K Number
K242437

Validate with FDA (Live)

Device Name
Smile Dx®
Manufacturer
Date Cleared
2025-05-14

(271 days)

Product Code
Regulation Number
892.2070
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Smile Dx® is a computer-assisted detection (CADe) software designed to aid dentists in the review of digital files of bitewing and periapical radiographs of permanent teeth. It is intended to aid in the detection and segmentation of suspected dental findings which include: caries, periapical radiolucencies (PARL), restorations, and dental anatomy.

Smile Dx® is also intended to aid dentists in the measurement (in millimeter and percentage measurements) of mesial and distal bone levels associated with each tooth.

The device is not intended as a replacement for a complete dentist's review or their clinical judgment that takes into account other relevant information from the image, patient history, and actual in vivo clinical assessment.

Smile Dx® supports both digital and phosphor sensors.

Device Description

Smile Dx® is a computer assisted detection (CADe) device indicated for use by licensed dentists as an aid in their assessment of bitewing and periapical radiographs of secondary dentition in adult patients. Smile Dx® utilizes machine learning to produce annotations for the following findings:

  • Caries
  • Periapical radiolucencies
  • Bone level measurements (mesial and distal)
  • Normal anatomy (enamel, dentin, pulp, and bone)
  • Restorations
AI/ML Overview

The provided FDA 510(k) Clearance Letter for Smile Dx® outlines the device's acceptance criteria and the studies conducted to prove it meets those criteria.

Acceptance Criteria and Device Performance

The acceptance criteria are implicitly defined by the performance metrics reported in the "Performance Testing" section. The device's performance is reported in terms of various metrics for both standalone and human-in-the-loop (MRMC) evaluations.

Here's a table summarizing the reported device performance against the implied acceptance criteria:

Table 1: Acceptance Criteria and Reported Device Performance

Feature/MetricAcceptance Criteria (Implied)Reported Device Performance
Standalone Testing:
Caries DetectionHigh Dice, SensitivityDice: 0.74 [0.72 0.76] Sensitivity (overall): 88.3% [83.5%, 92.6%]
Periapical Radiolucency (PARL) DetectionHigh Dice, SensitivityDice: 0.77 [0.74, 0.80] Sensitivity: 86.1% [80.2%, 91.9%]
Bone Level Detection (Bitewing)High Sensitivity, Specificity, Low MAESensitivity: 95.5% [94.3%, 96.7%] Specificity: 94.0% [91.1%, 96.6%] MAE: 0.30 mm [0.29mm, 0.32mm]
Bone Level Detection (Periapical)High Sensitivity, Specificity, Low MAE (percentage)Sensitivity: 87.3% [85.4%, 89.2%] Specificity: 92.1% [89.9%, 94.1%] MAE: 2.6% [2.4%, 2.8%]
Normal Anatomy DetectionHigh Dice, Sensitivity, SpecificityDice: 0.84 [0.83, 0.85] Sensitivity (Pixel-level): 86.1% [85.4%, 86.8%] Sensitivity (Contour-level): 95.2% [94.5%, 96%] Specificity (Contour-level): 93.5% [91.6%, 95.8%]
Restorations DetectionHigh Dice, Sensitivity, SpecificityDice: 0.87 [0.85, 0.90] Sensitivity (Pixel-level): 83.1% [80.3%, 86.4%] Sensitivity (Contour-level): 90.9% [88.2%, 93.9%] Specificity (Contour-level): 99.6% [99.3%, 99.8%]
MRMC Clinical Evaluation - Reader Improvement:
Caries Detection (wAFROC Δθ)Statistically significant improvement+0.127 [0.081, 0.172] (p < 0.001)
PARL Detection (wAFROC Δθ)Statistically significant improvement+0.098 [0.061, 0.135] (p < 0.001)
Caries Detection (Sensitivity Improvement)Increased sensitivity with device assistance19.6% [12.8%, 26.4%] increase (from 64.3% to 83.9%)
PARL Detection (Sensitivity Improvement)Increased sensitivity with device assistance19.1% [13.6%, 24.7%] increase (from 70.7% to 89.8%)
Caries Detection (Specificity Improvement)Maintained or improved specificity with device assistance16.7% [13.5%, 19.9%] increase (from 73.6% to 90.2%)
PARL Detection (Specificity Improvement)Maintained or improved specificity with device assistance4.7% [3%, 6.4%] increase (from 92.6% to 97.3%)

Study Details for Device Performance Proof:

1. Sample Sizes for the Test Set and Data Provenance

  • Standalone Testing:
    • Caries and Periapical Radiolucency Detection: 867 cases.
    • Bone Level Detection and Bone Loss Measurement: 352 cases.
    • Normal Anatomy and Restorations: 200 cases.
  • MRMC Clinical Evaluation: 352 cases.
  • Data Provenance: All test sets were collected from "multiple U.S. sites." The data is retrospective, as it's used in a "retrospective study" for the MRMC evaluation. Sub-group analysis also included "imaging hardware" and "patient demographics (i.e., age, sex, race)," indicating diversity in data.

2. Number of Experts and Qualifications for Ground Truth

  • Standalone Testing (Implicit): Not explicitly stated how ground truth for standalone testing was established, but it is likely derived from expert consensus, similar to the MRMC study.
  • MRMC Clinical Evaluation: Ground truth was established by the "consensus labels of four US licensed dentists."

3. Adjudication Method for the Test Set

  • MRMC Clinical Evaluation: The ground truth for the MRMC study was established by the "consensus labels of four US licensed dentists." This implies a form of consensus adjudication, likely where all four experts reviewed and reached agreement on the findings. The specific method (e.g., majority vote, 2+1, 3+1) is not explicitly detailed beyond "consensus labels." For standalone testing, the adjudication method for ground truth is not specified.

4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Yes, an MRMC comparative effectiveness study was done.
  • Effect Size of Human Readers' Improvement with AI vs. Without AI Assistance:
    • Caries Detection:
      • wAFROC Δθ: +0.127 [0.081, 0.172] (p < 0.001)
      • Sensitivity Improvement: 19.6% increase (from 64.3% without device to 83.9% with device).
      • Specificity Improvement: 16.7% increase (from 73.6% without device to 90.2% with device).
    • Periapical Radiolucency (PARL) Detection:
      • wAFROC Δθ: +0.098 [0.061, 0.135] (p < 0.001)
      • Sensitivity Improvement: 19.1% increase (from 70.7% without device to 89.8% with device).
      • Specificity Improvement: 4.7% increase (from 92.6% without device to 97.3% with device).
  • The study design was a "fully-crossed, multiple-reader multiple-case (MRMC) evaluation method" with "at least 13 US licensed dentists (Smile Dx® had 14 readers)." Half of the data set contained unannotated images, and the second half contained radiographs that had been processed through the CADe device. Radiographs were presented to readers in alternating groups throughout two different sessions, separated by a washout period.

5. Standalone Performance Study (Algorithm Only)

  • Yes, a standalone performance study was done.
  • It evaluated the algorithm's performance for:
    • Caries and Periapical Radiolucency Detection (Dice, Sensitivity)
    • Bone Level Detection and Bone Loss Measurement (Sensitivity, Specificity, Mean Absolute Error)
    • Normal Anatomy and Restorations Detection (Dice, Pixel-level Sensitivity, Contour-level Sensitivity, Contour-level Specificity)

6. Type of Ground Truth Used

  • Explicitly for MRMC Study: "Consensus labels of four US licensed dentists." This indicates expert consensus was used for ground truth for the human-in-the-loop evaluation and likely for the standalone evaluation's ground truth as well. There is no mention of pathology or outcomes data.

7. Sample Size for the Training Set

  • The document does not explicitly state the sample size for the training set. It only mentions the test set sizes.

8. How Ground Truth for Training Set Was Established

  • The document does not explicitly state how ground truth for the training set was established. It mentions the model "utilizes machine learning to produce annotations" and "training data" is used, but provides no details on its annotation process. It's highly probable that expert annotation was also used for the training data, similar to the test set, but this is not confirmed in the provided text.

FDA 510(k) Clearance Letter - Smile Dx®

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

May 14, 2025

Cube Click, Inc.
℅ Meritxell Martinez
Project Manager and Regulatory Specialist
Innolitics
1101 West 34th St #550
AUSTIN, TX 78705

Re: K242437
Trade/Device Name: Smile Dx®
Regulation Number: 21 CFR 892.2070
Regulation Name: Medical Image Analyzer
Regulatory Class: Class II
Product Code: MYN, LLZ
Dated: April 6, 2025
Received: April 7, 2025

Dear Meritxell Martinez:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K242437 - Meritxell Martinez Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K242437 - Meritxell Martinez Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K242437

Device Name: Smile Dx®

Indications for Use (Describe)

Smile Dx® is a computer-assisted detection (CADe) software designed to aid dentists in the review of digital files of bitewing and periapical radiographs of permanent teeth. It is intended to aid in the detection and segmentation of suspected dental findings which include: caries, periapical radiolucencies (PARL), restorations, and dental anatomy.

Smile Dx® is also intended to aid dentists in the measurement (in millimeter and percentage measurements) of mesial and distal bone levels associated with each tooth.

The device is not intended as a replacement for a complete dentist's review or their clinical judgment that takes into account other relevant information from the image, patient history, and actual in vivo clinical assessment.

Smile Dx® supports both digital and phosphor sensors.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary

1. CONTACT INFORMATION

FieldInformation
Company NameCube Click, Inc.
Address105 East 34 Street, Suite 231, New York, NY 10016
Phone Number+1.646.224.5908
Company RepresentativeRichard Ricci, D.D.S, M.S., F.A.G.D, President, Cube Click
Primary Correspondent(s)Meritxell Martinez, Regulatory Specialist, InnoliticsYujan Shrestha, MD, Partner and Co-Founder, Innolitics
Primary Correspondent Emailfda@innolitics.com
Date PreparedMay 13th, 2025
Documentation Prepared UsingInnolitics Medtech OS Framework

2. DEVICE INFORMATION

FieldInformation
Trade NameSmile Dx®
Common NameComputer-Assisted Detection Device
Classification NameMedical Image Analyzer
Primary Product CodeMYN
Primary Regulation Number21 CFR 892.2070
Secondary Product CodeLLZ

3. PRIMARY PREDICATE DEVICE INFORMATION

FieldInformation
Predicate Device NameOverjet Caries Assist
Predicate Device K NumberK212519

4. REFERENCE DEVICE(S) INFORMATION

FieldInformation
Predicate Device NameSecond Opinion

Page 6

FieldInformation
Predicate Device K NumberK210365
Predicate Device NameOverjet Dental Assist
Predicate Device K NumberK210187

5. DEVICE DESCRIPTION

Smile Dx® is a computer assisted detection (CADe) device indicated for use by licensed dentists as an aid in their assessment of bitewing and periapical radiographs of secondary dentition in adult patients. Smile Dx® utilizes machine learning to produce annotations for the following findings:

  • Caries
  • Periapical radiolucencies
  • Bone level measurements (mesial and distal)
  • Normal anatomy (enamel, dentin, pulp, and bone)
  • Restorations

6. INDICATIONS FOR USE

Smile Dx® is a computer-assisted detection (CADe) software designed to aid dentists in the review of digital files of bitewing and periapical radiographs of permanent teeth. It is intended to aid in the detection and segmentation of suspected dental findings which include: caries, periapical radiolucencies (PARL), restorations, and dental anatomy.

Smile Dx® is also intended to aid dentists in the measurement (in millimeter and percentage measurements) of mesial and distal bone levels associated with each tooth.

The device is not intended as a replacement for a complete dentist's review or their clinical judgment that takes into account other relevant information from the image, patient history, and actual in vivo clinical assessment.

Smile Dx® supports both digital and phosphor sensors.

7. INTENDED PATIENT POPULATION

The intended patient population of Smile Dx® are patients that have permanent adult dentition and are at least 22 years of age.

Page 7

8. SUBSTANTIAL EQUIVALENCE SUMMARY

Smile Dx® is similar to the predicate device and reference devices in the following ways:

  • Intended use: All devices are intended to be used to identify suspected areas of interest in intra-oral radiographs.
  • Technology characteristics: All devices employ machine learning algorithms to output detections.
  • Safety: As all devices are CADe and/or SaMD systems, and don't pose any direct safety hazard to the patient.
  • Clinical Performance: All devices have undergone clinical evaluations which demonstrate statistically significant improvement in aided reader performance.

8.1. Device Comparison Table

For technological characteristics comparison, the following table is provided:

CharacteristicSmile Dx®Predicate Device: Overjet Caries Assist (K212519)Reference Device: Second Opinion (K210365)Reference Device: Overjet Dental Assist (K210187)Comparison
ManufacturerCube Click, Inc.Overjet, Inc.Pearl Inc.Overjet, Inc.N/A
Regulation Number892.2070892.2070892.2070892.2050Smile Dx® falls under the same regulation (892.2070) as primary predicate (Overjet Caries Assist).
Regulation NameMedical image analyzerMedical image analyzerMedical image analyzerPicture archiving and communications systemSmile Dx® falls under the same regulation (892.2070) as predicate (Overjet Caries Assist).
Regulatory ClassClass IIClass IIClass IIClass IISame
Product CodeMYN; LLZMYNMYNLLZSmile Dx® has the same primary product MYN as predicate (Overjet Caries Assist) and reference device

Page 8

CharacteristicSmile Dx®Predicate Device: Overjet Caries Assist (K212519)Reference Device: Second Opinion (K210365)Reference Device: Overjet Dental Assist (K210187)Comparison
(Second Opinion). Additionally, Smile Dx®'s secondary product code LLZ is the same as reference device (Overjet Dental Assist). Thus, the subject device has the same product codes as the identified predicate and reference devices.
Device PropertySaMD/CADeSaMD/CADeSaMD/CADeSaMDSame
Indications for UseSmile Dx® is a computer-assisted detection (CADe) software designed to aid dentists in the review of digital files of bitewing and periapical radiographs of permanent teeth. It is intended to aid in the detection and segmentation of suspected dental findings which include: caries, periapical radiolucencies (PARL), restorations, and dental anatomy.Smile Dx® is also intended to aid dentists in the measurement (in millimeter and percentage measurements) of mesial and distal bone levels associated with each tooth.The Overjet Caries Assist (OCA) is a radiological, automated, concurrent read, computer-assisted detection software intended to aid in the detection and segmentation of caries on bitewing radiographs. The device provides additional information for the dentist to use in their diagnosis of a tooth surface suspected of being carious. The device is not intended as a replacement for a complete dentist's review or their clinical judgment that takes into account other relevantSecond Opinion® is a computer aided detection ("CADe") software to identify and mark regions in relation to suspected dental findings which include Caries, Discrepancy at the margin of an existing restoration, Calculus, Periapical radiolucency, Crown (metal, including zirconia & non-metal), Filling (metal & non-metal), Root canal, Bridge and Implants.It is designed to aid dental health professionals to review bitewing and periapical radiographs of permanent teeth in patients 12 yearsOverjet Dental Assist is a radiological semi-automated image processing software device intended to aid dental professionals in the measurements of mesial and distal bone levels associated with each tooth from bitewing and periapical radiographs.It should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. The system is to be used by trained professionals including, but not limited to, dentists and dental hygienists.Intended Patient Population:Smile Dx® has similar indications for use to predicate K212519 for detecting caries, reference device K210365 for detecting PARLs, and reference device K210187 for measurement of bone levels. Thus, the subject device combines the indications for use of the three identified predicate and reference devices. All three devices share the same intended use to identify suspected areas of interest in intra-oral radiographs.

Page 9

CharacteristicSmile Dx®Predicate Device: Overjet Caries Assist (K212519)Reference Device: Second Opinion (K210365)Reference Device: Overjet Dental Assist (K210187)Comparison
each tooth.The device is not intended as a replacement for a complete dentist's review or their clinical judgment that takes into account other relevant information from the image, patient history, and actual in vivo clinical assessment.Smile Dx® supports both digital and phosphor sensors.information from the image, patient history, and actual in vivo clinical assessment.of age or older as a second reader.The intended patient population of the device is patients living in the United States, who are 22 years old or older, and that do not have any remaining primary teeth. Overjet has not evaluated the performance of the device on primary dentition.(Second Opinion). Additionally, Smile Dx®'s secondary product code LLZ is the same as reference device (Overjet Dental Assist). Thus, the subject device has the same product codes as the identified predicate and reference devices.
Device PropertySaMD/CADeSaMD/CADeSaMD/CADeSaMDSame
Indications for UseSmile Dx® is a computer-assisted detection (CADe) software designed to aid dentists in the review of digital files of bitewing and periapical radiographs of permanent teeth. It is intended to aid in the detection and segmentation of suspected dental findings which include: caries, periapical radiolucencies (PARL), restorations, and dental anatomy.Smile Dx® is also intended to aid dentists in the measurement (in millimeter and percentage measurements) of mesial and distal bone levels associated withThe Overjet Caries Assist (OCA) is a radiological, automated, concurrent read, computer-assisted detection software intended to aid in the detection and segmentation of caries on bitewing radiographs. The device provides additional information for the dentist to use in their diagnosis of a tooth surface suspected of being carious. The device is not intended as a replacement for a complete dentist's review or their clinical judgment that takes into account other relevantSecond Opinion® is a computer aided detection ("CADe") software to identify and mark regions in relation to suspected dental findings which include Caries, Discrepancy at the margin of an existing restoration, Calculus, Periapical radiolucency, Crown (metal, including zirconia & non-metal), Filling (metal & non-metal), Root canal, Bridge and Implants.It is designed to aid dental health professionals to review bitewing and periapical radiographs of permanent teeth in patients 12 yearsOverjet Dental Assist is a radiological semi-automated image processing software device intended to aid dental professionals in the measurements of mesial and distal bone levels associated with each tooth from bitewing and periapical radiographs.It should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. The system is to be used by trained professionals including, but not limited to, dentists and dental hygienists.Intended Patient Population:Smile Dx® has similar indications for use to predicate K212519 for detecting caries, reference device K210365 for detecting PARLs, and reference device K210187 for measurement of bone levels. Thus, the subject device combines the indications for use of the three identified predicate and reference devices. All three devices share the same intended use to identify suspected areas of interest in intra-oral radiographs.

Page 10

CharacteristicSmile Dx®Predicate Device: Overjet Caries Assist (K212519)Reference Device: Second Opinion (K210365)Reference Device: Overjet Dental Assist (K210187)Comparison
Detections and/or measurements• Normal anatomy• Caries• Periapical radiolucencies• Bone level measurements (mesial and distal)• RestorationsCariesCaries, Margin discrepancy, Calculus, Periapical radiolucencies, RestorationsBone level measurements (mesial and distal)Smile Dx® detects caries similar to predicate Overjet Caries Assist K212519.Smile Dx® also detects PARLs and restorations, similar to reference device Second Opinion K210365. Second Opinion also detects other features such as margin discrepancy and calculus, however, these are not supported by subject device. The absence of these features in Smile Dx® does not create a new intended use.Smile Dx® also measures mesial and distal bone levels, same as reference device Overjet

Page 11

CharacteristicSmile Dx®Predicate Device: Overjet Caries Assist (K212519)Reference Device: Second Opinion (K210365)Reference Device: Overjet Dental Assist (K210187)Comparison
each tooth.The device is not intended as a replacement for a complete dentist's review or their clinical judgment that takes into account other relevant information from the image, patient history, and actual in vivo clinical assessment.Smile Dx® supports both digital and phosphor sensors.information from the image, patient history, and actual in vivo clinical assessment.of age or older as a second reader.The intended patient population of the device is patients living in the United States, who are 22 years old or older, and that do not have any remaining primary teeth. Overjet has not evaluated the performance of the device on primary dentition.The intended patient population of the device is patients living in the United States, who are 22 years old or older, and that do not have any remaining primary teeth. Overjet has not evaluated the performance of the device on primary dentition.Dental Assist K210187.
Intended UserDentistsDentistsDentistsDental professionals including, but not limited to, dentists and dental hygienists.Smile Dx® is intended for use by licensed dentists, same as all of the identified predicate and reference devices. The reference device Overjet Dental Assist is also intended for use by other dental professionals, such as dental hygienists. However, use of Smile Dx® has been limited to use by licensed dentists only.
Patient PopulationPatients requiring dental services, all sexes, at least 22 years of age, and with permanent dentition.Patients requiring dental services, all sexes, at least 18 years of age, and with permanent dentition.Patients requiring dental services, all sexes, at least 12 years of age, and with permanent dentition.Patients requiring dental services, all sexes, at least 22 years of age, and with permanent dentition.The intended patient population for Smile Dx® is limited to patients with permanent dentition who are at least 22 years of age. This is identical to the reference device Second Opinion. This intended patient population age has been supported by use of representative patient data in testing.
Rx or OTCRx-onlyRx-onlyRx-onlyRx-onlySame

Technical Characteristics

CharacteristicSmile Dx®Predicate Device: Overjet Caries Assist (K212519)Reference Device: Second Opinion (K210365)Reference Device: Overjet Dental Assist (K210187)Comparison
PlatformWeb - Edge, Chrome, FireFox, SafariWeb - Edge, Chrome, FirefoxLocal computer application (Cloud-based installation)Web - Edge, Chrome, FirefoxThe subject device is a web-based application, similar to Overjet Caries Assist and Overjet

Page 12

CharacteristicSmile Dx®Predicate Device: Overjet Caries Assist (K212519)Reference Device: Second Opinion (K210365)Reference Device: Overjet Dental Assist (K210187)Comparison
Dental Assist K210187.
Intended UserDentistsDentistsDentistsDental professionals including, but not limited to, dentists and dental hygienists.Smile Dx® is intended for use by licensed dentists, same as all of the identified predicate and reference devices. The reference device Overjet Dental Assist is also intended for use by other dental professionals, such as dental hygienists. However, use of Smile Dx® has been limited to use by licensed dentists only.
Patient PopulationPatients requiring dental services, all sexes, at least 22 years of age, and with permanent dentition.Patients requiring dental services, all sexes, at least 18 years of age, and with permanent dentition.Patients requiring dental services, all sexes, at least 12 years of age, and with permanent dentition.Patients requiring dental services, all sexes, at least 22 years of age, and with permanent dentition.The intended patient population for Smile Dx® is limited to patients with permanent dentition who are at least 22 years of age. This is identical to the reference device Second Opinion. This intended patient population age has been supported by use of representative patient data in testing.
Rx or OTCRx-onlyRx-onlyRx-onlyRx-onlySame

Technical Characteristics

CharacteristicSmile Dx®Predicate Device: Overjet Caries Assist (K212519)Reference Device: Second Opinion (K210365)Reference Device: Overjet Dental Assist (K210187)Comparison
PlatformWeb - Edge, Chrome, FireFox, SafariWeb - Edge, Chrome, FirefoxLocal computer application (Cloud-based installation)Web - Edge, Chrome, FirefoxThe subject device is a web-based application, similar to Overjet Caries Assist and Overjet Dental Assist. It is compatible with similar browsers and is also compatible with Safari, which is a commonly used browser.

Page 13

CharacteristicSmile Dx®Predicate Device: Overjet Caries Assist (K212519)Reference Device: Second Opinion (K210365)Reference Device: Overjet Dental Assist (K210187)Comparison
OSAnyAnyWindows 7 or higherAnyThe subject device is compatible with any operating system, same as predicate Overjet Caries Assist and reference device Overjet Dental Assist.
User InterfaceMouse, Keyboard, TrackpadMouse, Keyboard, TrackpadMouse, KeyboardMouse, Keyboard, TrackpadSame
Image ModalityRadiographRadiographRadiographRadiographSame
Radiograph TypeBitewing and periapicalBitewing and periapicalBitewing and periapicalBitewing and periapicalSame
Image Input SourcesImages imported from local computer, from a range of manufacturersImages imported from the radiographic device, or from the practice management system, from Carestream or Schick sensorsRadiography devices from a range of manufacturersImages imported from the radiographic device, or from the practice management systemImages are manually imported from the dentist user's local computer into Smile Dx®. The subject device does not connect to practice management systems like Overjet Caries Assist and Overjet Dental Assist. However, this difference does not raise any new questions of safety and effectiveness.Additionally, Smile Dx® is compatible with images from a range of radiography device

Page 14

CharacteristicSmile Dx®Predicate Device: Overjet Caries Assist (K212519)Reference Device: Second Opinion (K210365)Reference Device: Overjet Dental Assist (K210187)Comparison
manufacturers, similar to Second Opinion. This has been supported by using data from a range of manufacturers during testing.
Image Formatjpg, png, tif, tig, dex, gif, bitmap/bmp, and rawjpg, png, eop, jif, dicomjpeg, png, rvg, dcm, tiff, and dicjpg, png, jfif, eop, etp, jifSmile Dx® supports image formats that are commonly used in dental radiographs. The difference in supported image formats between the subject and predicate devices is considered to be minor and does not interfere with the ability of Smile Dx® to achieve its intended use.
Processing ArchitectureThree layers:- The Network layer contains a login module and a data ingestion module. Radiographs are manually uploaded from local computer to the server.- The Detection and Measurement layer processes the image and classifies it as either bitewing or periapical, and then annotates it via the algorithm- The visualization layer displays the annotated image via Smile Dx®'sThree layers:- The Network layer works with the practice PACS or EMR to transmit the image and meta-data to Overjet.- The decision layer processes the image to ensure it is the correct data type, and then annotates it via the algorithm- The presentation layer displays the annotated image in a non diagnostic viewer. The dentist can filter,The Second Opinion Client continuously monitors a local or networked resource for dental radiographs. When new images are found, it sends them to cloud-based Computer Vision Models (CV Models) for analysis. These models generate metadata describing the nature and location of detected features in the radiographs. The metadata is returned to the Second Opinion® Client, whichUnknownSmile Dx® has a similar processing architecture to the predicate Overjet Caries Assist (K212519) and reference device Second Opinion (K210365). All devices obtain radiographs, annotate them, and present annotations to via a user interface.One of the main differences in processing architecture is that the predicate device can establish connections with external sources to upload images, whereas the subject

Page 15

CharacteristicSmile Dx®Predicate Device: Overjet Caries Assist (K212519)Reference Device: Second Opinion (K210365)Reference Device: Overjet Dental Assist (K210187)Comparison
user interface. The dentist can adjust image settings, display and hide the presented annotations.display, hide, create and edit the annotations presented.displays it within its user interface. Detected features are highlighted with color-coded boundary boxes overlaid on the original radiograph.device exclusively uploads images from the local computer. Similarly, users of Smile Dx® are required to manually upload images to the software, whereas the reference device Second Opinion automatically uploads images. Nevertheless, these technological differences do not raise new questions of safety and effectiveness. The absence of these features in the subject device does not impede its ability to fulfill its intended use.
AlgorithmUtilizes computer vision machine learning algorithm(s).Utilizes computer vision machine learning algorithm(s).Utilizes computer vision neural network algorithms, developed from open-source models using supervised machine learning techniques.Utilizes computer vision techniques.Same. Smile Dx® utilizes computer vision techniques, specifically, machine learning algorithms to detect suspected findings, same as the identified predicate and reference devices.
OutputSuspected findings (i.e., caries, PARLs) and measurements (bone levels) are overlaid on the original radiographCaries detection and segmentation on radiograph resulting in outline of suspected cariesDetected features (e.g., caries, PARLs) are highlighted with color-coded boundary boxes overlaid on the original radiographBone-level annotations on radiographThe output of Smile Dx® is similar to the outputs of all of the identified predicate and reference devices, as they all overlay the findings on the original radiograph.
Marker Type/SizeContour segmentations for caries and periapicalPresents suspected carious lesions as segmentedColor-coded bounding boxes / FixedColor-coded lines for bone levelsThe subject device has similar marker types to its predicate and reference

Page 16

CharacteristicSmile Dx®Predicate Device: Overjet Caries Assist (K212519)Reference Device: Second Opinion (K210365)Reference Device: Overjet Dental Assist (K210187)Comparison
radiolucencies. Color-coded lines for bone levelspolygons outlining the predictiondevices.For suspected caries, the subject device creates a contour of the suspected caries, similar to the predicate (Overjet Caries Assist).For periapical radiolucencies, the subject device creates contours. This is different to reference device's (Second Opinion) bounding boxes. The performance and accuracy of the segmentations has been supported with appropriate performance testing.For bone levels, the subject device outputs color-coded lines depending on the level of measured boneloss, similar to reference device Overjet Dental Assist.
Image MeasurementLinear distanceN/AN/ALinear distanceSmile Dx® is similar to reference device Overjet Dental Assist in that it creates a linear distance for measured bone levels.
Image ViewingFull, ThumbnailFull, ThumbnailFull, ThumbnailFull, ThumbnailSame
Image ManipulationImage adjustment tools (e.g., brightness, contrast, opacity, rotation prior to upload)AnnotationsImage adjustment tools (e.g., brightness, contrast, zoom, invert, rotate)Annotation (line)Smile Dx® has similar image adjustment tools as reference device Second Opinion

Page 17

CharacteristicSmile Dx®Predicate Device: Overjet Caries Assist (K212519)Reference Device: Second Opinion (K210365)Reference Device: Overjet Dental Assist (K210187)Comparison
(e.g., contrast and brightness). Smile Dx® introduces a new tool (i.e., opacity adjustment), however, this does not raise any new questions of safety and effectiveness and is a common image adjustment tool. Smile Dx® lacks some of the features found in the predicate and reference devices (e.g., zoom, invert, and manual annotation tools). However, these tools are not needed to accomplish the intended use of the device and do not raise new questions of safety and effectiveness.
Performance Testing• Standalone performance study for pathologic feature (i.e., caries and periapical radiolucency) detection, non-pathologic feature detection (i.e., normal anatomy and restorations) and bone level measurement performance.• Multiple-Reader, Multiple-Case (MRMC) study for pathologic dental features• Analysis included:• wAFROC analysis for primary endpoint• Sensitivity and specificity evaluations• Standalone study for pathologic feature detection performance.• Multiple-Reader, Multiple-Case (MRMC) study for pathologic dental features• Analysis included:• Sensitivity and specificity evaluations• Dice coefficient analysis• Standalone study for pathological and non-pathologic feature detection performance.• Multiple-Reader, Multiple-Case (MRMC) study for pathologic dental features• Analysis included:• wAFROC-FOM analysis for primary endpoints• Determination of the changes in sensitivity and change in number of false positive dental pathologies of a given type per image (FPPI)• Bench testing evaluated precision and recall against labeled keypoints within radiographs.• Retrospective clinical performance testing that compared adjudicated measurements against Overjet Dental Assist's predicted measurements. Analysis included sensitivity and specificity.Smile Dx® underwent similar evaluations as predicate and reference devices, following recommendations in 2022 FDA Guidance "Computer-Assisted Detection Devices Applied to Radiology Images and Radiology Device Data - Premarket Notification [510(k)] Submissions".

Page 18

9. PERFORMANCE TESTING

9.1. Software Verification and Validation

As Smile Dx® is a SaMD, software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Device Software Functions".

9.2. CADe Testing

Smile Dx® was subjected to clinical performance testing according to 2022 FDA Guidance "Computer-Assisted Detection Devices Applied to Radiology Images and Radiology Device Data - Premarket Notification [510(k)] Submissions" Testing involved both standalone performance testing and a fully-crossed multi-reader multi-case (MRMC) reader clinical evaluation. As CADe devices, predicate device Overjet Caries Assist (K212519) and Second Opinion (K210365) also underwent the recommended testing.

9.2.1. Standalone Testing

Caries and Periapical Radiolucency Detection

Smile Dx®'s caries and periapical radiolucency detection performance was evaluated using a test dataset of 867 cases collected from multiple U.S. sites. The test results were as follows:

MetricCariesPeriapical Radiolucency
Dice0.74 [0.72 0.76]0.77 [0.74, 0.80]
Sensitivityoverall: 88.3% [83.5%, 92.6%]86.1% [80.2%, 91.9%]

Bone Level Detection and Bone Loss Measurement

Page 19

Smile Dx®'s bone level detection and bone loss measurement performance was evaluated using a test dataset of 352 cases collected from multiple U.S. sites. The test results were as follows:

Radiograph TypeBone Level Detection - SensitivityBone Level Detection - SpecificityBone Loss Measurement (Mean Absolute Error [95% CI]
Bitewing95.5% [94.3%, 96.7%]94.0% [91.1%, 96.6%]0.30 mm [0.29mm, 0.32mm]
Periapical87.3% [85.4%, 89.2%]92.1% [89.9%, 94.1%]2.6% [2.4%, 2.8%]

Normal Anatomy and Restorations

Smile Dx®'s performance for normal anatomy and restoration detection was evaluated using a test dataset of 200 cases collected from different U.S. sites. The test results were as follows:

Non-pathologic RegionDiceSensitivity (Pixel-level)Sensitivity (Contour-level)Specificity (Contour-level)
Normal Anatomy0.84 [0.83, 0.85]86.1% [85.4%, 86.8%]95.2% [94.5%, 96%]93.5% [91.6%, 95.8%]
Restorations0.87 [0.85, 0.90]83.1% [80.3%, 86.4%]90.9% [88.2%, 93.9%]99.6% [99.3%, 99.8%]

9.2.2. MRMC Clinical Evaluation - Reader Improvement

The clinical performance assessment for Smile Dx® used a fully-crossed, multiple-reader multiple-case (MRMC) evaluation method. The test dataset consisted of 352 cases collected from multiple U.S. sites. Smile Dx® underwent clinical evaluation using methods similar to those used for the primary predicate device Overjet Caries Assist (K212519):

Both devices were evaluated in a multi-reader, multi-case (MRMC) retrospective study with at least 13 US licensed dentists (Smile Dx® had 14 readers). Ground truth was established by the consensus labels of at least three US licensed dentists (the ground truth for Smile Dx®'s study was established by four US licensed dentists). Half of the data set contained unannotated images, and the second half contained radiographs that had been processed through the CADe device. The radiographs were presented to the readers in alternating groups throughout two different sessions, which were separated by a washout period. The results were compared against the consensus ground truth, and the sensitivity, specificity, and alternative free response receiver operating characteristic (AFROC) was evaluated to characterize the performance of the readers with and without viewing the model annotations.

Primary Endpoint: wAFROC Analysis

The weighted alternative free-response receiver operating characteristic (wAFROC) figure of merit serves as the primary performance metric, giving equal importance to each case while reducing the influence of cases with multiple lesions.

Page 20

The wAFROC analysis revealed statistically significant improvements, with an increase of Δθ for both caries detection (+0.127 [0.081, 0.172]) and PARL detection (+0.098 [0.061, 0.135]). These results demonstrate that Smile Dx® enhances reader sensitivity without compromising specificity, thereby meeting the acceptance criteria for effective detection of caries and PARLs.

θcontrolθCADΔθ [95% CI]p-value
Caries0.7810.908+0.127 [0.081, 0.172]< 0.001
PARLs0.8490.947+0.098 [0.061, 0.135]< 0.001

Secondary Endpoint: Sensitivity and Specificity

Sensitivity metrics were determined at the lesion level. Specificity metrics were determined at the case level.

Sensitivity

For caries detections, reader sensitivity improved from 64.3% without device to 83.9% with device, an increase of 19.6% [12.8%, 26.4%]. For PARL detections, reader sensitivity improved from 70.7% without device to 89.8% with device, an increase of 19.1% [13.6%, 24.7%].

Specificity

For caries detections, reader specificity improved from 73.6% without device to 90.2% with device, a difference of 16.7% [13.5%, 19.9%]. For PARL detections, reader specificity improved from 92.6% without device to 97.3% with device, a difference of 4.7% [3%, 6.4%].

Sub-group Analysis

Sub-group analyses were performed for patient demographics (i.e., age, sex, race), imaging hardware, and primary vs. secondary classification (caries). No distinct differences in performance were observed among sub-groups that raise concerns about the effectiveness of Smile Dx®.

10. CONCLUSION

Smile Dx® is substantially equivalent to the identified predicate device. Specifically, the subject device has similar indications and technological characteristics when compared to the predicate device. Any differences in technological characteristics do not raise new questions of safety or effectiveness.

Additionally, the results of the performance testing (i.e., software verification and validation, standalone performance assessment and MRMC clinical evaluation) of Smile Dx® support that the subject device's performance is as effective as the predicate and reference devices.

§ 892.2070 Medical image analyzer.

(a)
Identification. Medical image analyzers, including computer-assisted/aided detection (CADe) devices for mammography breast cancer, ultrasound breast lesions, radiograph lung nodules, and radiograph dental caries detection, is a prescription device that is intended to identify, mark, highlight, or in any other manner direct the clinicians' attention to portions of a radiology image that may reveal abnormalities during interpretation of patient radiology images by the clinicians. This device incorporates pattern recognition and data analysis capabilities and operates on previously acquired medical images. This device is not intended to replace the review by a qualified radiologist, and is not intended to be used for triage, or to recommend diagnosis.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithms including a description of the algorithm inputs and outputs, each major component or block, and algorithm limitations.
(ii) A detailed description of pre-specified performance testing methods and dataset(s) used to assess whether the device will improve reader performance as intended and to characterize the standalone device performance. Performance testing includes one or more standalone tests, side-by-side comparisons, or a reader study, as applicable.
(iii) Results from performance testing that demonstrate that the device improves reader performance in the intended use population when used in accordance with the instructions for use. The performance assessment must be based on appropriate diagnostic accuracy measures (
e.g., receiver operator characteristic plot, sensitivity, specificity, predictive value, and diagnostic likelihood ratio). The test dataset must contain a sufficient number of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, concomitant diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Appropriate software documentation (
e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results; and cybersecurity).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use.
(ii) A detailed description of the intended reading protocol.
(iii) A detailed description of the intended user and user training that addresses appropriate reading protocols for the device.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and imaging protocols.
(vi) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
e.g., poor image quality or for certain subpopulations), as applicable.(vii) Device operating instructions.
(viii) A detailed summary of the performance testing, including: test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.