K Number
K243235

Validate with FDA (Live)

Date Cleared
2025-03-03

(144 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

LVivo platform is intended for non-invasive processing of ultrasound images to detect, measure, and calculate relevant medical parameters of structures and function of patients with suspected disease. In addition, it has the ability to provide Quality Score feedback.

Device Description

The LVivo platform is a software system for automated analysis of ultrasound examinations. Automated analysis of echocardiographic examinations is done using DICOM movies. The LVivo platform supports global and segmental evaluation of the left ventricle (LV) of the heart. The global LV function is evaluated from two of the apical views: four-chamber (4CH) and two-chamber (2CH) by ejection fraction (EF). The segmental LV function is done from three apical views 4CH, 2CH and three chamber (3CH) and supports wall motion evaluation and strain. LVivo CE-EF (Contrast EF) extends the current toolbox of the LVivo platform by providing the ability to process Ultrasonic DICOM images which acquire by Ultrasound Equipment in which the patient was prescribed a contrast agent. In addition to the LV analysis, the cardiology toolbox includes a module for automated evaluation of the Right Ventricular function. The LVivo platform includes one additional non-cardiac module for the measurement of the bladder volume.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

Acceptance Criteria and Reported Device Performance

ParameterAcceptance Criteria (Implied by Study Design)Reported Device Performance
EDV 4CH Correlation (Automated vs. Manual)High correlation (implicitly, r > 0.90)r = 0.96, CI [0.94, 0.97]
EDV 4CH Bland-Altman Mean Difference (Bias)Close to 0 (implicitly, within acceptable clinical limits)-2.68 ml
EDV 4CH Bland-Altman Limits of Agreement (LOA)Narrow range (implicitly, clinically acceptable spread)(-36.02, 30.66)
ESV 4CH Correlation (Automated vs. Manual)High correlation (implicitly, r > 0.90)r = 0.98, CI [0.97, 0.99]
ESV 4CH Bland-Altman Mean Difference (Bias)Close to 0-3.87 ml
ESV 4CH Bland-Altman Limits of Agreement (LOA)Narrow range(-25.58, 17.82)
EF 4CH Correlation (Automated vs. Manual)High correlation (implicitly, r > 0.90)r = 0.95, CI [0.93, 0.97]
EF 4CH Bland-Altman Mean Difference (Bias)Close to 01.26% points
EF 4CH Bland-Altman Limits of Agreement (LOA)Narrow range(-8.42, 10.96)
EDV 2CH Correlation (Automated vs. Manual)High correlation (implicitly, r > 0.90)r = 0.96, CI [0.94, 0.97]
EDV 2CH Bland-Altman Mean Difference (Bias)Close to 0-5.69 ml
EDV 2CH Bland-Altman Limits of Agreement (LOA)Narrow range(-36.02, 24.44)
ESV 2CH Correlation (Automated vs. Manual)High correlation (implicitly, r > 0.90)r = 0.98, CI [0.97, 0.99]
ESV 2CH Bland-Altman Mean Difference (Bias)Close to 0-3.87 ml
ESV 2CH Bland-Altman Limits of Agreement (LOA)Narrow range(-25.58, 17.82)
EF 2CH Correlation (Automated vs. Manual)High correlation (implicitly, r > 0.90)r = 0.93, CI [0.90, 0.95]
EF 2CH Bland-Altman Mean Difference (Bias)Close to 0-0.54% points
EF 2CH Bland-Altman Limits of Agreement (LOA)Narrow range(-12.18, 11.1)
BP EDV Correlation (Automated vs. Manual)High correlation (implicitly, r > 0.90)r = 0.98, CI [0.97, 0.99]
BP EDV Bland-Altman Mean Difference (Bias)Close to 0-4.1 ml
BP EDV Bland-Altman Limits of Agreement (LOA)Narrow range(-29.04, 20.84)
BP ESV Correlation (Automated vs. Manual)High correlation (implicitly, r > 0.90)r = 0.98, CI [0.97, 0.99]
BP ESV Bland-Altman Mean Difference (Bias)Close to 0-2.77 ml
BP ESV Bland-Altman Limits of Agreement (LOA)Narrow range(-19.67, 14.13)
BP EF Correlation (Automated vs. Manual)High correlation (implicitly, r > 0.90)r = 0.96, CI [0.94, 0.97]
BP EF Bland-Altman Mean Difference (Bias)Close to 00.39% points
BP EF Bland-Altman Limits of Agreement (LOA)Narrow range(-8.26, 9.05)
Automatic Processing RateHigh rate (implicitly, > 80%)90% (91/101 exams)

Study Information

  1. Sample size used for the test set and the data provenance:

    • Sample Size: 101 patient exams.
    • Data Provenance: The data was collected from multiple sources: Beth Israel, Soroka, Hadassah (presumably Israel, based on the manufacturer's location), and UCMC (likely a US center given the race information collection). The text indicates that 69 of the 101 patients were collected from the US. The study is retrospective, as it refers to collected "patient exams."
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: 3 sonographers.
    • Qualifications: The text explicitly states "3 sonographers" were used. While their specific years of experience are not mentioned, the implication is that they are qualified medical professionals capable of performing and interpreting ultrasound measurements. The subsequent review by a cardiologist also suggests oversight and validation of their measurements.
  3. Adjudication method for the test set:

    • The ground truth was "comprised of the measurements by the 3 sonographers." This suggests a consensus or averaging approach among the three sonographers. It further states, "No further changes by the cardiologist to the measurements were needed following the cardiologist's review," implying the cardiologist reviewed and implicitly approved the sonographers' consensus measurements. This could be interpreted as a form of expert consensus adjudication, where the sonographers' measurements formed the basis, and a cardiologist provided final validation.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance was not described. This study directly compared the algorithm's performance to human (sonographer) ground truth measurements rather than assessing human reader improvement with AI assistance.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, a standalone performance evaluation of the algorithm was done. The study compares the "automated results" of the LVivo Software Application to the established "manual measurements" (ground truth). The fact that the algorithm "processed automatically 91/101 (90%) of the exams" and these automated results were compared to the manually derived ground truth confirms a standalone evaluation.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth used was expert consensus based on "manual measurements" performed by "3 sonographers," which were then reviewed and confirmed by a cardiologist.
  7. The sample size for the training set:

    • The document does not provide the sample size for the training set. It only discusses the performance evaluation using the test set of 101 patient exams.
  8. How the ground truth for the training set was established:

    • The document does not provide any information on how the ground truth for the training set was established. The focus of this submission is on the performance evaluation of the final device using a designated test set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

DiA Imaging Analysis Ltd. % George Hattub Medical Device Regulatory Affairs Specialist Medicsense USA LLC 291 Hillside Avenue Somerset, Massachusetts 02726

March 3, 2025

Re: K243235

Trade/Device Name: LVivo Software Application Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: OIH Dated: February 3, 2025 Received: February 3, 2025

Dear George Hattub:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Re"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See

{2}------------------------------------------------

the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb

Jessica Lamb, Ph.D. Assistant Director DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K243235

Device Name

LVivo Software Application

Indications for Use (Describe)

L Vivo platform is intended for non-invasive processing of ultrasound images to detect, measure, and calculate relevant medical parameters of structures and function of patients with suspected disease. In addition, it has the ability to provide Quality Score feedback.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K243235 510(k) Summary

Pursuant to CFR 807.92, the following 510(k) Summary is provided:

1. (a)SubmitterAddress:George J. HattubMedicsense USA LLC291 Hillside AvenueSomerset, MA 02726ghattub@comcast.net
1. (b)ManufacturerAddress:DiA Imaging Analysis LtdHaEnergia Street 77Beer-Sheva, Israel 8470912
Mfg. Phone:Tel.: +972 77 7648318
Contact Person:Mrs. Michal Yaacobi
Date:February 28, 2025
2.Device &ClassificationName:Medical Image Management and Processing System -classified as Class 2 QIH, Regulation Number 21 CFR 892.2050LVivo Software Application
3.Predicate Device:Reference Device:LVivo Software Application: K240553ASUSON Sequoia and Select Diagnostic Ultrasound System: K232145
4.Description:The LVivo platform is a software system for automated analysis ofultrasound examinations. Automated analysis of echocardiographicexaminations is done using DICOM movies. The LVivo platform supportsglobal and segmental evaluation of the left ventricle (LV) of the heart. Theglobal LV function is evaluated from two of the apical views: four-chamber(4CH) and two-chamber (2CH) by ejection fraction (EF). The segmental LVfunction is done from three apical views 4CH, 2CH and three chamber(3CH) and supports wall motion evaluation and strain.
LVivo CE-EF (Contrast EF) extends the current toolbox of the LVivo platformby providing the ability to process Ultrasonic DICOM images which acquireby Ultrasound Equipment in which the patient was prescribed a contrasagent.
In addition to the LV analysis, the cardiology toolbox includes a module forautomated evaluation of the Right Ventricular function. The LVivo platformincludes one additional non-cardiac module for the measurement of thebladder volume.
5.Indications forUse:LVivo platform is intended for non-invasive processing of ultrasoundimages to detect, measure, and calculate relevant medical parameters of

{5}------------------------------------------------

structures and function of patients with suspected disease. In addition, it has the ability to provide Quality Score feedback.

    1. Comparison of With respect to technology and intended use, DiA's LVivo Software Technological Application is substantially equivalent to its predicate devices. Based upon Characteristics: the outcomes from the risk analysis and Performance Testing Evaluation, DiA believes that the addition of LVivo Contrast EF module to the LVivo Software Application predicate device does not raise additional safety of efficacy concerns. The following comparison table depicts the changes.
Submitted DevicePredicate Device
Features/CharacteristicsLVivo SoftwareApplicationLVivo Software Application
Product CodesameQIH
Indication for UsesameLVivo platform isintended for non-invasive processingof ultrasoundimages to detect,measure, andcalculate relevantmedical parameters ofstructures andfunction of patientswith suspected disease. Inaddition, it has the ability toprovide Quality Scorefeedback.
ModulesLVivo CE-EF, LVivoEF, LVivo PLAX, LVivoSG, LVivo SAX, LVivoRV, LVivo Seamless,LVivo IQS and LVivoBladderLVivo EF, LVivo PLAX, LVivoSG, LVivo SAX, LVivo RV,LVivo Seamless, LVivo IQSand LVivo Bladder
Automationsameyes
Manual Adjustmentsameyes
Bi plane EF evaluationsameyes
Simultaneous 2CH and4CH evaluationsameyes
Off-line LV RV andBladder evaluation usingDICOM clips of anyvendorsameyes
Automated ED and ESframes selectionsameyes
Dynamic left ventricularsameyes
Manual editing byuser capabilitysameyes
Visually confirmresultssameyes
Automated rejectionof false resultssameyes
Volume calculation bystandard Simpson'smethod of discs for EFsameyes
Volume curvePresentationsameyes
EF, Strain, SWM, RV,SAX, Bladder resultspresentationsameyes
Enables presentation ofcardiac function resultsfor different cyclessameyes
AlgorithmAdded Supportfor Contrast EFyes
Calculation speedsameyes
Capability or a part of abigger package (device)for LV functionevaluation and Bladdersameyes
Segmental LongitudinalStrain Measuresameyes
Global LongitudinalStrain Measuresameyes
Segmental wall motionevaluationsameyes
Operating SystemWindow 10 orhigherWindows/Linux
510(k) #K243235K240553

{6}------------------------------------------------

{7}------------------------------------------------

Performance Evaluation:

A summary of the Performance Evaluation, which was based upon wellestablished test methods, demonstrated conformity to the intended use.

The overall patient population included 58% patients with Normal LV function 2% patients with preserved LV function and 40% patients with Abnormal LV function, among them 10% with mildly impaired LV function, 14% with moderately impaired LV function and 16% with Severely impaired LV function. LV was enlarged in 26% of the patients. 32% of the patients had LV hypertrophy.

Gender and Age were available for 94 (93%) of the patients. 52/94 (55%) were female and 59/94 male (62%). Age ranged from 20 to 100 years of aqe. For 47 patient exams collected from Beth Israel, Soroka and Hadassah the average age was 66±16.4 years old, average BMI was 28±8, range 13 to 56. For 47 patients reported from UCMC the average age was 63±15.8 years old, average BMI was 31.98, range 17.9 to 54. Race information was available for total of 69/101 (69%) patient examinations collected from US, among them 21/69 (30%) were indicated as Black or African Americans, 46/69 (67%) were indicated as white and 2/69 (3%) were Asian.

Of the 101 patient exams that were included in the study, the algorithm processed automatically 91/101 (90%) of the exams. The final ground truth was comprised of the measurements by the 3 sonographers. No further changes by the cardiologist to the measurements were needed following the cardiologist's review. The automated results were compared to the ground truth. The comparison between automated results and manual measurements for EDV, ESV and EF for A4CH, A2CH views separately and biplane (BP) are presented below (BA=Bland-Altman):

EDV 4CH correlation of r=0.96, CI [0.94, 0.97] was obtained between automated and manual results, BA: -2.68ml, LOA (-36.02, 30.66).

ESV 4CH correlation of r=0.98, CI [0.97, 0.99] was obtained between automated and manual results, BA: -3.87ml, LOA (-25.58, 17.82).

EF 4CH correlation of r=0.95, CI [0.93, 0.97] was obtained between automated and manual results, BA: 1.26% points, LOA (-8.42, 10.96).

{8}------------------------------------------------

EDV 2CH correlation of r=0.96, CI [0.94, 0.97] was obtained between automated and manual results, BA: -5.69 ml, LOA (-36.02, 24.44)

ESV 2CH correlation of r=0.98, CI [0.97, 0.99] was obtained between automated and manual results, BA: -3.87ml, LOA (-25.58, 17.82).

EF 2CH correlation of r=0.93, CI [0.90, 0.95] was obtained between automated and manual results, BA: -0.54% points, LOA (-12.18, 11.1)

BP EDV correlation of r=0.98, CI [0.97, 0.99] was obtained between automated and manual results, BA: -4.1ml. LOA (-29.04, 20.84).

BP ESV correlation of r=0.98, CI [0.97, 0.99] was obtained between automated and manual results, BA: -2.77ml. LOA (-19.67, 14.13), BA: 0.39% points, LOA (-8.26, 9.05).

BP EF correlation of r=0.96, Cl [0.94, 0.97] was obtained between automated and manual results, BA: 0.39% points, LOA (-8.26, 9.05)

    1. Conclusion: The intended use and the technological characteristics in the current device are the same as those in the predicate device. Likewise, the modifications do not affect the safety and effectiveness of the device. The performance tests have been completed and successfully verify the validated performance of the modified device. Therefore. DiA Imaging Analysis has concluded that the modified LVivo Software Application is substantially equivalent to its predicate device.

N/A