K Number
K242258
Date Cleared
2024-11-26

(118 days)

Product Code
Regulation Number
892.1000
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Whole body magnetic resonance diagnostic medical device intended for the production of images for diagnostic purposes with transverse, coronal, sagittal and obligue orientation relating to the internal structures of the examined body areas, also including the examination of the brain through the DWI sequence, with limitations regarding cardiac, breast, lung and upper abdominal imaging.

Device Description

MROpen Evo is a whole body magnetic resonance diagnostic medical device that acquires high-resolution medical images of various parts of the body. In particular, it produces images of the entire body except: upper abdomen, heart, lungs and breasts with transverse, coronal, sagittal and oblique orientation relating to the internal structures of the examined body areas, also including the examination of the brain through the DWI sequence. The standard composition of the MROpen Evo device is: Magnetic Unit Operator PC- Operator PC - Electronic Cabinet - Penetration Filters - Major Software MR GUI-PRO 3.1. Coil set (Body MAX, Hand MAX, Cervical MAX, Knee MAX, Multipurpose Linear, Flat Flex, Spine, Head-Neck, Long Spine, Shoulder) Cushions' Set The device consists of two models. 03-2000-00. The 03-4000-00 model of the MROpen Evo device has a very similar composition to the 03-2000-00 model, but most of the subassemblies that constitute it are replaced with others of equal functionality, in order to reduce production costs. The MROpen Evo device allows the execution of standing imaging for those cases that require this need or facilitates acquisitions on patients suffering from movement disorders or suffering from claustrophobia. It is also possible to undergo this investigation in patients who have a high weight, but in any case, not exceeding 200 kg. The software called MR GU-PRO represents the set of SW modules necessary for the MROpen Evo MR scanner based on the I-box spectrometer. The main modules act as a graphic interface for the execution of the exam. The MR GUI-PRO software has external interface elements, on the one hand the operator who interacts via monitor, keyboard and mouse and the other HW modules of the system and mainly: Gradient Amplifier, RF reception signal amplifier.

AI/ML Overview

The provided text is a 510(k) summary for the MROpen Evo device. It states that the MROpen Evo is "substantially equivalent" to a predicate device, MROpen EVO (K193116). The document describes changes in the subject device, primarily related to an updated spectrometer (I-box 07), a new cryocooler, and the introduction of a "Compressed Sensing" (CS) algorithm for faster MRI acquisition. The acceptance criteria and the study that proves the device meets the acceptance criteria are mainly focused on demonstrating this substantial equivalence.

Here's an breakdown of the information requested, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Stated or Implied for Equivalence)Reported Device Performance and Evidence
Medical Checklist Pass Criteria: Absence of unexpected artifacts, demonstrating equivalence across configurations.Images rated similar when comparing the three configurations (I-box 07 with CS, I-box 07 without CS, and the predicate's configuration). No artifacts were found due to new characteristics.
Software Performance: Limited reconstruction errors and no introduction of artifacts compared to standard processing."CS software module shows limited recontruction errors and does not introduce artefacts in image processing that does not use this reconstruction algorithm." Analysis showed predicate and MROpen Evo software performances are substantially equivalent.
Biocompatibility: Not cytotoxic, not irritant, not sensitizing.Testing results confirmed the samples in their finished form are: Not cytotoxic, Not irritant, Not sensitizing.
Electrical Safety and EMC: Compliance with IEC 60601-1, IEC 60601-1-2, and IEC 62471.Testing results are positive; no "fail" result appears on the test reports.
Software Verification and Validation: Compliance with IEC 62304 and FDA guidance for software.All test results are "Passed." Analysis showed predicate and MROpen Evo software performances are substantially equivalent.
Performance and Safety: Equivalent acoustic noise, unchanged electromagnetic characteristic of inner part of the magnetic unit, equivalent gradient output, equivalent Specific Absorption Rate (SAR), equivalent dynamic forces on patient support.Analysis of performance data demonstrates that the legally marketed Predicate Device performances are substantially equivalent for the assessed endpoints.
Usability: Compliance with IEC 60601-1-6 and IEC 62366-1.Testing results are positive; no "fail" result appears on the test reports. This demonstrates safety, effectiveness, and substantial equivalence.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated as a number of patients or cases. The "medical checklist" and "in vivo" testing for the Compressed Sensing algorithm are mentioned. For the in vivo testing, it states "comparing the and without undersampling," implying a comparison of images. It does not provide the number of patients or images used for this comparison.
  • Data Provenance: Not explicitly stated (e.g., country of origin). The testing seems to be internal to ASG Superconductors S.p.A. The "in vivo" testing suggests prospective data collection in a clinical setting, but the details are sparse.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • This information is not provided in the document. The evaluation of image similarity and artifact absence was likely performed by qualified personnel, but their number and specific qualifications are not specified.

4. Adjudication Method for the Test Set

  • This information is not provided in the document. The process for rating images and determining "similarity" or "absence of artifacts" is not detailed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a MRMC comparative effectiveness study is not explicitly mentioned. The document focuses on demonstrating "substantial equivalence" of the device itself (hardware and software), rather than a comparative effectiveness study of human readers with vs. without AI assistance. The "in vivo" testing compares images produced by the device under different configurations, not human reader performance.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • Yes, a form of standalone performance evaluation was conducted for the Compressed Sensing (CS) algorithm. The text states:
    • "Compressed Sensing algorithm was tested in vivo, comparing the and without undersampling. The images were rated similar when comparing the three configurations artifacts were found due to the new characteristics implemented."
    • "the introduction of CS software module shows limited recontruction errors and does not introduce artefacts in image processing that does not use this reconstruction algorithm."
      This indicates an assessment of the algorithm's output (images) without necessarily involving human interpretation in a diagnostic workflow comparison.

7. Type of Ground Truth Used

  • For the "medical checklist" and "in vivo" testing of image quality (artifacts, similarity): The ground truth appears to be based on expert consensus/visual assessment of image quality and the presence or absence of artifacts. This is inferred from statements like "images were rated similar" and "no artifacts were found."
  • For other non-clinical tests (biocompatibility, electrical safety, software V&V, performance & safety, usability): The ground truth is based on compliance with recognized international standards (e.g., ISO 10993, IEC 60601, IEC 62304, IEC 62366) and the specific criteria outlined within those standards.

8. Sample Size for the Training Set

  • This information is not provided in the document. The document describes the testing of the device and its algorithms, but not the development or training of any machine learning components. While the Compressed Sensing algorithm might involve learned parameters, the training data details are not disclosed.

9. How the Ground Truth for the Training Set Was Established

  • This information is not provided as the details of any algorithm training are absent from the document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 26, 2024

ASG Superconductors S.P.A. % Silvia Tamarri QA/RA Consultant Thema S.r.1. Via Giuseppe Saragat, 5 Imola, BO 40026 Italy

Re: K242258

Trade/Device Name: MROpen Evo (03-2000-00); MROpen Evo (03-4000-00) Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH Dated: September 3, 2024 Received: September 3, 2024

Dear Silvia Tamarri:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K242258

Device Name

MROpen Evo (03-2000-00); MROpen Evo (03-4000-00)

Indications for Use (Describe)

Whole body magnetic resonance diagnostic medical device intended for the production of images for diagnostic purposes with transverse, coronal, sagittal and obligue orientation relating to the internal structures of the examined body areas, also including the examination of the brain through the DWI sequence, with limitations regarding cardiac, breast, lung and upper abdominal imaging.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) #:K242258510(k) SummaryPrepared on: 2024-10-04
-------------------------------------------------------------

21 CFR 807.92(a)(1)

Contact Details
Applicant NameASG Superconductors S.P.A.
Applicant AddressCorso Ferdinando Maria Perrone, 73/R Genova GE 16152 Italy
Applicant Contact Telephone+39 0106489350
Applicant ContactMr. Leonardo Bertora
Applicant Contact Emailbertora.leonardo@as-g.it
Correspondent NameThema S.r.l.
Correspondent AddressVia Giuseppe Saragat, 5 Imola BO 40026 Italy
Correspondent Contact Telephone+39 0542 643496
Correspondent ContactMrs. Silvia Tamarri
Correspondent Contact Emailsilvia.tamarri@thema-med.com

21 CFR 807.92(a)(2)

Device Name
Device Trade NameMROpen Evo ( 03-2000-00);MROpen Evo (03-4000-00)
Common NameMagnetic resonance diagnostic device
Classification NameSystem, Nuclear Magnetic Resonance Imaging
Regulation Number892.1000
Product Code(s)LNH

21 CFR 807.92(a)(3)
Legally Marketed Predicate Devices

Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K193116MROpen EVOLNH

21 CFR 807.92(a)(4)
Device Description Summary
MROpen Evo is a whole body magnetic resonance diagnostic medical device that acquires high-resolution medical images of various
parts of the body. In particular, it produces images of the entire body except: upper abdomen, heart, lungs and breasts with transverse,
coronal, sagittal and oblique orientation relating to the internal structures of the examined body areas, also including the examination of
the brain through the DWI sequence.
The standard composition of the MROpen Evo device is:
Magnetic Unit
Operator PC- Operator PC

  • Electronic Cabinet
  • Penetration Filters
  • Major Software MR GUI-PRO 3.1.

{5}------------------------------------------------

Coil set (Body MAX, Hand MAX, Cervical MAX, Knee MAX, Multipurpose Linear, Flat Flex, Spine, Head-Neck, Long Spine, Shoulder)

Cushions' Set

The device consists of two models. 03-2000-00. The 03-4000-00 model of the MROpen Evo device has a very similar composition to the 03-2000-00 model, but most of the subassemblies that constitute it are replaced with others of equal functionality, in order to reduce production costs.

The MROpen Evo device allows the execution of standing imaging for those cases that require this need or facilitates acquisitions on patients suffering from movement disorders or suffering from claustrophobia. It is also possible to undergo this investigation in patients who have a high weight, but in any case, not exceeding 200 kg.

The software called MR GU-PRO represents the set of SW modules necessary for the MROpen Evo MR scanner based on the I-box spectrometer. The main modules act as a graphic interface for the execution of the exam. The MR GUI-PRO software has external interface elements, on the one hand the operator who interacts via monitor, keyboard and mouse and the other HW modules of the system and mainly: Gradient Amplifier, RF reception signal amplifier.

Intended Use/Indications for Use

21 CFR 807.92(a)(5)

21 CFR 807.92(a)(5)

Whole body magnetic resonance diagnostic medical device intended for the production of images for diagnostic purposes with transverse, coronal, sagittal and oblique orientation relating to the internal structures of the examined body areas, also including the examination of the brain through the DWI sequence, with limitations regarding cardiac, breast, lung and upper abdominal imaging.

Indications for Use Comparison

The Predicate Device MROpen EVO (K193116) is a diagnostic total body imaging device with the following limitations: no cardiac imaging, no breast imaging.

The subject device MROpen Evo is developed for same applications. As the Predicate Device, it is a whole body magnetic resonance diagnostic medical device intended for the production of images for diagnostic purposes with transverse, coronal, sagittal and oblique orientation relating to the internal structures of the examined body areas. The subject device also has limitations in lung imaging and upper abdomen imaging, but this leds to a narrower scope of applications for use of the Predicate Device. In every cases, MROpen Evo indications for use fall within the indications for use of the Predicate Device and the performances, the intended use and also the reference users are substantially equivalent.

Technological Comparison

21 CFR 807.92(a)(6)

The main change involving the technical and performance characteristics of the MROpen Evo device compared to the Predicate Device concerns the acquisition and reconstruction of images.

I-box 07 spectrometer, with a separate server, and Leybold 5000 cryocooler were introduced in the new acquisition/reconstruction algorithm, named Compressed Sensing, was used to speed up MRI acquisition by acquiring less data through subsampling of k-space.

The I-box 01-1800-06 spectrometer of the Predicate Device integrated two logical components within a single enclosure: the reconstruction PC and the spectrometer boards. The proposed device contains only the spectrometer boards with the PC board being replaced by an off-the-shelf server PC.

The two configurations are considered equivalent in terms of performance is demonstrated by testing, with a Medical checklist (MedicalCheckList_3.1.0), different acquisitions, with all sequences interested by the change, that cover a large amount of the space of parameters that can be implemented. The checklist is repeated three times: with I-box 07 with Compressed Sensing and I-box 07 without Compressed Sensing.

For the MROpen Evo device, the Leybold 5000 cryocooler was introduced in addition to the models aready available and used by the Predicate Device:

  • 6000 H MD (compressor) + 10 MD (Cold Head) supplied by Leybold GmbH;
    F70 (compressor) + CH210L (Cold Gead) supplied by Sumitomo GmbH.

The cryocooler is a service component of the magnetic resonance system designed to cool the superconducting cable at cryogenic temperatures and does not affect the performance of the medical device. Functionality and periormance remain equivalent in all three options and safety has been proven through dedicated tests carried out in accordance with IEC 60601-1.

Then, the reconstruction algorithm, named Compressed Sensing (CS), was introduced in the subject device to speed up MRI acquisition by acquiring less data through subsampling of k-space.

The tests carried out to compare the software performances of MROpen Evo with the Predicate of CS software module leads to limited reconstruction errors and does not introduce artefacts in image processing compared to standard processing that does not use this reconstruction as shown in the software test reports carried out in accordance with lEC 62304. So, the predicate and MROpen Evo performances are substantially equivalent.

{6}------------------------------------------------

The substantial equivalence between the legally marketed Predicate Device MROpen EVO (K193116) and the proposed device MRopen Evo is demonstrated by testing with a Medical checklist (MedicalCheckList_3.1.0).

The Medical Checklist consists in a list of acquisitions, with all sequences interested by the change to be tested, that cover a large amount of the space of parameters that can be implemented, with particular focus on the test of the behavior when the acquisition parameters are set to their boundaries, such as maximum number of slices, maximum and minimum voxel. The pass criteria are the absence of unexpected artifacts.

The medical checklist (MedicalCheckList_3.1.0) is repeated three times: with I-box 07 with Compressed Sensing and I-box 07 without Compressed Sensing. This medical checklist is used to prove equivalence between the three configurations. Moreover, at a validation stage, Compressed Sensing algorithm was tested in vivo, comparing the and without undersampling. The images were rated similar when comparing the three configurations artifacts were found due to the new characteristics implemented; so, acceptance criteria were met and performance that the subject device is as safe and effective and is substantially equivalent to the Predicate Device.

Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

Listed below are the non clinical tests performed in support of this Submission 510(k) on the MROpen Evo device (models 03-2000-00 and 03-4000-00), in particular:

  • the biocompatibility of the device including the identification of all materials in direct contact with the patient or user; 1.
    1. electrical safety and electromagnetic compatibility;
    1. software verification and validation;
    1. the shelf life of the device;
    1. performance and safety;
    1. usability.

BIOCOMPATIBILITY

A biological evaluation of the materials that come into contact with the patient and/or the user was carried out in compliance with the requirements of the reference standard ISO 1099-1: 2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.

According to this standard, MROpen Evo is a surface-contacting device and the contact duration is a "limited exposure" (A).

The patient can come into contact with the superficial parts of the receiving coils, made of ABS, that constitute the device and with the fabric covered in plastic/synthetic material of support cushions, through the intact skin surfaces of the body.

In order to demonstrate the biological safety of the device, tests were conducted to evaluate the endpoints of cytotoxicity, sensitization and irritation on materials in contact with in accordance with the following FDA-recognized technical standards: ISO 10993-5: 2009 (Biological evaluation of medical devices - Part 5: Tests for in vitro cytoxicity); -

  • ISO 10993-10: 2010 (Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization);

  • ISO 10993-23: 2021 (Biological evaluation of medical devices - Part 23: Tests for irritation).

Testing results in the following table confirm that the samples tested in their finished form can be considered:

  • Not cytotoxic .
  • Not irritant
  • Not sensitizing

and no additional testing are required.

ELECTRICAL SAFETY AND EMC

In order to demonstrate the EMC and Electrical safety of the MROpen Evo device, ASG Superconductors S.p.A. has performed tests in compliance with the standards below:

IEC 60601-1:2005 + A1:2012 + A2:2020 (Medical electrical equipment - Part 1: General requirements for basic safety and essential performance):

  • IEC 60601-1-2:2015 + A1:2020 (Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests);
    IEC 62471: 2006 (Photobiological safety of lamps and lamp systems).

Testing results are positive and no "fail" result appears on the test reports.

SOFTWARE VERIFICATION AND VALIDATION

The life cycle assessment of the MR-GUI PRO software in the current revision was carried out according to the requirements of the IEC 62304:2006 + A1:2015 standard (Medical device software life cycle processes). Documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff relevant to the software: "Content of Premarket Submissions for Device Software Functions".

To evaluate MROpen Evo software performances, ASG Superconductor S.p.A. has produced Unit test and System and Integration test on SW release 3.1.0 and 3.1.1. With the adoption of the MR GU-PRO software release 3.1.0 and 3.1.1, the image reconstruction technique called Compressed Sensing (CS) for 2D and 3D is introduced. CS is a method to speed up MR acquisition by acquiring less data through subsampling of k-space.

All test results are Passed because they met the introduction of CS software module shows limited recontruction

{7}------------------------------------------------

errors and does not introduce artefacts in image processing that does not use this reconstruction algorithm. The analysis of results showed that predicate and MROpen Evo software performances are substantially equivalent.

PERFORMANCE AND SAFETY

ASG Superconductor S.p.A. also caried out performance and safety tests on the device MROpen Evo in conformity with the requirements of the standard IEC 60601-2-33:2010 + A1:2013 + A2:2015 (Medical electrical equipment - Part 2-33: Particular requirements for the basic safety and essential performance equipment for medical diagnosis). In particular, performance tests were conducted in order to evaluate the following endpoints:

  • Acoustic Noise

  • Unchanged electromagnetic characteristic of inner part of the magnetic unit

  • Gradient output

  • Specific Absorption Rate (SAR)

  • Dynamic forces on patient support

The analysis of performance data demonstrates that the legally marketed Predicate Device performances are substantially equivalent for the assessed endpoints.

USABILITY

Regarding the usability of the device MROpen Evo, tests were caried out in conformity with the requirements of the standards below: - IEC 60601-1-6:2010 + A 1:2013 + A 2:2020 (Medical electrical equirements for basic safety and essential performance - Collateral standard: Usability);

  • IEC 62366-1:2015 + A 1: 2020 (Medical devices - Part 1: Application of usability engineering to medical devices).

Testing results are positive and no "fail" result appears on the test reports. This demonstrates that the subject device is as safe and effective and is substantially equivalent to the legally marketed Predicate Device.

CONCLUSIONS

Concerning the technical characteristics, it has been demonstrated that the substantially equivalent to the Predicate Device. Then, the non-clinical data support the substantial equivalence of the device with the predicate. The hardware and software verification and validation demonstrate that the subject device should perform as intended in the specified use conditions. Performance data demonstrate the device is substantially equivalent to the legally marketed Predicate Devices.

Therefore, refer to the technical and performance comparison, hazard analysis and performance test aspects, it is demonstrated that MROpen Evo is substantially equivalent and can be considered as safe and as effective as the Predicate Device.

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.