K Number
K242379

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-11-07

(87 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LL Implant System is intended to be use in the treatment of missing mandibular central and lateral incisors to support dental prosthetic devices, such as artificial teeth, in order to restore chewing function in partially edentulous patients. It is intended for single use only. It is intended for delayed loading.

Device Description

The LL Implant System consists of a one-piece implant and temporary cap. The LL Mini Onebody Implant is made of CP Ti Grade 4(ASTM F67) with the surface treated by the SLA method. It has several design characteristics: submerged type, tapered body, sided cutting edge. The LL Temporary Cap is made Polyoxymethylene (ASTM F1855) without any surface treatment.

AI/ML Overview

The provided text does not contain detailed acceptance criteria and a study that proves the device meets them in the format requested. The document discusses the substantial equivalence of the "LL Implant System" to predicate devices, primarily focusing on materials, design, and general non-clinical testing performed according to ISO standards and FDA guidance.

Here's an analysis based on the information available and what is missing regarding your specific questions:

1. Table of Acceptance Criteria and Reported Device Performance

  • Information Available: The document mentions that several non-clinical tests were performed (Fatigue Testing, Sterilization Validation, Shelf-Life Testing, Biocompatibility, Bacterial Endotoxin Testing, Surface Modification, and MR Safety). For some tests, it states that "the results have met the acceptance criteria" or "demonstrated equivalence to the predicate device."
  • Missing Information: Specific numerical acceptance criteria for each test and the corresponding quantitative results/performance from the LL Implant System are not provided. For example, for "Fatigue Testing," it says it was done "according to ISO 14801:2016" and "expected to function properly," but it doesn't state the specific load cycles, force applied, and the device's measured endurance limits to compare against predefined acceptance thresholds. Similarly for "Bacterial Endotoxin Testing," it says "met the acceptance criteria," but the criteria itself and the measured endotoxin levels are absent.

2. Sample Size Used for the Test Set and Data Provenance

  • Missing Information: The document does not specify the sample sizes used for any of the non-clinical tests. It refers to "worst-case scenario" for several tests but doesn't quantify how many samples were included in these scenarios. Additionally, there's no mention of the data provenance (e.g., country of origin, retrospective or prospective) as these are non-clinical lab tests and not human subject studies.

3. Number of Experts Used to Establish Ground Truth and Qualifications

  • Not Applicable: This question is relevant for studies involving human data and expert review for ground truth establishment (e.g., diagnostic AI models). The studies described here are non-clinical, laboratory-based tests on a physical medical device (dental implant). Thus, there's no "ground truth" to be established by human experts in this context.

4. Adjudication Method for the Test Set

  • Not Applicable: Similar to point 3, adjudication methods are typically used in clinical studies or studies involving human readers/experts to resolve discrepancies in ground truth labeling. This is not pertinent to the non-clinical device testing described.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Not Performed: The document does not mention any MRMC comparative effectiveness study, nor does it discuss human readers or AI assistance. The device is a physical dental implant, not an AI software.

6. Standalone (Algorithm Only) Performance Study

  • Not Performed: This question is referring to AI algorithm performance. The device is a physical dental implant, not an AI algorithm.

7. Type of Ground Truth Used

  • Not Applicable (for clinical ground truth): As these are non-clinical, laboratory tests, the "ground truth" aligns with established scientific and engineering principles, international standards (e.g., ISO 14801 for fatigue, ISO 11137 for sterilization), and FDA guidance documents. The "truth" is whether the device physically performs according to these standards, not an expert consensus or pathology on a biological sample.

8. Sample Size for the Training Set

  • Not Applicable: This question typically relates to AI/ML models. The LL Implant System is a physical device, and therefore, there is no "training set" in the context of an algorithm.

9. How the Ground Truth for the Training Set Was Established

  • Not Applicable: As there is no training set for an AI model, this question is not relevant.

Summary of what is available (and implied in relation to "acceptance criteria" and "proof"):

  • Acceptance Criteria (Implied): The acceptance criteria are largely implied by compliance with specific ISO standards (e.g., ISO 14801:2016 for fatigue, ISO 11137-1 for sterilization, ISO 10993-1 for biocompatibility) and FDA guidance documents. The document states that the test results "met the acceptance criteria" or "demonstrated substantial equivalence" to predicate devices, indicating that the predefined thresholds within these standards were achieved.
  • Study That Proves the Device Meets Criteria: Various non-clinical studies were conducted:
    • Fatigue Testing: Performed on the subject device according to ISO 14801:2016 and FDA guidance.
    • Sterilization Validation: Gamma sterilization (implants) validated per ISO 11137 series; End-user sterilization (caps) validated per ISO 17665 series and FDA guidance. These were leveraged from previous submissions (K223924, K231079).
    • Shelf-Life Test: Performed per ASTM F1980 (leveraged from K223924).
    • Biocompatibility Testing: Performed per ISO 10993-1 and FDA guidance, including Cytotoxicity per ISO 10993-5 (leveraged from K223924, K231079).
    • Bacterial Endotoxin Testing: Performed per USP <85> and <161> and FDA guidance.
    • Surface Modification Information: Evaluation per FDA guidance, including surface roughness, composition analysis, SEM, and ICP analysis.
    • MR Safety: Worst-case MRI review performed using scientific rationale and published literature, addressing magnetically induced displacement force and torque based on FDA guidance.

In short, the document confirms that required non-clinical tests were conducted according to recognized standards and guidance, and the results were deemed to meet acceptance criteria, thereby demonstrating substantial equivalence to predicate devices. However, it does not provide the specific quantitative details of these acceptance criteria or the measured performance results for each test.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the agency's name, "U.S. Food & Drug Administration," in blue text.

Ossvis Co., Ltd. Woojin Lee Regulatory Affairs Specialist 38, Burim-ro 170beon-gil, Dongan-gu 6F, 7F and B1 Anyang-si, Gyeonggi-do 14055 REPUBLIC OF KOREA

November 7, 2024

Re: K242379

Trade/Device Name: LL Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: August 9, 2024 Received: August 12, 2024

Dear Woojin Lee:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Andrew I. Steen -S

Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K242379

Device Name LL Implant System

Indications for Use (Describe)

The LL Implant System is intended to be use in the treatment of missing mandibular central and lateral incisors to support dental prosthetic devices, such as artificial teeth, in order to restore chewing function in partially edentulous patients. It is intended for single use only. It is intended for delayed loading.

Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "Ossvis" in a bold, green font. The letters are closely spaced together, creating a compact and unified appearance. Above the "i" is a curved, gray line.

510(K) Summary

Submitter Information

Ossvis Co., Ltd. Woo Jin Lee 6F, 7F and B1, 38, Burim-ro 170beon-gil, Dongan-gu, Anyang-si, Gyeonggi-do, Republic of Korea Email: wjl316@huvitz.com Tel. +82-31-360-0082 Fax. +82-31-360-0058

Device Information

  • Trade Name: LL Implant System
  • Common Name: Endosseous Dental Implant ●
  • . Classification Name: Implant, Endosseous, Root-Form
  • Primary Product Code: DZE .
  • . Secondary Product Code: NHA
  • Panel: Dental
  • . Regulation Number: 21 CFR 872.3640
  • Device Class: Class II
  • Date Prepared: October 16,2024

Predicate Devices:

The subject device is substantially equivalent to the following predicate devices:

Primary Predicate

  • . K191201, EM SA Implant System by Hiossen, Inc.
    Reference Device

  • K170394, Kisses Mini by ACHIMHAI Medical Corporation ●

  • K112540, S-MiNi Implant System by Neobiotech Co., Ltd. ●

  • K223924, LW Implant System by Ossvis Co., Ltd. ●

  • K231079, LW Retraction Cap by Ossvis Co., Ltd.

  • K182091, Osstem Abutment System by OSSTEM IMPLANT Co., Ltd. ●

Indication for Use:

The LL Implant System is intended to be use in the treatment of missing mandibular central and lateral incisors to support dental prosthetic devices, such as artificial teeth, in order to restore chewing function in partially edentulous patients. It is intended for single use only. It is intended for delayed loading.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "Ossvis" in a bold, green font. There is a curved line above the "i" in "Ossvis" that is a light gray color. The word is centered and takes up most of the image. The background is white.

Device Description:

The LL Implant System consists of a one-piece implant and temporary cap.

The LL Mini Onebody Implant is made of CP Ti Grade 4(ASTM F67) with the surface treated by the SLA method. It has several design characteristics: submerged type, tapered body, sided cutting edge. The LL Temporary Cap is made Polyoxymethylene (ASTM F1855) without any surface treatment.

NoDevice NameDimension
1LL Mini Onebody ImplantØ 2.7, 3.1 (D) x 8.5, 10.0, 11.5, 13.0 mm (L)
2LL Temporary CapØ 4.0 (D) x 9.6 mm (L)

The dimensions of subject devices are as following:

The below table outlines the information of intended uses and surface treatment:

Device NameUsesSurface Treatment
LL Mini Onebody ImplantImplant is intended to replace the patient'snatural tooth and restore chewing function.SLA
LL Temporary CapUsed to protect the upper part of LL MiniOnebody Implant in the oral cavityNone

The LL Mini Onebody Implant is provided sterile. And LL Temporary Cap is provided non-sterile, which are required to be sterilized by the end-user before use.

Materials:

  • Dental Implant is fabricated from Pure titanium of ASTM F67
  • Temporary Cap is fabricated from POM(Polyoxymethylene) of ASTM F1855 ●

{6}------------------------------------------------

Ossvīs

Ossvis Co., Ltd. 6F, 7F and B1, 38, Burim-ro 170beon-gil, Dongan-gu, Anyang-si, Gyeonggi-do, Republic of Korea

Summaries of Technological Characteristics & Substantial Equivalence Discussion

LL Implant System

LL Mini Onebody Implant
Subject DevicePredicate DeviceReference DeviceReference Device
510(k) #K242379K191201K170394K112540
Device NameLL Implant SystemEM SA Implant SystemKisses MiniS-MiNi Implant System
Implant Name/TypeLL Mini Onebody ImplantNarrow Ridge TypePost TypeCement Type
ManufacturerOssvis Co., Ltd.Hiossen, Inc.Achimhai Medical CorporationNeobiotech Co., Ltd.
Product CodeDZEDZEDZEDZE
Regulation21 CFR 872.364021 CFR 872.364021 CFR 872.364021 CFR 872.3640
AppearanceImage: LL Mini Onebody ImplantImage: EM SA Implant SystemImage: Kisses MiniImage: S-MiNi Implant System
Indications for Use StatementThe LL Implant System isintended to be use in thetreatment of missing mandibularcentral and lateral incisors tosupport dental prosthetic devices,such as artificial teeth, in order torestore chewing function inpartially edentulous patients. It isintended for single use only. It isintended for delayed loading.The EM SA Implant System(DENTURE) is intended to beplace in the bone of the upper orlower jaw arches providingsupport to dental prostheticdevices, specifically for denturestabilization to restore a patient'schewing function. It is intendedfor single use only.The EM SA Implant (NARROWRIDGE) is intended to be use inthe treatment of missingmandibular central and lateralincisors to support dentalprosthetic devices, such asartificial teeth, in order to restorechewing function in partiallyedentulous patients. It is intendedThe Kisses Mini is indicated for usein partially or fully edentulousmandibles and maxilla, in supportof single or multiple-unitrestorations including; cementedretained, screw retained, oroverdenture restorations, andterminal or intermediate abutmentsupport for fixed bridgework. TheKisses Mini is for single and twostage surgical procedures. It is fordelayed loading.The intended use of the 3.2 mmInternal Hex Fixture is limited tothe replacement of maxillary lateralincisors and mandibular incisors.The intended use of the single-piecepost type fixture is limited to theThe S-MiNi Implant System isdivided into two types:- Cement TypeThe Cement type is indicated foruse in the treatment of missingmaxillary lateral incisors or themandibular central and lateralincisors to serve as temporarysupport prosthetic devices duringthe healing phase of permanentendosseous dental implant, such asartificial teeth, in order to restorechewing function in partiallyedentulous patients.- Ball Type
for single use only. It is intendedfor delayed loading.replacement of mandibular centraland lateral incisors.The ball-type single-piece implantis intended for stabilization andretention of overdentures.The Ball type is designed for use indental implant surgery. Ball type isintended for use in partially or fullyedentulous mandibles and maxiliaein support of overdentures. Balltype implants are for temporary useonly.
• Integrated Implant System (Onebody)• Integrated Implant System(One body)• Integrated Implant System(One body)• Integrated Implant System(One body)
Structure• Tapered & Straight body shape• Threaded body design• Cutting edge with self- tapping• Tapered & Straight body shape• Threaded body design• Cutting edge with self- tapping• Tapered & Straight body shape• Threaded body design• Cutting edge with self- tapping• Tapered & Straight body shape• Threaded body design• Cutting edge with self -tapping
Diameter (Ø)2.7, 3.12.5, 3.02.8, 3.02.0, 2.5, 3.0, 3.5
Length (mm)8.5, 10.0, 11.5, 13.08.5, 10.0, 11.5, 13.0, 15.08.0, 9.5, 11.0, 12.5, 14.57.0, 8.5, 10.0, 11.5, 13.0, 15.0
MaterialPure Titanium Grade 4(ASTM F67)Ti 6Al 4V ELI(ASTM F136)Pure Titanium Grade 4(ASTM F67)Pure Titanium Grade 4(ASTM F67)
SterilizationGamma irradiationGamma irradiationGamma irradiationGamma irradiation
Surface treatmentSLASLASLARBM
Substantial Equivalence Discussion
The II Mini Onebody Implant has the same intended use, design features, surface treatment, and sterilization as the primary predicate K101201. Both devices feature a one-

LL Mini Onebody Implant

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the word "Ossvis" in a bold, green font. There is a curved line above the "is" portion of the word. The background is white.

Ossvis Co., Ltd. 6F, 7F and B1, 38, Burim-ro 170beon-gil, Dongan-gu, Anyang-si, Gyeonggi-do, Republic of Korea

110 - 22 min other intention and mich clearly, and on the man and mind may produced on one producer on the species the may predicate devices are the material and dimensions The prince is made of T-6A1-V EL and does not include diameters above 3.1 mm, whereas the subject de vice is made of Pure Titanium Grade 4. However, these differences are overed by the reference devices. Reference devices K170394 is made of Pare Titanium Grade 4 and undergoes SLA surface treation. Reference device K112540 offers a vider range of dimensions than the subject device, supporting substantial equivalence. Therefore, the subject device is considered substantially equivalent.

{8}------------------------------------------------

Ossvis Co., Ltd. 6F, 7F and B1, 38, Burim-ro 170beon-gil, Dongan-gu, Anyang-si, Gyeonggi-do, Republic of Korea

Ossvīs

LL Temporary Cap

Subject DeviceReference DeviceReference Device
510(k) #K242379K231079K182091
Device NameLL Implant SystemLW Retraction CapOsstem Abutment System
Abutment NameLL Temporary CapLW Retraction CapTemporary Cap (Narrow Ridge)
ManufacturerOssvis Co., Ltd.Ossvis Co., Ltd.OSSTEM IMPLANT Co., Ltd.
Product CodeNHANHANHA
Regulation21 CFR 872.363021 CFR 872.363021 CFR 872.3630
AppearanceImage: LL Temporary CapImage: LW Retraction CapImage: Temporary Cap (Narrow Ridge)
Intended UseUsed in the oral cavity to protect the upper partof compatible one-piece implant from foreignsubstances such as plaque.Used in the oral cavity to protect compatibleabutments from foreign substances such asplaque.Used in the oral cavity to protect compatibleabutments from foreign substances such asplaque.
Diameter (Ø)4.05.0, 6.0, 6.1, 6.7, 7.2, 7.7, 8.2, 8.74.0
Length (mm)9.65.5, 7.0, 8.59.6
MaterialPOM (Polyoxymethylene)(ASTM F1855)POM (Polyoxymethylene)(ASTM F1855)Poly Carbonate
SterilizationN/AN/AN/A
Surface treatmentN/AN/AN/A
Substantial Equivalence Discussion
The LL Temporary Cap has the same indication for use, material, design feature, structure, surface treatment, and sterilization as the reference devices.The difference between the subject and reference device(K231079) is dimensions. However, it is used temporarily in the oral cavity and the size is designed to fit itscompatible implant considering the intended use of the device. Also the dimensional difference is covered by the reference device(K182091). Therefore, we believe that thedifference does not raise any serious issues in performance or safety, and the subject device is substantial equivalence to the predicate device.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the word "Ossvis" in a bold, green font. There is a curved line above the "i" in "Ossvis". The background is white.

Non-Clinical Test Data

Below tests were performed on the subject device:

  • Fatigue Testing according to ISO 14801:2016 .
    The Below tests were performed for our previous submission devices and leveraged for the subject device:

  • Gamma Sterilization Validation Test on LW Fixture according to ISO 11137-1, ISO 11137-2, . and ISO 11137-3 referenced in K223924 (LW Implant System)

  • . End User Sterilization Validation Test Report on LW Retraction Cap according to ISO 17665-1, and ISO 17665-2 referenced in K231079 (LW Retraction Cap)

  • . Shelf-Life Test on LW Fixture according to ASTM F1980 referenced in K223924 (LW Implant Svstem)

  • Biocompatibility testing on LW Fixture according to ISO 10993-1:2018 referenced in K223924 . (LW Implant System)

  • Biocompatibility testing on LW Retraction Cap according to ISO 10993-1:2018 referenced in ● K231079 (LW Retraction Cap)

  • LAL endotoxin testing on LW Fixture according to USP 43, <161> referenced in K223924 (LW Implant System)

  • . Surface Modification Information on LW Fixture referenced in K223924 (LW Implant System)

A brief description of the leveraged information is as follows.

[Surface Modification Information]

The surface treatment evaluation has been performed in accordance with 'Section 11 of Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutment'. The surface modification with SLA (sand-blasted, large-grit, acid-etched) was provided. To compare surface modification between the subject and predicate device, surface roughness, surface composition analysis, SEM imaging, and ICP analysis were provided, and it demonstrates substantial equivalence.

[Gamma Sterilization Validation]

For devices delivered in a sterile state, such as Implants, a sterility assurance level (SAL) of 10 has been validated in accordance with ISO 11137-1 "Sterilization of health care products – Radiation – Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices". The validation took into account the worst-case scenario, and the results prove equivalence to the predicate device.

[End User Sterilization Validation]

The non-sterile abutments are intended to be sterilized by the end user. The recommended sterilization conditions have been validated through the end-user sterilization, following ISO 17665-1 "Sterilization of health care products - Moist heat - part 1: Requirements for the development. validation, and routine control of a sterilization process for medical devices", ISO 17665-2 "Sterlization of health care products – Moist heat – part 2: Guidance on the application of ISO 17665-1", and the FDA guidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, issued on March 17, 2015". The validation considered the worst-case scenario, and the results demonstrated equivalence to the predicate device.

{10}------------------------------------------------

Ossvis

[Biocompatibility Testing]

Biocompatibility Testing was performed in accordance with ISO 10993-1 "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" and the FDA Guidance document, "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process', Guidance for Industry and Food and Drug Administration Staff". Cytotoxicity testing was performed according to ISO 10993-5.

[Shelf-life Testing]

The devices provided in a sterile state underwent Shelf-life testing in accordance with ASTM F1980, "Standard Guide for Accelerated Aging of Sterile Barrier Systems and Medical Devices". The worstcase scenario was tested, and the results demonstrated that the devices are equivalent to the predicate devices. The shelf-life for the Implant is 5 years, and throughout this time frame, the devices will function as intended without any degradation. Please note that the devices will not be labeled as 'nonpyrogenic ' when they are marketed.

[Fatigue Testing]

To evaluate performance of the subject device, Dynamic Fatigue, and Static Compression Strength tests were conducted according to the FDA guidance document "Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments" and ISO 14801:2016, "Dentistry – Implants – Dynamic fatigue test for endosseous dental implants" under the worst-case scenario. As a result, our dental implant system is expected to function properly for its intended use, which is to replace the patient's natural tooth and restore chewing function.

[Bacterial Endotoxin Testing]

To address the presence of bacterial endotoxins, the devices corresponding to implants should meet the pyrogen limit specifications as specified in the FDA guidance document "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile" and USP <161> "Medical Devices – Bacterial Endotoxin and Pyrogen Tests". Therefore, our dental implants were tested for bacterial endotoxin under USP <85> "Bacterial

Endotoxins Test", and USP <161> "Medical Devices – Bacterial Endotoxin and Pyrogen Tests". The test results have met the acceptance criteria and demonstrated the substantial equivalence with the predicate device.

[MR Safety]

Non-clinical worst-case MRI review was performed to evaluate the subject devices in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 4 9.2 (2 01 9): 7 8 3-795), based on the entire system including all implants and abutments and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque.

The non-clinical testing results demonstrate that the subject devices are substantially equivalent to the predicate devices.

Conclusion

The documentation submitted in this premarket notification demonstrates the LL Implant System is substantially equivalent to the primary predicate and reference devices.

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.