K Number
K242190

Validate with FDA (Live)

Date Cleared
2025-03-05

(223 days)

Product Code
Regulation Number
862.1205
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Access Cortisol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems. A cortisol (hydrocortisone and hydroxycorticosterone) test system is a device intended to measure the cortisol hormones secreted by the adrenal gland in serum, plasma and urine. Measurements of cortisol are used in the diagnosis and treatment of disorders of the adrenal gland.

The DxC 500i Clinical Analyzer combines the DxC 500 AU Clinical Chemistry Analyzer and the Access 2 Immunoassay System into a single instrument presentation. The system is for in vitro diagnostic use only.

The chemistry module of the DxC 500i Clinical Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (OC) material and other accessories. The immunoassay module of the DxC 500i Clinical Analyzer is an in-vitro diagnostic device used for the quantitative, semiquantitative, or qualitative determination of various analyte concentrations found in human body fluids.

Device Description

The Access Cortisol assay is a competitive binding immuno-enzymatic assay designed for use on Beckman Coulter's Access immunoassay analyzers in a clinical laboratory setting.

The DxC 500i Clinical Analyzer is an integrated chemistry-immunoassay work cell that combines Beckman Coulter's DxC 500 AU Clinical Chemistry Analyzer and the Access 2 Immunoassay System into a single instrument presentation. The DxC 500i instrument has a single user interface and common point of entry for sample racks; the sample handling unit operates as a parallel processor and sample manager for both sides of the instrument. The DxC 500i operates in conjunction with the existing reagents, calibrators, controls, and system solutions for the AU and Access instrument families.

AI/ML Overview

The provided text describes the Beckman Coulter Access Cortisol assay on the DxC 500i Clinical Analyzer and its comparison to a predicate device. Here's a breakdown of the acceptance criteria and study information:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
Method ComparisonSlope criteria of 1.00 ± 0.12 (using Weighted Deming regression analysis when compared to predicate device)Serum: Slope = 0.974 (95% CI: 0.952 - 0.996) Urine: Slope = 1.002 (95% CI: 0.976 - 1.029)
LinearityLinear throughout the analytical measuring range.Determined to be linear throughout the analytical measuring range (2.3 - 60.0 µg/dL).
Imprecision (Repeatability & Total)Allowable imprecision of <12% at approximately 5 µg/dL and <10% for higher concentrations.Control L1 (approx 3 µg/dL): Instrument 1: Repeatability CV = 4.8%, Total CV = 6.4% Instrument 2: Repeatability CV = 5.6%, Total CV = 7.2% Control L2 (approx 17.7 µg/dL): Instrument 1: Repeatability CV = 3.2%, Total CV = 3.6% Instrument 2: Repeatability CV = 2.8%, Total CV = 4.1% Control L3 (approx 27 µg/dL): Instrument 1: Repeatability CV = 3.0%, Total CV = 3.7% Instrument 2: Repeatability CV = 2.9%, Total CV = 3.6% Serum Pool (approx 47-49 µg/dL): Instrument 1: Repeatability CV = 2.7%, Total CV = 3.8% Instrument 2: Repeatability CV = 3.0%, Total CV = 5.0%
Detection CapabilityNot explicitly stated as acceptance criteria, but results are provided for LoB, LoD, and LoQ.Limit of Blank (LoB): ≤ 0.4 µg/dL Limit of Detection (LoD): ≤ 0.4 µg/dL Limit of Quantitation (LoQ) ≤ 20% within-lab CV: ≤ 0.8 µg/dL

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Method Comparison:
    • Serum: N = 104
    • Urine: N = 115
  • Imprecision: For each of the two instruments and four control levels/serum pools, N = 80 measurements were performed (this implies 20 days x 2 replicates x 2 runs, or similar, over the course of the study).
  • Provenance: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective. It does refer to "patient samples" for the method comparison, suggesting real-world samples.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable as the device is an in-vitro diagnostic test for cortisol levels, not an imaging device requiring expert interpretation of results for ground truth. The "ground truth" for comparison and performance evaluation is generally established by the predicate device's measurements and reference methods.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable for this type of in-vitro diagnostic device. Adjudication methods like 2+1 or 3+1 are typically used in studies involving subjective human interpretation (e.g., radiology reads) where discrepancies between assessors need to be resolved. For quantitative laboratory tests, "ground truth" is typically established by comparing against established reference methods or predicate devices.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable as the device is an in-vitro diagnostic test for cortisol levels, not an AI-assisted diagnostic tool for human readers.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This concept is not directly applicable in the same way it would be for an AI algorithm. The Access Cortisol assay on the DxC 500i Clinical Analyzer is an automated in-vitro diagnostic system. The "standalone" performance is essentially what is being evaluated in the method comparison, linearity, imprecision, and detection capability studies, where the system itself generates the quantitative results without subjective human interpretation as part of the core measurement. Human operators are involved in running the instrument and interpreting the numerical results, but the measurement itself is automated.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" in these studies is established by:

  • Predicate Device Measurements: For the method comparison study, the Access Cortisol assay run on the predicate Access 2 Immunoassay System serves as the comparator or "reference" for evaluating the candidate device (Access Cortisol assay on the DxC 500i Clinical Analyzer).
  • Reference Materials/Established Protocols: For linearity, imprecision, and detection capability, the ground truth is often tied to known concentrations in control materials, spiked samples, and established analytical performance specifications derived from CLSI guidelines. The calibrators use "Human serum containing cortisol (purified chemical compound) at levels of 0 and approximately 2, 5, 10, 25, and 60 µg/dL," with traceability to "USP Reference Material."

8. The sample size for the training set

This information is not provided and is likely not explicitly documented in the submission for an IVD device of this nature. The instrument doesn't explicitly 'train' in the machine learning sense on a dataset. Its operational parameters are set during its development and validated through studies like those described.

9. How the ground truth for the training set was established

As there's no explicitly mentioned "training set" in the context of machine learning, this question is not directly applicable. The device's operational characteristics and performance specifications are established through rigorous analytical verification and validation using known standards, controls, and patient samples, rather than a "training set" in the AI sense.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it.

March 5, 2025

Beckman Coulter, Inc. Mary Beth Tang Senior Staff Regulatory Affairs 250 South Kraemer Boulevard Brea, California 92821

Re: K242190

Trade/Device Name: Access Cortisol; DxC 500i Clinical Analyzer Regulation Number: 21 CFR 862.1205 Regulation Name: Cortisol (Hydrocortisone and Hydroxycorticosterone) Test System Regulatory Class: Class II Product Code: CGR, JJE Dated: January 24, 2025 Received: January 24, 2025

Dear Mary Beth Tang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Paula V. Caposino -S

Paula V. Caposino, Ph.D. Deputy Division Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K242190

Device Name Access Cortisol; DxC 500i Clinical Analyzer

Indications for Use (Describe)

The Access Cortisol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems. A cortisol (hydrocortisone and hydroxycorticosterone) test system is a device intended to measure the cortisol hormones secreted by the adrenal gland in serum, plasma and urine. Measurements of cortisol are used in the diagnosis and treatment of disorders of the adrenal gland.

The DxC 500i Clinical Analyzer combines the DxC 500 AU Clinical Chemistry Analyzer and the Access 2 Immunoassay System into a single instrument presentation. The system is for in vitro diagnostic use only.

The chemistry module of the DxC 500i Clinical Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (OC) material and other accessories. The immunoassay module of the DxC 500i Clinical Analyzer is an in-vitro diagnostic device used for the quantitative, semiquantitative, or qualitative determination of various analyte concentrations found in human body fluids.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, followed by the company name "BECKMAN COULTER" in bold, black letters. The word "BECKMAN" is stacked on top of the word "COULTER".

Submitted By:

Mary Beth Tang Senior Staff Regulatory Affairs Beckman Coulter, Inc. 250 S. Kraemer Boulevard Brea, CA 92821 Telephone: 714-961-3728 Fax: 714-961-4234 Email: mbtang@beckman.com

Date of Preparation:

March 4, 2025

Device Name(s):

Proprietary Name:Access Cortisol
Common Name:Cortisol (hydrocortisone and hydroxycorticosterone) test system
Classification Name:Radioimmunoassay, Cortisol
Class:Class II
Regulation Number:21 CFR 862.1205
Product Code:CGR
Proprietary Name:DxC 500i Clinical Analyzer
Common Name:Discrete photometric analyzer for clinical use
Classification Name:Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Class:Class I
Regulation Number:21 CFR 862.2160
Product Code:JJE

Predicate Device:

Candidate Test SystemPredicate Test SystemPredicate Docket
Access Cortisol on the DxC 500iClinical AnalyzerAccess Cortisol on the Access 2Immunoassay SystemK050202

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, resembling a stylized wave or heartbeat. To the right of the circle are the words "BECKMAN" and "COULTER" stacked on top of each other in bold, black font. The overall design is clean and modern.

Device Description:

The Access Cortisol assay is a competitive binding immuno-enzymatic assay designed for use on Beckman Coulter's Access immunoassay analyzers in a clinical laboratory setting. Table 1 summarizes the device characteristics of the Access Cortisol assay on the DxC 500i Clinical Analyzer.

The DxC 500i Clinical Analyzer is an integrated chemistry-immunoassay work cell that combines Beckman Coulter's DxC 500 AU Clinical Chemistry Analyzer and the Access 2 Immunoassay System into a single instrument presentation. The DxC 500i instrument has a single user interface and common point of entry for sample racks; the sample handling unit operates as a parallel processor and sample manager for both sides of the instrument. The DxC 500i operates in conjunction with the existing reagents, calibrators, controls, and system solutions for the AU and Access instrument families.

Indications for Use:

The Access Cortisol assay is a paramagnetic particle, chemiluminescent immunoassay for the guantitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems. A cortisol (hydrocortisone and hydroxycorticosterone) test system is a device intended to measure the cortisol hormones secreted by the adrenal gland in serum, plasma and urine. Measurements of cortisol are used in the diagnosis and treatment of disorders of the adrenal gland.

The DxC 500i Clinical Analyzer combines the DxC 500 AU Clinical Chemistry Analyzer and the Access 2 Immunoassay System into a single instrument presentation. The system is for in vitro diagnostic use only. The chemistry module of the DxC 500i Clinical Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. The immunoassay module of the DxC 500i Clinical Analyzer is an in-vitro diagnostic device used for the quantitative, or qualitative determination of various analyte concentrations found in human body fluids.

Comparison to the Predicate:

The Access Cortisol assay run on the DxC 500i Clinical Analyzer was compared to the Access Cortisol assay run on the stand-alone Access 2 Immunoassay System. Tables 1 and 2 compare the assay and instrument platform characteristics, respectively. The information from the predicate device was derived from the predicate device 510(k) Summary and product labeling.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, followed by the company name in black, with "BECKMAN" on top of "COULTER". The font is bold and sans-serif.

FeaturePredicate Device:Access Cortisol assayon the Access 2 Immunoassay System(K050202)Candidate Device:Access Cortisol assayon the DxC 500iClinical Analyzer
Intended Use/Indications for UseThe Access Cortisol assay is a paramagneticparticle, chemiluminescent immunoassay for thequantitative determination of cortisol levels inhuman serum, plasma (heparin, EDTA) and urineusing the Access Immunoassay Systems. Acortisol (hydrocortisone and hydroxycorticosterone)test system is a device intended to measure thecortisol hormones secreted by the adrenal gland inserum, plasma and urine. Measurements of cortisolare used in the diagnosis and treatment ofdisorders of the adrenal gland.Same
MethodologyCompetitive binding immunoassay withchemiluminescence detection (automated)Same
Solid PhaseParamagnetic particles coated with goat anti-rabbitlgGSame
ConjugateCortisol-alkaline phosphatase (bovine) conjugateSame
SubstrateAccess SubstrateSame
CalibratorsHuman serum containing cortisol (purified chemicalcompound) at levels of 0 and approximately 2, 5,10, 25, and 60 µg/dLSame
TraceabilityUSP Reference MaterialSame
Calibration Stability28 daysSame
Reagent Open-PackStabilityStable at 2 to 10°C for 14 days after initial useSame
Sample TypeSerum, plasma, urineSame
Measuring Range0.4 to 60.0 µg/dL2.3 to 60.0 µg/dL
InstrumentAccess 2Access 2 and DxC 500iClinical Analyzer
Reagent IFUA33262D12398A

Table 1 Predicate Device Comparison (Assay)

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red oval shape with two white curved lines inside, followed by the company name "BECKMAN COULTER" in bold, black letters. The word "BECKMAN" is stacked on top of "COULTER".

FeaturePredicate Device: K121214,Access 2 Immunoassay SystemCandidate Device:DxC 500i Clinical Analyzer
Intended UseThe Access 2 Analyzer is an in-vitrodiagnostic device used for thequantitative, semi-quantitative, orqualitative determination of variousanalyte concentrations found inhuman body fluids.The DxC 500i Clinical Analyzer combinesthe DxC 500 AU Clinical ChemistryAnalyzer and the Access 2 ImmunoassaySystem into a single instrumentpresentation. The system is for in vitrodiagnostic use only. The chemistry moduleof the DxC 500i Clinical Analyzer is anautomated chemistry analyzer thatmeasures analytes in samples, incombination with appropriate reagents,calibrators, quality control (QC) material andother accessories. The immunoassaymodule of the DxC 500i Clinical Analyzer isan in-vitro diagnostic device used for thequantitative, semi-quantitative, or qualitativedetermination of various analyteconcentrations found in human body fluids.
AnalyticalModulesAccess 2 Immunoassay Analyzerwith closed chemiluminescencemoduleA modified Access 2 system serves as theimmunoassay module in the DxC 500i workcell configuration. Beckman Coulter's DxC500 AU Clinical Chemistry Analyzer servesas the chemistry module in the DxC 500iwork cell configuration. The analyticalprocesses on both analytical modules areidentical to those on the respective stand-alone analyzers; technologicalcharacteristics, sample and reagentdispensing, chemistry analysis, and datareduction modules are all unchanged.
Test SystemsMenu of cleared and exempt Accessassays; user-enabled systemconfiguration.Menu of cleared and exempt Access(immunoassay) and AU (chemistry) assays;user-enabled system configuration.
AssayParametersDatabase of Access Assay ProtocolFiles for Access 2 SystemsSame for immunoassay module; chemistrymodule uses current AU assay settings.
System ConsoleAccess 2 console for hardwareconfiguration as currently marketedShared console for the integrated hardwareconfiguration; manages combined chemistryand immunoassay module operations.
FeaturePredicate Device: K121214, Access 2 Immunoassay SystemCandidate Device: DxC 500i Clinical Analyzer
GUI ApplicationUser interface unique to Access 2 SystemsUser interface common to new Beckman Coulter analyzers
Specimen TypesSerum, plasma, urine, amniotic fluid, whole blood hemolysate (off-line preparation)Same; chemistry module includes whole blood (online hemolysate preparation)
Sample InputSample rack load area with a sample wheel that accommodates all sample types (routine, STAT, QC, calibration)Shared sample input area: Continuous sample loading (routine, STAT, QC, calibration, maintenance) from single point of entry on the shared sample handler.
Sample HandlingOpen primary tubes/cups loaded in a unique 10-tube rack, continuous loading (patient, STAT, QC, calibration, maintenance samples)Open primary tubes/cups loaded in a 7-tube rack common to new Beckman Coulter analyzers, continuous loading (patient, STAT, QC, calibration, maintenance samples)
Sample IdentificationStationary internal barcode readerDynamic barcode reader attached to the rack shuttle on the shared sample handler
Reagent LoadingInternal fixed barcode reader with refrigerated onboard storage; reagents can be loaded during instrument measurement cycle.Same
Results ManagementThe analytical module collects and sends the response data to the instrument console and calculates the final test results for reportingSame; the shared console collates the chemistry and immunoassay test results.
Track ConnectivityNoSame

Table 2 Predicate Device Comparison (Instrument)

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, followed by the words "BECKMAN" and "COULTER" stacked on top of each other. The text is in a bold, sans-serif font.

K242190 510(k) Summary

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, resembling a stylized wave or flow pattern. To the right of the circle, the words "BECKMAN" and "COULTER" are stacked vertically in a bold, sans-serif font.

Comparison Testing:

Substantial equivalence was demonstrated through non-clinical (bench) studies. CLSI study protocols were used to ensure that the technological differences between the candidate and predicate analyzer models did not adversely affect safety and effectiveness; these performance evaluations included method comparison, linearity, imprecision, and detection limit studies. Additional studies were conducted for carryover and thermal sensitivity.

Summary of Performance Data:

The method comparison studies between the Access Cortisol assay on the DxC 500i Clinical Analyzer and the Access Cortisol assay on the Access 2 Immunoassay System were designed in accordance with CLSI Guideline EP09C-ED3, Measurement Procedure Comparison and Bias Estimation Using Patient Samples: Third Edition. The study evaluated both serum and urine using test samples that spanned the measuring range of the assay (2.3 - 60.0 µg/dL). The test results met the slope criteria of 1.00 ± 0.12 using Weighted Deming regression analysis.

SampleTypeConcentrationRange(µg/dL)NSlopeSlope 95% CIIntercept(µg/dL)CorrelationCoefficient(r)
Serum2.3 – 56.21040.9740.952 - 0.9960.330.993
Urine*3.1 - 59.01151.0020.976- 1.0290.180.992

*Unextracted, random collection

Linearity studies were designed in accordance with CLSI Guideline EP06-ED2:2020, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline, 2nd Edition. The Access Cortisol assay was determined to be linear throughout the analytical measuring range on the DxC 500i analyzer.

Repeatability (within-run) and within-laboratory (total) precision studies followed CLSI guideline EP05-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition. The Access Cortisol assay met the established performance specification for allowable imprecision of <12% at approximately 5 µg/dL and <10% for higher concentrations for all test samples when run on the DxC 500i analyzer.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, resembling a stylized wave or heartbeat. To the right of the circle, the words "BECKMAN" and "COULTER" are stacked vertically in a bold, sans-serif font.

K242190 510(k) Summary Access Cortisol on DxC 500i

DxC 500iInstrumentTest SampleNMeanValue(ug/dL)Repeatability(Within-run)Total(Within-laboratory)
SD(ug/dL)CV(%)SD(ug/dL)CV(%)
1Control L1803.040.154.80.196.4
Control L28017.70.563.20.633.6
Control L38027.00.803.01.003.7
Serum Pool8049.31.332.71.883.8
2Control L1802.950.165.60.217.2
Control L28017.70.502.80.724.1
Control L38026.80.772.90.983.6
Serum Pool8047.21.403.02.345.0

Detection capability studies followed the CLSI guideline EP17-A2 "Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures Approved Guideline - Second Edition". The Access Cortisol assay on the DxC 500i Clinical Analyzer produced the following results:

Parameterµg/dLnmol/L
Limit of Blank (LoB)≤ 0.4≤ 11
Limit of Detection (LoD)≤ 0.4≤ 11
Limit of Quantitation (LoQ) ≤ 20% within-lab CV≤ 0.8≤ 22

Substantial Equivalence Conclusion

The data contained in the Premarket Notification supports a finding of substantial equivalence to the measurand test systems already in commercial distribution. The performance testing presented in this submission provides evidence that the device is safe and effective in its Intended Use.

§ 862.1205 Cortisol (hydrocortisone and hydroxycorticosterone) test system.

(a)
Identification. A cortisol (hydrocortisone and hydroxycorticosterone) test system is a device intended to measure the cortisol hormones secreted by the adrenal gland in plasma and urine. Measurements of cortisol are used in the diagnosis and treatment of disorders of the adrenal gland.(b)
Classification. Class II.