K Number
K242926

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-05-30

(248 days)

Product Code
Regulation Number
890.5850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The XBODY Go USA and XBODY Pro USA are machines with electronic muscle stimulation based on EMS technology. The devices are specifically designed as an addition to other sports and for training muscles. They must be used only for healthy muscles and clients, not for rehabilitation purposes.

Both models are intended to stimulate healthy muscles in order to improve or facilitate muscle performance. Neither model is intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. None of the XBODY Go USA or XBODY Pro USA training programs are designed for injured or ailing muscles and its use on such muscles is contraindicated.

For both models, electrical impulses allow the triggering of action potentials on motoneurons of motor nerves (excitations). These excitations of motoneurons are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles.

The various types of muscle work that the XBODY Go USA and XBODY Pro USA can impose on the stimulated muscles are able to improve or facilitate muscle performance. Both models may therefore be considered a technique of muscle training.

Device Description

The XBody Go USA and the XBody Pro USA generate electronic muscle stimulation based on electrical muscle stimulation (EMS) technology. The devices are designed as additions to other sports and for training muscles. They are not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. They are intended for use only by persons with healthy muscles, not for rehabilitation purposes.

AI/ML Overview

Based on the provided FDA 510(k) Clearance Letter and Submission Summary, here's a description of the acceptance criteria and the study that proves the device meets them:

Important Note: The provided document is a 510(k) submission summary, primarily focused on establishing substantial equivalence to a predicate device. It details the similarities and minor differences between the new device and the predicate. For devices like the XBody Go USA and XBody Pro USA, which are "Powered Muscle Stimulators," the primary "performance" and "acceptance criteria" revolve around demonstrating equivalence in electrical safety, electromagnetic compatibility, and fundamental stimulation characteristics, rather than a clinical study measuring a specific physiological outcome (like muscle strength improvement). The document explicitly states, "Any differences that exist between the subject and predicate devices do not affect safety or performance." This implies that the acceptance criteria are largely met by demonstrating adherence to recognized safety and performance standards and showing that any changes do not negatively impact these.

Therefore, the "proof" described is not a typical clinical trial demonstrating superiority or non-inferiority in effectiveness, but rather a demonstration of continued safety and equivalent performance to a device already deemed safe and effective.


Acceptance Criteria and Device Performance Study

The acceptance criteria for the XBody Go USA and XBody Pro USA, as implied by the 510(k) submission, are primarily centered around demonstrating essential safety and equivalent performance to their predicate devices, particularly regarding electrical muscle stimulation parameters and the introduction of new accessories (the XBody DrySuit and DryWear). Since this is a 510(k) for a powered muscle stimulator, the "performance" proved is not a clinical outcome like muscle strength gain, but rather the device's ability to safely and effectively deliver electrical impulses as intended, consistent with the predicate.

1. Table of Acceptance Criteria and Reported Device Performance

Given the nature of this 510(k) (substantial equivalence for a powered muscle stimulator), the "acceptance criteria" are predominantly related to compliance with recognized standards and maintaining equivalent technical specifications and safety profiles. The "device performance" is therefore demonstrated by meeting these standards and showing that the subject device's operation is not adversely affected by the minor changes.

Acceptance Criteria (Implied)Reported Device Performance (as demonstrated)
Electrical Safety (IEC 60601-1, 60601-2-10)Validated through independent testing. Product meets max output voltage, current, phase charge, current density, power density, net charge, waveform shape specifications, and incorporates safety circuits (overload, no-load trip, battery monitoring, watchdog).
Electromagnetic Compatibility (IEC 60601-1-2)Validated through independent testing. Device is compliant.
Usability (IEC 60601-1-6, IEC 62366-1)User interface is described as "large and easily controllable" with clear stimulation parameters and channel identification. Operator training and certification are required. Updated control units maintain ease of use.
Software Life Cycle (IEC 62304)Software differences are minor, aiming to enhance user interface without impacting safety/effectiveness. Compliance with IEC 62304 is stated.
Biocompatibility (ISO 10993-1, -5, -10)Cotton underwear is "biocompatibility certified." New DryWear and DrySuit materials are implied to be compliant through risk management and biocompatibility standards.
Maintaining Intended Use & Indications for UseDevices have "same in indication for use, intended use performance, design, dimensions, and materials as the predicate devices." The new DrySuit is stated to have "acceptable similarity in intended use."
Equivalent Output SpecificationsOutput voltage, current, phase charge, current density, power density, net charge, waveform, burst mode parameters, output frequency, and pulse widths are all "Same" as the predicate.
Equivalent Portability/Environment of Use"Portable with ease," "indoors," "home healthcare environment." Same as predicate.
Safe Use of New Accessories (DrySuit/DryWear)Performance testing of the XBody DrySuit demonstrated "acceptable similarity in intended use." Instructions for use (e.g., wearing undergarments) ensure safe conductivity.

2. Sample Size Used for the Test Set and the Data Provenance

The document does not specify a "sample size" in the context of a data-driven test set for complex performance metrics like accuracy or sensitivity/specificity for an AI/ML device. For this type of device, the "test set" is more conceptual, referring to the various configurations and conditions under which the device's electrical safety, EMC, and fundamental performance characteristics were validated through bench testing.

  • Sample Size: Not applicable in the typical sense for a clinical data set. The validation is based on engineering tests performed on production units or prototypes.
  • Data Provenance: The data comes from "independent testing" of the XBody Actiwear G2 main unit, the control unit, the XBody Training Suit 2.1, the XBody Training Suit 3.0, and the new XBody DrySuit. This implies lab-based testing. The country of origin of the data/testing is not specified, but the manufacturer is XBody Hungary Kft. The testing would be considered prospective in the sense that it's conducted to demonstrate compliance for the submission, rather than retrospective analysis of existing clinical data.

3. Number of Experts Used to Establish Ground Truth and Qualifications

Not applicable. For a powered muscle stimulator's safety and performance testing, "ground truth" is established by calibrated instruments and adherence to engineering standards (e.g., peak voltage, current, waveform shape, and safety circuit functionality). There is no "expert consensus" on an image or clinical finding that requires adjudication.

4. Adjudication Method for the Test Set

Not applicable. There is no human interpretation or subjective assessment of data requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. This is not an AI/ML diagnostic or image-based device that would typically undergo an MRMC study to assess human reader improvement with AI assistance. The device's function is direct electrical muscle stimulation, not assisting in interpretation.

6. Standalone (Algorithm Only) Performance Study

Not applicable. This device is a physical electromedical device, not a standalone algorithm. Its performance is intrinsically linked to its hardware and software working together to deliver electrical impulses.

7. Type of Ground Truth Used

The "ground truth" is based on:

  • Engineering specifications and standards: Adherence to output parameters (voltage, current, frequency, pulse width) as defined by electrical engineering principles and the device's design.
  • Safety standards: Compliance with IEC 60601 series standards for electrical safety and electromagnetic compatibility.
  • Predicate device characteristics: The predicate device itself effectively serves as a "ground truth" or benchmark for acceptable performance and safety, as the submission aims to demonstrate substantial equivalence.

8. Sample Size for the Training Set

Not applicable. This is not an AI/ML device that learns from a "training set" of data in the typical sense of machine learning. The device's behavior is deterministic, based on its programmed logic and hardware design.

9. How the Ground Truth for the Training Set was Established

Not applicable, as there is no "training set" for this type of device. The ground truth for the device's design and functionality is rooted in established electrical engineering principles, medical device safety standards, and the performance characteristics of the legally marketed predicate device.

FDA 510(k) Clearance Letter and Submission Summary

Page 1

May 30, 2025

XBody Hungary Kft.
℅ Bhoomika Joyappa
Senior Regulatory Consultant
Medical Device Academy, Inc.
345 Lincoln Hill Road
Shrewsbury, Vermont 05738

Re: K242926
Trade/Device Name: XBody Go USA, XBody Pro USA
Regulation Number: 21 CFR 890.5850
Regulation Name: Powered Muscle Stimulator
Regulatory Class: Class II
Product Code: NGX
Dated: September 20, 2024
Received: September 24, 2024

Dear Bhoomika Joyappa:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Page 2

K242926 - Bhoomika Joyappa Page 2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K242926 - Bhoomika Joyappa Page 3

Sincerely,

Tushar Bansal -S

for Heather Dean, PhD
Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K242926

Device Name
XBody Go USA, XBody Pro USA

Indications for Use (Describe)

The XBODY Go USA and XBODY Pro USA are machines with electronic muscle stimulation based on EMS technology. The devices are specifically designed as an addition to other sports and for training muscles. They must be used only for healthy muscles and clients, not for rehabilitation purposes.

Both models are intended to stimulate healthy muscles in order to improve or facilitate muscle performance. Neither model is intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. None of the XBODY Go USA or XBODY Pro USA training programs are designed for injured or ailing muscles and their use on such muscles is contraindicated.

For both models, electrical impulses allow the triggering of action potentials on motoneurons of motor nerves (excitations). These excitations of motoneurons are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles.

The various types of muscle work that the XBODY Go USA and XBODY Pro USA can impose on the stimulated muscles are able to improve or facilitate muscle performance. Both models may therefore be considered a technique of muscle training.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.

Page 5

Premarket Notification [510(k) Summary]

K242926

A. General Information

Sponsor's Name: XBody Hungary Kft.
Sponsor's Contact Person: Orsolya Balog (Regulatory Compliance)
Address: György Aladár u. 35-39 Budapest, Hungary 1125
orsolya.balog@XBodyworld.com
www.XBodyworld.com

Submission Contact Person: Bhoomika Joyappa
Senior Regulatory Consultant
Address: Medical Device Academy, Inc
345 Lincoln Hill Rd,
Shrewsbury, VT 05738
bhoomika@fdaestar.com

Date Prepared: 5/28/2025

B. Device

Trade Names of Companion Products: XBody Go USA, XBody Pro USA
Common Name: Powered Muscle Stimulator
Product Classification Code: NGX
Regulatory Class: 2
Device Classification Name: Stimulator, Muscle, Powered, For Muscle Conditioning
Regulation Name: Powered Muscle Stimulator
Regulation Number: 21 CFR 890.5850

C. Identification of Legally Marketed Predicate Device

Predicate Device: XBody Go USA, XBody Pro USA
Manufacturer: XBody Hungary Kft
Predicate 510(k) Number: K221200

Page 6

D. Description of the Devices

The XBody Go USA and the XBody Pro USA generate electronic muscle stimulation based on electrical muscle stimulation (EMS) technology. The devices are designed as additions to other sports and for training muscles. They are not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. They are intended for use only by persons with healthy muscles, not for rehabilitation purposes.

E. Indication for Use

The XBody Go USA and XBody Pro USA are machines with electronic muscle stimulation based on EMS technology. The devices are specifically designed as an addition to other sports and for training muscles. They must be used for only healthy muscles and clients, not for rehabilitation purposes.

Both models are intended to stimulate healthy muscles in order to improve or facilitate muscle performance. Neither model is intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. None of the XBody Go USA or XBody Pro USA training programs are designed for injured or ailing muscles and its use on such muscles is contraindicated.

For both models, electrical impulses allow the triggering of action potentials on motoneurons of motor nerves (excitations). These excitations of motoneurons are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles.

The various types of muscle work that the XBody Go USA and XBody Pro USA can impose on the stimulated muscles are able to improve or facilitate muscle performance. Both models may therefore be considered a technique of muscle training.

F. Technological Characteristics

Compared to the predicate devices, the XBody Go USA and XBody Pro USA are the same in indication for use, intended use performance, design, dimensions, and materials as the predicate devices. The new devices meet the same standards for safety and performance as the predicate devices.

The differences cited in the comparison tables between subject and predicate devices do not affect safety and performance of the subject devices when compared for equivalence to the predicate devices.

Parameter/applicationSubject Device XBody Go USA and XBody Pro USAPredicate Device XBody Go USA and XBody Pro USA (K221200)Assessment of substantial equivalence between Subject and Predicate Devices
510k NumberNAK221200NA

Page 7

ParameterSubject DevicePredicate DeviceAssessment
SubmitterXBody Hungary KftXBody Hungary KftSame
Common or Usual NamePowered muscle stimulator.Powered muscle stimulator.Same
Product codeNGXNGXSame
Classification NameStimulator, Muscle, Powered, For Muscle ConditioningStimulator, Muscle, Powered, For Muscle ConditioningSame
Product ClassificationClass IIClass IISame
Review PanelPhysical MedicinePhysical MedicineSame
Regulated voltageYesYesSame
Regulation Name21 CFR 890.585021 CFR 890.5850Same
Type of usePrescription UsePrescription UseSame
Indications for useThe XBody Go USA and XBody Pro USA are machines with electronic muscle stimulation based on EMS technology. The devices are specifically designed as an addition to other sports and for training muscles. They must be used for only healthy muscles and people (clients), not for rehabilitation purposes. The XBody Go USA and XBody Pro USA are intended to stimulate healthy muscles in order to improve or facilitate muscle performance The XBody Go USA and XBody Pro USA are not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. None of the XBody Go USA or XBody Pro USA training programs are designed for injured or ailing muscles and its use on such muscles is contraindicated.The XBody Go USA and XBody Pro USA are machines with electronic muscle stimulation based on EMS technology. The devices are specifically designed as an addition to other sports and for training muscles. They must be used for only healthy muscles and people (clients), not for rehabilitation purposes. The XBody Go USA and XBody Pro USA are intended to stimulate healthy muscles in order to improve or facilitate muscle performance The XBody Go USA and XBody Pro USA are not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. None of the XBody Go USA or XBody Pro USA training programs are designed for injured or ailing muscles and its use on such muscles is contraindicated.Same

Page 8

ParameterSubject DevicePredicate DeviceAssessment
Indications for use (continued)The XBody electrical impulses allow the triggering of action potentials on motoneurons of motor nerves (excitations). These excitations of motoneurons are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles. The various types of muscle work that the XBody Go USA or XBody Pro USA can impose on the stimulated muscles are able to improve or facilitate muscle performance. The XBody Go USA and the XBody Pro USA may therefore be considered as a technique of muscle training.The XBody electrical impulses allow the triggering of action potentials on motoneurons of motor nerves (excitations). These excitations of motoneurons are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles. The various types of muscle work that the XBody Go USA or XBody Pro USA can impose on the stimulated muscles are able to improve or facilitate muscle performance. The XBody Go USA and the XBody Pro USA may therefore be considered as a technique of muscle training.Same
Portability/Mobile UseThe device is portable with ease. It is not a mobile device; its intended environment is indoors. Its intended use requires a qualified and trained operator.The device is portable with ease. It is not a mobile device; its intended environment is indoors. Its intended use requires a qualified and trained operator.Same
User interfaceThe device can be controlled using the graphical windows appearing on the touchscreen of the external control unit (Microsoft Surface Go 2, Microsoft Surface Go 3 or Microsoft Surface Go 4 for XBody Go USA / Microsoft Surface Pro 7, Microsoft Surface Pro 9 or Microsoft Surface Pro 10 for XBody Pro USA). On the training screen where stimulation controls can be used the START/STOP buttons are large and easily controllable. Stimulation controls for adjusting channel intensities, and all other stimulation parameters are clearly visible and easily controllable. Channel identification is supported with big pictures showing the selected muscle groups. When the stimulation is on, the STOP button is always visible and accessible.The device can be controlled using the graphical windows appearing on the touchscreen of the external control unit (Microsoft Surface Go 2 for XBody Go USA / Microsoft Surface Pro 7 for XBody Pro USA). On the training screen where stimulation controls can be used the START/STOP buttons are large and easily controllable. Stimulation controls for adjusting channel intensities, and all other stimulation parameters are clearly visible and easily controllable. Channel identification is supported with big pictures showing the selected muscle groups. When the stimulation is on, the STOP button is always visible and accessible.Similar. Subject devices have more options for control unit devices. The control units are equipped with the same specialized software.

Page 9

ParameterSubject DevicePredicate DeviceAssessment
Menu/SettingsEasy-to-use multi-choice menu for registered and certified trainers to customize training parameters and stimulation Programs.Easy-to-use multi-choice menu for registered and certified trainers to customize training parameters and stimulation Programs.Same
OperatorTo operate the devices the trainer must complete an XBody US EMS Trainer Course. The certification data received at the end of the course is required when XBody registers trainers in the device database. Only registered trainers can start training stimulation Programs using a passcode.To operate the devices the trainer must complete an XBody US EMS Trainer Course. The certification data received at the end of the course is required when XBody registers trainers in the device database. Only registered trainers can start training stimulation Programs using a passcode.Same
DisplayGO: <10.5" LCD display on the XBody Go USA tablet> PRO: <12.3" or 13" LCD display on the XBody Pro USA tablet>GO: <10.5" LCD display on the XBody Go USA tablet> PRO: <12.3" LCD display on the XBody Pro USA tablet>Similar. The XBody Pro USA subject device has optional control units equipped with slightly bigger displays.
Statistical FunctionsTraining data (trainer, client, date, duration). Client related data. Number of training sessions (today, yesterday, this week, this month, total).Training data (trainer, client, date, duration). Client related data. Number of training sessions (today, yesterday, this week, this month, total).Same
Output specificationsMax Output Voltage = 20.8V @500ΩMax Output Voltage = 20.8V @500ΩSame
Max Output Current = 41.6mA @500ΩMax Output Current = 41.6mA @500ΩSame
Max Phase Charge = 16.64µC@500ΩMax Phase Charge = 16.64µC@500ΩSame
Max Current Density =0.65mA/cm2 @500ΩMax Current Density =0.65mA/cm2 @500ΩSame
Max Power Density =3.46mW/cm2 @500ΩMax Power Density =3.46mW/cm2 @500ΩSame
Net Charge (µC per pulse)0 @500Ω (each phase uses symmetric waveform)0 @500Ω (each phase uses symmetric waveform)Same
ShapeRectangularRectangularSame

Page 10

ParameterSubject DevicePredicate DeviceAssessment
Number of Output channels6 output channels, but maximum 12 independently regulated outputs in case of Training Suit (TS) 2.1 and TS 3.0. 5 output channels, but maximum 10 independently regulated outputs in case of DrySuit6 output channels, but 12 independently regulated outputsSimilar. The subject devices have the option to use the DrySuit which has only 10 channels.
WaveformSymmetric biphasicSymmetric biphasicSame
Burst mode - Pulses per burst100 * 10 = 1000100 * 10 = 1000Same
Burst mode - Bursts per second0.10.1Same
Burst mode - Burst duration (seconds)1111Same
Burst mode - ON Time (seconds)1010Same
Burst mode - OFF Time (seconds)11Same
Burst mode - Duty Cycle: Line (d)/(Line (d)+ Line (e))*0.910.91Same
Additional Features (specify, if applicable)N/AN/ASame
Output frequency1-150Hz1-150HzSame
Positive pulse width50-500usec50-500usecSame
Negative pulse width50-500usec50-500usecSame
Power source – BatteryLi-ion 4x 3.7V (3.4 Ah)Li-ion 4x 3.7V (3.4 Ah)Same

Page 11

ParameterSubject DevicePredicate DeviceAssessment
Size of the electrodesPredefined electrode sizes inside the training suit described in User ManualPredefined electrode sizes inside the training suit described in User ManualSame
Safety circuitsOverload trip detects short-circuit, No-load trip detects circuit break, battery voltage monitoring, hardware error detection at startup, and watchdog monitoring.Overload trip detects short-circuit, No-load trip detects circuit break, battery voltage monitoring, hardware error detection at startup, and watchdog monitoring.Same
PlugsThe XBody Go USA and XBody Pro USA tablets and the XBody Actiwear G2 wireless stimulation unit are connected wirelessly. The XBody Actiwear G2 and the XBody Training Suit are connected with magnetic connectors. The internal cable of the training suit connects to snap fasteners in the suit to which detachable electrodes are attached via waterproof connections.The XBody Go USA and XBody Pro USA tablets and the XBody Actiwear G2 wireless stimulation unit are connected wirelessly. The XBody Actiwear G2 and the XBody Training Suit are connected with magnetic connectors. The internal cable of the training suit connects to snap fasteners in the suit to which detachable electrodes are attached via waterproof connections.Same
Lead Wires – CablesTS2.1: PVC coated ultra-flexible LIFY 0,50 mm2 (256 x 0,05 mm) cables and LiYV 0,56 mm2 (7 x 0,32 mm) in the training suit. Cables are compliant with protected lead wire and patient cable safety requirements TS3.0: Suit cable: Ultra-flexible Microminiature & Miniature PVC Insulated Lead Wire Textile material (yarn): Polyester DS: PFA coated stainless steel filaments cored yarn (1.0mm~1.05mm) 275F*6, 316LTS2.1: PVC coated ultra-flexible LIFY 0,50 mm2 (256 x 0,05 mm) cables and LiYV 0,56 mm2 (7 x 0,32 mm) in the training suit. Cables are compliant with protected lead wire and patient cable safety requirements TS3.0: Suit cable: Ultra-flexible Microminiature & Miniature PVC Insulated Lead Wire Textile material (yarn): PolyesterSimilar. The subject devices have the option to use the DrySuit with its own lead wire

Page 12

ParameterSubject DevicePredicate DeviceAssessment
Conductivity of the ElectrodesWhen using the TS 2.1 or TS 3.0, the client has two options: wearing an XBody cotton underwear (biocompatibility certified), or wearing the DryWear with lightweight thin undergarments that do not obstruct the connection between the electrodes and the skin on the glutes. When using the cotton underwear, the electrodes are contained in cotton covers which must be watered using normal tap water to create conductive media. The cotton textiles hold enough water to provide conductivity during the training. The electrodes are washable and can be disinfected, as described in User Manual. When using the XBody DryWear with undergarments, the electrodes of the TS2.1 or TS3.0 are fixed in the training suit with Velcro attachments, instead of cotton covers. Conductive media in this case is the DryWear. The XBody DrySuit must be worn with lightweight thin undergarments that do not obstruct the connection between the electrodes and the skin on the glutes. The XBody DrySuit transmits electrical stimulation to the surface of the skin without the use of water.The client must wear an XBody cotton underwear (biocompatibility certified). The electrodes are contained in cotton covers which must be watered using normal tap water to create conductive media. The cotton textiles hold enough water to Provide conductivity during the training. The electrodes are washable and can be disinfected, as described in User Manual.Similar
Soldering of the Printed Circuit BoardsIn accordance with the ROHS directive, no lead solder material used.In accordance with the ROHS directive, no lead solder material used.Same
Placement of the electrodesThe electrodes are located at fixed positions in the training suit ensuring proper placement.The electrodes are located at fixed positions in the training suit ensuring proper placement.Same
Reusable padsYesYesSame
Number of programsXBody Go USA: 5 XBody Pro USA: 6XBody Go USA: 5 XBody Pro USA: 6Same
Treatment duration1 min to 60 min maximum1 min to 60 min maximumSame
Environment(s) of useHome healthcare environment, according to IEC 60601-1-11Home healthcare environment, according to IEC 60601-1-11Same
Pulse duration1-10 s1-10 sSame

Page 13

ParameterSubject DevicePredicate DeviceAssessment
Compliance standardsIEC 60529:1989+ A2:2013+C1:2019 IEC 60601-1:2005+ A1:2012+A2:2020 IEC 60601-1-2:2014 +A1:2020 IEC 60601-1-6:2010 +A1:2013+A2:2020IEC 60529:1989+ A2:2013+C1:2019 IEC 60601-1:2005+ A1:2012+A2:2020 IEC 60601-1-2:2014 +A1:2020 IEC 60601-1-6:2010 +A1:2013+A2:2020Same

Page 14

ParameterSubject DevicePredicate DeviceAssessment
Compliance standards (continued)IEC 60601-1-11:2015 +A1:2020 IEC 60601-2-10:2012 +A1:2016 IEC 62304:2006 +A1:2015 IEC 62366-1:2015 +A1:2020IEC 60601-1-11:2015 +A1:2020 IEC 60601-2-10:2012 +A1:2016 IEC 62304:2006 +A1:2015 IEC 62366-1:2015 +A1:2020Same
Compliance with 21 CFR 898?YESYESSame
Automatic No-Load TripYESYESSame
Method of Line Current IsolationN/A (battery-operated device)N/A (battery-operated device)Same
Patient leakage current – normal condition (µA)N/A (battery-operated device)N/A (battery-operated device)Same
Patient leakage current – single fault condition (µA)N/A (battery-operated device)N/A (battery-operated device)Same
Synchronous or alternating?N/A (battery-operated device)N/A (battery-operated device)Same
Method of Channel IsolationHardware and timingHardware and timingSame
Regulated Current or Regulated Voltage?Regulated current and voltageRegulated current and voltageSame
Software/Firmware/Microprocessor Control?YesYesSame
Automatic Over current Trip? Yes/noYesYesSame
Patient Override Control? Yes/noYesYesSame
Patient Override Control methodDisconnect suit connectorsDisconnect suit connectorsSame
Indicator DisplayYesYesSame
Number of Output ModesOne output mode, but with varying stimulation frequency and duty cycle rangesOne output mode, but with varying stimulation frequency and duty cycle rangesSame
Total Dimensions (in.) [W x H x D]9.65 X 6.9 X 0.33 (tablet) 1.96 X 1.96 X 0.98 (Actiwear G2)9.65 X 6.9 X 0.33 (tablet) 1.96 X 1.96 X 0.98 (Actiwear G2)Same
Housing Materials and ConstructionPlasticPlasticSame

Page 15

Table 2

Comparison of Predicate and Subject Training Suits: XBody Training Suit 2.1, XBody Training Suit 3.0 and XBody DrySuit

Training suitsXBody Training Suit 2.1 & Training Suit 3.0 Predicate training suits for XBody Go USA and XBody Pro USA (K221200)XBody DrySuitAssessment
Compatible productXBody Go USA XBody Pro USAXBody Go USA XBody Pro USASame
PartsVest, pants, optional bandsJacket, pantsSame
MaterialTS 2.1: 75% Polyamide, 25% Polyester (Same as TS 2.0) TS 3.0: 40 % Polyamide, 30 % Acrylic, 25 % Polyester, 5 % Elastane75% Polyester, 15% Spandex, 15% ElastaneSimilar
Underwear required during useYes. Option 1: same underwear for XBody Newave USA (K190038). Underwear serves as barrier to direct contact with the body during use Option 2: XBody DryWear worn over lightweight thin undergarments that do not obstruct the connection between the electrodes and the skin on the glutes. XBody DryWear is new to this submission.Yes. The XBody DrySuit must be worn over lightweight thin undergarments that do not obstruct the connection between the electrodes and the skin on the glutes. The XBody DrySuit contacts the skin of the client directly.Different
Number of channels1210Similar, but less available channels
Number of electrodes24 pieces per suit20 pieces per suitSimilar, but less available electrodes
Weight with electrodes and cablesTS2.1: 2.7 – 3.3 kg (6 – 7.3 lbs.) TS3.0: 1.8 - 2.4 kg (4 – 5.3 lbs.)1 kg (2.2 lbs.)Similar
Available SizesTS2.1: XXXS, XS, S, M, L, XL, XXL TS3.0: 2, 3, 3-L, 4, 4-L, 5, 5-W, 6, 7, 8XXS/2, XS/3, S/4, M/5, L/6, XL/7, XXL/8, XXXL/9Similar

Page 16

ParameterSubject DevicePredicate DeviceAssessment
Available Electrode Sizes and ShapesTS2.1: 8 different electrode shapes TS3.0: 10 different electrode shapes9 different electrodesSimilar

Table 3

Similarities of XBody Go USA and XBody Pro USA and Comparison to Predicate Control Unit Parameters

Page 17

ParameterXBody Go USA (Subject Device)XBody Pro USA (Subject Device)Comparison of Subject DevicesSubject Devices vs. Predicate Devices
Control UnitMicrosoft SurfaceMicrosoft SurfaceSameSame
Dimensions of Control Unit9.65" x 6.9" x 0.33" (245 mm x 175 mm x 8.3 mm)11.5" x 7.9" x 0.33" (292 mm x 201 mm x 8.5 mm)SimilarSame
Wireless Stimulation UnitXBody Actiwear G2XBody Actiwear G2SameSame
Training SuitAble to use XBody Training Suit 2.1, XBody Training Suit 3.0 and XBody DrySuitAble to use XBody Training Suit 2.1, XBody Training Suit 3.0 and XBody DrySuitSameSimilar. Predicate can only use XBody Training Suit 2.1 and XBody Training Suit 3.0
Requirement for UnderwearXBody Cotton underwear or XBody DryWear worn over undergarments must be worn during use with XBody Training Suit 2.1 and XBody Training Suit 3.0. Undergarments must be worn with the XBody DrySuit.XBody Cotton underwear or XBody DryWear worn over undergarments must be worn during use with XBody Training Suit 2.1 and XBody Training Suit 3.0. Undergarments must be worn with the XBody DrySuit.SameSimilar. Predicate can only use XBody Training Suit 2.1 and XBody Training Suit 3.0
Training ProgramsManual settings, Muscle Development, Endurance, Relax, Professional training programsManual settings, Muscle Development, Endurance, Relax, Professional, XBeat training programsSimilarSimilar
Professional Training FeaturesIn person, Virtual trainer, Video editorIn person, Virtual trainer, Video editorSameSame
User ManualYes, content varies based on device designYes, content varies based on device designSimilarSimilar
Maximum number of Simultaneous Clients1-2 persons1-6 personsSimilarSimilar
Control Unit StorageTablet: 64 GBTablet: Solid-state drive: 128GBSimilarSimilar
Control Unit Display10.5-inch Tablet Screen12.3-inch Tablet ScreenSimilarSimilar

Page 18

ParameterXBody Go USA (Subject Device)XBody Pro USA (Subject Device)Comparison of Subject DevicesSubject Devices vs. Predicate Devices
Control Unit Battery LifeUp to 9 hours of typical device usageUp to 15 hours of typical device usageSimilarSimilar

Page 19

ParameterXBody Go USA (Subject Device)XBody Pro USA (Subject Device)Comparison of Subject DevicesSubject Devices vs. Predicate Devices
Control Panel Camera5.0MP front-facing camera with 1080p Skype HD video; 8.0MP rear-facing autofocus camera with 1080p HD video Dual Studio Mics 2W stereo speakers with Dolby® Audio5.0MP front-facing camera with 1080p full HD video; 8.0MP rear-facing autofocus camera with 1080p full HD video Dual far-field Studio Mics 1.6W stereo speakersSimilarSimilar
Wireless typeWi-Fi: IEEE 802.11a/b/g/n/ac/ax Bluetooth Wireless 5.0 technologyWi-Fi 6: 802.11ax compatible Bluetooth Wireless 5.0 technologySimilarSimilar
Weight of Control Unit1.2 lbs. (544 g)1.70 lbs. (775 g)SimilarSimilar

The following is a comparison of the similarities of the subject and predicate devices as listed in Table 3.

Control unit: For the predicate devices, the control devices were limited to the Microsoft Surface Go 2 for the XBody Go USA and the Microsoft Surface Pro 7 for the XBody Pro USA. For the subject devices, the available control devices include the Microsoft Surface Go 2, Microsoft Surface Go 3, and Microsoft Surface Go 4 for the XBody Go USA, and the Microsoft Surface Pro 7, Microsoft Surface Pro 9, and Microsoft Surface Pro 10 for the XBody Pro USA. This change does not impact safety or performance, as the newly introduced control units operate the same software developed by XBody Hungary Kft., have similar specifications, and exhibit identical performance.

Number of clients: Using the predicate or the subject device versions of the XBody Go USA, the same number of clients can train at a time. Using the predicate or the subject device versions of the XBody Pro USA, the same number of clients can train at a time.

Assessment: The subject devices and the predicate devices are the same in terms of intended use, performance, and design. Similarities derive from the introduction of the XBody DrySuit and DryWear. The new device meets the same standards for safety as the predicate device.

SpecificationMicrosoft Surface Go 2Microsoft Surface Go 3Microsoft Surface Go 4
Execution platformIntel® Pentium® Gold Processor 4425Y 8th GenIntel® Pentium® Gold Processor 6500Y 10th GenIntel N200
CPU Cores224
RAM4 GB4 GB8 GB
GraphicsIntel® UHD Graphics 615Intel® UHD Graphics 615Intel UHD Graphics
Display10.5", 1920 x 1280, 10-point multi-touch10.5", 1920 x 1280, 10-point multi-touch10.5", 1920 x 1280, 10-point multi-touch
StorageeMMC drive: 64GBeMMC drive: 64GBUFS Drive: 64GB, 128GB, 256GB
Connections1 USB Type-C, HDMI over1 USB Type-C, HDMI over1 USB Type-C, HDMI over

Page 20

ParameterMicrosoft Surface Go 2Microsoft Surface Go 3Microsoft Surface Go 4
Connections (continued)docking adapter or Type-Cdocking adapter or Type-Cdocking adapter or Type-C
Wireless communicationWi-Fi 6, BT/BLE 5Wi-Fi 6, BT/BLE 5Wi-Fi 6, BT/BLE 5.1
SecurityTPM 2.0, UEFI secure bootTPM 2.0, UEFI secure bootTPM 2.0, UEFI secure boot
OS supportLinux kernel supportedLinux kernel supportedLinux kernel supported
ConstructionHandheld tablet with integrated batteryHandheld tablet with integrated batteryHandheld tablet with integrated battery
SpecificationMicrosoft Surface Pro 7Microsoft Surface Pro 9Microsoft Surface Pro 10
Execution platformIntel® i5-1035G4 10th GenIntel® i5-1245U 12th GenIntel Core Ultra 5 Processor 135U Intel Core Ultra 7 Processor 165U
CPU Cores4412
RAM8 GB8 GB8GB, 16GB, 32GB, 64GB
GraphicsIntel® Iris Plus Graphics 615Intel® Iris XE Graphics 615Intel Graphics
Display12.3", 2736 x 1824, 10-point multi-touch13.0", 2880 x 1920, 10-point multi-touch13.0", 2880 x 1920, 10-point multi-touch
StorageSSD drive: 128GBSSD drive: 256GBSSD drive: 256GB, 512GB, 1TB
Connections1 USB Type-C, HDMI over docking adapter or Type-C1 USB Type-C, HDMI over docking adapter or Type-C2 USB Type-C, HDMI over docking adapter or Type-C
Wireless communicationWi-Fi 6, BT/BLE 5Wi-Fi 6E, BT/BLE 5.1Wi-Fi 6E, BT/BLE 5.3
SecurityTPM 2.0, UEFI secure bootTPM 2.0, UEFI secure bootTPM 2.0, UEFI secure boot
OS supportLinux kernel supportedLinux kernel supportedLinux kernel supported
ConstructionHandheld tablet with integrated batteryHandheld tablet with integrated batteryHandheld tablet with integrated battery

Table 6

Standard Number and Title Used to Support XBody Go USA and XBody Pro USA Safety and Performance

Standard NumberApplied Standards by Title

Page 21

Standard NumberApplied Standards by Title
ANSI/AAMI 60601-1:2005/(R)2012 and A1:2012.Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2:2014+A1:2020Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
IEC 60601-1-6:2010+ A1:2013+A2:2020Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance -Collateral standard: Usability
IEC 60601-1-11:2015+A1:2020Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance – Collateral standard: Requirements for medical electrical equipment and medical electrical systemsused in the home healthcare environment
IEC 60601-2-10:2012+A1:2016Medical electrical equipment - Part 2-10: Particularrequirements for the basic safety and essential performance of nerve and muscle stimulators
IEC 62366-1:2015+A1:2020Medical devices - Part 1: Application of usability engineering to medical devices
IEC 62304:2006+A1:2015Medical device software - Software life cycle Processes
IEC 60529:1989+A2:2013+ C1:2019Degrees of Protection Provided by enclosures (IP Code)
EN ISO 14971:2019Medical devices - Application of risk management to medical devices.
ISO 10993-1Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process (Biocompatibility).
ISO 10993-5Biological evaluation of medical devices - part 5: testsfor in vitro cytotoxicity.
ISO 10993-10Biological evaluation of medical devices - part 10: tests for irritation and skin sensitization.
EN ISO 14971:2019Medical devices - Application of risk management to medical devices.
EN ISO 13485:2016Medical Devices – Quality Management Systems –requirements for regulatory purposes

Page 22

Standard NumberApplied Standards by Title
FDA GuidanceFood and Drug Administration Guidance for theContent of Premarket Submissions for Software Contained in Medical Devices 2005
FDA GuidanceGuidance for Industry, FDA Reviewers/Staff and Compliance Guidance Document for Powered MuscleStimulator 510(k)s, June 9, 1999

G. Conclusion of Substantial Equivalence

Electrical safety and electromagnetic compatibility of the XBody Go USA and XBody Pro USA devices have been validated through independent testing in accordance with international standards. This testing covered the XBody Actiwear G2 main unit, the control unit, the XBody Training Suit 2.1, the XBody Training Suit 3.0 (all four used in both the subject and predicate devices), and the XBody DrySuit, which is newly included in this submission.

The subject devices and the predicate devices have the same software programming except for minor functionalities. The software of the subject devices has been updated to include a feature that allows users to select the DrySuit among the available training suits. Independent electrical and electromagnetic compatibility testing, along with performance testing of the subject devices, confirms that these software differences are aimed at enhancing the user interface without impacting safety and effectiveness.

Performance testing of the XBody DrySuit demonstrated acceptable similarity in intended use to the XBody Training Suit 2.1 and the XBody Training Suit 3.0, both of which were previously cleared for the XBody Go USA and XBody Pro USA predicate devices.

The predicate devices' sole applied part is the cotton underwear, which was required to be worn under the XBody Training Suit 2.1 and the XBody Training Suit 3.0. The subject devices have optional applied parts, cotton underwear (same as the predicates') and the XBody DryWear, both of which can be interchangeably worn under the XBody Training Suit 2.1 and the XBody Training Suit 3.0. Additionally, the newly introduced applied parts, the XBody DrySuit and DryWear must be worn with undergarments that do not obstruct the connection between the electrodes and the skin on the glutes.

The Power source of the subject and predicate devices is the same rechargeable battery, supported by bench and usability studies.

Conclusion: XBody Go USA and XBody Pro USA are substantially equivalent to their predicate devices. Any differences that exist between the subject and predicate devices do not affect safety or performance.

END OF DOCUMENT.

§ 890.5850 Powered muscle stimulator.

(a)
Identification. A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.(b)
Classification. Class II (performance standards).