K Number
K242922

Validate with FDA (Live)

Date Cleared
2025-06-18

(267 days)

Product Code
Regulation Number
876.5990
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Extracorporeal Shock Wave Lithotripter, model: U200, is intended to fragment urinary stones in the kidney (renal pelvis and renal calyces) and ureter (upper, middle, and lower ureter).

Device Description

The Extracorporeal Shock Wave Lithotripter, U200, is intended to treat urinary tract stones using ESWL. The Extracorporeal Shock Wave Lithotripter generates high-energy shock waves using a shock wave generator that is focused to produce a highly concentrated stress area at the focal point. These pressure pulses are focused on a specific point in the body where the Urinary calculi are located using assisted movement and manual localization through B-ultrasound. The pressure generated by the shock wave causes the human urinary tract stones to produce physical effects to achieve the therapeutic purpose of crushing stones.

The Extracorporeal Shock Wave Lithotripter, U200, is composed of main unit, control console, patient table, electrical cabinet, power supply unit, and computer.

Control platform is integrated control of electric cabinet, main engine and treatment bed.

Electrical cabinet is used to provide power control of the whole machine, and shock wave source function control.

Main Unit contains the L-shape arm and the support matrix of the impact wave source, and realize the movement of L-shape arm up and down, rotation and oscillation of the impact wave source inside and outside.

Power box is power supply of the whole machine, power supply input voltage 120 V, 60 Hz, 3000VA.

Treatment bed is used for patient support and treatment position, and provides three-dimensional movement to allow easy positioning of the stone in the shock wave focus for lithotripsy and urological procedures.

PC(OptiPlex 7000 Tower) is used for installation of Extracorporeal Shock Wave Lithotripter Computer Control System, and to realize the Movement Control Support Function and the calibration of shock wave target, and image handling.

Principle of operation:

The extracorporeal shock wave lithotripter generates pressure pulses using an electromagnetic shock wave generator, with water serving as the transmission medium. These high energy shock waves with peak acoustic pressure up to 35MPa are focused on a specific point using a lens. In addition, the lithotripter often incorporates an ultrasound-based assisted movement system which locates and aligns the shock wave focus with the urinary tract stone. The shock waves create a stress effect on the stone. After several hundred to around two thousand discharges, the stones are fragmented into smaller pieces which can be excreted from the body.

AI/ML Overview

This document is a 510(k) clearance letter for an Extracorporeal Shock Wave Lithotripter (U200). It describes the device, its intended use, and the studies conducted to demonstrate its safety and effectiveness.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated as quantitative thresholds for the clinical study's primary effectiveness endpoint. Instead, the clinical study aims to demonstrate that the device has a "safety and effectiveness profile that is similar to the predicate device." The reported device performance is presented as successful outcomes in various subgroups. The key performance metrics are "effective rate" for stone fragmentation.

Acceptance Criteria (Inferred)Reported Device Performance (U200)
Primary Effectiveness Endpoint: Achieve an "effective rate" comparable to or better than the predicate device for stone fragmentation. (No specific numerical target provided in the text, but the reported rates are presented as positive outcomes demonstrating effectiveness).Overall Effective Rate: - FAS (Full Analysis Set): 79.2% - PPS (Per-Protocol Set): 86.4%
Localization Accuracy: Auxiliary positioning target error, number of auxiliary positioning operations, and total time of auxiliary positioning operations should demonstrate relative speed, accuracy, and convenience.Localization Performance: - Auxiliary positioning target error: within 3mm - Number of auxiliary positioning operations: within 3 times - Total time of auxiliary positioning operations: within 15min (Stated as "relatively fast, accurate, and convenient, and the system ergonomics performance is good.")
Safety: Device should be safe with no device malfunctions reported in the clinical investigations.Safety: "The incidence of device malfunction does not happen in these clinical investigations." "patients treated by the Extracorporeal Shock Wave Lithotripter (model: U200) are safe."
Subgroup Effectiveness (Inferred, as results are presented): Demonstrate effectiveness across different stone sizes, locations, and densities.Stone Size: - Stones 5-10mm (FAS): 92.9% effective rate - Stones 10-15mm (FAS): 60.0% effective rate - Stones 5-10mm (PPS): 100% effective rate - Stones 10-15mm (PPS): 66.7% effective rate
Stone Location: - Kidney stones (FAS): 80.0% effective rate (Renal calyx: 75.0%, Renal pelvis: 100%) - Ureteral stones (FAS): 78.6% effective rate (Upper ureteral: 70.0%, Middle ureteral: 100%, Lower ureteral: 100%) - Kidney stones (PPS): 80.0% effective rate (Renal calyx: 75.0%, Renal pelvis: 100%) - Ureteral stones (PPS): 91.7% effective rate (Upper ureteral: 87.5%, Middle ureteral: 100%, Lower ureteral: 100%)
Stone Density: - Density < 500HU (FAS): 50% effective rate - Density 500-1000HU (FAS): 81.8% effective rate - Density > 1000HU (FAS): 81.8% effective rate - Density < 500HU (PPS): 50% effective rate - Density 500-1000HU (PPS): 100% effective rate - Density > 1000HU (PPS): 81.8% effective rate

Study Details from the Provided Text:

  1. Sample Size Used for the Test Set and Data Provenance:

    • Clinical Test Set Sample Size: 25 subjects enrolled, 22 subjects completed the study. (4 subjects failed screening, 3 dropped out). The study used both a Full Analysis Set (FAS) and a Per-Protocol Set (PPS) for effectiveness calculation, implying some missing data handled by intent-to-treat (FAS) and per-protocol (PPS).
    • Data Provenance: The clinical investigation was a "multicenter, open-label clinical design." The specific countries of origin for the clinical data are not mentioned, but the manufacturer and correspondent are based in China. The retrospective or prospective nature is not explicitly stated, but "clinical investigations" with "subjects screened," "enrolled," and "followed for a period of follow-up" strongly suggest a prospective study design.
  2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

    • The document does not explicitly state the number of experts or their qualifications used to establish "ground truth" for stone fragmentation. The primary effectiveness endpoint, "effective rate," is defined based on the outcome of the ESWL treatment, likely assessed clinically (e.g., successful fragmentation and clearance/passage of stones). This assessment would typically be done by treating physicians (urologists) or through imaging (e.g., ultrasound, X-ray) reviewed by radiologists, but the document does not specify.
  3. Adjudication Method for the Test Set:

    • The document does not describe any specific adjudication method (e.g., 2+1, 3+1 consensus) for assessing the "effective rate" or other clinical outcomes. It suggests that the clinical outcomes were observed and reported as part of the study.
  4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs. without AI assistance:

    • No. This device is an Extracorporeal Shock Wave Lithotripter, which is a physical medical device for breaking up stones, not an AI-assisted diagnostic or imaging device. Therefore, an MRMC study comparing human readers with and without AI assistance is not applicable and was not performed.
  5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable in the AI sense. The device's primary function is a physical therapy. However, the document does mention "PC(OptiPlex 7000 Tower) is used for installation of Extracorporeal Shock Wave Lithotripter Computer Control System, and to realize the Movement Control Support Function and the calibration of shock wave target, and image handling." It also states "Software Verification and Validation was performed, and it was demonstrated that the software performs as intended." This implies a standalone software performance assessment, but not a standalone diagnostic algorithm performance assessment. The device's "localization/stone targeting system" appears to be an integral part of its operation, not a separate diagnostic algorithm.
  6. The Type of Ground Truth Used:

    • For the clinical effectiveness endpoint ("effective rate"), the ground truth is based on clinical outcomes (successful stone fragmentation and presumed clearance) observed in patients after receiving ESWL treatment. This would typically be assessed by follow-up imaging and clinical evaluation.
    • For Localization Accuracy, the ground truth was likely established by physical measurements comparing the "target location" to the "shock wave focus."
  7. The Sample Size for the Training Set:

    • The document does not mention a separate training set or its sample size for the clinical study. This is a medical device clearance, not an AI model development project where explicit training sets are typically discussed. If any software component used machine learning, details of its training would be relevant, but those are not provided. The clinical study described appears to be a validation/test set for the entire system's clinical performance.
  8. How the Ground Truth for the Training Set Was Established:

    • Since no explicit training set for a machine learning model is mentioned, the method for establishing ground truth for a training set is not applicable based on the provided text.

FDA 510(k) Clearance Letter - Extracorporeal Shock Wave Lithotripter

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID # 04017.07.05
Silver Spring, MD 20993
www.fda.gov

June 16, 2025

Shenzhen Wikkon Precision Technologies Co., Ltd.
℅ Yijie You
General Manager
Qimmiq Medical Consulting Service Co., Ltd.
RM.406, Building C, Run Science Park, No.18 Shenzhou Road,
Huangpu
Guangzhou, Guangdong 510663
CHINA

Re: K242922
Trade/Device Name: Extracorporeal Shock Wave Lithotripter (U200)
Regulation Number: 21 CFR 876.5990
Regulation Name: Extracorporeal Shock Wave Lithotripter
Regulatory Class: Class II
Product Code: LNS
Dated: May 14, 2025
Received: May 15, 2025

Dear Yijie You:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.


U.S. FOOD & DRUG ADMINISTRATION

June 16, 2025

Shenzhen Wikkon Precision Technologies Co., Ltd.
℅ Yijie You
General Manager
Qimmiq Medical Consulting Service Co., Ltd.
RM.406, Building C, Run Science Park, No.18 Shenzhou Road,
Huangpu
Guangzhou, Guangdong 510663
CHINA

Re: K242922
Trade/Device Name: Extracorporeal Shock Wave Lithotripter (U200)
Regulation Number: 21 CFR 876.5990
Regulation Name: Extracorporeal Shock Wave Lithotripter
Regulatory Class: Class II
Product Code: LNS
Dated: May 14, 2025
Received: May 15, 2025

Dear Yijie You:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Page 2

K242922 - Yijie You Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K242922 - Yijie You Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark J. Antonino -S

Mark J. Antonino, M.S.
Assistant Director
DHT3B: Division of Reproductive,
Gynecology, and Urology Devices
OHT3: Office of GastroRenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K242922

Device Name
Extracorporeal Shock Wave Lithotripter (U200)

Indications for Use (Describe)
The Extracorporeal Shock Wave Lithotripter, model: U200, is intended to fragment urinary stones in the kidney (renal pelvis and renal calyces) and ureter (upper, middle, and lower ureter).

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k)#: K242922

510(k) Summary

Prepared on: 2025-06-17

Contact Details

21 CFR 807.92(a)(1)

FieldInformation
Applicant NameShenzhen Wikkon Precision Technologies Co., Ltd.
Applicant AddressRoom 1503, 15th Floor, Building A, Chuanglingtong Science and Technology Building, No.1 Shihua Road, Futian Bonded Area Shenzhen Guangdong 518026 China
Applicant Contact Telephone86-755-25515460
Applicant ContactMr. Jack Kong
Applicant Contact EmailJackkong@wikkon.com
Correspondent NameQimmiq Medical Consulting Service Co., Ltd.
Correspondent AddressRM.406, Building C, Run Science Park, No.18 Shenzhou Road, Huangpu Guangzhou Guangdong 510663 China
Correspondent Contact Telephone(+86)020-822458
Correspondent ContactMr. Yijie You
Correspondent Contact Emailjet.you@qimmiq-med.com

Device Name

21 CFR 807.92(a)(2)

FieldInformation
Device Trade NameExtracorporeal Shock Wave Lithotripter (U200)
Common NameExtracorporeal shock wave lithotripter
Classification NameLithotriptor, Extracorporeal Shock-Wave, Urological
Regulation Number876.5990
Product Code(s)LNS

Legally Marketed Predicate Devices

21 CFR 807.92(a)(3)

Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K103217Extracorporeal Shock Wave Lithotripter, LM-9200 ELMALNS

Device Description Summary

21 CFR 807.92(a)(4)

The Extracorporeal Shock Wave Lithotripter, U200, is intended to treat urinary tract stones using ESWL. The Extracorporeal Shock Wave Lithotripter generates high-energy shock waves using a shock wave generator that is focused to produce a highly concentrated stress area at the focal point. These pressure pulses are focused on a specific point in the body where the Urinary calculi are located using

Page 6

assisted movement and manual localization through B-ultrasound. The pressure generated by the shock wave causes the human urinary tract stones to produce physical effects to achieve the therapeutic purpose of crushing stones.

The Extracorporeal Shock Wave Lithotripter, U200, is composed of main unit, control console, patient table, electrical cabinet, power supply unit, and computer.

Control platform is integrated control of electric cabinet, main engine and treatment bed.

Electrical cabinet is used to provide power control of the whole machine, and shock wave source function control.

Main Unit contains the L-shape arm and the support matrix of the impact wave source, and realize the movement of L-shape arm up and down, rotation and oscillation of the impact wave source inside and outside.

Power box is power supply of the whole machine, power supply input voltage 120 V, 60 Hz, 3000VA.

Treatment bed is used for patient support and treatment position, and provides three-dimensional movement to allow easy positioning of the stone in the shock wave focus for lithotripsy and urological procedures.

PC(OptiPlex 7000 Tower) is used for installation of Extracorporeal Shock Wave Lithotripter Computer Control System, and to realize the Movement Control Support Function and the calibration of shock wave target, and image handling.

Principle of operation:

The extracorporeal shock wave lithotripter generates pressure pulses using an electromagnetic shock wave generator, with water serving as the transmission medium. These high energy shock waves with peak acoustic pressure up to 35MPa are focused on a specific point using a lens. In addition, the lithotripter often incorporates an ultrasound-based assisted movement system which locates and aligns the shock wave focus with the urinary tract stone. The shock waves create a stress effect on the stone. After several hundred to around two thousand discharges, the stones are fragmented into smaller pieces which can be excreted from the body.

Intended Use/Indications for Use

21 CFR 807.92(a)(5)

The Extracorporeal Shock Wave Lithotripter, model: U200, is intended to fragment urinary stones in the kidney (renal pelvis and renal calyces) and ureter (upper, middle, and lower ureter).

Indications for Use Comparison

21 CFR 807.92(a)(5)

The proposed device, model: U200, has the same indication for use as the primary predicate device Extracorporeal Shock Wave Lithotripter (model: LM-9200 ELMA).

Technological Comparison

21 CFR 807.92(a)(6)

The proposed device, model: U200, has the similar characteristics as the primary predicate device Extracorporeal Shock Wave Lithotripter (model: LM-9200 ELMA), such as control console, shock wave generator, Water System,Localization/Imaging System, Patient Table, and acoustic Output Characteristics . Moreover, the bench testing and clinical testing contained in this submission supplied demonstrate that the differences existed between U200 and the primary predicate device Extracorporeal Shock Wave Lithotripter (model: LM-9200 ELMA), do not raise any new questions of safety or effectiveness. Thus, extracorporeal Shock Wave Lithotripter, model: U200, is Substantially Equivalent (SE) to the predicate devices.

Non-Clinical and/or Clinical Tests Summary & Conclusions

21 CFR 807.92(b)

Shenzhen Wikkon performed performance testing to demonstrate and support safety and effectiveness when compared to the predicate and the applicable standards.

The following non-clinical testing was provided in this 510(k) submission:

1) Biocompatibility Testing
According to ISO 10993-1:2018, we performed the biocompatibility evaluation and test of skin surface direct-contacting components with duration of less than 24 hours.

  • (1) In vitro Cytotoxicity test according to ISO 10993-5:2009
  • (2) Skin Sensitization test according to ISO 10993-10:2021 tested
  • (3) Irritation test according to ISO 10993-23:2021 tested

2) Electrical Safety and Electromagnetic Compatibility Testing
The Device was tested and complied with the applicable requirements of the following standards: ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-36 and IEC 61846 demonstrated that the basic safety and performance of the device met the requirements.

3) Software Verification and Validation
Software verification and validation was performed, and it was demonstrated that the software performs as intended according to the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
Cybersecurity verification and validation was performed, and it was demonstrated that the cybersecurity performs as intended according to the FDA Guidance for Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions.

K242922 Page 2 of 3

Page 7

K242922 Page 3 of 3

4) Shock Wave Characterization Measurements testing
The acoustic output characteristics performance with minimum, typical, and maximum shock wave generator output settings of Extracorporeal Shock Wave Lithotripter(model: U200) was tested and complied with the applicable requirements of the standards of IEC 61846 demonstrated that the performance of the device met the requirements.

5) Assessment of Localization Accuracy testing
The maximum deviation of the "target location" from the "shock wave focus" for the localization/stone targeting system was verified to demonstrate the localization/stone targeting system is capable of locating the shock wave focus with sufficient accuracy to target stones as performance claimed.

The clinical investigations adopted a multicenter, open-label clinical design to confirm the functionality and labeling adequacy of the U200 Extracorporeal Shock Wave Lithotripter

A total of 29 subjects were screened in the two centers, 4 subjects failed to be screened, 25 subjects were enrolled, 22 subjects completed the study, and 3 subjects dropped out, with a dropout rate of 12%. The Extracorporeal Shock Wave Lithotripter was used to treat urinary tract stones by ESWL. The stones sizes treated were between 0.5 cm and 1.5 cm. Using ultrasound assisted movement function, the device focuses shock waves on a specific point within the body where the stone is located and break the stone into pieces by the physical effect of the shock waves. Patients were followed for a period of follow-up 2(Day 14 ± 2 days after Treatment 1) and Follow-up 4 (Day 14 ± 2 days after Treatment 2).

When using the research device for lithotripsy treatment, the auxiliary positioning target error of the treatment is within 3mm, the number of auxiliary positioning operations is within 3 times, and the total time of auxiliary positioning operations is within 15min.The positioning operation of the research device is relatively fast, accurate, and convenient, and the system ergonomics performance is good.

Primary effectiveness endpoint:
After the subjects completed the treatment, the effective rate of FAS was 79.2%, the effective rate in the PPS was 86.4%.
In the Full Analysis Set, there were 14 subjects with stones between 5 and 10 mm in long axis, the treatment effective rate of 92.9%, and there were 10 subjects with stones between 1015mm in long axis, the treatment effectiveness rate of 60.0%.
In the Per-Protocol Set, there were 13 patients with stones between 5 and 10 mm in long axis, the treatment effective rate was 100%.
There were 9 subjects with stones between 10
15mm in long axis, the treatment effective rate was 66.7%.
In the Full Analysis Set, there were 10 cases of kidney stones, the treatment efficiency of 80.0%, the treatment efficiency of the renal calyx stones was 75.0%,and the treatment efficiency of the renal pelvis stones was 100%; there were 14 cases of ureteral stones, the treatment efficiency was 78.6% among which the treatment efficacy of upper ureteral stones was 70.0%,that of middle ureteral stones was 100%, and that of lower ureteral stones was 100%.
In the Per-Protocol Set, there were 10 subjects with kidney stones, the treatment efficiency of 80.0% among which the treatment efficiency of renal calyx stones was 75.0%,and that of renal pelvis stones was 100%; and there were 12 subjects with ureteral stones, the treatment efficiency of 91.7% among which the treatment efficiency of upper ureteral stones was 87.5%,that of middle ureteral stones was 100%,and that of lower ureteral stones was 100%.
In the Full Analysis Set, there were 2 subjects with a stone density less than 500HU, the treatment effectiveness rate of 50%. There were 11 subjects with a stone density between 500 and 1000 HU, the treatment effective rate was 81.8%. There were 11 subjects with a stone density greater than 1000 HU, the treatment effective rate of 81.8%.
In the Per-Protocol Set, there were 2 subjects with a stone density less than 500HU, the treatment effectiveness rate of 50%. There were 9 subjects with a stone density between 500 and 1000 HU, the treatment effectiveness rate of 100%. There were 11 subjects with a stone density greater than 1000 HU, the treatment effective rate of 81.8%.

The experiences of physicians have shown that patients treated by the Extracorporeal Shock Wave Lithotripter (model: U200) are safe and having high expectations for the device function. And the user's manual is adequate for the safe operation of Extracorporeal Shock Wave Lithotripter, model: U200. The incidence of device malfunction does not happen in these clinical investigations.

Based on the clinical performance as documented in the clinical study, the Extracorporeal Shock Wave Lithotripter (model: U200) was found to have a safety and effectiveness profile that is similar to the predicate device.

The proposed device, model: U200, has the same intended use and similar characteristics as the primary predicate device Extracorporeal Shock Wave Lithotripter (model: LM-9200 ELMA). Moreover, bench testing and clinical testing contained in this submission supplied demonstrate that the differences existed between U200 and the primary predicate device Extracorporeal Shock Wave Lithotripter (model: LM-9200 ELMA), do not raise any new questions of safety or effectiveness. Thus, extracorporeal Shock Wave Lithotripter, model: U200, is Substantially Equivalent (SE) to the predicate devices.

§ 876.5990 Extracorporeal shock wave lithotripter.

(a)
Identification. An extracorporeal shock wave lithotripter is a device that focuses ultrasonic shock waves into the body to noninvasively fragment urinary calculi within the kidney or ureter. The primary components of the device are a shock wave generator, high voltage generator, control console, imaging/localization system, and patient table. Prior to treatment, the urinary stone is targeted using either an integral or stand-alone localization/imaging system. Shock waves are typically generated using electrostatic spark discharge (spark gap), electromagnetically repelled membranes, or piezoelectric crystal arrays, and focused onto the stone with either a specially designed reflector, dish, or acoustic lens. The shock waves are created under water within the shock wave generator, and are transferred to the patient's body using an appropriate acoustic interface. After the stone has been fragmented by the focused shock waves, the fragments pass out of the body with the patient's urine.(b)
Classification. Class II (special controls) (FDA guidance document: “Guidance for the Content of Premarket Notifications (510(k)'s) for Extracorporeal Shock Wave Lithotripters Indicated for the Fragmentation of Kidney and Ureteral Calculi.”)