AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Embryo Transfer Catheter is intended for introduction of in vitro fertilized embryo(s) into the uterine cavity.

The stylet is intended to be used with Vitrolife's Embryo Transfer Catheters to assist in uterine access of the guide during an embryo transfer procedure.

Device Description

Vitrolife´s Embryo Transfer Catheter (ETC) Assortment consists of three embryo transfer catheters (ETCs), one pre-curved variant (Model 17500, 230 mm) and two straight variants (REFs 17501, 17502) in different lengths (180 mm and 230 mm); and two stylets (Stylets) (REFs 17510, 17511) in different lengths (180 mm and 230 mm). ETC and Stylet are sterile, single-use devices used to deliver in vitro fertilized embryos to the uterine cavity.

Embryo Transfer Catheter consists of:

  • A guide, also referred as outer sheath. The guide is used to navigate through the cervix canal. The guide has a distance marking, a stopper and a rounded tip to facilitate proper positioning. The guide is available in two different variants:
    • Pre-curved - stiff
    • Straight - soft and malleable
  • A catheter, also referred as inner catheter. The catheter is loaded with the embryo(s) in a small volume of transfer medium and inserted through the guide to gently deposit the embryo(s) into the uterine cavity. The transfer catheter is soft, flexible and approximately 50 mm longer than the guide. The tip of the catheter has an echogenic marking to enable ultrasound guidance.

Embryo Transfer Catheter Stylet consists of:

  • A plastic-coated metal wire, which is malleable. The stylet is an accessory that can be used together with Vitrolife's guides to make them stiffer.

The 17500 variant includes a stylet. The 17501 and 17502 variants do not include stylets, but the stylet can be purchased separately.

AI/ML Overview

The FDA Clearance Letter for Vitrolife Sweden AB's Embryo Transfer Catheter (ETC) Assortment primarily describes the non-clinical performance testing conducted to demonstrate substantial equivalence to a predicate device. This letter does not describe an AI/ML-based device, nor does it detail studies involving human experts, ground truth adjudication, or MRMC studies typically associated with AI/ML medical devices.

Therefore, many of the requested details regarding acceptance criteria and study information (particularly points 2-9 related to AI/ML device testing) cannot be extracted from the provided text.

However, I can provide the acceptance criteria and reported performance for the non-clinical tests that were performed, as outlined in the document.

Summary of Device Acceptance Criteria and Performance (Based on Non-Clinical Testing):

The acceptance criteria for this device are based on its physical properties, sterility, biocompatibility, and functionality, demonstrating its substantial equivalence to an existing (predicate) device. The studies described are primarily bench performance tests, material tests, and sterilization/packaging validations.

Acceptance Criteria CategorySpecific TestAcceptance CriteriaReported Device Performance
SterilizationSterilization Validation (E-Beam)Compliance with ISO 11137-1:2006/(R)2015 and ISO 11137-2:2013Performed according to standards. (Implied compliance for clearance.)
Packaging IntegrityVisual InspectionCompliance with ASTM F1886/F1886M-16Met predetermined acceptance criteria.
Dye PenetrationCompliance with ASTM F1929-15Met predetermined acceptance criteria.
Seal StrengthCompliance with ASTM F88/F88M-23Met predetermined acceptance criteria.
TransportationTransportation Simulation TestingCompliance with ASTM 4169-22 and subsequent package integrity testingMet predetermined acceptance criteria.
BiocompatibilityCytotoxicityNon-cytotoxic (per ISO 10993-5)Non-cytotoxic.
SensitizationNon-sensitizing (per ISO 10993-10)Non-sensitizing.
IrritationNon-irritating (per ISO 10993-23)Non-irritating.
EndotoxinBacterial Endotoxin Assay (BEA)< 20 EU/device (per USP <85> and USP <161>)< 20 EU/device.
EmbryotoxicityMouse Embryo Assay (MEA)≥80% embryos developed to expanded blastocyst at 96h (One-Cell MEA) (per 2021 FDA guidance)≥80% embryos developed to expanded blastocysts within 96 hours.
Bench Performance (General)Visual Inspection (device, packaging, labels)Met predetermined acceptance criteriaMet predetermined acceptance criteria.
Dimensional VerificationMet predetermined acceptance criteriaMet predetermined acceptance criteria.
Compatibility Testing (catheter, guide, stylet)Met predetermined acceptance criteriaMet predetermined acceptance criteria.
Usable Length Verification (assembled guide and catheter)Met predetermined acceptance criteriaMet predetermined acceptance criteria.
Echogenic Marking Verification by UltrasoundMet predetermined acceptance criteriaMet predetermined acceptance criteria.
Luer Lock Connection VerificationCompliance with ISO 80369-7:2021Met predetermined acceptance criteria.
Aspiration, Flush, Leak, Flowrate Tests (catheter)Compliance with EN 1618Met predetermined acceptance criteria.
Distance Marking Location and DurabilityMet predetermined acceptance criteriaMet predetermined acceptance criteria.
Curvature Verification (Pre-curved guide)Met predetermined acceptance criteriaMet predetermined acceptance criteria.
Dislodgement Force (positioning stopper)Met predetermined acceptance criteriaMet predetermined acceptance criteria.
Tensile Strength Testing (catheter)Met predetermined acceptance criteriaMet predetermined acceptance criteria.
Echogenic Marking Visibility Test (catheter)Met predetermined acceptance criteriaMet predetermined acceptance criteria.
**Bench Performance (Specific) **Tip Drop Angle of Simulated Loaded Catheter when held horizontallyMet predetermined acceptance criteria (with mitigation)"All test articles met the predetermined acceptance criteria in the tests, except for 'tip drop when held horizontally.'" The device labeling includes a precaution.

Regarding the specific questions asked for an AI/ML device:

  1. A table of acceptance criteria and the reported device performance: Provided above for the non-clinical tests described.
  2. Sample sized used for the test set and the data provenance: Not applicable. These were non-clinical, benchtop, and laboratory tests. No "test set" in the context of patient data or clinical images is mentioned.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for these physical/chemical tests is established by laboratory standards and measurements, not expert human interpretation of data.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/ML device.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI/ML device.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): For the non-clinical tests, ground truth was derived from standardized testing methods, material specifications, and validated laboratory procedures (e.g., spectrophotometry for cytotoxicity, bacterial growth for endotoxin, visual inspection, dimensional measurements, mechanical testing).
  8. The sample size for the training set: Not applicable. This is not an AI/ML device, so there is no "training set."
  9. How the ground truth for the training set was established: Not applicable.

In conclusion, the provided FDA clearance letter pertains to a conventional medical device (Embryo Transfer Catheter) and not an AI/ML artificial intelligence medical device. Therefore, the detailed requirements for AI/ML device testing (e.g., data sets, expert review, MRMC studies) are not found in this document. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the rigorous non-clinical, laboratory, and benchtop performance testing that ensured the device's safety and effectiveness compared to a legally marketed predicate device.

U.S. Food & Drug Administration FDA Clearance Letter

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID # 04017.08.00
Silver Spring, MD 20993
www.fda.gov

July 23, 2025

Vitrolife Sweden AB
Ann-Cathrine Ericson
RA Manager
Gustaf Werners gata 2
Gothenburg, 421 32
SWEDEN

Re: K243373

Trade/Device Name: Embryo Transfer Catheter (ETC) Assortment:

  • EmbryoCath Embryo Transfer Catheter - Pre-curved 230 mm (REF 17500)
  • EmbryoCath Embryo Transfer Catheter - Straight Malleable 180 mm (REF 17501)
  • EmbryoCath Embryo Transfer Catheter - Straight Malleable 230 mm (REF 17502)
  • EmbryoCath Stylet - 180 mm (REF 17510)
  • EmbryoCath Stylet - 230 mm (REF 17511)

Regulation Number: 21 CFR 884.6110
Regulation Name: Assisted Reproduction Catheters
Regulatory Class: II
Product Code: MQF
Dated: June 13, 2025
Received: June 13, 2025

Dear Ann-Cathrine Ericson:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

Page 2

K243373 - Ann-Cathrine Ericson Page 2

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

Page 3

K243373 - Ann-Cathrine Ericson Page 3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Monica D. Garcia -S

Monica D. Garcia, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known): K243373

Device Name:
Embryo Transfer Catheter (ETC) Assortment:

  • EmbryoCath Embryo Transfer Catheter - Pre-curved 230 mm (REF 17500)
  • EmbryoCath Embryo Transfer Catheter - Straight Malleable 180 mm (REF 17501)
  • EmbryoCath Embryo Transfer Catheter - Straight Malleable 230 mm (REF 17502)
  • EmbryoCath Stylet - 180 mm (REF 17510)
  • EmbryoCath Stylet - 230 mm (REF 17511)

Indications for Use (Describe):

The Embryo Transfer Catheter is intended for introduction of in vitro fertilized embryo(s) into the uterine cavity.

The stylet is intended to be used with Vitrolife's Embryo Transfer Catheters to assist in uterine access of the guide during an embryo transfer procedure.

Type of Use (Select one or both, as applicable):
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

K243373 Page 1 of 5

Embryo Transfer Catheter (ETC) Assortment

510(k) Summary – K243373

Submitted by: Vitrolife Sweden AB
Gustaf Werners gata 2
SE-421 32 Västra Frölunda
Sweden

Contact Person: Ann-Catherine Ericson
Vitrolife Sweden AB
Gustaf Werners gata 2
SE-421 32 Västra Frölunda
Sweden
Phone: +46 31 721 80 00
Fax: +46 31 721 80 90
Email: aericson@vitrolife.com

Date prepared: July 23, 2025

1 Device Identification

Trade Name: Embryo Transfer Catheter (ETC) Assortment:

  • EmbryoCath Embryo Transfer Catheter - Pre-curved 230 mm (REF 17500)
  • EmbryoCath Embryo Transfer Catheter - Straight Malleable 180 mm (REF 17501)
  • EmbryoCath Embryo Transfer Catheter - Straight Malleable 230 mm (REF 17502)
  • EmbryoCath Stylet - 180 mm (REF 17510)
  • EmbryoCath Stylet - 230 mm (REF 17511)

Common name: Embryo Transfer Catheters
Regulatory Class: II
Regulation number: 21 CFR 884.6110
Regulation name: Assisted Reproduction Catheters
Product code: MQF (Catheter, Assisted Reproduction)

2. Predicate Device

Kitazato ET Catheters (K162878) manufactured by Kitazato Corporation.

The predicate device has not been subject to a design-related recall.

3. Device Description

Vitrolife´s Embryo Transfer Catheter (ETC) Assortment consists of three embryo transfer catheters (ETCs), one pre-curved variant (Model 17500, 230 mm) and two straight variants (REFs 17501, 17502) in different lengths (180 mm and 230 mm); and two stylets (Stylets) (REFs 17510, 17511) in different lengths (180 mm and 230 mm). ETC and Stylet are sterile, single-use devices used to deliver in vitro fertilized embryos to the uterine cavity.

Embryo Transfer Catheter consists of:

  • A guide, also referred as outer sheath. The guide is used to navigate through the cervix canal. The guide has a distance marking, a stopper and a rounded tip to facilitate proper positioning. The guide is available in two different variants:
    • Pre-curved - stiff

Page 6

K243373 Page 2 of 5

Embryo Transfer Catheter (ETC) Assortment

  • Straight - soft and malleable
  • A catheter, also referred as inner catheter. The catheter is loaded with the embryo(s) in a small volume of transfer medium and inserted through the guide to gently deposit the embryo(s) into the uterine cavity. The transfer catheter is soft, flexible and approximately 50 mm longer than the guide. The tip of the catheter has an echogenic marking to enable ultrasound guidance.

Embryo Transfer Catheter Stylet consists of:

  • A plastic-coated metal wire, which is malleable. The stylet is an accessory that can be used together with Vitrolife's guides to make them stiffer.

The 17500 variant includes a stylet. The 17501 and 17502 variants do not include stylets, but the stylet can be purchased separately.

4. Indications for Use

The Embryo Transfer Catheter is intended for introduction of in vitro fertilized embryo(s) into the uterine cavity.

The stylet is intended to be used with Vitrolife´s Embryo Transfer Catheters to assist in uterine access of the guide during an embryo transfer procedure.

5. Comparison of Intended Use and Technological Characteristics of the Subject and Predicate Devices

CharacteristicsSubject Device K243373Predicate Device K162878Comparison
ManufacturerVitrolife Sweden ABKitazato CorporationN/A
Trade nameEmbryo Transfer Catheter (ETC) Assortment: EmbryoCath Embryo Transfer Catheter - Pre-curved 230 mm (REF 17500); EmbryoCath Embryo Transfer Catheter - Straight Malleable 180 mm (REF 17501); EmbryoCath Embryo Transfer Catheter - Straight Malleable 230 mm (REF 17502); EmbryoCath Stylet - 180 mm (REF 17510); EmbryoCath Stylet - 230 mm (REF 17511)Kitazato ET CathetersN/A
Indications for useThe Embryo Transfer Catheter is intended for introduction of in vitro fertilized embryo(s) into the uterine cavity. The Stylet is intended to be used with Vitrolife´s Embryo Transfer Catheters to assist in uterine access of the guideThe Kitazato ET Catheters are intended for ultrasound guided introduction of embryos into the uterine cavity following in vitro fertilisation. Kitazato ET Catheters are provided in various configurations (Type 1 through Type 5), whichThe indications for use statements subject and predicate devices are not identical; however, they have the same intended use (i.e., transfer of embryos into the uterine cavity).

Page 7

K243373 Page 3 of 5

Embryo Transfer Catheter (ETC) Assortment

CharacteristicsSubject Device K243373Predicate Device K162878Comparison
during an embryo transfer procedure.consist of the following components: • Transfer Catheter, for delivery of embryos into the uterine cavity; • Guide Catheter, to guide the insertion of the Transfer Catheter and reinforce it during use: • Trial Catheter, used to confirm the curvature of the cervix and if the cervix is passable; • Stylet Sheath, used to increase the strength of the Guide Catheter during insertion.
Device ComponentsEmbryo transfer catheter, guide and styletTransfer catheter, guide catheter, trial catheter, and a stylet sheathDifferent. The differences in device components do not raise different questions of safety and effectiveness (S&E).
MaterialsStraight Guide: Polyethylene (PE), Silicone Pre-curved Guide: Thermoplastic polyurethane (TPU), Silicone Catheter: Thermoplastic polyurethane (TPU), Polyamide (PA 12) Stylet: Stainless steel, Polyamide (PA 12) Luer Connector: PEBACatheter/Trial shaft: Polyurethane Guide: Fluoric resin Outer Stiffener: Stainless steel Luer Connector: ABSDifferent. The differences in device materials do not raise different questions of S&E.
MarkingsDistance markings on embryo transfer catheter and guide.Depth marks on catheter and guide.Same
Ultrasound visibilityYes – Echogenic markingYes – Echo line for Type3 and Type4.Same
Single useYesYesSame
DimensionLength: Catheter: 20 cm, 25 cm Guide: 13 cm, 18 cm Stylet: 15 cm, 20 cm OD: Catheter: 1.45 mm Guide: 2.35 mm Stylet: 1.2 mmLength: Catheter: 18 cm, 23 cm Guide: 14 cm, 19 cm OD: Catheter: 1.55 mm (4.7 Fr) Guide: Unknown Stylet: UnknownDifferent. The differences in device dimension do not raise different questions of S&E.
SterilizationSterilized by E-Beam radiationSterilized by ethylene oxide (ETO)Different. The differences in device

Page 8

K243373 Page 4 of 5

Embryo Transfer Catheter (ETC) Assortment

CharacteristicsSubject Device K243373Predicate Device K162878Comparison
sterilization methods do not raise different questions of S&E.
Shelf-life3 months3 yearsDifferent. The differences in shelf-life duration do not raise different questions of S&E
Mouse Embryo Assay (MEA)One-cell MEA: ≥80% embryos developed to expanded blastocyst at 96hOne-cell MEA: ≥80% embryos developed to expanded blastocyst at 96hSame
Bacterial Endotoxin Assay (BEA)<20 EU/device<20 EU/deviceSame

The subject and predicate device have differences in their indications for use statements; however, their intended uses are the same (i.e., transfer of embryos to the uterine cavity). There are different technological characteristics, including differences in device components, materials, dimension, sterilization method, and shelf-life. The different technological characteristics do not raise different questions of safety and effectiveness.

6. Summary of Non-Clinical Performance Testing

The following studies have been performed to support substantial equivalence to the predicate device

Sterilization validation performed according to ISO 11137-1:2006/(R)2015 and ISO 11137-2:2013.

Packaging integrity testing according to ISO 11607-1:2019 and ISO 11607-2:2019

  • Visual inspection per ASTM F1886/F1886M-16
  • Dye penetration per ASTM F1929-15
  • Seal strength per ASTM F88/F88M-23

Transportation simulation testing per ASTM 4169-22 and package integrity testing

Biocompatibility evaluation conducted in accordance with the 2023 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices –Part 1: Evaluation and testing within a risk management process." Testing included the following assessments:

  • Cytotoxicity per ISO 10993-5
  • Sensitization per ISO 10993-10
  • Irritation per ISO 10993-23

Testing showed the device material to be non-cytotoxic, non-sensitizing, and non-irritating.

Endotoxin testing (BEA) performed according to USP <85> and USP <161> with acceptance criteria < 20 EU/device.

Mouse embryo assay (MEA) performed per the 2021 FDA guidance document Mouse Embryo Assay for Assisted Reproduction Technology Devices with specification: One-Cell MEA: ≥80% embryos developed to expanded blastocysts within 96 hours.

Bench performance assessed before and after accelerated aging to the equivalent of three months of real-time aging in accordance with ASTM F1980-21:

Page 9

K243373 Page 5 of 5

Embryo Transfer Catheter (ETC) Assortment

  • Visual inspection of device, packaging and labels
  • Dimensional verification
  • Compatibility testing with catheter, guide, and stylet
  • Useable length verification of assembled guide and catheter
  • Echogenic marking verification by ultrasound
  • Verification of Luer lock connection per ISO 80369-7:2021
  • Aspiration, flush, leak and flowrate tests of catheter per EN 1618
  • Distance marking location and durability
  • Curvature verification of Pre-curved guide
  • Dislodgement force of positioning stopper
  • Tensile strength testing of catheter
  • Echogenic marking visibility test of catheter
  • Tip drop angle of simulated loaded catheter when held horizontally

All test articles met the predetermined acceptance criteria in the tests, except for "tip drop when held horizontally." The device labeling includes a precaution "Due to the soft nature of the catheter, when loaded, the catheter may bend. Ensure the tip of the catheter is properly aligned with the guide before inserting it into the guide during the embryo transfer procedure" to mitigate embryo loss risks.

7. Conclusion

The results of the performance testing described above demonstrate that the EmbryoCath Embryo Transfer Catheters (REFs 17500, 17501, 17502) and EmbryoCath Stylets (REFs 17510, 17511) are as safe and effective as the predicate device and supports a determination of substantial equivalence.

§ 884.6110 Assisted reproduction catheters.

(a)
Identification. Assisted reproduction catheters are devices used in in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures to introduce or remove gametes, zygote(s), preembryo(s), and/or embryo(s) into or from the body. This generic type of device may include catheters, cannulae, introducers, dilators, sheaths, stylets, and component parts.(b)
Classification. Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing).