, 2025

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

DRTECH Corporation % Juhee Lee Official Correspondent Suite No.1, 2 Floor / Suite No.2, 3 Floor, 29, Dunchon-daero 541beon-gil, Jungwon-gu, Seongnam-si Gyeonggi-do, 13216 SOUTH KOREA

Re: K243443

Trade/Device Name: Expd 4343n; Expd 4343n; Expd 4343nu; Expd 4343np; Expd 3643n1; Expd 3643n; Expd 3643u1; Expd 3643nu; Expd 3643np Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: Class II Product Code: MQB Dated: November 12, 2024 Received: February 19, 2025

Dear Juhee Lee:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed Gabriela for by Gabriela M. M. Rodal -S Rodal -S

Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)K243443
Device NameEXPD 4343N1;
EXPD 4343N;
EXPD 4343U1;
EXPD 4343NU;
EXPD 4343NP;
EXPD 3643N1;
EXPD 3643N;
EXPD 3643U1;
EXPD 3643NU;
EXPD 3643NP
Indications for Use (Describe)The Digital X-ray detector, EXPD-N Series, is designed for use in digital imaging solutions for general radiographic diagnosis of human anatomy. This device is intended for use in all general diagnostic procedures, replacing film or screen-based radiographic systems for both adult and paediatric patients. It is not intended for use in mammography.
Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

DRTECH Corporation Suite No. 1, 2 Floor/ Suite No. 2, 3 Floor, 29, Dunchon-Daero 541beon-gil, Jungwon-gu, Seognam-si, Gyeonggi-do, 13216, Republic of Korea

510(k) Summary (K243443)

These 510(k) summaries of safety and effectiveness information is prepared in accordance with 21 CFR 807.92.

1. Date Prepared [21 CFR 807.92(a) (1)]

March 19, 2025

.

.

2. Submitter's Information [21 CFR 807.92(a) (1)]

  • DRTECH Corporation ● Company Name:
    . Address: Suite No. 1, 2 Floor/ Suite No. 2, 3 Floor, 29, Dunchon-Daero 541beon-gil, Jungwon-gu, Seognamsi, Gyeonggi-do, 13216, Republic of Korea

  • . Contact Person: Juhee Lee

  • . + 82-31-779-7742 Phone:

  • . Fax: +82-31-779-7790

  • . Email: drtechra@drtech.com

3. Trade Name, Common Name, Classification [21 CFR 807.92(a) (2)|

EXPD 4343N1, EXPD 4343N, EXPD 4343U1,

  • EXPD 4343NU, EXPD 4343NP,
  • Trade/ Device Name: EXPD 3643N1, EXPD 3643N, EXPD 3643U1,
  • EXPD 3643NU, EXPD 3643NP
  • . Common Name: Digital Flat Panel X-ray Detector
  • . Classification Name: Stationary X-ray System
  • . Classification Panel: Radiology
  • Product Code: MOB
  • . Regulation Number: 21 CFR 892.1680
  • . Device Class: II

Identification of Predicate Device(s) [21 CFR 807.92(a) (3)] 4.

The identified predicate device(s) within this submission are as follows:

  • 510(k) Number: K193017
  • . Manufacturer: DRTECH Corporation
    • Common Name: Digital Flat Panel X-ray Detector

II

  • EVS 4343W, EVS 4343WG, EVS 4343WP, . Trade Name:
    • EVS 3643W, EVS 3643WG, EVS 3643WP
  • . Classification Name: Stationary X-ray System
  • . Classification Panel: Radiology
  • . Product Code: MOB
  • Regulation Number: 21 CFR 892.1680 ●
  • Device Class:

{5}------------------------------------------------

Description of the Device [21 CFR 807.92(a) (4)] 5.

In comparison to existing devices, the new detectors incorporate a Flexible a-Si in the TFT material within the panel. The primary difference from the conventional glass a-Si panel is that the electronic circuits, such as silicon, are deposited on a plastic substrate instead of a glass substrate during the manufacturing of the TFT panel. Since only the material of the substrate on which the silicon is deposited changes, the overall image performance remains unaffected.

Another difference is the pixel pitch. While existing products feature only a pixel pitch of 140μm, the new models include an option with a pixel pitch of 100um. The resolution of an X-ray detector has a significant impact on MTF (Modulation Transfer Function) and sensitivity.

Indications for Use [21 CFR 807.92(a) (5)] 6.

The Digital X-ray detector, EXPD-N Series, is designed for use in digital imaging solutions for general radiographic diagnosis of human anatomy. This device is intended for use in all general diagnostic procedures, replacing film or screen-based radiographic systems for both adult and paediatric patients.

It is not intended for use in mammography.

7. Technological Characteristics [21 CFR 807.92(a) (6)]

The fundamental operational principle of the subject device is the same as that of the predicate device. The EXPD-N series detector is a square plate-shaped indirect conversion device that converts incoming X-rays into visible light. This visible light is subsequently captured by an optical sensor, which produces an electric charge representation of the spatial distribution of the incoming X-ray quanta.

Through thin film transistors, the charges are transformed into a modulated electrical signal is amplified, then changed from an analog to digital form (from voltage to signal) so that it can be printed out, sent for remote viewing, or saved as an electronic data file for later viewing.

ItemSubject Device(EXPD 4343S)Predicate Device(EXPD 4343P)
510(k) No.K243443K193017
ManufacturerDRTECH CorporationDRTECH Corporation
Trade/ device NameEXPD 4343N1, EXPD 4343N,EXPD 4343U1, EXPD 4343NU,EXPD 4343NP,EXPD 3643N1, EXPD 3643N,EXPD 3643U1, EXPD 3643NU,EXPD 3643NPEVS 4343W, EVS 4343WG, EVS 4343WPEVS 3643W, EVS 3643WG, EVS 3643WP
Classification NameStationary X-ray SystemSame
ClassificationRegulation21 CFR 892.1680Same
Product CodeMQBSame
Device ClassClass IISame
ItemSubject Device(EXPD 4343S)Predicate Device(EXPD 4343P)
Intended UseThe Digital X-ray detector, EXPD-N Series,is designed for use in digital imagingsolutions for general radiographic diagnosisof human anatomy. This device is intendedfor use in all general diagnostic procedures,replacing film or screen-based radiographicsystems for both adult and paediatricpatients.It is not intended for use in mammography.EVS 4343W / EVS 4343WG / EVS 4343WP/ EVS 3643W / EVS 3643WG / EVS3643WP Digital X-ray detector is indicatedfor digital imaging solution designed forproviding general radiographic diagnosis ofhuman anatomy. This device is intended toreplace film or screen based radiographicsystems in all general purpose diagnosticprocedures. This device is not intended formammography applications.
Anatomical SitesGeneral RadiographySame
Dimensions(mm)- EXPD 3643N/ EXPD 3643NP/ EXPD 3643NU/ EXPD 3643N1/ EXPD 3643U1:460(W) x 386(L) x 15.5(H)- EXPD 4343N/ EXPD 4343NP/ EXPD 4343NU/ EXPD 4343N1/ EXPD 4343U1:460(W) x 460(L) x 15.5(H)mm- EVS 3643W/ EVS 3643WG/ EVS 3643WP:460(W) x 386(L) x 15(H)mm- EVS 4343W/ EVS 4343WG/ EVS 4343WP:460(W) x 460(L) x 15(H)mm
Pixel Pitch- EXPD 4343N / EXPD 4343NP /EXPD 4343NU / EXPD 3643N /EXPD 3643NP / EXPD 3643NU : 140 μm- EXPD 4343N1 / EXPD 4343U1 /EXPD 3643N1 / EXPD 3643U1 : 100 μm140μm
Image Size(pixels)- EXPD 4343N / EXPD 4343NP /EXPD 4343NU : 3,072 × 3,072- EXPD 3643N / EXPD 3643NP /EXPD 3643NU : 2,560 × 3,072- EXPD 4343N1 / EXPD 4343U1 : 4,302 × 4,302- EXPD 3643N1 / EXPD 3643U1 :3,534 × 4,302- EVS 3643W/ EVS 3643WG/ EVS 3643WP:2,560 x 3,072- EVS 4343W/ EVS 4343WG/ EVS 4343WP:3,072 x 3,072
Active Area(mm)- EXPD 4343N / EXPD 4343NP /EXPD 4343NU : 430.2mm × 430.2mm- EXPD 3643N / EXPD 3643NP /EXPD 3643NU : 353.4mm × 430.2mm- EXPD 4343N1 / EXPD 4343U1 :430.08mm × 430.08mm- EXPD 3643N1 / EXPD 3643U1 :358.4mm × 430.08mm- EVS 3643W/ EVS 3643WG/ EVS 3643WP:430mm x 358- EVS 4343W/ EVS 4343WG/ EVS 4343WP:430 x 430
TFT Material- EXPD 4343N1/ EXPD 4343N/ EXPD 3643 N1/EXPD 3643N: a-Si- EXDP 4343U1/ EXPD 4343NU/ EXPD 3643U1/ EXPD 3643 NU: Flexible a-Si- EXPD 4343NP/ EXPD 3643NP: IGZO- EVS 3643W/ EVS 4343W/EVS 3643WG/ EVS 4343WG: a-Si- EVS 4343WP/ EVS 3643WP: IGZO
Cycle Time< 8.5 sec.Same
Power Supply100240V, 50/60 HzSame
Dynamic Range16 bitsSame
MTF- EXPD 4343N/ EXPD 4343NP/ EXPD 4343NU/EXPD 3643N/ EXPD 3643NP/EXPD 3643NU: 35% *2.0lp/mm- EXPD 4343N1 / EXPD 4343U1 /EXPD 3643N1 / EXPD 3643U1: 40% *2.0lp/mm- EVS 4343W: 50.0% at 2.0 lp/mm- EVS 3643W: 42.5% at 2.0 lp/mm- EVS 4343WP: 48.4% at 2.0 lp/mm- EVS 3643WP: 42.9% at 2.0 lp/mm- EVS 4343WG: 50.1% at 2.0 lp/mm- EVS 3643WG: 47.8% at 2.0 lp/mm
DQE- EXPD 4343N /EXPD 4343NP /EXPD 4343NU/EXPD 3643N /EXPD 3643NP /EXPD 3643NU: 60% *0.5lp/mm- EXPD 4343N1 /EXPD 4343U1 /- EVS 4343W: 52.8% at 1.0 lp/mm- EVS 3643W: 53.3% at 1.0 lp/mm- EVS 4343WP: 50.0% at 1.0 lp/mm- EVS 3643WP: 53.1% at 1.0 lp/mm
ItemSubject Device(EXPD 4343S)Predicate Device(EXPD 4343P)
EXPD 3643N1 / EXPD 3643U1: 60% *0.5lp/mm- EVS 3643WG: 25.9% at 1.0 lp/mm
Resolution- EXPD 4343N / EXPD 4343NP /EXPD 4343NU / EXPD 3643N /EXPD 3643NP / EXPD 3643NU : 3.5 lp/mm- EXPD 4343N1 / EXPD 4343U1 / EXPD 3643N1/ EXPD 3643U1 : 5 lp/mm3.5 lp/mm
CommunicationMethodWire communicationSame
Wireless communication: IEEE 802.11n/ac (2.4 GHz / 5 GHz)Wireless communication: IEEE 802.11a/g/n (2.4 GHz / 5 GHz)
Software(Image Processingprogram)EConsole 2(K240243)EConsole 1(K152172)

8. Substantial Equivalence [21 CFR 807.92(b)]

{6}------------------------------------------------

DRTECH Corporation

Suite No. 1, 2 Floor/ Suite No. 2, 3 Floor, 29, Dunchon-Daero 541beon-gil, Jungwon-gu, Seognam-si, Gyeonggi-do, 13216, Republic of Korea

DRTECH

{7}------------------------------------------------

DRTECH Corporation

Suite No. 1, 2 Floor/ Suite No. 2, 3 Floor, 29, Dunchon-Daero 541beon-gil, Jungwon-gu, Seognam-si, Gyeonggi-do, 13216, Republic of Korea

The table above presented similarities(or identicality) between the subject device and predicate devices in:

  • Intended Use
  • Applied Anatomical Sites
  • Technological Characteristics and Operating Principle •
  • Communication Method •
  • . Size (Dimensions)
  • Power Supply
  • . Dynamic Range

Following parts that shown differences between two devices,

  • Image size .
  • Active area .
  • Pixel pitch and Resolution ●
  • . MTF and DQE
  • . TFT material
  • . Components

9. Summary of Non-Clinical Data [21 CFR 807.92(b)(1)]

Non-clinical performance testing in major physical values, such as MTF and DQE, performed in comparison with the predicate device for evaluation. Additionally, clinical images acquired from various anatomical sites were reviewed by qualified expert clinicians to assess imaging performance. The results demonstrated that the device produces accurate X-ray images suitable for diagnostic purposes.

Besides performance testing in comparison with predicate device, Electrical Safety testing and Electromagnetic compatibility testing was performed in accordance with IEC 60601-1 and IEC 60601-1-2, and results met the requirements.

10. Summary of Clinical Data [21 CFR 807.92(b)(2)]

The clinical studies are not required as there are no significant changes that requires assess of the performance made in the subject device, such as new indication, and non-clinical testing demonstrated equivalency to the predicate device sufficiently.

{8}------------------------------------------------

DRTECH

11. Conclusion [21 CFR 807.92(b)(3)]

The subject device, EXPD 4343N1, EXPD 4343N, EXPD 4343U1, EXPD 4343NU, EXPD 4343NP, EXPD 3643N1, EXPD 3643N, EXPD 3643U1, EXPD 3643NU, and EXPD 3643NP is substantially equivalent to the currently marketed predicate device in terms of technical characteristics, design features, operating principles, basis of functional and performance characteristics. Furthermore, the modification does not change to its intended use and indications for use.

N/A