K Number
K242948

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-12-23

(89 days)

Product Code
Regulation Number
892.1650
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Adora DRFi is a direct radiography and radioscopy, universal and permanently installed diagnostic X-ray system with a flat panel detector, which allows the acquisition of a wide range of Xray examinations. It is intended for use in generating diagnostic and radioscopic images of human anatomy. The Adora DRFi X-rav system is intended for use by trained and qualified personnel at the entire body of both adult and pediatric patients, and exposures may be acquired with the patient in supine, standing or sitting positions as well as with the patient in e.g., a wheelchair or on a stretcher. Radiographic procedures can take place with the detector in free position as well as fixed in the docking station: whereas radioscopic procedures may only take place when the detector is fixed in the docking station.

The Adora DRFi is not for mammography examinations.

Device Description

The Adora DRFi x-ray system is a ceiling suspended universal x-ray system using direct radiography and radioscopy on flat panels for diagnostic purposes. The system can be used for acquisition of a wide range of radiographic and radioscopic x-ray examinations of the whole body, on both adult and pediatric patients. The patients may be lying on the Adora table, in a hospital bed or on a stretcher, using the flat panel's ability to be placed either inside or outside its housing for radiographic examinations. Radioscopic examinations, however, can only be made with the flat panel in the housing suspended in one of the two telescopes. The patients may also be standing on the floor, sitting on a chair or in a wheelchair. Examinations which can be performed on Adora X-ray systems; Radiography for Pelvis Supine, Pelvis Standing, Knee, and Radioscopy for Esophagus, Tube placement, knee.

The positioning of the ceiling suspended X-ray tube and the flat panel relative to the patient and each other are controlled by the operator, either through a joystick or via pre-programmed auto-positions activated by the operator.

The Adora X-ray system is a main powered, permanently installed system for use in hospitals and x-ray clinics and is intended solely for use by healthcare professionals, trained in and qualified for the use of medical xray equipment for diagnostic purposes.

AI/ML Overview

This FDA 510(k) summary for the Adora DRFi X-ray system does not contain the detailed acceptance criteria or a study proving the device meets an acceptance criterion for an AI/ML medical device.

The document describes a conventional X-ray system (Adora DRFi) and compares its technical specifications to a predicate device (Siemens Multitom Rax). The focus is on demonstrating substantial equivalence based on shared indications for use, target population, and comparable technical aspects (detector technology, generator performance, imaging modalities, positioning capabilities, and table specifications).

Therefore, I cannot provide the requested information, particularly regarding AI/ML device performance, acceptance criteria, sample sizes for test/training sets, expert adjudication, or MRMC studies, as these elements are not detailed in the provided text.

The document indicates that the Adora DRFi utilizes the Canon CXDI-RF Wireless B1 with the Canon CXDI imaging software for image acquisition and post processing, previously cleared under K232298. This suggests that any software-related "AI" aspects (if they exist) would have been cleared under a separate submission (K232298) and are not the primary focus of this 510(k) for the Adora DRFi hardware system.

The "Conclusions" section reiterates that the device is substantially equivalent and that "Testing demonstrates that this device will perform in a manner that is as safe and effective or better than the predicate device." However, the details of this testing are not provided in this summary, especially not in a way that aligns with the typical requirements for an AI/ML device study.

{0}------------------------------------------------

Build Correspondence

Image /page/0/Picture/2 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 23, 2024

NRT X-Ray A/S % Mogens Ravn Managing Director Birkegaardsvej 16 8361 Hasselager Denmark

Re: K242948

Trade/Device Name: Adora DRFi Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: JAA Dated: September 25, 2024 Received: September 25, 2024

Dear Mogens Ravn:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Gabriela M. Rodal -S

for Digitally signed by Gabriela M. Rodal -S

Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K242948

Device Name

Adora DRFi

Indications for Use (Describe)

The Adora DRFi is a direct radiography and radioscopy, universal and permanently installed diagnostic X-ray system with a flat panel detector, which allows the acquisition of a wide range of Xray examinations. It is intended for use in generating diagnostic and radioscopic images of human anatomy. The Adora DRFi X-rav system is intended for use by trained and qualified personnel at the entire body of both adult and pediatric patients, and exposures may be acquired with the patient in supine, standing or sitting positions as well as with the patient in e.g., a wheelchair or on a stretcher. Radiographic procedures can take place with the detector in free position as well as fixed in the docking station: whereas radioscopic procedures may only take place when the detector is fixed in the docking station.

The Adora DRFi is not for mammography examinations.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for NRT. The logo consists of a circular icon with concentric circles emanating from the center, resembling a signal or wave pattern. To the right of the icon, the letters 'NRT' are displayed in a stylized, sans-serif font, with a blue color scheme used throughout the logo.

This summary document has been prepared to meet the conditions of 21 CFR 807.92.

Date of summary:

Dec. 23, 2024

Subject device:

Manufacturer:

  • Trade name: Classification name: Classification: CFR Section: Product Code:
    Adora DRFi Image-intensified fluoroscopic x-ray system Class II 892.1650 JAA
Company name:NRT X-RAY A/S
Address:Birkegaardsvej 168361 HasselagerDenmark
Phone no.:+45 8628 3500
E-mail:nrt@nrtxray.com
Contact person:Mogens Ravn
Phone no.:+45 5123 0330
E-mail:mr@nrtxray.com

Predicate device:

  • Trade name: Classification name: Classification: CFR Section: Product Code: Premarket Notification No.: Manufacturer:

Siemens Multitom Rax

Image-intensified fluoroscopic x-ray system Class II 892.1650 OWB, JAA K152928 SIEMENS AG Sector Healthcare Siemensstraße 1 D-91301 Forchheim, Germany

Device description for new device:

The Adora DRFi x-ray system is a ceiling suspended universal x-ray system using direct radiography and radioscopy on flat panels for diagnostic purposes. The system can be used for acquisition of a wide range of radiographic and radioscopic x-ray examinations of the whole body, on both adult and pediatric patients. The patients may be lying on the Adora table, in a hospital bed or on a stretcher, using the flat panel's ability to be placed either inside or outside its housing for radiographic examinations. Radioscopic examinations, however, can only be made with the flat panel in the housing suspended in one of the two telescopes. The

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for NRT. The logo consists of a circular icon on the left and the text "NRT" on the right. The circular icon is made up of concentric circles, with the innermost circle being the darkest and the outermost circle being the lightest. The text "NRT" is in a sans-serif font and is the same color as the circular icon.

patients may also be standing on the floor, sitting on a chair or in a wheelchair. Examinations which can be performed on Adora X-ray systems; Radiography for Pelvis Supine, Pelvis Standing, Knee, and Radioscopy for Esophagus, Tube placement, knee.

The positioning of the ceiling suspended X-ray tube and the flat panel relative to the patient and each other are controlled by the operator, either through a joystick or via pre-programmed auto-positions activated by the operator.

The Adora X-ray system is a main powered, permanently installed system for use in hospitals and x-ray clinics and is intended solely for use by healthcare professionals, trained in and qualified for the use of medical xray equipment for diagnostic purposes.

The Adora DRFi device will be offered in one main variant:

NRT REF No.:Device name:
04550010Adora DRFi

Electrical safety, EMC and QMS:

The Adora DRFi device conforms to applicable international standards such as IEC 60601-1 including collateral and particular standards, and FDA performance standards 21 CFR 1020.30, 31 and 32. Design and production controls are carried out under a Quality Management System in compliance with 21 CFR 820 and ISO 13485 requirements

The following applied standards have been used developing the Adora DRFi device:

Standard no.Standard titleFDArecognition no.VersionDate
IEC 60601-1 Edition 3.2(2020)Medical electrical equipment -Part 1: General requirements for basic safety and essentialperformance+ US national differences19-49Edition 3.2(2020-08)
IEC 60601-1-2 Edition 4.1(2020)Medical electrical equipment -Part 1-2: General requirements for basic safety and essentialperformance. Collateral standard: Electromagnetic disturbances- Requirements and tests19-36Edition 4.1(2020-09)
IEC 60601-1-3 Edition 2.2(2021)Medical electrical equipment -Part 1-3: General requirements for basic safety and essentialperformance. Collateral Standard: Radiation protection indiagnostic X-ray equipment12-336Edition 2.2(2021-01)
60601-1-6 Edition 3.2(2020)Medical electrical equipment -Part 1-6: General requirements for basic safety and essentialperformance. Collateral standard: Usability5-132Edition 3.2(2020-07)
IEC 60601-2-54 Edition 2.0(2022)Medical electrical equipment -Part 2-54: Particular requirements for the basic safety andessential performance of X-ray equipment for radiography andradioscopy12-348Edition 2.0(2022-09)

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for NRT. The logo consists of a circular icon on the left and the text "NRT" on the right. The circular icon is made up of concentric circles in shades of blue. The text "NRT" is also in blue and has a modern, stylized font.

Standard no.Standard titleFDArecognition no.VersionDate
IEC 60825-1 Edition 3(2014)Safety of laser products - Part 1: Equipment classification andrequirements12-273Edition 3.0(2014-05)
IEC 62304 Edition 1.1(2015)Medical device software - Software life-cycle processes13-79Edition 1.1(2015-06)
IEC 62366-1 Edition 1.1(2020)Medical devices - Part 1: Application of usability engineeringto medical devices5-129Edition 1.1(2020-06)
21CFR Part 1020All applicable sectionsFDA Performance Standards for Ionizing Radiation EmittingProductsN/AN/AN/A
ISO 10993-1: 2018(EN ISO 10993-1:2020)Biological evaluation of medical devices,Part 1: Evaluation and testing within a risk management process2-258Edition 5(2018-08)
ISO 10993-10: 2010(EN ISO 10993-10:2013)Biological evaluation of medical devices,Part 10: Tests for irritation and skin sensitization2-174Edition 3(2010-08)
EN ISO 13485:2016Medical devices - Quality management systems - Requirementsfor regulatory purposesN/AEdition 3(2016-03)
21CFR Part 820FDA Quality System RegulationN/AN/AN/A
EN ISO 14971:2019Medical Devices - Application of risk management to medicaldevices5-125Edition 3(2019-12)
EN ISO 15223-1:2021Medical devices - Symbols to be used with information to besupplied by the manufacturer - Part 1: General requirements5-134Edition 4(2021-07)
DS/EN ISO 20417:2021Information to be supplied by the manufacturer5-135Edition 1(2021-12)

FDA guidance documents used for this submission:

FDA Software Guidance: "Content of Premarket Submissions for Device Software For Industry and Food and Drug Administration Staff," dated June 2023

FDA Cybersecurity Guidance: "Cybersecurity in Medical Devices: Quality System Considerations and content of Premarket Submissions, Guidance for Industry and Food and Drug Administration Staff," dated September 27, 2023

FDA Pediatric Guidance: "Pediatric Information for X-ray Imaging Device Premarket Notifications, Guidance for Industry and Food and Drug Administration Staff," dated November 2017.

Comparison with predicate device:

The Adora DRFi and Siemens Multitom Rax are very similar in all comparable parameters. The predicate device comparison in "Technilogical Differences" section of this summary, contains a comparison between the Adora DRFi and its predicate device, the Siemens Multitom Rax. For the purpose of this 510(k) Summary, please find a comparison of the indications for Use, target population and selected technology aspects below:

Comparison table, Indications for Use and target population:

Adora DRFiMultitom Rax
Indicationsfor use:The Adora DRFi is a direct radiography and radioscopy, universal and permanently installed diagnostic x-ray system with a flat panel detector, which allows the acquisition of a wide range of x-ray examinations. It is intended for use in generatingThe Multitom RAX is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. The Multitom

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for NRT. The logo consists of a circular icon with concentric circles emanating from the center, resembling a signal or wave pattern. To the right of the icon, the letters 'NRT' are displayed in a stylized, sans-serif font, with a blue color scheme used for both the icon and the text.

diagnostic radiographic and radioscopic images ofhuman anatomy. The Adora DRFi x-ray system isintended for use by trained and qualified personnelat the entire body of both adult and pediatricpatients, and exposures may be acquired with thepatient in supine, standing or sitting positions as wellas with the patient in e.g., a wheelchair or on astretcher. Radiographic procedures can take placewith the detector in free position as well as fixed inthe docking station; whereas radioscopic proceduresmay only take place when the detector is fixed in thedocking station.The Adora DRFi is not for mammographyexaminations.RAX is a device intended to visualize anatomicalstructures by converting a pattern of X-ray into a visibleimage.The system has medical applications ranging from butnot limited to gastrointestinal examinations, cranial,skeletal, thoracic and lung exposures as well asexamination of the urogenital tract. The unit may also beused in lymphography, endoscopy, myelography,venography, pediatrics, arthrography, interventionalradiography, digital angiography and digital subtractionangiography (DSA). The Multitom RAX may be used foroutpatient and emergency treatment, as well as formobile transport (wheelchair and bed) examinations.The Multitom RAX is not for mammographyexaminations.
Targetpopulation:Adults and pediatricsAdults and pediatrics

Technological differences:

Both the Adora DRFi and the Siemens Multitom Rax are stationary X-ray systems for radiography and fluoroscopy. They consist of a floor mounted patient table and ceiling suspended X-ray tube and a ceiling suspended flat panel detector. Together with an X-ray generator and a digital imaging system they provide comprehensive image acquisition modes to support radiographic and fluoroscopic imaging procedures. Xray tube and flat panel detector movements are synchronized to provide rotation around a center. The Adora DRFi utilizes the Canon CXDI-RF Wireless B1 with the Canon CXDI imaging software for

image acquisition and post processing, previously cleared under K232298.

Data categoryAdora DRFiSiemens Multitom Rax
General designPatient table with floating top, height adjustable andwith an optional 340° rotation around the baseensures easy patient access all around the table.Suitable for access with wheelchairs and beds andprovides optimal cleaning conditions.Height-adjustable table with table height of 50 cm.360° patient access, including equipment forinterventions.
ModelModel
DetectorCanon CXDI-RF Wireless B1 (FDA K232298)RAXdetector
Technology: a-Si / CslTechnology: a-Si / Csl
Active area: 43 x 42 cmActive area: 43 x 42 cm
Pixel array: 160 micronsPixel array: 148 microns
GeneratorModelModel
CPI Indico IQ, 80kWSiemens Polydoros F80, 80kW
FluoroscopyFluoroscopy
Generatorperformance40 kV to 125 kV / 5 mA to 99 mA Pulsed fluoro40 kV to 110 kV / 4 mA to 84 mA Pulsed fluoro
RadiographyRadiography
40 kV to 150 kV40 kV to 150 kV
ImagingmodalitiesRadiographyRadiography
FluoroscopyFluoroscopy
CineCine
3D image acquisitionDSA (roadmap)3D image acquisition and reconstruction
Technical dataPositioning:Positioning:

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for NRT. The logo consists of a circular icon with concentric circles emanating from the center, followed by the letters "NRT" in a sans-serif font. The color of the logo is a shade of blue.

Up to 999 APRs and associated auto-positions with allimage acquisition parameters can be programmed tomatch users' preferences.The system can move into more than 1000 pre-defined positions.
TabletopTabletop size:Tabletop size:
Width: 70 cmWidth: 75 cm
Length: 293 cmLength: 219 cm
Material: Carbon fiberMaterial: Carbon fiber
Tabletop movement:Tabletop movement:
Longitudinal: ± 10 cmLongitudinal: N/A (fixed)
Transverse: ± 20 cmTransverse: N/A (fixed)
Tabletop movement:Tabletop movement:
Vertical movement: 57 to 92 cm.Vertical movement: 50 to 92 cm.
Table rotation: ± 340° (optional)Table rotation: 0°.
Max. table load: 250 kg.Max. table load: 240 kg.

Comparison table, Design, Modality and Technology

Conclusions:

Based upon the above information we believe that the Adora DRFi is substantially equivalent to the predicate device in that, though there are minor difference, the Indication for Use statement is functionally the same and the intended use of the device is in line with the predicate Siemens Multitom Rax. Testing demonstrates that this device will perform in a manner that is as safe and effective or better than the predicate device. The major difference between the two systems is related to the design of components and that the data and information provided supports a decision of substantial equivalence. The differences between the systems are mainly due to the difference in design philosophy. The Adora system has been developed with the specific aim of improving ergonomics and workflow, while the Siemens competitor is more traditional in their approach. However, the differences in specifications are small and both systems are capable of a similar range of examination procedures. Our conclusion is that new issues of safety or effectiveness are not raised with the introduction of the Adora DRFi device.

§ 892.1650 Image-intensified fluoroscopic x-ray system.

(a)
Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.