K Number
K243305

Validate with FDA (Live)

Device Name
Masimo W1
Manufacturer
Date Cleared
2025-04-03

(164 days)

Product Code
Regulation Number
870.2340
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Masimo W1™ and the integrated Masimo W1 Module are intended to record, store and transfer single-channel electrocardiogram (ECG) rhythms. The Masimo W1 also displays ECG rhythms, and the Masimo W1 Module detects the presence of atrial fibrillation. The Masimo W1 and the integrated Masimo W1 Module are intended for use by healthcare professionals, patients with known or suspected heart conditions, and health-conscious individuals.

Masimo W1 Module ECG software is an over-the-counter (OTC) software that operates on the Masimo W1 Module that can be used with compatible watches (e.g., Masimo W1). The software is intended to create, record, store, transfer, and display a single channel electrocardiogram (ECG) for informational use only in adults 22 years and older. It supports the classification of either atrial fibrillation (AFib) or sinus rhythm with the intention of aiding heart rhythm identification; it is not intended to replace traditional methods of diagnosis or treatment. The software is not intended for users with other known arrhythmias and users should not interpret or take clinical action based on the device output without consultation of a qualified healthcare professional.

The Masimo W1™ and the integrated Masimo W1 Module are also intended for the spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The Masimo W1 and the integrated Masimo W1 Module are indicated for adults in hospitals, clinics, long-term care facilities, and homes.

Device Description

The Masimo W1 is a watch that incorporates the Masimo W1 Module, which is the device that is responsible for the physiological signal detection and algorithm used to support the different parameters. The Masimo W1 Module incorporates spot check ECG functionality and Masimo SET Pulse Oximetry technology so that it can provide both ECG and Masimo SET pulse oximetry parameters. As part of this submission, Masimo is requesting clearance for an automated atrial fibrillation "AFib" Classification Feature that is used to analyze the single channel ECG waveform.

The parameter outputs from the Masimo W1 Module are communicated and displayed on the watch screen so that the data can be viewed and recorded. Masimo W1 also supports wireless communication of monitored data to a compatible smart device application. The sharing of the parameter data to the applications allows the users to see and track their data using their smart phones. Smart phone applications can also help to share information to caregivers and healthcare professionals.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) Clearance Letter for the Masimo W1 with the Atrial Fibrillation (AFib) Classification Feature:


1. Table of Acceptance Criteria and Reported Device Performance

For the Atrial Fibrillation Classification Feature:

Performance MetricAcceptance Criteria (Not explicitly stated as "acceptance criteria" but implied by comparison to predicates and clinical study results)Reported Device Performance (Masimo W1 with AFib Classification Feature)
Atrial Fibrillation (AFib) Classification
SensitivityComparable to or better than predicate devices (Withing Scan Monitor: 96.3%; Samsung ECG Monitor: 98.1%)99.3% [96.3%, 100%]
Specificity (Sinus Rhythm)Comparable to or better than predicate devices (Withing Scan Monitor: 100%; Samsung ECG Monitor: 100%)100% [97.8%, 100%]
Positive Predictive Value (PPV)(Not explicitly compared in table, but reported as a key performance metric)100% [97.5%, 100%]
Unclassified Rate(Not explicitly compared in table)5.0%
Noise Rate(Not explicitly compared in table)1.7%
ECG Waveform Quality Analysis
Qualitative Agreement (with 12-lead ECG Lead I)High agreement by qualified clinicians98% [96% - 98%]
Quantitative Similarity (Key ECG features like QRS amplitude, QRS width)Similar to Lead I of a gold-standard 12-lead ECG referenceSimilar to Lead I of a gold-standard 12-lead ECG reference

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: The document states "Prospective clinical testing was conducted to validate the AFib Classification Feature on adult subjects from 4 different sites." While a specific number of subjects is not provided, it indicates a multi-site study.
  • Data Provenance: Prospective clinical testing. The country of origin is not explicitly stated in the provided text.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: Unspecified (referred to as "qualified clinicians" for the qualitative assessment of ECG waveforms).
  • Qualifications of Experts: "Qualified clinicians" were used for the qualitative assessment of the ECG waveforms. Further specific qualifications (e.g., cardiologist, years of experience) are not detailed in the provided text.

4. Adjudication Method for the Test Set

The document does not explicitly state the adjudication method used for establishing the ground truth for the test set (e.g., 2+1, 3+1). It only mentions that "qualified clinicians" made an agreement assessment.


5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was it done? No, a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with AI assistance versus without AI assistance was not described in the provided text. The study focused on the standalone performance of the AI AFib classification feature and the quality of its ECG waveform compared to a gold standard.
  • Effect Size: Not applicable, as an MRMC study was not described.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • Was it done? Yes, a standalone performance study was clearly conducted for the AFib Classification Feature. The "Masimo W1 ECG AFib Classification feature Performance" section directly reports sensitivity, specificity, and PPV for the algorithm's output. The "Masimo W1 Module ECG software" is described as supporting the classification, indicating an algorithmic assessment.

7. Type of Ground Truth Used

  • For AFib Classification: The gold standard for AFib classification is not explicitly stated, but clinical validation for ECG rhythm typically uses expert-adjudicated 12-lead ECG recordings. The document mentions "comparing similarity between Masimo W1 and gold-standard 12 lead ECG as reference" for waveform quality, which strongly implies 12-lead ECGs were used as a reference for rhythm classification as well.
  • For ECG Waveform Quality: Gold-standard 12-lead ECG as reference.

8. Sample Size for the Training Set

The document does not specify the sample size for the training set. It only describes the test set used for clinical validation.


9. How the Ground Truth for the Training Set Was Established

The document does not specify how the ground truth for the training set was established. It focuses solely on the validation/test set.

FDA 510(k) Clearance Letter - Masimo W1

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

April 3, 2025

Masimo Corporation
Sindura Penubarthi
Associate Director, Regulatory Affairs
52 Discovery
Irvine, California 92618

Re: K243305
Trade/Device Name: Masimo W1
Regulation Number: 21 CFR 870.2340
Regulation Name: Electrocardiograph
Regulatory Class: Class II
Product Code: DPS, DQA, DXH, QDA
Dated: February 25, 2025
Received: February 25, 2025

Dear Sindura Penubarthi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K243305 - Sindura Penubarthi Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K243305 - Sindura Penubarthi Page 3

Sincerely,

Jennifer W. Shih -S

Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K243305

Device Name
Masimo W1

Indications for Use (Describe)

Masimo W1™ and the integrated Masimo W1 Module are intended to record, store and transfer single-channel electrocardiogram (ECG) rhythms. The Masimo W1 also displays ECG rhythms, and the Masimo W1 Module detects the presence of atrial fibrillation. The Masimo W1 and the integrated Masimo W1 Module are intended for use by healthcare professionals, patients with known or suspected heart conditions, and health-conscious individuals.

Masimo W1 Module ECG software is an over-the-counter (OTC) software that operates on the Masimo W1 Module that can be used with compatible watches (e.g., Masimo W1). The software is intended to create, record, store, transfer, and display a single channel electrocardiogram (ECG) for informational use only in adults 22 years and older. It supports the classification of either atrial fibrillation (AFib) or sinus rhythm with the intention of aiding heart rhythm identification; it is not intended to replace traditional methods of diagnosis or treatment. The software is not intended for users with other known arrhythmias and users should not interpret or take clinical action based on the device output without consultation of a qualified healthcare professional.

The Masimo W1™ and the integrated Masimo W1 Module are also intended for the spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The Masimo W1 and the integrated Masimo W1 Module are indicated for adults in hospitals, clinics, long-term care facilities, and homes.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D) ☒ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary

MASIMO CORPORATION
52 Discovery
Irvine, CA 92618

Submitter and Address of Manufacturing Facility:Masimo Corporation52 DiscoveryIrvine, CA 92618Phone: (949) 297-7000FAX: (949) 297-7592
Date:April 2nd 2025
Contact:Sindura PenubarthiAssociate Director, Regulatory AffairsMasimo CorporationPhone: (949) 396-4041
Trade Name:Masimo W1
Common Name:Electrocardiograph
Classification Regulation/Product Code:21 CFR 870.2340, Class II/DPS
Additional Product Code:DQA, DXH, QDA
Establishment Registration Number:3011353843
Reason for Premarket Notification:Update to add Atrial Fibrillation (AFib) Classification Feature
Primary Predicate:K240229 – Masimo W1
Secondary Predicate:K201456 - Withing Scan Monitor
Tertiary PredicateK201168 – Samsung ECG Monitor

1 Device Description

The Masimo W1 is a watch that incorporates the Masimo W1 Module, which is the device that is responsible for the physiological signal detection and algorithm used to support the different parameters. The Masimo W1 Module incorporates spot check ECG functionality and Masimo SET Pulse Oximetry technology so that it can provide both ECG and Masimo SET pulse oximetry parameters. As part of this submission, Masimo is requesting clearance for an automated atrial fibrillation "AFib" Classification Feature that is used to analyze the single channel ECG waveform.

The parameter outputs from the Masimo W1 Module are communicated and displayed on the watch screen so that the data can be viewed and recorded. Masimo W1 also supports wireless communication of monitored data to a compatible smart device application. The sharing of the parameter data to the applications allows the users to see and track their data using their smart phones. Smart phone applications can also help to share information to caregivers and healthcare professionals.

The Masimo W1 specifications are provided in Table 1-1 below:

Table 1-1 - Specifications

FeatureSpecifications
Continuous Display of Parameter DataYes
Performance Specifications
SpO2, No Motion/ Low Perfusion (70-100%)2% Arms adults

K243305
Page 1 of 13

Page 6

510(k) Summary

Table 1-1 - Specifications

FeatureSpecifications
Pulse Rate (25-240 bpm)3 bpm Arms adults
Heart Rate (25-240 bpm)≤5 bpm adults
ECG Rhythm ClassificationsAtrial Fibrillation, Sinus Rhythm
Electrical Specifications
BatteryInternal Rechargeable Li-Ion
Mechanical Specifications
Size40 mm (1.57")
Display TypeTouchscreen
Weight54 g (including watchband)
Environmental Specifications
Operating Temperature0 to 35 °C (32 to 95°F)
Operating Humidity10% to 95% RH (non-condensing)
Storage/Transport Temperature-20°C to 60°C (-4°F to 140°F)
Storage/Transport Humidity10% to 95% RH (non-condensing)
Classification per IEC 60601-1
Electrical SafetyIEC 60601-1
EMCIEC 60601-1-2
Electrical Isolation TypeInternally Powered
Applied Part TypeCF Applied Part
Ingress ProtectionIP24
Mode of OperationContinuous

2 Indications for Use

Masimo W1™ and the integrated Masimo W1 Module are intended to record, store and transfer single-channel electrocardiogram (ECG) rhythms. The Masimo W1 also displays ECG rhythms, and the Masimo W1 Module detects the presence of atrial fibrillation. The Masimo W1 and the integrated Masimo W1 Module are intended for use by healthcare professionals, patients with known or suspected heart conditions, and health-conscious individuals.

Masimo W1 Module ECG software is an over-the-counter (OTC) software that operates on the Masimo W1 Module that can be used with compatible watches (e.g., Masimo W1). The software is intended to create, record, store, transfer, and display a single channel electrocardiogram (ECG) for informational use only in adults 22 years and older. It supports the classification of either atrial fibrillation (AFib) or sinus rhythm with the intention of aiding heart rhythm identification; it is not intended to replace traditional methods of diagnosis or treatment. The software is not intended for users with other known arrhythmias and users should not interpret or take clinical action based on the device output without consultation of a qualified healthcare professional.

The Masimo W1™ and the integrated Masimo W1 Module are also intended for the spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The Masimo W1 and the integrated Masimo W1 Module are indicated for adults in hospitals, clinics, long-term care facilities, and homes.

K243305
Page 2 of 13

Page 7

510(k) Summary

3 Technological Characteristics

Principle of Operation

There were no changes made to the principles of operation from the previously cleared features of the Masimo W1 under K240229. However, as part of this submission, an AFib Classification Feature is added that expands upon the principles of operation for the ECG feature.

Electrocardiogram (ECG)

Same as under K240229, the ECG feature on Masimo W1 relies on the principle that the electrical signals can be detected as the heart contracts and relaxes during each cardiac cycle allowing for the estimation of heart rate (HR). The depolarization and repolarization of the heart muscle creates electrical signals that are propagated for recording on the skin surface at the wrist.

The Masimo W1 Module ECG software can also analyze these electrical signals for signs of AFib. To detect AFib, the software relies on the understanding that AFib rhythms are defined as "irregularly irregular" heartbeat patterns.

Pulse oximetry-based parameters

The Masimo W1 still relies on Masimo SET pulse oximetry, which relies on the Beer-Lambert law and the following principles of pulse oximetry to provide estimates of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate:

  • Oxyhemoglobin (oxygenated blood) and deoxyhemoglobin (non-oxygenated blood) differ in their absorption of red and infrared light (spectrophotometry).
  • The amount of arterial blood in tissue changes with your pulse (photoplethysmography). Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well.

Mechanism of Action for Achieving the Intended Effect

There were no changes to the mechanism of action for the Masimo W1 from the previous clearance under K240229.

The Masimo W1 still achieves its intended effect through the integrated Masimo W1 Module that provides both the electrical and optical sensing technology and parameter algorithms. The Masimo W1 Module is integrated into the Masimo W1 watch so that the sensing components contact the skin on the wrist. The watch is provided with an electrical sensing pad that is contacted by fingers on the opposing hand. Contacting the electrical sensing pad allows for recording of the ECG signal when the ECG feature is activated on the watch touchscreen. Once the ECG feature is activated, it detects and calculates the heart rate and determines the rhythm classification (e.g., Sinus Rhythm, Atrial Fibrillation).

K243305
Page 3 of 13

Page 8

510(k) Summary

The Masimo SET pulse oximetry-based parameters are supported by the Masimo W1 Module's optical sensing components located on the bottom of the module that contacts the skin on the wrist. The Masimo W1 Module continuously detects and processes the optical signals that change with the transmission of LED light into the wrist tissue. The Masimo W1 Module utilizes multiple wavelengths of light and advanced signal processing techniques to isolate the arterial signal from other static factors (e.g., skin pigment) to establish the ratio used in the estimation of the SpO2. The pulse rate is determined by the periodic changes in the photoplethysmograph (PPG).

The monitored parameters are then continuously updated and displayed on the watch so that it can be viewed, recorded and/or transferred. Masimo W1 is also provided with Bluetooth connectivity to support the wireless communication of the monitored data to a smart device application for supplemental display. The use of the Masimo W1 watch can be discontinued by taking off the watch or deactivating the continuous parameters from the watch's touchscreen.

4 Discussion of Similarities and Differences Between Primary Predicate and Subject Device

Similarities and Differences between Predicate and Subject Device

The subject device, Masimo W1 with AFib Classification Feature, and the predicate device, Masimo W1 (K240229), have the following key similarities:

  • Both devices have similiar intended use.
  • Both devices rely on the same principles of operation.

The subject device, Masimo W1 with AFib Classification Feature, and the predicate device, Masimo W1 (K240229), have the following key difference:

  • The subject device is provided with an AFib Classification Feature.

The key difference between the subject device and predicate device is the addition of the AFib Classification Feature. The intended use of this feature is substantially equivalent to the predicate devices. To support there are no technological characteristic differences that raise different questions of safety and effectiveness, clinical performance data of the AFib classification feature is provided as part of this submission. The test results support the substantial equivalence of the subject device.

K243305
Page 4 of 13

Page 9

510(k) Summary

Feature 510(k) NumberMasimo W1 Subject DeviceMasimo W1 Primary Predicate (K240229)Withing Scan Monitor Secondary Predicate (K201456)Samsung ECG Monitor Tertiary Predicate (K201168)Comparison
General Information
ClassificationClass II, ElectrocardiographClass II, ElectrocardiographClass II, ElectrocardiographClass II, Electrocardiograph software for Over-The-Counter UseSame.
Regulation/ Product Code21 CFR 870.2340, Class II/ DPS21 CFR 870.2340, Class II/ DPS21 CFR 870.2340, Class II/ DPS21 CFR 870.2345, Class II/ QDASame.
Additional Product Code(s)DQA DXH QDADQA DXHDQA DXHQDAThe subject device includes an additional product code QDA to support atrial fibrillation classification feature. The addition of the product code is the same as the tertiary predicate device that has the same intended use.
Intended UseMasimo W1™ and the integrated Masimo W1 module are intended to record, store and transfer single-channel electrocardiogram (ECG) rhythmsThe Masimo W1 and the integrated Masimo W1 module records, stores, transfers, and displays the single-channel electrocardiogram (ECG).The Scan Monitor is intended to record, store and transfer single-channel electrocardiogram (ECG) rhythms.The app is intended to create, record, store, transfer, and display a single channel electrocardiogram (ECG).Same.
Indications for UseMasimo W1 also displays ECG rhythms, and the Masimo W1 Module detects the presence of atrial fibrillation.The Masimo W1 and the integrated Masimo W1 module records, stores, transfers, and displays the single-channelThe Scan Monitor also displays ECG rhythms and detects the presence of atrial fibrillation (when the monitorThe Samsung ECG Monitor Application is an over-the-counter (OTC) software-only, mobile medicalDifferent. The subject device is indicated for an atrial fibrillation classification feature. The indication for the AFib feature of the subject device is similar to the secondary

Page 10

510(k) Summary

Feature 510(k) NumberMasimo W1 Subject DeviceMasimo W1 Primary Predicate (K240229)Withing Scan Monitor Secondary Predicate (K201456)Samsung ECG Monitor Tertiary Predicate (K201168)Comparison to Primary Predicate
The Masimo W1 and the integrated Masimo W1 Module are intended for use by healthcare professionals, patients with known or suspected heart conditions, and health-conscious individuals. Masimo W1 Module ECG software is an over-the-counter (OTC) software that operates on the Masimo W1 Module that can be used with compatible watches (e.g., Masimo W1). The software is intended to create, record, store, transfer, and display a single channel electrocardiogram (ECG) for informational use only in adults 22 years and older. It supports the classification of either atrial fibrillation (AFib) or sinus rhythm with the intention of aiding heart rhythm identification; it is not intended to replace traditional methods ofECG for the manual interpretation of heart rate. It is worn on the wrist and also provides other continuous parameters technologies (e.g., pulse oximetry). The Masimo W1™ and the integrated Masimo W1 module are also intended for the spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The Masimo W1 and the integrated Masimo W1 Module are indicated for adults in hospitals, clinics, long-term care facilities, and homes.is prescribed or used under the care of a physician). The Scan Monitor is intended for use by healthcare professionals, patients with known or suspected heart conditions and health-conscious individuals.application operating on a compatible Samsung Galaxy Watch and Phone. The app is intended to create, record, store, transfer, and display a single channel electrocardiogram (ECG), similar to a Lead I ECG, for informational use only in adults 22 years and older. Classifiable traces are labeled by the app as either atrial fibrillation (AFib) or sinus rhythm with the intention of aiding heart rhythm identification; it is not intended to replace traditional methods of diagnosis or treatment. The app is not intended for users with otherand tertiary predicate devices under the same intended use.

Page 11

510(k) Summary

Feature 510(k) NumberMasimo W1 Subject DeviceMasimo W1 Primary Predicate (K240229)Withing Scan Monitor Secondary Predicate (K201456)Samsung ECG Monitor Tertiary Predicate (K201168)Comparison to Primary Predicate
diagnosis or treatment. The software is not intended for users with other known arrhythmias and users should not interpret or take clinical action based on the device output without consultation of a qualified healthcare professional. The Masimo W1™ and the integrated Masimo W1 Module are also intended for the spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The Masimo W1 and the integrated Masimo W1 Module are indicated for adults in hospitals, clinics, long-term care facilities, and homes.known arrhythmias and users should not interpret or take clinical action based on the device output without consultation of a qualified healthcare professional.
Technological Characteristics
Principles of OperationThe ECG feature on Masimo W1 relies on the principle that the electrical signals can be detected as the heart contracts andThe ECG feature on Masimo W1 still relies on the principle that the electrical signals can beThe iECG software analyzes the patterns in the ECG waveform to determine differentUnavailable.Same.

Page 12

510(k) Summary

Feature 510(k) NumberMasimo W1 Subject DeviceMasimo W1 Primary Predicate (K240229)Withing Scan Monitor Secondary Predicate (K201456)Samsung ECG Monitor Tertiary Predicate (K201168)Comparison to Primary Predicate
relaxes during each cardiac cycle allowing for the estimation of heart rate (HR). The depolarization and repolarization of the heart muscle creates electrical signals that are propagated for recording on the skin surface at the wrist. The Masimo W1 Module ECG software can also analyze these electrical signals for signs of AFib. To detect AFib, the software relies on the understanding that AFib rhythms are defined as "irregularly irregular" heartbeat patterns. The Masimo W1 still relies on Masimo SET pulse oximetry, which relies on the Beer-Lambert law and the following principles of pulse oximetry to provide estimates of functional oxygen saturation ofdetected as different parts of the heart contract and relax during a cardiac cycle allowing the detection of heart activity and for the estimation of heart rate (HR). The change in polarization of the heart muscles creates the electrical signals that are propagated so that they can be detected at the skin surface of the wrist. The Masimo SET pulse oximeter technology relies on the Beer-Lambert law and the following principles of pulse oximetry: • Oxyhemoglobin (oxygenated blood) andtypes of known heart activity (e.g., Heart Rate, Atrial Fibrillation). Scan Monitor pulse oximetry technology relies on the differential absorption by blood of red (660nm) and infrared light (940nm) which relies on the Beer-Lambert law for the principles of pulse oximetry.

Page 13

510(k) Summary

Feature 510(k) NumberMasimo W1 Subject DeviceMasimo W1 Primary Predicate (K240229)Withing Scan Monitor Secondary Predicate (K201456)Samsung ECG Monitor Tertiary Predicate (K201168)Comparison to Primary Predicate
arterial hemoglobin (SpO2) and pulse rate: • Oxyhemoglobin (oxygenated blood) and deoxyhemoglobin (non-oxygenated blood) differ in their absorption of red and infrared light (spectrophotometry). • The amount of arterial blood in tissue changes with your pulse (photoplethysmography). Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well.deoxyhemoglobin (non-oxygenated blood) differ in their absorption of red and infrared light (spectrophotometry). • The amount of arterial blood in tissue changes with your pulse (photoplethysmography). Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well.
Supported Measured Pulse Oximetry

Page 14

510(k) Summary

Feature 510(k) NumberMasimo W1 Subject DeviceMasimo W1 Primary Predicate (K240229)Withing Scan Monitor Secondary Predicate (K201456)Samsung ECG Monitor Tertiary Predicate (K201168)Comparison to Primary Predicate
Parameters
Supported Measured Pulse OximetrySpO2, PRSpO2, PRSpO2, PRN/ASame.

Page 15

510(k) Summary

Feature 510(k) NumberMasimo W1 Subject DeviceMasimo W1 Primary Predicate (K240229)Withing Scan Monitor Secondary Predicate (K201456)Samsung ECG Monitor Tertiary Predicate (K201168)Comparison to Primary Predicate
Parameters
Supported Measured ECG ParametersHRHRHR, HRVHRSame.
Supported Rhythm ClassificationsAtrial Fibrillation, Sinus RhythmN/AAtrial Fibrillation, Sinus RhythmAtrial Fibrillation, Sinus RhythmDifferent. The subject device provides an Atrial Fibrillation classification feature similar to the secondary and tertiary predicate devices. Testing is provided to support the substantial equivalence.
Supported Calculated featuresPiPiN/AN/ASame.
LeadA single channel ECG, similar to Lead 1 ECG.A single channel ECG, similar to Lead 1 ECG.A single channel ECG, similar to Lead 1 ECG.A single channel ECG, similar to Lead 1 ECG.Same.
Performance specifications
SpO2, (70-100%)2%, adults (No Motion/ Low Perfusion)2%, adults (No Motion/ Low Perfusion)3%, adults (No Motion)N/ASame.
Pulse Rate (25-240 bpm)3 bpm3 bpmNot supportedN/ASame.
Heart Rate (25-240 bpm)5 bpm5 bpmNot KnownN/ASame.
Rhythm ClassificationAtrial Fibrillation Sensitivity: 99.3%N/AAtrial Fibrillation Sensitivity: 96.3%Atrial Fibrillation Sensitivity: 98.1%AFib feature is being added. The performance has been evaluated to be

Page 16

510(k) Summary

Feature 510(k) NumberMasimo W1 Subject DeviceMasimo W1 Primary Predicate (K240229)Withing Scan Monitor Secondary Predicate (K201456)Samsung ECG Monitor Tertiary Predicate (K201168)Comparison to Primary Predicate
PerformanceSinus Rhythm Specificity: 100%Sinus Rhythm Specificity: 100%Sinus Rhythm Specificity: 100%substantially equivalent to the secondary and tertiary predicates.
Electrical Specifications
BatteryInternal RechargeableInternal RechargeableInternal RechargeableN/ASame.
Mechanical Specifications
Watch Face Size40 mm (1.57")40 mm (1.57")38 mm or 42 mmN/ASame.
Weight54 g (including watchband)54 g (including watchband)58 g or 83gN/ASame.
Environmental Specifications
Operating Temperature0 to 35 °C (32 to 95°F)0 to 35 °C (32 to 95°F)-10 to 45 °C (14 to 113°F)N/ASame.
Operating Humidity10% to 95%, non-condensing10% to 95%, non-condensingNot KnownN/ASame.
Storage/Transport Temperature-20 to 60°C (-4 to 140°F)-20 to 60°C (-4 to 140°F)-20 to 85°C (4 to 185°F)N/ASame.
Storage/Transport Humidity10% to 95%, non-condensing10% to 95%, non-condensingNot KnownN/ASame.
Classification per IEC 60601-1
Electrical SafetyIEC 60601-1IEC 60601-1IEC 60601-1N/ASame.
EMCIEC 60601-1-2IEC 60601-1-2IEC 60601-1-2N/ASame.
Ingress ProtectionIP24IP24Not Known.N/ASame.
Mode of OperationContinuousContinuousContinuousN/ASame.

Page 17

510(k) Summary

5 Performance Data

As part of this submission, clinical and non-clinical testing was included to support addition of the Atrial Fibrillation Classification Feature.

Performance Bench Testing

Bench testing for the Masimo W1 was included in this submission to support the addition of the Atrial Fibrillation Classification Feature.

Biocompatibility Testing

As there were no changes made to the patient contacting materials of the subject device from the previous clearance under K240229, no new biocompatibility testing was included in this submission.

Electromagnetic Compatibility, Electrical Safety, Environmental, Mechanical and Cleaning

As there were no hardware changes made to the subject device from the previous clearance under K240229, no new electrical safety, environmental, mechanical, and cleaning testing was included as part of this submission.

Software Verification and Validation Testing

Software verification and validation testing were conducted and provided as part of this submission as recommended by the FDA Guidance, "Content of Premarket Submissions for Software Device Software Functions", June 2023 to support the addition of the AFib detection feature.

The Masimo W1 software was found to fit the category of products that would require Basic Documentation Level because the failure or latent flaw of the device software function would not present a hazardous situation with a probable risk of death or serious injury to either a patient, user of the device or others in the environment of use prior to the implementation of the risk controls.

Wireless Testing

As there were no changes made to the subject device from the previous clearance under K240229, no new wireless testing was included as part of this submission.

Cybersecurity Testing

As there were no changes made to the subject device wireless communication from the previous clearance under K240229, thus no new cybersecurity testing was included as part of this submission.

Human Factors and Usability Testing

Human factors testing demonstrated the acceptability of the human factors and usability risks.

K243305
Page 12 of 13

Page 18

510(k) Summary

Clinical Testing

To support the performance of the Masimo W1 with Atrial Fibrillation Classification Feature, clinical validation testing is provided as part of this submission.

Masimo W1 ECG AFib Classification feature Performance

Prospective clinical testing was conducted to validate the AFib Classification Feature on adult subjects from 4 different sites. The testing included subjects with wide range of age and BMI.

The breakdown of the performance results, including the 95% Confidence Interval, are provided below in Table 5-1.

Table 5-1

Performance MetricsPerformance Result
Sensitivity99.3% [96.3%,100%]
Specificity100% [97.8%, 100%]
PPV100% [97.5%, 100%]
Unclassified Rate5.0%
Noise Rate1.7%

Masimo W1 ECG Waveform Quality Analysis

To assess the quality of Masimo W1 ECG waveform in adults, the ECG waveform was accessed qualitatively and quantitatively by comparing similarity between Masimo W1 and gold-standard 12 lead ECG as reference. The qualitative assessment found a 98% [95% CI: 96% -98%] agreement by qualified clinicians in the analysis of the Masimo W1 ECG waveforms to that of the reference Lead I of a gold-standard 12-lead ECG reference.

The quantitative analysis showed key ECG features (QRS amplitude and QRS width) of the Masimo W1 ECG waveforms were similar to Lead I of a gold-standard 12-lead ECG reference.

6 Conclusion

The data supported the substantial equivalence of the subject device.

K243305
Page 13 of 13

§ 870.2340 Electrocardiograph.

(a)
Identification. An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.(b)
Classification. Class II (performance standards).