K Number
K242394

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-09-09

(27 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Digital Flat Panel X-Ray Detector is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. Not to be used for mammography.

Device Description

1717WCE, 1717WCE-HR, 1717WCE-HS, 1717WCE-GF X-ray detectors, are wired/wireless digital solid state X-ray detectors that are based on flat panel technology. The wireless LAN (IEEE 802.11 n/ac) communication signals images captured to the system and improves the user operability through high speed processing. These radiographic image detectors are processing unit consist of a scintillator coupled to an TFT sensor. The flat-panel detectors need to be integrated with an x-ray generator (not part of the submission), so it can be utilized to capture and digitize x-ray images for radiographic diagnosis.

1717WCE. 1717WCE-HR. 1717WCE-HS. 1717WCE-GF includes the software (firmware) of basic level of concern. It's the same Image Acquisition and Operating Software used for the predicate device is used but modified to include additional detector models in comparison with the predicate device. Full software documentation has been submitted, as well as the necessary sections to demonstrate device cybersecurity.

The RAW files can be further processed as DICOM compatible image files by separate consol SW (K190866, XmaruView V1 / Rayence Co.,Ltd) for a radiographic diagnosis and analysis.

1717WCE is the basic model. 1717WCE-HR is identical with the basic model except for pixel pitch not related to safety. 1717WCE-HS is identical with the basic model except for marking of sheet. 1717WCE-GF is identical with the basic model except for case color and pixel pitch.

AI/ML Overview

The given text describes a 510(k) submission for a Digital Flat Panel X-ray Detector. The submission aims to demonstrate substantial equivalence to a predicate device. However, the document does not describe a study that uses an AI algorithm as a device or an AI assistance to human readers, so most of the requested information regarding AI-specific criteria (like MRMC studies, standalone AI performance, training set details, or ground truth establishment for AI) is not present.

The document focuses on the technical and clinical performance comparison of the subject device (new X-ray detector models) against a predicate device (older X-ray detector models) through non-clinical and image quality assessments by human reviewers.

Here's the available information based on the provided text, addressing the points where information is available and noting where it's not:

1. Table of acceptance criteria and reported device performance:

The document doesn't present a formal table of "acceptance criteria" for the entire device as one might see for an AI algorithm's specific performance metrics (e.g., AUC, sensitivity, specificity). Instead, it demonstrates substantial equivalence by comparing its technological characteristics and performance to a predicate device. The performance is described qualitatively through comparisons of image quality.

CharacteristicSubject Device (1717WCE, 1717WCE-HR, 1717WCE-HS, 1717WCE-GF)Predicate Device (1417WCE, 1417WCE-HR, 1417WCE-HS, 1417WCE-GF)Reported Performance/Comparison
Intended UseDigital imaging for general radiographic system, human anatomy, replaces film/screen systems. Not for mammography.Digital imaging for general radiographic system, human anatomy, replaces film/screen systems. Not for mammography.Same
Detector TypeAmorphous Silicon, TFT (1717WCE, 1717WCE-HS); Amorphous Silicon, TFT, Indium Gallium Zinc Oxide with TFT (1717WCE-HR, 1717WCE-GF)Amorphous Silicon, TFTSimilar (some models of subject device use advanced TFT)
ScintillatorCsI:TlCsI:TlSame
Imaging Area17 x 17 inches14 x 17 inchesSimilar (Subject device has larger area)
Pixel Pitch (WCE, HS)140 µm140 µmSame
Pixel Pitch (WCE-HR, GF)99.9 µm100 µmSame (effectively)
Total Pixel Matrix (WCE, HS)3072 x 30722500 x 3052Similar
Total Pixel Matrix (WCE-HR, GF)4302 x 43023534 x 4302Similar
Resolution3.57 lp/mm (WCE, HS); 5.00 lp/mm (WCE-HR, GF)3.57 lp/mm (WCE, HS); 5.00 lp/mm (WCE-HR, GF)Same
DQE (@1lp/mm)Typ. 69% (WCE, HS); Typ. 62% (WCE-HR, GF)Typ. 63% (WCE, HS); Typ. 62% (WCE-HR, GF)Similar (some models of subject device show improvement)
MTF (@1lp/mm)Typ. 62% (WCE, HS); Typ. 66% (WCE-HR, GF)Typ. 66% (WCE, HS); Typ. 61% (WCE-HR, GF)Similar (some models of subject device show improvement)
A/D Conversion16 bits16 bitsSame
Dimensions460 x 460 x 15mm384 x 460 x 15mmSimilar
Weight3.3 kg2.7 kgSimilar
Viewer SoftwareXmaruView V1 (K190866/ Rayence Co.,Ltd)XmaruView V1 (K190866/ Rayence Co.,Ltd)Same

Qualitative Image Quality Assessment (summarized from section 6):

  • 1717WCE 99.9um vs. 1417WCE 100um: Subject device (1717WCE 99.9um) showed overall better image quality, clearer anatomical structures (bony and soft tissues of upper and lower extremities). Predicate device (1417WCE 100um) had decreased sharpness, more overexposed appearance, and higher noise.
  • 1717WCE 140um vs. 1417WCE 140um: Subject device (1717WCE 140um) showed overall better image quality, clearer anatomical structures. Predicate device (1417WCE 140um) had less sharpness, more overexposed appearance, and higher noise.
  • Conclusion: Both 1717WCE 140um and 1717WCE 99.9um demonstrated sufficient image quality for diagnostic purposes, with better image quality than the predicate device.

2. Sample size used for the test set and the data provenance:

  • Sample Size: The document states, "After comparing a broad review of plain radiographic images taken with 1717WCE... and 1417WCE images obtained equivalent quality for the same view obtained from a similar patient." It further mentions reviewing "plain radiographic images taken with 1717WCE 99.9um and the 1417WCE 100um" and "plain radiographic images taken with 1717WCE 140um and the 1417WCE 140um." No specific numerical sample size (e.g., number of images, number of patients) is provided for this qualitative review.
  • Data Provenance: Not explicitly stated (e.g., country of origin). The studies appear to be internal performance assessments rather than large-scale clinical trials. The data is retrospective in the sense that existing images were compared.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • The document mentions "Upon review of the plain radiographic images..." suggesting a human review. However, it does not specify the number of experts, their qualifications (e.g., radiologist with X years of experience), or how ground truth was established by them. The "ground truth" here seems to be subjective human judgment of image quality for diagnostic purposes rather than a definitive disease presence/absence.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • No adjudication method is described. The qualitative image quality assessment is presented as a singular conclusion derived from "review."

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, an MRMC comparative effectiveness study was not done. This submission is for an X-ray detector, not an AI algorithm assisting human readers. The qualitative image review is a comparison of image detector performance, not human reader performance with or without AI.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. The device is a digital X-ray detector, which produces images. It's not a standalone AI algorithm designed to interpret those images without human involvement.

7. The type of ground truth used:

  • Qualitative Human Assessment of Image Quality. The "ground truth" for the performance study is based on visual assessment by unspecified reviewers that the image quality of the subject device is "better" or "sufficient" for diagnostic purposes compared to the predicate device. It is not tied to a confirmed diagnosis (e.g., pathology, surgical findings, or long-term outcomes data). The document also mentions "non-clinical test report for the subject device were prepared and submitted to FDA... by using the identical test equipment and same analysis method described by IEC 62220-1" for metrics like MTF, DQE, and NPS, which are objective image quality measurements.

8. The sample size for the training set:

  • Not applicable. This device is an X-ray detector, not an AI or machine learning model that requires a training set.

9. How the ground truth for the training set was established:

  • Not applicable. As the device is not an AI/ML model, there is no "training set" or "ground truth for the training set."

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.

September 9, 2024

Rayence Co., Ltd % Mr. Dave Kim Regulatory Affairs Consultant Mtech Group LLC 7505 Fannin St. Suite 610 HOUSTON, TX 77054

Re: K242394

Trade/Device Name: Digital Flat Panel X-ray Detector (1717WCE, 1717WCE-HR, 1717WCE-HS, 1717WCE-GF) Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MOB Dated: May 30, 2024 Received: August 13, 2024

Dear Mr. Dave Kim:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang

Lu Jiang, Ph.D. Assistant Director Diagnostoc X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K242394

Device Name Digital Flat Panel X-Ray Detector (1717WCE, 1717WCE-HR, 1717WCE-HS, 1717WCE-GF)

Indications for Use (Describe)

Digital Flat Panel X-Ray Detector is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. Not to be used for mammography.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the word "rayence" in red font. The font is sans-serif and appears to be bolded. The word is written in lowercase letters.

510(k) Summary K242394

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date 510k summary prepared: July 30, 2024

Submitter's Name :Rayence Co., Ltd.
Submitter's Address:14, Samsung 1-ro 1-gil, Hwaseong-si, Gyeonggi-do, Korea
Submitter's Telephone:+82-31-8015-6459
Contact person:Ms. Brianna Park / Manager / +82-31-8015-6327
Official Correspondent:Dave Kim (davekim@mtechgroupllc.com)
Address:7505 Fannin St. Ste 610-V111, Houston, TX 77054
Telephone:+713-467-2607

Name of the device, including the trade or proprietary name if applicable, the common or usual name and the classification name, if known:

Subject Device:
Trade/proprietary name: Digital Flat Panel X-ray Detector (1717WCE, 1717WCE-HR,1717WCE-HS, 1717WCE-GF)
Regulation Number: 21 CFR 892.1680
Regulation Name: Stationary X-ray System
Regulatory Class: Class II
Product Code: MQB
Predicate Device :
Trade/Device Name:: Digital Flat Panel X-ray Detector (Model: 1417WCE, 1417WCE-HR,1417WCE-HS, 1417WCE-GF)
510(k) Number: K231467
Regulation Number: 21 CFR 892.1680
Regulation Name: Stationary X-ray System
Regulatory Class: Class II

{5}------------------------------------------------

Product Code : MQB

2. Device Description

1717WCE, 1717WCE-HR, 1717WCE-HS, 1717WCE-GF X-ray detectors, are wired/wireless digital solid state X-ray detectors that are based on flat panel technology. The wireless LAN (IEEE 802.11 n/ac) communication signals images captured to the system and improves the user operability through high speed processing. These radiographic image detectors are processing unit consist of a scintillator coupled to an TFT sensor. The flat-panel detectors need to be integrated with an x-ray generator (not part of the submission), so it can be utilized to capture and digitize x-ray images for radiographic diagnosis.

1717WCE. 1717WCE-HR. 1717WCE-HS. 1717WCE-GF includes the software (firmware) of basic level of concern. It's the same Image Acquisition and Operating Software used for the predicate device is used but modified to include additional detector models in comparison with the predicate device. Full software documentation has been submitted, as well as the necessary sections to demonstrate device cybersecurity.

The RAW files can be further processed as DICOM compatible image files by separate consol SW (K190866, XmaruView V1 / Rayence Co.,Ltd) for a radiographic diagnosis and analysis.

1717WCE is the basic model. 1717WCE-HR is identical with the basic model except for pixel pitch not related to safety. 1717WCE-HS is identical with the basic model except for marking of sheet. 1717WCE-GF is identical with the basic model except for case color and pixel pitch.

ModelColorPixel pitch
1717WCEWhite140um
1717WCE-HRWhite99.9um
1717WCE-GFBlack99.9um
1717WCE-HSWhite & Green Mark140um

3. Indication for use

Digital Flat Panel X-Ray Detector is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. Not to be used for mammography.

{6}------------------------------------------------

4. Summary of Design Control Risk management

1717WCE, 1717WCE-HR, 1717WCE-HS, 1717WCE-GF digital X-ray detectors are modification of 1417WCE (K231467).

The risks and the hazardous impact of the device modification were analyzed with FMEA method. The specific risk control and protective measures to mitigate the risks from the modification were reviewed and implemented in the new product design phase. The overall assessment concluded that all risks and hazardous conditions identified arising from the design change were successfully mitigated and accepted.

5. Summary of the technological characteristics of the device compared to the predicate device:

Detector described in this 510(k) has the same indications for use and similar technical characteristics as its predicate devices (K231467).

5.1 Scintillator layer

  • scintillator layer. ( * scintillator : a phosphor that produces scintillations)
Scintillator TypeProposedPredicate
CsI (Cesium Iodide)1717WCE, 1717WCE-HR,1717WCE-HS, 1717WCE-GF1417WCE, 1417WCE-HR,1417WCE-HS, 1417WCE-GF(K231467)

{7}------------------------------------------------

5.2 Comparison table

Subject devicePredicate device
Model1717WCE1717WCE-HS1717WCE-HR1717WCE-GF1417WCE1417WCE-HS1417WCE-HR1417WCE-GFSimilarity
FeatureImage: 1717WCEImage: 1717WCE-HRImage: 1417WCEImage: 1417WCE-HR
Image: 1717WCE-HSImage: 1717WCE-GFImage: 1417WCE-HSImage: 1417WCE-GF
IntendedUseDigital Flat Panel X-Ray Detector isindicated for digital imaging solutiondesigned for general radiographic system forhuman anatomy. It is intended to replacefilm or screen based radiographic systems inall general purpose diagnostic procedures.Not to be used for mammography.Digital Flat Panel X-Ray Detector isindicated for digital imaging solutiondesigned for general radiographic system forhuman anatomy. It is intended to replacefilm or screen based radiographic systems inall general purpose diagnostic procedures.Not to be used for mammography.Same
DetectorTypeAmorphous Silicon,TFTAmorphous Silicon,TFTIndium Gallium ZincOxide with TFT(Single panel)Amorphous Silicon, TFTSimilar
ScintillatorCsI:TlCsI:TlSame
ImagingArea17 x 17 inches14 x 17 inchesSimilar
Pixel pitch140 µm99.9 µm140 µm100 µmSame
TotalPixelmatrix3072 X 30724302 X 43022500 X 30523534 X 4302Similar
Resolution3.57 lp/mm5.00 lp/mm3.57 lp/mm5.00 lp/mmSame
DQE(@1lp/mm)Typ.69%Typ.62 %Typ.63%Typ.62 %Similar
MTF(@1lp/mm)Typ.62%Typ.66%Typ.66%Typ.61%Similar
A/Dconversion16 bits16 bitsSame
Dimensions460 X 460 X 15mm384 X 460 X 15mmSimilar
Weight3.3 kg (incl. 1 battery)2.7 kg (incl. 1 battery)Similar
ViewerSoftware(option)XmaruView V1 (K190866/ RayenceCo.,Ltd)XmaruView V1 (K190866/ RayenceCo.,Ltd)Same

{8}------------------------------------------------

6. Summary of Performance Testing

1717WCE, 1717WCE-HR, 1717WCE-HS, 1717WCE-GF X-ray detectors have same indications for use, scintillator material (Csl:T1), general specifications and same risk analysis characteristics compared to the predicate device, 1417WCE (K231467). The pixel matrix and pixel pitch sizes are different imaging areas but the differences do not raise new concerns for the safety and effectiveness of the subject device.

The clinical and non-clinical test report for the subject device were prepared and submitted to FDA to demonstrate the substantial equivalency of the subject device performance compared to the predicate device.

After comparing a broad review of plain radiographic images taken with 1717WCE, 1717WCE-HR, 1717WCE-HS, 1717WCE-GF and 1417WCE images obtained equivalent quality for the same view obtained from a similar patient .

Upon review of the plain radiographic images taken with 1717WCE 99.9um and the 1417WCE 100um, 1717WCE 99.9um has presented overall a better image quality of the same anatomical position in the separate patients. The anatomical structures, both bony and soft tissues of the upper and lower extremities were seen with better clarity in the 1717WCE 99.9um. The disadvantages of 1417WCE 100um were decreased sharpness, more overexposed appearance, and higher noise level in general. In conclusion, 1717WCE 99.9um had a clearer view than 1417WCE 100um, which will provide aid for diagnosis purposes.

Upon review of the plain radiographic images taken with 1717WCE 140um and the 1417WCE 140um, 1717WCE 140um has presented overall a better image quality of the same anatomical position in the separate patients. The anatomical structures, both bony and soft tissues of the upper and lower extremities were seen with better clarity in the 1717WCE 140um. The disadvantages of 1417WCE 140um were less sharpness, more over exposed appearance and higher noise level in general. In conclusion, 1717WCE 140um had a clearer view than 1417WCE 140um, which will provide aid for diagnosis purposes.

In conclusion, both 1717 WCE 140 um and 1717 WCE 99.9um have demonstrated sufficient image quality which will provide aid for diagnostic purposes.

The Radio Frequency Wireless Coexistence Test was conducted for the subject device. 1717WCE

{9}------------------------------------------------

wireless implements IEEE802.11 a/g/n to establish bi-directional communication between the X-ray detector and Access point. 1717WCE wirelss has been found capable of transmitting data within twenty seconds in a normal deployment. The wireless transmission feature is unchanged from the predicate device.

The non-clinical test report contains the MTF, DQE and NPS performance test comparison between the subject device, and the predicate device, by using the identical test equipment and same analysis method described by IEC 62220-1.

Based on the non-clinical consideration evaluation, the sponsor can claim the substantial equivalency between the subject device and the predicate device in terms of diagnostic image quality.

7. Summary for any testing and reference guidance:

  • A Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1:2005, AMD1:2012, AMD2:2020 (Medical electrical equipment Part 1:General requirements for basic safety and essential performance)
  • EMC testing were conducted in accordance with standard IEC 60601-1-2: 2014+AMD1:2020

  • A IEC 62220-1:2015 Ed 1.0 Medical electrical equipment-Characteristics of digital X-ray imaging devices Part 1-1: Determination of the detective quantum efficiency Detectors used in radiographic imaging
  • Non-clinical consideration according to FDA Guidance "Guidance for the Submissions A of 510(k)'s for Solid State X-ray Imaging Devices"
  • A "Guidance for the Contents of Premarket Submission for Software Contained in Medical Device".
  • Pediatric Information for X-ray Imaging Device Premarket Notifications A
  • A Content of Premarket Submissions for Management of Cybersecurity in Medical Devices

8. Conclusions:

In accordance with the performance outcomes, 1717WCE, 1717WCE-HR, 1717WCE-HS, 1717WCE-GF X-ray detectors demonstrated equivalent or better performance compared to 1417WCE. Therefore, Rayence claims the substantial equivalency between the subject device and the predicate device in terms of diagnostic image quality about safety and effectiveness.

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.