K Number
K243130

Validate with FDA (Live)

Date Cleared
2025-06-27

(270 days)

Product Code
Regulation Number
872.6640
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Integral Dental Unit is intended to supply power to and serve as a base for dental devices and accessories. This device includes a dental chair and is intended for use in the dental clinic environment and is designed for use by trained dental professionals, dentists and/or dental assistants.

Device Description

The Integral Dental Unit (Model: YD-A4) is a dental operative unit specially designed and provided for a qualified dentist to be used in a professional clinic or hospital facility to carry out dental procedures.

Integral Dental Unit is intended to be used in a professional environment for dental diagnosis, treatment, or operation.

It includes a treatment chair, dentist element, assistant element and a dental light as offering several additional options and electronically-controlled chair movements with software and water/air unit functions.

The dental unit consists of an electronically operated dental chair and integrated control unit control for electricity, water and air supply to handpieces or some other dental instrument. The device is equipped with an instrument tray, pipes for water supply and tube air supply, a mouth lamp, a saliva aspirator, a spittoon, a three-way syringe, a film viewer, a foot switch and chair with driving motors and armrest.

The dental chair is intended to be used with dental hand pieces, cure light, ultrasonic scaler, camera system or other doctor stool, which is not provided by the manufacturer. The user will select the dental instruments and accessories with FDA clearance by themself. So, the device in the submission does not include these parts and accessories. The connector standard type complies with ISO 9168.

Basic parameters/use conditions/power supply specifications is as follows:

◆Noise <70 dBA

◆Base box Power supply: 115/230 Vac, 50/60Hz, single-phase 3core, protective grounding.
Power input: 380 VA
Water filter hole diameter: <90µm
Air filter hole diameter: <25µm

◆Saliva ejector Weak saliva:

  • vacuum degree >15 kPa;
  • water pumping rate>80 mL/min

Strong saliva:

  • vacuum degree >25 kPa;
  • water pumping rate>1000 mL/min

◆Instrument tray Rotating angle: >270º
Up-down moving range: >440mm
Max. Load: <5 Kg

◆Film viewer Power supply: 24 V ac Inside power supply
Power input: 80 VA

◆Operating light Power supply: 12 V ac Inside power supply
Power input: 50 VA
Illuminance: 5 000~20 000 Lx
Radiant heat: <200 W/m² @ max. illuminance
Color rendering index: >85 Ra

◆Foot switch Tripping force: >10N and <30N
Degree of protection against harmful ingress of water: IPX4(foot switch)
Service life: >25 000 repeats

◆Dental chair Power supply: 24 V dc Inside power supply
Loading capacity: 1323N (about 135 Kg)
Loading capacity of headrest: 300N (about 30Kg)
Moving range of headrest: 120mm
Range of backrest when going backwards: 90º~170º
Seat cushion's maximum height away from ground: 730mm
Seat cushion's minimum height away from ground: 450mm

Attachment
----- Amalgam separation device
It has a medical device product registration card
Attachment parameters are reflected in its operating instructions

◆Work space L: ≥3 000 mm; W: ≥2 000 mm; H: ≥2 500 mm

◆Environment for operation
Temperature: +5°C to +40°C
Relative humidity: 30% - 80%
Atmospheric pressure: 86kPa ~ 106kPa

AI/ML Overview

The provided FDA 510(k) clearance letter pertains to the Integral Dental Unit, a Class I medical device. This device is an "Operative Dental Unit and Accessories" and is intended to supply power to and serve as a base for other dental devices and accessories, including a dental chair, for use by trained dental professionals in a dental clinic environment.

The document states that clinical performance testing was not performed for this device. As such, information regarding acceptance criteria derived from a clinical study, sample size for test sets (including data provenance, ground truth establishment, expert qualifications, and adjudication methods), MRMC studies, or standalone algorithm performance, and training set details are not provided in this submission.

Instead, the device's acceptance criteria are based on conformance to established international standards for electrical safety, electromagnetic compatibility, biocompatibility, software verification and validation, software life cycle processes, and general performance for stationary dental units and dental patient chairs.

Here's the information extracted from the document:

1. Table of Acceptance Criteria and Reported Device Performance

Criteria Category / Specific ParameterAcceptance Criteria (Standard / Pre-defined Value)Reported Device Performance (Reference to Standard Conformance)
Electrical SafetyIEC 60601-1:2012Complied with IEC 60601-1
Electromagnetic CompatibilityIEC 60601-1-2:2014Complied with IEC 60601-1-2
BiocompatibilityISO 10993-1:2018Complied with ISO 10993-1
Software Verification & ValidationFDA "Guidance for Premarket Submissions and for Software Contained in Medical Devices"Complied with FDA Guidance
Software Life Cycle ProcessesIEC 62304:2006+AMD1:2015Complied with IEC 62304
Performance - General Requirements (Dental Unit & Patient Chair)ISO 7494-1:2018Complied with ISO 7494-1
Performance - Air, Water, Suction & Wastewater SystemsISO 7494-2:2018Complied with ISO 7494-2
Noise< 70 dBA< 70 dBA
Base box - Power supply115/230 Vac, 50/60Hz, single-phase 3core, protective grounding115/230 Vac, 50/60Hz, single-phase 3core, protective grounding
Base box - Power inputNot explicitly stated (Predicate: AC 100-120/220-240V)380 VA
Base box - Water filter hole diameter< 90µm< 90µm
Base box - Air filter hole diameter< 25µm< 25µm
Saliva ejector - Weak saliva vacuum degree> 15 kPa> 15 kPa
Saliva ejector - Weak saliva water pumping rate> 80 mL/min> 80 mL/min
Saliva ejector - Strong saliva vacuum degree> 25 kPa> 25 kPa
Saliva ejector - Strong saliva water pumping rate> 1000 mL/min> 1000 mL/min
Instrument tray - Rotating angle> 270º> 270º
Instrument tray - Up-down moving range> 440mm> 440mm
Instrument tray - Max. Load< 5 Kg< 5 Kg
Film viewer - Power supply24 V ac Inside power supply24 V ac Inside power supply
Film viewer - Power input80 VA80 VA
Operating light - Power supply12 V ac Inside power supply12 V ac Inside power supply
Operating light - Power input50 VA50 VA
Operating light - Illuminance5 000~20 000 Lx5 000~20 000 Lx
Operating light - Radiant heat< 200 W/m² @ max. illuminance< 200 W/m² @ max. illuminance
Operating light - Color rendering index> 85 Ra> 85 Ra
Foot switch - Tripping force> 10N and < 30N> 10N and < 30N
Foot switch - Degree of protection against harmful ingress of waterIPX4IPX4
Foot switch - Service life> 25 000 repeats> 25 000 repeats
Dental chair - Power supply24 V dc Inside power supply24 V dc Inside power supply
Dental chair - Loading capacity1323N (about 135 Kg)1323N (about 135 Kg)
Dental chair - Loading capacity of headrest300N (about 30Kg)300N (about 30Kg)
Dental chair - Moving range of headrest120mm120mm
Dental chair - Range of backrest when going backwards90º~170º90º~170º
Dental chair - Seat cushion's maximum height away from ground730mm730mm (Predicate: 795±10mm)
Dental chair - Seat cushion's minimum height away from ground450mm450mm (Predicate: 365±10mm)
Work space - Length≥ 3 000 mm≥ 3 000 mm
Work space - Width≥ 2 000 mm≥ 2 000 mm
Work space - Height≥ 2 500 mm≥ 2 500 mm
Environment for operation - Temperature+5°C to +40°C+5°C to +40°C
Environment for operation - Relative humidity30% - 80%30% - 80%
Environment for operation - Atmospheric pressure86kPa ~ 106kPa86kPa ~ 106kPa
Air Pressure (Comparison to Predicate)0.6MPa~0.80 MPa0.6MPa~0.80 MPa (Predicate: 500kPa(min)/750kPa(max))
Water Pressure (Comparison to Predicate)0.20 MPa ~0.40 MPa0.20 MPa ~0.40 MPa (Predicate: 250kPa(min)/600 kPa(max))

2. Sample size used for the test set and the data provenance

Not applicable. Clinical performance testing was not performed. The testing conducted was primarily engineering and bench testing against recognized standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. Ground truth as typically understood for clinical studies (e.g., expert interpretation of medical images) was not established. Performance was assessed against engineering specifications and international standards by relevant testing laboratories.

4. Adjudication method for the test set

Not applicable. No expert review or adjudication method was described as clinical performance testing was not performed.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. No MRMC study was conducted as this device is an Integral Dental Unit, not an AI-assisted diagnostic or therapeutic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is not an algorithm or AI system.

7. The type of ground truth used

The "ground truth" for this device's performance is based on conformance to international engineering and safety standards (e.g., IEC, ISO) and the device's adherence to its own design specifications and functional parameters. There is no clinical "ground truth" derived from patient outcomes or expert consensus in this submission.

8. The sample size for the training set

Not applicable. This is not an AI/Machine Learning device that requires a training set.

9. How the ground truth for the training set was established

Not applicable. This is not an AI/Machine Learning device.

FDA 510(k) Clearance Letter - Integral Dental Unit

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

June 27, 2025

Guangdong Yadeng Medical Apparatus Co., Ltd.,
℅ Jett Lee
Official Applicant
Guangdong Jianda Medical Technology Co., Ltd.
906 Room
Longxiang Garden, Tianhe district
GuangZhou, Guangdong 510000
CHINA

Re: K243130
Trade/Device Name: Integral Dental Unit
Regulation Number: 21 CFR 872.6640
Regulation Name: Dental Operative Unit And Accessories
Regulatory Class: Class I, reserved
Product Code: EIA, KLC
Dated: May 26, 2025
Received: May 27, 2025

Dear Jett Lee:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K243130 - Jett Lee Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K243130 - Jett Lee Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

MICHAEL E. ADJODHA -S

Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K243130

Device Name
Integral Dental Unit

Indications for Use (Describe)
The Integral Dental Unit is intended to supply power to and serve as a base for dental devices and accessories. This device includes a dental chair and is intended for use in the dental clinic environment and is designed for use by trained dental professionals, dentists and/or dental assistants.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

June 23, 2025

K243130 - 510(k) Summary

1. Submitter's Information

Sponsor

  • Company Name: GUANGDONG YADENG MEDICAL APPARATUS CO., LTD.
  • Address: Factory 2, No.12, Fenggong Road, Nansha Industrial Zone, Danzao Town, Nanhai District, Foshan, Guangdong, 528216, P.R. China
  • Phone: +86- 15916558838
  • Fax: +86-0757-86603813
  • Contact Person (including title): Weng changming
  • E-mail: 541405168@qq.com

Application Correspondent

  • Company: Guangdong Jianda Medical Technology Co., Ltd
  • Address: 906 Room, Longxiang Garden, Tianhe district, Guangzhou, China
  • Contact Person: Jett Lee
  • Tel: 86-13512755282
  • Email: jianda-lee@foxmail.com

2. Subject Device Information

  • Type of 510(k) submission: Traditional
  • Common Name: Unit, Operative Dental
  • Trade Name: Integral Dental Unit
  • Model: YD-A4
  • Classification Name: Dental Operative Unit and Accessories
  • Review Panel: Dental
  • Product Code: EIA
  • Regulation Number: 872.6640
  • Regulation Class: I

3. Predicate Device Information

Predicate Device
SponsorOsstem Implant Co., Ltd.
Device NameK3
Model/
510(k) NumberK183347

Page 6

Product CodeEIA
Regulation Number872.6640
Regulation ClassI

4. Device Description

The Integral Dental Unit (Model: YD-A4) is a dental operative unit specially designed and provided for a qualified dentist to be used in a professional clinic or hospital facility to carry out dental procedures.

Integral Dental Unit is intended to be used in a professional environment for dental diagnosis, treatment, or operation.

It includes a treatment chair, dentist element, assistant element and a dental light as offering several additional options and electronically-controlled chair movements with software and water/air unit functions.

The dental unit consists of an electronically operated dental chair and integrated control unit control for electricity, water and air supply to handpieces or some other dental instrument. The device is equipped with an instrument tray, pipes for water supply and tube air supply, a mouth lamp, a saliva aspirator, a spittoon, a three-way syringe, a film viewer, a foot switch and chair with driving motors and armrest.

The dental chair is intended to be used with dental hand pieces, cure light, ultrasonic scaler, camera system or other doctor stool, which is not provided by the manufacturer. The user will select the dental instruments and accessories with FDA clearance by themself. So, the device in the submission does not include these parts and accessories. The connector standard type complies with ISO 9168.

Basic parameters/use conditions/power supply specifications is as follows:

◆Noise <70 dBA

◆Base box Power supply: 115/230 Vac, 50/60Hz, single-phase 3core, protective grounding.
Power input: 380 VA
Water filter hole diameter: <90µm
Air filter hole diameter: <25µm

◆Saliva ejector Weak saliva:

  • vacuum degree >15 kPa;
  • water pumping rate>80 mL/min

Strong saliva:

  • vacuum degree >25 kPa;
  • water pumping rate>1000 mL/min

◆Instrument tray Rotating angle: >270º
Up-down moving range: >440mm
Max. Load: <5 Kg

◆Film viewer Power supply: 24 V ac Inside power supply
Power input: 80 VA

Page 7

◆Operating light Power supply: 12 V ac Inside power supply
Power input: 50 VA
Illuminance: 5 000~20 000 Lx
Radiant heat: <200 W/m² @ max. illuminance
Color rendering index: >85 Ra

◆Foot switch Tripping force: >10N and <30N
Degree of protection against harmful ingress of water: IPX4(foot switch)
Service life: >25 000 repeats

◆Dental chair Power supply: 24 V dc Inside power supply
Loading capacity: 1323N (about 135 Kg)
Loading capacity of headrest: 300N (about 30Kg)
Moving range of headrest: 120mm
Range of backrest when going backwards: 90º~170º
Seat cushion's maximum height away from ground: 730mm
Seat cushion's minimum height away from ground: 450mm

Attachment
----- Amalgam separation device
It has a medical device product registration card
Attachment parameters are reflected in its operating instructions

◆Work space L: ≥3 000 mm; W: ≥2 000 mm; H: ≥2 500 mm

◆Environment for operation
Temperature: +5°C to +40°C
Relative humidity: 30% - 80%
Atmospheric pressure: 86kPa ~ 106kPa

5. Intended Use / Indications for Use

The Integral Dental Unit is intended to supply power to and serve as a base for dental devices and accessories. This device includes a dental chair and is intended for use in the dental clinic environment and is designed for use by trained dental professionals, dentists and/or dental assistants.

6. Test Summary

The Integral Dental Unit was evaluated for conformance to recognized international standards. The following is a list of these evaluations and tests that were found to be in conformance:

Page 8

  • Electrical safety test
    IEC 60601-1:2012, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance
    IEC 80601-2-60:2019, Medical electrical equipment Part 2-60: Particular requirements for the basic safety and essential performance of dental equipment

  • Electromagnetic compatibility test
    IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility

  • Biocompatibility test
    ISO 10993-1:2018, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process

  • Software verification and validation test
    FDA "Guidance for Premarket Submissions and for Software Contained in Medical Devices"

  • Software Life Cycle Processes
    IEC 62304:2006+AMD1:2015 Medical Device Software

  • Performance Testing
    ISO 7494-1:2018, Dentistry – Stationary dental units and dental patient chairs, Part 1: General requirements
    ISO 7494-2:2018, Dentistry – Stationary dental units and dental patient chairs, Part 2: Air, water, suction and wastewater systems

7. Comparison to predicate device and conclusion

The subject device Integral Dental Unit is substantially equivalent to the predicated device based on intended use, design, specifications and performance. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.

Information for predicate device was obtained from publicly available sources. A technical comparison to the predicate is provided below.

Elements of ComparisonSubject DevicePredicate Device IVerdict
510(k) NumberK243130K183347--
Device NameIntegral Dental Unit (Model: YD-A4)K3--
Product CodeEIAEIASame
Regulation Number21 CFR 872.664021 CFR 872.6640Same
Regulation ClassIISame
PrescriptionPrescription usePrescription useSame

Page 9

Intended UseThe Integral Dental Unit is intended to supply power to and serve as a base for dental devices and accessories. This device includes a dental chair and is intended for use in the dental clinic environment and is designed for use by trained dental professionals, dentists and/or dental assistants.K3 is intended to supply power to and serve as a base for dental devices and accessories. This device includes a dental chair and is intended for use in the dental clinic environment and is designed for use by trained dental professionals, dentists and/or dental assistants.Same
Power & Utility SupplyAC 230V, 50/60Hz, compressed air and waterAC 100-120/220-240V, 50/60Hz, compressed air and waterSimilar Note 1
Main ComponentsGround box, foot controller, patient chair, weak suction system, strong suction system, Triple function handpiece(hot), cuspidor: water flues& supply, cabinet box, operation lamp, film viewer, instrument control system, instrument tray, Triple function handpiece(cold), cup collector, glass plateChair, Unit, Table, Seat, Stool, Monitor Arm*, Hanaro Console* (Note: K3 Cart* model applied ONLY)Similar Note 1
Syringe3-way syringe3-way syringeSame
Control of water and airUses pneumatically controlled vales to water control the flow of air and water. On/off and intensity controlled by foot pedal.Uses pneumatically controlled vales to water control the flow of air and water. On/off and intensity controlled by foot pedal.Same
Air Pressure0.6MPa~0.80 MPa500kPa(min)/750kPa(max)Similar Note 1
Water Pressure0.20 MPa ~0.40 MPa250kPa(min)/600 kPa(max)Similar Note 1
Water SystemCity water supplyCity water supplySame
Patient LoadMax. 135kgMax. 135kgSame
Chair HeightMax. 730±10mm, Min. 450±10mmMax. 795±10mm, Min. 365±10mmSimilar Note 1
Lift MotorHydraulic electromotorHydraulic electromotorSame
Electrical SafetyComplied with IEC 60601-1Complied with IEC 60601-1Same
Electromagnetic compatibilityComplied with IEC 60601-1-2Complied with IEC 60601-1-2Same

Summary for clinical test

Clinical performance testing was not performed for this device.

Conclusion

Based on the above performance as documented in this application, Integral Dental Unit is substantially equivalent to the predicate device, K3 (K183347). Thus, the subject device is substantially equivalent to the predicate devices.

§ 872.6640 Dental operative unit and accessories.

(a)
Identification. A dental operative unit and accessories is an AC-powered device that is intended to supply power to and serve as a base for other dental devices, such as a dental handpiece, a dental operating light, an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to a dental chair.(b)
Classification. Class I (general controls). Except for dental operative unit, accessories are exempt from premarket notification procedures in subpart E of part 807 of this chapter subject to § 872.9.