K Number
K243081

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-02-14

(137 days)

Product Code
Regulation Number
892.1750
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Green X 21 (Model: PHT-90CHO) is intended to produce panoramic, cephalometric, or 3D digital Xray images. It provides diagnostic details of the dento-maxillofacial, sinus, TMJ, and ENT for adult and pediatric patients. The system also utilizes carpal images for orthodontic treatment. The device is to be operated by healthcare professionals.

Device Description

Green X 21 (Model : PHT-90CHO) is an advanced 6-in-1 digital X-ray system specifically designed for 2D and 3D dental radiography. This system features six imaging modalities: PANO, CEPH (optional), DENTAL CT, ENT CT, MODEL, and FACE SCAN, all integrated into a single unit.

Green X 21's imaging system is based on digital TFT detectors, accompanied by imaging viewers and an X-ray generator.

The materials, safety characteristics, X-ray source, indications for use and image reconstruction/MAR(Metal Artifact Reduction) algorithm of the subject device are same as the predicate device (PHT-75CHS (K210329)). The subject device differs from the predicate device in the following ways: It is equipped with new X-ray detectors for CT/PANO and CEPH modification results in a different maximum FOV provided in a single scan for CT mode compared to the predicate device. For CEPH modality, the subject device utilizes a one- shot imaging capture method.

Additionally, the subject device includes new modalities such as ENT CT and FACE SCAN with a face scanner, along with new software functions, including Auto Pano, Smart Focus, and Scout.

AI/ML Overview

The provided document describes a 510(k) premarket notification for the "Green X 21 (PHT-90CHO)" digital X-ray system. The core of the submission is to demonstrate substantial equivalence to a predicate device, the "Green X 18 (PHT-75CHS)". The document focuses heavily on comparing technical specifications and performance data to support this claim, particularly for the new components and features of the Green X 21.

However, it's crucial to note that this document does not describe a study that proves the device meets specific acceptance criteria in the context of an AI/human-in-the-loop performance study. Instead, it focuses on engineering performance criteria related to image quality and safety for an X-ray imaging device, and comparing these to a predicate device. The "acceptance criteria" discussed are primarily technical specifications and performance benchmarks for the X-ray system components (detectors, imaging quality metrics), rather than clinical performance metrics in disease detection with AI assistance.

Therefore, many of the requested items (e.g., sample size for test set, number of experts for ground truth, adjudication method, MRMC study, standalone performance) are not applicable or not detailed in this submission because the device in question is an X-ray imaging system, not an AI-based diagnostic tool. The "new software functions" mentioned (Auto Pano, Smart Focus, Scout) are described as image reconstruction or manipulation tools, not AI algorithms for clinical diagnosis.

Here's a breakdown based on the information available, addressing the points where possible and noting when information is absent or not relevant to the provided text:

Acceptance Criteria and Reported Device Performance

Given that this is an X-ray imaging device and not an AI diagnostic tool, the acceptance criteria are generally related to image quality, safety, and functional performance, benchmarked against standards and the predicate device.

Table of Acceptance Criteria and Reported Device Performance (as inferred from the document):

Acceptance Criteria CategorySpecific Criteria (Inferred)Reported Device Performance (Green X 21)
I. Imaging Performance (New X-ray Detectors)
1. CT/PANO Detector (Jupi1012X)- Modulation Transfer Function (MTF) & Detective Quantum Efficiency (DQE) & Noise to Power Spectrum (NPS): Performance comparable or superior to predicate (Xmaru1524CF Master Plus OP).- MTF (CT/PANO): Jupi1012X showed more stable or superior performance for DQE, MTF, and NPS, particularly better stability in high-frequency regions. Jupi1012X could distinguish up to 3.5 line pairs (MTF 10% criterion), compared to 2.5 line pairs for predicate.
- Pixel Size: Similar to predicate device.- Pixel Size (CT/PANO): "Very similar" to predicate. Image test patterns demonstrated test objects across the same spatial frequency range without aliasing.
2. CEPH Detector (Venu1012VD)- MTF, DQE, NPS: Performance comparable or superior to predicate (Xmaru2602CF), despite predicate having lower NPS (noise).- MTF (CEPH): Venu1012VD exhibited superior performance in DQE, MTF, and NPS (except predicate's better NPS due to lower noise). Higher MTF values indicate sharper images.
- Pixel Size: Similar to predicate device.- Pixel Size (CEPH): Similar to predicate's 100 µm (non-binning). Image test patterns demonstrated test objects across the same spatial frequency range without aliasing.
3. Overall Diagnostic Image Quality- Equivalent to or better than predicate device.- "Equivalently or better than the predicate device in overall image quality."
II. Compliance with Standards and Guidelines
1. IEC 61223-3-5 (CT)- Quantitative testing for noise, contrast, CNR, MTF 10%.- All parameters met the standards specified.
2. IEC 61223-3-4 (Dental X-ray)- Quantitative assessment for line pair resolution and low contrast performance (PANO/CEPH).- All parameters met the standards specified.
3. Software/Firmware (Basic Documentation Level)- Adherence to FDA Guidance "Content of Premarket Submissions for Device Software Functions."- Software verification and validation conducted and documented.
4. Safety & EMC- Compliance with IEC 60601-1, IEC 60601-1-3, IEC 60601-2-63 standards for electrical, mechanical, environmental safety and performance; IEC 60601-1-2 for EMC.- Testing performed and results were satisfactory.
III. Functional Equivalence / Performance of New Modalities/Functions
1. ENT CT Modality- Meet image quality standards of Dental CT; limit radiation exposure to ENT region.- "Adheres to the same image quality standards as the Dental CT modality." Specifically designed to limit radiation exposure to ENT region.
2. FACE SCAN Modality- Intended for aesthetic consultations, not clinical diagnostic; meets internal criteria.- "Not designed for clinical diagnostic purposes." "Meets the internally established criteria and has been designed to perform as intended."
3. Auto Pano, Smart Focus, Scout Modes- Should function as intended, similar to previously cleared devices (Green X 12, K231796).- Evaluated according to IEC 61223-3-4 and IEC 61223-3-5; "both standard requirements were met."
IV. Dosimetric Performance (DAP)
1. PANO Modality DAP- Similar DAP to predicate device under same exposure conditions.- "DAP measurement results are similar" when tested under same exposure conditions (High Resolution Mode).
2. CEPH Modality DAP- Performance balance between increased DAP (due to one-shot type) and reduced exposure time/motion artifacts.- DAP "more than twice that of the predicate device" (due to one-shot vs. scan-type), but "utilizes a one-shot type, operating with approximately one-fourth the exposure time... This reduces motion artifacts."
3. CT Modality DAP- Overall DAP performance balanced with FOV and image quality.- "Slight increase in DAP for the subject device" for most comparable/equivalent FOVs. However, "maximum FOV provided by the subject device demonstrated a reduced radiation dose compared to the predicate device."

Study Details (as far as applicable and available)

  1. Sample size used for the test set and the data provenance:

    • Test Set Description: The "test set" in this context is referring to data collected through bench testing using phantoms for image quality assessment, and potentially clinical images for subjective comparison, rather than a clinical trial cohort.
    • Sample Size: Not specified in terms of number of patient images. The testing was conducted on the device itself using phantom studies (e.g., test patterns for MTF, line pairs for resolution, phantoms for low contrast).
    • Data Provenance: The bench tests were conducted in a laboratory ("in a laboratory using the same test protocol as the predicate device"). The "Clinical consideration" section for image quality evaluation implies some clinical image review, but the origin (e.g., country, retrospective/prospective) of these potential clinical images is not detailed.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable in the traditional sense of an AI study. Ground truth for an X-ray imaging device's performance is typically established through quantitative measurements using phantoms and technical specifications, not expert consensus on medical findings.
    • The document mentions "Image Quality Evaluation Report and Clinical consideration" and concludes "the subject device performed equivalently or better than the predicate device in overall image quality." This implies some form of expert review for subjective image quality, but the number or qualifications of these "experts" are not detailed.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable. As noted above, this is not an AI diagnostic study with a human-in-the-loop component requiring adjudication of disease findings. The evaluations are primarily technical and quantitative measurements comparing physical properties and image output.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. An MRMC study was not described. This is an X-ray imaging system, not an AI diagnostic software that assists human readers. The new software functions (Auto Pano, Smart Focus, Scout) described are image manipulation/reconstruction features, not AI for diagnostic assistance.
  5. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done:

    • Not applicable in the context of an AI algorithm. The device is an X-ray system; its "standalone" performance refers to its ability to produce images according to technical specifications, which was assessed through bench testing (MTF, DQE, NPS, etc.). There is no mention of a diagnostic AI algorithm that operates standalone.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Quantitative Phantom Measurements and Technical Specifications: For image quality (MTF, DQE, NPS, line pair resolution, contrast, noise, CNR), the ground truth is established by physical measurements using standardized phantoms and reference values as per IEC standards.
    • Predicate Device Performance: A key "ground truth" for substantial equivalence is the established performance of the predicate device. The new device's performance is compared against this benchmark.
    • Internal Criteria: For functionalities like FACE SCAN, "internally established criteria" were used as a benchmark.
  7. The sample size for the training set:

    • Not Applicable. This document describes an X-ray imaging device, not an AI model that requires a training set. The software functions mentioned (Auto Pano, Smart Focus, Scout) are described as computational algorithms for image reconstruction or enhancement, not machine learning models that learn from a training dataset.
  8. How the ground truth for the training set was established:

    • Not Applicable. No training set for an AI model is described.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 14, 2025

VATECH Co., Ltd % Dave Kim RA Consultant Mtech Group LLC 7505 Fannin St. Suite 610 HOUSTON, TX 77054

Re: K243081

Trade/Device Name: Green X 21 (PHT-90CHO) Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: OAS Dated: January 14, 2025 Received: January 15, 2025

Dear Dave Kim:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang

Lu Jiang, Ph.D. Assistant Director Diagnostic X-ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K243081

Device Name

Green X 21 (PHT-90CHO)

Indications for Use (Describe)

Green X 21 (Model: PHT-90CHO) is intended to produce panoramic, cephalometric, or 3D digital Xray images. It provides diagnostic details of the dento-maxillofacial, sinus, TMJ, and ENT for adult and pediatric patients. The system also utilizes carpal images for orthodontic treatment. The device is to be operated by healthcare professionals.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary K243081

1. 510(k) Summary

This summary of 510(k) information is being submitted in accordance with requirements of 21 CFR Part 807.92.

2. Date 510K Summary prepared: February 7, 2025

3. Administrative Information

Official Correspondent:Dave Kim / Mtech Group
Address: 7505 Fannin Street, Suite 610, Houston, TX 77054
Tel: +713-467-2607
Contact person: Mr. Dave Kim (davekim@mtechgroupllc.com)
510(k) Submitter:VATECH Co., Ltd.
Address: 13, Samsung 1-ro 2-gil, Hwaseong-si, Gyeonggi-do, 18449, Korea
Tel: +82-31-379-9492
Fax: +82-31-379-9400
Contact person: Daniel Kim / Manager (daniel.kim@vatech.co.kr)

4. Device Information

Type of 510(k) Submission:Traditional
Trade or Proprietary Name:Green X 21 (PHT-90CHO)
Common or Usual Name:X-Ray, Tomography, Computed, Dental
Regulation Classification:Computed tomography x-ray system(21 CFR 892.1750)
Product Code:OAS
Class of Device:Class II
Panel:Radiology

5. Predicate Device Information

Manufacturer:VATECH Co., Ltd.
Trade or Proprietary Name:Green X 18 (Model : PHT-75CHS)
Common or Usual Name:X-Ray, Tomography, Computed, Dental
Regulation Classification:Computed tomography x-ray system(21 CFR 892.1750)
Product Code:OAS
Class of Device:Class II
Panel:Radiology
510(k) Number:K210329

{5}------------------------------------------------

6. Reference Device Information

Manufacturer:VATECH Co., Ltd.
Trade or Proprietary Name:Green X 12 (Model : PHT-75CHS)
Common or Usual Name:X-Ray, Tomography, Computed, Dental
Regulation Classification:Computed tomography x-ray system(21 CFR 892.1750)
Product Code:OAS
Class of Device:Class II
Panel:Radiology
510(k) Number:K231796

7. Device Description

Green X 21 (Model : PHT-90CHO) is an advanced 6-in-1 digital X-ray system specifically designed for 2D and 3D dental radiography. This system features six imaging modalities: PANO, CEPH (optional), DENTAL CT, ENT CT, MODEL, and FACE SCAN, all integrated into a single unit.

Green X 21's imaging system is based on digital TFT detectors, accompanied by imaging viewers and an X-ray generator.

The materials, safety characteristics, X-ray source, indications for use and image reconstruction/MAR(Metal Artifact Reduction) algorithm of the subject device are same as the predicate device (PHT-75CHS (K210329)). The subject device differs from the predicate device in the following ways: It is equipped with new X-ray detectors for CT/PANO and CEPH modification results in a different maximum FOV provided in a single scan for CT mode compared to the predicate device. For CEPH modality, the subject device utilizes a one- shot imaging capture method.

Additionally, the subject device includes new modalities such as ENT CT and FACE SCAN with a face scanner, along with new software functions, including Auto Pano, Smart Focus, and Scout.

8. Indications for use

Green X 21 (Model: PHT-90CHO) is intended to produce panoramic, or 3D digital X-ray images. It provides diagnostic details of the dento-maxillofacial, sinus, TMJ, and ENT for adult and pediatric patients. The system also utilizes carpal images for orthodontic treatment. The device is to be operated by healthcare professionals.

{6}------------------------------------------------

9. Comparison of Technological characteristics with the predicate device

Subject DevicePredicate Device
Device NameGreen X 21 (Model : PHT-90CHO)Green X 18 (Model : PHT-75CHS)
Applicant NameVATECH Co., Ltd.VATECH Co., Ltd.
510(k) NumberK243081K210329
Device ClassificationNameX-Ray, Tomography, Computed, DentalX-Ray, Tomography, Computed, Dental
Classification ProductCodeOASOAS
Regulation Number21 CFR 892.175021 CFR 892.1750
Indications for UseGreen X 21 (Model: PHT-90CHO) isintended to produce panoramic,cephalometric, or 3D digital X-rayimages.It provides diagnostic details of thedento- maxillofacial, sinus, TMJ, andENT for adult and pediatric patients.The system also utilizes carpalimages for orthodontic treatment.The device is to be operated byhealthcare professionals.Green X 18 (Model: PHT-75CHS) isintended to produce panoramic,cephalometric, or 3D digital x-rayimages.It provides diagnostic details of thedento- maxillofacial, sinus, TMJ, andENT for adult and pediatric patients.The system also utilizes carpalimages for orthodontic treatment.The device is to be operated byhealthcare professionals.
PerformanceSpecificationPanoramic, Cephalometric and computedtomographyPanoramic, Cephalometric andcomputed tomography
Input VoltageAC 100 - 240 VAC 100 - 240 V
X-ray sourceD-052SBD-052SB
Tube Voltage60 - 99 kV60 - 99 kV
Tube Current4 - 16 mA4 - 16 mA
Focal Spot Size0.5 x 0.5 mm0.5 x 0.5 mm
Exposure TimeMax. 18.0 sMax. 18.0 s
Slice WidthMin. 0.1 mmMin. 0.1 mm
Total FiltrationMin. 2.5 mm AlMin. 2.5 mm Al
MechanicalCompact designCompact design
ElectricalLDCP logic circuitLDCP logic circuit
SoftwareDICOM 3.0 Format compatibleDICOM 3.0 Format compatible
2D Image ViewingProgramEzDent-i (K241114)EzDent-i (K241114)
3D Image ViewingProgramEz3D-i (K231757)Ez3D-i (K231757)
Anatomical SitesMaxillofacialMaxillofacial
X-ray CT&PANOdetector CEPHJupi1012XVenu1012VDXmaru1524CF Master Plus OPXmaru2602CF
Size of Imaging VolumeJupi1012XMax. 210 x 190 mmXmaru1524CFMaster Plus OPMax. 180 x 150 mm
X-raydetectorActiveJupi1012X249.6 x 300.8 mmVenu1012VD250 x 300 mmXmaru1524CFMaster Plus OP233.04 x 145.73 mmXmaru2602CF259 x 15.60 mm
area
Pixel SizeCT&PANO100 µm (1x1 binning)200 µm (2x2 binning)99 μm -2X2binning 198 μm -4X4 binning
CEPH125 µm100 μm -non binning200 µm -2X2 binning

{7}------------------------------------------------

10. Performance Data

Summary of Performance Testing -

The Green X 21 (Model : PHT-90CHO) digital X-ray system described in this 510(k) is identical to the predicate device in terms of indications for use, materials, safety characteristics, X-ray source and image reconstruction /MAR(Metal Artifact Reduction) process algorithm.

The following information further substantial equivalence between the subject device and the predicate device : The fundamental technological characteristics of the subject and predicate device are similar. The imaging modalities are similar; PANO, CEPH (Optional), CBCT (Dental/ENT CT) and MODEL. All viewing software programs have been cleared with previous 510k submissions; EzDent-i (K241114) and Ez3D-i (K231757).

The differences between the subject device and predicate device:

    1. New X-ray detectors: CT/PANO (Jupi1012X) and CEPH(Venu1012VD)
    1. New modalities: ENT CT and FACE SCAN
    1. New Software Functions: Auto Pano, Smart Focus, and Scout mode

First difference, New X-ray detectors

The subject device is equipped with New X-ray detectors(CT/PANO modality; Jupi1012X and CEPH modality; Venu1012VD).

The subject device was tested and evaluated in a laboratory using the same test protocol as the predicate device to compare non-clinical and clinical performance. Non-clinical tests were conducted for the subject device's new X-ray detectors, comparing their performance with the predicate device in terms of Modulation Transfer Function (MTF), Detective Quantum Efficiency (DQE), and Noise to Power Spectrum (NPS).

According to the results of non-clinical tests, the pixel size of the new CT/PANO X-ray detector(Jupi1012X) was found to be very similar to that of the predicate device(Xmaru1524CF Master Plus OP). Consequently, the image test patterns of the Jupi1012X clearly demonstrate test objects across the same spatial frequency range as the Xmaru1524CF Master Plus OP without any aliasing artifacts. The Jupi1012X showed more stable or superior performance compared to the Xmaru1524CF Master Plus OP in terms of DQE, MTF, and NPS, with particularly better stability in high-frequency regions than in low-frequency regions. Additionally, based on the MTF 10% criterion, the Jupi1012X could distinguish up to 3.5 line pairs, while the Xmaru1524CF Master Plus OP could only distinguish up to 2.5 line pairs.

{8}------------------------------------------------

In CEPH modality, the pixel size of the new X-ray detector(Venu1012VD) was also similar to the 100 µm (non binning) pixel size supported by the predicate device's CEPH x-ray detector(Xmaru2602CF). The image test patterns obtained with the Venu1012VD also demonstrate test objects across the same spatial frequency range as the Xmaru2602CF without aliasing artifacts.

While the Venu1012VD exhibited superior performance in DQE, MTF, and NPS, the Xmaru2602CF demonstrated better performance in NPS due to its lower noise level. However, Venu1012VD's higher MTF values indicate its capability to deliver sharper images.

In conclusion, the diagnostic image quality of the new X-ray detectors was equivalent to or better than that of the predicate device, with no significant differences in efficiency and safety.

Second difference, New modalities

The subject device includes additional modalities: ENT CT and FACE SCAN.

The CT mode of the subject device is subdivided into DENTAL CT and ENT CT. The newly added ENT CT modality adheres to the same image quality standards as the Dental CT modality. It is specifically designed to limit radiation exposure to the ENT region (Facial Bone, Petrous Bone, Airway, Cochlear, Single Ear) for diagnostic purposes.

The FACE SCAN modality is an optional feature that overlays 3D photographs onto the patient's Dental CT images. This modality does not require X-ray exposure for its operation. Additionally, FACE SCAN is intended for aesthetic consultations, such as discussing facial appearance before and after procedures or orthodontic treatments, and is not designed for clinical diagnostic purposes.

The FACE SCAN feature is only available when the FOV is set to 210x150, or 120x90 (mm). The new functionality of the subject device meets the internally established criteria and has been designed to perform as intended.

Third difference, New Software functions

The new software functions (Auto Pano, Smart Focus, and Scout) are the same as those previously cleared in the reference device, Green X 12 (PHT-75CHS, K231796).

The Auto Pano function reconstructs 3D CBCT data to create 2D panoramic images without separate X-ray scans.

Smart focus function provides a high-resolution CBCT image in the FOV 40x40 mm that the user can designate from 1 to 3 images in the console SW after acquiring a projection of the entire tooth. Full arch CT (FOV 120x90 mm) and Auto Pano images are additionally provided as user options.

Scout is a function provided in FOV120x50, 80x50, 50x50 mm and Endo(40x40 mm) mode, which allows the user to preview and adjust the position of teeth before actual image acquisition.

The acceptance test was conducted in accordance with the requirements of 21 CFR Part 1020.30, 1020.31, 1020.33, IEC 61223-3-5 (Evaluation and routine testing in medical imaging departments -Part 3- 5: Acceptance and constancy tests - Imaging performance of computed tomography X-ray equipment), and IEC 61223-3-4 (Evaluation and routine testing in medical imaging departments - Part 3-4: Acceptance tests - Imaging performance of dental X-ray equipment).

The CT modality was evaluated through quantitative testing to verify and measure four key parameters representing CT image quality: noise, contrast, CNR, and MTF 10%. These evaluations

{9}------------------------------------------------

were conducted using the provided reconstruction algorithms, FDK(back projection) and CS(iterative). For PANO and CEPH modalities, quantitative assessments were performed with an emphasis on line pair resolution and low contrast performance using phantoms. The evaluation also included the newly added software functions, all of which met the standards specified in IEC 61223-3-4 and IEC 61223-3-ഗ്

In addition, the dosimetric performance of the subject device and predicate device was compared in terms of DAP. For the PANO modality, although the X-ray detectors differ, both devices demonstrate the technical characteristics of having the same FDD (Focal Spot to Detector Distance) and exposure area.

When tested under the same exposure conditions (image option: High Resolution Mode), the DAP measurement results are similar.

For the CEPH modality, both devices have the same FDD (Focal spot to Detector Distance). However, the predicate device utilizes a scan-type X-ray detector, while the subject device employs a one-shot type X- ray detector. As a result, differences in X-ray detector specifications and exposure conditions make direct comparison under identical conditions difficult.

Although the exposure conditions for the two devices are not equivalent, a DAP evaluation was conducted using the default settings of image option (High Resolution Mode) with the same name to compare DAP values. The results indicate that the DAP of the subject device was more than twice that of the predicate device. However, the subject device utilizes a one-shot type, operating with approximately one-fourth the exposure time of the predicate device. This reduces motion artifacts during X-ray imaging and enhances patient convenience.

For the CT modality, while both devices have the same FDD (Focal Spot to Detector Distance), they differ in X-ray detector specifications. Consequently, their exposure conditions and image acquisition sequences are designed differently. The FOV provided by the two devices also differs, and the variations in exposure conditions make a direct comparison of DAP under identical conditions difficult. However, for

most comparable or equivalent FOVs, the results confirmed a slight increase in DAP for the subject device. Nonetheless, the maximum FOV provided by the subject device demonstrated a reduced radiation dose compared to the predicate device.

Additionally, PANO/CEPH/CT images from the subject and predicate device are evaluated in the Image Quality Evaluation Report and Clinical consideration. The results demonstrated that the subject device performed equivalently or better than the predicate device in overall image quality.

{10}------------------------------------------------

Software/Firmware Verification and Validation Testing

Software verification and validation were conducted and documented as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions." The software for this device was considered as a "Basic Documentation Level", since a failure or latent flaw in the software would not present a hazardous situation with a probable risk of death or serious injury to either a patient, user of the device, or others in the environment of use, prior to the implementation of risk control measures.

Green X 21 (Model: PHT-90CHO) provides the following imaging viewer programs; 2D Image viewing program: EzDent-i (K241114) 3D Image viewing program: Ez3D-i (K231757)

Safety and EMC and Performance Data

Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601- 1:2005+AMD1:2012+AMD2:2020(Edition 3.2), IEC 60601-1-3:2008+AMD1:2013+AMD2:2021 (Edition 2.2),

IEC 60601-2-63:2012+AMD1:2017+AMD2:2021 (Edition 1.2) were performed, and EMC testing were conducted in accordance with standard IEC 60601-1-2:2014+AMD1:2020 (Edition 4.1).

Bench Testing

Non-clinical consideration report was provided in accordance with the FDA guidelines "Guidance for the submissions of 510(k)'s for Solid State X-ray Imaging Devices". Bench testing according to FDA Guidance "Format for Traditional and Abbreviated 510(k)s, Performance Testing - Bench" were performed. Image Quality evaluation report according to IEC 61223-3-5 and IEC 61223-3-5 were also performed.

Applied FDA Guidance

"Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submission" "Off-the-Shelf Software Use in Medical Devices"

"Pediatric Information for X-ray Imaging Device Premarket Notifications" All test results were satisfactory.

{11}------------------------------------------------

11. Conclusions

Comprehensive verification of the subject device has been performed and evaluated as the basis of a determination of substantial equivalence.

The subject device and the predicate device have the same indications for use and exhibit similar technical characteristics. The performance data for the subject device demonstrates equivalence or superiority to the predicate device in several aspects.

The newly applied X-ray detector has been proven in performance tests to perform equivalently or better compared to the predicate device across various performance parameters, including DQE, MTF, and NPS. Additionally, the image quality of the new X-ray detectors, including the newly added modality ENT CT, has been evaluated in compliance with IEC 61223-3-4 and IEC 61223-3-5. Both standard requirements were met. The FACE SCAN modality is not intended for clinical diagnosis but is designed to assist with patient consultations during treatment. This modality has been evaluated according to internal criteria and confirmed to fulfill its intended purpose.

Auto Pano, Smart Focus and Scout modes have already been incorporated in previous devices which obtained premarket clearance from FDA (K231796). It was confirmed that both standard requirements were met through evaluation according to IEC 61223-3-4 and IEC 61223-3-5.

Quality assurance procedures are adhered to, and the specifications and functional requirements have been verified.

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, VATECH Co., Ltd. concludes that Green X 21 (Model: PHT-90CHO) is substantially equivalent to the predicate device as described herein.

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.