K Number
K242802

Validate with FDA (Live)

Date Cleared
2024-11-08

(52 days)

Product Code
Regulation Number
862.3650
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CLUNGENE Fentanyl Home Test Cassette is competitive binding, lateral flow immunochromatographic assay for the qualitative detection of Fentany] in human urine at the cut off concentration of 1.0 ng/mL. This test provides only a preliminary result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas Chromatography-Mass Spectrometry (GC-MS) or Liquid Chromatography-Mass Spectrometry (LC-MS) is the preferred confirmatory method. Evaluate preliminary positive results carefully. For in vitro diagnostic use only.

The CLUNGENE Fentanyl Test Cassette is competitive binding, lateral flow immunochromatographic assay for the qualitative detection of Fentanyl in human urine at the cut off concentration of 1.0 ng/mL.

This test provides only a preliminary result. A more specific alternative chemical must be used to obtain a confirmed presumptive positive result. Gas Chromatography-Mass Spectrometry (GC-MS). Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results. For in vitro diagnostic use only.

Device Description

The CLUNGENE Fentanyl Tests are immunoassays intended for the qualitative detection of fentanyl in human urine. Each CLUNGENE Fentanyl Test device consists of a Test Cassette, a Dropper and a package insert. Each Test Cassette is sealed with sachets of desiccant in an aluminum pouch.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the CLUNGENE Fentanyl Home Test Cassette and CLUNGENE Fentanyl Test Cassette, based on the provided FDA 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly list "acceptance criteria" in a structured table. However, based on the performance characteristics presented, we can infer the criteria considered for demonstrating substantial equivalence. The reported performance is presented in the context of these implied criteria.

For qualitative detection of Fentanyl in human urine at a cut-off concentration of 1.0 ng/mL, the key performance characteristics evaluated were:

Acceptance Criteria (Implied)Reported Device Performance
Precision (agreement around cutoff)Lot 1: -100%, -75%, -50% cut off: 100% negative (60-/0+) -25% cut off: 96.7% negative (58-/2+) Cut off: 55% negative, 45% positive (33-/27+) +25%, +50%, +75%, +100% cut off: 100% positive (60+/0-) Lot 2: -100%, -75%, -50%, -25% cut off: 100% negative (60-/0+) Cut off: 43.3% negative, 56.7% positive (26-/34+) +25%, +50%, +75%, +100% cut off: 100% positive (60+/0-) Lot 3: -100%, -75%, -50% cut off: 100% negative (60-/0+) -25% cut off: 96.7% negative (58-/2+) Cut off: 53.3% negative, 46.7% positive (32-/28+) +25%, +50%, +75%, +100% cut off: 100% positive (60+/0-)
StabilityStable at 39-86º F for 24 months (based on accelerated stability study).
Interference (no impact from common substances)No interference observed at 100µg/mL (or specified concentrations) for a wide range of physiological, pathological, and common drug substances listed. This includes substances like Acetone (1000 mg/dL), Ethanol (1%), Glucose (3000 mg/dL), Ibuprofen, Caffeine, etc.
Specificity (cross-reactivity with similar compounds)Demonstrated defined cross-reactivity profiles for Fentanyl analogs and related substances (e.g., Acetyl fentanyl (100%), Isobutyryl fentanyl (40%), Carfentanil (2%), Sufentanil (4%)). No cross-reactivity for a list of 27 other opioid compounds tested at 100 µg/mL.
Effect of Urine Specific Gravity and pH (robustness to urine variability)All samples at and above +50% Cut-Off were positive; all samples at and below -50% Cut-Off were negative for urine specific gravities from 1.000 to 1.035 and pH from 4 to 9.
Method Comparison (agreement with confirmatory method)Overall agreement near cutoff: For samples near the cutoff (between -50% and +50% of cutoff), there were a few discordant results, indicating expected variability around the cutoff. - Example: 1 positive result for a true negative sample (0.953 ng/mL) by Operator 1. - Example: 2 negative results for true positive samples (1.073 ng/mL, 1.083 ng/mL) by Operator 1 and 2. Key finding: Generally high agreement for samples significantly below or above the cutoff.
Lay-User Performance (ease of use and accuracy by target users)- 100% correct results for -100%, -75%, -50%, +25%, +50%, +75% of cutoff concentrations. - 95% correct results for -25% of cutoff (1 positive result out of 20 samples). - All lay users indicated the instructions were easy to follow. - Package insert has a Flesch-Kincaid reading grade level less than 7.

2. Sample Size Used for the Test Set and Data Provenance

  • Precision Study: For each of 9 fentanyl concentrations (-100% to +100% cut off), 60 tests were performed per lot (2 tests/day/operator for 10 days, across 3 lots).
  • Interference Study: Urine samples were prepared with various interfering substances, with target fentanyl at 50% below and 50% above Cut-Off levels. Tested using three batches of each device. Specific sample sizes per substance are not given, but refers to "compounds that showed no interference at a concentration of 100µg/mL or specified concentrations".
  • Specificity Study: Drug metabolites and other components were tested. Not explicitly stated as "test set" samples, but these are part of the analytical validation.
  • Effect of Urine Specific Gravity and pH: Urine samples with varying specific gravity and pH were spiked with fentanyl at 50% below and 50% above cut-off levels. Tested using three lots of devices.
  • Method Comparison Studies (Clinical Samples): 80 unaltered clinical urine samples (40 negative and 40 positive) were used.
  • Lay-user Study: 140 lay persons tested samples. 20 samples were prepared per concentration level spanning -100% to +75% of the cutoff.

Data Provenance:
The document does not explicitly state the country of origin for the clinical samples or for the samples used in the precision and lay-user studies. The method comparison study used "unaltered clinical samples." The precision and lay-user studies used "spiked fentanyl in negative samples" or "spiked fentanyl into drug free-pooled urine specimens," which implies these were lab-prepared samples. The overall study appears to be retrospective in terms of sample collection for clinical samples, but the testing itself would be prospective for evaluating the device.

3. Number of Experts and Qualifications for Ground Truth for Test Set

  • Method Comparison Studies (Clinical Samples): The ground truth was established by LC/MS (Liquid Chromatography-Mass Spectrometry). This is a highly accurate analytical method and is stated as the "preferred confirmatory method" in the Indications for Use. No human experts are listed for establishing this ground truth, as it is an instrumental analysis.
  • Precision Study, Lay-user Study, Interference, Specificity, Effect of Urine Specific Gravity and pH: The fentanyl concentrations in spiked samples were "confirmed by LC/MS". Again, LC/MS served as the ground truth.

4. Adjudication Method for the Test Set

  • Method Comparison Studies (Clinical Samples): Samples were "blind labeled" and compared to LC/MS results. This suggests a direct comparison method. Discrepancies (discordant results) were noted, but there's no mention of an adjudication process by human experts to resolve these. The LC/MS result is considered the definitive truth.
  • Precision, Interference, Specificity, Effect of Urine Specific Gravity and pH, Lay-user Study: The ground truth was based on the known spiked concentrations confirmed by LC/MS. No human adjudication method is described or necessary for these types of analytical validation studies.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was done. The study compares the device's performance against LC/MS (a gold standard analytical method) and evaluates lay-user performance, but not in a comparative effectiveness study with human readers with and without AI assistance. This device is a rapid diagnostic test cassette, not an AI-assisted diagnostic tool.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance)

Yes, this is essentially a standalone (device-only) performance assessment. The "CLUNGENE Fentanyl Test Cassette" is a lateral flow immunoassay. Its performance is read visually from the lines on the cassette. While human interpretation is involved in reading the result, the core "performance" (sensitivity, specificity, precision, etc.) is inherent to the chemical reactions and design of the device itself. The lay-user study evaluates the human interpretation aspect too.

7. Type of Ground Truth Used

The ground truth used throughout the studies was analytical confirmation by LC/MS (Liquid Chromatography-Mass Spectrometry). This is considered a highly reliable and quantitative method for determining the presence and concentration of fentanyl.

8. Sample Size for the Training Set

The document does not describe a "training set" in the context of machine learning or AI. This is a traditional immunoassay device, which typically does not involve machine learning models that require training sets. The studies described are for analytical and clinical validation of the device's inherent performance characteristics.

9. How the Ground Truth for the Training Set Was Established

As there is no mention of a "training set" for a machine learning algorithm, this question is not applicable to the provided document.

{0}------------------------------------------------

November 8, 2024

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is a stylized representation of the Department of Health & Human Services seal. To the right of the symbol is a blue square containing the acronym "FDA" in white letters, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.

Hangzhou Clongene Biotech Co.,Ltd. % Jenny Xia Director LSI International Inc 504E Diamond Ave., Suite H Gaithersburg, Maryland 20877

Re: K242802

Trade/Device Name: CLUNGENE Fentanyl Home Test Cassette: CLUNGENE Fentanyl Test Cassette Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate Test System Regulatory Class: Class II Product Code: NGL Dated: September 14, 2024 Received: September 17, 2024

Dear Jenny Xia:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Joseph A. Digitally signed by Joseph A Kotarek -S Kotarek -S Date: 2024.11.08 08:51:37 -05'00' Joseph Kotarek, Ph.D. Branch Chief Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K242802

Device Name

CLUNGENE Fentanyl Home Test Cassette; CLUNGENE Fentanyl Test Cassette

Indications for Use (Describe)

The CLUNGENE Fentanyl Home Test Cassette is competitive binding, lateral flow immunochromatographic assay for the qualitative detection of Fentany] in human urine at the cut off concentration of 1.0 ng/mL. This test provides only a preliminary result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas Chromatography-Mass Spectrometry (GC-MS) or Liquid Chromatography-Mass Spectrometry (LC-MS) is the preferred confirmatory method. Evaluate preliminary positive results carefully. For in vitro diagnostic use only.

The CLUNGENE Fentanyl Test Cassette is competitive binding, lateral flow immunochromatographic assay for the qualitative detection of Fentanyl in human urine at the cut off concentration of 1.0 ng/mL.

This test provides only a preliminary result. A more specific alternative chemical must be used to obtain a confirmed presumptive positive result. Gas Chromatography-Mass Spectrometry (GC-MS). Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results. For in vitro diagnostic use only.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY K242802

1. Date:September 13, 2024
2. Submitter:Hangzhou Clongene Biotech Co., Ltd.No.1 Yichuang Road, Yuhang DistrictHangzhou, China
3. Contact person:Jenny XiaLSI International Inc.504E Diamond Ave., Suite HGaithersburg, MD 20877Telephone: 301-525-6856Email: jxia@lsi-consulting.org
4. Device Names:CLUNGENE Fentanyl Home Test CassetteCLUNGENE Fentanyl Test Cassette
Classification:Class 2
Product CodeClassificationRegulation SectionPanel
NGLII21 CFR § 862.3650Opiate Test SystemToxicology (91)
    1. Predicate Devices:
      AllTest Fentanyl Urine Test Cassette (K233417)
    1. Indications for Use
      The CLUNGENE Fentanyl Home Test Cassette is competitive binding, lateral flow immunochromatographic assay for the qualitative detection of Fentanyl in human urine at the cut off concentration of 1.0 ng/mL.

This test provides only a preliminary result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas Chromatography-Mass Spectrometry (GC-MS) or Liquid Chromatography-Mass Spectrometry (LC-MS) is the preferred confirmatory method. Evaluate preliminary positive results carefully. For in vitro diagnostic use only.

binding, The Fentanyl Test Cassette is competitive CLUNGENE lateral flow immunochromatographic assay for the qualitative detection of Fentanyl in human urine at the cut off concentration of 1.0 ng/mL.

This test provides only a preliminary result. A more specific alternative chemical method must be used to obtain a confirmed presumptive positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem massspectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results. For in vitro diagnostic use only.

    1. Device Description
      The CLUNGENE Fentanyl Tests are immunoassays intended for the qualitative detection of fentanyl in human urine. Each CLUNGENE Fentanyl Test device consists of a Test Cassette, a

{5}------------------------------------------------

Dropper and a package insert. Each Test Cassette is sealed with sachets of desiccant in an aluminum pouch.

    1. Substantial Equivalence Information
      A summary comparison of features of the CLUNGENE Fentanyl Test and the predicate device is provided in following table.
ItemDevicePredicate - K233417
Indication(s)for UseFor the qualitative determination offentanyl in human urine.Same
Calibrator and Cut-OffValuesFentanyl (FTY)1 ng/mlSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays basedon the principle of antigen antibodyimmunochemistry.Same
Type of TestQualitativeSame
Specimen TypeHuman UrineSame
Intended UseFor OTC useSame
ConfigurationsCassetteSame
Storage4-30°CSame

Table 1: Features Comparison of CLUNGENE Fentanyl Test and the Predicate Device

    1. Test Principle
      The CLUNGENE Fentanyl Tests are immunoassays based on the principle of competitive binding.

During testing, a urine specimen migrates upward by capillary action. Fentanyl, if present in the urine specimen below 1 ng/mL, will not saturate the binding sites of antibody-coated particles in the test device. The antibody-coated particles will then be captured by immobilized fentanyl conjugate and a visible colored line will show up in the test line region. The colored line will not form in the test line region if the fentanyl level exceeds 1 ng/mL because it will saturate all the binding sites of anti-fentanyl antibodies.

To serve as a procedural control, a colored line will always appear at the control line region indicating that proper volume of specimen has been added and membrane wicking has occurred.

10. Performance Characteristics

    1. Analytical Performance
    • Precision a.

Precision studies were carried out for samples with concentrations of -100% cut off, -75% cut off, -50% cut off, -25% cut off, cut off, +25% cut off, +50% cut off, +75% cut off and +100% cut off. These samples were prepared by spiking fentanyl in negative samples. Each fentanyl concentration was confirmed by LC/MS. All sample aliquots were blindly labeled by the person who prepared the samples and didn't take part in the

{6}------------------------------------------------

sample testing. For each concentration, tests were performed two tests per lot per day per operator for 10 days in a randomized order.

Lot Number-100% cut off-75% cut off-50% cut off-25% cutoffcut off+25% cut off+50% cut off+75% cut off+100% cut off
Lot 160-/0+60-/0+60-/0+58-/2+33-/27+60+/0-60+/0-60+/0-60+/0-
Lot 260-/0+60-/0+60-/0+60-/0+26-/34+60+/0-60+/0-60+/0-60+/0-
Lot 360-/0+60-/0+60-/0+58-/2+32-/28+60+/0-60+/0-60+/0-60+/0-

c. Stability

The devices are stable at 39-86º F for 24 months based on the accelerated stability study.

d. Interference

Potential interfering substances found in human urine of physiological or pathological conditions were added to drug-free urine and target drug fentanyl urine with concentrations at 50% below and 50% above Cut-Off levels. These urine samples were tested using three batches of each device. Compounds that showed no interference at a concentration of 100µg/mL or specified concentrations are summarized in the following tables.

AcetaminophenErythromycinOctopamine
Acetone (1000 mg/dL)Ethanol (1%)O-Hydroxyhippuric acid
AcetophenetidinFenofibrateOlanzapine
Acetylsalicylic acidFenoprofenOmeprazole
AcyclovirFluphenazineOxalic acid (100 mg/dL)
Albumin (100mg/dL)FurosemideOxazepam
AlbuterolGalactose (10 mg/dL)Oxolinic acid
AminopyrineGamma globulin (500 mg/dL)Oxymetazoline
AmitriptylineGatifloxacinPapaverine
AmobarbitalGentisic acidPenicillin G
AmoxicillinGlibenclamidePerphenazine
AmpicillinGliclazidePhencyclidine
ApomorphineGlucose (3000 mg/dL)Phenelzine
Ascorbic acidHemoglobinPhenobarbital
AspartameHydralazinePrednisone
AtropineHydrochlorothiazideProcaine
Benzilic acidHydrocortisonePromethazine
Benzoic acidHydroxytyraminePropoxyphene (50 mg/dL)
BenzoylecgonineIbuprofenPropranolol
BilirubinImipraminePropylthiouracil
Boric acid (1%)IsoproterenolPseudoephedrine
BupropionIsoxsuprineQuinine
CaffeineKetamineRanitidine
CaptoprilKetoprofenRibavirin
CarbamazepineLabetalolRiboflavin (10 mg/dL)
Chloral hydrateLevonorgestrelRifampicin
ChloramphenicolLidocaineSalicylic acid
ChlorothiazideLoperamideSecobarbital
ChlorpheniramineMaprotilineSerotonin (5-Hydroxytyramine)
ChlorpromazineMDMASimvastatin
CholesterolMeperidineSulfamethazine
ClarithromycinMeprobamateSulindac
ClomipramineMethamphetamineTetrahydrocortisone 3-(β- Dglucuronide)

{7}------------------------------------------------

ClonidineMethapyrileneTetrahydrocortisone 3-acetate
CortisoneMethaqualoneTetrahydrozoline
CotinineMethoxyphenamineTheophylline
CreatinineMetoprolol tartrateThiamine
CyclobenzaprineMetronidazole (300 µg/mL)Thioridazine
DeoxycorticosteroneMifepristoneTriamterene
DesipramineMontelukast sodiumTrifluoperazine
DextromethorphanN-AcetylprocainamideTrimethoprim
DiazepamNaCl (4000 mg/dL)Tyramine
DiclofenacNalidixic acidUrea (2000 mg/dL)
DiflunisalNaloxoneUric acid
DigoxinNaltrexoneValproic acid (250 µg/mL)
DiphenhydramineNaproxenVenlafaxine
DL-TryptophanNiacinamideVerapamil
DL-TyrosineNicotineZomepirac
DoxepinNifedipineβ-Estradiol
Ecgonine methyl esterNorethindroneΔ9-THC
EphedrineNortriptyline/
Epinephrine hydrochlorideNoscapine

e. Specificity

To test specificity, drug metabolites and other components that are likely to interfere in urine samples were tested using three batches of device. The lowest concentration that caused a positive result for each compound are listed below.

Fentanyl (Cutoff=1ng/mL)Minimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Acetyl fentanyl1100%
Acrylfentanyl1100%
Isobutyryl fentanyl2.540%
Ocfentanil520%
Butyryl fentanyl520%
Furanyl fentanyl1010%
Valeryl fentanyl520%
(±) β-hydroxythiofentanyl2.540%
4-Fluoro-isobutyrylfentanyl1010%
Para-fluorobutyryl fentanyl520%
Para-fluoro fentanyl2.540%
Carfentanil502%
Sufentanil254%
Alfentanil75000.01%
ω-1-Hydroxy fentanyl25000.04%
(±)-3-cis-methyl fentanyl751.33%
Despropionyl fentanyl (4-ANPP)20000.05%
β-hydroxyfentanyl1001%
Thiofentanyl502%
Cyclopropyl Fentanyl1010%
Trazodone10000.1%
Remifentanil> 100000<0.001%

{8}------------------------------------------------

Norcarfentanil> 100000<0.001%
Norfentanyl> 100000<0.001%
Acetyl norfentanyl> 100000<0.001%

The following opioid compounds were tested at a concentration of 100 µg/mL. A negative result was obtained for all these compounds. There is no cross-reactivity for these compounds using the CLUNGENE® Fentanyl Test Cassette.

6-Acetyl morphineNaloxone
AmphetamineNaltrexone
BuprenorphineNorbuprenorphine
BuprenorphineglucuronideNorcodeine
CodeineNorketamine
DextromethorphanNormeperidine
DihydrocodeineNormorphine
EDDPNoroxycodone
EMDPOxycodone
FluoxetineOxymorphone
HeroinPentazocine (Talwin)
HydrocodonePipamperone
HydromorphoneRisperidone
KetamineTapentadol
LevorphanolThioridazine
MeperidineTilidine
MethadoneTramadol
MorphineTramadol-O-Desmethyl
Morphine-3-glucuronideTramadol-N-Desmethyl

f. Effect of Urine Specific Gravity and Urine pH

To investigate the effect of urine specific gravity and urine pH, urine samples, with 1.000 to 1.035 specific gravity or urine samples with pH 4 to 9 were spiked with target fentanyl at 50% below and 50% above Cut-Off levels. These samples were tested using three lots of device. Results were all positive for samples at and above +50% Cut-Off and all negative for samples at and below -50% Cut-Off.

2. Comparison Studies

Method comparison studies for the CLUNGENE Fentanyl Test were performed by three different operators. Operators ran 80 (40 negative and 40 positive) unaltered clinical samples. The samples were blind labeled and compared to LC/MS results. The results are presented in the tables below.

NegativeLow Negative by LC/MS (less than -50%)Near Cutoff Negative by LC/MS (Between -50% and cutoff)Near Cutoff Positive by LC/MS (Between the cutoff and +50%)High Positive by LC/MS (greater than +50%)
Operator 1Positive0012118
Negative10161310
Operator 2Positive0012018
Negative10161320

{9}------------------------------------------------

Operator
Positive0012018
3Negative10161320
Discordant Results
OperatorSample IDLC/MS Result (ng/mL)Rapid Test Result
Operator 1FYL570.953Positive
Operator 2FYL660.987Positive
Operator 3FYL660.987Positive
Operator 1FYL381.073Negative
Operator 2FYL191.083Negative
Operator 2FYL551.004Negative
Operator 3FYL241.073Negative
Operator 3FYL551.004Negative

3. Lay-user study

A lay user study was performed at three testing sites representative of intended use settings with 140 lay persons. They had diverse educational and professional backgrounds and ranged in age from 19 to >50 years. Urine samples were prepared at the following concentrations; -100%, +/-75%, +/-50%, +/-25% of the cut-off by spiking fentanyl into drug free-pooled urine specimens. The concentrations of the samples were confirmed by LC/MS. Each sample was aliquoted into individual containers, blind-labeled and randomized. Each participant was provided with the package insert, 1 blind labeled sample and a device. The results are summarized below:

% of CutoffNumberofsamplesFentanyl Concentrationby LC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff200.23020100
-50% Cutoff200.53020100
-25% Cutoff200.7511995
+25% Cutoff201.20200100
+50% Cutoff201.46200100
+75% Cutoff201.77200100

Lay-users were also given surveys on the ease of understanding the package insert instructions. All lay users indicated that the device instructions can be easily followed. A Flesch-Kincaid reading analysis was performed on the package insert and the score revealed a reading grade level of less than 7.

    1. Clinical Studies
      Not applicable.
    1. Conclusion
      Based on the test principle and acceptable performance characteristics including precision, cut-off, interference, specificity, method comparison and Lay-user studies of the devices, it's concluded that the CLUNGENE Fentanyl Test is substantially equivalent to the predicate.

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).