K Number
K242294

Validate with FDA (Live)

Date Cleared
2025-05-09

(280 days)

Product Code
Regulation Number
866.3215
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

DiaSys Procalcitonin FS assay is a particle enhanced immunoturbidimetric test intended for the quantitative in vitro determination of procalcitonin (PCT) in human serum and lithium heparin plasma on automated Abbott ARCHITECT c8000 analyzer.

Measurement of PCT in conjunction with other laboratory findings and clinical assessments aids in the risk assessment of critically ill patients on their first day of Intensive Care Unit (ICU) admission for progression to severe sepsis and septic shock.

The TruCal Procalcitonin Calibrator Set is intended for in vitro use for calibration of the DiaSys Procalcitonin FS assay.

TruLab Procalcitonin Bi-Level Controls are an assayed quality control material for monitoring the performance of quantitative in vitro determination of Procalcitonin (PCT) for the DiaSys Procalcitonin FS assay.

For in vitro diagnostic use only.

Device Description

The DiaSys Procalcitonin FS assay is a particle enhanced turbidimetric immunoassay (PETIA) test for in vitro quantitative determination of procalcitonin (PCT) levels in human serum and lithium heparin plasma on automated photometric systems.

The test utilizes anti-human PCT polyclonal antibodies (goat) covalently bound to polystyrene particles.

PCT in the sample binds to the anti-PCT antibodies on the particles and causes agglutination. The degree of the turbidity caused by agglutination is optically measured at 660 nm by the automated photometric system and is proportional to the amount of PCT in the sample. DiaSys Procalcitonin FS Reagent specifies settings for an automated photometric analyzer, the Abbott ARCHITECT c8000.

The DiaSys TruCal Procalcitonin Calibrator Set is provided in 6 levels, traceable to a commercial assay for use in the calibration of the DiaSys Procalcitonin FS assay.

The DiaSys TruLab Procalcitonin Bi-Level Controls, Levels 1 (low) and 2 (high) are intended for use as quality control for the DiaSys Procalcitonin FS assay.

DiaSys PCT Calibrators and Bi-Level Controls are ready-to-use, stable, aqueous solutions containing recombinant human full-length PCT and biological additives from bovine origin.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the DiaSys Procalcitonin FS assay, extracted and organized from the provided FDA 510(k) clearance letter:

Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaReported Device Performance
Precision/Reproducibility
Total Precision (Serum)CV ≤ 10% (0.2-1.0 ng/mL)CV ≤ 7% (1.0-3.0 ng/mL)CV ≤ 5% (3.0-20 ng/mL)Met:0.2-1.0 ng/mL: 5.64%1.0-3.0 ng/mL: 2.34%3.0-20 ng/mL: 1.93%
Total Precision (Plasma)CV ≤ 10% (0.2-1.0 ng/mL)CV ≤ 7% (1.0-3.0 ng/mL)CV ≤ 5% (3.0-20 ng/mL)Met:0.2-1.0 ng/mL: 8.31%1.0-3.0 ng/mL: 3.24%3.0-20 ng/mL: 2.61%
Within-run Precision (Serum)CV ≤ 10% (0.2 ng/mL range)CV ≤ 7% (1.0-3.0 ng/mL range)CV ≤ 5% (3.0-20 ng/mL range)Met (except for LoQ proximity):0.252 ng/mL: 16.62% (Determined acceptable given proximity to LoQ)0.555 ng/mL: 3.89%2.02 ng/mL: 2.78%9.68 ng/mL: 2.24%18.8 ng/mL: 2.16%
Within-run Precision (Plasma)CV ≤ 10% (0.2 ng/mL range)CV ≤ 7% (1.0-3.0 ng/mL range)CV ≤ 5% (3.0-20 ng/mL range)Met:0.314 ng/mL: 8.00%0.475 ng/mL: 6.94%1.910 ng/mL: 2.93%9.851 ng/mL: 1.94%17.953 ng/mL: 1.96%
Linearity/Assay Reportable RangeLinearity < 0.5 ng/mL: ± 0.1 ng/mLLinearity 0.5-5 ng/mL: ± 20%Linearity > 5 ng/mL: ± 10%Met (Human serum and plasma evaluations)
Detection LimitsLoB ≤ 0.2 ng/mLLoQ ≤ 0.5 ng/mLMet:Limit of Blank (LOB): 0.081 ng/mLLimit of Quantitation (LOQ): 0.23 ng/mL (with %CV ≤ 20%)
InterferenceNo significant interference (shift in PCT concentration > ±15%)Met: (No significant interference observed for substances tested at specified concentrations, including Ascorbic acid, Bilirubin, Lipemia, HAMA, Rheumatoid factor, Hemolysis, and various drugs. N-Terminus interferes.)
Stability (Reagent)
Real-time Stability24 months stabilityConfirmed: 25 months stability from date of production.
In-use Stability24 months stabilityConfirmed: At least 25 months stability after opening.
Specimen StabilityWithin ± 15% of concentration at time zeroSlope of regression line within ± 10%Met: for 24 hours at 20-25°C, 5 days at 2-8°C, and 14 days at -20°C (for both serum and plasma, and measurements >10% over specified durations).
Freeze-Thaw Cycle (2 cycles)Within 15% of concentration at time zeroMet: for both serum and plasma samples after one and two freeze-thaw cycles.
Hook Effect (Prozone)No hook/prozone effectConfirmed: No hook/prozone effect up to 1000 ng/mL.
Method Comparison (Study 1, 2021)Slope: 0.85 – 1.15Intercept: ± 1.0 ng/mLCorrelation coefficient R: ≥ 0.95n ≥ 100Met:N = 120Slope = 1.08Intercept = 0.105 ng/mLR = 0.991PCT Cutoff 0.5 ng/mL: NPA 93.9%, PPA 100%PCT Cutoff 2.0 ng/mL: NPA 93.3%, PPA 97.8%
Method Comparison (Study 2, 2025)Slope: 0.85 – 1.15Intercept: ± 1.0 ng/mLCorrelation coefficient R: ≥ 0.95n ≥ 100Met:N = 210Slope = 0.940Intercept = 0.017 ng/mLR = 0.965PCT Cutoff 0.5 ng/mL: NPA 98%, PPA 100%PCT Cutoff 2.0 ng/mL: NPA 100%, PPA 94%
Matrix ComparisonSlope: 0.85 – 1.15Intercept: ± 1.0 ng/mLCorrelation coefficient R ≥ 0.98n ≥ 20Met: (Human serum vs. lithium heparin plasma, for at least 20 native patient samples)

Study Details:

  1. Sample Size used for the test set and the data provenance:

    • Precision/Reproducibility (Total Precision): 3 human serum samples, 3 human plasma samples.
    • Precision/Reproducibility (Within-run Precision): 5 human serum samples, 5 human plasma samples.
    • Linearity/Assay Reportable Range: Stock solutions assayed in quadruplicate.
    • Detection Limits (LOB, LOD, LOQ): Evaluated in human serum and plasma.
    • Interference: Human samples at approximate PCT concentrations of 0.5 and 2.0 ng/mL.
    • Specimen Stability: 15 human samples (serum and lithium heparin plasma).
    • Freeze-Thaw Cycle Stability: 15 human samples (serum and plasma).
    • Method Comparison (Study 1): 120 native patient samples.
    • Method Comparison (Study 2): 210 native patient samples.
    • Matrix Comparison: At least 20 native patient samples (serum and lithium heparin plasma).
    • Data Provenance: Not explicitly stated as retrospective or prospective, nor country of origin for patient/human samples. The nature of "native patient samples" generally implies retrospective analysis of collected clinical samples, but this is not definitively stated.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable for this type of in vitro diagnostic device (analyte measurement). The "ground truth" for the test set samples is typically established by reference methods or established values/concentrations for controls/calibrators, rather than expert clinical consensus.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. This is an assay for quantitative measurement of an analyte. "Adjudication" typically refers to the process of multiple readers/experts reviewing images or clinical cases.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is an in vitro diagnostic assay, not an AI-assisted diagnostic imaging or clinical decision support system. The comparison studies are against a predicate device (another PCT assay), not human readers.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Yes, the performance listed in the table effectively represents the standalone performance of the DiaSys Procalcitonin FS assay when run on the automated Abbott ARCHITECT c8000 analyzer, without human intervention in the result generation itself. The results are quantitative measurements of the Procalcitonin analyte.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Analytical Performance: Ground truth is based on known concentrations of analytes in controls and calibrators, and comparative measurements against a legally marketed predicate device (VIDAS B.R.A.H.M.S. PCT ASSAY, K071146).
    • Clinical Efficacy/Impact: The device aids in risk assessment of critically ill patients for progression to severe sepsis and septic shock, "in conjunction with other laboratory findings and clinical assessments." This implies that the actual clinical "ground truth" (e.g., diagnosis of severe sepsis or septic shock, patient outcome) relies on a broader clinical picture, not solely on the PCT measurement. The study primarily focuses on analytical equivalence, not a direct measure of clinical outcomeprediction compared to a gold standard outcome.
  7. The sample size for the training set:

    • The document primarily describes validation and verification studies for the device's analytical performance and equivalence to a predicate. It does not mention a "training set" in the context of machine learning or AI algorithm development, as this device appears to be a traditional IVD assay.
  8. How the ground truth for the training set was established:

    • Not applicable, as there is no apparent "training set" in the machine learning sense. The device is a chemical reagent-based assay.

FDA 510(k) Clearance Letter for DiaSys Procalcitonin FS

Page 1

U.S. Food & Drug Administration

10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

May 9, 2025

DiaSys Diagnostic Systems GmbH
℅ Stephen Gorski
Official Correspondent
Imagenix, Inc.
S65W35739 Piper Road
Eagle, Wisconsin 53119

Re: K242294
Trade/Device Name: DiaSys Procalcitonin FS; DiaSys TruCal Procalcitonin Calibrator Set; DiaSys TruLab Procalcitonin Bi-Level Controls
Regulation Number: 21 CFR 866.3215
Regulation Name: Device To Detect And Measure Non-Microbial Analyte(S) In Human Clinical Specimens To Aid In Assessment Of Patients With Suspected Sepsis
Regulatory Class: Class II
Product Code: PTF
Dated: July 15, 2024
Received: August 2, 2024

Dear Stephen Gorski:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K242294 - Stephen Gorski Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K242294 - Stephen Gorski Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Noel J. Gerald -S

Noel J. Gerald, Ph.D.
Deputy Division Director
Division of Microbiology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K242294

Device Name
DiaSys Procalcitonin FS; DiaSys TruCal Procalcitonin Calibrator Set; DiaSys TruLab Procalcitonin Bi-Level Controls

Indications for Use (Describe)
DiaSys Procalcitonin FS assay is a particle enhanced immunoturbidimetric test intended for the quantitative in vitro determination of procalcitonin (PCT) in human serum and lithium heparin plasma on automated Abbott ARCHITECT c8000 analyzer.

Measurement of PCT in conjunction with other laboratory findings and clinical assessments aids in the risk assessment of critically ill patients on their first day of Intensive Care Unit (ICU) admission for progression to severe sepsis and septic shock.

The TruCal Procalcitonin Calibrator Set is intended for in vitro use for calibration of the DiaSys Procalcitonin FS assay.

TruLab Procalcitonin Bi-Level Controls are an assayed quality control material for monitoring the performance of quantitative in vitro determination of Procalcitonin (PCT) for the DiaSys Procalcitonin FS assay.

For in vitro diagnostic use only.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 5

510(k) Summary in accordance with 21 CFR 807.92

Page 1 of 11

(a) (1) Submitted by: DiaSys Diagnostic Systems GmbH
Alte Strasse 9
65558 Holzheim
Germany
Tel.: +49 6432 9146-0
Fax: +49 6432 9146-32
Jan.Gorka (at) diasys.de

Contact Person: Mr. Jan Gorka

Position/Title: CEO

Date of Preparation: May 9, 2025

(2) Trade Name: DiaSys Procalcitonin FS assay; DiaSys TruCaI Procalcitonin Calibrator Set; and DiaSys TruLab Procalcitonin Bi-Level Controls

Common/Classification Name: Device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis

Product Code(s): 21 CFR §866.3215; PTF (reagent)

Class: Class II

(3) Predicate Device(s): Substantial Equivalence to:

K NumberModelManufacturer
K071146VIDAS B.R.A.H.M.S. PCT ASSAYbioMérieux, Inc.

Reference Device(s):

K NumberModelManufacturer
K162297Diazyme Procalcitonin (pct) Assay, Diazyme Procalcitonin (pct) Calibrator Set, Diazyme Procalcitonin (pct) Control SetDiazyme Laboratories

Reason for Submission: New Device(s)

(4) Description of Device:

Page 6

Page 2 of 11

The DiaSys Procalcitonin FS assay is a particle enhanced turbidimetric immunoassay (PETIA) test for in vitro quantitative determination of procalcitonin (PCT) levels in human serum and lithium heparin plasma on automated photometric systems.

The test utilizes anti-human PCT polyclonal antibodies (goat) covalently bound to polystyrene particles.

PCT in the sample binds to the anti-PCT antibodies on the particles and causes agglutination. The degree of the turbidity caused by agglutination is optically measured at 660 nm by the automated photometric system and is proportional to the amount of PCT in the sample. DiaSys Procalcitonin FS Reagent specifies settings for an automated photometric analyzer, the Abbott ARCHITECT c8000.

The DiaSys TruCal Procalcitonin Calibrator Set is provided in 6 levels, traceable to a commercial assay for use in the calibration of the DiaSys Procalcitonin FS assay.

The DiaSys TruLab Procalcitonin Bi-Level Controls, Levels 1 (low) and 2 (high) are intended for use as quality control for the DiaSys Procalcitonin FS assay.

DiaSys PCT Calibrators and Bi-Level Controls are ready-to-use, stable, aqueous solutions containing recombinant human full-length PCT and biological additives from bovine origin.

(5) Intended use:

The intended use for the DiaSys Procalcitonin FS assay is the same as the predicate device: quantitative in vitro determination of procalcitonin (PCT) in human serum or lithium heparin plasma.

Indications for Use:

DiaSys Procalcitonin FS assay is a particle enhanced immunoturbidimetric test intended for the quantitative in vitro determination of procalcitonin (PCT) in human serum and lithium heparin plasma on automated Abbott ARCHITECT c8000 analyzer.

Measurement of PCT in conjunction with other laboratory findings and clinical assessments aids in the risk assessment of critically ill patients on their first day of Intensive Care Unit (ICU) admission for progression to severe sepsis and septic shock.

The TruCal Procalcitonin Calibrator Set is intended for in vitro use for calibration of the DiaSys Procalcitonin FS assay.

TruLab Procalcitonin Bi-Level Controls are an assayed quality control material for monitoring the performance of quantitative in vitro determination of Procalcitonin (PCT) for the DiaSys Procalcitonin FS assay.

For in vitro diagnostic use only.

(6) Technological Characteristics:

Page 7

Page 3 of 11

Comparison of Subject Intended Use and Technological Features to Predicate Device:

Product/FeatureSubject Device: DiaSys Procalcitonin FS AssayPredicate Device: VIDAS BRAHMS PCT AssayComment
ManufacturerDiaSys Diagnostic Systems GmbHbioMérieux, Inc.---
Model Number(s)Procalcitonin FS; Cat. No. 1 7318 99 43 018TruCal Procalcitonin Calibrator Set; Cat. No. 1 7310 99 43 082TruLab Bi-Level Controls Cat. No. 5 9980 99 43 086VIDAS ® B·R·A·H·M·S PCT™ (PCT) Cat. 30450-01 (60 tests)Kit (30450) contains C1/C2 Controls and S1/S2 CalibratorsEquivalent function - subject device separately specifies the calibrators and controls, the predicate device does not, however they are included in the reagent kit.
510(k) NumberK242294K071146---
Application/Intended use:Quantitative in vitro determination of procalcitonin (PCT) in human serum or Lithium heparin plasma on automated photometric systemDetermination of human procalcitonin in human serum or plasma (lithium heparin)Same
Analyte MeasuredProcalcitonin (PCT)Procalcitonin (PCT)Same
Environment of UseIVD: Clinical Laboratory use onlyIVD: Clinical Laboratory use onlySame

Page 8

Page 4 of 11

Product/FeatureSubject Device: DiaSys Procalcitonin FS AssayPredicate Device: VIDAS BRAHMS PCT AssayComment
Indications for UseDiaSys Procalcitonin FS assay is a particle enhanced immuno-turbidimetric test intended for the quantitative in vitro determination of procalcitonin (PCT) in human serum and lithium heparin plasma on automated Abbott ARCHITECT c8000 analyzer. Measurement of PCT in conjunction with other laboratory findings and clinical assessments aids in the risk assessment of critically ill patients on their first day of Intensive Care Unit (ICU) admission for progression to severe sepsis and septic shock.The TruCal Procalcitonin Calibrator Set is intended for in vitro use for calibration of the DiaSys Procalcitonin FS assay.TruLab Procalcitonin Bi-Level Controls are an assayed quality control material for monitoring the performance of quantitative in vitro determination of Procalcitonin (PCT) for the DiaSys Procalcitonin FS assay.For in vitro diagnostic use only.VIDAS BRAHMS PCT is an automated test for use on the VIDAS instruments for the determination of human procalcitonin in human serum or plasma (lithium heparin) using the ELFA (Enzyme-Linked Fluorescent Assay) technique. The VIDAS BRAHMS PCT assay is intended for use in conjunction with other laboratory findings and clinical assessments to aid in the risk assessment of critically ill patients on their first day of ICU admission, for progression to severe sepsis and septic shock.Equivalent function: Similar warnings to the effect that the PCT assay is not to be used as the sole determinant for diagnosis of sepsis, and should be used in conjunction with other clinical signs and symptoms.The subject device warns against the use as an aid in decision making on antibiotic therapy for patients. The predicate device warns that decisions regarding antibiotic therapy should NOT be based solely on procalcitonin concentrations.
Measurement Characteristics
Assay Principle(s)Particle enhanced immunoturbidimetric method – for use on automated photometric system(s)Immunoassay based on sandwich Enzyme-Linked Fluorescent Assay (ELFA) – for use on automated assayEquivalent function with some differences in assay principle: Immunoturbidimetric principle of subject assay is different from the predicate device (ELFA)
Specified Analyzer(s)ARCHITECT c8000™VIDAS and miniVIDAS family of analyzersEquivalent function on different automated chemistry analyzer(s)
Sample typeHuman serum or lithium heparin plasmaHuman serum or lithium heparin plasmaSame sample matrices

Page 9

Page 5 of 11

Product/FeatureSubject Device: DiaSys Procalcitonin FS AssayPredicate Device: VIDAS BRAHMS PCT AssayComment
Sample volume10 μL200 μLSubject device requires less sample volume
Time to obtain readings10 minutes20 minutesEquivalent function
Measurement range0.23 up to 50 ng/mL (dependent on value of highest calibrator)0.05 to 200 ng/mLEquivalent function: assays span the clinical reference range of 0.5 ng/ml to 2.0 ng/ml
Limit of Blank (LOB)0.081 ng/mL (max value of 4 series)Limit of Blank (LoB) 0.01 ng/mL (from insert, decision summary does not specify)Equivalent function: subject assay LOB is below the clinical decision range of 0.5 ng/mL
Limit of Quantitation (LOQ)0.23 ng/mL (%CV ≤ 20%)0.05 ng/mL (bias ≤ 10%, % CV ≤ 20%)Equivalent function: subject assay LOQ is below the clinical decision range of 0.5 ng/mL
Hook(Prozone) EffectNo prozone effect up to 1000 ng/mLNo hook effect found up to concentrations of 2600 ng/mlEquivalent function for clinical evaluation range
Precision/ ReproducibilityPrecision within acceptance criteria:CV ≤ 10% at 0.2-1.0 ng/mL (actual 5.64%);CV ≤ 7% at 1.0-3.0 ng/mL (actual 2.34%);CV ≤ 5% at 3.0-20 ng/mL (actual 1.93%)total precision: 6.17 - 15.31% CVPrecision values are expressed differently (CV, SD), however generally equivalent magnitude
Reagent
Reagent measurement volume120 μl Reagent 140 μl Reagent 2Self-contained reagents integrated into proprietary sample stripEquivalent function; different packaged volumes
Primary Reagent ConstituentsReagent 1: 0.1 mol/L Tris-buffer, pH 6.5Reagent 2: 0.1 mol/L Tris-buffer, pH 9.0; Polyclonal antibodies (goat) against human PCT covalently bound to polystyrene particles; Sodium azide (0.9 g/L) as preservativeSolid phase receptacle (SPR): Interior coated with mouse monoclonal anti-human procalcitonin immunoglobulins; proprietary reagent strip contains substrate (4-Methyl-umbelliferyl phosphate) and rinse agents for cycling in and out of the SPREquivalent function; different packaged reagent constituents; Subject device has two step reagent construction, predicate device has proprietary reagent strip.

Page 10

Page 6 of 11

Product/FeatureSubject Device: DiaSys Procalcitonin FS AssayPredicate Device: VIDAS BRAHMS PCT AssayComment
Calibrators and Controls
Calibrators: Number and Volume6 x 1 mL Six-level Calibrator2 x 2 mL two-level Calibrator S1 and S2Equivalent function: number of calibrators varies
Controls: Packaged Volume & Quantity6 x 1 mL Low Level6 x 1 mL High Level2 x 2.0 mL C12 x 2.0 mL C2Equivalent function

Considerations of 513(i)(1)(A) of the FD&C Act and 21 CFR 807.87(f)] include the following:

  • The intended use for the subject DiaSys Procalcitonin FS assay is the same as the predicate VIDAS BRAHMS PCT Assay: quantitative determination of procalcitonin (PCT) levels.
  • The subject DiaSys Procalcitonin FS assay specifies a particle enhanced immunoturbidimetric test method, the predicate device specifies the ELFA (Enzyme-Linked Fluorescent Assay) technique. The test methods provide equivalent results.
  • Different technological characteristics, where identified, do not raise different questions of safety and effectiveness than the predicate device.

Slight differences in the wording of the subject and predicate device indications for use are not critical to the intended use for assay Procalcitonin (PCT) and do not affect the safety and effectiveness of the device when used as labeled for the following reasons:

  • Differences in wording of the DiaSys Procalcitonin FS assay indications for use include the definitions of their test methods:
    • DiaSys Procalcitonin FS assay is a particle enhanced immuno-turbidimetric test intended for the quantitative in vitro determination of procalcitonin (PCT) in human serum and lithium heparin plasma on automated Abbott ARCHITECT c8000 analyzer.
    • VIDAS BRAHMS PCT is an automated test for use on the VIDAS instruments for the determination of human procalcitonin in human serum or plasma (lithium heparin) using the ELFA (Enzyme-Linked Fluorescent Assay) technique.
  • As described above and elsewhere in this document, the different assay methods provide equivalent function for determination of procalcitonin (PCT).

Therefore, in consideration of the above, the differences identified are not critical to the intended use of the device and do not affect the safety and effectiveness of the device when used as labeled.

(b) (1),(2) Summary of Studies

The DiaSys Procalcitonin (PCT) Assay(s) have been evaluated in accordance with current recognized standards/guidelines per the following:

Analytical Performance:

Precision/Reproducibility per the method(s) of CLSI EP05-A3 and EP15-A3, Precision and Estimation of Bias:
Total Precision

Page 11

Page 7 of 11

3 samples (human serum) with different values over the measuring range assayed in duplicate over 20 days using the Abbot ARCHITECT c8000 analyzer:
Reported Precision within acceptance criteria:
CV ≤ 10% at 0.2-1.0 ng/mL (actual 5.64%);
CV ≤ 7% at 1.0-3.0 ng/mL (actual 2.34%);
CV ≤ 5% at 3.0-20 ng/mL (actual 1.93%)

3 samples (human plasma) with different values over the measuring range assayed in duplicate over 20 days using the Abbot ARCHITECT c8000 analyzer:
Reported Precision within acceptance criteria:
CV ≤ 10% at 0.2-1.0 ng/mL (actual 8.31%);
CV ≤ 7% at 1.0-3.0 ng/mL (actual 3.24%);
CV ≤ 5% at 3.0-20 ng/mL (actual 2.61%)

Within-run Precision
Five samples (human serum) with values spanning the measurement and reference ranges assayed twenty times each using the Abbot ARCHITECT c8000 analyzer:
CV ≤ 10% at 0.252 ng/mL (actual 16.62%);
CV ≤ 10% at 0.555 ng/mL (actual 3.89%);
CV ≤ 7% at 2.02 ng/mL (actual 2.78%)
CV ≤ 5% at 9.68 ng/mL (actual 2.24%)
CV ≤ 5% at 18.8 ng/mL (actual 2.16%)

Five samples (human plasma) with values spanning the measurement and reference ranges assayed twenty times each using the Abbot ARCHITECT c8000 analyzer:
CV ≤ 10% at 0.314 ng/mL (actual 8.00%);
CV ≤ 10% at 0.475 ng/mL (actual 6.94%);
CV ≤ 7% at 1.910 ng/mL (actual 2.93%)
CV ≤ 5% at 9.851 ng/mL (actual 1.94%)
CV ≤ 5% at 17.953 ng/mL (actual 1.96%)

For all measured concentrations the CV is within acceptance criteria except for the CV for the human serum sample close to the assay LoQ (Limit of Quantitation), which is determined to be acceptable given its proximity to the LoQ.

Linearity/Assay Reportable Range per the method of CLSI EP06-A
Linearity in human serum and plasma is evaluated by assaying dilutions of stock solutions at least in quadruplicate. Mean values (y-axis) versus target values (x-axis) are plotted and linear regression is applied. Acceptance criteria were met:
Linearity < 0.5 ng/mL is with ± 0.1 ng/mL,
between 0.5 ng/mL to 5 ng/mL within ± 20%,
at > 5 ng/mL within ± 10%.

Detections Limits: LOB, LOD, LOQ per the method of CLSI EP17-A2
LOB and LOQ were evaluated in human serum and plasma. Acceptance criteria were LoB ≤ 0.2 ng/mL; LoQ ≤ 0.5 ng/mL – acceptance criteria were met as follows:
Limit of Blank (LOB): 0.081 ng/mL (max value of 4 series)

Page 12

Page 8 of 11

Limit of Detection (LOD): 0.23 ng/mL (LOD = LOQ) (%CV ≤ 20%)
Limit of Quantitation (LOQ) 0.23 ng/mL (%CV ≤ 20%)

Interference: per the method of CLSI EP07-A3
Sequential dilutions of the interfering substances are mixed in equal parts with aliquots of human samples at approximate PCT concentrations of 0.5 and 2.0 ng/mL, and each mixture is tested in five replicates. The mean measured PCT values (y-axis) are plotted against the concentration of the interfering substance (x-axis). No significant interference was observed, defined as a shift in PCT concentration greater than ±15%, when the following substances were tested at the concentrations indicated in the table below.

Interference byInterferent concentration tested
Ascorbic acid151 mg/dL
α-CGRP12 μg/mL
Azithromycin1.44 mg/dL
β-CGRP12 μg/mL
Bilirubin (conjugated)72.5 mg/dL
Bilirubin (unconjugated)71.4 mg/dL
Calcitonin10.8 ng/mL
Cefotaxime189 mg/dL
Cromolyn28.8 mg/L
Dobutamine22.9 μg/mL
Dopamine27.3 mg/dL
Doxycycline6.61 mg/dL
Enoxaparin24000 U/L
Ethanol720 mg/dL
Furosemide4.2 mg/dL
HAMA480 ng/mL
Hemolysis1200 mg/dL
Ibuprofen63.1 mg/dL
Imipenem2.52 mg/mL
Katacalcin6 ng/mL
Lipemia (triglycerides)1910 mg/dL
Noradrenalin4.2 μg/mL
Pantoprazole4.32 mg/dL
Rheumatoid factor1020 IU/mL
Salmeterol Xinafoate104 ng/mL

Page 13

Page 9 of 11

Interference byInterferent concentration tested
Scopolamine-N-butyl bromide72 mg/L
Total Protein139 g/L
Vancomycin3.78 mg/mL

N-Terminus interferes.

There is the possibility that other substances (e.g. ingestion, drugs or during specimen preparation) and medical conditions may lead to deviating results. Detailed information on every possible interfering substance or medical condition is beyond the scope of this script.

Stability per the method(s) of CLSI EP25-A
Real time stability
Real time stability testing of Procalcitonin FS reagent at 2 - 8°C was performed with three manufacturing lots - stability was checked by recovery of controls. Each lot was tested after manufacturing and subsequently within specific intervals. Recovery in % was calculated by dividing the value of the tested controls stored at 2 - 8°C by the target value, followed by multiplying the result by 100. Results of the continued real-time stability study confirm a stability of 25 months.

Results therefore support the stability claim of 24 months from the date of production.

In-use Stability
In-use stability of Procalcitonin FS reagent was checked by recovery of controls. One lot was tested after manufacturing and subsequently within specified intervals. After first opening, the tested controls were stored with closed cap at 2 - 8°C. Recovery in % was calculated by dividing the value of the tested controls stored at 2 - 8°C by the target, followed by multiplying the result by 100. Measured data fulfill the stability criteria for in-use stability. Results of the in-use stability study confirm a stability of at least 25 months.

Results support the in-use stability claim of 24 months from the date after opening.

Specimen Stability
Specimen stability testing were performed to the conditions specified on the DiaSys Procalcitonin FS reagent package insert:

  • 24 hours at 20-25°C
  • 5 days at 2-8°C
  • 14 days at -20°C

15 human samples covering the entire analytical range of the DiaSys Procalcitonin FS test were measured at pre-determined times using the specified Abbot ARCHITECT c8000 analyzer. Studies were performed for both sample matrices, human serum and lithium heparin plasma, and including measurements at least 10% over the specified durations. For each test series, the acceptance criteria was within ± 15% of the concentration at time zero, and the slope of the regression line is accepted if within ± 10%.

Page 14

Page 10 of 11

Test results for specimen stability for the three stated conditions met the acceptance criteria.

Stability Test – Additional Freeze-Thaw Cycle
As with the stability testing above, 15 human samples covering the analytical range of the Procalcitonin FS assay were measured using the specified Abbot ARCHITECT c8000 analyzer, at time 0, then after one freeze/thaw cycle, and again after a second freeze thaw cycle. Both serum and plasma samples were tested.

All samples tested at the two freeze/thaw cycles met the acceptance value of within 15% of the concentration at time zero.

Hook Effect
No hook/prozone effect up to 1000 ng/mL.

Comparison Studies
Method Comparison
Two method comparison studies were performed per the method of CLSI EP09-A3 was performed with predicate device – VIDAS Brahms PCT Assay (K071146) and the Abbott ARCHITECT c8000 analyzer.

First method comparison study [2021]
At least 100 native patient samples were measured two times with each method. Analysis of the mean patient recovery measured by validated method (x-axis) against mean patient recovery measured by evaluation method (y-axis) by Passing & Bablok.

Results met acceptance criteria of Slope: 0.85 – 1.15; Intercept: ± 1.0 ng/mL; Correlation coefficient R: ≥ 0.95; n ≥ 100 as follows:
N = 120
Slope = 1.08
Intercept = 0.105 ng/mL
R = 0.991

PCT CutoffNegative Percent Agreement (95% CI)Positive Percent Agreement (95% CI)
0.5 ng/mL93.9% (79.8% to 99.3%)100% (95.9% to 100%)
2.0 ng/mL93.3% (85.1% to 97.8%)97.8% (88.2% to 99.9%)

Second method comparison study [2025]
At least 200 native patient samples were measured two times with each method. Analysis of the mean patient recovery measured by validated method (x-axis) against mean patient recovery measured by evaluation method (y-axis) by Passing/Bablok.

Results met acceptance criteria of Slope: 0.85 – 1.15; Intercept: ± 1.0 ng/mL; Correlation coefficient R: ≥ 0.95; n ≥ 100 as follows:
N = 210
Slope = 0.940
Intercept = 0.017 ng/mL
R = 0.965

Page 15

Page 11 of 11

PCT CutoffNegative Percent Agreement (95% CI)Positive Percent Agreement (95% CI)
0.5 ng/mL98% (92.9% to 99.8%)100% (96.7% to 100%)
2.0 ng/mL100% (97.5% to 100%)94% (85.4% to 98.4%)

Matrix Comparison per the method of CLSI EP09-A3 was performed with human serum and lithium heparin plasma.
At least 20 native patient samples were measured two times for each matrix. Analysis of the mean patient recovery measured in serum matrix (x-axis) against mean patient recovery measured in heparin plasma matrix (y-axis) by Passing & Bablok.

Results of matrix study met acceptance criteria of Slope: 0.85 – 1.15; Intercept: ± 1.0 ng/mL; Correlation coefficient R ≥ 0.98; n ≥ 20.

FDA Special Controls
The DiaSys Procalcitonin (PCT) Assay(s) have been analyzed to the requirements of FDA Special Controls per 21 CFR § 866.3215.

Risk Management
Risk management has been performed in accordance with the following standard and includes the following risk management documentation:

  • ISO 14971:2019, Medical devices—Risk management—Application of risk management to medical devices
  • Procalcitonin (PCT) Assay Risk Analysis performed in accordance with ISO 14971
  • Procalcitonin (PCT) Assay FMEA

(3) Conclusions from Tests:

As described in (b)(1) and (b)(2) above, the DiaSys Procalcitonin FS assay is equivalent to the predicate device as supported by the referenced verification and validation testing.

The results of all tests demonstrate that the DiaSys Procalcitonin FS assay, DiaSys TruCaI Procalcitonin Calibrator Set; and DiaSys TruLab Procalcitonin Bi-Level Controls meet specified requirements for performance and substantial equivalence to the referenced predicate device.

§ 866.3215 Device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis.

Link to an amendment published at 90 FR 55982, Dec. 4, 2025. (a)
Identification. A device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis is identified as an in vitro device intended for the detection and qualitative and/or quantitative measurement of one or more non-microbial analytes in human clinical specimens to aid in the assessment of patients with suspected sepsis when used in conjunction with clinical signs and symptoms and other clinical and laboratory findings.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the device's detailed Indications for Use statement describing what the device detects and measures, the results provided to the user, whether the measure is qualitative and/or quantitative, the clinical indications for which the test is to be used, and the specific population(s) for which the device use is intended.
(2) Premarket notification submissions must include detailed documentation of the device description, including (as applicable), all device components, software, ancillary reagents required but not provided, explanation of the device principle and methodology, and for molecular devices include detailed documentation of the primer/probe sequence, design, and rationale for sequence selection.
(3) Premarket notification submissions must include detailed documentation of applicable analytical studies, such as, analytical sensitivity (Limit of Detection, Limit of Blank, and Limit of Quantitation), precision, reproducibility, analytical measuring range, interference, cross-reactivity, and specimen stability.
(4) Premarket notification submissions must include detailed documentation of a prospective clinical study or, if appropriate, results from an equivalent sample set. This detailed documentation must include the following information:
(i) Results must demonstrate adequate device performance relative to a well-accepted comparator.
(ii) Clinical sample results must demonstrate consistency of device output throughout the device measuring range likely to be encountered in the Intended Use population.
(iii) Clinical study documentation must include the original study protocol (including predefined statistical analysis plan), study report documenting support for the Indications for Use(s), and results of all statistical analyses.
(5) Premarket notification submissions must include evaluation of the level of the non-microbial analyte in asymptomatic patients with demographic characteristics (
e.g., age, racial, ethnic, and gender distribution) similar to the Intended Use population.(6) As part of the risk management activities performed under 21 CFR 820.30 design controls, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument.
(7) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling, and a detailed explanation of the interpretation of the limitations of the samples (
e.g., collected on day of diagnosis) must be included in the device's 21 CFR 809.10(b)(10) compliant labeling.