K Number
K243291

Validate with FDA (Live)

Date Cleared
2025-01-17

(91 days)

Product Code
Regulation Number
884.2675
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Goldtrace FSE is indicated for patients requiring continuous electronic intrapartum monitoring of the fetal heart rate during labor and delivery.

Device Description

Goldtrace FSE is a sterile, single use device transmitting the fetal ECG signal for continuous electronic intrapartum monitoring. The device consists of gold-plated stainless steel spiral needle electrode that attaches to the fetal presenting part. The device includes a reference plate to measure the reference voltage level in the amniotic fluid. A guide tube and drive tube are used for application to the fetal presenting part. The subject device is connected to a compatible monitor (e.g., Philips Avalon FM30) via Philips DECG Reusable Legplate adaptor cable to provide a fetal ECG signal to the monitor. The subject device does not perform any conversion, analysis, quantification, or modifications to the fetal ECG.

AI/ML Overview

This document is a 510(k) Summary for the Goldtrace Fetal Spiral Electrode (FSE), which is a medical device. The summary outlines the device's indications for use, its comparison to a predicate device, and the non-clinical performance data provided to support its substantial equivalence.

Here's an analysis to answer your questions:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state "acceptance criteria" with numerical targets for the comparative performance testing. Instead, it states that the devices are "equivalent" and "perform equally."

For other performance areas, the acceptance criteria are implied by adherence to recognized standards and successful completion of tests.

Acceptance Criteria CategoryReported Device Performance
Biocompatibility- Cytotoxicity: Non-cytotoxic (ISO 10993-5:2009)
- Sensitization: Non-sensitizing (ISO 10993-10:2010)
- Irritation: Non-irritating (ISO 10993-10:2010)
- Systemic Toxicity: Not systemically toxic (ISO 10993-11:2006)
- Pyrogenicity: Non-pyrogenic (USP38-NF33:2015 <151>)
Electrical Safety & EMCComplies with IEC 60601-1:2020 and IEC 60601-1-2:2014 + AMD:2020. Also tested (with deviations consistent with device's intended use) according to ANSI/AAMI EC12:2000/(R)2015 and ANSI/AAMI EC53:2013/(R)2020.
Sterility & Shelf-Life- Sterilization: Validated per ISO 11135:2014 and FDA guidance.
- Transportation: Passed simulated transportation and package integrity testing per ASTM D4169-22.
- Shelf-life: 5 years (accelerated aging per ASTM F1980:21, including package integrity tests: dye penetration per ASTM F1929-15, seal strength per ASTM F88/F88M-21, visual assessment per ASTM F1886/F1886M-16).
Mechanical Testing- Pull Apart Test: Demonstrated successful performance for Goldtrace FSE connector and Legplate.
- Three-point Bend Test: Demonstrated material consistency and pliability for drive tube and guide tube.
Comparative PerformanceGoldtrace FSE and predicate device (K030691) are equivalent in electrical properties, in their ability to transmit fetal ECG signals, and perform equally when used with Philips Avalon FM30 fetal monitor.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document is a 510(k) summary for non-clinical performance data. It does not provide details about clinical studies, test set sample sizes, or data provenance in terms of patient data. The "test sets" for the non-clinical tests would refer to the number of device units tested for each specific test (e.g., how many devices were tested for biocompatibility, electrical safety, mechanical strength), which is not specified in this summary. The data provenance is from non-clinical testing, likely conducted by the manufacturer or accredited labs on their behalf. The country of origin for the data is not explicitly stated, but the manufacturer is Neoventa Medical AB in Sweden.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This document is about a hardware medical device (fetal spiral electrode) and its non-clinical performance. It does not involve interpretation of medical images or data by human experts for establishing ground truth in the way an AI algorithm for diagnostics would. Therefore, this question is not applicable to the information provided. The ground truth for the device's function is its ability to transmit fetal ECG signals accurately, which is assessed through instrumental measurements rather than expert consensus on medical findings.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

As this document describes non-clinical performance testing of a hardware device, there is no adjudication method involving multiple human readers as would be typical for clinical studies or AI diagnostic algorithms. Performance is assessed against technical standards and comparisons to a predicate device via instrumental measurements.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not done. This is a 510(k) summary for a hardware medical device (fetal spiral electrode), not an AI-powered diagnostic tool. The document focuses on the device's physical and electrical performance, and its equivalence to a predicate device, not on human reader performance with or without AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable. The Goldtrace Fetal Spiral Electrode is a hardware device described as transmitting a fetal ECG signal to a monitor. The document explicitly states: "The subject device does not perform any conversion, analysis, quantification, or modifications to the fetal ECG." It does not contain an algorithm for standalone performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For this device, the "ground truth" for its performance is established by:

  • Adherence to recognized technical standards: This includes biocompatibility, electrical safety, sterility, and mechanical integrity standards (e.g., ISO, IEC, ASTM, USP). The "truth" is whether the device meets the specifications and limits defined by these standards.
  • Functional equivalence to a predicate device: For the comparative performance testing, the ground truth is that the device should achieve the "same" or "equivalent" performance in transmitting fetal ECG signals as the legally marketed predicate device (K030691) when used with a compatible monitor. This equivalence is assessed through direct electrical and signal integrity measurements, not a human-derived ground truth.

8. The sample size for the training set

This question is not applicable. This device is a hardware medical device and does not involve AI algorithms that require a "training set."

9. How the ground truth for the training set was established

This question is not applicable, as there is no training set for this hardware device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

January 17, 2025

Neoventa Medical AB Johan Sundberg Chief Quality and Regulatory Officer Norra Ågatan 32 Mölndal, Västra Götaland SE-431 35 SWEDEN

Re: K243291

Trade/Device Name: Goldtrace Fetal Spiral Electrode (FSE) (CNS000004) Regulation Number: 21 CFR 884.2675 Regulation Name: Fetal Scalp Circular (Spiral) Electrode and Applicator Regulatory Class: II Product Code: HGP Dated: October 18, 2024 Received: October 18, 2024

Dear Johan Sundberg:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

{2}------------------------------------------------

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Monica D. Garcia -S

Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K243291

Device Name

Goldtrace Fetal Spiral Electrode (FSE) (CNS000004)

Indications for Use (Describe)

Goldtrace FSE is indicated for patients requiring continuous electronic intrapartum monitoring of the fetal heart rate during labor and delivery.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Neoventa Medical. The logo consists of the word "neoventa" in a teal sans-serif font, followed by the tagline "A better start in life" in a smaller font. To the right of the text is a teal graphic of a mother and child in an embrace. The text "Page 1 of 4" is in the upper left corner of the image.

K243291

510(k) Summary - Goldtrace Fetal Spiral Electrode (K243291)

I. SUBMITTER

Neoventa Medical AB Norra Agatan 32 SE-431 35 Mölndal, Sweden Phone: +46(0)31 758 32 00 Fax: +46(0)31 758 32 99 Date Summary Prepared: January 17, 2025

Contact Person: Johan Sundberg Chief Quality and Regulatory Officer Email: johan.sundberg@neoventa.com

II. DEVICE

Device Trade Name: Goldtrace Fetal Spiral Electrode (Goldtrace FSE), CNS000004 Device Common Name: Fetal Spiral Electrode (FSE) Regulation number: 21 CFR 884.2675 Regulation Name: Fetal scalp circular (spiral) electrode and applicator Regulatory Class: II Product code: HGP (electrode, circular (spiral), scalp and applicator) Classification Panel: Obstetrics/Gynecology

III. PREDICATE DEVICE

Device Name: Fetal Spiral Electrode Manufacturer: Clinical Innovations, Inc 510(k) Number: K030691

The predicate device has not been subject to a design-related recall.

IV. DEVICE DESCRIPTION

Goldtrace FSE is a sterile, single use device transmitting the fetal ECG signal for continuous electronic intrapartum monitoring.

The Goldtrace FSE is intended for use in fetuses and mothers after rupture of the amniotic membranes requiring continuous electronic intrapartum monitoring of the fetal heart rate or analysis of the fetal ECG waveform. Goldtrace FSE is intended to be used by medically trained personnel at delivery wards.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Neoventa Medical. The logo includes the company name in a stylized font, with the tagline "A better start in life" underneath. To the right of the name is an image of a mother and child.

The device consists of gold-plated stainless steel spiral needle electrode that attaches to the fetal presenting part. The device includes a reference plate to measure the reference voltage level in the amniotic fluid. A guide tube and drive tube are used for application to the fetal presenting part. The subject device is connected to a compatible monitor (e.g., Philips Avalon FM30) via Philips DECG Reusable Legplate adaptor cable to provide a fetal ECG signal to the monitor. The subject device does not perform any conversion, analysis, quantification, or modifications to the fetal ECG.

The Goldtrace FSE materials conform to FDA recognized consensus standard for medical use (ISO 10993-1: 2018) and the device is sterilized using Ethylene Oxide sterilization (EO).

  • V. INDICATIONS FOR USE
    Goldtrace FSE is indicated for patients requiring continuous electronic intrapartum monitoring of the fetal heart rate during labor and delivery.

COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS VI. WITH THE PREDICATE DEVICE

The subject device and predicate device are intended to be used on patients during labor requiring electronic intrapartum monitoring of the fetal heart rate by means of a spiral electrode.

Following table provides comparison of the intended use and technological characteristics of the subject device with the predicate device:

SpecificationSubject Device (K243291)Predicate Device (K030691)
Intended UseGoldtrace FSE is indicated forpatients requiring continuouselectronic intrapartum monitoring ofthe fetal heart rate during labor anddelivery.For patients requiring fetal heartrate monitoring during labor.
Principle of OperationTransmit fetal ECG signal forrecording, analysis andquantification of the monitorTransmit fetal ECG signal forrecording, analysis, andquantification of the monitor
Electrode DesignSingle spiral needle with twistedtwo-wire concept. Guide tube anddrive tube used for application.Single spiral needle with twistedtwo-wire concept. Guide tube anddrive tube used for application.
FSE componentsSpiral needle, Reference Plate,Connector, Guide Tube, Drive Tube,Drive Tube Grip, WiresSpiral needle, Reference Plate,Connector, Guide Tube, DriveTube, Drive Tube Grip, Wires
Materials used forspiral electrodeSS316 stainless steel for electrodeswith 10-25 µm gold-plated surfaceStainless steel for electrodes
Spiral needle diameterDiameter: 4.7mm (0.185 inches) ±0.13mm (0.005 inches) (O.D.)Outside Diameter: 0.55 mm (0.021inches) ± 0.05 mm (0.002 inches)Not publicly available
HubOutside diameter:0.219 inches ± 0.005 inches (O.D.)Not publicly available
WiresLength: 60 cm (23.6 inches) ± 2 cmNot publicly available

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the document number K243291 and page number 3 of 4. Below this is the logo for Neoventa, which is written in a teal color. Underneath the logo is the text "A better start in life".

Shelf life5 yearsNot publicly available
SterileYesYes
Single useYesYes

The subject and predicate device have similar indications for use statements and have the same intended use - continuous intrapartum monitoring of fetal ECG. The subject device (Goldtrace Fetal Spiral Electrode) has similar fundamental scientific technology to the predicate device (Fetal Scalp Electrode). The subject device differs from the predicate device in terms of device materials, dimensions, and electrical properties. The different technological features of the subject device do not raise different questions of safety and effectiveness as compared to the predicate device.

VII. SUMMARY OF NON-CLINICAL PERFORMANCE DATA

The following performance data are provided in support of the substantial equivalence determination.

Biocompatibility

Biological evaluation of the subject device has been performed in accordance with the 2024 FDA guidance "Use of International standard ISO 10993-1, Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process" as well as ISO 10993- 1:2018 "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". Biocompatibility of the subject device is supported by performed tests of the following biological endpoints:

  • . Cytotoxicity (ISO 10993-5:2009)
  • . Sensitization (ISO 10993-10:2010)
  • Irritation (ISO 10993-10:2010)
  • Systemic Toxicity (ISO 10993-11:2006)
  • Material Mediated Pyrogenicity (USP38-NF33:2015 <151> Pyrogen Test) .

The subject device was found to be non-cytotoxic, non-sensitizing, non-irritating, not systemically toxic, and non-pyrogenic (material mediated).

Electrical Safety and Electromagnetic Compatibility

Testing was conducted in accordance with IEC 60601- 1:2020, Medical electrical equipment -Part 1: General requirements for basic safety and essential performance, and IEC 60601-1-2:2014 + AMD:2020, Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests, to support the electrical safety and electromagnetic compatibility of the subject device. Device was also tested (with deviations consistent with the device's intended use) according ANSI/AAMI EC12 EC12:2000/(R)2015 - Disposable ECG electrodes and ANSI/AAMI EC53 2013/(R)2020 - ECG trunk cables and patient leadwires.

Sterility and Shelf-Life

Following testing was conducted to support sterility and shelf-life of the subject device:

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Neoventa Medical. The logo includes the text "neoventa" in a teal color, followed by the tagline "A better start in life" in a smaller font. To the right of the text is a stylized image of a mother and child in a teal color. The image also includes the text "K243291 Page 4 of 4" in the upper left corner.

  • . Ethylene Oxide (EO) sterilization and validation testing per ISO 11135:2014 and the 2024 FDA guidance "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile Guidance for Industry and Food and Drug Administration Staff".
  • . Simulated transportation and associated package integrity testing in accordance with ASTM D4169-22.
  • Shelf-life testing (accelerated aging) per ASTM F1980:21, including package integrity ● testing (dye penetration test per ASTM F1929-15, seal strength testing per ASTM F88/F88M-21 and visual assessment per ASTM F1886/F1886M-16).

Mechanical Testing

Testing to support structural and mechanical specifications of the subject device was conducted, including pull apart test for the Goldtrace FSE connector and Legplate and threepoint bend test for the drive tube and the guide tube. The bend tests demonstrated the material consistency over batches and that the material in the tubes was pliable within the load range.

Comparative Performance Testing

A comparative test was performed between the Goldtrace FSE and the predicate device (K030691). The two devices are equivalent in electrical properties, in their ability to transmit fetal ECG signals and perform equally when used together with Philips Avalon FM30 fetal monitor.

CONCLUSION VIII.

The performance data described above demonstrate that the Goldtrace FSE is as safe and effective as the predicate device, Fetal Spiral Electrode (K030691) and supports determination of substantial equivalence.

§ 884.2675 Fetal scalp circular (spiral) electrode and applicator.

(a)
Identification. A fetal scalp circular (spiral) electrode and applicator is a device used to obtain a fetal electrocardiogram during labor and delivery. It establishes electrical contact between fetal skin and an external monitoring device by a shallow subcutaneous puncture of fetal scalp tissue with a curved needle or needles. This generic type of device includes nonreusable spiral electrodes and reusable circular electrodes.(b)
Classification. Class II (performance standards).