(254 days)
Probe: The DxTx e²Coil probes are recommended for high-resolution magnetic resonance imaging, including spectroscopy, of the prostate and surrounding pelvic tissue in the male human anatomies and are specifically designed for use with the appropriate e²Coil Interface Device and Intermediate Cable Assembly. The one-time use, disposable e²Coil probe has been designed to be inserted into the human rectum. The e²Coil probe may be inserted by a trained healthcare professional, depending on state or country law, who has training in e²Coil probe insertion.
Interface Device: The Siemens 3.0T Endo Interface II with Tim 4G SlideConnect (e²Coil™) provides interface and support functions to allow the use of 3.0T disposable e²Coil probes with Siemens 3.0T MRI Systems with a Tim 4G connector. Only trained healthcare professionals are intended to operate this device.
The e²Coil™ Imaging System for the Siemens MAGNETOM 3.0T MRI Systems with Tim 4G Connector is a combination of the following two devices for improving the diagnostic MR image of the prostate and its surrounding pelvic tissues:
- Siemens 3.0T Endo Interface Gen II with Tim 4G Slide Connect (e2Coil)
- Endorectal MRI Coil II for 3.0T MR Imaging of the Prostate (e2Coil) – Siemens
- Coil contains two loops and four channels
The intended contact duration of the device is less than 60 minutes.
No components of this device contain medicinal substances, tissues or blood products.
The patient contacting device component materials are Latex, PVC, and Polymer.
The provided FDA 510(k) clearance letter and summary describe the acceptance criteria and the study that proves the device meets those criteria for the e²Coil™ Imaging System.
Here's an breakdown:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Criteria/Tests | Reported Device Performance |
|---|---|---|
| Biocompatibility | - Cytotoxicity | Complies with FDA Guidance 635 and ISO 10993-1:2018 |
| - Irritation | Complies with FDA Guidance 635 and ISO 10993-1:2018 | |
| - Sensitization | Complies with FDA Guidance 635 and ISO 10993-1:2018 | |
| Electrical Safety & EMC | - IEC 60601-1 Standard | Complies with IEC 60601-1 standard for EMC |
| Mechanical Testing | - Push | Complies with IEC 60601-1 standard for mechanical strength |
| - Impact | Complies with IEC 60601-1 standard for mechanical strength | |
| - Drop | Complies with IEC 60601-1 standard for mechanical strength | |
| - Moulding Stress Relief | Complies with IEC 60601-1 standard for mechanical strength | |
| MR Safety | - IEC 60601-2-33 | MR Conditional compliant to the 3.0T environment per IEC 60601-2-33 |
| Signal to Noise Ratio (SNR) | - Higher than scanner body coil alone (NEMA MS 1 2.3.2.4 Method 4) | Consistently higher than the SNR of the scanner body coil alone on the Siemens 3.0T Skyra scanner |
| Uniformity | - Good signal uniformity (NEMA MS 3 and IEC 62464-1 in axial, coronal, and sagittal axes) | Signal showed good uniformity per NEMA MS 3 and IEC 62464-1 in the axial, coronal, and sagittal axes |
2. Sample size used for the test set and the data provenance:
The document does not explicitly state the specific sample sizes for each test set (e.g., number of coils for mechanical testing, number of measurements for SNR). The nature of the device (an MRI coil) suggests that testing would involve a representative number of manufactured units rather than patient data.
The data provenance is internal testing performed by the manufacturer, DxTx Medical, Inc., for regulatory submission. There is no mention of external data or patient data provenance in this summarized document.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not provided in the document. The studies performed (biocompatibility, electrical safety, mechanical, MR safety, SNR, uniformity) are primarily objective engineering and scientific tests based on established standards (e.g., ISO, IEC, NEMA), rather than interpretations by medical experts to establish a "ground truth" in the diagnostic sense.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
This is not applicable to the types of performance tests conducted for this device. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies where expert consensus is needed to establish ground truth for diagnostic accuracy, which is not the focus of these engineering and safety tests.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This is not applicable. The e²Coil™ Imaging System is an MRI coil, a hardware component, not an AI-powered diagnostic software. Therefore, an MRMC study related to human readers improving with AI assistance would not be relevant to this device.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
This is not applicable, as the device is an MRI coil, not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The "ground truth" for the performance characteristics measured in this submission is based on established engineering and scientific standards and physical properties. For example:
- Biocompatibility: Adherence to ISO 10993-1:2018 and FDA Guidance 635.
- Electrical Safety & EMC: Compliance with IEC 60601-1.
- Mechanical Testing: Compliance with IEC 60601-1.
- MR Safety: Compliance with IEC 60601-2-33.
- SNR & Uniformity: Adherence to NEMA MS 1 and NEMA MS 3/IEC 62464-1 standards.
These are objective, measurable criteria, not subjective interpretations requiring expert consensus or clinical outcomes data in the usual sense.
8. The sample size for the training set:
Not applicable. This device is an MRI coil, not a machine learning model, so there is no "training set."
9. How the ground truth for the training set was established:
Not applicable, as there is no training set for this device.
FDA 510(k) Clearance Letter - K243428
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.08.00
July 17, 2025
DxTx Medical, Inc.
Paul Lawson
Director of Quality and Regulatory, R&D Manager
639 Alpha Dr
Pittsburgh, Pennsylvania 15238
Re: K243428
Trade/Device Name: e²Coil™ Imaging System for the Siemens MAGNETOM 3.0T MRI Systems with Tim 4G Connector
Regulation Number: 21 CFR 892.1000
Regulation Name: Magnetic Resonance Diagnostic Device
Regulatory Class: Class II
Product Code: MOS
Dated: June 11, 2025
Received: June 12, 2025
Dear Paul Lawson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Page 2
K243428 - Paul Lawson
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
Page 3
K243428 - Paul Lawson
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ningzhi Li -S Digitally signed by Ningzhi Li -S
for
Daniel M. Krainak, Ph.D.
Assistant Director
Magnetic Resonance and Nuclear Medicine Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Indications for Use
510(k) Number (if known): K243428
Device Name: e²Coil™ Imaging System for the Siemens MAGNETOM 3.0T MRI Systems with Tim 4G Connector
Indications for Use (Describe)
Probe: The DxTx e²Coil probes are recommended for high-resolution magnetic resonance imaging, including spectroscopy, of the prostate and surrounding pelvic tissue in the male human anatomies and are specifically designed for use with the appropriate e²Coil Interface Device and Intermediate Cable Assembly. The one-time use, disposable e²Coil probe has been designed to be inserted into the human rectum. The e²Coil probe may be inserted by a trained healthcare professional, depending on state or country law, who has training in e²Coil probe insertion.
Interface Device: The Siemens 3.0T Endo Interface II with Tim 4G SlideConnect (e²Coil™) provides interface and support functions to allow the use of 3.0T disposable e²Coil probes with Siemens 3.0T MRI Systems with a Tim 4G connector. Only trained healthcare professionals are intended to operate this device.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
Page 5
510(k) Summary
e²Coil™ Imaging System for the Siemens MAGNETOM 3.0T MRI Systems with Tim 4G Connector
I. Sponsor Information
DxTx Medical, Inc.
639 Alpha Drive
Pittsburgh, PA 15238
Contact Person: Paul Lawson
Phone: 412-435-3867
Email: paul.lawson@dxtxmedical.com
https://dxtxmedical.com/
Date of 510(k) Preparation: July 14, 2025
II. Subject Device:
Trade name: e²Coil™ Imaging System for the Siemens MAGNETOM 3.0T MRI Systems with Tim 4G Connector
Common Name: Prostate Endo Coil
Classification Name: Coil, Magnetic Resonance, Specialty
Regulation number: 21 CFR 892.1000
Product Code: MOS
Regulatory class: II
Review Panel: Radiology
III. Predicate Device
510(k): K051349
Trade name: 3.0T Prostate Imaging System
Common Name: Prostate Endo Coil
Classification Name: Coil, Magnetic Resonance, Specialty
Regulation number: 21 CFR 892.1000
Product Code: MOS
Regulatory class: II
Review Panel: Radiology
Page 6
IV. Device Description
The e²Coil™ Imaging System for the Siemens MAGNETOM 3.0T MRI Systems with Tim 4G Connector is a combination of the following two devices for improving the diagnostic MR image of the prostate and its surrounding pelvic tissues:
- Siemens 3.0T Endo Interface Gen II with Tim 4G Slide Connect (e2Coil)
- Endorectal MRI Coil II for 3.0T MR Imaging of the Prostate (e2Coil) – Siemens
- Coil contains two loops and four channels
The intended contact duration of the device is less than 60 minutes.
No components of this device contain medicinal substances, tissues or blood products.
The patient contacting device component materials are Latex, PVC, and Polymer.
V. Intended Use / Indications for Use
Probe: The DxTx e²Coil probes are recommended for high-resolution magnetic resonance imaging, including spectroscopy, of the prostate and surrounding pelvic tissue in the male human anatomies and are specifically designed for use with the appropriate e²Coil Interface Device and Intermediate Cable Assembly. The one-time use, disposable e²Coil probe has been designed to be inserted into the human rectum. The e²Coil probe may be inserted by a trained healthcare professional, depending on state or country law, who has training in e²Coil probe insertion.
Interface Device: The Siemens 3.0T Endo Interface II with Tim 4G SlideConnect (e2Coil™) provides interface and support functions to allow the use of 3.0T disposable e2Coil probes with Siemens 3.0T MRI Systems with a Tim 4G connector. Only trained healthcare professionals are intended to operate this device.
VI. Summary of Technological Characteristics Compared to Predicate Device
The intended use and principles of operation of the e²Coil™ Imaging System for the Siemens MAGNETOM 3.0T MRI Systems with Tim 4G Connector and the DxTx Medical (formerly MEDRAD/Bayer) 3.0T Prostate Imaging System predicate device are substantially equivalent. Both devices meet the regulatory definition for Coil, Magnetic Resonance, Specialty outlined in 21 CFR 892.1000.
Only trained healthcare professionals are intended to operate this device.
The technological differences between the e²Coil™ and its predicate eCoil raise no new issues of safety or image quality effectiveness.
Page 7
| Feature | Predicate Device | DxTx Medical Device System |
|---|---|---|
| Trade Name | 3.0T Prostate Imaging SystemModel Numbers: 08624855 (Interface Device), BPX 30 (Disposable Probe)510(k) Holder: DxTx Medical, Inc. (transferred from Bayer Healthcare in 2019)510(k) Number: K051349 | e²Coil™ Imaging System for the Siemens MAGNETOM 3.0T MRI Systems with Tim 4G ConnectorDevice Model Numbers - Trade Names in System:4000518 - Siemens 3.0T Endo Interface Gen II with Tim 4G Slide Connect (e2Coil)BPX 30 NG-S - Endorectal MRI Coil II for 3.0T MR Imaging of the Prostate (e2Coil) – Siemens |
| Indications for Use | The 3.0T Siemens Prostate Imaging System ("the device") is part of the Prostate Imaging Offerings from DxTx and is intended for use as a Magnetic Resonance Diagnostic Device (MRDD) for high-resolution magnetic resonance imaging, including spectroscopy, of the human prostate gland and surrounding pelvic tissue. The disposable eCoil Endorectal Coil (eCoil) is intended for one time use and is specifically designed to be used with an appropriate interface device. The 3.0T Siemens Prostate Imaging System is intended for use with Siemens 3.0T Tim Trio scanner platforms only. Only trained healthcare professionals are intended to operate this device. The 3.0T Siemens Prostate Imaging System is comprised of the Siemens Interface Device and the 3.0T eCoil (eCoil). | Probe: The DxTx e²Coil probes are recommended for high-resolution magnetic resonance imaging, including spectroscopy, of the prostate and surrounding pelvic tissue in the male human anatomies and are specifically designed for use with the appropriate e²Coil Interface Device and Intermediate Cable Assembly. The one-time use, disposable e²Coil probe has been designed to be inserted into the human rectum. The e²Coil probe may be inserted by a trained healthcare professional, depending on state or country law, who has training in e²Coil probe insertion.Interface Device: The Siemens 3.0T Endo Interface II with Tim 4G SlideConnect (e2Coil™) provides interface and support functions to allow the use of 3.0T disposable e2Coil probes with Siemens 3.0T MRI Systems with a Tim 4G connector. Only trained healthcare professionals are intended to operate this device. |
| Size | Interface Device with cable: 28 ¼" longBPX 30 disposable coil: 25" long | Interface Device with cable + Intermediate Cable 38 ¼"Disposable coil: 14 ½" |
| Technology Characteristics | Soft TipThin Wall BalloonSemi-flexible shaftDepth stopperFlexible syringe connector tubingOne coil (loop) in disposable coilOne channel | Soft TipThin Wall Balloon (same part as BPX 30)Semi-flexible shaft (same part as BPX 30)Depth stopper (same part as BPX 30)Flexible syringe connector tubing (same part as BPX 30)Two coils (loops) in disposable coilFour channels |
| Accessories | 50-60 ml Syringe | 50-60 ml SyringeIntermediate Cable |
| Safety | No product related hazards were reported in the FDA MAUDE database in the most recent Post Market | Equivalent low-risk safety design |
Page 8
| Feature | Predicate Device | DxTx Medical Device System |
|---|---|---|
| Surveillance Report 01-Oct-2019 to 30-Sep-2023Coil BiocompatibilityMR ≤ 3.0T Conditional | Coil is the same mechanically as BPX 30 so has the same biocompatibilityMR ≤ 3.0T ConditionalThe disposable coil includes a fuse as an additional level of electrical protection, which the predicate system does not. | |
| Effectiveness | Performance Tests:Scanner Validation on Siemens 3.0T Skyra MR scannerCoil Leakage, Tensile strength, Burst strength, Depth stopper resistance | Performance Tests:Scanner Validation on Siemens 3.0T Skyra MR scannerCoil is the same mechanically as BPX 30 so has the same mechanical performance |
| Material | Thin Wall Double Balloon: Natural Latex | Thin Wall Balloon: Same balloon |
| Energy Source | MR scanner | MR scanner |
VII. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
Biocompatibility testing
The biocompatibility evaluation for the Endorectal MRI Coil II for 3.0T MR Imaging of the Prostate (e2Coil) was conducted in accordance with FDA Guidance 635 "Immunotoxicity Testing Guidance" May 6, 1999 and International Standard ISO 10993-1:2018 "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" as recognized by the FDA. The battery of testing includes the following tests:
- Cytotoxicity
- Irritation
- Sensitization
Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the e²Coil™ Imaging System for the Siemens MAGNETOM 3.0T MRI, consisting of the Siemens 3.0T Endo Interface Gen II with Tim 4G Slide Connect and the Endorectal MRI Coil II for 3.0T MR Imaging of the Prostate. The system complies with the IEC 60601-1 standard for EMC.
Mechanical testing
Mechanical strength for portable equipment testing, consisting of the following listed tests, was conducted on the e²Coil™ Imaging System for the Siemens MAGNETOM 3.0T MRI, consisting of the Siemens 3.0T Endo Interface Gen II with Tim 4G Slide Connect and the Endorectal MRI Coil II for 3.0T MR Imaging of the Prostate:
- Push
- Impact
- Drop
- Moulding Stress Relief
Page 9
The system complies with the IEC 60601-1 standard for mechanical strength.
MR Safety testing
MR safety testing was conducted on the e²Coil™ Imaging System for the Siemens MAGNETOM 3.0T MRI, consisting of the Siemens 3.0T Endo Interface Gen II with Tim 4G Slide Connect and the Endorectal MRI Coil II for 3.0T MR Imaging of the Prostate. The system is MR Conditional compliant to the 3.0T environment per IEC 60601-2-33.
SNR testing
Signal to Noise Ratio (SNR) testing was conducted on the e²Coil™ Imaging System for the Siemens MAGNETOM 3.0T MRI, consisting of the Siemens 3.0T Endo Interface Gen II with Tim 4G Slide Connect and the Endorectal MRI Coil II for 3.0T MR Imaging of the Prostate.
The SNR using the e²Coil™ Imaging System was consistently higher than the SNR of the scanner body coil alone on the Siemens 3.0T Skyra scanner per NEMA MS 1 2.3.2.4 Method 4.
Uniformity testing
Signal uniformity testing was conducted on the e²Coil™ Imaging System for the Siemens MAGNETOM 3.0T MRI, consisting of the Siemens 3.0T Endo Interface Gen II with Tim 4G Slide Connect and the Endorectal MRI Coil II for 3.0T MR Imaging of the Prostate. The signal showed good uniformity per NEMA MS 3 and IEC 62464-1 in the axial, coronal, and sagittal axes.
VIII. Conclusion
The DxTx e²Coil is designed to have safety and effectiveness substantially equivalent to the predicate device. In particular, the subject device has the same or similar indications, technological characteristics, and principles of operation as the predicate device. The minor differences between the two devices do not raise any new issues of safety and effectiveness when the device is used as labeled.
N/A