K Number
K243312

Validate with FDA (Live)

Device Name
CER-S
Manufacturer
Date Cleared
2025-02-20

(121 days)

Product Code
Regulation Number
870.1425
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CER-S is a medical device stand-alone software intended for the analysis, editing, review and reporting of continuous ECG digital recordings, to support the clinician in the interpretation of the trace and the diagnosis of cardiac rhythm disorders in adult and pediatric patients.

CER-S is intended for use in clinical and hospital settings only by qualified medical personnel or adequately trained professional personnel working under the supervision and responsibility of a clinician. Users must undergo a thorough software training before using the medical device. This device does not provide an automated interpretation and is not intended for use as the only diagnostic tool.

CER-S analysis provides indications for evaluation of:

  • Patients with rhythm disturbances (cardiac arrhythmias),
  • Patients with transient myocardial ischemia,
  • Patients with pacemaker (only if pacing detection is available from the input recording),
  • Patients needing HRV evaluation.
  • Newborn patients limited to QRS detection.
Device Description

CER-S is a tool, designed to offer a framework for the interaction of different software-modules, providing advanced solutions for Continuous ECG Recording (CER).
Different modules provide:

  • ECG Beat detection and classification,
  • analysis of ECG rhythm, arrhythmia detection,
  • interactive Viewer and set of tools to perform editing of ECG beats, Rhythm annotations and noise Windows,
  • interactive Continuous ECG Viewer,
  • interactive display/management of ECG Templates,
  • holter-like report for analyzed Continuous ECG records,
  • record exportation, in ISHNE format,
  • generation of aECG FDA HL7 XML (v. 2).

Note: The automatic analysis is limited to data acquired from electrodes with conductive paste/gel (dry and dry/metal electrodes are not intended to be used) placed on standard location in compatible formats from any device used for the arrhythmia diagnostics such as Holter, event recorder, 12-lead ambulatory ECG devices when assessment of the rhythm is necessary.
Moreover, CER-S allows the automatic analysis for the following patch location:

  • two-electrode patches positioned in the left upper chest area at a roughly inclined angle
  • three-electrode triangular shape patches positioned on the patient's left upper chest area below the 1st rib, at an inclined angle
  • three-electrode T-shaped patches positioned in the center-thoracic position between the upper part of the chest (manubrium) and the sternum
    In all cases, patch placement must strictly follow the indication provided by the manufacturer of the certified-patch device.
AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the CER-S device, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document (a 510(k) Premarket Notification) does not explicitly state acceptance criteria in terms of specific performance metrics with numerical thresholds (e.g., sensitivity > X%, specificity > Y%). Instead, it outlines the scope of performance testing and concludes that the device is "substantially equivalent" to its predicate and meets the relevant standards.

The document focuses on demonstrating that the device adheres to recognized industry standards and its performance is comparable to the predicate.

Acceptance Criteria (Implied from the document)Reported Device Performance
Compliance with ECG performance testing standards.Non-clinical tests were conducted according to: • IEC 60601-2-47:2012 (Recognition Number 3-155) • IEC 60601-2-25:2011 (Recognition Number 3-105) The testing demonstrated that CER-S is substantially equivalent to the predicate for the proposed intended use.
Compliance with software development lifecycle standards.Software development life cycle was conducted according to: • IEC 62304:2015 (Recognition Number 13-79)
Compliance with usability engineering process standards.Usability engineering process was conducted according to: • IEC 62366-1:2020 (Recognition Number 5-129)
Compliance with risk management standards.Potential hazards were evaluated and controlled through Risk Management activities, according to: • ISO 14971:2019 (Recognition Number 5-125)
Equivalence in intended use, technological characteristics, and safety/effectiveness vs. predicate.The subject device has the same intended use as the predicate device (K213861). Any differences in technological characteristics (Enhanced software usability, Option to activate and run pacemaker analysis, Overall optimization of embedded algorithms for beat and noise detection, rhythm analysis and beat measurements, Strengthened cybersecurity measures) do not raise different questions of safety or effectiveness. The device has been shown to be substantially equivalent to the legally marketed predicate devices based on indications for use, technological characteristics, and comparison.

2. Sample Size Used for the Test Set and Data Provenance

The document states: "No clinical tests were performed to demonstrate the substantial equivalence of CER-S." This implies that there isn't a "test set" in the traditional sense of a clinical dataset used to evaluate the device's diagnostic performance against a ground truth.

The performance evaluation relied on non-clinical tests against recognized standards rather than a clinical dataset with individual patient records. Therefore, information about sample size, country of origin, or retrospective/prospective nature of a clinical test set is not available in this document.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

As there were no clinical tests performed with a ground truth established by experts, this information is not applicable/not available from the provided text.

4. Adjudication Method for the Test Set

Since no clinical test set was used with expert review, an adjudication method is not applicable/not available from the provided text.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Improvement with AI vs. Without AI Assistance

The document explicitly states, "No clinical tests were performed to demonstrate the substantial equivalence of CER-S." Therefore, no MRMC comparative effectiveness study was conducted, and information about human reader improvement with AI assistance is not available. The CER-S is a standalone software, not an AI-assisted reading tool for human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, a standalone performance evaluation was done through non-clinical testing against recognized standards. The device itself is described as a "stand-alone software" intended to "support the clinician," not replace them, and explicitly states it "does not provide an automated interpretation and is not intended for use as the only diagnostic tool." The reported performance focused on compliance with ECG performance standards and software/risk management standards, indicating the algorithm's functionality and safety were evaluated.

7. The Type of Ground Truth Used

The "ground truth" for the non-clinical performance evaluation appears to be the requirements and specifications within the referenced industry standards (IEC 60601-2-47, IEC 60601-2-25 for ECG performance; IEC 62304 for software; IEC 62366-1 for usability; ISO 14971 for risk management). The device's algorithms and functions were tested to ensure they met the technical specifications and safety/performance criteria outlined by these standards, rather than a clinical ground truth like pathology or expert consensus on patient cases.

8. The Sample Size for the Training Set

The document does not provide information about the training set size for the CER-S algorithms. It mentions "Overall optimization of embedded algorithms for beat and noise detection, rhythm analysis and beat measurements," which implies an underlying development and training process, but no details are given.

9. How the Ground Truth for the Training Set Was Established

The document does not provide information on how the ground truth for the training set was established.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

February 20, 2025

CardioCalm Srl Fabio Badilini President Corso Martiri della Libertà, nr. 40 Montichiari, BS 25018 Italy

Re: K243312

Trade/Device Name: CER-S Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK Dated: January 22, 2025 Received: January 22, 2025

Dear Fabio Badilini:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Jennifer W. Shih -S

Jennifer Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)K242242
---------------------------------------

K243312 Device Name

CER-S

Indications for Use (Describe)

CER-S is a medical device stand-alone software intended for the analysis, editing, review and reporting of continuous ECG digital recordings, to support the clinician in the interpretation of the trace and the diagnosis of cardiac rhythm disorders in adult and pediatric patients.

CER-S is intended for use in clinical and hospital settings only by qualified medical personnel or adequately trained professional personnel working under the supervision and responsibility of a clinician. Users must undergo a thorough software training before using the medical device. This device does not provide an automated interpretation and is not intended for use as the only diagnostic tool.

CER-S analysis provides indications for evaluation of:

  • · Patients with rhythm disturbances (cardiac arrhythmias),
  • · Patients with transient myocardial ischemia,
  • · Patients with pacemaker (only if pacing detection is available from the input recording),
  • · Patients needing HRV evaluation.
  • · Newborn patients limited to QRS detection.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

(as required by 21 CFR 807.92)
SubmitterCardioCalm Srl
Corso Martiri della Libertà, nr. 40
25018 Montichiari (BS)
Telephone+ 39 030 9650745
Fax+ 39 030 9650572
Contact PersonFabio Badilini
President & Legal Representative
AddressCorso Martiri della Libertà, nr. 40
Telephone+ 39 030 9650745
Fax+ 39 030 9650572
Emailbadilini@cardiocalm.com
Date Prepared21/10/2024
Trade NameCER-S
Common NameContinuous ECG Recording - Suite
Panel CodeCardiovascular/74
Classification NameComputer diagnostic, programmable
ClassClass II
Regulation Number21 CFR 870.1425
Regulation DescriptionProgrammable diagnostic computer
Product CodeDQK
Name of Predicate Device510(k) #Manufacturer
CER-SK213861CardioCalm Srl
Reason forSubmissionIntention to introduce into commercial distribution a device, CER-S, that has undergone significantchanges that could affect the substantial equivalence with the predicate device.
DeviceDescriptionCER-S is a tool, designed to offer a framework for the interaction of different software-modules,providing advanced solutions for Continuous ECG Recording (CER).Different modules provide:
ECG Beat detection and classification,
analysis of ECG rhythm, arrhythmia detection,
interactive Viewer and set of tools to perform editing of ECG beats, Rhythm annotationsand noise Windows,
interactive Continuous ECG Viewer,
interactive display/management of ECG Templates,
holter-like report for analyzed Continuous ECG records,
record exportation, in ISHNE format,generation of aECG FDA HL7 XML (v. 2).
Note: The automatic analysis is limited to data acquired from electrodes with conductive paste/gel(dry and dry/metal electrodes are not intended to be used) placed on standard location incompatible formats from any device used for the arrhythmia diagnostics such as Holter, eventrecorder, 12-lead ambulatory ECG devices when assessment of the rhythm is necessary.Moreover, CER-S allows the automatic analysis for the following patch location:
two-electrode patches positioned in the left upper chest area at a roughly inclined angle
three-electrode triangular shape patches positioned on the patient's left upper chest areabelow the 1st rib, at an inclined angle
three-electrode T-shaped patches positioned in the center-thoracic position between theupper part of the chest (manubrium) and the sternum
In all cases, patch placement must strictly follow the indication provided by the manufacturer of thechanges that could affect the substantial equivalence with the predicate device.
DeviceDescriptionCER-S is a tool, designed to offer a framework for the interaction of different software-modules, providing advanced solutions for Continuous ECG Recording (CER).Different modules provide:• ECG Beat detection and classification,• analysis of ECG rhythm, arrhythmia detection,• interactive Viewer and set of tools to perform editing of ECG beats, Rhythm annotations and noise Windows,• interactive Continuous ECG Viewer,• interactive display/management of ECG Templates,• holter-like report for analyzed Continuous ECG records,• record exportation, in ISHNE format,• generation of aECG FDA HL7 XML (ν. 2).Note: The automatic analysis is limited to data acquired from electrodes with conductive paste/gel (dry and dry/metal electrodes are not intended to be used) placed on standard location in compatible formats from any device used for the arrhythmia diagnostics such as Holter, event recorder, 12-lead ambulatory ECG devices when assessment of the rhythm is necessary. Moreover, CER-S allows the automatic analysis for the following patch location:• two-electrode patches positioned in the left upper chest area at a roughly inclined angle• three-electrode triangular shape patches positioned on the patient's left upper chest area below the 1st rib, at an inclined angle• three-electrode T-shaped patches positioned in the center-thoracic position between the upper part of the chest (manubrium) and the sternumIn all cases, patch placement must strictly follow the indication provided by the manufacturer of the certified-patch device.
Indications foruseCER-S is a medical device stand-alone software intended for the analysis, editing, review and reporting of continuous ECG digital recordings, to support the clinician in the interpretation of the trace and the diagnosis of cardiac rhythm disorders in adult and pediatric patients.CER-S is intended for use in clinical and hospital settings only by qualified medical personnel or adequately trained professional personnel working under the supervision and responsibility of a clinician. Users must undergo a thorough software training before using the medical device.

{5}------------------------------------------------

Traditional 510(k) Premarket Notification

CER-S

This device does not provide an automated interpretation and is not intended for use as the onlydiagnostic tool.CER-S analysis provides indications for evaluation of:
Patients with rhythm disturbances (cardiac arrhythmias)
Patients with transient myocardial ischemia
Patients with pacemaker (only if pacing detection is available from the input recording)
Patients needing HRV evaluation
Newborn patients limited to QRS detection.
SubstantialEquivalenceThe fundamental scientific principles and technological characteristic, including the indications foruse and general design are equivalent to the predicate device, as summarized hereafter:• Software type, Holter Analyzer, PC based SaMD, working in Windows Operative System, as the predicate,• Patient population, adults and pediatrics, as the predicate,• User groups: qualified medical personnel or adequately trained professional personnel working under the supervision and responsibility of a clinician, equivalent to predicate device,• Indications for use, equivalent to the predicate,• Environmental use, equivalent to the predicate.
The subject device introduces significant improvements over the predicate device. The mainmodifications can be summarized as follows:• Enhanced software usability,• Option to activate and run pacemaker analysis,• Overall optimization of embedded algorithms for beat and noise detection, rhythm analysis and beat measurements,• Strengthened cybersecurity measures.In conclusion, the subject device has the same intended use as the predicate device, and anydifferences in technological characteristics do not raise different questions of safety oreffectiveness.
PerformanceDataNon-clinical tests have been conducted on the Subject Device:ECG performance testing, according to:• IEC 60601-2-47:2012 (Recognition Number 3-155)• IEC 60601-2-25:2011 (Recognition Number 3-105)Software development life cycle, according to:• IEC 62304:2015 (Recognition Number 13-79)Usability engineering process, according to:• IEC 62366-1:2020 (Recognition Number 5-129)Any potential hazards have been evaluated and controlled through Risk Managementactivities, according to:• ISO 14971:2019 (Recognition Number 5-125)The testing demonstrated that CER-S is substantially equivalent to the predicate for the proposedintended use.No clinical tests were performed to demonstrate the substantial equivalence of CER-S.
ConclusionBased on the indications for use, technological characteristics, and comparison to predicatedevices, the Subject device CER-S has been shown to be substantially equivalent to the legallymarketed predicate devices.

§ 870.1425 Programmable diagnostic computer.

(a)
Identification. A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.(b)
Classification. Class II (performance standards).