AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Hair Removal Device is an over-the-counter device intended for removal of unwanted body hair and/or facial hair.

Device Description

Hair Removal Device (Model: R3C16-P, R3C16-G; R3C16-P Pro, R3C16-W Pro, R3C16-G Pro, R3505-W, R3505-B, R3505-W Pro, R3505-B Pro), is an over-the-counter, home-use and personal device for hair reduction based on Intense Pulsed Light (IPL). It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. Hair Removal Device includes two series (R3C16 series) with a total of 10 models. The two series products adopt identical intended use, similar performance and operation, consisting of IPL host and power adapter, which the host is mainly composed of lamp tube, filter, display screen, buttons, thermoelectric cooler, fan, and DC socket. The device is only powered by the external power adapter. The device is fitted with a skin sensor to detect appropriate skin contact, if the treatment window of the device is not in full contact with the skin, the device cannot emit light pulses, and the IPL emission is by manual finger switch or auto light emission. All models of the Hair Removal Device have a cooling function for a better hair removal experience, but the main difference is that the model with "Pro" in the model name is equipped with sapphire while the model without "Pro" uses metal treatment window surface.

AI/ML Overview

This 510(k) summary provides limited information regarding specific acceptance criteria and detailed study results for the Hair Removal Device. However, based on the provided text, here's an analysis of what can be extracted.

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state quantitative acceptance criteria in a table format with corresponding reported device performance values. Instead, it relies on a comparison to predicate devices and adherence to various safety and performance standards.

We can infer the intent of some acceptance criteria by looking at the "Comparison Elements" table on pages 7-8 and the "Non-Clinical Testing" section on page 9. The underlying acceptance is that the subject device's characteristics and performance are "similar" or "same" to the predicate devices and meet the specified international standards for medical electrical equipment and photobiological safety.

Acceptance Criteria (Inferred)Reported Device Performance (from Comparison Table)
Intended Use"Hair Removal Device is an over-the-counter device intended for removal of unwanted body hair and/or facial hair." (Same as Predicate 1, similar to others which also mention permanent reduction)
Prescription or OTCOTC (Same as all listed predicates/references)
Source EnergyAn external power supply (Same as all listed predicates/references)
Power Supply100~240V, 50/60Hz (Same as Predicate 1 and Reference 1, similar to Predicate 2)
Dimensions (mm)R3C16 series: 1857040mm; R3505 series: 19811751mm (Different from Predicate 1 and Reference 1, but this is a size difference, not necessarily a performance failure)
SterilizationNot required (Same as all listed predicates/references)
Light SourceIntense Pulsed Light (Same as all listed predicates/references)
Energy MediumXenon Arc Flashlamp (Same as all listed predicates/references)
Wavelength Range530~1200nm (Same as Predicate 1, similar to others in the 510-1200nm range)
Energy DensityR3C16: 1.56J/cm²4.69J/cm²; R3C16 Pro: 1.39J/cm²4.17J/cm²; R3505: 3.57J/cm²7.5J/cm²; R3505 Pro: 3.03J/cm²6.36J/cm² (Similar to predicate devices which range from 1.8 J/cm² to 10 J/cm²)
Spot SizeR3C16: 3.2±0.25cm²; R3C16 Pro: 3.6±0.25cm²; R3505: 2.8±0.25cm²; R3505 Pro: 3.3±0.25cm² (Similar to predicate devices which range from 3cm² to 4.2cm²)
Pulse DurationR3C16 series: 6-11ms; R3505 series: 11~12ms (Similar to predicate devices which range from 2ms to 25ms)
Pulsing ControlFinger switch (Same as all listed predicates/references)
Delivery DeviceDirect illumination to tissue (Same as all listed predicates/references)
Output Intensity Level9 levels (Same as Predicate 1 and Reference 1)
Software/Firmware/Microprocessor Control?Yes (Same as Predicate 1, 2, 3 and Reference 1)
Biological Evaluation (Cytotoxicity)Compliant with ISO 10993-5: 2009
Biological Evaluation (Skin Sensitization)Compliant with ISO 10993-10: 2021
Biological Evaluation (Irritation)Compliant with ISO 10993-23: 2021
Electrical SafetyCompliant with IEC 60601-1: 2020
Electromagnetic Compatibility (EMC)Compliant with IEC 60601-1-2: 2020
Home Healthcare Environment SafetyCompliant with IEC 60601-1-11: 2020
Home Light Therapy Equipment SafetyCompliant with IEC 60601-2-83: 2022
Photobiological SafetyCompliant with IEC 62471: 2006
UsabilityCompliant with IEC 60601-1-6: 2020
Overall Safety and Effectiveness"as safe, as efforms as well as the legally marketed predicate devices."

2. Sample Size Used for the Test Set and Data Provenance

The document explicitly states "VIII. Clinical Testing: Not applicable."
This means there was no clinical test set for evaluating the device's hair removal efficacy and safety on human subjects as part of this 510(k) submission. Therefore, details like sample size, country of origin, or retrospective/prospective nature are not provided for efficacy testing.

The "non-clinical testing" mentioned primarily refers to bench testing and adherence to international safety and performance standards for medical devices and photobiological safety. For these tests, specific sample sizes (e.g., number of devices tested for electrical safety, materials tested for biocompatibility) are typically defined by the respective standards, but these details are not provided in this summary.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Since no clinical testing was applicable or performed for this 510(k) submission, there was no clinical test set for which ground truth needed to be established by experts for efficacy. The evaluation was based on non-clinical characteristic comparisons and technical standard compliance.

4. Adjudication Method for the Test Set

Not applicable, as no clinical test set requiring expert adjudication for ground truth was performed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Done

No. The document explicitly states "VIII. Clinical Testing: Not applicable." An MRMC study would be a type of clinical comparative effectiveness study, which was not done.

6. Standalone Performance Done

The performance evaluation in this 510(k) is primarily a standalone assessment of the device's technical specifications and compliance with safety standards. The comparison to predicates is to demonstrate "substantial equivalence" based on similar characteristics and the device meeting the same technical and safety requirements. The non-clinical tests (e.g., biocompatibility, electrical safety, photobiological safety) are assessments of the device itself (standalone).

7. Type of Ground Truth Used

Given the "Clinical Testing: Not applicable" statement, there was no clinical "ground truth" related to hair removal efficacy established or used in this submission. The "ground truth" for the non-clinical tests would be the requirements and limits defined by the referenced international standards (e.g., IEC, ISO standards). For example, the ground truth for cytotoxicity testing is whether the material shows cytotoxic effects according to the criteria in ISO 10993-5.

8. The Sample Size for the Training Set

Not applicable. The device is a hardware product for hair removal, not an AI/ML algorithm that requires a training set in the typical sense. The "software/firmware/microprocessor control" mentioned refers to embedded control systems for device operation (e.g., intensity levels, safety features), not for learning or diagnostic purposes that would require a large training dataset.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no mention of an AI/ML algorithm training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 6, 2024

Shenzhen Jianchao Intelligent Technology Co., Ltd. % Riley Chen RA Engineer Feiying Drug & Medical Consulting Technical Service Group Rm 2401 Zhenye International Business Center, No. 3101-90 Qianhai Road Shenzhen, Guangdong 518052 China

Re: K242710

Trade/Device Name: Hair Removal Device (R3C16-P, R3C16-G, R3C16-G, R3C16-P Pro, R3C16-W Pro, R3C16-G Pro, R3505-W, R3505-B, R3505-W Pro, R3505-B Pro) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: September 5, 2024 Received: September 9, 2024

Dear Riley Chen:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

{2}------------------------------------------------

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Yan Fu -S
Digitally signed by Yan Fu
Date: 2024.12.06 13:50:30

-05'00' for Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K242710

Device Name

Name of Device: Hair Removal Device

Model(s): R3C16-P, R3C16-W, R3C16-P Pro, R3C16-W Pro, R3C16-G Pro, R3505-W. R3505-B, R3505-W Pro, R3505-B Pro

Indications for Use (Describe)

Hair Removal Device is an over-the-counter device intended for removal of unwanted body hair and/or facial hair.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary: K242710

"510(k) Summary" as required by 21 CFR Part 807.92.

I. Submitter

Shenzhen Jianchao Intelligent Technology Co., Ltd. Rm301, Bldg.25, No.68 Hexiu West Road, Fuhai St., Baoan, Shenzhen, CN Post code: 518107 Fax: +86-0755-86961489

Fred Li Title: Director Tel.: +86 18675507170 Email: fred@vellcolife.com

Date of preparation: 2024-09-06

II. Device

Name of Device: Hair Removal Device

Model(s): R3C16-P, R3C16-W, R3C16-G, R3C16-P Pro, R3C16-W Pro, R3C16-G Pro, R3505-W, R3505-B, R3505-W Pro, R3505-B Pro Common or Usual Name: Light Based Over-The-Counter Hair Removal Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: OHT Regulation Number: 21 CFR 878.4810

III. Predicate Device & Reference Device

Predicate device:

Predicate DeviceManufacturer510(k) NumberProduct code
HairRemoval Device(R2815-G Pro, R2815-G)Shenzhen Jianchao IntelligentTechnology Co., Ltd.K232575OHT
Ipulse Smoothskin Gold HairRemoval DeviceCyden LimitedK160968OHT, GEX
IPL Hair Removal DeviceSHENZHEN BORRIATECHNOLOGY CO., LTD.K232939OHT

Reference device:

Reference DeviceManufacturer510(k) NumberProduct code
IPL Hair Removal Device JoyVersion (CB-027)Shen Zhen CosBeauty Co.,LtdK173813OHT
iPulse Hair Removal SystemCyDen, Ltd.K130315OHT

{5}------------------------------------------------

IV. Device Description

Hair Removal Device (Model: R3C16-P, R3C16-G; R3C16-P Pro, R3C16-W Pro, R3C16-G Pro, R3505-W, R3505-B, R3505-W Pro, R3505-B Pro), is an over-the-counter, home-use and personal device for hair reduction based on Intense Pulsed Light (IPL). It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain.

Hair Removal Device includes two series (R3C16 series) with a total of 10 models. The two series products adopt identical intended use, similar performance and operation, consisting of IPL host and power adapter, which the host is mainly composed of lamp tube, filter, display screen, buttons, thermoelectric cooler, fan, and DC socket.

The device is only powered by the external power adapter. The device is fitted with a skin sensor to detect appropriate skin contact, if the treatment window of the device is not in full contact with the skin, the device cannot emit light pulses, and the IPL emission is by manual finger switch or auto light emission. All models of the Hair Removal Device have a cooling function for a better hair removal experience, but the main difference is that the model with "Pro" in the model name is equipped with sapphire while the model without "Pro" uses metal treatment window surface. The difference of design between models are listed below:

Model (s)Enclosure colorCooling functionProduct drawing
R3C16-PPurpleMetal coolingImage: Product drawing 1
R3C16-WWhiteMetal coolingImage: Product drawing 1
R3C16-GGoldenMetal coolingImage: Product drawing 1
R3C16-P ProPurpleSapphire coolingImage: Product drawing 1
R3C16-W ProWhiteSapphire coolingImage: Product drawing 1
R3C16-G ProGoldenSapphire coolingImage: Product drawing 1
R3505-WWhiteMetal coolingImage: Product drawing 2
R3505-BBlackMetal coolingImage: Product drawing 2
R3505-W ProWhiteSapphire coolingImage: Product drawing 2
R3505-B ProBlackSapphire coolingImage: Product drawing 2

V. Indications for Use

Hair Removal Device is an over-the-counter device intended for removal of unwanted body hair and/or facial hair.

Comparison of Indications for Use: The indications for use of the subject device is comparable to the indications for use of the predicate device.

VI. Comparison of Technological Characteristics With the Predicate Device

The Hair Removal Device has the same indications for use, mode of action and similar operational characteristics as the predicate devices. Any minor differences between the subject device and the listed predicate devices do no raise any issues of safety or efficacy. Performance data supports that the device

{6}------------------------------------------------

is safe and as effective as the predicate devices for its intended use. Therefore, the Hair Removal Device may be found substantially equivalent to its predicate devices.

Hair Removal Device is compared with the following predicate devices and reference devices in terms of intended use, design, specifications and performance:

{7}------------------------------------------------

ComparisonElementsSubject DevicePredicate Device 1Predicate Device 2Predicate device 3Reference Device 1Reference Device 2Remark
510(k)NumberK242710K232575K160968K232939K173813K130315/
Trade nameHair Removal DeviceHair Removal DeviceiPulse SmoothSkinGoldHair Removal SystemIPL Hair Removal DeviceIPL Hair Removal DeviceJoy VersioniPulse Hair RemovalSystem/
Model (s)R3C16-P, R3C16-W, R3C16-G, R3C16-P Pro, R3C16-WPro, R3C16-G ProR3505-W, R3505-B, R3505-WPro, R3505-B ProR2815-G Pro, R2815-G/BR223, BR225CB-027//
Indication foruse/IntendeduseHair Removal Device is anover-the-counter deviceintended for removal ofunwanted body hair and/orfacial hair.Hair Removal Device is anover-the-counter deviceintended for removal ofunwanted body hair and/orfacial hair.The iPulse SmoothSkinGold Hair Removal Systemis indicated for the removalof unwanted hair. The iPulseSmoothskin Gold is alsoindicated for the permanentreduction in hair regrowth,defined as the long-term,stable reduction in thenumber of hairs regrowingwhen measured at 6, 9 and12 months after thecompletion of a treatmentregime.The IPL Hair RemovalDevice is indicated for theremoval for unwantedhair. The IPL HairRemoval Device is alsoindicated for thepermanent reduction inhair regrowth, defined asthe long-term, stablereduction in the number ofhairs regrowing whenmeasured at 6, 9 and 12months after thecompletion of thetreatment regime.The IPL Hair RemovalDevice Joy Version isindicated for the removalof unwanted hair. Thedevice is also indicated forthe permanent reduction inhair regrowth, defined asthe long-term, stablereduction in the number ofhairs regrowing whenmeasured at 6, 9 and 12months after thecompletion of a treatmentregime. The device is usedfor adults with Fitzpatrickskin types I - IV.The iPulse HairRemoval System isan Over-the-counterdevice intended forthe removal ofunwanted hair.Same
Prescription orOTCOTCOTCOTCOTCOTCOTCSame
Device design
Source energyAn external power supplyAn external power supplyExternal Power supplyExternal Power supplySupplied by externaladapterSupplied by externalpower converterSame
Power supply100~240V, 50/60Hz100~240V, 50/60Hz110V or 230V, 50/60HzNot public100~240V,50/60HzNot publicSame
Dimension(mm)R3C16 series: 1857040mmR3505 series: 19811751mm19411745mmNot publicNot public12678200mmNot publicDifferent
SterilizationNot requiredNot requiredNot requiredNot requiredNot requiredNot publicSame
Output specification
ComparisonElementsSubject DevicePredicate Device 1Predicate Device 2Predicate device 3Reference Device 1Reference Device 2Remark
Light sourceIntense Pulsed LightIntense Pulsed LightIntense Pulsed LightIntense Pulsed LightIntense Pulsed LightIntense Pulsed LightSame
EnergymediumXenon Arc FlashlampXenon Arc FlashlampXenon Arc FlashlampXenon Arc FlashlampXenon Arc FlashlampXenonArcFlashlampSame
Wavelengthrange530~1200nm530-1200nm510-1100nm510〜1100nm510-1200nm530-1100nmSame
Energy densityR3C16-P, R3C16-W, R3C16-G: 1.56J/cm²4.69J/cm² ;R3C16-P Pro, R3C16-W Pro,R3C16-GPro:1.39J/cm²4.17J/cm²R3505-W,R3505-B:3.57J/cm²7.5J/cm²;R3505-W Pro, R3505-B Pro:3.03J/cm²6.36J/cm²Max. 5.2J/cm²(rangein1.2J/cm²~5.2J/cm²)3-6J/cm²Max.6.0J/cm²1.8~5.1J/cm²7-10J/cm²Similar
Spot sizeR3C16-P, R3C16-W, R3C16-G: 3.2±0.25cm²R3C16-P Pro, R3C16-W Pro,R3C16-G Pro: 3.6±0.25cm²R3505-W,R3505-B:2.8±0.25cm²;R3505-W Pro, R3505-B Pro:3.3±0.25cm²3.4±0.25 cm²3cm² (3cm by 1cm)3.36cm²Body: 4.2cm²Bikini and face: 2.0cm²3cm² (1.3cm by 2.4cm)Similar
Pulse durationR3C16 series: 6-11msR3505 series: 11~12ms8.5±2.5ms(range in 6ms~11ms)2ms to 10ms8~12ms9.2~11.2msVariable -Single pulse: 25msDouble Pulse: 20mson, 60ms off.Similar
Pulsing controlFinger switchFinger switchFinger switchFinger switchFinger switchFinger switchSame
DeliverydeviceDirect illumination to tissueDirectilluminationtotissueDirect illumination to tissueDirectilluminationtotissueDirectilluminationtotissueDirect IlluminationTo TissueSame
Outputintensity level9 levels9 levelsNot publicNot public5 levelsNot publicSame
Software/Firmware/Microprocessor Control?YesYesYesYesYesNot publicSame

{8}------------------------------------------------

{9}------------------------------------------------

VII. Non-Clinical Testing

The following performance data were provided in support of the substantial equivalence determination.

  • ISO 10993-5: 2009, Biological evaluation of medical devices -Par t 5: Tests for in vitro cytotoxicity

  • ISO 10993-10: 2021, Biological evaluation of medical devices Part 10: Tests for skin sensitization

  • ISO 10993-23: 2021, Biological evaluation of medical devices Part 23: Tests for irritation

  • IEC 60601-1: 2020, Medical electrical equipment Part 1: General requirements for basic safety and essential performance

  • IEC 60601-1-2: 2020, Medical electrical equipments for basic safety and essential performance -

Collateral Standard: Electromagnetic disturbances - Requirements and tests

IEC 60601-1-11: 2020. Medical Electrical Equipment -Part 1-11: General Requirements for Basic Safety and Essential Performance -Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment

IEC 60601-2-83: 2022, Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment

  • IEC 62471: 2006, Photobiological safety of lamps and lamp systems

IEC 60601-1-6: 2020, Medical electrical equipments for basic safety and essential performance -Collateral standard: Usability

VIII. Clinical Testing

Not applicable.

IX. Conclusions

Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device is as safe, as efforms as well as the legally marketed predicate devices.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.