K Number
K242610

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-03-11

(189 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Endo Wrist Monopolar Curved Scissors is intended to be used with the da Vinci Xi System or the da Vinci X Surgical System for endoscopic manipulation of tissue, including: cutting, blunt and sharp dissection, electrocautery.

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Surgical System, Model IS4000 and da Vinci Surgical System, Model IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

Device Description

The da Vinci X/Xi 8 mm monopolar scissors instrument (470179) is used in conjunction with its intended surgical robot for cutting, cauterizing, coagulation, manipulating and blunt dissue. This instrument consists of the housing, shaft, wrist, and tip. The shaft and wrist allow for different axis of rotation, and the instrument with the patient tissue. This instrument is reusable and is provided non-sterile. The instrument is used with a single use tip cover accessory.

Endo Wrist Instruments are designed to provide surgeons with natural dexterity and a greater range of motion than even the human hand. This allows for greater precision in a minimally invasive environment. EndoWrist instruments, in conjunction with the applicable surgical robot, are designed to support rapid and precise suturing, dissection, and tissue manipulation in surgical procedures.

The design, materials, and intended use of the Reusable Surgical Instruments, after an additional ten (10) reuse cycles, are substantially equivalent to the predicate device in form, fit, and function. The reusable device is identical to the predicate device in that the same standard mechanicals, and size are utilized. There are no changes to the claims, intended use, clinical applications, patient population, performance specifications, or method of operation.

AI/ML Overview

The provided text is a 510(k) premarket notification for a medical device: an 8mm Monopolar Curved Scissors (470179) for use with the da Vinci Xi/X Surgical System. The notification states that the device is a remanufactured/reprocessed version of an existing device, increasing its usability from 10 to 20 cycles through a refurbishment process.

The document discusses acceptance criteria and testing, but not in the context of an AI-powered diagnostic or assistive technology. Instead, the acceptance criteria and study described relate to the safety and effectiveness of the remanufactured physical surgical instrument itself.

Therefore, most of the requested information regarding AI device performance (e.g., sample size for test set, data provenance, number of experts, adjudication method, MRMC study, standalone performance, type of ground truth, training set size, ground truth for training set) is not applicable to this submission.

Here's a breakdown of the relevant information provided:

1. A table of acceptance criteria and the reported device performance:

Acceptance Criteria (Tests Performed)Reported Device Performance / Conclusion
BiocompatibilityConcluded that the reprocessed devices are as safe and effective as the predicate.
Functional Performance TestingConcluded that the reprocessed devices are as safe and effective as the predicate and operate as originally intended.
Bioburden, Cleaning, and Bacterial EndotoxinsConcluded that the reprocessed devices are as safe and effective as the predicate.
Electrical SafetyConcluded that the reprocessed devices are as safe and effective as the predicate.
Validation of ReprocessingImplicitly met, as performance testing demonstrates reprocessed devices are safe and effective.

2. Sample size used for the test set and the data provenance:

  • Sample Size for Test Set: Not explicitly stated. The document refers to "design verification and design validation testing." This typically involves testing a representative number of units, but the specific count is not provided.
  • Data Provenance: Not applicable in the context of clinical data for an AI device. The data comes from laboratory and engineering testing of the physical remanufactured surgical instruments. The country of origin of the data is not specified, but it would be from the manufacturer's testing facilities. The tests are prospective in nature, as they are conducted on the remanufactured devices to prove their performance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. The "ground truth" here is objective physical and electrical properties, and biological safety, measured through standardized tests, not expert interpretation of medical images or data. Engineering and quality control personnel would be involved in conducting and evaluating these tests.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable. Adjudication methods like 2+1 are used for expert consensus on subjective assessments. The tests described are objective and rely on measurement against predefined specifications.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. This device is a physical surgical instrument, not an AI-powered diagnostic or assistive technology. Therefore, an MRMC study is not relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is a remanufactured physical surgical tool.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • The "ground truth" used for this device's acceptance is based on predefined engineering specifications, material properties, electrical safety standards, and biocompatibility requirements, as well as the performance characteristics of the predicate device. It's objective, measurable criteria from a combination of regulatory standards and the original device's design.

8. The sample size for the training set:

  • Not applicable. This is not an AI/machine learning device.

9. How the ground truth for the training set was established:

  • Not applicable. This is not an AI/machine learning device.

In summary: The provided FDA submission for the 8mm Monopolar Curved Scissors focuses on demonstrating the substantial equivalence and safety/effectiveness of a remanufactured physical surgical instrument. The acceptance criteria and studies are related to the physical, electrical, and biological performance of this instrument, not to the performance of an AI algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 11, 2025

Iconocare Health Rick Ferreira President 7825 East Redfield Rd. Suite 103 Scottsdale, Arizona 85260

Re: K242610

Trade/Device Name: 8mm Monopolar Curved Scissors (470179) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: OSM, NAY Dated: February 4, 2025 Received: February 4, 2025

Dear Rick Ferreira:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2025.03.11
10:41:30-04'00' Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K242610

Device Name

8mm Monopolar Curved Scissors (470179)

Indications for Use (Describe)

The Endo Wrist Monopolar Curved Scissors is intended to be used with the da Vinci Xi System or the da Vinci X Surgical System for endoscopic manipulation of tissue, including: cutting, blunt and sharp dissection, electrocautery.

The Intuitive Surqical Endoscopic Instrument Control System (da Vinci Surgical System, Model IS4000 and da Vinci Surgical System, Model IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) #:K242610
510(k) Summary
Prepared on: 2025-02-03
Contact Details21 CFR 807.92(a)(1)
--------------------------------------
Applicant NameIconocare Health
Applicant Address7825 East Redfield Rd. Suite 103 Scottsdale AZ 85260 US
Applicant Contact Telephone480-467-8517
Applicant ContactMr. Rick Ferreira
Applicant Contact Emailrferreira@alliancehcpartners.com
Device Name21 CFR 807.92(a)(2)
----------------------------------
Device Trade Name8mm Monopolar Curved Scissors (470179)
Common Name876.1500
Classification NameSystem, Surgical, Computer Controlled Instrument, Remanufactured
Regulation Number21 CFR §876.1500
Product Code(s)QSM

Legally Marketed Predicate Devices

21 CFR 807.92(a)(3)
-----------------------
Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K2200238mm Monopolar Curved ScissorsNAY
K171294da Vinci X Surgical SystemNAY
Device Description Summary21 CFR 807.92(a)(4)
-------------------------------------------------

Device Description Summal y

The da Vinci X/Xi 8 mm monopolar scissors instrument (470179) is used in conjunction with its intended surgical robot for cutting, cauterizing, coagulation, manipulating and blunt dissue. This instrument consists of the housing, shaft, wrist, and tip. The shaft and wrist allow for different axis of rotation, and the instrument with the patient tissue. This instrument is reusable and is provided non-sterile. The instrument is used with a single use tip cover accessory.

Endo Wrist Instruments are designed to provide surgeons with natural dexterity and a greater range of motion than even the human hand. This allows for greater precision in a minimally invasive environment. EndoWrist instruments, in conjunction with the applicable surgical robot, are designed to support rapid and precise suturing, dissection, and tissue manipulation in surgical procedures.

The design, materials, and intended use of the Reusable Surgical Instruments, after an additional ten (10) reuse cycles, are substantially equivalent to the predicate device in form, fit, and function. The reusable device is identical to the predicate device in that the same standard mechanicals, and size are utilized. There are no changes to the claims, intended use, clinical applications, patient population, performance specifications, or method of operation.

In accordance with the Design control process, risk analysis was conducted to evaluate the impact of the predicate device. Design verification and design validation testing were confirm that the design outputs meet the design input requirements and that the device is safe and effective for its included the following tests and/or rationale-based evaluations:

{5}------------------------------------------------

· Biocompatibility

· Functional performance testing

• Bioburden, cleaning and bacterial endotoxins

· Electrical safety

The performance testing demonstrates that remanufactured devices are as safe and effective as the predicate device and operate as originally intended. K210478 is a reference device to support the testing methodology

Intended Use/Indications for Use

21 CFR 807.92(a)(5)

The Endo Wrist Monopolar Curved Scissors is intended to be used with the da Vinci Xi System or the da Vinci X Surgical System for endoscopic manipulation of tissue, including: cutting, blunt and sharp dissection, electrocautery.

The Intuitive Surgical Endoscopic Instrument Control System, Model IS4000 and da Vinci Surgical System, Model IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatic use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

Indications for Use Comparison

Based on the intended use, indications for use, technological characteristics, and performance data, the subject 8mm Monopolar Curved Scissors is substantially equivalent to the predicate device.

Technological Comparison

The da Vinci Xi system is a multiport robotic system that is designed for a wide spectrum of procedures and specialties. This system is compatible with multiple types of surgical instruments. The 8 mm monopolar scissors instrument (470179) and is indicated for 10 cycles of use. The subject process proposed by Iconocare will increase the cycles to 10 through a refurbishment process.

The design, materials, and intended use of Reusable Surgical Instruments, after an additional ten (10) reuse cycles are equivalent to the predicate device. The mechanism of the reusable device is identical to the predicate device in that the same standard mechanical design, materials, and sizes are utilized. There are no changes to the claims, intended use, clinical applications, patient population, performance specifications, or method of operation. Each individual device is tested for appropriate function of its components prior to packaging and labeling operations.

Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

In accordance with the Design Control process, risk analysis was conducted to evaluate the impact of modifications to the predicate device. Design verification and design validation testing were confirm that the design outputs meet design input requirements and that the device is safe and effective for its included the following tests: · Biocompatibility

  • · Validation of Reprocessing
  • · Functional Performance Testing

· Electrical Safety Testing

The performance testing demonstrates that reprocessed devices are as safe and effective as the predicate and operate as originally intended.

21 CFR 807.92(a)(5)

21 CFR 807.92(a)(6)

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.