K Number
K242277

Validate with FDA (Live)

Device Name
Crown HT
Manufacturer
Date Cleared
2024-10-01

(61 days)

Product Code
Regulation Number
872.3690
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SprintRay Crown HT is a light-curable polymerizable resin intended to be used for the fabrication of; individual and fixed definitive full single crowns; definitive partial crowns in anterior and posterior area, individual and fixed single veneers; artificial teeth for dental prostheses, which are used for removable definitive full dentures; and individual and removable monolithic full and partial dentures in dental offices and laboratories. The material is an alternative to traditional restorative dental material.

Device Description

SprintRay Crown HT is a photo-polymer methacrylate resin material used in conjunction with a 3D printer and a scanned 3D image in a dental office to build dental prosthetics by 3D printing layer upon layer of the composite material. SprintRay Crown HT resin is offered in various shades such as Bleach. A1, A2, A3, and B1. Crown HT is an alternative to traditional dental prosthesis material that is intended exclusively for professional dental work.

SprintRay Crown HT resin is intended exclusively for professional dental work. Fabrication of dental prosthetics with SprintRay Crown HT resin requires computer-aided design and CAD/CAM manufacturing system that includes the following components not part of the device: Crown HT file created in an optical impression system, 3D printer, and curing light equipment.

Crown HT resin is designed to meet appropriate ISO standards for flexibility and sorption, to withstand prolonged use in the oral cavity. It is delivered non-sterile, and instructions on curing, and cleaning the final device prior to providing it to a patient are provided in the device instruction for use (IFU) document.

AI/ML Overview

The provided text describes the 510(k) summary for SprintRay Inc.'s "Crown HT" dental resin. It includes acceptance criteria for the device based on performance testing but does not describe a study involving human readers, ground truth consensus, or AI assistance for diagnostic purposes. The device being discussed is a material for fabricating dental prosthetics, not a diagnostic imaging AI system.

Therefore, many of the requested elements (e.g., sample size for test set, number of experts for ground truth, MRMC study, effect size of human reader improvement with AI, type of ground truth for AI, training set size for AI, ground truth establishment for AI training) are not applicable to this submission.

However, I can extract the relevant acceptance criteria and performance data for the material properties of the Crown HT resin, as well as information about the studies conducted to prove these properties.

Here's the information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally implied by the requirement to perform testing based on ISO standards and to show comparable or better performance than the predicate device.

Acceptance Criteria (from Predicate/ISO Standards)Reported Crown HT Device Performance
Material Properties:
Flexural Strength (>50.0 MPa)>124 MPa
Flexural Modulus>7000 MPa
Sorption (<40 µg/mm³)11.15 ± 2.5 µg/mm³
Solubility (<7.5 µg/mm³)0.49 ± 0.59 µg/mm³
Monomer Methyl Methacrylate (<2.2S%)Not detectable
Biocompatibility (based on ISO 10993 Series):
Genotoxicity, carcinogenicity, reproductive toxicityTested (functional as intended)
In Vitro CytotoxicityTested (functional as intended)
Systemic ToxicityTested (functional as intended)
Skin SensitizationTested (functional as intended)
Skin Irritation (ISO 10993-23)Tested (functional as intended)
Additional Bench Testing:
Radio Opacity (ISO-4049)Tested (functional as intended)
Print Accuracy and dimension stability TestTested (functional as intended)
Surface Finish (ISO 10477)Tested (functional as intended)
Shear Bond Strength (ISO 10477)Tested (functional as intended)
TranslucencyTested (functional as intended)
Depth of CureTested (functional as intended)
ViscosityTested (functional as intended)
ShadeTested (functional as intended)
Material ReuseTested (functional as intended)

2. Sample size used for the test set and the data provenance

  • Sample Size for Testing: The specific sample sizes for each test (e.g., number of specimens for flexural strength) are not explicitly stated in the provided text. The text only mentions that "samples of the Crown HT resin that had been formed into dental appliances using a 3D printer" were used.
  • Data Provenance: Not specified in terms of country of origin. The studies are described as "Biocompatibility testing" and "Additional following bench testing." It is implied that these were conducted by or on behalf of SprintRay Inc. as part of their 510(k) submission. No indication of retrospective or prospective human subject data, as this is a material property study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not Applicable. This submission is for a dental material, not a diagnostic device requiring expert interpretation of medical images or clinical data for ground truth establishment. The "ground truth" for this device relates to its physical, chemical, and biological properties, which are measured using established laboratory methods and validated against ISO standards.

4. Adjudication method for the test set

  • Not Applicable. As per point 3, this is not a study requiring adjudication of expert interpretations.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This device is a dental material, not an AI-enabled diagnostic tool requiring MRMC studies.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Partially Applicable (in a different context): The performance data presented (e.g., flexural strength, sorption) represents the standalone performance of the material itself when fabricated into a dental appliance. There is no algorithm being tested. The material's properties are measured independently of a human "in the loop" for diagnostic purposes.

7. The type of ground truth used

  • Not Applicable (in the requested context of AI/diagnostics).
  • For material testing: The "ground truth" is defined by the objective measurements obtained through standardized laboratory tests (e.g., ISO 10477, ISO 10993 series) for physical, chemical, and biological properties. These standards define the methods and acceptance limits that establish whether the material meets its intended performance requirements.

8. The sample size for the training set

  • Not Applicable. This is not an AI/machine learning device that requires a "training set."

9. How the ground truth for the training set was established

  • Not Applicable. As per point 8, no training set was used.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 1, 2024

SprintRay Inc. Sara Moghtadernejad Regulatory Affairs 2710 Media Center Drive. Suite 100A Los Angeles, California 90065

Re: K242277

Trade/Device Name: Crown HT Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF, EBI, EBG, ELM, PZY Dated: August 1, 2024 Received: August 1, 2024

Dear Sara Moghtadernejad:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael E. Adjodha -S

Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K242277

Device Name

Crown HT

Indications for Use (Describe)

SprintRay Crown HT is a light-curable polymerizable resin intended to be used for the fabrication of; individual and fixed definitive full single crowns; definitive partial crowns in anterior and posterior area, individual and fixed single veneers; artificial teeth for dental prostheses, which are used for removable definitive full dentures; and individual and removable monolithic full and partial dentures in dental offices and laboratories. The material is an alternative to traditional restorative dental material.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K242277

510(k) SUMMARY

Crown HT

Submitter:SprintRay Inc.2710 Media Center Drive, Suite 100ALos Angeles, CA 90065
--------------------------------------------------------------------------------------------

(800) 914-8004 Phone:

Contact Person: Sara Moghtadernejad

Date Prepared: August 01, 2024

Trade/Device Name:Crown HT
Common or Usual Name:Tooth Shade Resin Material
Regulation Number:21 CFR 872.3690
Regulation Name:Tooth Shade Resin Material
Regulatory Class:Class II
Product Code:EBF
Secondar Product Code:EBI, ELM
Predicate DevicesDigital Crown (K222623)

Device Description

SprintRay Crown HT is a photo-polymer methacrylate resin material used in conjunction with a 3D printer and a scanned 3D image in a dental office to build dental prosthetics by 3D printing layer upon layer of the composite material. SprintRay Crown HT resin is offered in various shades such as Bleach. A1, A2, A3, and B1. Crown HT is an alternative to traditional dental prosthesis material that is intended exclusively for professional dental work.

SprintRay Crown HT resin is intended exclusively for professional dental work. Fabrication of dental prosthetics with SprintRay Crown HT resin requires computer-aided design and CAD/CAM manufacturing system that includes the following components not part of the device: Crown HT file created in an optical impression system, 3D printer, and curing light equipment.

Crown HT resin is designed to meet appropriate ISO standards for flexibility and sorption, to withstand prolonged use in the oral cavity. It is delivered non-sterile, and instructions on curing, and cleaning the final device prior to providing it to a patient are provided in the device instruction for use (IFU) document.

Intended Use / Indications for Use

SprintRay Crown HT is a light-curable polymerizable resin intended to be used for the fabrication of; individual and fixed definitive full single crowns; definitive

{5}------------------------------------------------

partial crowns in anterior and posterior area, individual and fixed single veneers; artificial teeth for dental prostheses, which are used for removable definitive full dentures; and individual and removable monolithic full and partial dentures in dental offices and laboratories. The material is an alternative to traditional restorative dental material.

Summary of Technological Characteristics

Light-based curing of a 3D printed acrylate resin is the technological principle for both the subject and predicate devices. The Crown HT resin is poured into a 3D printer, which relies on scanned images of the patient's oral cavity to produce a dental appliance. At a high level, the subject and predicate devices are based on the following same technological elements:

  • are a pourable acrylate resin .
  • are used in conjunction with 3D printers, which rely on common 3D . images to define the fabricated dental appliance
  • . are cured prior to final trimming and cleaning
  • . are used for the fabrication of orthodontic and dental appliances

The following technological differences exist between the subject and predicate devices:

  • · differences in acrylate resin material composition

Performance Data

The subject device is considered tissue contacting for a period longer than 30 Days. The following Biocompatibility testing was performed on samples of the Crown HT resin that had been formed into dental appliances using a 3D printer in accordance with the FDA Blue Book Memorandum #G95-1 and International Standard ISO 10993-1 and ISO 7405, as recognized by FDA.

  • ISO 10993-3 Genotoxicity, carcinogenicity and reproductive toxicity ●
  • ISO 10993-5 In Vitro Cytotoxicity ●
  • ISO 10993-11 Systemic Toxicity ●
  • ISO 10993-10 skin sensitization ●
  • ISO 10993-23 skin irritation ●

Additional following bench testing based on the test steps laid out in ISO 10477 was performed using dental appliance fabricated from Crown HT resin.

  • Ultimate Flexural Strength (ISO 10477) ●
  • . Water Sorption and Water Solubility (ISO 10477)
  • Radio Opacity (ISO-4049)
  • Print Accuracy and dimension stability Test ●
  • Surface Finish (ISO 10477) ●
  • Shear Bond Strength (ISO 10477)
  • Translucency
  • . Free Monomer Extraction (ISO 20795-1)
  • Tolerance
  • Depth of Cure ●
  • Viscosity
  • Shade ●
  • Material Reuse

{6}------------------------------------------------

SprintRay, Inc.'s Crown HT resin Substantial Equivalence Chart

Digital Crown (Predicate Device)Crown HT (Subject Device)
Intended Use & Indicationsfor UseSprintRay Digital Crown is a light-curable polymerizable resin intended tobe used for the fabrication of; individualand fixed definitive full single crowns;definitive partial crowns in anterior andposterior area, individual and fixedsingle veneers; artificial teeth for dentalprostheses, which are used forremovable definitive full dentures; andindividual and removable monolithic fulland partial dentures in dental officesand laboratories. The material is analternative to traditional restorativedental material.SprintRay Crown HT is a light- curablepolymerizable resin intended to beused for the fabrication of; individualand fixed definitive full single crowns;definitive partial crowns in anterior andposterior area, individual and fixedsingle veneers; artificial teeth for dentalprostheses, which are used forremovable definitive full dentures; andindividual and removable monolithicfull and partial dentures in dentaloffices and laboratories. The materialis an alternative to traditionalrestorative dental material.
User PopulationClinicians in dental offices providedpermanent partial crowns and artificialteeth as indicated in theabove IFU for patients.Same
Material TypeLight-curable polymerizable resinSame
AccessoriesSprintRay Commercially available3D printersSame
Volume Provided250g bottle250g bottle or 4ml Midas Capsule
Shelf Life1.5 years1.5 years
BiocompatibilityTested to ISO-10993-1, -3, -5, -10and -11Tested to ISO-10993-1, -3, -5, -10,-11, and 23
Performance TestingTesting performed using methods laidout in ISO- 10477Same
Flexural Strength (>50.0 MPa)125-174 MPa>124 MPa
Flexural Modulus6116-7714MPa>7000 MPa
Sorption (<40 µg/mm³)17.35 ± 2.56 µg/mm311.15 ± 2.5 µg/mm3
Solubility (<7.5 µg/mm³)2.16 ±1.30 µg/mm30.49 ±0.59 µg/mm3
Monomer MethylMethacrylate (<2.2S%)Not detectableNot detectable

In all instances, the Crown HT resin functioned as intended and the outcomes were as expected.

Conclusions

The Crown HT resin is as safe and effective as its predicate device. The Crown HT resin has the same intended use and indication, and similar technological characteristics, and principles of operation as its predicate device. The minor technological differences between the Crown HT resin and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the Crown HT resin is as safe and effective as the predicate device. Thus, the Crown HT resin is substantially equivalent.

§ 872.3690 Tooth shade resin material.

(a)
Identification. Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.(b)
Classification. Class II.