K Number
K242498

Validate with FDA (Live)

Date Cleared
2024-10-01

(40 days)

Product Code
Regulation Number
862.3100
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Pocguide Multi-Drug Test Panel OTC is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5dimethyl-3,3-diphenylpyrrolidine, Methylenedioxy-methamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline and Marijuana in human urine at the cutoff concentrations of:

Drug ( Identifier)Cut-off level
Amphetamine (AMP)1000 ng/mL or 500 ng/mL
Buprenorphine (BUP)10 ng/mL
Secobarbital (BAR)300 ng/mL
Oxazepam (BZO)300 ng/mL
Benzoylecognine (COC)300 ng/mL or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300 ng/mL
Methamphetamine (MET)1000 ng/mL or 500 ng/mL
Methylenedioxymethamphetamine (MDMA)500 ng/mL
Morphine (OPI2000/MOP300)2000 ng/mL or 300 ng/mL
Methadone (MTD)300 ng/mL
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Nortriptyline (TCA)1000 ng/mL
Marijuana (THC)50 ng/mL

The single or multi-test panels can consist of up to the above listed analytes in any combination. The tests provide only a preliminary result. A more specific alternative chemical must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.

For over-the-counter use. For in vitro diagnostic use only

Device Description

Pocguide™ Multi-Drug Test Panel and Pocguide™ Multi-Drug Test Panel OTC are immunochromatographic assays that use a lateral flow system for the qualitative detection of single drugs in human urine at or above the cut-off levels as indicated. The products are single use in vitro diagnostic devices.

This device is a dipcard format in which the test strips are integrated into the plastic dipcard. After removing the cap of the dipcard, the absorbent end of the test strips is exposed and can be in direct contact with the urine sample. The device is in a ready-to-use format and no longer requires assembly before use.

AI/ML Overview

The provided document describes the Pocguide Multi-Drug Test Panel and Pocguide Multi-Drug Test Panel OTC, which are in vitro diagnostic devices for qualitative and simultaneous detection of various drugs in human urine.

Here's an analysis of the acceptance criteria and the study proving the device meets those criteria, based on the provided text:


1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for this type of qualitative diagnostic device are typically related to its analytical performance, specifically precision (reproducibility) around the cutoff concentration, and its ability to correctly identify positive and negative samples when compared to a confirmed method (method comparison).

Acceptance Criteria (Implied based on study design and regulatory context for qualitative drug tests):

  • Precision/Reproducibility: A high percentage of agreement (e.g., typically >80%) for samples near the cutoff (e.g., +/- 25% cutoff, cutoff itself) over multiple lots and runs, and 100% agreement for samples far from the cutoff (e.g., +/- 100% cutoff).
  • Method Comparison: High overall agreement (sensitivity and specificity) with a gold standard confirmatory method (LC/MS or GC/MS) for clinical samples, especially for true positive and true negative samples. Acceptable performance for samples near the cutoff where some discordance is expected due to the nature of qualitative assays.
  • Analytical Specificity (Cross-Reactivity): No significant cross-reactivity with common substances or structurally similar compounds to avoid false positives.
  • Interference: No interference from common physiological substances found in urine.
  • Lay-User Study (for OTC devices): High agreement with trained professionals and ease of use for the intended lay user.

Reported Device Performance (from "Precision/Reproducibility" and "Method Comparison" sections):

Test CharacteristicDrug (Cutoff) ExamplesReported Device Performance
Precision/ReproducibilityAMP 500 ng/mL, BUP 10 ng/mL, etc.+100%, +75%, +50%, +25% Cutoff: 100.0% Positive (Across all tested drugs and cutoffs) -100%, -75%, -50%, -25% Cutoff: 100.0% Negative (Across all tested drugs and cutoffs) Cutoff: * AMP 500: 82.0% Positive, 18.0% Negative * BUP 10: 84.0% Positive, 16.0% Negative * BAR 300: 82.7% Positive, 17.3% Negative * And similar ranges for other drugs listed in Table 2. Each drug showed similar performance around the cutoff.
Method ComparisonAMP 1000 ng/mL, AMP 500 ng/mL, BAR 300 ng/mL, etc.Excellent agreement for Drug-Free, Low Negative, and High Positive Samples: Typically 100% correct classification by the device for these categories (e.g., "Viewer A Positive" for High Positive by LC/MS and "Viewer A Negative" for Drug-Free by LC/MS are 100% for almost all drugs). Expected Discordance Near Cutoff: As anticipated for qualitative tests, some samples near the cutoff (especially -25% and +25%) show mixed results (discordance) between the device and LC/MS, as detailed in Table 5 and Table 6 (Discordant results). These are typically within acceptable ranges for qualitative tests, acknowledging the inherent variation around a precise cutoff.
Analytical SpecificityAMP, BUP, BAR, etc.Tested numerous substances. Most showed no cross-reactivity or very low percentages at very high concentrations, indicating good specificity. Specific cross-reactivity percentages are provided in Table 3.
InterferenceN/A - broadly testedNo interference observed for a wide range of common substances and physiological conditions (urine specific gravity 1.000-1.035, pH 4-9) as listed in Tables 4.
Lay-User StudyAll Configuration 1 & 2 DrugsAgreement (%): * -100%, -75%, -50% Cutoff: 100% negative calls. * +25%, +50%, +75% Cutoff: Mostly 100% positive calls, some 95% for +25% cutoff. * -25% Cutoff: 95% negative calls for most drugs. * Raw numbers show 19/20 or 20/20 correct calls for most categories. Ease of Use: All participants indicated instructions were easy to understand and follow (Flesch-Kincaid Grade Level 7).

2. Sample Sizes and Data Provenance

  • Test Set Sample Sizes:
    • Precision/Reproducibility: For each drug and each cutoff, 50 samples were tested at each concentration level (-100%, -75%, -50%, -25%, Cutoff, +25%, +50%, +75%, +100%). This was done across 3 lots, so $50 \text{ samples/level} \times 9 \text{ levels} \times 3 \text{ lots} = 1350$ tests per drug. (For AMP alone, this would be $1350 \times 2 \text{ cutoffs} = 2700$ tests).
    • Method Comparison: 100 unaltered clinical samples were used for each target drug (40 negative, 40 positive, and an additional 20 samples around the cutoff as seen in the breakdown of results). So, for 13 drug analytes, this would be $13 \times 100 = 1300$ clinical samples.
    • Cross-Reactivity / Interference: Specific numbers for each substance are not given, but samples were spiked at various concentrations and tested using three lots of each device.
    • Lay-User Study: For Configuration 1, 140 participants (58 male, 82 female). For Configuration 2, 140 participants (56 male, 84 female). Each participant tested 1 blind-labeled sample. For each drug within each configuration, 20 samples were prepared per concentration level (-100%, -75%, -50%, -25%, +25%, +50%, +75%).
  • Data Provenance: The document does not explicitly state the country of origin for the data. The consulting firm is in Shanghai, China, and the applicant's address is Irvine, CA, USA. Given the FDA 510(k) submission, it's implied that the data is intended to be representative and valid for the US market. The studies are described as retrospective as they involve samples prepared at specific concentrations or existing clinical samples compared to a gold standard.

3. Number of Experts and their Qualifications

  • For Precision/Reproducibility, Cross-Reactivity, Interference, and Method Comparison: The document does not explicitly state the number of "experts" used to establish ground truth or interpret results. These are quantitative/analytical laboratory tests where the ground truth (concentration by LC/MS or GC/MS) is established by analytical instrumentation. The "Viewers" (A, B, C) mentioned in the Method Comparison section appear to be individuals performing the visual interpretation of the device results, not necessarily independent experts establishing ground truth. Their qualifications are not specified but are implied to be trained laboratory personnel.
  • For Lay-User Study: No "experts" were used to establish ground truth for the lay-user study. The ground truth for the samples used in this study was established by LC-MS/MS confirming the spiked drug concentrations.

4. Adjudication Method for the Test Set

  • For Precision/Reproducibility, Cross-Reactivity, Interference: No adjudication method is described. The results are reported as counts of positive/negative readings against a known (spiked) concentration.
  • For Method Comparison: No explicit "adjudication" among multiple readers is described. Results for the candidate device were observed by "Viewer A, B, C." The comparison is directly between the "Candidate Device Result" (presumably individual Viewer results, though aggregated in Table 5) and the LC/MS reference method. The discordant results in Table 6 specify which viewers made the discordant call (e.g., "Viewer A, B," "Viewer C"). This suggests independent readings by three viewers, but no formal adjudication process to resolve disagreements among them is mentioned; the individual viewer calls are presented relative to the ground truth.
  • For Lay-User Study: No adjudication method is mentioned. The tables report the number of positive/negative results per concentration.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No MRMC comparative effectiveness study was done. The study design is an analytical performance study and a comparison study against a laboratory reference method, along with a lay-user study for OTC claims. It assesses device performance in a standalone or simulated user setting, not direct human reader improvement with AI assistance. The device is a lateral flow immunoassay, not an AI-powered diagnostic.

6. Standalone (Algorithm Only) Performance

  • This question is not applicable as the device is a lateral flow immunochromatographic assay, not an algorithm or software-based diagnostic. Its performance is inherent to the chemical reactions on the test strip and visual interpretation, not an algorithm.

7. Type of Ground Truth Used

  • The primary ground truth used for performance evaluation (Precision/Reproducibility, Method Comparison, Lay-User Study) is analytical confirmation by Gas Chromatography-Mass Spectrometry (GC-MS) or Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions (LC-MS/MS), which are stated as the "preferred confirmatory methods." This is a highly accurate and quantitative method for determining drug concentrations.

8. Sample Size for the Training Set

  • This question is not applicable. The device is a qualitative diagnostic test based on immunoassay principles, not a machine learning or AI-based device that requires a "training set" in the computational sense. The development of the immunoassay itself relies on antigen-antibody binding characteristics and optimization, not a deep learning model.

9. How the Ground Truth for the Training Set Was Established

  • This question is not applicable for the same reasons as #8. The "ground truth" for developing the test and optimizing its performance would be established through standard immunoassay R&D processes, involving controlled experiments with known concentrations of analytes and cross-reactants, guided by established analytical chemistry principles.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a larger font size than the rest of the name. The logo is simple and professional, reflecting the agency's role in protecting public health.

Aicheck Biotech, Inc. % Dylan Wu IVD Senior Consultant Shanghai Sungo Management Consulting Co., Ltd. 14th Floor, 1500# Century Avenue Shanghai, 200122, China

Re: K242498

Trade/Device Name: Pocguide Multi-Drug Test Panel, Pocguide Multi-Drug Test Panel OTC Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine test system Regulatory Class: Class II Product Code: NFT, NGL, PTH, NFV, NFY, PTG, NGG, NGM, QAW, NFW Dated: August 22, 2024 Received: August 22, 2024 Dear Dylan Wu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely. Digitally signed by Joseph A. Joseph A. Kotarek -S Kotarek -S Date: 2024.10.0 10:15:07 -04'00' Joseph Kotarek Branch Chief for Toxicology Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K242498

Device Name

Pocguide Multi-Drug Test Panel OTC Pocguide Multi-Drug Test Panel

Indications for Use (Describe)

Pocquide Multi-Drug Test Panel OTC is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5dimethyl-3,3-diphenylpyrrolidine, Methylenedioxy-methamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline and Marijuana in human urine at the cutoff concentrations of:

Drug ( Identifier)Cut-off level
Amphetamine (AMP)1000 ng/mL or 500 ng/mL
Buprenorphine (BUP)10 ng/mL
Secobarbital (BAR)300 ng/mL
Oxazepam (BZO)300 ng/mL
Benzoylecognine (COC)300 ng/mL or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300 ng/mL
Methamphetamine (MET)1000 ng/mL or 500 ng/mL
Methylenedioxymethamphetamine (MDMA)500 ng/mL
Morphine (OPI2000/MOP300)2000 ng/mL or 300 ng/mL
Methadone (MTD)300 ng/mL
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Nortriptyline (TCA)1000 ng/mL
Marijuana (THC)50 ng/mL

The single or multi-test panels can consist of up to the above listed analytes in any combination. The tests provide only a preliminary result. A more specific alternative chemical must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.

For over-the-counter use. For in vitro diagnostic use only

Pocguide Multi-Drug Test Panel is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methylenedioxy-methamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline and Marijuana in human urine at the cutoff concentrations of:

Drug ( Identifier)Cut-off level
Amphetamine (AMP)1000 ng/mL or 500 ng/mL
Buprenorphine (BUP)10 ng/mL
Secobarbital (BAR)300 ng/mL
Oxazepam (BZO)300 ng/mL
Benzoylecognine (COC)300 ng/mL or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300 ng/mL
Methamphetamine (MET)1000 ng/mL or 500 ng/mL
Methylenedioxymethamphetamine (MDMA)500 ng/mL
Morphine (OPI2000/MOP300)2000 ng/mL or 300 ng/mL
Methadone (MTD)300 ng/mL
Oxycodone (OXY)100 ng/mL

{3}------------------------------------------------

Phencyclidine (PCP)25 ng/mL
Nortriptyline (TCA)1000 ng/mL
Marijuana (THC)50 ng/mL
The single or multi-test panels can consist of up to the above listed analytes in any combination.
The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended
to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should
be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.
The tests provide only a preliminary result. A more specific alternative chemical method must be used to obtain a
confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass
Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful
consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating
preliminary positive results.
For in vitro diagnostic use only

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) Summary

K242498

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement by 21 CFR 807.92

Date prepared: 15th, Aug 2024

1 Submitter's Information

Name: Aicheck Biotech, Inc. Address: 17701 Cowan Ste 230, Irvine, CA, USA, 92614 Contact Person: Lisa Liu Title: Manager of the quality department Tel: +86-571-89055299 Email: ra.rep@aichek.com

2 Official Contact Person Information

Primary contact: Mr. Dylan Wu Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-18616797137 Email: haiyang.wu@sungoglobal.com Secondary contact: Mr. Raymond Luo Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: zxfda@sungoglobal.com

3 Subject Device

3.1 Trade Name and Regulatory Information

Pocguide™ Multi-Drug Test Panel Pocguide™ Multi-Drug Test Panel OTC

3.2 Regulatory Information

Drug (Identifier)Code (s)ClassificationRegulation SectionPanel
Amphetamine (AMP)NFTClass II21 CFR § 862.3100Amphetamine Test SystemClinicalToxicology
Buprenorphine (BUP)NGLClass II21 CFR § 862.3650Opiate Test SystemClinicalToxicology
Secobarbital (BAR)PTHClass II21 CFR § 862.3150Barbiturate test systemClinicalToxicology
Oxazepam (BZO)NFVClass II21 CFR § 862.3170Benzodiazepine test systemClinicalToxicology
Benzoylecognine (COC)NFYClass II21 CFR § 862.3250Cocaine and Cocaine Metabolites Test SystemClinicalToxicology
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)PTGClass II21CFR § 862.3620Methadone Test SystemClinicalToxicology
Methamphetamine (MET)NGGClass II21 CFR § 862.3610Methamphetamine Test SystemClinicalToxicology

{5}------------------------------------------------

Drug (Identifier)Code (s)ClassificationRegulation SectionPanel
Methylenedioxymethamphetamine (MDMA)NGGClass II21 CFR § 862.3610Methamphetamine Test SystemClinicalToxicology
Morphine (OPI/MOP)NGLClass II21 CFR § 862.3650Opiate Test SystemClinicalToxicology
Methadone (MTD)PTGClass II21CFR § 862.3620Methadone Test SystemClinicalToxicology
Oxycodone (OXY)NGLClass II21 CFR § 862.3650Opiate Test SystemClinicalToxicology
Phencyclidine (PCP)NGMUnclassified
Nortriptyline (TCA)QAWClass II21 CFR § 862.3910Tricyclic antidepressant drugs test systemClinicalToxicology
Marijuana (THC)NFWClass II21 CFR § 862.3870Cannabinoid test systemClinicalToxicology

CA IICA 92614

4 Predicate device

K202567

Wondfo T-Dip® Multi-Drug Urine Test Panel Wondfo T-Dip® Multi-Drug Urine Test Panel Rx

5 Indications for use/Intended use

Pocguide™ Multi-Drug Test Panel is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5dimethyl-3,3-diphenylpyrrolidine, Methylenedioxy-methamphetamine, Morphine, Mothadone, Oxycodone, Phencyclidine, Nortriptyline and Marijuana in human urine at the cutoff concentrations of:

Drug (Identifier)Cut-off level
Amphetamine (AMP)1000 ng/mL or 500 ng/mL
Buprenorphine (BUP)10 ng/mL
Secobarbital (BAR)300 ng/mL
Oxazepam (BZO)300 ng/mL
Benzoylecognine (COC)300 ng/mL or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300 ng/mL
Methamphetamine (MET)1000 ng/mL or 500 ng/mL
Methylenedioxymethamphetamine (MDMA)500 ng/mL
Morphine (OPI2000/MOP300)2000 ng/mL or 300 ng/mL
Methadone (MTD)300 ng/mL
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Nortriptyline (TCA)1000 ng/mL
Marijuana (THC)50 ng/mL

The single or multi-test panels can consist of up to the above listed analytes in any combination.

The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. The tests provide only a preliminary result. A more specific alternative chemical method must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.

For in vitro diagnostic use only.

{6}------------------------------------------------

Pocguide™ Multi-Drug Test Panel OTC is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Secobarbital, Oxazepam, Cocaine, 2ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methamphetamine, Methylenedioxy-methamphetamine, Morphine, Methadone, Oxycodone, Phencycline and Marijuana in human urine at the cutoff concentrations of:

Drug (Identifier)Cut-off level
Amphetamine (AMP)1000 ng/mL or 500 ng/mL
Buprenorphine (BUP)10 ng/mL
Secobarbital (BAR)300 ng/mL
Oxazepam (BZO)300 ng/mL
Benzoylecognine (COC)300 ng/mL or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300 ng/mL
Methamphetamine (MET)1000 ng/mL or 500 ng/mL
Methylenedioxymethamphetamine (MDMA)500 ng/mL
Morphine (OPI2000/MOP300)2000 ng/mL or 300 ng/mL
Methadone (MTD)300 ng/mL
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Nortriptyline (TCA)1000 ng/mL
Marijuana (THC)50 ng/mL

The single or multi-test panels can consist of up to the above listed analytes in any combination.

The tests provide only a preliminary result. A more specific alternative chemical must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.

For over-the-counter (OTC) use. For in vitro diagnostic use only.

б Device Description

Pocguide™ Multi-Drug Test Panel and Pocguide™ Multi-Drug Test Panel OTC are immunochromatographic assays that use a lateral flow system for the qualitative detection of single drugs in human urine at or above the cut-off levels as indicated. The products are single use in vitro diagnostic devices.

This device is a dipcard format in which the test strips are integrated into the plastic dipcard. After removing the cap of the dipcard, the absorbent end of the test strips is exposed and can be in direct contact with the urine sample. The device is in a ready-to-use format and no longer requires assembly before use.

7 Principle of Operation

Pocguide™ Multi-Drug Test Panel is a competitive immunoassay that is used to screen for the presence of drugs of abuse in urine. It is chromatographic absorbent device in which drugs or drug metabolites in a sample competitively combined to a limited number of antibody-dye conjugate binding sites. When the absorbent end of the test device is immersed into the urine sample, the urine is absorbed into the device by capillary action, mixes with the antibody

{7}------------------------------------------------

dye conjugate, and flows across the pre-coated membrane.

When sample drug levels are at or above the target cutoff (the detection sensitivity of the drug in the sample binds to the antibody-dye conjugate preventing the antibody-dye conjugate from binding to the drug-protein pre-coated in the test region (T). This prevents the development of a distinct colored band in the test region indicating a potentially positive result.

When sample drug levels are zero or below the target cutoff, antibody-dye conjugate binds to the drug-protein precoated in the test region (T) of the device. This produces a colored test line that, regardless of its intensity, indicates a negative result.

To serve as a procedural control, a colored line will always appear at the control region (C) indicating that proper volume of specimen has been added and membrane wicking has occurred.

{8}------------------------------------------------

Substantial Equivalence Information 8

8.1 Technological characteristics

Table 1 Comparison of Technological characteristics
DeviceProposed DevicePredicate Device
510K numberK242498K202567
Device namePocguide™ Multi-Drug Test PanelPocguide™ Multi-Drug Test Panel OTCWondfo T-Dip® Multi-Drug Urine Test PanelWondfo T-Dip® Multi-Drug Urine Test Panel Rx
Indication (s) for useFor the qualitative determination of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Propoxyphene, Nortriptyline and Marijuana in human urine.Same
MethodologyCompetitive binding, lateral flow immunochromatographic assay based on antigen-antibody reactionSame
Type of TestQualitativeSame
Specimen TypeHuman UrineSame
Target Drugand CutoffValueTarget DrugCutoff (ng/mL)
Amphetamine (AMP)1000 or 500
Buprenorphine (BUP)10
Secobarbital (BAR)300
Oxazepam (BZO)300
Benzoylecognine (COC)300 or 150
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300
Methamphetamine (MET)1000 or 500
Methylenedioxymethamphetamine (MDMA)500
Morphine (OPI2000/MOP300)2000 or 300
Methadone (MTD)300
Oxycodone (OXY)100
Phencyclidine (PCP)25
Nortriptyline (TCA)1000
Marijuana (THC)50The number of drugs detected are different, the predicate device also includes Propoxyphene (PPX) 300 ng/mL
ConfigurationsTest PanelSame

{9}------------------------------------------------

д Summary of Non-Clinical Testing

9.1 Analytical performance

Precision / Reproducibility a

Precision studies were carried out for samples with contract, - 75% atoff, -25% cutoff, -25% cutoff, -15% cutoff, +50% cutoff +15% cutoff +15% cutoff and +100% cutoff. Samples with concentration of -100% cutoff were prepared by spiking target drug in drug-free urines samples. Each drug concentration was confirmed by LC/MS. All sample aliquots were tested in a blinded fashion, tests were performed in 2 runs per day for 25 days using 3 lots of test panels. The results obtained are summarized in Table 2.

Drug testedClaimed Cutoff%cutoffLot 1Lot 2Lot 3Total Result%Positive%Negative
+-+-+-+-
AMP500+100% cutoff5005005001500100.0%0.0%
+75% cutoff5005005001500100.0%0.0%
+50% cutoff5005005001500100.0%0.0%
+25% cutoff5005005001500100.0%0.0%
AMP500cutoff41942840101232782.0%18.0%
AMP500-25% cutoff050050050015000.0%100.0%
-50% cutoff050050050015000.0%100.0%
-75% cutoff050050050015000.0%100.0%
-100% cutoff050050050015000.0%100.0%
AMP1000+100% cutoff5005005001500100.0%0.0%
+75% cutoff5005005001500100.0%0.0%
+50% cutoff5005005001500100.0%0.0%
+25% cutoff5005005001500100.0%0.0%
AMP1000cutoff4284374371282285.3%14.7%
AMP1000-25% cutoff050050050015000.0%100.0%
-50% cutoff050050050015000.0%100.0%
-75% cutoff050050050015000.0%100.0%
-100% cutoff050050050015000.0%100.0%
BUP10+100% cutoff5005005001500100.0%0.0%
+75% cutoff5005005001500100.0%0.0%
+50% cutoff5005005001500100.0%0.0%
+25% cutoff5005005001500100.0%0.0%
BUP10cutoff4374194281262484.0%16.0%
BUP10-25% cutoff050050050015000.0%100.0%
-50% cutoff050050050015000.0%100.0%
-75% cutoff050050050015000.0%100.0%
Negative050050050015000.0%100.0%
BAR300+100% cutoff5005005001500100.0%0.0%
+75% cutoff5005005001500100.0%0.0%
+50% cutoff5005005001500100.0%0.0%
+25% cutoff5005005001500100.0%0.0%
BAR300cutoff4284194191242682.7%17.3%
BAR300-25% cutoff050050050015000.0%100.0%
-50% cutoff050050050015000.0%100.0%
-75% cutoff050050050015000.0%100.0%
-100% cutoff050050050015000.0%100.0%
BZO300+100% cutoff5005005001500100.0%0.0%
+75% cutoff5005005001500100.0%0.0%
+50% cutoff5005005001500100.0%0.0%
+25% cutoff5005005001500100.0%0.0%
BZO300cutoff4374194281262484.0%16.0%
BZO300-25% cutoff050050050015000.0%100.0%
-50% cutoff050050050015000.0%100.0%
-75% cutoff050050050015000.0%100.0%
-100% cutoff050050050015000.0%100.0%
COC300+100% cutoff5005005001500100.0%0.0%
+75% cutoff5005005001500100.0%0.0%
+50% cutoff5005005001500100.0%0.0%
+25% cutoff5005005001500100.0%0.0%
COC300cutoff4374284191262484.0%16.0%
COC300-25% cutoff050050050015000.0%100.0%
-50% cutoff050050050015000.0%100.0%
-75% cutoff050050050015000.0%100.0%
-100% cutoff050050050015000.0%100.0%
Table 2 Precision / Reproducibility results of Pocquide™ Multi-Drug Test Panel
--------------------------------------------------------------------------------------

{10}------------------------------------------------

Aicheck Biotech, Inc. 17701 Cowan Ste 230, Irvine, CA, USA, 92614

Drug testedLot 1Lot 2Lot 3Total Result%Positive%Negative
Claimed Cutoff%cutoff+-+-+-+
+100% cutoff5005005001500100.0%0.0%
+75% cutoff5005005001500100.0%0.0%
+50% cutoff5005005001500100.0%0.0%
+25% cutoff5005005001500100.0%0.0%
COCcutoff4194194281242682.7%17.3%
150-25% cutoff05005005001500.0%100.0%
-50% cutoff05005005001500.0%100.0%
-75% cutoff05005005001500.0%100.0%
-100% cutoff05005005001500.0%100.0%
+100% cutoff5005005001500100.0%0.0%
+75% cutoff5005005001500100.0%0.0%
+50% cutoff5005005001500100.0%0.0%
+25% cutoff5005005001500100.0%0.0%
EDDPcutoff4284374371282285.3%14.7%
300-25% cutoff05005005001500.0%100.0%
50500500
-50% cutoff001500.0%100.0%
-75% cutoff05005005001500.0%100.0%
-100% cutoff05005005001500.0%100.0%
+100% cutoff5005005001500100.0%0.0%
+75% cutoff5005005001500100.0%0.0%
+50% cutoff5005005001500100.0%0.0%
MET+25% cutoff5005005001500100.0%0.0%
1000cutoff40104194191222881.3%18.7%
-25% cutoff05005005001500.0%100.0%
-50% cutoff05005005001500.0%100.0%
-75% cutoff05005005001500.0%100.0%
-100% cutoff05005005001500.0%100.0%
+100% cutoff5005005001500100.0%0.0%
+75% cutoff5005005001500100.0%0.0%
+50% cutoff5005005001500100.0%0.0%
+25% cutoff5005005001500100.0%0.0%
METcutoff4374194281262484.0%16.0%
500-25% cutoff05005005001500.0%100.0%
-50% cutoff05005005001500.0%100.0%
-75% cutoff05005005001500.0%100.0%
Negative05005005001500.0%100.0%
+100% cutoff5005005001500100.0%0.0%
+75% cutoff5005005001500100.0%0.0%
+50% cutoff5005005001500100.0%0.0%
+25% cutoff5005005001500100.0%0.0%
MDMAcutoff4194374281262484.0%16.0%
500-25% cutoff05005005001500.0%100.0%
-50% cutoff05005005001500.0%100.0%
-75% cutoff05005005001500.0%100.0%
-100% cutoff05005005001500.0%100.0%
+100% cutoff5005005001500100.0%0.0%
+75% cutoff5005005001500100.0%0.0%
+50% cutoff5005005001500
100.0%0.0%
OPI+25% cutoff5005005001500100.0%0.0%
2000cutoff4194374281262484.0%16.0%
-25% cutoff05005005001500.0%100.0%
-50% cutoff05005005001500.0%100.0%
-75% cutoff05005005001500.0%100.0%
-100% cutoff05005005001500.0%100.0%
+100% cutoff5005005001500100.0%0.0%
MOP300+75% cutoff5005005001500100.0%0.0%
+50% cutoff5005005001500100.0%0.0%
+25% cutoff5005005001500100.0%0.0%
cutoff42842840101242682.7%17.3%
-25% cutoff05005005001500.0%100.0%
-50% cutoff05005005001500.0%100.0%
-75% cutoff05005005001500.0%100.0%
Drug tested%cutoffLot 1Lot 2Lot 3Total Result%Positive%Negative
Claimed Cutoff+-+-+-+-
-100% cutoff05005005001500.0%100.0%
MTD300+100% cutoff5005005001500100.0%0.0%
+75% cutoff5005005001500100.0%0.0%
+50% cutoff5005005001500100.0%0.0%
+25% cutoff5005005001500100.0%0.0%
cutoff4284284371272384.7%15.3%
-25% cutoff05005005001500.0%100.0%
-50% cutoff05005005001500.0%100.0%
-75% cutoff05005005001500.0%100.0%
-100% cutoff05005005001500.0%100.0%
OXY100+100% cutoff5005005001500100.0%0.0%
+75% cutoff5005005001500100.0%0.0%
+50% cutoff5005005001500100.0%0.0%
+25% cutoff5005005001500100.0%0.0%
cutoff4374464371302086.7%13.3%
-25% cutoff05005005001500.0%100.0%
-50% cutoff05005005001500.0%100.0%
-75% cutoff05005005001500.0%100.0%
-100% cutoff05005005001500.0%100.0%
PCP25+100% cutoff5005005001500100.0%0.0%
+75% cutoff5005005001500100.0%0.0%
+50% cutoff5005005001500100.0%0.0%
+25% cutoff5005005001500100.0%0.0%
cutoff4284374281272384.7%15.3%
-25% cutoff05005005001500.0%100.0%
-50% cutoff05005005001500.0%100.0%
-75% cutoff05005005001500.0%100.0%
-100% cutoff05005005001500.0%100.0%
TCA1000+100% cutoff5005005001500100.0%0.0%
+75% cutoff5005005001500100.0%0.0%
+50% cutoff5005005001500100.0%0.0%
+25% cutoff5005005001500100.0%0.0%
cutoff4374194281262484.0%16.0%
-25% cutoff05005005001500.0%100.0%
-50% cutoff05005005001500.0%100.0%
-75% cutoff05005005001500.0%100.0%
-100% cutoff05005005001500.0%100.0%
THC50+100% cutoff5005005001500100.0%0.0%
+75% cutoff5005005001500100.0%0.0%
+50% cutoff5005005001500100.0%0.0%
+25% cutoff5005005001500100.0%0.0%
cutoff4284374371282285.3%14.7%
-25% cutoff05005005001500.0%100.0%
-50% cutoff05005005001500.0%100.0%
-75% cutoff05005005001500.0%100.0%

{11}------------------------------------------------

Aicheck Biotech, Inc. 17701 Cowan Ste 230, Irvine, CA, USA, 92614

Linearity/assay reportable range b

Linearity is not applicable since this is a qualitative test.

Analytical specificity/Interference C

c.1 Cross-Reactivity

To test cross-reactivity, drug netabolites and other compounds that are likely to cross-react in urine samples were spiked into drug-free unine sample and were tested using three lots of each device. The lowed a positive result for each compound is listed below along with the corresponding % crossreactivity. The highest concentration tested is shown if no cross reactivity was observed.

Drug (Cutoff)SubstancesMinimum concentration requiredto obtain a positive result (ng/mL)% Cross-Reactivity
AMP 500d-Amphetamine500100%
l-Amphetamine150003.30%
dl- Amphetamine150033.30%
(+/-) 3,4-Methylenedioxyamphetamine (MDA)500010%

Table 3 Cross-Reactivity results of Pocguide™ Multi-Drug Test Panel

{12}------------------------------------------------

Drug (Cutoff)SubstancesMinimum concentration requiredto obtain a positive result (ng/mL)% Cross-Reactivity
Phentermine150033.30%
Hydroxyamphetamine80006.25%
d-Methamphetamine>100000--
l-Methamphetamine>100000--
(+/-) 3,4-Methylenedioxyethylamphetamine (MDE)>100000--
(+/-) 3,4-Methylenedioxymethamphetamine (MDMA)>100000--
Ephedrine>100000--
β-Phenylethylamine1000000.50%
Tyramine1000000.50%
p-Hydroxynorephedrine1000000.50%
Phenylpropanolamine>100000--
(±) Phenylpropanolamine>100000--
p-Hydroxyamphetamine1000000.50%
d/l-Norephedrine1000000.50%
Benzphetamine>100000--
l-Ephedrine>100000--
l-Epinephrine>100000--
d/l-Epinephrine>100000--
d-Amphetamine1000100%
AMP 1000l-Amphetamine300003.30%
dl- Amphetamine300033.30%
(+/-)3,4-Methylenedioxyamphetamine (MDA)500020%
Phentermine300033.30%
Hydroxyamphetamine800012.50%
d-Methamphetamine> 100000--
l-Methamphetamine> 100000--
(+/-)3,4-Methylenedioxyethylamphetamine (MDEA)> 100000--
(+/-)3,4-Methylenedioxymethamphetamine (MDMA)> 100000--
Ephedrine>100000--
β-Phenylethylamine1000001%
Tyramine1000001%
p-Hydroxynorephedrine1000001%
Phenylpropanolamine> 100000--
(±) Phenylpropanolamine> 100000--
p-Hydroxyamphetamine1000001%
d/l-Norephedrine1000001%
Benzphetamine> 100000--
l-Ephedrine> 100000--
l-Epinephrine> 100000--
d/l-Epinephrine> 100000--
BUP 10Buprenorphine10
Buprenorphine -3-D-Glucuronide1566.67%
Norbuprenorphine2050%
Norbuprenorphine-3-D-Glucuronide10010%
Morphine> 100000-
Oxymorphone> 100000-
Hydromorphone> 100000-
Secobarbital300100%
Amobarbital50006%
BAR 300Alphenol150200%
Aprobarbital200150%
Butabarbital150200%
Butethal50600%
Butalbital250012%
Cyclopentobarbital60050%
Pentobarbital200015%
Phenobarbital50006%
BZO 300Oxazepam300100%
Alprazolam200150%
a-Hydroxyalprazolam100030%
Bromazepam50060%
Chlordiazepoxide150020%

Aicheck Biotech, Inc. 17701 Cowan Ste 230, Irvine, CA, USA, 92614

{13}------------------------------------------------

Drug (Cutoff)SubstancesMinimum concentration required to obtain a positive result (ng/mL)% Cross-Reactivity
Clobazam100300%
Clonazepam300010%
Clorazepate dipotassium200150%
Delorazepam150020%
Desalkylflurazepam40075%
Diazepam200150%
Estazolam50060%
Flunitrazepam100030%
Midazolam50006%
Nitrazepam100030%
Norchlordiazepoxide200150%
Nordiazepam50060%
Temazepam250120%
Triazolam120025%
Demoxepam200015%
Flurazepam50060%
D, L-Lorazepam150020%
COC 300Benzoylecgonine300100%
Cocaine100030%
Cocaethylene150002%
Ecgonine300001%
Ecgonine methyl ester>100000--
Norcocaine>100000--
COC 150Benzoylecgonine150100%
Cocaine50030%
Cocaethylene62502.40%
Ecgonine160000.94%
Ecgonine methyl ester> 100000--
Norcocaine> 100000--
EDDP 3002-ethylidene-1,5-dimethyl-3,3- diphenylpyrrolidine300100%
Methadone2000000.15%
EMDP3000000.10%
Doxylamine> 100000--
Disopyramide> 100000--
LAAM (Levo-alpha-acetylmethadol) HCl> 100000--
Alpha Methadol> 100000--
D (+)-Methamphetamine1000100%
(+/-)3,4-Methylenedioxy-n-ethylamphetamine (MDEA)200050%
D/L-Methamphetamine1000100%
p-Hydroxymethamphetamine300003.30%
D-Amphetamine> 100000--
L-Amphetamine1000001%
Chloroquine200005%
(+/-) -Ephedrine500002%
MET 1000(-) -Methamphetamine250004%
(+/-)3,4-Methylenedioxyamphetamine (MDA)250040%
(+/-)3,4-Methylenedioxymethamphetamine (MDMA)400025%
ẞ-Phenylethylamine500002%
Trimethobenzamide1000010%
d,l-Amphetamine1000001%
Mephetermine500002%
(1R,2S)-(-)-Ephedrine> 100000--
1-phenylephrine> 100000--
L-Methamphetamine200005%
MET 500D (+) -Methamphetamine500100%
(+/-)3,4-Methylenedioxy-n-ethylamphetamine (MDEA)200025%
D/L-Methamphetamine500100%
MET 500p-Hydroxymethamphetamine(+/-)3,4-Methylenedioxy-n-ethylamphetamine (MDEA)1500020003.33%25%
D/L-Methamphetamine500100%
p-Hydroxymethamphetamine150003.33%
D-Amphetamine500001%
L-Amphetamine1000000.50%
Chloroquine100005%
(+/-)-Ephedrine250002%

Aicheck Biotech, Inc. 17701 Cowan Ste 230, Irvine, CA, USA, 92614

{14}------------------------------------------------

Drug (Cutoff)SubstancesMinimum concentration requiredto obtain a positive result (ng/mL)% Cross-Reactivity
(-)-Methamphetamine125004%
(+/-)3,4-Methylenedioxyamphetamine (MDA)200025%
(+/-)3,4-Methylenedioxymethamphetamine (MDMA)200025%
β-Phenylethylamine250002%
Trimethobenzamide500010%
d,l-Amphetamine750000.67%
Mephetermine250002%
(1R,2S)- (-)-Ephedrine500001%
l-phenylephrine1000000.50%
L-Methamphetamine100005%
3,4-Methylenedioxymethamphetamine (MDMA)500100%
3,4-Methylenedioxyamphetamine HCl (MDA)300016.67%
3,4-Methylenedioxyethylamphetamine (MDEA)500100%
MDMA 500d-methamphetamine> 100000--
d-amphetamine> 100000--
l-methamphetamine500001%
l-amphetamine> 100000--
Morphine2000100%
Normorphine500004%
Codeine2000100%
s-Monoacetylmorphine2000100%
Ethyl Morphine1500133.30%
OPI 2000Heroin2000100%
Hydrocodone1250016%
Hydromorphone350057.10%
Morphinie-3-β-d-glucuronide2000100%
Oxycodone250008%
Oxymorphone250008%
Thebaine500040%
Levorphanol750002.70%
6-Monoacetylmorphine (6-MAM)1500133.30%
Norcodeine1250016%
Procaine1500001.30%
MOP 300Morphine300100%
Normorphine250120%
Codeine300100%
s-Monoacetylmorphine300100%
Ethyl Morphine100300%
Heroin300100%
Hydrocodone50006%
Hydromorphone200015%
Morphinie-3-β-d-glucuronide100030%
Oxycodone> 100000--
Oxymorphone1000000.30%
Thebaine300010%
Levorphanol50006%
6-Monoacetylmorphine (6-MAM)150200%
Norcodeine65004.60%
Procaine1000000.30%
MTD 300Methadone300100%
Doxylamine300010%
EDDP> 100000--
EMDP> 100000--
LAAM> 100000--
Alpha Methadol> 100000--
OXY 100Oxycodone100100%
Dihydrocodeine200000.50%
Hydrocodone100001%
Oxymorphone100010%
Codeine1000000.10%
Hydromorphone320000.30%
Morphine>100000--

Aicheck Biotech, Inc. 17701 Cowan Ste 230, Irvine, CA, USA, 92614

{15}------------------------------------------------

Drug (Cutoff)SubstancesMinimum concentration requiredto obtain a positive result (ng/mL)% Cross-Reactivity
Acetylmorphine>100000--
Buprenorphine>100000--
Ethylmorphine>100000--
Thebaine>100000--
PCP 25Phencyclidine25100%
4-Hydroxyphencyclidine125000.20%
Nortriptyline1000100%
Nordoxepine1000100%
Trimipramine200050%
Amitriptyline150066.70%
Promazine150066.70%
Desipramine400250%
TCA 1000Imipramine400250%
Clomipramine125008%
Doxepin1000100%
Maprotiline200050%
Promethazine250004%
Cyclobenzaprine150066.70%
Norclomipramine1000010%
THC 5011-nor-Δ9-THC-9-COOH50100%
(-)-11-nor-9-carboxy-Δ9-THC50100%
11-nor-Δ8-THC-9-COOH30166.67%
11-nor-Δ9-THC-carboxy glucuronide10050%
Cannabinol200000.25%
Cannabidiol1000000.05%
Δ8- Tetrahydrocannabinol13003.80%
Δ9- Tetrahydrocannabinol50001%
511-hydroxy-Δ9-Tetrahydrocannabinol50001%

Aicheck Biotech, Inc. 17701 Cowan Ste 230, Irvine, CA, USA, 92614

c.2 Interference Study

Potential interfering substances found in human urine of physiological conditions were added to drug-free urine and urine samples with concentrations of the target drugs at 25% below and 25% above Cutoff levels. These urine three lots of each device. The compounds were tested at 100 µg/ml (unless otherwise noted in the table). No interference was observed for anyounds tested, and the list of compounds is provided in the following table.

Table 4 Interfering substances

AcetaminophenCiprofloxacin Hydrochloride1% EthanolLidocaine HydrochlorideOndansetran
AcetophenetidinCitalopramFenofibrateLisinoprilPaliperidone
Acetylsalicylic AcidClarithromycinFenoprofenLithium CarbonatePantoprazole
AcyclovirClonidineFentanyl CitrateLiveritePapaverine
AfrinClopidogrel Hydrogen SulphateFluoxetine HydrochlorideLoperamideParoxetine Hydrochloride
Albumin(100mg/dL)ClozapineFluvoxamineLoratadinePenfluridol
AminophyllineConjugated EstrogensFurosemideMagnesiumPenicillinV Potassium
AminopyrineCortisoneGabapentinMeperidinePenicillin-G
Amiodarone HydrochlorideCreatinineGentisic AcidMeprobamatePhenelzine
Amlodipine Mesylate(-) CotinineGlibenclamideMetoprolol TartratePioglitazone Hydrochloride
AmoxicillinchlorpheniramineGliclazideMifepristonePiracetam
AmpicillinD, L-OctopamineGlipizideMirtazapinePravastatin Sodium
ApomorphineD, L-PropranololGlucoseMontelukast SodiumPrednisone
AripiprazoleD, L-TyrosineHaloperidolMosapride CitratePropylthiouracil
AspartameDeoxy- corticosteroneHemoglobinMinocyclineQuetiapine Fumarate
AtomoxetineDextromethorphanHydrochlorothiazideNalidixic AcidQuinine
Atorvastatin CalciumDiclofenacHydrocortisoneNaproxenRanitidine
AtropineDiflunisal3-HydroxytyramineNiacinamideRifampicin
Benzilic AcidDigoxinIsosorbide DinitrateNifedipineRisperidone
Benzoic AcidDiphenhydramineIsoxsuprineNikethamideSalicylic Acid
BilirubinDirithromycinIbuprofenNimodipineSerotonin
BupropionDomperidoneKetoconazoleNitroglycerinSertraline Hydrochloride
CaptoprilD-PseudoephedrineKetoprofenNorethindroneSildenafil Citrate
CarbamazepineDuloxetineKetamineN-Acetylprocain-amideSimvastatin
CefradineDicyclomineKratom powder0-Hydroxyhippuric AcidSodium Valproate
CephalexinEffexorLabetalolOlanzapineSpironolactone
Chloral HydrateEnalapril MaleateLamotrigineOmeprazoleSulfamethazine
ChloramphenicolErythromycinLevofloxacin HydrochlorideOxalic AcidSulindac

{16}------------------------------------------------

ChlorothiazideEsomeprazole MagnesiumLevonorgestrelOxolinic AcidTetracycline
Cholesterolß-EstradiolLevothyroxine SodiumOxymetazolineTetrahydrocortisone 3 -acetate
Tetrahydrocortisone 3-(ß-D glucuronide)ThioridazineTrazodone HydrochlorideTrimethoprimVerapamil
TetrahydrozolineTopiramateTriamtereneUric AcidVitamin B2
ThiamineTramadol HydrochlorideTrifluoperazineValproateVitamin C
ChloroquineEcgonine Methyl EsterDoxylamineEpinephrine HydrochlorideMaprotiline
Promethazine

c.3 Effect of urine specific gravity and pH

To investigate the effect of urine specific gravity and urine samples, with specific gravity ranging from 1.000 to 1.035 or urine samples with pH ranging from 4 to 9 were spiked with the target drugs to concentrations at -25% Cutoff levels. These samples were tested using three lots of each device. Results vere all postive for samples at +25% Cutoff and all negative for samples demonstrated that urine specific gravity levels of 1000 to 1.035 and urine pH levels of 4 to 9 do not affect the results of the assays.

Traceability, Stability, Expected values (controls, calibrators, or methods): d

d.1 Traceability

The device is traceable to commercially available materials.

d.2 Stability

The stability data supports that the products have the shelf life of 24 months when stored at 4-30°C.

Real time Stability

A 27-month real time stability test is planned to vertify of the device as 24 morths. Three batches for each configuration in sealed foil pouch are currently stable for 21 months at 4°C and 30°C, and the real time stability study is still on-going.

Accelerated Stability

Stable at 4-30°C for 24 months based on the accelerated stability study at 55°C for 31 days and real time stability determination at both 4°C and 30°C.

e Detection limit

Not Applicable.

Assay Cut-off f

Refer to section 9.1.a.

9.2 Comparison Studies

Method comparison studies a

Method comparison studies were performed by the 100 unaltered clinical samples (40 negative and 40 positive) were run for each target drug. The samples were tested in a blinded fashion and compared to LC/MS results are presented in Table 5 below, and the discordant results are summarized in below.

DrugcutoffCandidate Device ResultDrug-FreeLow Negative by LC/MS(less than - 50%)Near Cutoff Negative by LC/MS(Between -50% and cutoff)Near Cutoff Positive by LC/MS(Between cutoff and +50%)High Positive by LC/MS(greater than +50%)
AMP 1000Viewer APositive0011425
Negative7151710
Viewer BPositive0011525
Negative7151700
Viewer CPositive0011425
Negative7151710
AMP 500Viewer APositive0021523
Negative1371820
Viewer BPositive0021623
Negative1371810
Viewer CPositive0011523
Negative1371920
BAR 300Viewer APositive0011919
Negative10151420
Viewer BPositive0011919
Negative10151420
Viewer CPositive0011919
Negative10151420
BUP 10Viewer APositive0032018
Negative10151220
Viewer BPositive0032018
Negative10151020

Table 5 The Results of Method comparison with predicate device

Viewer CPositive0022018
Negative10151320
BZO 300Viewer APositive0021820
Negative10151320
Viewer BPositive0021820
Negative10151320
Viewer CPositive0031820
Negative10151220
COC 300Viewer APositive0001821
Negative9161510
Viewer BPositive0001721
Negative9161520
Viewer CPositive0011921
Negative9161400
COC 150Viewer APositive0002115
Negative10151540
Viewer BPositive0012315
Negative10151420

12

2

าก

0

1Q

15

Negative

Docitive

10

0

{17}------------------------------------------------

Aicheck Biotech, Inc. 17701 Cowan Ste 230, Irvine, CA, USA, 92614

Drug cutoffCandidate Device ResultDrug-FreeLow Negative by LC/MS (less than - 50%)Near Cutoff Negative by LC/MS (Between -50% and cutoff)Near Cutoff Positive by LC/MS (Between cutoff and +50%)High Positive by LC/MS (greater than +50%)
EDDP 300Viewer A Positive0002115
Viewer A Negative10141540
Viewer A Positive0021920
Viewer A Negative10141410
Viewer B Positive0031920
Viewer B Negative10141310
Viewer C Positive0022020
Viewer C Negative10141400
MET 1000Viewer A Positive0011622
Viewer A Negative8131820
Viewer B Positive0021722
Viewer B Negative8131710
Viewer C Positive0021722
Viewer C Negative8131710
MET 500Viewer A Positive0012116
Viewer A Negative8201130
Viewer B Positive0022316
Viewer B Negative8201010
Viewer C Positive0002116
Viewer C Negative8201230
OPI 2000Viewer A Positive0012218
Viewer A Negative10151400
Viewer B Positive0012018
Viewer B Negative10151420
Viewer C Positive0022018
Viewer C Negative10151320
MOP 300Viewer A Positive0012315
Viewer A Negative10141520
Viewer B Positive0022315
Viewer B Negative10141420
Viewer C Positive0012315
Viewer C Negative10141520
MTD 300Viewer A Positive0011424
Viewer A Negative10181120
Viewer B Positive0011524
Viewer B Negative10181110
Viewer C Positive0001424
Viewer C Negative10181220
OXY 100Viewer A Positive0022315
Viewer A Negative10151320
Viewer B Positive0022315
Viewer B Negative10151320
Viewer C Positive0012215
Viewer C Negative10151430
PCP 25Viewer A Positive0012018
Viewer A Negative9161420
Viewer B Positive0012018
Viewer B Negative9161420
Viewer C Positive0002018
Viewer C Negative9161520
TCA 1000Viewer A Positive0022018
Viewer A Negative10151320
Viewer B Positive0011918
Viewer B Negative10151430
Viewer C Positive0022018
Viewer C Negative10151320
THC 50Viewer A Positive0011820
Viewer A Negative10151420
Viewer B Positive0021920
Viewer B Negative10151310
Viewer C Positive0011720
Viewer C Negative10151430
Viewer A PositiveViewer A0Positive0010201181820
Negative10151420
THC 50Viewer BPositive0021920
Negative10151310
Viewer CPositive0011720
Negative10151430
Viewer APositive0012018
Negative10151420
MDMA 500
Viewer BPositive0012118
Negative10151410
Viewer CPositive0012018
Negative10151420

{18}------------------------------------------------

Table 6 Discordant results
Drug cutoffOperatorSample IDLC/MS Results(ng/mL)Test ResultDrug cutoffOperatorSample IDLC/MS Results(ng/mL)Test Result
AMP 1000Viewer CAH621968.6PositiveMET 500Viewer BML217457.3Positive
Viewer A, BAH618970.1PositiveViewer A, BML177491Positive
Viewer A, CAH5621101NegativeViewer A, CML224506Negative
Viewer BAL554472.6PositiveViewer B, CML216512Negative
Viewer AAL486475.7PositiveViewer A, CML168518Negative
AMP 500Viewer A, B, CAL546489PositiveViewer AML233523Negative
Viewer A, CAL491501.2NegativeViewer A, COP0941869.3Positive
Viewer B, CAL551508.9NegativeOPI 2000Viewer B, COP0841983Positive
Viewer AAL483536.2NegativeViewer B, COP1492045Negative
Viewer A, B, CBA436296PositiveViewer B, COP0872080Negative
BAR 300Viewer A, CBA445302.1NegativeViewer BM0033287.5Positive
Viewer A, B, CBA474307.2NegativeViewer A, B, CM0039298Positive
Viewer BBA403319.83NegativeMOP 300Viewer A, B, CM0028301Negative
Viewer A, BBU0018.36PositiveViewer A, CMO009301.6Negative
Viewer A, B, CBU0529.21PositiveViewer BM0071319.5Negative
BUP 10Viewer A, B, CBU0539.42PositiveViewer COX23687Positive
Viewer A, B, CBU00710.2NegativeViewer BOX22892.3Positive
Viewer A, B, CBU00310.5NegativeViewer A, BOX20495Positive
Viewer B, CBZ100282PositiveOXY 100Viewer AOX18596Positive
Viewer A, CBZ112291.2PositiveViewer A, B, COX199100
Viewer A, B, CBZ149298PositiveViewer A, COX209102
BZO 300Viewer CBZ103301NegativeViewer B, COX195103Negative
Viewer A, B, CBZ096305NegativeViewer APC25223.16Positive
Viewer BBZ108309NegativeViewer BPC26023.97Positive
Viewer ABZ155317.6NegativePCP 25Viewer A, B, CPC28625.39Negative
Viewer CCH945294.8PositiveViewer A, B, CPC29725.9
COC 300Viewer A, BCH948328.5NegativeViewer CTA752849Positive
Viewer BCH893338.5NegativeViewer A, CTA776892Positive
Viewer BCL801145PositiveViewer A, BTA724934Positive
Viewer BCL827154NegativeTCA 1000Viewer ATA7741025Negative
Viewer A, CCL806156NegativeViewer B, CTA7611058
COC 150Viewer A, B, CCL839159NegativeViewer B, CTA7671069Negative
Viewer A, CCL830160.1NegativeViewer A, BTA7991079Negative
Viewer A, CCL858163.5NegativeViewer B, CTH33448Positive
Viewer BED366280.8PositiveViewer A, BTH37548.8Positive
Viewer A, B, CED350281PositiveTHC 50Viewer A, CTH37450.35Negative
EDDP 300Viewer A, B, CED335284.1PositiveViewer A, CTH33953.46
Viewer AED331325NegativeViewer B, CTH39854.45Negative
Viewer BED358335.7NegativeViewer BMD642476.1Positive
Viewer A, B, CMH307981.1PositiveMDMA 500Viewer A, CMD708482.34Positive
MET 1000Viewer B, CMH291995.8PositiveViewer A, CMD706502
Viewer A, BMH2461039.7NegativeViewer B, CMD681527.89Negative
Viewer A, CMH3021098NegativeViewer AMD710532Negative
Viewer A, BMT464278.8Positive
MTD 300Viewer A, B, CMT424326.5Negative
Viewer A, CMT475329.8Negative

b Matrix comparison

Not applicable. This device is for testing with human urine only.

10 Clinical studies

Clinical Sensitivity a

Not applicable

Clinical specificity b

Not applicable

Other clinical supportive data (when a. and b. are not applicable) C

c.1 Lay-User Study

The lay-user study was performed at three intensed by and 82 female tested Porguite™ Multi-Drug Test Panel Configuration 1 (including AMP 500, MET 500, NET 500, NET 500, NET MOP 300, COC 150); 56 male and 84 female tested Poguide™ Multi-Drug Test Panel Configuration 2 (including AMP 1000, MET 1000). Ney had

direrse educational and occupational backgrounds and the > 50. Urine samples were prepared at the following concentrations: -100%, +/-75%, +/-50%, +/-25% of the cutoff by spiking drug free-pooled urine specimens. The concentrations of the sample were confirmed by LC-MS/MS Each sample was aliquoted into individual containers and blind-label. Each participant was provided with the package insert, 1 blind label sample and a device. The results are summarized in below.

DrugCutoff(ng/mL)ResultsConcentration
AMP500Negative20202019100
AMPPositive0001192020
AMPTotal20202020202020
AMPAgreement (%)100%100%100%95.00%95.00%100%100%
BAR300Negative20202020100
BARPositive0000192020

Table 7 Result of Pocquide™ Multi-Drug Test Panel Configuration 1 (including AMP 500, MET 500, MOP 300, COC 150)

{19}------------------------------------------------

CutoffConcentration
Drug(ng/mL)Results-100% cutoff-75% cutoff-50% cutoff-25% cutoff+25% cutoff+50% cutoff+75% cutoff
BZO300Negative20202019000
Positive0001202020
Total20202020202020
Agreement (%)100%100%100%95.00%100%100%100%
BUP10Negative20202019100
Positive0001192020
Total20202020202020
Agreement (%)100%100%100%95.00%95.00%100%100%
COC150Negative20202019000
Positive0001202020
Total20202020202020
Agreement (%)100%100%100%95.00%100%100%100%
EDDP300Negative20202019000
Positive0001202020
Total20202020202020
Agreement (%)100%100%100%95.00%100%100%100%
MDMA500Negative20202020100
Positive0000192020
Total20202020202020
Agreement (%)100%100%100%100%95.00%100%100%
MET500Negative20202019100
Positive0001192020
Total20202020202020
Agreement (%)100%100%100%95.00%95.00%100%100%
MOP300Negative20202019000
Positive0001202020
Total20202020202020
Agreement (%)100%100%100%95.00%100%100%100%
MTD300Negative20202020100
Positive0000192020
Total20202020202020
Agreement (%)100%100%100%100%95.00%100%100%
OXY100Negative20202019100
Positive0001192020
Total20202020202020
Agreement (%)100%100%100%95.00%95.00%100%100%
PCP25Negative20202019000
Positive0001202020
Total20202020202020
Agreement (%)100%100%100%95.00%100%100%100%
TCA1000Negative20202019000
Positive0001202020
Total20202020202020
Agreement (%)100%100%100%95.00%100%100%100%
THC50Negative20202019000
Positive0001202020
Total20202020202020
Agreement (%)100%100%100%95.00%100%100%100%
Agreement (%)100%100%100%95.00%100%100%100%
-----------------------------------------------------------
DrugCutoff(ng/mL)ResultsConcentration
-100% cutoff-75% cutoff-50% cutoff-25% cutoff+25% cutoff+50% cutoff+75% cutoff
AMP1000Negative20202019000
Positive0001202020
Total20202020202020
Agreement (%)100%100%100%95.00%100.00%100%100%
BAR300Negative20202019100
Positive0001192020
Total20202020202020
Agreement (%)100%100%100%95.00%95.00%100%100%
DrugCutoff(ng/mL)ResultsConcentration
-100% cutoff-75% cutoff-50% cutoff-25% cutoff+25% cutoff+50% cutoff+75% cutoff
BZO300Negative20202019100
Positive0001192020
Total20202020202020
Agreement (%)100%100%100%95.00%95.00%100%100%
BUP10Negative20202020100
Positive0000192020
Total20202020202020
Agreement (%)100%100%100%100.00%95.00%100%100%
COC300Negative20202020100
Positive0000192020
Total20202020202020
Agreement (%)100%100%100%100.00%95.00%100%100%
EDDP300Negative20202019100
Positive0001192020
Total20202020202020
Agreement (%)100%100%100%95.00%95.00%100%100%
MDMA500Negative20202020100
Positive0000192020
Total20202020202020
Agreement (%)100%100%100%100%95.00%100%100%
MET1000Negative20202020100
Positive0000192020
Total20202020202020
Agreement (%)100%100%100%100.00%95.00%100%100%
OPI2000Negative20202019100
Positive0001192020
Total20202020202020
Agreement (%)100%100%100%95.00%95.00%100%100%
MTD300Negative20202020100
Positive0000192020
Total20202020202020
Agreement (%)100%100%100%100%95.00%100%100%
OXY100Negative20202020100
Positive0000192020
Total20202020202020
Agreement (%)100%100%100%100.00%95.00%100%100%
PCP25Negative20202020100
Positive0000192020
Total20202020202020
Agreement (%)100%100%100%100.00%95.00%100%100%
TCA1000Negative20202019100
Positive0001192020
Total20202020202020
Agreement (%)100%100%100%95.00%95.00%100%100%
THC50Negative20202019000
Positive0001202020
Total20202020202020
Agreement (%)100%100%100%95.00%100%100%100%

Table 8 Result of Pocguide™ Multi-Drug Test Panel Configuration 2 (including AMP 1000, MET 1000, MOP 2000 (OPI), COC 300)

{20}------------------------------------------------

Participants were given surveys on the ease of under on to use. All participants indicated that the device instruction is easy to understand and follow. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.

Clinical Cut-off d

Not applicable

11 Conclusion

Based on the test principle and performance charactive including precision, cut-off, interference, specificity, method comparison and lay-user studies of the devices, it's concluded that Pocguide™ Multi-Drug Test Panel OTC are substantially equivalent to the predicate devices. The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).