(211 days)
The MENTOR™ PliaForm™ Breast Tissue Expander w/ Suture Tabs can be utilized for breast reconstruction after mastectomy, correction of an underdeveloped breast, scar revision and tissue defect procedures. The expander is intended for temporary subcutaneous or submuscular implantation and is not intended for use beyond six months.
The PliaForm™ Breast Tissue Expander w/ Suture Tabs is designed for breast reconstruction, designed for thinner components and a reduction of reinforcement features and minimized insertion profile. The integrated port design consists of a plastic needle guard and a smaller magnetic element to improve MR compatibility (MR Conditional). The next generation design control technology (DCT) facilitates predictable expansion and is shape-optimized to better match the entire implant portfolio. An improved Bufferzone features a new sheeting-based self-sealing technology that offers protection from inadvertent needle punctures across the entire coverage area, and a more flexible design, which improves the overall device flexibility. An internally fixated injection port reduces device bulkiness. The injection port will incorporate a new plastic needle guard with a silicone backing and smaller magnet with reduced magnetic field strength compared to the Artoura™ breast tissue expander devices. This feature reduction improves device compatibility with MR-CT imaging, and external beam radiation (such as photon beam).
Identification of the injection port site is accomplished by use of the CENTERSCOPE™ 2.0 Magnetic Injection Port Locator, which is provided with the Tissue Expander. When the Centerscope™ 2.0 device is placed on top of the skin, the magnetic arm points to the center of the tissue expander's injection port. Injections into the injection port area are made using the provided infusion needle set to inject sterile, pyrogen-free Sodium Chloride U.S.P. Solution.
The MENTOR™ PliaForm™ Breast Tissue Expander w/ Suture Tabs incorporates suture tabs to provide surgeons with the option to attach the device to surrounding tissue for enhanced device stability.
The MENTOR™ PliaForm™ Breast Tissue Expander w/ Suture Tabs are provided sterile in three styles (Low Height, Medium Height, and Tall Height) and various sizes.
The following accessories are packaged with the MENTOR™ PliaForm™ Breast Tissue Expander w/ Suture Tabs:
- Centerscope™ 2.0 Magnetic Injection Port Finder
- Winged Infusion Set (21G)
The provided FDA 510(k) clearance letter is for a physical medical device, the MENTOR™ PliaForm™ Breast Tissue Expander w/ Suture Tabs, and not for an AI/software as a medical device (SaMD).
Therefore, the requested information regarding AI/SaMD specific aspects like:
- A table of acceptance criteria and reported device performance for an AI model
- Sample sizes, data provenance, expert ground truth, adjudication methods for training and test sets of an AI model
- Multi-Reader Multi-Case (MRMC) studies or standalone algorithm performance for an AI model
- Type of ground truth and how it was established for training and test sets of an AI model
cannot be extracted from this document.
The document describes acceptance criteria and studies for a physical medical device, focusing on its mechanical properties, biocompatibility, and compatibility with imaging modalities (MR/CT/RT).
Here's the information that can be extracted, tailored to the context of a physical device:
Acceptance Criteria and Study Proving Device Meets Acceptance Criteria for MENTOR™ PliaForm™ Breast Tissue Expander w/ Suture Tabs
This document describes the premarket notification for a physical medical device, the MENTOR™ PliaForm™ Breast Tissue Expander w/ Suture Tabs. The acceptance criteria and supporting studies are focused on the device's physical and material performance, safety, and compatibility, rather than the performance of an AI algorithm.
1. Table of Acceptance Criteria and Reported Device Performance
The document states that all mechanical performance testing results met their pre-determined acceptance criteria. While specific numerical acceptance criteria (e.g., minimum bond strength in Newtons) are not explicitly detailed in the summary, the categories of tests and the conclusion that they were met serve as the reported performance against the criteria.
| Acceptance Criteria Category (Test Description) | Reported Device Performance |
|---|---|
| Biocompatibility | Concluded no change in biocompatibility safety profile compared to predicate device after full analytical characterization. |
| Mechanical Testing | All pre-determined acceptance criteria were met for tests listed below. |
| Valve Functionality and Device Leakage | Met acceptance criteria. |
| Joint Bond Strength (Shell/Bufferzone) | Met acceptance criteria. |
| Joint Bond Strength (Shell/Patch) | Met acceptance criteria. |
| Joint Bond Strength (Shell/Insert) | Met acceptance criteria. |
| Tensile/Elongation Properties of Elastic Materials (Legs/Bufferzone Joint) | Met acceptance criteria. |
| Tensile/Elongation Properties of Elastic Materials (Legs/Base Joint) | Met acceptance criteria. |
| Shell Tension Set | Met acceptance criteria. |
| Tensile/Elongation Properties of Elastic Materials (Shell) | Met acceptance criteria. |
| Needle Stop Penetration | Met acceptance criteria. |
| Injection Site (Dome) Leakage | Met acceptance criteria. |
| Bufferzone Self-Sealing | Met acceptance criteria. |
| Needle Guard Detachment Force | Met acceptance criteria. |
| Suture Tab Tear | Met acceptance criteria. |
| MR/CT/RT Qualification & Physical Integrity | |
| MR Safety (Induced Force, Torque, Heating) | Qualified. |
| Device Integrity Post Photon Beam Radiation (80 Gy) | Met physical property testing per ASTM F1441-03 (e.g., overexpansion, injection port competence, shell break force, shell tensile set, joint testing, Bufferzone self-sealing, port location) |
| Radiation Dose Distribution Assessment (Phantom Model) | Results for PliaForm™ were within 5% of dosimetry values compared to sham control (phantom without expander). |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes for each mechanical, biocompatibility, or MR/CT/RT test. For a physical device, testing is typically done on a defined number of manufactured units to demonstrate design validation and manufacturing quality. The data provenance is derived from these laboratory tests, not from patient data, and is prospective based on manufacturing and testing protocols. No specific country of origin for test data is mentioned, as is common for laboratory-based performance testing of medical devices.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts
This question is not applicable in the context of this device. "Ground truth" for a physical device is established through engineering specifications, material science principles, and adherence to industry standards (e.g., ASTM F1441-03 for soft-tissue expanders). The "experts" involved would be engineers, material scientists, and quality assurance personnel responsible for designing, manufacturing, and testing the device against these objective criteria.
4. Adjudication Method for the Test Set
This concept is not applicable. Adjudication methods (like 2+1, 3+1) are used in clinical studies or AI evaluations to resolve discrepancies in human interpretations or expert annotations. For physical device testing, results are typically quantitative and objectively measured against pre-defined engineering thresholds. Deviations from these thresholds would be considered failures, not subject to subjective adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
Not applicable. MRMC studies are used to assess the diagnostic performance of human readers, often comparing performance with and without AI assistance. This document pertains to a physical implantable device, not a diagnostic AI system or an imaging modality.
6. Standalone (i.e., algorithm only without human-in-the-loop performance) Performance
Not applicable. This device is not an algorithm or AI system. Its "performance" is inherent to its physical properties and function when implanted.
7. The Type of Ground Truth Used
The "ground truth" for this physical device is defined by:
- Engineering Specifications: Design parameters, material properties, and functional requirements established during product development.
- Industry Standards: Compliance with recognized standards such as ASTM F1441-03 for soft-tissue expanders.
- Biocompatibility Standards: Adherence to ISO 10993 series for medical device biocompatibility.
All test methods measure against these predefined, objective criteria.
8. The Sample Size for the Training Set
Not applicable. This device is not an AI model requiring a training set. Its design and manufacturing processes are developed through engineering R&D, not machine learning.
9. How the Ground Truth for the Training Set Was Established
Not applicable. As this is not an AI model, there is no "ground truth for a training set." The design and manufacturing are based on established engineering principles, material science, and regulatory requirements.
FDA 510(k) Clearance Letter - MENTOR™ PliaForm™ Breast Tissue Expander w/ Suture Tabs
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.07.05
April 24, 2025
Mentor Worldwide LLC
Alicia Botham
Senior Program Lead
31 Technology Drive
Irvine, California 92614
Re: K242963
Trade/Device Name: MENTOR™ PliaForm™ Breast Tissue Expander w/ Suture Tabs
Regulatory Class: Unclassified
Product Code: LCJ
Dated: March 14, 2025
Received: March 19, 2025
Dear Alicia Botham:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Page 2
K242963 - Alicia Botham
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Page 3
K242963 - Alicia Botham
Page 3
Sincerely,
Alicia Hemphill -S (Digitally signed by Alicia Hemphill -S Date: 2025.04.24 15:43:44 -05'00')
Alicia Hemphill (Johnson), M.S.
Assistant Director
DHT4B: Division of Plastic and Reconstructive Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known): K242963
Device Name: MENTOR™ PliaForm™ Breast Tissue Expander w/ Suture Tabs
Indications for Use (Describe):
The MENTOR™ PliaForm™ Breast Tissue Expander w/ Suture Tabs can be utilized for breast reconstruction after mastectomy, correction of an underdeveloped breast, scar revision and tissue defect procedures. The expander is intended for temporary subcutaneous or submuscular implantation and is not intended for use beyond six months.
Type of Use (Select one or both, as applicable):
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 5
510(k) SUMMARY (K242963)
I. SUBMITTER
Mentor Worldwide LLC
31 Technology Drive
Irvine, CA 92614
Contact Person:
Alicia Botham
Regulatory Sr. Program Lead
Phone: 828-775-0302
ABotham1@its.jnj.com
Date Prepared: April 24, 2025
II. DEVICE
| Name of Device | MENTOR™ PliaForm™ Breast Tissue Expander w/ Suture Tabs |
|---|---|
| Common Device Name | Expander, Skin, Inflatable |
| Classification | Unclassified, Pre-Amendment |
| Regulation | Unclassified, Pre-Amendment |
| Panel | General & Plastic Surgery |
| Product Code | LCJ |
III. PREDICATE DEVICE
K161176, Mentor Artoura™ Smooth Breast Tissue Expanders
This predicate has not been subjected to a design-related recall.
No reference devices were used in this submission.
Page 6
IV. DEVICE DESCRIPTION
The PliaForm™ Breast Tissue Expander w/ Suture Tabs is designed for breast reconstruction, designed for thinner components and a reduction of reinforcement features and minimized insertion profile. The integrated port design consists of a plastic needle guard and a smaller magnetic element to improve MR compatibility (MR Conditional). The next generation design control technology (DCT) facilitates predictable expansion and is shape-optimized to better match the entire implant portfolio. An improved Bufferzone features a new sheeting-based self-sealing technology that offers protection from inadvertent needle punctures across the entire coverage area, and a more flexible design, which improves the overall device flexibility. An internally fixated injection port reduces device bulkiness. The injection port will incorporate a new plastic needle guard with a silicone backing and smaller magnet with reduced magnetic field strength compared to the Artoura™ breast tissue expander devices. This feature reduction improves device compatibility with MR-CT imaging, and external beam radiation (such as photon beam).
Identification of the injection port site is accomplished by use of the CENTERSCOPE™ 2.0 Magnetic Injection Port Locator, which is provided with the Tissue Expander. When the Centerscope™ 2.0 device is placed on top of the skin, the magnetic arm points to the center of the tissue expander's injection port. Injections into the injection port area are made using the provided infusion needle set to inject sterile, pyrogen-free Sodium Chloride U.S.P. Solution.
The MENTOR™ PliaForm™ Breast Tissue Expander w/ Suture Tabs incorporates suture tabs to provide surgeons with the option to attach the device to surrounding tissue for enhanced device stability.
The MENTOR™ PliaForm™ Breast Tissue Expander w/ Suture Tabs are provided sterile in three styles (Low Height, Medium Height, and Tall Height) and various sizes.
The following accessories are packaged with the MENTOR™ PliaForm™ Breast Tissue Expander w/ Suture Tabs:
- Centerscope™ 2.0 Magnetic Injection Port Finder
- Winged Infusion Set (21G)
V. INDICATIONS FOR USE
The MENTOR™ PliaForm™ Breast Tissue Expander w/ Suture Tabs can be utilized for breast reconstruction after mastectomy, correction of an underdeveloped breast, scar revision and tissue
Page 7
defect procedures. The expander is intended for temporary subcutaneous or submuscular implantation and is not intended for use beyond six months.
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE
The technological principle for both the proposed and predicate devices is the same. The expansion in both devices is based on incremental filling of a silicone shell with saline fluid to stretch the surrounding tissue. Filling is achieved via an integral injection site (port) with needle guard, and a magnetic injection port finder to locate the port.
The following table summarizes the technological and design features in the predicate and the proposed devices:
| FEATURE | PROPOSED DEVICE (PliaForm™) | PREDICATE DEVICE (Artoura™, K161176) |
|---|---|---|
| Injection Port | With following updates:• Redesigned for internal attachment• Needle guard material change to plastic with silicone backing• Smaller magnet for expander to be MR Conditional | Injection port with magnet and needle guard, surrounded by silicone bufferzone, compatible with 21G needle |
| Self-Sealing Buffer Zone | With following updates:• use of new sheeting-based self-sealing technology• Updated geometry | Silicone elastomer/Dacron mesh subassembly filled with silicone gel |
| Design Control Technology (DCT) Components | With following update(s):• Updated geometry to accommodate shell shape and updated materials | Internal Legs and Insert to provide structural support |
| Shell | With following update(s):• Thinner three-layer smooth silicone elastomer sheeting with identical patient-contacting surface | Smooth Silicone Elastomer Sheeting |
| Orientation Mark | With following update(s):• Blue silicone ink pad printed on 6 o'clock tab (covered by layer of patient-contacting silicone elastomer | Palpable mark (indent) on 6 o'clock suture tab |
Page 8
| Suturing Tabs | With the following update(s):• Additional tabs with suturing holes at 1:30, 4:30, 7:30, and 10:30. | Tabs at 3, 6, 9, o'clock with suturing holes |
VII. PERFORMANCE DATA
Biocompatibility Testing:
The MENTOR™ PliaForm™ Breast Tissue Expander is an implantable device, the contact category according to ISO 10993-1 is: Implant, tissue contacting, permanent (> 30 days). All patient-contacting materials used in the MENTOR™ PliaForm™ Breast Tissue Expander are identical to the materials used in the predicate device. Mentor qualified the introduction of new materials with biocompatibility testing, including full analytical characterization. Results from testing and biocompatibility risk analysis concluded no change in the biocompatibility safety profile, compared to the predicate device.
Mechanical Testing:
Mechanical testing was conducted in accordance with ASTM F1441-03, Standard Specification for Soft-Tissue Expanders. The following mechanical testing were performed:
| ASTM Reference (if applicable) | Document | Description |
|---|---|---|
| ASTM F1441-03 -Section 9.2.5 | TM000388 | Valve Functionality and Device Leakage |
| ASTM F1441-03 -Section 9.2.8.2 | TM000401 | Determination of Joint Bond Strength (Shell/Bufferzone) |
| ASTM F1441-03 -Section 9.2.8.1 | TM000401 | Determination of Joint Bond Strength (Shell/Patch) |
| ASTM F1441-03 -Section 9.2.8.2 | TM000401 | Determination of Joint Bond Strength (Shell/Insert) |
| ASTM F1441-03 -Section 9.2.8.2 / Section 9.2.2.2 | TM000019 | Determination of Tensile/Elongation Properties of Elastic Materials (Legs/Bufferzone Joint) |
| ASTM F1441-03 -Section 9.2.8.2 / Section 9.2.2.2 | TM000019 | Determination of Tensile/Elongation Properties of Elastic Materials (Legs/Base Joint) |
| ASTM F1441-03 -Section 9.2.2.1 | TM000406 | Shell Tension Set |
| ASTM F1441-03 -Section 9.2.2.2 | TM000019 | Determination of Tensile/Elongation Properties of Elastic Materials (Shell) |
| ASTM F1441-03 -Section 9.2.7 | 101031807 | Needle Stop Penetration |
| ASTM F1441-03 -Section 9.2.4 | TM000386 | Injection Site (Dome) Leakage |
| N/A | TM000412 | Bufferzone Self-Sealing |
| N/A | 101040589 | Needle Guard Detachment Force |
| N/A | 100899657 | Suture Tab Tear |
Page 9
All mechanical performance testing results met their pre-determined acceptance criteria, thus demonstrating that the proposed device is substantially equivalent to the predicate device.
Non-Clinical Testing - MR/CT/RT Qualification:
Mentor reduced specific features in the predicate design (Artoura™) to improve device compatibility with MR-CT imaging, and external beam radiation (e.g., photon beam) for radiation therapy. Mentor qualified this capability by performing MR safety related tests including induced force, induced torque and induced heating. Furthermore, Mentor conducted device integrity testing post MR, CT and RT exposures.
Mentor performed physical device integrity testing on PliaForm™ expanders after expanders were exposed to photon beam radiation therapy dosage of 80 Gy. Radiation treatments post mastectomy typically are 1.8 – 2Gy per fraction, with a total radiation treatment dose of approximately 66.6 Gy assuming 33 treatments. Accounting for 120% hot spot, 80 Gy was taken as worst-case representation of clinical radiation therapy exposure. Post radiation exposure, PliaForm™ expanders were shown to meet physical property testing per ASTM F1441-03 "Standard Specification for Soft-Tissue Expander Devices", including tests related to overexpansion, injection port competence, shell break force, shell tensile set and joint testing. Additional testing included self-sealing of the BUFFERZONE™ patch and the ability to sufficiently locate the injection port using the CENTERSCOPE™ 2.0 port locator.
Mentor performed an assessment of radiation dose distribution in a preclinical phantom model using Volumetric Modulated Arc Therapy (VMAT) using commercial radiation planning software. Three metrics were evaluated in one experiment using two different size expanders, including homogeneity index (HI), conformity index (CI) and quality of coverage (QOC) to quantify dosimetry distribution. Results for PliaForm™ were compared to results of the sham control (phantom without any expander) and were found to be within 5% of dosimetry values.
VIII. CONCLUSION
The MENTOR™ PliaForm™ Breast Tissue Expander is substantially equivalent to the legally marketed predicate device, MENTOR™ Artoura™ Breast Tissue Expander with Smooth surface (K161176). The MENTOR™ PliaForm™ Breast Tissue Expander has the same intended use, indications for use, operating principle and technological characteristics as the predicate device. Like the predicate, the MENTOR™ PliaForm™ Breast Tissue Expander meets all performance testing specifications. The successful performance evaluations demonstrate the subject device is substantially equivalent to the predicate device.