K Number
K242266

Validate with FDA (Live)

Date Cleared
2024-10-21

(81 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PhotonBlade 3 is a monopolar radiofrequency (RF) device coupled with illumination that is indicated for cutting and coagulation of soft tissue during general, plastic and reconstructive (including but not limited to skin incisions and development of skin flaps), ENT, gynecologic, orthopaedic, arthroscopic, spinal, neurological, and breast procedures (including mastectomy and lumpectomy).

The PhotonBlade 3 Smoke Evacuation device is a monopolar RF device coupled with illumination that is indicated for cutting and coagulation of soft tissue during general, plastic and reconstructive (including but not limited to skin incisions and development of skin flaps), ENT, gynecologic, orthopaedic, arthroscopic, spinal, neurological, and breast procedures (including mastectomy and lumpectomy), and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system.

Device Description

The PhotonBlade 3 (Model No: 0730-101-000) and PhotonBlade 3 Smoke Evacuation (Model No: 0730-102-000) devices are sterile, single use, electrosurgical devices have a monopolar electrode at the distal end, which delivers radiofrequency (RF) energy for cutting and coagulation of tissue. The electrode is located at the distal end of a rotatable and extendable shaft. The devices have integrated LED-based illumination. In addition, the PhotonBlade 3 Smoke Evacuation features integrated smoke evacuation tubing for removal of smoke generated by electrosurgery when connected to a smoke evacuation universal connector attaches the device to an IEC 60601 compliant electrosurgical generator.

AI/ML Overview

The provided text is a 510(k) Summary for the Stryker Instruments PhotonBlade 3 and PhotonBlade 3 Smoke Evacuation devices. It discusses the regulatory clearance process and comparison to predicate devices, but does not contain the detailed information required to answer your specific questions regarding acceptance criteria and a study proving the device meets those criteria.

While it mentions "pre-defined acceptance criteria for the above tests were met" in the "Performance Testing" section, it does not explicitly state what those acceptance criteria were for each test, nor does it provide a detailed study design or results from such a study.

Therefore, I cannot extract the information requested. The document primarily focuses on establishing "substantial equivalence" to a predicate device based on similar indications for use and technological characteristics, and by referencing compliance with various standards and general performance testing without providing the specific details of a study that directly proves the device meets predefined acceptance criteria for its performance metrics.

To specifically address your questions, the following would need to be present in the document:

  1. A table of acceptance criteria and reported device performance: This would require a table explicitly listing criteria (e.g., "minimum lumen output 50 lumens") and the actual measured performance (e.g., "Device A measured 65 lumens").
  2. Sample size and data provenance for the test set: The document vaguely mentions "simulated use life testing," "coag verification test," etc., but no sample sizes or origin of data (e.g., animal, human, in vitro, country).
  3. Number and qualifications of experts for ground truth: This type of information is usually relevant for studies involving subjective assessment (e.g., image interpretation). For a device like an electrosurgical unit, ground truth often comes from objective measurements or established physical/biological parameters. The document does not provide this.
  4. Adjudication method for the test set: Not applicable based on the type of device and testing described.
  5. MRMC comparative effectiveness study: Not conducted or mentioned. This type of study is typically for diagnostic aids involving human interpretation.
  6. Standalone performance: The described "performance testing" implies standalone testing of the device's physical and electrical characteristics, but the specific metrics and study details are missing.
  7. Type of ground truth used: Ground truth would be based on objective measurements against established engineering, physical, and biological standards (e.g., thermal spread in porcine tissue). The document describes what was tested but not the specific "ground truth" values used for comparison or how they were established.
  8. Sample size for the training set: Not applicable, as this is not an AI/ML device that requires a training set.
  9. How ground truth for the training set was established: Not applicable.

The closest the document comes to providing performance data related to "acceptance criteria" is under "Performance Testing," which states: "All pre-defined acceptance criteria for the above tests were met. Results from the the subject devices perform as intended and support a determination of substantial equivalence to the predicate." However, it fails to specify what those acceptance criteria were or the actual results that met them.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration" in a stacked format.

October 21, 2024

Stryker Instruments Miriam Giltinan Senior Staff Regulatory Affairs Specialist 1941 Stryker Way Portage, Michigan 49002

Re: K242266

Trade/Device Name: PhotonBlade 3; PhotonBlade 3 Smoke Evacuation Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: August 1, 2024 Received: August 1, 2024

Dear Miriam Giltinan:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Long H. Chen--> Digitally signed by Long H.Chen-S
Date: 2024.10.21 12:58:00 -04'00'

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K242266

Device Name

PhotonBlade 3;

PhotonBlade 3 Smoke Evacuation

Indications for Use (Describe)

The PhotonBlade 3 is a monopolar radiofrequency (RF) device coupled with illumination that is indicated for cutting and coagulation of soft tissue during general, plastic and reconstructive (including but not limited to skin incisions and development of skin flaps). ENT, avnecologic, orthopaedic, arthroscopic, spinal, neurological, and breast procedures (including mastectomy and lumpectomy).

The PhotonBlade 3 Smoke Evacuation device is a monopolar RF device coupled with illumination that is indicated for cutting and coaqulation of soft tissue during general, plastic and reconstructive (including but not limited to skin incisions and development of skin flaps), ENT, gynecologic, orthopaedic, arthroscopic, spinal, neurological, and breast procedures (including mastectorny and lumpectomy), and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

Submitter Information

510(k) Owner/Submitter:Stryker Instruments1941 Stryker WayPortage, MI 49002USA
Contact Person:Miriam GiltinanSenior Staff RA SpecialistEmail: miriam.giltinan@stryker.com
Registration Number:3015967359
Date Summary Prepared:August 01, 2024

Device Name

Trade Name(s)/Model number(s):PhotonBlade 3/0730-101-000
PhotonBlade 3 Smoke Evacuation/0730-102-000
Classification Name:Electrosurgical cutting and coagulation device andaccessories (21 CFR 878.4400)
Classification Panel:General & Plastic Surgery
Device Class:Class 2
Product Code:GEI

Predicate Device

Predicate: PlasmaBlade X3.0S Light (K193579)

Reference: PhotonBlade with Adaptive Smoke Evacuation (K191583)

Device Description

The PhotonBlade 3 (Model No: 0730-101-000) and PhotonBlade 3 Smoke Evacuation (Model No: 0730-102-000) devices are sterile, single use, electrosurgical devices have a monopolar electrode at the distal end, which delivers radiofrequency (RF) energy for cutting and coagulation of tissue. The electrode is located at the distal end of a rotatable and extendable shaft. The devices have integrated LED-based illumination. In addition, the PhotonBlade 3 Smoke Evacuation features integrated smoke evacuation tubing for removal of smoke generated by electrosurgery when connected to a smoke evacuation universal connector attaches the device to an IEC 60601 compliant electrosurgical generator.

{5}------------------------------------------------

Indications for Use

The PhotonBlade 3 is a monopolar radiofrequency (RF) device coupled with illumination that is indicated for cutting and coagulation of soft tissue during general, plastic and reconstructive (including but not limited to skin incisions and development of skin flaps), ENT, gynecologic, arthroscopic, spinal, neurological, and breast procedures (including mastectomy and lumpectomy).

The PhotonBlade 3 Smoke Evacuation device is a monopolar RF device coupled with illumination that is indicated for cutting and coagulation of soft tissue during general, plastic and reconstructive (including but not limited to skin incisions and development of skin flaps), ENT, gynecologic, arthroscopic, spinal, neurological, and breast procedures (including mastectomy and lumpectomy), and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system.

Substantial Equivalence Comparison

The predicate device has been identified as the PlasmaBlade X3.0S Light (K193579). Both subject and predicate devices share the same intended use, and similar indications for use and technological characteristics.

Intended Use Comparison

PhotonBlade 3PhotonBlade 3 SmokeEvacuationPlasmaBladeLightX3.0S
Subject DeviceSubject DevicePredicate Device
ClassificationClass 2Class 2
Product CodeGEIGEI
Regulation21 CFR 878.440021 CFR 878.4400
Regulation NameElectrosurgical cutting and coagulation device and accessoriesElectrosurgical cutting and coagulation device and accessories
Indications for UseThe PhotonBlade 3 is a monopolar radiofrequency (RF) device coupled with illumination that is indicated for cutting and coagulation of soft tissue during general, plastic and reconstructive (including but not limited to skin incisions and development of skin flaps), ENT, gynecologic, orthopaedic, arthroscopic, spinal, neurological, and breast proceduresThe PhotonBlade 3 Smoke Evacuation device is a monopolar RF device coupled with illumination that is indicated for cutting and coagulation of soft tissue during general, plastic and reconstructive (including but not limited to skin incisions and development of skin flaps), ENT, gynecologic, orthopaedic, arthroscopic, spinal, neurological, and breast proceduresThe surgery system is indicated for cutting and coagulation of soft tissue during general, plastic and reconstructive (including but not limited to skin incisions and development of skin flaps), ENT, gynecologic, orthopaedic, arthroscopic, spinal and neurological procedures.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase and black. The letters are closely spaced together, giving the word a solid, block-like appearance.

PhotonBlade 3PhotonBlade 3 Smoke EvacuationPlasmaBlade Light
Subject DeviceSubject DevicePredicate Device
(including mastectomyand lumpectomy).(including mastectomyand lumpectomy), and forremoving smokegenerated byelectrosurgery when usedin conjunction with aneffective smokeevacuation system.
Type of UseRx onlyRx only
Conditions of UseSingle UseSingle Use

The subject and predicate devices share identication regulation and intended use. The subject devices add an indication for "breast procedures (including mastectomy and lumpectomy)" which falls within the intended use of the predicate device. The indication is supported by a systematic literature review.

The purpose of the systematic literature review was to demonstrate the safety and efficacy of monopolar RF devices for the indication "breast procedures (including mastectomy and lumpectomy)". A thorough review of the literature explored 1) the medical field and targeted treated populations applicable for the subject devices as well as those of the predicate and other relevant monopolar RF devices; 2) clinical data for predicate and other relevant monopolar RF devices; and 3) vigilance databases to assess safety. Performance was assessed using reported outcomes including drain output, flap perfusion, thermal tissue injury, length of hospital stay, and procedural success. Safety was assessed by reviewing adverse events/complications reported in both peer-reviewed publications and vigilance databases and sought to identify events that were specifically related to the device or procedure.

Stryker identified a total of 63 unique publications utilizing search criteria specific to predefined indications. These were screened for inclusion using predetermined criteria and a weighted appraisal scoring system. In addition to vigilance databases, a thorough full-text review of 10 manuscripts was performed providing supportive clinical efficacy and safety data that covered the stated indication. An additional 34 publications describing relevant safety information (not specific to the breast indication) were included for a comprehensive product safety assessment.

Results from the systematic literature review concluded that monopolar RF devices are safe and effective for use in the breast surgery indications. No new or unexpected risks were identified.

The data supports the assertion that the indication for "breast procedures (including mastectomy and lumpectorny)" falls within the intended use (cutting and coagulation of soft tissue) of the predicate and does not introduce new questions of safety and effectiveness.

{7}------------------------------------------------

K242266

Image /page/7/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are black and the background is white. The word is written in all lowercase letters.

Technological Characteristics Comparison

PhotonBlade 3PhotonBlade 3 Smoke EvacuationPlasmaBlade LightX3.0S
Subject DeviceSubject DevicePredicate Device
Mechanism of action/Principle of operationRF Monopolar EnergyRF Monopolar Energy
Compatibility with Electrosurgical Unit (ESU)Standard ESU complying to IEC 60601 standardsDedicated ESU
Illumination SourceLEDLED
Illumination LocationDistal ShaftMain handpiece body
Illumination Power Source2x3V batteryUSB plug straight to generator
Light ColorWhiteWhite
Smoke Evacuation DesignN/AIntegratedIntegrated
Smoke Evacuation Mode of OperationN/AConnect to commercially available smoke evacuation unit or suction/vacuum sourceConnect to commercially available smoke evacuation unit or suction/vacuum source
SterilizationSterile, single use (EtO)Sterile, single use (EtO)
Type of Patient ContactExternal Communicating Device <24 hours, tested to ISO 10993-1 (Limited contact)External Communicating Device <24 hours, tested to ISO 10993-1 (Limited contact)
Blade MaterialCoated Stainless SteelCoated Stainless Steel

The subject and predicate devices share or similar fundamental technological characteristics:

  • Mechanism of action/principle of operation RF Monopolar Energy .
  • . Electrode - Coated Stainless Steel
  • Patient Contact - Limited contact
  • Illumination LED ●
  • Configuration sterile (EO), single use ●
  • . Smoke Evacuation (applicable for PhotonBlade 3 Smoke Evacuation only) - Integrated smoke evacuation design.

Differences in technological characteristics between subject and predicate devices were noted for:

  • . Illumination location and power source – The location of illumination is at the distal shaft for the subject devices compared to the predicate device where it is located in the main handpiece body. The source of illumination for the subject devices is by 2 x 3V batteries compared to the predicate device which uses a USB plug to the generator.
    The reference device (PhotonBlade with Adaptive Smoke Evacuation, K191583) also contains illumination at the distal shaft which is powered by a 3V battery.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are black and the background is white. The word is written in all lowercase letters.

  • Compatibility with ESU The subject devices are to be connected with any IEC 60601 compliant ESU. ● The predicate device is designed to be used with its own dedicated ESU.
    Results of verification and validation demonstrate that these differences do not introduce new questions of safety and effectiveness. The subject devices are at least as safe and effective as the predicate device.

Summary of Non-Clinical Testing

A suite of performance testing was conducted to demonstrate substantial equivalence with the predicate device. Testing requirements were determined by relevant standards and FDA Guidance Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery, issued March 9, 2020.

Sterilization, Biocompatibility, Electrical Safety and EMC Testing were conducted to the following standards:

StandardTitle
ISO 11135:2014Sterilization of health-care products - Ethylene oxide -Requirements for the development, validation and routine controlof a sterilization process for medical devices
ISO 10993-7:2008Biological evaluation of medical devices - Part 7: Ethylene oxidesterilization residuals
ISO 11737-1: 2018Sterilization of health care products. Microbiological methods.Determination of a population of microorganisms on products
ISO 11737-2: 2019Sterilization of medical devices. Microbiological methods. Testsof sterility performed in the definition, validation andmaintenance of a sterilization process
ISO 10993-5:2009Biological evaluation of medical devices Part 5: Tests for in vitrocytotoxicity
ISO 10993-10:2021Biological evaluation of medical devices Part 10: Tests for skinsensitization
ISO 10993-23:2021Biological evaluation of medical devices Part 23: Tests forirritation
ISO 10993-11:2017Biological evaluation of medical devices Part 11: Tests forsystemic toxicity
ANSI AAMI ES60601-1:2005/(R)2012 &A1:2012, C1:2009/(R)2012 &A2:2010/(R)2012 (Cons. Text) [Incl.AMD2:2021]Medical electrical equipment - Part 1: General requirements forbasic safety and essential performance (IEC 60601-1:2005,MOD) [Including Amendment 2 (2021)]
ANSI AAMI IEC 60601-1-2:2014[Including AMD 1:2021]Medical electrical equipment - Part 1-2: General requirements forbasic safety and essential performance - Collateral Standard:Electromagnetic disturbances - Requirements and tests [IncludingAmendment 1 (2021)]
ANSI AAMI IEC 60601-2-2: 2017Medical electrical equipment - Part 2-2: Particular requirementsfor the basic safety and essential performance of high frequencysurgical equipment and high frequency surgical accessories
AIM Standard 7351731Rev. 3.00 2021-06-04Medical Electrical Equipment and System ElectromagneticImmunity Test for Exposure to Radio Frequency IdentificationReaders - An AIM Standard

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are all lowercase, and the word is black. The background is white. The font is a custom font that is used by Stryker Corporation, a medical technology company.

Performance Testing

A range of performance testing was conducted to confirm the subject devices perform as intended and are safe and effective for their intended use. The testing included:

  • Simulated Use Life Testing to verify that the subject devices perform as intended during use. ●
  • Coag Verification Test to verify that the subject devices are effective in spot coagulation.
  • . Telescope Locking Force Test to verify that the subject devices' telescoping mechanism locks and does not extend or collapse under reasonable force.
  • Telescope Bending Test to verify that the telescope of the subject devices withstands reasonable bending forces.
  • Light (Lumen) Test to verify that the subject devices provide adequate brightness of lighting.
  • Blade Bending Test to verify that the subject devices meet the pre-defined electrode bending criteria. ●
  • Smoke Evacuation Test (for PhotonBlade 3 Smoke Evacuation only) to verify that the device evacuates . smoke.
  • . Thermal Spread Testing to evaluate the penetrating thermal damage of the subject devices. Thermal Spread testing (ex vivo) was completed across three porcine tissue types (muscle, skin and liver tissue) in accordance with FDA Guidance Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery, issued March 9, 2020.

All pre-defined acceptance criteria for the above tests were met. Results from the the subject devices perform as intended and support a determination of substantial equivalence to the predicate.

Usability/Human Factors

The subject devices demonstrated safe and effective use for the intended use, and use environments in accordance with FDA Guidance "Applying Human Factors and Usability Engineering to Medical Devices", issued February 3, 2016.

Conclusion

The subject devices, in comparison with the legally marketed predicate, share the same intended use and operating principles. Performance testing demonstrate that the subject devices are and effective as the predicate. Any differences between the subject and predicate devices do not raise any new or different types of questions of safety and effectiveness. A determination of substantial equivalence is supported.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.