K Number
K242805

Validate with FDA (Live)

Date Cleared
2025-01-09

(114 days)

Product Code
Regulation Number
892.1750
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

FinScan F350 is CBCT, panoramic and cephalometric x-ray imaging system. The device is intended to radiographic examination of the dento-maxillofacial and TMJ structure for diagnostic support for adult and pediatric patients. Cephalometric image also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment.

The device is to be operated and used by dentists or other legally qualified health care professionals.

Device Description

Dental Cone-beam Computed Tomography (model: FinScan F350) uses cone-beam computed tomography (CBCT) technology through X-ray cone-beam, panoramic radiography, cephalometric radiography to produce images of the dento-maxillofacial and TMJ structure to provides diagnostic details for dental clinics and dental hospitals.

This product mainly consists of column, rotating unit, tube head assembly, CBCT/panoramic detector, cephalometric detector, the workstation and image processing software.

AI/ML Overview

The provided text describes the 510(k) premarket notification for the Dental Cone-beam Computed Tomography (model: FinScan F350) and its substantial equivalence to a predicate device. However, the document does not contain explicit acceptance criteria or a comprehensive study demonstrating that the device meets specific performance criteria in a quantitative manner as typically expected for device performance studies.

The document focuses on regulatory compliance, outlining how the FinScan F350 is substantially equivalent to the predicate device RCT700 (K213226) based on technological characteristics and adherence to recognized standards. It mentions "Performance bench testing" and "Clinical images acquired using FinScan F350 were evaluated by two US board certified experienced dentists to be of acceptable clinical effectiveness," but it does not provide specific metrics or the study details requested.

Here's an analysis based on the available information, noting what is not present:

1. Table of Acceptance Criteria and Reported Device Performance:

The document mentions performance bench testing that measured spatial resolution, CNR (Contrast-to-Noise Ratio), and homogeneity for CT image quality. It states that "The results demonstrate that the subject device is as effective, and perform as well as the predicate device." However, specific numerical acceptance criteria or the reported performance values for these metrics are NOT provided in the text.

2. Sample Size Used for the Test Set and Data Provenance:

The document briefly mentions "Clinical images acquired using FinScan F350 were evaluated..." but does not specify the sample size (number of images or cases) used for this evaluation, nor does it provide information on the data provenance (e.g., country of origin, retrospective or prospective nature of the image acquisition).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

"Clinical images acquired using FinScan F350 were evaluated by two US board certified experienced dentists to be of acceptable clinical effectiveness for the proposed indications for use."

  • Number of experts: 2
  • Qualifications: US board certified experienced dentists. (Specific years of experience are not mentioned).

4. Adjudication Method for the Test Set:

The document states "evaluated by two US board certified experienced dentists," but it does not specify an adjudication method (e.g., whether consensus was required, if a third expert was used in case of disagreement, or if individual evaluations were aggregated).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

No, an MRMC comparative effectiveness study is NOT mentioned in the provided text. The document describes a substantial equivalence submission, which typically focuses on demonstrating that a new device is as safe and effective as a legally marketed predicate device, rather than quantifying performance improvements with AI assistance. The device is described as an imaging system, not an AI-based diagnostic tool that assists human readers.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

The FinScan F350 is a medical imaging device (CBCT, panoramic, cephalometric x-ray system) that produces images for diagnostic support. It includes "image processing software." The performance evaluation mentioned (bench testing and clinical image evaluation by dentists) is about the general image quality and clinical effectiveness of the imaging system itself. The concept of "standalone algorithm performance" (without human-in-the-loop) as typically applied to AI/CADe systems is not directly applicable or discussed for this device based on the provided text, as it's an image acquisition system, not primarily a standalone diagnostic algorithm.

7. The Type of Ground Truth Used:

For the "clinical images evaluation report," the ground truth seems to be implicitly expert judgment/consensus of the two US board-certified experienced dentists, who deemed the image quality "acceptable clinical effectiveness." Pathology or outcomes data are not mentioned as ground truth sources for this evaluation.

8. The Sample Size for the Training Set:

The document does not mention any training set size. This is expected as the document describes a hardware imaging device with associated standard image processing software, not a machine learning or AI algorithm that requires a "training set" in the context of deep learning. The software mentioned (FinScan and FinScanFW) performs control, image acquisition, reconstruction, storage, browsing, and processing, which are typically deterministic software functions, not machine learning models.

9. How the Ground Truth for the Training Set was Established:

As no training set is mentioned as applicable to this type of device (traditional imaging system with processing software, not an AI/ML algorithm), this information is not applicable and not provided.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 9, 2025

Changzhou Finno Medical Technology Co., Ltd. % Jie Zhu RA Manager First Floor East and Third Floor, Building 5 No. 28 Yandanghe Road, Xinbei District Changzhou, Jiangsu 213000 CHINA

Re: K242805

Trade/Device Name: Dental Cone-beam Computed Tomography (model: FinScan F350) Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: OAS Dated: September 13, 2024 Received: December 11, 2024

Dear Jie Zhu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang

Lu Jiang, Ph.D. Assistant Director Diagnsotic X-ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K242805

Device Name

Dental Cone-beam Computed Tomography (model: FinScan F350)

Indications for Use (Describe)

FinScan F350 is CBCT, panoramic and cephalometric x-ray imaging system. The device is intended to radiographic examination of the dento-maxillofacial and TMJ structure for diagnostic support for adult and pediatric patients. Cephalometric image also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment.

The device is to be operated and used by dentists or other legally qualified health care professionals.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains the logo for FINNO. To the left of the name is a graphic of blue squares arranged in a pattern. To the right of the name is the company name in Chinese.

510(k) Summary

1. Submitter

Manufacturer:Changzhou Finno Medical Technology Co., Ltd.First Floor East and Third Floor, Building 5, No. 28 Yandanghe Road,Xinbei District, 213000 Changzhou, Jiangsu, China
Phone:+86 400-677-9698
Contact Person:E-mail:Jie Zhuamanda@sifary.com
Date Prepared:September 13, 2024

2. Device

Name of Device:Dental Cone-beam Computed Tomography (model: FinScan F350)
Common Name:Dental Cone-beam Computed Tomography
Classification Name:Computed tomography x-ray system
Regulation Number:21 CFR Part 892.1750
Regulatory Class:Class II
Product Code:OAS
Review Panel:Radiology

3. Predicate Device

Trade/Device Name:RCT700
Classification Name:Computed tomography x-ray system
Regulation Number:21 CFR Part 892.1750
Regulatory Class:Class II
Product Code:OAS
Review Panel:Radiology
Manufacturer:RAY Co., Ltd
510 (k) Number:K213226

No reference devices were used in this submission.

4. Device Description

Dental Cone-beam Computed Tomography (model: FinScan F350) uses cone-beam computed tomography (CBCT) technology through X-ray cone-beam, panoramic radiography, cephalometric radiography to produce images of the dento-maxillofacial and TMJ structure to provides diagnostic details for dental clinics and dental hospitals.

This product mainly consists of column, rotating unit, tube head assembly, CBCT/panoramic detector,

{5}------------------------------------------------

cephalometric detector, the workstation and image processing software.

5. Indications for Use

FinScan F350 is CBCT, panoramic and cephalometric x-ray imaging system. The device is intended to radiographic examination of the dento-maxillofacial and TMJ structure for diagnostic support for adult and pediatric patients. Cephalometric image also includes wrist to obtain carpus images for growth and maturity assessment for orthodontic treatment.

The device is to be operated and used by dentists or other legally qualified health care professionals.

6. Intended patient population

The patient population can be the possible person who can be taken X-ray diagnostic radiation exposure.

There is no restriction for ethnic group, Gender, weight, health, or condition.

We recommend patients for X-ray diagnostic radiation exposure to be over 5 years old.

Radiation exposure is a concern in both adults and children. However, children are more sensitive to radiation than adults and have a longer life expectancy.

Radiation risk is higher in young patients, as they have more rapidly dividing cells than adults.

The vounger the patient, the more sensitive they are. Using the same exposure parameters on a child as used on an adult may result in larger doses to the child. There is no need for these larger doses to children, and X-ray settings can be adjusted to reduce dose sigmificantly while maintaining diagnostic image quality.

Please refer the web pages regarding additional pediatric information.

FDA's Pediatric X-ray Imaging webpage:

http://www.fda.gov/Radiation-EmittingProducts/RadiationEmittingProductsandProcedures

7. Comparison of Technological Characteristics with the Predicate Device

The comparison between the overall specifications of predicate device (RCT700) and the new device (FinScan F350) is shown in the following table.

ParameterSubject DevicePredicate Device
ManufacturerChangzhou Finno MedicalTechnology Co., Ltd.RAY Co., Ltd.
Device nameFinScan F350RCT700
510(K) Number/K213226
Indications for UseFinScan F350 is CBCT,panoramic and cephalometricx-ray imaging system. Thedevice is intended toradiographic examination ofthe dento-maxillofacial andTMJ structure for diagnosticRCT700 is CBCT andpanoramic x-ray imagingsystem with cephalometric.Which is intended toradiographic examination of thedento-maxillofacial, sinus,TMJ, Airway and ENT
support for adult and pediatricpatients. Cephalometric imagealso includes wrist to obtaincarpus images for growth andmaturity assessment fororthodontic treatment.The device is to be operatedand used by dentists or otherlegally qualified health careprofessionals.structure for diagnostic supportfor adult and pediatric patients.And a model scan is included asan option. Cephalometric imageis also includes wrist to obtaincarpus images for growth andmaturity assessment fororthodontic treatment.The device is to be operatedand used by dentists or otherlegally qualified heath careprofessionals.
Mode of OperationContinuousoperation/intermittent loadingContinuous operation withintermittent, stated permissibleloading
3D TechnologyCBCT Cone beam ComputedTomographyCBCT Cone beam ComputedTomography
Performance Specification1) CBCT Computedtomography2) Panoramic3) Cephalometric (scan type)1) CBCT Computedtomography- Patient2) Panoramic3) Cephalometric (optional)- One shot type- Scan type
Detector TypeCTTFTTFT
PanoramicTFTTFT
CephalometricCMOS (Scan type)CMOS (Scan type)TFT (One shot type)
Main Components- Column- Handle- Touchpad- Chin rest and bite fork, Chinrest (TMJ), Chin rest(Toothless)- Temple support- CBCT/ Panoramic detector- Laser light area- X-ray exposure light- Rotating unit- Tube head assembly- Nasal rod- Ear rod- Cephalometric detector- Ceph Apparatus- Vertical Carriage- Rotator- X-RAY Generator- X-ray tube- High Frequency Generator- Column- Touch monitor (panel)- Detector- Chinrest- Head rest- Automatic Collimator- Exposure switch- Emergency stop switch- Console PC set
- Emergency stop switch- On/off switch- Workstation
Display TypeNo displayTFT LCD type(Normally black)*1280x800 pixel
ClassClass I with type B appliedparts according to IEC 60601-1Class I with type B appliedparts according to IEC 60601-1
Focal size0.50.5
Field of View(CT)Max.160x100 mmMax.160x100 mm
X-ray Tube Voltage60~100kVp60~100kVp
X-ray Tube Current6mA~10mA1~17mA
Total FiltrationMin. 2.8 mm Al equivalentMin. 2.8 mm Al equivalent
DetectorPixel SizeCT100μmFXDD-0606CA: 119µmJupi0606X: 100μm
Panoramic100μmFXDD-0606CA: 119µmJupi0606X: 100μm
Cephalometric100µmCeph (Scan)XID-C24DC: 100µm
Ceph(One shot)FXDD-1012CA: 124µmFXRD-1717VA: 140µm
Scan TimeCTGeneral:18sFast: 9sbelow 14sec
PanoramicArch:14sTMJ: 6sbelow 14sec
CephalometricLateral, Frontal:8sCarpus: 6sCeph[Scan type] : below 20secCeph[One shot type]: below2sec
Format CompatibleDICOM 3.0 Format compatibleDICOM 3.0 Format compatible
Image Viewing SoftwareFinScan (2D and 3D imageprocessing software installedon the workstation);FinScanFW (embeddedsoftware in the single chip ofthe host)RayScan(Cleared under K182614)
Image AcquisitionGiga-Ethernet NetworkGiga-Ethernet Network
Total HeightMax 2,281mmMax 2,296mm
Weight220kg (485lb)±10%Max 212.5kg (468.5lb) ± 10%
Applicable StandardsIEC 60601-1IEC 60601-1-3IEC 60601-2-63IEC 60601-1IEC 60601-1-3IEC 60601-2-63

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image contains a logo with a blue color scheme. On the left side of the logo, there is a square-shaped graphic made up of smaller squares. To the right of the graphic, the word "FINNO" is written in a sans-serif font, and below that, there are three Chinese characters.

{7}------------------------------------------------

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for FINNO. The logo consists of a blue square pattern on the left, followed by the word "FINNO" in blue. Below the word "FINNO" are two Chinese characters, also in blue. The logo is simple and modern, and the blue color gives it a professional look.

IEC 60601-1-2IEC 60601-1-2
--------------------------------

The above comparison table show that the subject and predicate devices have substantially similar technology characteristics. Differences show up in the display type, detector pixel size, scan time, and image viewing software. All of the parameters of the subject device comply with 21 CFR 1020.30. 1020.31, 1020.33 and related IEC standards. The differences of the device do not raise new issues of safety and effectiveness.

8. Performance Data

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility

For the biocompatibility of the parts that can come into contact with the patient, ISO 10993-1:2018 "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" has been considered. In clinical use, a disposable plastic firm is required to cover the Bite fork. The Bite fork is does not come in direct contact with the patient's oral tissue and teeth normally on the chance that contact does occur if the disposable cover is broken. The Chin rest (TMJ), Chin rest (Toothless), Temple support, Ear rod, Nasal rod come in direct contact with patient's intact skin surfaces. These parts are made of Polycarbonate (PC). The battery of testing included the following tests:

  • . Cytotoxicity
  • . Irritation
  • . Sensitization

According to the test results, no biocompatibility problem has been observed.

Electrical Safety and Electromagnetic Compatibility (EMC)

The electrical safety and EMC data included in the submission is in compliance with the following FDA recognized standards:

  • o IEC 60601-1 Edition 3.2 2020-08
  • ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 [Incl. AMD2:2021]
  • . IEC 60601-1-2 Edition 4.1 2020-09
  • IEC/TR 60601-4-2 Edition 1.0 2016-05 .
  • IEC 60601-1-3 Edition 2.2 2021-01 .
  • . IEC 60601-2-63 Edition 1.2 2021-05
  • IEC 60601-1-6 Edition 3.2 2020-07 o

The laser light was tested to be conformed to IEC 60825-1:2014.

{9}------------------------------------------------

Software Verification and Validation Testing

Software architecture of FinScan F350 can be divided into two groups:

  • Software installed on the workstation, which is used to control the imaging system to complete . the case management and patient image acquisition, reconstruction, storage, browsing, processing and three-dimensional image reconstruction and processing process.
  • . Software embedded in the single chip of the host. which is used to control the motion of the device and the acquisition of image data.

Software verification and validation testing has been conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions", and "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions". The software documentation level for this device was considered as Basic Documentation Level, since a failure or flaw of software function could not present a hazardous situation with a probable risk of death or serious injury, either to a patient, user of the device, or others in the environment of use.

Bench Testing

Performance bench testing was conducted as part of design control to ensure the substantial equivalence of FinScan F350 to the predicate device. The testing was performed according to the requirements of 21 CFR Part 1020.30, 1020.31, 1020.33, IEC 61223-3-4 and IEC 61223-3-7, and FDA Guidance "Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices". The representative indicators for CT image quality, such as spatial resolution, CNR, and homogeneity were measured. The results demonstrate that the subject device is as effective, and perform as well as the predicate device.

Clinical Considerations

Clinical images acquired using FinScan F350 were evaluated by two US board certified experienced dentists to be of acceptable clinical effectiveness for the proposed indications for use. The results have been summarized in the clinical images evaluation report. The overall image quality was acceptable for all cases and image types in various imaging mode.

9. Conclusions

In reference to the comparison information provided in the substantial equivalence table, most of the technological characteristics of the subject device are similar with the predicate device. The differences have been addressed and analyzed with no new issues of safety and effectiveness raised. We believe that Dental Cone-beam Computed Tomography (model: FinScan F350) is safe and effective, and has no new indication for use. FinScan F350 is substantially equivalent to the predicate device.

N/A