K Number
K242837

Validate with FDA (Live)

Device Name
BriefCase-Triage
Date Cleared
2024-10-18

(29 days)

Product Code
Regulation Number
892.2080
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

BriefCase-Triage is a radiological computer aided triage and notification software indicated for use in the analysis of CT scans that include the cervical spine, in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of linear lucencies in the cervical spine bone in patterns compatible with fractures.

BriefCase-Triage uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

The results of BriefCase-Triage are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

Device Description

BriefCase-Triage is a radiological computer-assisted triage and notification software device.

The software is based on an algorithm programmed component and is intended to run on a linuxbased server in a cloud environment.

The BriefCase-Triage receives filtered DICOM Images, and processes them chronologically by running the algorithms on each series to detect suspected cases. Following the Al processing, the output of the algorithm analysis is transferred to an image review software (desktop application). When a suspected case is detected, the user receives a pop-up notification and is presented with a compressed, low-quality, grayscale image that is captioned "not for diagnostic use, for prioritization only" which is displayed as a preview function. This preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification.

Presenting the users with worklist prioritization facilitates efficient triage by prompting the user to assess the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.

The algorithm was trained during software development on images of the pathology. As is customary in the field of machine learning algorithm development consisted of training on manually labeled ("tagged") images. In that process, critical findings were tagged in all CTs in the training data set.

AI/ML Overview

The Aidoc BriefCase-Triage device, intended for triaging cervical spine CT scans for fractures, underwent a retrospective, blinded, multicenter study to evaluate its performance.

Here's a breakdown of the acceptance criteria and study details:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Sensitivity (performance goal ≥ 80%)92.1% (95% CI: 87.5%, 95.4%)
Specificity (performance goal ≥ 80%)92.6% (95% CI: 89.0%, 95.4%)
Time-to-Notification (comparable to predicate)15.1 seconds (95% CI: 14.1-16.2)

2. Sample size used for the test set and the data provenance

  • Sample Size: 487 cases
  • Data Provenance: Retrospective, multicenter study from 6 US-based clinical sites. The cases collected for the pivotal dataset were all distinct in time or center from the cases used to train the algorithm.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Number of Experts: Three
  • Qualifications: Senior board-certified radiologists

4. Adjudication method for the test set

The document does not explicitly state the adjudication method (e.g., 2+1, 3+1). However, it implies that the ground truth was "as determined by three senior board-certified radiologists," suggesting a consensus-based approach among these experts.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There is no MRMC comparative effectiveness study presented in this document that evaluates human readers' improvement with AI assistance versus without AI assistance. The study focuses on the standalone performance of the AI device and its time-to-notification compared to a predicate device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, a standalone algorithm-only performance study was done. The primary endpoints (sensitivity and specificity) and secondary endpoints (time-to-notification, PPV, NPV, PLR, NLR) directly measure the performance of the BriefCase-Triage software itself.

7. The type of ground truth used

The ground truth was established by expert consensus (determined by three senior board-certified radiologists).

8. The sample size for the training set

The document states, "The algorithm was trained during software development on images of the pathology." However, it does not specify the sample size for the training set. It only mentions that the test pivotal study data was sequestered from algorithm development activities.

9. How the ground truth for the training set was established

The ground truth for the training set was established by manually labeled ("tagged") images. The document states, "critical findings were tagged in all CTs in the training data set." This implies expert annotation of the training data.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Aidoc Medical, Ltd. % John J. Smith Partner Hogan Lovells U.S. LLP 555 Thirteenth Street NW WASHINGTON DC 20004

October 18, 2024

Re: K242837

Trade/Device Name: BriefCase-Triage Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological Computer Aided Triage And Notification Software Regulatory Class: Class II Product Code: QAS Dated: September 19, 2024 Received: September 19, 2024

Dear John J. Smith:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb

Jessica Lamb, Ph.D. Assistant Director DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K242837

Device Name

BriefCase-Triage

Indications for Use (Describe)

BriefCase-Triage is a radiological computer aided triage and notification software indicated for use in the analysis of CT scans that include the cervical spine, in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of linear lucencies in the cervical spine bone in patterns compatible with fractures.

BriefCase-Triage uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

The results of BriefCase-Triage are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

| Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

Aidoc Medical, Ltd.'s BriefCase-Triage

Submitter:

Aidoc Medical, Ltd.3 Aminadav St.Tel-Aviv, IsraelPhone:+972-73-7946870
Contact Person:Amalia Schreier, LL.M
Date Prepared:October 17, 2024
Name of Device:BriefCase-Triage
Classification Name:Radiological computer-assisted triage and notification softwaredevice
Regulatory Class:Class II
Product Code:QAS
Primary Predicate Device:BriefCase-Triage for CSF (K203508)
Reference Device:BriefCase-Triage for PE (K241727)
Device Description:

BriefCase-Triage is a radiological computer-assisted triage and notification software device.

The software is based on an algorithm programmed component and is intended to run on a linuxbased server in a cloud environment.

The BriefCase-Triage receives filtered DICOM Images, and processes them chronologically by running the algorithms on each series to detect suspected cases. Following the Al processing, the output of the algorithm analysis is transferred to an image review software (desktop application). When a suspected case is detected, the user receives a pop-up notification and is presented with a compressed, low-quality, grayscale image that is captioned "not for diagnostic use, for prioritization only" which is displayed as a preview function. This preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification.

Presenting the users with worklist prioritization facilitates efficient triage by prompting the user to assess the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.

The algorithm was trained during software development on images of the pathology. As is customary in the field of machine learning algorithm development consisted of training on manually

{5}------------------------------------------------

labeled ("tagged") images. In that process, critical findings were tagged in all CTs in the training data set.

Intended Use / Indications for Use

BriefCase-Triage is a radiological computer aided triage and notification software indicated for use in the analysis of CT scans that include the cervical spine, in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of linear lucencies in the cervical spine bone in patterns compatible with fractures.

BriefCase-Triage uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

The results of BriefCase-Triage are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

Summary of Technological Characteristics

The subject BriefCase-Triage for CSF and the algorithm analysis module for the primary predicate BriefCase-Triage for CSF (K203508) are identical in most aspects and differ mostly with respect to their algorithm performance. In addition, the SW architecture was changed to separate the image communication platform from the BriefCase-Triage SW. The subject device consists of only the algorithm analysis module which can be integrated with image communication platforms that meet the BriefCase-Triage input and output requirements.

Both the primary predicate and subject devices are radiological computer-aided triage and notification software programs. Both devices are artificial intelligence, deep-learning algorithms incorporated in software packages for use with DICOM compliant CT scanners, PACS, and radiology workstations.

Both devices are intended to aid in triage and prioritization of radiological images and utilize the same design of deep learning algorithm trained on medical images. Both devices are intended to provide the specialists with notifications and unannotated, compressed, low-quality, and grayscale preview images of suspect studies for the purpose of preemptive triage.

The subject and predicate BriefCase-Triage devices raise the same types of safety and effectiveness questions, namely, accurate triage of findings within the processed study. It is important to note that, like the predicate, the subject device neither removes cases from the standard of care reading queue nor de-prioritized cases. Both devices operate in parallel with the standard of care, which remains the default option for all cases.

A table comparing the key features of the subject and the primary predicate devices is provided below.

{6}------------------------------------------------

Subject DeviceAidoc BriefCase-Triage for CSFPredicate DeviceAidoc BriefCase-Triage for CSF(K203508)
IntendedUse /Indicationsfor UseBriefCase-Triage is a radiologicalcomputer aided triage and notificationsoftware indicated for use in theanalysis of CT scans that include thecervical spine, in adults or transitionaladolescents aged 18 and older. Thedevice is intended to assist hospitalnetworks and appropriately trainedmedical specialists in workflow triage byflagging and communication of linearlucencies in the cervical spine bone inpatterns compatible with fractures.BriefCase-Triage uses an artificialintelligence algorithm to analyze imagesand highlight cases with detectedfindings in parallel to the ongoingstandard of care image interpretation.The user is presented with notificationsfor cases with suspected findings.Notifications include compressedpreview images that are meant forinformational purposes only and notintended for diagnostic use beyondnotification. The device does not alterthe original medical image and is notintended to be used as a diagnosticdevice.The results of BriefCase-Triage areintended to be used in conjunction withother patient information and based ontheir professional judgment, to assistwith triage/prioritization of medicalimages. Notified clinicians areresponsible for viewing full images perthe standard of care.BriefCase is a radiological computer aidedtriage and notification software indicatedfor use in the analysis of cervical spine CTimages. The device is intended to assisthospital networks and appropriatelytrained medical specialists in workflowtriage by flagging and communication ofsuspected positive findings of linearlucencies in the cervical spine bone inpatterns compatible with fractures.BriefCase uses an artificial intelligencealgorithm to analyze images and highlightcases with detected findings on astandalone desktop application in parallelto the ongoing standard of care imageinterpretation. The user is presented withnotifications for cases with suspectedfindings. Notifications include compressedpreview images that are meant forinformational purposes only and notintended for diagnostic use beyondnotification. The device does not alter theoriginal medical image and is not intendedto be used as a diagnostic device.The results of BriefCase are intended tobe used in conjunction with other patientinformation and based on theirprofessional judgment, to assist withtriage/prioritization of medical images.Notified clinicians are responsible forviewing full images per the standard ofcare.
UserpopulationHospital networks and appropriatelytrained medical specialistsHospital networks and appropriatelytrained medical specialists
Subject DeviceAidoc BriefCase-Triage for CSFPredicate DeviceAidoc BriefCase-Triage for CSF(K203508)
Anatomicalregion ofinterestCervical SpineCervical Spine
DataacquisitionprotocolCT with or without contrast that includesthe cervical spineNon-contrast cervical spine CT scan
Notification-only(/notificationalerts),parallelworkflowtoolYesYes
ImagesformatDICOMDICOM
InterferencewithstandardworkflowNo. No cases are removed fromdesktop app or deprioritizedNo. No cases are removed fromdesktop app or deprioritized
Inclusion/Exclusioncriteria forclinicalperformancetestingInclusion Criteria:CT with or without contrast that• includes the cervical spine(bone and standard kernel)• Scans performed onadults/transitional adults ≥ 18years of age.• Slice thickness: 0.3 mm - 5.0mmExclusion Criteria• All studies that have aninadequate field of viewInclusion Criteria:• Non-enhanced cervical spine CT(bone kernel only).• Scans performed onadults/transitional adults ≥ 18years of age.• Scans performed on CT scannerswith 64 or greater number ofdetectors.• Slice thickness: 0.625 mm - 2.5mm axial or 2-3 mm sagittal or 2-3mm. coronal reconstructions withbone kernel.Exclusion Criteria• All scans that are technicallyinadequate, including motionartifacts, severe metal artifacts, oran inadequate field of view.
AdditionalOperatingPoints4 Additional Operating PointsN/A
AlgorithmArtificial intelligence algorithm withdatabase of images.Artificial intelligence algorithm withdatabase of images.
Subject DeviceAidoc BriefCase-Triage for CSFPredicate DeviceAidoc BriefCase-Triage for CSF(K203508)
Structure- Integrated with image routing modulevia image communication platform (ICP)(image acquisition).- Algorithm module (imageprocessing)- AHS module ( image acquisition);- ACS module (image processing);
- Integrated with desktop applicationfor workflow integration (feed andnon-diagnostic Image Viewer).- Aidoc Desktop Application forworkflow integration (Feed/Worklist(alternate names) and non-diagnosticImage Viewer).

Table 1. Key Feature Comparison

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Aidoc. The logo is in blue and consists of the word "aidoc" in a sans-serif font. There is a small orange circle to the right of the "c" in "aidoc."

{8}------------------------------------------------

Performance Data

Pivotal Study Summary

Aidoc conducted a retrospective, blinded, multicenter study with the BriefCase-Triage software to evaluate the software's performance in identifying CT scans that include the cervical spine images containing Cervical Spine Fracture (CSF) in 487 cases from 6 US-based clinical sites, compared to the ground truth as determined by three senior board-certified radiologists. The cases collected for the pivotal dataset were all distinct in time or center from the cases used to train the algorithm. Test pivotal study data was sequestered from algorithm development activities, and use of the data is managed by appropriate Quality Management System procedures.

Primary endpoints were sensitivity and specificity with an 80% performance goal. Secondary endpoints were BriefCase-Triage time-to-notification compared to the predicate device. Positive Predictive Value (PPV), Negative Predictive Value (NPV), Positive Likelihood Ratio (PLR), and Negative Likelihood Ratio (NLR) were also assessed.

Primary Endpoint

Sensitivity and specificity exceeded the 80% performance goal. Sensitivity was 92.1% (95% Cl: 87.5%, 95.4%) and Specificity was 92.6% (95% Cl: 89.0%, 95.4%).

Secondary Endpoint

In addition, the time-to-notification metric observed for the BriefCase-Triage software, when integrated with a compatible image communication platform, was compared to the equivalent metric of the predicate devices. The BriefCase-Triage time-to-notification includes the time to get the DICOM exam, de-identify it, upload it to the cloud, analyze and send a notification on a positive suspect case back to the desktop application.

The BriefCase-Triage time-to-notification was measured for all True Positive cases (i.e., identified as positive both by the reviewers as well as the BriefCase-Triage device) and is given in Table 2 below. The Table also displays the same metric reported for the predicate BriefCase-Triage PE.

{9}------------------------------------------------

The time-to-notification results obtained for the subject BriefCase-Triage device show comparability with the primary predicate with regard to time savings to the standard of care review. The BriefCase-Triage mean time-to-notification for the subject CSF triage was 15.1 seconds (95% Cl: 14.1-16.2). The time-to-notification for the predicate CSF was 234 seconds (95% Cl 228-246).

Time -to-NotificationMeanEstimateN95% LowerCL95% UpperCLMedian
(seconds)
Predicate K203508Processing Time23448228246234
BriefCase-Triage +ImageCommunicationPlatform Time-To-Notification15.120214.116.213.4

Table 2. Time-to- notification comparison for BriefCase-Triage devices (Seconds)

NPV was 99.1% (95% Cl: 98.5%- 99.4%) and PPV was 58.1% (95% Cl: 47.9%-67.7%) PLR was 12.4965 (95% Cl: 8.263 - 18.898) and NLR was 0.0855 (95% Cl: 0.053 - 0.137).

Table 3. Sensitivity and Specificity Poolability Across Slice Thicknesses
-------------------------------------------------------------------------------
Slice Thickness[mm]SensitivitySpecificity
0.3-1.093.9% (95% CI: 85.0%-98.3%)94.2% (95% CI: 87.8%-97.8%)
1.0-2.093.5% (95% CI: 85.5%-97.9%)88.1% (95% CI: 80.2%-93.7%)
2.0-5.088.3% (95% CI: 77.4%-95.2%)96.3% (95% CI: 89.6%-99.2%)

Thus, the reported similar time-to-notification data demonstrates that when using the subject BriefCase-Triage for CSF the clinician may have the same benefit in time saving as with the predicate BriefCase-Triage for CSF.

As can be seen in Table 4 the mean age of patients whose scans were reviewed for CSF was 62.0 years, with a standard deviation of 19.6 years. Gender distribution was 47.2% male, and 49.3% female (Table 5). Scanner distribution can also be found in Table 6 below.

{10}------------------------------------------------

Table 4. Descriptive Statistics for Age

MeanStdMinMedianMaxN
Age(Years)62.019.618.065.090.0487

Table 5. Frequency Distribution of Gender *

GroundTruthResultsGender
FemaleMaleAll
N%N%N%
Positive8016.4%10822.2%18838.6%
Negative16032.9%12225.1%28257.9%
All24049.3%23047.2%487*100%
  • 17 cases were unknown for gender, 3 negative and 14 positive.

Table 6. Frequency Distribution of Manufacturer

ManufacturerN%
GE MEDICALSYSTEMS15030.8%
SIEMENS13227.1%
TOSHIBA12225.1%
Philips8317.0%
Total487100%

Clinical Subgroups and Confounders:

Pathologies present in negative cases: Oncology; Trauma; Neuro & Vascular; Chronic Spinal Diseases and None of the above.

Additional Operating Points:

In addition to the default operating point that was selected to maximize both sensitivity and specificity, four additional operating points (AOP) were selected to maximize specificity while maintaining a lower bound 95% confidence interval of 80% for sensitivity and spectively:

{11}------------------------------------------------

Operating PointsSensitivity % (95% CI)Specificity % (95% CI)
AOP195.5% (95% CI: 91.7%- 97.9%)86.0% (95% CI: 81.4%- 89.8%)
AOP286.1% (95% CI: 80.6%- 90.6%)95.1% (95% CI: 91.9%- 97.3%)
AOP393.6% (95% CI: 89.3%- 96.5%)89.5% (95% CI: 85.3%- 92.8%)
AOP487.6% (95% CI: 82.3%- 91.8%)93.7% (95% CI: 90.2%- 96.2%)

Table 7. Additional Operating Points

In summary, performance goals were achieved for the default and four additional operating points. Combined with the comparison results of time-to-notification metric with the predicate device, these data establish the achievement by the subject BriefCase-Triage of preemptive triage in the range of several minutes.

Conclusions

The subject BriefCase-Triage for CSF and the predicate BriefCase-Triage for CSF (K203508) are intended to aid in prioritization and triage of radiological images for the indications for suspected positive findings of Cervical Spine Fracture pathologies. Both devices are software packages consisting of deep learning Al algorithms that process images and produce analysis results, which are displayed to the user by a prioritization alert and a compressed, low-quality, grayscale, unannotated preview image. In both devices, the labeling clearly states that the devices are not for diagnostic use and instructs the user to further evaluate and diagnose based only on the original images in the local PACS.

Both devices operate in parallel to the standard of care workflow in the sense that they do not change the original image, do not provide any marking on the output preview, do not remove images from the standard of care FIFO queue and do not de-prioritize cases, thus not disturbing standard interpretation of the images. Both devices notify the radiologist of time-sensitive critical cases within the range of several minutes, and thus contribute similarly to the standard of care workflow turnaround time reduction through preemptive triage.

The technological characteristics of both devices are identical with the exception of additional operating points and minor changes in inclusion and exclusion criteria. These differences do not raise new questions of safety and effectiveness. Thus, the subject device is substantially equivalent. The subject BriefCase-Triage device for CSF is thus substantially equivalent to the primary predicate BriefCase-Triage for CSF (K203508).

N/A