K Number
K243103

Validate with FDA (Live)

Device Name
Denture Base
Date Cleared
2024-12-23

(84 days)

Product Code
Regulation Number
872.3760
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Denture Base resin is a light-curable polymerizable resin intended to be used for the fabrication and repair of full and partial removable dentures and baseplates. The material is an alternative to traditional denture base material.

Device Description

Fabrication of dental prosthetics with Denture Base resin requires computer-aided design and manufacturing system that includes the following components not part of the device: oral casting impression, digital denture base file created in an optical impression system, 3D printer, and curing light equipment.

The Denture Base consists of a curable dental acrylate resin that is designed to be used in conjunction with a scanned 3D image of a patient's teeth, and 3D printer assembly, to locally manufacture out a dental appliance in dental offices based on the clinician's judgment of patient need.

Denture Base resin is intended exclusively for professional dental work.

Denture Base is designed to meet appropriate ISO standards for flexibility, sorption, and solubility to withstand prolonged use in the oral cavity. It is delivered non-sterile, and instructions are provided on cleaning the material prior to providing it to a patient. Curing is performed with a UV lamp. The appliance is then cleaned, trimmed, and verified to fit in the dental office before the patient leaves.

AI/ML Overview

I am sorry, but the provided text does not contain the information needed to answer your request. The document is an FDA 510(k) clearance letter and summary for a denture base resin, focusing on the substantial equivalence to a predicate device through biocompatibility and bench testing.

Specifically, the document does not describe acceptance criteria for a study, nor does it detail a study that proves a device meets such criteria in the way you've outlined for AI/software-based medical devices.

The document describes material testing for a physical product (denture resin), not the performance of an algorithm or AI model. Therefore, information such as sample sizes for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, type of ground truth, training set size, and how training ground truth was established, are not present.

The "performance data" mentioned in the document refers to:

  • Biocompatibility Testing: According to ISO 10993-1, covering tests like Cytotoxicity, Genotoxicity, Acute Systemic Toxicity, Subchronic toxicity, Irritation, and Sensitization.
  • Bench Testing: According to ISO 20795-1, covering tests like Ultimate Flexural Strength and Modulus, Water Sorption and Solubility, Residual Methyl Methacrylate Monomers, Homogeneity, Surface Characteristics, Shape Capability, Translucency, Polishability, Colour, Freedom from Porosity, Color Stability, Maximum stress intensity factor, Total fracture work, and Bonding to synthetic polymer teeth.

These are tests of the physical properties and biological safety of the material, not algorithmic performance.

{0}------------------------------------------------

December 23, 2024

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a seal on the left and the FDA name on the right. The seal includes the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The FDA name is written in blue and includes the words "U.S. FOOD & DRUG ADMINISTRATION".

Riton 3D Technology Co., Ltd Reanny Wang General Manager Room C101, Building 2, No. 21, Hejing South Road Liwan District Guangzhou, Guangdong 51038 China

Re: K243103

Trade/Device Name: Denture Base Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, Or Rebasing Resin Regulatory Class: Class II Product Code: EBI Dated: September 25, 2024 Received: September 30, 2024

Dear Reanny Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bobak Shirmohammadi -S

For Michael E. Adjodha, M.ChE., RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K243103

Device Name

Denture Base

Indications for Use (Describe)

Denture Base resin is a light-curable polymerizable resin intended to be used for the fabrication and repair of full and partial removable dentures and baseplates. The material is an alternative to traditional denture base material.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) SUMMARY

Denture Base (K243103)

Submitter:Riton 3D Technology Co., LtdRoom C101, Building 2, No. 21, Hejing South RoadLiwan District Guangzhou, Guangdong CN 51038
Phone:+86-20-81509265
Contact Person:YongMing Sun
Date Prepared:Sep. 25, 2024
Name of Device:Denture Base
Common or Usual Name:Denture Base, Prescription
Classification Name:Resin, Denture, Relining, Repairing, Rebasing
Regulation Description:Denture relining, repairing, or rebasing resin
Device Class:Class 2
Regulation Number:21 CFR 872.3760
Product Code:EBI
Primary Predicate Device:SprintRay High Impact Denture Base(K221678)
Device Description:

Fabrication of dental prosthetics with Denture Base resin requires computer-aided design and manufacturing system that includes the following components not part of the device: oral casting impression, digital denture base file created in an optical impression system, 3D printer, and curing light equipment.

The Denture Base consists of a curable dental acrylate resin that is designed to be used in conjunction with a scanned 3D image of a patient's teeth, and 3D printer assembly, to locally manufacture out a dental appliance in dental offices based on the clinician's judgment of patient need.

Denture Base resin is intended exclusively for professional dental work.

Denture Base is designed to meet appropriate ISO standards for flexibility, sorption, and solubility to withstand prolonged use in the oral cavity. It is delivered non-sterile, and instructions are provided on cleaning the material prior to providing it to a patient. Curing is performed with a UV lamp. The appliance is then cleaned, trimmed, and verified to fit in the dental office before the patient leaves.

Intended Use / Indications for Use

Denture Base resin is a light-curable polymerizable resin intended to be used for the fabrication and repair of full and partial removable dentures and baseplates. The material is an alternative to traditional denture base material.

Summary of Technological Characteristics

{5}------------------------------------------------

Light curing of a 3D printed acrylate resin is the technological principle for both the subject and predicate devices. The Denture Base is poured into a 3D printer, which relies on scanned images of the patient's oral cavity to produce a dental appliance.

The following technological differences exist between the subject and predicate devices:

  • . Differences in acrylate resin material
    The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility Testing

The biocompatibility evaluation for Denture Base was conducted in accordance with the International Standard ISO 10993-1, as recognized by FDA. The battery of testing included the following tests:

  • Cytotoxicity ●
  • Genotoxicity ●
  • Acute Systematic Toxicity ●
  • Subchronic toxicity ●
  • Irritation ●
  • Sensitization ●

Denture Base is categorized as Surface Devices in contact with mucosal membrane with a contact duration of long time>30 days

Bench Testing

Denture Base was tested for conformity with the industry consensus standard ISO 20795-1. The battery of testing included the following tests:

  • Ultimate Flexural Strength and Modulus ●
  • Water Sorption and Solubility ●
  • Residual Methyl Methacrylate Monomers ●
  • Homogeneity ●
  • Surface Characteristics ●
  • Shape Capability
  • Translucency, and Polishability
  • Colour ●
  • Freedom from Porosity ●
  • Color Stability ●
  • Maximum stress intensity factor
  • Total fracture work ●
  • Bonding to synthetic polymer teeth ●

In all instances, Denture Base functioned as intended and the outcomes were as expected.

{6}------------------------------------------------

FeatureSubject Device (K243103)Primary Predicate Device (K221678)Comparison
Product NameDenture BaseSprintRay High Impact Denture BaseN/A
Product CodeEBIEBISame
Regulation21 CFR 872.376021 CFR 872.3760Same
Indications for Use/Intended UseDenture Base resin is a light-curable polymerizable resin intended to be used for the fabrication and repair of full and partial removable dentures and baseplates. The material is an alternative to traditional denture base material.The SprintRayHigh Impact Denture Base resin is a light-curable polymerizable resin intended to be used for the fabrication and repair, of full and partial removable dentures and baseplates. The material is an alternative to traditional denture base material.Same
User PopulationClinicians in dental officesClinicians in dental officesSame
Chemical DescriptionMethacrylate-based resinMethacrylate-based resinSimilar
Material TypeLight-curable ResinLight-curable ResinSame
Curing MethodUV LightUV LightSame
Product StateLiquidLiquidSame
Manufacturing Technology TypeAdditiveAdditiveSame
Volume provided1kg/bottle1kg bottleSame
Shelf life2 years1.5 yearsSimilar
StandardsISO 20795-01ISO 20795-01Same
Physical and Mechanical PropertiesUltimate Flexural Strength and ModulusFlexural Strength and ModulusSimilar
Water Sorption and SolubilityWater Sorption and Solubility
Residual Methyl Methacrylate MonomersResidual Methyl Methacrylate Monomers
HomogeneityHomogeneity
Surface CharacteristicsSurface Characteristics
Shape CapabilityTranslucency, and PolishabilityColourFreedom from PorosityShape Capability,Translucency, and PolishabilityFreedom from PorosityColor Stability
Color Stability
Maximum stress intensity factor
Total fracture work
Bonding to synthetic polymer teeth
Tested to ISO-10993-1Tested to ISO7405, ISO-10993-1
BiocompatibilityGenotoxicity (Part 3)Genotoxicity (Part 3)
Cytotoxicity(Part 5)Cytotoxicity (Part 5)
Sensitization (Part 10)Acute Systematic Toxicity(Part 11)Similar
Acute Systematic Toxicity(Part 11)Sensitization (Part 10)
Subchronic toxicity(Part 11)Irritation (Part 10)
Irritation (Part 23)
AdditiveManufacturingTesting, according to FDA'sguidance Technical Considerationsfor Additive Manufactured MedicalDevices, was performed and resultswere provided in the 510(k). Thesetests included evaluation of allrelevant properties of the printedTesting, according to FDA's guidanceTechnical Considerations for AdditiveManufactured Medical Devices, wasperformed and results were provided inthe 510(k). These tests includedevaluation of all relevant properties of theprinted resin using the permittedSame
resin using the permitted machines.Further, tests based on considerationsof the orientation duringmanufacturing were performed.machines. Further, tests based onconsiderations of the orientation duringmanufacturing were performed.
PrinterDeviceRXDent-L230 3D printerSprintRay Pro 95Similar
Post-CureDeviceRXDent-W90(s)SprintRay ProCure 2Similar
SterilityNon-sterileNon-sterileSame

{7}------------------------------------------------

Conclusions

The Denture Base resin is as safe and effective as its predicate device. The Denture Base has the same intended use and indication, and similar technological characteristics, and principles of operation as its predicate device. The minor technological differences between the Denture Base and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the Denture Base is as safe and effective as the predicate device. Thus, the Denture Base is substantially equivalent.

§ 872.3760 Denture relining, repairing, or rebasing resin.

(a)
Identification. A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.(b)
Classification. Class II.