K Number
K242521

Validate with FDA (Live)

Device Name
Estar-ZE
Date Cleared
2024-11-20

(89 days)

Product Code
Regulation Number
872.6660
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Estar-ZE is a ceramic intended to manufacture dental restorations, including inlays, artificial teeth, crowns and bridges.

Device Description

Estar-ZE is a ceramic intended to manufacture dental restorations, including inlays, artificial teeth, crowns and bridges.

Estar-ZE is prefabricated ceramic block (pre-sintered yttrium-stabilized zirconium oxide) which is to be milled and sintered in the furnace to produce the final dental restorations. After sintering, it forms polycrytstalline oxide ceramic consisted of Tetragonal Zirconium Oxide Polycrystal (TZP). In accordance with ISO 6872:2015, it is classified as Type II Class 5 zirconia.

Estar-ZE is provided in non-sterile and available in various shades and thickness of disk shape and block shape

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Estar-ZE device, based on the provided FDA 510(k) summary:

Device: Estar-ZE (Porcelain Powder for Clinical Use)

1. Table of Acceptance Criteria and Reported Device Performance

ParameterAcceptance Criteria (Standard)Reported Device Performance
Material TypeISO 6872:2015 ClassificationType II Class 5 zirconia
Flexural Strength>800 MPa>800 MPa
BiocompatibilityISO 10993 standardsMeets requirements

Study Proving Device Meets Acceptance Criteria:

The provided document describes non-clinical performance testing to demonstrate substantial equivalence of the Estar-ZE device.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the exact sample sizes for the mechanical properties or biocompatibility testing. It only states that the device was "designed and tested" and that "all tests have passed the evaluation criteria."

The data provenance is not explicitly stated as retrospective or prospective, nor is the country of origin specified for the test data itself (although the manufacturer is based in Korea).

3. Number of Experts Used to Establish Ground Truth and Qualifications

This information is not applicable as the device is a ceramic material and the testing described is non-clinical (mechanical properties, biocompatibility), not involving human diagnostic interpretation or ground truth established by experts.

4. Adjudication Method

This information is not applicable for the reasons stated above.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

A MRMC comparative effectiveness study was not performed. This type of study is typically relevant for AI/software devices where human readers provide diagnoses or interpretations.

6. Standalone (Algorithm Only) Performance Study

A standalone performance study was not performed. This is also relevant for AI/software devices. The Estar-ZE is a physical dental material.

7. Type of Ground Truth Used

The "ground truth" for this device's performance is based on established international standards:

  • ISO 6872:2015 Dentistry - Ceramic Materials for mechanical properties (e.g., flexural strength, classification).
  • ISO 10993 suite of standards for biocompatibility.

8. Sample Size for the Training Set

This information is not applicable. The Estar-ZE is a physical ceramic material, not an AI/machine learning algorithm requiring a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the reasons stated above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 20, 2024

Osstem Implant Co., Ltd. % Peter Lee RA/QA Manager Hiossen Inc. 85 Ben Fairless Dr. Fairless Hills, Pennsylvania 19030

Re: K242521

Trade/Device Name: Estar-ZE Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: July 25, 2024 Received: August 23, 2024

Dear Peter Lee:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/2/Picture/3 description: The image contains the text "MICHAEL E. ADJODHA-S". The text is written in a simple, sans-serif font and is arranged in two lines. The first line contains "MICHAEL E." and the second line contains "ADJODHA-S".

Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K242521

Device Name

Estar-ZE

Indications for Use (Describe)

Estar-ZE is a ceramic intended to manufacture dental restorations, including inlays, artificial teeth, crowns and bridges.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Osstem Implant. The logo is orange and features the word "OSSTEM" in a bold, sans-serif font. Below the word "OSSTEM" is the word "IMPLANT" in a smaller, sans-serif font. The logo is simple and modern.

Osstem Implant Co., Ltd.

A-dong, 51, Mayu-ro 238beon-gil, Siheung-si, Gyeonggi-do 15079 Korea Tel: +82 70 4394 7896 Fax: +82 31 498 0824 www.osstem.com

510(k) Summary

K242521

Date: November 12, 2024

1. Company and Correspondent making the submission

-Submitter's Name:Osstem Implant Co., Ltd.
-Address:A-dong, 51, Mayu-ro 238beon-gil, Siheung-si, Gyeonggi-do15079 Korea, Republic of Korea
-Contact:Ms. Seungju Kang
-Phone:+82-70-4871-0203
-Correspondent's Name:Hiossen Inc.
-Address:85 Ben Fairless Dr. Fairless Hills, PA 19030
-Contact:Mr. Peter Lee
-Phone:+1-267-759-7031
Proposed Device

Powder, Porcelain

21 CFR 872.6660

Class II

EIH

:

:

:

:

2.

  • Trade or (Proprietary) Name Estar-ZE -:
  • -Classification Name
  • -Regulation Number
    • Device Classification
    • Classification Product Code

3. Predicate Device(s)

Primary Predicate

Trade or (Proprietary) Name: Estar-Z (K210097)
Classification Name: Powder, Porcelain
Regulation Number: 21 CFR 872.6660
Device Classification: Class II
Classification Product Code: EIH

4. Description

Estar-ZE is a ceramic intended to manufacture dental restorations, including inlays, artificial teeth, crowns and bridges.

Estar-ZE is prefabricated ceramic block (pre-sintered yttrium-stabilized zirconium oxide) which is to be milled and sintered in the furnace to produce the final dental restorations. After sintering, it forms polycrytstalline oxide ceramic consisted of Tetragonal Zirconium Oxide Polycrystal (TZP). In accordance with ISO 6872:2015, it is classified as Type II Class 5 zirconia.

Estar-ZE is provided in non-sterile and available in various shades and thickness of disk shape and block shape

5. Substantial Equivalence Matrix

Subject DevicePrimary PredicateComparison
ManufacturerOsstem Implant Co., Ltd.Osstem Implant Co., Ltd.-

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for OSSTEM IMPLANT. The logo is in orange and features the word "OSSTEM" in a bold, sans-serif font. Below the word "OSSTEM" is the word "IMPLANT" in a smaller, sans-serif font. The logo is simple and modern.

Osstem Implant Co., Ltd.
A-dong, 51, Mayu-ro 238bon-gil, Siheung-si, Gyeonggi-do 15079 Korea
Tel: +82 70 4394 7896 Fax: +82 31 498 0824 www.osstem.com

Device NameEstar-ZEEstar-Z HTDifferent
510(k) NumberK242521K210097-
Regulation NamePorcelain powder for clinical usePorcelain powder for clinical useSame
Regulation Number21 CFR 872.666021 CFR 872.6660Same
Product CodeEIHEIHSame
Device ClassIIIISame
Indications for UseEstar- ZE is a ceramic intended tomanufacture dental restorations,including inlays, artificial teeth,crowns and bridgesEstar-Z is a ceramic intended tomanufacture dental restorations,including inlays, artificial teeth,crowns and bridgesSame
Principle ofOperationsThis partial sintered zirconia block ismilled and finally sintered to makedental prosthesis.This partial sintered zirconia block ismilled and finally sintered to makedental prosthesis.Same
MaterialscompositionZirconium dioxideFerric OxideErbium(III) oxideManganese(II) oxideHolder: Stainless Steel, AluminumAlloyZirconium dioxideIron(III) oxideErbium(III) oxideCobalt(II,III) oxideDifferent
Physical Properties
StandardConformedISO 6872:2015ISO 6872:2015Same
ClassificationType II Class 5Type II Class 5Same
FlexuralStrength>800 MPa>800 MPaSame
DeliveryForm(s)Disk typeBlock typeDisk typeDifferent
GeometryØ 98 mm disc in different heightsfrom 10 – 30 mmHeight 13 – 40 block in differentlength from 14 – 85 mmØ98 mm disc in different heights from10 – 25 mmDifferent
Shade(s)A0, A1, A2, A3, A3.5, A4, B1, B2,B3, B4, C1, C2, C3, C4, D2, D3, D4,S4, Gradation A1, Gradation A2,Gradation A3, Gradation A3.5,Gradation A4, Gradation B1,Gradation B2, Gradation B3,Gradation B4, Gradation C1,Gradation C2, Gradation C3,Gradation C4, Gradation D2,Gradation D3, Gradation D4A0, A1, A2, A3, A3.5, A4,B1, B2, B3, B4,C1, C2, C3, C4,D2, D3, D4, Z4,Multi A1, Multi A2, Multi A3, MultiA3.5, Multi A4, Multi B1, Multi B2,Multi B3, Multi B4,Multi C1, Multi C2, Multi C3, MultiC4, Multi D2, Multi D3, Multi D4Different
SterileNon sterileNon sterileSame
SubstantialEquivalenceSimilaritiesThe subject device has the same indications for use, principle of operations, manufacturer, andmanufacturing process etc. In addition, the subject device is classified in the same Type II Class 5as its precated Estar-Z HT according to the ISO 6872, and both products meet the requirementsfrom ISO 6872 standard.DifferencesThe material composition is different from that of primary predicate due to the difference in thetype of coloring additives. However, zirconia powder, the main raw material that makes up mostof the subject device, is the same, and we conducted biocompatibility tests of subject device fordifferences in coloring additives. According to the test results, the subject device meets therequirements from ISO 10993 standards.The delivery forms differ from primary predicate due to differences in the way it is combinedwith the milling machine and how it is stored and managed after use.However, since the disk type and block type have the same raw material, principle of operations,intended use, manufacturing process, etc., the performance of disk type and block type is same,and we conducted performance test of disk type of subject device.The subject device has a different geometry and shades from primary predicate for variety ofproducts, but we demonstrate substantial equivalence through performance test andbiocompatibility test of subject device.: Therefore, the Estar-ZE is substantially equivalent previously cleared primary predicate.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Osstem Implant. The logo is orange and features the word "OSSTEM" in a bold, sans-serif font. Below the word "OSSTEM" is the word "IMPLANT" in a smaller, italicized font. There is a small circle above the right side of the word "OSSTEM".

Osstem Implant Co., Ltd.

A-dong, 51, Mayu-ro 238beon-gil, Siheung-si, Gyeonggi-do 15079 Korea

Tel: +82 70 4394 7896 Fax: +82 31 498 0824 www.osstem.com

6. Indications for Use

Estar-ZF is a ceramic intended to manufacture dental restorations, including inlays, artificial teeth, crowns and bridges

7. Summary of Non-clinical Performance Testing

Non-clinical testing data are submitted to demonstrate substantial equivalence.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Osstem Implant. The logo is in orange and gray. The word "OSSTEM" is in orange, and the word "IMPLANT" is in gray and is located below the word "OSSTEM".

Osstem Implant Co., Ltd.

A-dong, 51, Mayu-ro 238beon-gil, Siheung-si, Gyeonggi-do 15079 Korea Tel: +82 70 4394 7896 Fax: +82 31 498 0824 www.osstem.com

Biocompatibility Evaluation

The biocompoatibility testing was considered followed the FDA Guidance Document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process, " and the ISO 10993 suite of standards. The biocompatibility for the proposed device was found to be substantially equivalent to the predicate devices as a result.

Mechanical Properties

Proposed Estar-ZE has been designed and tested in accordance with ISO 6872 Dentistry - Ceramic Materials and ISO 13356 Implants for surgery – Ceramic materials based on vttria-stabilized tetragonal zirconia (Y-TZP). All tests have passed the evaluation criteria and met the requirement of product-specifics for Class 5 dental ceramics. The mechanical properties were found to be substantially equivalent to the predicate devices as a result.

Sterilization Validation and Shelf-life

Proposed Estar-Z Multi is delivered in non-sterilization validation and shelf-life testing was not considered.

8. Summary of Clnical Testing

No clinical studies are submitted.

9. Conclusion

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, Osstem Implant Co., Ltd. concludes that the Estar-ZF is substantially equivalent to the predicated device as herein.

§ 872.6660 Porcelain powder for clinical use.

(a)
Identification. Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.(b)
Classification. Class II.