K Number
K242732

Validate with FDA (Live)

Device Name
Tristel OPH
Date Cleared
2025-05-09

(241 days)

Product Code
Regulation Number
880.6886
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Tristel OPH is a high level disinfectant foam for reprocessing ophthalmic devices. It can be used to high level disinfect devices that may contact mucous membranes during use, such as diagnostic and laser lenses, tonometer prisms, pachymeters, retinal imaging probes, A-scan and B-scan probes.

Tristel OPH is a high level disinfectant when used in accordance with the Directions for Use, including a disinfection contact time of 2 minutes, at room temperature and a concentration at or above Minimum Recommended Concentration (MRC).

Tristel OPH Foam must be applied on the surface of an ophthalmic device using Tristel OPH Wipes.

Tristel OPH Wipes are intended for application of Tristel OPH Foam on the surface of ophthalmic devices and to remove residue of the foam after high level disinfection.

Each dose of Tristel OPH Foam and each Tristel OPH Wipe are single use.

The semi-critical ophthalmic devices processed by Tristel OPH must first be cleaned according to a validated cleaning protocol or standard and following the device manufacturers' instructions. Tristel OPH Foam bottle is designed in a way that ensures measured dose delivery with each application generating active ingredient above the MRC.

Minimum Recommended Concentration (MRC): ~90% v/v (~280ppm) at point of use. MRC of Tristel OPH may be verified using Tristel Test Strips.

Device Description

Tristel OPH is a high level disinfectant foam for reprocessing ophthalmic devices. It can be used to high level disinfect devices that may contact mucous membranes during use, such as diagnostic and laser lenses, tonometer prisms, pachymeters, retinal imaging probes, and A-scan and B-scan probes.
Tristel OPH Foam is packaged in a hand-held dispenser bottle. Tristel OPH Foam is made of two solutions, Tristel OPH Activator solution and Tristel OPH Base solution, which are mixed when dispensed to generate active ingredient chlorine dioxide. In-use concentration of chlorine dioxide (ClO2) is 0.032% (320ppm).
Tristel OPH Foam Activator and Base solutions are contained in a dual compartment bottle with a dispenser pump and are ready to use.
Tristel OPH Foam is supplied with Tristel OPH Wipes. Tristel OPH Wipes are dry wipes made of a non-woven and low-linting material. Tristel OPH Wipes are intended to apply Tristel OPH Foam on the surface of ophthalmic devices and to remove residue of the foam after high level disinfection once the contact time of 2 minutes has elapsed.
Each dose of Tristel OPH Foam and each Tristel OPH Wipe are single-use.
Minimum Recommended Concentration (MRC): ~90% v/v (~280ppm) at point of use. MRC of Tristel OPH may be verified using Tristel Test Strips.

AI/ML Overview

The provided FDA 510(k) clearance letter and its attachments describe the device Tristel OPH, a high-level disinfectant foam. However, the document focuses on demonstrating substantial equivalence to a predicate device (Tristel Duo ULT) through non-clinical performance testing. It meticulously compares the two devices based on their indications for use, technological characteristics, and performance data from various laboratory efficacy tests (bactericidal, fungicidal, virucidal, mycobactericidal, sporicidal, and additional clinically relevant efficacy) as well as simulated-use, in-use, compatibility, biocompatibility, residue, and inhalation studies.

Crucially, the document does not describe a study that involves analyzing a test set of data with ground truth, nor does it refer to acceptance criteria in the context of diagnostic performance (e.g., sensitivity, specificity, AUC) for a medical image analysis or AI/ML device.

The "acceptance criteria" referred to in the document are primarily the performance benchmarks met during chemical and microbiological efficacy testing, as well as safety and compatibility tests. The "study that proves the device meets the acceptance criteria" refers to a series of non-clinical, in-vitro, and simulated-use laboratory studies designed to demonstrate the disinfectant's potency, safety, and proper functioning.

Therefore, many of the requested elements in your prompt, which are typical for studies evaluating AI/ML-based diagnostic devices, cannot be extracted from this document. I will focus on the information that is available.


Here's an interpretation of the provided information, framed to address as many of your prompt's points as possible within the context of a disinfectant's clearance:

Acceptance Criteria and Device Performance for Tristel OPH (High-Level Disinfectant)

The "acceptance criteria" for Tristel OPH are not diagnostic performance metrics, but rather the successful demonstration of its high-level disinfection capabilities, safety, and functional equivalence to its predicate device under various laboratory and simulated-use conditions.

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria CategorySpecific Acceptance Criteria (Implied from Testing)Reported Device Performance (Tristel OPH)
Potency - BactericidalBactericidal against Pseudomonas aeruginosa, Staphylococcus aureus, Salmonella enterica in 2.0 minutes exposure at 20°C (AOAC 955.15, 964.02, 955.14).Bactericidal against Pseudomonas aeruginosa, Staphylococcus aureus, and Salmonella enterica in 2.0 minutes exposure at 20°C.
Potency - FungicidalFungicidal against Trichophyton interdigitale in 1.0 minute exposure at 20°C (AOAC 955.17).Fungicidal against Trichophyton interdigitale in 1.0 minute exposure at 20°C.
Potency - VirucidalVirucidal against Poliovirus type 1, Herpes simplex virus type 1, Adenovirus type 5, and Feline calicivirus in 30 seconds exposure at 20°C (ASTM E1053).Virucidal against Poliovirus type 1, Herpes simplex virus type 1, Adenovirus type 5, and Feline calicivirus in 30 seconds exposure at 20°C.
Potency - MycobactericidalMycobactericidal against Mycobacterium terrae in 1.0 minute exposure at 20°C (Ascenzi et al., 1987).Mycobactericidal against Mycobacterium terrae in 1.0 minute exposure at 20°C.
Potency - SporicidalSporicidal against Clostridium sporogenes and Bacillus subtilis in 12 hours exposure at 20°C (AOAC 966.04).Sporicidal against Clostridium sporogenes and Bacillus subtilis in 12 hours exposure at 20°C.
Potency - Additional Clinically Relevant EfficacyPassing AOAC and ASTM1053 against specified clinically relevant bacteria, fungi, and viruses, and inactivation of HPV types 16 & 18.Passed AOAC and ASTM1053 against Candida albicans, Aspergillus brasiliensis, Streptococcus agalacticae, Carbapenem-resistant Klebsiella pneumoniae, Extended Spectrum Beta-Lactamase (ESBL) Escherichia coli, Haemophilus influenza, Vancomycin-Resistant Enterococcus faecalis (VRE), Multi-Drug Resistant Streptococcus pneumoniae, Staphylococcus epidermidis, Human Coronovirus SARS-CoV-2, Human hepatitis B virus surrogate (duck hepatitis B virus), Human immunodeficiency virus type 1, Influenza A Virus. Inactivates Human papillomavirus (HPV) type 16 and type 18.
Simulated-Use>6 log reduction mycobactericidal efficacy on worst-case inoculated devices in <2 minutes exposure at 20°C at end of shelf-life, end of use-life, and MRC.Provides >6 log reduction mycobactericidal efficacy on worst-case inoculated devices in <2 minutes exposure at 20°C at end of shelf-life, end of use-life and MRC.
In-UseNo evidence of growth of wild type bacteria or fungi recovered from clinically used instruments following disinfection.No evidence of growth of wild type bacteria or fungi recovered from devices following disinfection on clinically used instruments.
Device CompatibilityNo corrosion or other visible damage to tested devices/materials; inclusion in compatible disinfectant lists.Tested and approved by several device manufacturers and included in the List of Compatible disinfectants. Does not produce any corrosion or other visible damage in the devices and materials tested.
Biocompatibility ToxicityNon-mutagenic, non-clastogenic, non-oral toxic, not a skin sensitizer, negligible irritant to skin, no cytotoxic effect (in use with wipe).Tristel OPH foam is non-mutagenic, non-clastogenic in maturing erythrocytes, and non-oral toxic. It is not a skin sensitizer. It is classified as a slight irritant to the skin and an eye irritant in direct contact. In use with the Tristel OPH Wipe, Tristel OPH foam is not considered to have a cytotoxic effect, is negligible irritating to the skin and has no sensitizing potential.
ResiduesPost-reprocessing residuals quantified and pose no additional risk to patient/user; non-irritating to dermal and eye tissues.Residue testing showed that Tristel OPH germicide poses no additional risk to the patient or user of the device. Any potential residues remaining post reprocessing are non-irritating to dermal and eye tissues.
Inhalation ExposureAverage 8-hour TWA readings do not exceed U.S. OSHA PEL of 0.1ppm for chlorine dioxide; average 15-minute STEL readings do not exceed California Division of Occupational Safety and Health PEL of 0.3ppm.In a study simulating 51 disinfection procedures, the average 8-hour TWA readings did not exceed the U.S. OSHA permissible exposure limit of 0.1ppm for chlorine dioxide, and the results of the average 15-minute STEL readings did not exceed the California Division of Occupational Safety and Health PEL of 0.3ppm.
Human Factors/UsabilityEvaluated for intended users, uses, and use environments.Tristel OPH has been evaluated for the intended users, uses, and use environments.

2. Sample sizes used for the test set and the data provenance

  • Test Set (for Efficacy/Safety/Compatibility): The document does not provide explicit sample sizes in terms of "number of devices" or "number of tests" for each specific study (e.g., number of inoculated items for simulated-use, number of materials for compatibility). However, it implies standardized testing (e.g., AOAC, ASTM, ISO standards) which would inherently include appropriate sample sizes for replicates and controls.
  • Data Provenance: The studies are non-clinical, laboratory-based tests. The country of origin for the data is not explicitly stated but is assumed to be from the manufacturer's labs or certified contract research organizations. These are prospective experiments conducted to demonstrate the device's performance against pre-defined standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. For a disinfectant, "ground truth" is established through standardized microbiological assays (e.g., presence or absence of viable microorganisms after exposure to the disinfectant) or chemical analyses, not through expert consensus on diagnostic images. The performance is objectively measured based on established scientific protocols.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. Adjudication methods like 2+1 or 3+1 are used for establishing ground truth in human observer studies, particularly in medical image interpretation. The studies for Tristel OPH are laboratory-based and involve objective measurements.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. MRMC studies are relevant for evaluating the impact of AI on human diagnostic performance. Tristel OPH is a chemical disinfectant, not an AI/ML diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This concept is also not applicable in the context of a disinfectant. The device (disinfectant) inherently performs its function (killing microbes) in a "standalone" chemical process, and its efficacy is tested in laboratory settings without human intervention in the disinfection process itself, though human compliance with the instructions for use is necessary.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the performance of this disinfectant is based on:

  • Microbiological Culture Results: Absence or significant reduction (e.g., >6 log reduction) of specific viable microorganisms after exposure to the disinfectant.
  • Chemical Analysis: Measurement of residual chemicals, concentration of active ingredients.
  • Physical Observation: Assessment of material integrity (corrosion, damage).
  • Environmental Monitoring: Measurement of airborne chemical concentrations.
    These are all objective, laboratory-derived measurements based on established scientific methodologies and standards.

8. The sample size for the training set

Not applicable. This device is a chemical disinfectant, not an AI/ML algorithm that requires a training set.

9. How the ground truth for the training set was established

Not applicable (as above).

FDA 510(k) Clearance Letter - Tristel OPH

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

May 09, 2025

Tristel Solutions Ltd.
Véronique Li
Unit 1b, Lynx Business Park, Fordham Road
Snailwell
Cambridgeshire, CB8 7NY
United Kingdom

Re: K242732
Trade/Device Name: Tristel OPH
Regulation Number: 21 CFR 880.6886
Regulation Name: Foam Or Gel Chemical Sterilant/High Level Disinfectant
Regulatory Class: Class II
Product Code: QWS
Dated: September 10, 2024
Received: April 8, 2025

Dear Véronique Li:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K242732 - Véronique Li
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K242732 - Véronique Li
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Katharine Segars -S

Katharine Segars, Ph.D.
Assistant Director
DHT4C: Division of Infection Control
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K242732

Device Name
Tristel OPH

Indications for Use (Describe)

Tristel OPH is a high level disinfectant foam for reprocessing ophthalmic devices. It can be used to high level disinfect devices that may contact mucous membranes during use, such as diagnostic and laser lenses, tonometer prisms, pachymeters, retinal imaging probes, A-scan and B-scan probes.

Tristel OPH is a high level disinfectant when used in accordance with the Directions for Use, including a disinfection contact time of 2 minutes, at room temperature and a concentration at or above Minimum Recommended Concentration (MRC).

Tristel OPH Foam must be applied on the surface of an ophthalmic device using Tristel OPH Wipes.

Tristel OPH Wipes are intended for application of Tristel OPH Foam on the surface of ophthalmic devices and to remove residue of the foam after high level disinfection.

Each dose of Tristel OPH Foam and each Tristel OPH Wipe are single use.

The semi-critical ophthalmic devices processed by Tristel OPH must first be cleaned according to a validated cleaning protocol or standard and following the device manufacturers' instructions. Tristel OPH Foam bottle is designed in a way that ensures measured dose delivery with each application generating active ingredient above the MRC.

Minimum Recommended Concentration (MRC): ~90% v/v (~280ppm) at point of use. MRC of Tristel OPH may be verified using Tristel Test Strips.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.

Page 5

510(k) Summary K242732

I. SUBMITTER

Tristel Solutions Ltd.
Tristel Solutions Ltd., Unit 1B, Lynx Business Park,
Fordham Road, Snailwell, Cambridgeshire,
CB8 7NY, United Kingdom Phone:
+44 (0)1638 721500

Contact Person: Julija Shabanova
Tristel Solutions Ltd., Unit 1B, Lynx Business Park,
Fordham Road, Snailwell, Cambridgeshire,
CB8 7NY, United Kingdom
Ph. +44 (0)1638 721500
Em. julijashabanova@tristel.com

Date Prepared: May 9, 2025

II. SUBJECT DEVICE

Brand Name: Tristel OPH
Common Name: High Level Disinfectant
Classification Name: Foam or gel chemical sterilant/high level disinfectant
Regulation Number: 21 CFR 880.6886
Product Code: QWS
Regulatory Class: II

III. PREDICATE DEVICE

Predicate Device: Tristel Duo ULT (DEN220041)

IV. DEVICE DESCRIPTION

Tristel OPH is a high level disinfectant foam for reprocessing ophthalmic devices. It can be used to high level disinfect devices that may contact mucous membranes during use, such as diagnostic and laser lenses, tonometer prisms, pachymeters, retinal imaging probes, and A-scan and B-scan probes.

Page 6

Tristel OPH Foam is packaged in a hand-held dispenser bottle (Figure 1). Tristel OPH Foam is made of two solutions, Tristel OPH Activator solution and Tristel OPH Base solution, which are mixed when dispensed to generate active ingredient chlorine dioxide. In-use concentration of chlorine dioxide (ClO2) is 0.032% (320ppm).

Tristel OPH Foam Activator and Base solutions are contained in a dual compartment bottle with a dispenser pump and are ready to use (Figure 2).

Figure 1. Pictures of the packaged Tristel OPH
Figure 2. Schematic of Tristel OPH Foam bottle

Tristel OPH Foam is supplied with Tristel OPH Wipes. Tristel OPH Wipes are dry wipes made of a non-woven and low-linting material. Tristel OPH Wipes are intended to apply Tristel OPH Foam on the surface of ophthalmic devices and to remove residue of the foam after high level disinfection once the contact time of 2 minutes has elapsed.

Each dose of Tristel OPH Foam and each Tristel OPH Wipe are single-use.

Minimum Recommended Concentration (MRC): ~90% v/v (~280ppm) at point of use. MRC of Tristel OPH may be verified using Tristel Test Strips.

V. INDICATIONS FOR USE

Tristel OPH is a high level disinfectant foam for reprocessing ophthalmic devices. It can be used to high level disinfect devices that may contact mucous membranes during use, such as diagnostic and laser lenses, tonometer prisms, pachymeters, retinal imaging probes, A-scan and B-scan biometry probes.

Tristel OPH is a high level disinfectant when used in accordance with the Directions for Use, including a disinfection contact time of 2 minutes, at room temperature and a concentration at or above the Minimum Recommended Concentration (MRC).

Tristel OPH Foam must be applied on the surface of an ophthalmic device using Tristel OPH Wipes. Tristel OPH Wipes are intended for the application of the Tristel OPH Foam on the surface of ophthalmic devices and to remove residue of the foam after high level disinfection.

Each dose of Tristel OPH Foam and each Tristel OPH Wipe are single use.

Page 7

The semi-critical ophthalmic devices processed by Tristel OPH must first be cleaned according to a validated cleaning protocol or standard and following the device manufacturers' instructions. Tristel OPH Foam bottle is designed in a way that ensures measured dose delivery with each application generating active ingredient above the MRC.

Minimum Recommended Concentration (MRC): ~90% v/v (280ppm) at point of use. MRC of Tristel OPH may be verified using Tristel Test Strips.

Both the subject and the predicate devices have the same intended use for the high level disinfection of semi-critical devices. The Indications for Use differ only in the types of semi-critical devices intended to be disinfected. The differences do not alter the intended.

Table 1 – Predicate Device Tristel Duo ULT and Tristel OPH Indications for Use Comparison Table

Tristel Duo ULT(DEN220041)Tristel OPH (K242732)
Indications for Use: Tristel Duo ULT is a high level disinfectant foam for reprocessing ultrasound probes. It can be used to high level disinfect endocavity transvaginal and transrectal probes, and skin surface transducers that may contact non-intact skin during use.Tristel Duo ULT is a high level disinfectant when used in accordance with the Directions for Use, including a disinfection contact time of 2 minutes, at room temperature and a concentration at or above Minimum Recommended Concentration (MRC).Tristel Duo ULT Foam must be applied on the surface of an ultrasound probe using Tristel Duo Wipes.Tristel Duo Wipes are intended for application of Tristel Duo ULT Foam on the surface of ultrasound probes and to remove residue of the foam after high level disinfection.Each dose of Tristel Duo ULT Foam and each Tristel Duo Wipe are single use.The semi-critical ultrasound probes processed by Tristel Duo ULT must first be cleaned according to a validated cleaning protocol or standard and following theTristel OPH is a high level disinfectant foam for reprocessing ophthalmic devices. It can be used to high level disinfect devices that may contact mucous membranes during use, such as diagnostic and laser lenses, tonometer prisms, pachymeters, retinal imaging probes, A-scan and B-scan probes.Tristel OPH is a high level disinfectant when used in accordance with the Directions for Use, including a disinfection contact time of 2 minutes, at room temperature and a concentration at or above Minimum Recommended Concentration (MRC).Tristel OPH Foam must be applied on the surface of an ophthalmic device using Tristel OPH Wipes.Tristel OPH Wipes are intended for application of Tristel OPH Foam on the surface of ophthalmic devices and to remove residue of the foam after high level disinfection.Each dose of Tristel OPH Foam and each Tristel OPH Wipe are single use.The semi-critical ophthalmic devices processed by Tristel OPH must first be cleaned according to a validated cleaning protocol or standard and following the

Page 8

Tristel Duo ULT(DEN220041)Tristel OPH (K242732)
device manufacturers' instructions. Tristel Duo ULT Foam bottle is designed in a way that ensures measured dose delivery with each application generating active ingredient above the MRC.Minimum Recommended Concentration (MRC): ~90% v/v (~280ppm) at point of use. MRC of Tristel Duo ULT may be verified using Tristel Test Strips.device manufacturers' instructions. Tristel OPH Foam bottle is designed in a way that ensures measured dose delivery with each application generating active ingredient above the MRC.Minimum Recommended Concentration (MRC): ~90% v/v (~280ppm) at point of use. MRC of Tristel OPH may be verified using Tristel Test Strips.

VI. TECHNOLOGICAL CHARACTERISTIC COMPARISON

A side-by-side comparison of the technological characteristics of the predicate device and Tristel OPH is shown in Table 2.

Table 2 – Predicate Device Tristel Duo ULT and Tristel OPH Technological Characteristics Comparison Table

Tristel Duo ULT (DEN220041)Tristel OPH (K242732)Comparison
General
Active Chemistry:Chlorine dioxideChlorine dioxideSame
Active Concentration:~320ppm~320ppmSame
Purpose:To achieve high level disinfection by destroying viable forms of microbial life, when used according to labelingTo achieve high level disinfection by destroying viable forms of microbial life, when used according to labelingSame
Labeling:High Level Disinfectant FoamHigh Level Disinfectant FoamSame
No. of Foam Doses Required for Each Use:42Smaller ophthalmic devices require less volume of foam
Prescription Device:YesYesSame
Human Factors:Dispensed ready-to-use foam and applied by a wipeDispensed ready-to-use foam and applied by a wipeSame
Design, Construction, Components:Aqueous solutions contained in two separate reservoirs within a single dispenser apparatus and mixed together when dosed onto a Tristel WipeAqueous solutions contained in two separate reservoirs within a single dispenser apparatus and mixed together when dosed onto a Tristel WipeSame
Reuse Characteristics:Each dose of Tristel Duo ULT Foam and each Tristel Duo Wipe are single use.Each dose of Tristel OPH Foam and each Tristel OPH Wipe are single use.Same
Chemical Indicator:Tristel Test StripsTristel Test StripsSame

Page 9

Tristel Duo ULT (DEN220041)Tristel OPH (K242732)Comparison
Contact Time for HLD:2 minutes at room temperature2 minutes at room temperatureSame
Mode of Action:Chemical oxidation of biomolecules, including the cell membrane and nucleic acids.Chemical oxidation of biomolecules, including the cell membrane and nucleic acids.Same
Packaging:Tristel Duo ULT Foam Activator and Base solutions are contained in a dual compartment bottle with a dispenser pump and are ready to use.Tristel OPH Foam Activator and Base solutions are contained in a dual compartment bottle with a dispenser pump and are ready to use.Same
Stability
Shelf Life:2 years unopened Shelf life6 month opened Shelf life2 years unopened Shelf life6 month opened Shelf lifeSame
Transport Stability:Conditions encountered for global road, sea, and air freight have no effect on product performance.Conditions encountered for global road, sea, and air freight have no effect on product performance.Same
Wipes
Physical Characteristics:Dry, non-woven, and low-lintingDry, non-woven, and low-lintingSame
No. of Wipes Supplied:Pack variant specific (e.g. 200)Pack variant specific (e.g.310)Pack variant specific
Linting Tests:<4 log coefficient of linting<4 log coefficient of lintingSame
Abrasion Tests:Withstand a minimum of 200 revolutionsWithstand a minimum of 200 revolutionsSame
Compatibility with Tristel Foam:CompatibleCompatibleSame

The subject device and its predicate device have an identical intended use and minimal difference in technological characteristics.

VII. PERFORMANCE DATA

Summary of Non-Clinical Testing

In addition to the non-clinical performance testing carried out on Tristel Duo ULT which are applicable to Tristel OPH, additional non-clinical performance testing for the intended ophthalmic use pattern were performed using the same test methods used to verify the predicate device. The testing demonstrated that the subject device met the acceptance criteria described in the standard/methodology.

Page 10

Table 3 - Predicate Device Tristel Duo ULT and Tristel OPH Non-Clinical Testing Comparison Table

Tristel Duo ULTTristel OPHComparison
Potency
Bactericidal Efficacy: AOAC 955.15, 964.02 and 955.14Bactericidal against Pseudomonas aeruginosa, Staphylococcus aureus and Salmonella enterica in 2.0 minutes exposure at 20°C.Bactericidal against Pseudomonas aeruginosa, Staphylococcus aureus and Salmonella enterica in 2.0 minutes exposure at 20°C.Same
Fungicidal Efficacy: AOAC 955.17Fungicidal against Trichophyton interdigitale in 1.0 minute exposure at 20°C.Fungicidal against Trichophyton interdigitale in 1.0 minute exposure at 20°C.Same
Virucidal Efficacy: ASTM E1053Virucidal against Poliovirus type 1, Herpes simplex virus type 1, Adenovirus type 5 and Feline calicivirus in 30 seconds exposure at 20°C.Virucidal against Poliovirus type 1, Herpes simplex virus type 1, Adenovirus type 5 and Feline calicivirus in 30 seconds exposure at 20°C.Same
Mycobactericidal Efficacy: Ascenzi et al., 1987Mycobactericidal against Mycobacterium terrae in 1.0 minute exposure at 20°C.Mycobactericidal against Mycobacterium terrae in 1.0 minute exposure at 20°C.Same
Sporicidal Efficacy: AOAC 966.04Sporicidal against Clostridium sporogenes and Bacillus subtilis in 12 hours exposure at 20°C.Sporicidal against Clostridium sporogenes and Bacillus subtilis in 12 hours exposure at 20°C.Same
Additional Clinically Relevant Efficacy:Passes AOAC and ASTM1053 against clinically relevant bacteria and viruses.Candida albicans, Human hepatitis B virus surrogate (duck hepatitis B virus), Human immunodeficiency virus type 1, Chlamydia trachomatis, Neisseria gonorrhoeae, Streptococcus agalacticae, Carbapenem-resistant Klebsiella pneumoniae, Extended Spectrum Beta-Lactamase (ESBL) Escherichia coli.Tristel Duo ULT inactivates Human papillomavirus(HPV) type 16 and type 18.Passes AOAC and ASTM1053 against clinically relevant bacteria, fungi and viruses.Candida albicans, Aspergillus brasiliensis, Streptococcus agalacticae, Carbapenem-resistant Klebsiella pneumoniae, Extended Spectrum Beta-Lactamase (ESBL) Escherichia coli, Haemophilus influenza, Vancomycin-Resistant Enterococcus faecalis (VRE), Multi-Drug Resistant Streptococcus pneumoniae, Staphylococcus epidermidis, Human Coronovirus SARS-CoV-2, Human hepatitis B virus surrogate (duck hepatitis B virus), Human immunodeficiency virus type 1, Influenza A Virus.Tristel OPH inactivates Human papillomavirus(HPV) type 16 and type 18.Different additional organisms tested depending on clinical area of use

Page 11

Tristel Duo ULTTristel OPHComparison
Simulated-Use Summary:Provides >6 log reduction mycobactericidal efficacy on worst-case inoculated devices in <2 minutes exposure at 20°C at end of self-life, and of use-life and MRC.Provides >6 log reduction mycobactericidal efficacy on worst-case inoculated devices in <2 minutes exposure at 20°C at end of shelf-life, end of use-life and MRC.Different devices tested depending on clinical area of use
In-Use Summary:No evidence of growth of wild type bacteria or fungi recovered from devices following disinfection on clinically used instrumentsNo evidence of growth of wild type bacteria or fungi recovered from devices following disinfection on clinically used instrumentsDifferent devices tested depending on clinical area of use
Compatibility
Device Compatibility:Tristel Duo ULT was tested and approved by several device manufacturers and included in the List of Compatible disinfectants.Tristel Duo ULT does not produce any corrosion or other visible damage in the devices and materials tested.Tristel OPH was tested and approved by several device manufacturers and included in the List of Compatible disinfectants.Tristel OPH does not produce any corrosion or other visible damage in the devices and materials tested.Different devices/materials tested depending on clinical area of use
Biocompatibility Toxicity
Summary:Testing was conducted following ISO 10993-5 and ISO 10993-10.Tristel Duo ULT foam is non-mutagenic, non-clastogenic in maturing erythrocytes, and non-oral toxic.It is not a skin sensitizer. It is classified as a slight irritant to the skin and an eye irritant in direct contact.In use with the Tristel Duo Wipe, Tristel Duo ULT foam is not considered to have a cytotoxic effect, is negligible irritating to the skin and has no sensitizing potential.Testing was conducted following ISO 10993-5 and ISO 10993-10.Tristel OPH foam is non-mutagenic, non-clastogenic in maturing erythrocytes, and non-oral toxic.It is not a skin sensitizer. It is classified as a slight irritant to the skin and an eye irritant in direct contact.In use with the Tristel OPH Wipe, Tristel OPH foam is not considered to have a cytotoxic effect, is negligible irritating to the skin and has no sensitizing potential.Same

Page 12

Tristel Duo ULTTristel OPHComparison
Residues
Residue Extractable and Leachable:Residue testing was performed to quantify any residuals remaining/extracted/leached after disinfection with Tristel Duo ULT. The treated device was extracted, and the eluates analyzed for residual components. The results show that Tristel Duo ULT germicide poses no additional risk to the patient or user of the device.Residue testing was performed to quantify any residuals remaining/extracted/leached after disinfection with Tristel OPH. The treated device/materials were extracted, and the eluates analyzed for residual components. The results show that Tristel OPH germicide poses no additional risk to the patient or user of the device.Different devices tested depending on clinical area of use
Biocompatibility of Residues:Any potential residues remaining post reprocessing are non-irritating to dermal, vaginal and rectal tissues.Any potential residues remaining post reprocessing are non-irritating to dermal and eye tissues.Different testing depending on clinical area of use
Inhalation
Inhalation:Tristel Duo ULT was evaluated for inhalation exposure during use. In a study simulating 40 disinfection procedures, the average 8-hour TWA readings did not exceed the U.S. OSHA permissible exposure limit of 0.1ppm for chlorine dioxide, and the results of the average 15-minute STEL readings did not exceed the California Division of Occupational Safety and Health PEL of 0.3ppm.Tristel OPH was evaluated for inhalation exposure during use. In a study simulating 51 disinfection procedures, the average 8-hour TWA readings did not exceed the U.S. OSHA permissible exposure limit of 0.1ppm for chlorine dioxide, and the results of the average 15-minute STEL readings did not exceed the California Division of Occupational Safety and Health PEL of 0.3ppm.Different testing depending on clinical area of use
Human Factors and Usability Engineering
Summary:Tristel Duo ULT has been evaluated for the intended users, uses, and use environments.Tristel OPH has been evaluated for the intended users, uses, and use environments.Same

VIII. CONCLUSION

The information on technological characteristics and data from non-clinical performance studies establish that the subject device, Tristel OPH, is as safe, as effective and performs at least as safely and effectively as the legally marketed predicate device.