K Number
K242485

Validate with FDA (Live)

Device Name
EveryWare
Manufacturer
Date Cleared
2025-05-20

(272 days)

Product Code
Regulation Number
868.5895
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EveryWare is indicated to support clinicians by managing data of patients who are prescribed compatible therapy devices in accordance with the intended use of those therapy devices. EveryWare provides remote patient data collection & viewing and is intended to be used by healthcare representatives in conjunction with compatible non-life support therapy devices to adjust prescription and/or performance settings. EveryWare is intended to be used in hospital, institutional, provider, and home care settings.

Device Description

EveryWare is a platform that gathers treatment information from Breas respiratory devices in a secure cloud-based system. The data from ventilator is sent over a cellular modem through a cellular network to a cloud-hosted secure data storage system. The authenticated users access this cloud hosted system from a browser in their computer.
EveryWare securely connects compatible medical devices located at the point of patient care to the cloud, and provides authorized healthcare representatives the means to manage patient and device information and settings. EveryWare does NOT alter the intended use of connected medical devices or provide functions to automate diagnosis or therapy.

AI/ML Overview

This document is a 510(k) clearance letter for the medical device "EveryWare." It focuses on establishing substantial equivalence to previously cleared predicate devices. Therefore, it does not contain the detailed study information typically found in a clinical trial report or a performance validation study for a novel device.

Based on the provided text, here's what can be extracted and what is NOT available:

1. Table of Acceptance Criteria and Reported Device Performance

  • Acceptance Criteria: Not explicitly stated as quantifiable metrics (e.g., "sensitivity $\ge 90%$"). The document primarily focuses on demonstrating that EveryWare's functionalities and technical characteristics are similar to its predicates, and that its differences "do not raise new questions of safety and effectiveness."
  • Reported Device Performance: Instead of performance metrics, the document lists functional and technological characteristics and states their similarity or difference compared to predicates. Performance is broadly assessed through "non-clinical testing" listed below.
CharacteristicAcceptance Criterion (Implied)EveryWare Performance (Reported)
Intended UseMust be within or similar to predicate devices."EveryWare is indicated to support clinicians by managing data of patients... provides remote patient data collection & viewing and is intended to be used by healthcare representatives... to adjust prescription and/or performance settings." Similar to predicates.
Environment of UseMust be similar to predicate devices."EveryWare is intended to be used in hospital, institutional provider, and home care settings." Similar to predicates.
User PopulationMust be similar to predicate devices."EveryWare is intended to be used by... Prescribing physicians, nurses, respiratory therapists, physiotherapists and technicians - Durable Medical Equipment (DME)/Homecare providers (HCPs)/Home Medical Equipment providers (HMPs)." Similar to predicates.
Application TypeMust be similar to predicate devices."Web-based application." Same as predicates.
Product Codes Supported by Remote Prescription Change FunctionalityMust support relevant product codes, with differences not raising new safety/effectiveness questions.MNS, MNT. Similar to K152356 (which supports MNS, MNT, BZD), but EveryWare does not support BZD. This difference is not presented as raising safety/effectiveness concerns.
Product Codes NOT Supported by Remote Prescription Change FunctionalityMust align with predicates or have justified differences.NOU, CBK. Similar to predicates.
Data Transfer TechnologyMust be robust and achieve remote data transfer.Wireless (cellular modem), SD Card, File upload/Internet. Uses iLink cellular modem. Similar to predicates.
FunctionalityMust include core features of data management and settings.Centralized database, Patient Management, Display therapy data, Generate reports, Settings management for non-life supporting devices. Same as predicates.
Therapy Mode Change (for non-life support devices)Justified difference from predicates if not raising new safety/effectiveness questions.No. Predicates Yes. Different, but "The absence of Therapy Mode Change does not raise new questions of safety and effectiveness."
Target Volume SettingMust be present if offered by predicates.Yes. Same as predicates.
IPAP & EPAP SettingsMust be present if offered by predicates.Yes. Same as predicates.
Breath Rate and Insp. Time SettingMust be present if offered by predicates.Yes. Same as predicates.
Rise Time & Ramp Pressure SettingMust be present if offered by predicates.Yes. Same as predicates.
Humidifier SettingJustified difference from predicates if not raising new safety/effectiveness questions.No. Predicates Yes. Different, but "The absence of Humidifier Setting does not raise new questions of safety and effectiveness."
High Minute Volume, Disconnection & Apnea alarmMust be present if offered by a predicate and relevant.Yes. Similar to K152356.
Mask Resistance & Mask Resistance LockMust align with predicates or have justified differences.No. Similar to K151901.
View Optional ScreenMust align with predicates or have justified differences.No. Similar to K151901.
Tubing Type & Tubing Type LockMust align with predicates or have justified differences.No. Similar to K151901.
Automated Airway ManagementMust align with predicates or have justified differences.No. Similar to K151901.
ReportsMust provide compliance reports.Detailed report (includes compliance information). Same as predicates.
LabelingMust include online help.Online help file within the application. Same as predicates.
Viewing of dataMust allow interactive viewing based on date range.Interactive viewing of data based on user selected date range. Same as predicates.
Safety and EffectivenessMust be demonstrated through compliance with recognized standards and lack of new safety/effectiveness questions.EveryWare was designed and subjected to performance testing in accordance with listed non-clinical standards (e.g., Software, Cybersecurity, Human Factors, EMC). "EveryWare, by itself, is not intended for diagnosis and treatment of disease or medical condition. EveryWare does not change the performance specification of the connected therapeutic devices. Therefore, no further performance data is required to demonstrate substantial equivalence."

2. Sample Size Used for the Test Set and Data Provenance

This information is not provided in the clearance letter. The document mentions "performance testing" and adherence to various standards but does not detail a specific test set, its sample size, or data provenance (e.g., country of origin, retrospective/prospective nature). The focus is on software and system validation rather than clinical dataset performance.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided. As the device is primarily a data management and remote settings platform aiming for "substantial equivalence" based on functional and technical characteristics, and not for diagnostic or therapeutic AI, there is no mention of experts establishing a clinical ground truth for a test set.

4. Adjudication Method for the Test Set

This information is not provided. Given the nature of the device and the presented documentation, it's unlikely an adjudication method for a clinical test set would be applicable or detailed here.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

A MRMC study was not performed nor mentioned. EveryWare is a data management and settings adjustment platform, not an AI-assisted diagnostic or therapeutic tool that would typically involve human readers interpreting AI output.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The device itself is a "platform that gathers treatment information from Breas respiratory devices in a secure cloud-based system" and allows "authorized healthcare representatives the means to manage patient and device information and settings." It does not involve a "standalone algorithm" in the typical sense of a diagnostic or predictive AI. Its performance is tied to its functionality in securely collecting, displaying, and allowing modification of therapy data, which relies on system integrity rather than algorithmic output interpretation. Therefore, a standalone algorithmic performance study in the context of AI was not done or reported.

7. The Type of Ground Truth Used

For EveryWare, the "ground truth" relates to:

  • The accurate and secure transmission, storage, and display of patient data from compatible devices.
  • The correct implementation of remote prescription/performance settings onto compatible devices.
  • Compliance with various software, cybersecurity, human factors, and electromagnetic compatibility standards.

This is fundamentally different from a clinical ground truth (e.g., pathology, outcomes data, or expert consensus for disease detection). The ground truth for this type of device would be defined by technical specifications, data integrity, security protocols, and successful functional validation against a known correct behavior.

8. The Sample Size for the Training Set

This information is not provided. EveryWare is described as a data management platform rather than a machine learning model that would require a "training set" in the context of AI. The development would involve software engineering and validation practices.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable as there is no "training set" mentioned in the context of an AI/ML model for this device. The development process would have involved establishing requirements and validating the software against those requirements, rather than training against a ground-truth dataset.

FDA 510(k) Clearance Letter - EveryWare

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

May 20, 2025

Breas Medical AB
Ivan Liljegren
Senior Vice President, Global QA/RA
Foretagsvagen 1
Molnlycke, 45333
Sweden

Re: K242485
Trade/Device Name: EveryWare
Regulation Number: 21 CFR 868.5895
Regulation Name: Continuous ventilator
Regulatory Class: Class II
Product Code: MOD, MNS, MNT, NOU, CBK
Dated: April 14, 2025
Received: April 16, 2025

Dear Ivan Liljegren:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K242485 - Ivan Liljegren Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K242485 - Ivan Liljegren Page 3

Sincerely,

Ethan L. Nyberg -S

Ethan Nyberg, Ph.D.
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K242485

Device Name
EveryWare

Indications for Use (Describe)

EveryWare is indicated to support clinicians by managing data of patients who are prescribed compatible therapy devices in accordance with the intended use of those therapy devices. EveryWare provides remote patient data collection & viewing and is intended to be used by healthcare representatives in conjunction with compatible non-life support therapy devices to adjust prescription and/or performance settings. EveryWare is intended to be used in hospital, institutional, provider, and home care settings.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) SUMMARY – K242485

Breas Medical
EveryWare

510(k) Owner
Breas Medical AB
Foretagsvagen 1
Molnlycke, Vastra Gotaland SWEDEN 435 33
Telephone: +46 31868800

Submission Correspondent
Ivan Liljegren
Sr. Vice President, Global QARA
Breas Medical AB
Foretagsvagen 1, Molnlycke,
Vastra Gotalands Lan (Se-14), Sweden 453 33

Date Prepared:
May 15, 2025

Trade Names of Device
EveryWare

Common or Usual Name
(Accessory to) Continuous Ventilator

Classification Name
Regulation: 21 C.F.R. §868.5895 Continuous Ventilator, Class II

Primary Predicate Device(s)
ResMed Ltd. AirView cleared in K151901

Secondary Predicate Device(s)
Respironics Inc. Care Orchestrator (Sapphire) cleared in K152356

Device Description

EveryWare is a platform that gathers treatment information from Breas respiratory devices in a secure cloud-based system. The data from ventilator is sent over a cellular modem through a cellular network to a cloud-hosted secure data storage system. The authenticated users access this cloud hosted system from a browser in their computer.

EveryWare is indicated to support clinicians by managing data of patients who are prescribed compatible therapy devices in accordance with the intended use of those therapy devices. EveryWare provides remote patient data collection & viewing and is intended to be used by

Page 6

Page 1 of 7

healthcare representatives in conjunction with compatible non-life support therapy devices to adjust prescription and/or performance settings. EveryWare is intended to be used in hospital, institutional, provider, and home care settings.EveryWare securely connects compatible medical devices located at the point of patient care to the cloud, and provides authorized healthcare representatives the means to manage patient and device information and settings. EveryWare does NOT alter the intended use of connected medical devices or provide functions to automate diagnosis or therapy.

Indications for Use

EveryWare is indicated to support clinicians by managing data of patients who are prescribed compatible therapy devices in accordance with the intended use of those therapy devices. EveryWare provides remote patient data collection & viewing and is intended to be used by healthcare representatives in conjunction with compatible non-life support therapy devices to adjust prescription and/or performance settings. EveryWare is intended to be used in hospital, institutional, provider, and home care settings.

Indications for Use Comparison

The Indications for Use for EveryWare fall fully within the indications for use of the primary predicate device and the secondary predicate device.

Technological Characteristics Compared to Predicate

Similar to ResMed AirView and Respironics Care Orchestrator, EveryWare is a medical device data management system comprised of a web-based graphical user interface and a component for wireless connection to the compatible therapy devices. Compatible therapy devices for EveryWare only include devices belonging to the MNS, MNT, NOU and CBK product codes, whereas AirView and Care Orchestrator are also compatible with therapy devices that are in BZD product code.

Aside from supporting clinicians to manage data on patients who are prescribed with compatible therapy devices in accordance with the intended use of these therapy devices, both EveryWare, AirView and Care Orchestrator allow settings in the compatible non-life support therapy devices to be changed remotely. EveryWare does not allow ventilation modes to be changed remotely.

The wireless connection of the therapy devices to AirView is through a cellular modem that can either be embedded in the therapy device, or provided by means of a separate connectivity module. The physical component for wireless connection of the therapy devices to Care Orchestrator is embedded within the therapy device itself, and can either be a bluetooth or cellular modem wireless connection. EveryWare uses a cellular modem called iLink that needs to be connected via USB to the therapy device.

The following table presents similarities and differences with the predicates. There are no additional technological features that the EveryWare has which are not already cleared in AirView and/or Care Orchestrator.

Page 2 of 7

Page 7

Page 3 of 7

CharacteristicBreas Medical AB, EveryWare (subject device)ResMed Ltd. AirView (K151901)Respironics, Inc Care Orchestrator (K152356)Discussion of Differences
Intended UseEveryWare is indicated to support clinicians by managing data of patients who are prescribed compatible therapy devices in accordance with the intended use of those therapy devices. EveryWare provides remote patient data collection & viewing and is intended to be used by healthcare representatives in conjunction with compatible non-life support therapy devices to adjust prescription and/or performance settings.AirView is a web based solution for healthcare specialists intended to assist in the diagnosis of sleep disordered breathing in adult patients through analysis of data recorded by an AirView compatible home sleep device. Transfer and display machine and therapeutic information that has been transmitted remotely from the patient's therapy device. It is intended to support the standard follow-up care of patients that have been prescribed a compatible ResMed therapy device. AirView also provides remote settings capabilities for non-life support devices only.Sapphire [Care Orchestrator] is intended to support clinicians by tracking data on patients who are prescribed compatible therapy devices in accordance with the intended use of those therapy devices. Sapphire [Care Orchestrator] provides remote patient data collection & viewing and is intended to be used by healthcare representatives (e.g.,Physicians, Clinicians, Durable Medical Equipment providers) in conjunction with compatible non-life support therapy devices to adjust prescription and/or performance settings.Similar
Environment of UseEveryWare is intended to be used in hospital, institutional provider, and home care settings.Healthcare professional's computer; Clinical users, home care providers (HCPs) and other healthcare specialists.Sapphire [Care Orchestrator] is intended to be used in hospital, institutional provider, and home care settings.Similar

Page 8

Page 4 of 7

CharacteristicBreas Medical AB, EveryWare (subject device)ResMed Ltd. AirView (K151901)Respironics, Inc Care Orchestrator (K152356)Discussion of Differences
User PopulationEveryWare is intended to be used by the following 3 user groups: - Prescribing physicians, nurses, respiratory therapists, physiotherapists and technicians - Durable Medical Equipment (DME)/Homecare providers (HCPs)/Home Medical Equipment providers (HMPs) including managers and administratorsClinical users, home care providers (HCPs) and other healthcare specialistsHealthcare representatives (e.g., Physicians, Clinicians, Durable Medical Equipment providers) and patientsSimilar
Application TypeWeb-based applicationWeb-based applicationWeb-based applicationSame
Product Codes Supported by Remote Prescription Change FunctionalityMNS MNTMNS BZDMNS MNT BZDSimilar to K152356 with respect to compatible ventilators supported.
Product Codes NOT SUPPORTED by Remote Prescription Change FunctionalityNOU CBKCBK MNRNOU CBKSimilar
Data Transfer TechnologyWireless (cellular modem) SD Card File upload/InternetWireless (cellular modem) SD Card File upload/Internet USB Stick/InternetWireless (Bluetooth or cellular modem) SD Card/InternetSimilar

Page 9

Page 5 of 7

CharacteristicBreas Medical AB, EveryWare (subject device)ResMed Ltd. AirView (K151901)Respironics, Inc Care Orchestrator (K152356)Discussion of Differences
FunctionalityCentralized database Patient Management Display therapy data Generate reports Settings management for non-life supporting devicesCentralized database Patient Management Display therapy data Generate reports Settings management for non-life supporting devicesCentralized database Patient Management Display therapy data Generate reports Settings management for non-life supporting devicesSame
Therapy Settings Management (for non-life support devices), including:Performance Settings PrescriptionPerformance Settings Prescription ModePerformance Settings Prescription ModeSimilar. The absence of Therapy Mode Change does not raise new questions of safety and effectiveness.
- Therapy Mode ChangeNoYesYesDifferent The absence of Therapy Mode Change does not raise new questions of safety and effectiveness.
- Target Volume SettingYesYesYesSame
- IPAP & EPAP SettingsYesYesYesSame
- Breath Rate and Insp. Time SettingYesYesYesSame
- Rise Time & Ramp Pressure SettingYesYesYesSame
- Humidifier SettingNoYesYesDifferent The absence of Humidifier Setting does not raise new questions of safety and effectiveness.

Page 10

Page 6 of 7

CharacteristicBreas Medical AB, EveryWare (subject device)ResMed Ltd. AirView (K151901)Respironics, Inc Care Orchestrator (K152356)Discussion of Differences
- High Minute Volume, Disconnection & Apnea alarmYesNoYesSimilar to K152356
- Mask Resistance & Mask Resistance LockNoNoYesSimilar to K151901.
- View Optional ScreenNoNoYesSimilar to K151901.
- Tubing Type & Tubing Type LockNoNoYesSimilar to K151901.
- Automated Airway ManagementNoNoYesSimilar to K151901.
ReportsDetailed report (includes compliance information)Detailed report (includes compliance information)Detailed report (includes compliance information)Same
LabelingOnline help file within the applicationOnline help file within the applicationOnline help file within the applicationSame
Viewing of dataInteractive viewing of data based on user selected date rangeInteractive viewing of data based on user selected date rangeInteractive viewing of data based on user selected date rangeSame

Non-clinical Testing

EveryWare was designed and subjected to performance testing in accordance with all applicable requirements specifications, guidance documents and recognized consensus standards:

  • Content of Premarket Submissions for Device Software Functions. Guidance for Industry and Food and Drug Administration Staff (June 14, 2023)
  • Off-the-Shelf Software Use in Medical Devices. Guidance for Industry and Food and Drug Administration Staff (August 11, 2023)
  • Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submission. Guidance for Industry and Food and Drug Administration Staff (September 27, 2023)

Page 11

Page 7 of 7

  • Applying Human Factors and Usability Engineering to Medical Devices. Guidance for Industry and Food and Drug Administration Staff (February 3, 2016)
  • IEC 60601-1 Edition 3.2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
  • IEC 60601-1-2 Edition 4.1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
  • IEC 62304 Edition 1.1 2015 Medical device software - Software life cycle processes

EveryWare, by itself, is not intended for diagnosis and treatment of disease or medical condition. EveryWare does not change the performance specification of the connected therapeutic devices. Therefore, no further performance data is required to demonstrate substantial equivalence.

Conclusion

EveryWare is substantially equivalent to the AirView and Care Orchestrator (Sapphire), as the devices share a common intended use and technological characteristics as demonstrated through performance testing.

§ 868.5895 Continuous ventilator.

(a)
Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.(b)
Classification. Class II (performance standards).