K Number
K242135

Validate with FDA (Live)

Date Cleared
2025-01-31

(193 days)

Product Code
Regulation Number
862.1155
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Synthgene Home Test HCG Test Midstream is used for the qualitative detection of human chorionic gonadotropin in urine to aid in the early detection of pregnancy.

Synthqene Home Test HCG Test Cassette is used for the qualitative detection of human chorionic gonadotropin in urine to aid in the early detection of pregnancy.

Synthqene Home Test HCG Test Strip is used for the qualitative detection of human chorionic gonadotropin in urine to aid in the early detection of pregnancy.

Device Description

Synthgene Home Test HCG Test is used for the qualitative detection of human chorionic gonadotropin in urine to aid in the early detection of pregnancy. The test devices are in three different formats: Strip, Cassette and Midstream. Each test kit contains the individually packaged test pen/strip/cassette, medical waste bag, urine cup and instructions for Use. The cassette format also contains a dropper. The device is in a readyto-use format and does not require assembly before use.

AI/ML Overview

The provided text describes the performance characteristics and acceptance criteria for the "Synthgene Home Test HCG Test" (Strip, Cassette, and Midstream formats) for qualitative detection of human chorionic gonadotropin in urine. This is a medical device for home use, so the study focuses on analytical performance and comparison with a predicate device, as well as a lay-user study for ease of use. This is not an AI/ML powered device, so "human readers improve with AI vs without AI assistance" or "algorithm only without human-in-the-loop performance" are not applicable.

Here's a breakdown of the requested information based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a "table of acceptance criteria" in the same format as an AI/ML performance study with specific sensitivity/specificity targets. Instead, it describes various analytical performance tests and method comparison studies to demonstrate that the device performs as intended and is substantially equivalent to a predicate device.

However, based on the studies conducted, we can infer the performance:

Inferred Acceptance Criteria & Reported Performance:

Performance MetricAcceptance Criteria (Inferred from Study Design)Reported Device Performance
Analytical Sensitivity (Limit of Detection)Detect hCG at 25 mIU/mL with 100% positivity and show 0% positivity at concentrations below 25 mIU/mL (e.g., 0, 12.5, 18.75 mIU/mL).For all formats (Dip, Midstream, Cassette, Strip):- 0, 12.5, 18.75 mIU/mL: 0% positive results (100% negative).- 22.5 mIU/mL: High positivity (e.g., Dip: 87.3%, Midstream: 84.9%, Cassette: 86.2%, Strip: 85.11%). This demonstrates the detection threshold is around 25 mIU/mL.- 25, 50, 100, 200 mIU/mL: 100% positive results. The sensitivity is demonstrated to be 25 mIU/mL.
Precision/ReproducibilityConsistent results across different lots, operators, sites, and times of day.No significant differences in performance were observed between lots, operators, sites, or times of day. The sensitivity data (Table 9.1.a) shows consistent results for each concentration across 3 different lots.
Cross-Reactivity/SpecificityNo interference from tested concentrations of LH, FSH, TSH, and hCG β-core fragment.No interference observed at 500 mIU/mL LH, 1000 mIU/mL FSH, 1000 mIU/L TSH, and 2,000,000 pmol/L hCG β-core fragment, for both negative and positive urine samples.
Interfering SubstancesNo interference from a list of common exogenous and endogenous substances at specified concentrations.No interferences observed from the comprehensive list of 30 tested substances (e.g., Acetaminophen, Aspirin, Vitamin C, Ethanol, Bilirubin, Glucose) at their specified concentrations for both negative and positive urine samples.
Effects of Urine pHNo interference in hCG detection across a physiological pH range.Changes in pH range of 4-9 do not interfere with results for either positive (25 mIU/mL) or negative (0 mIU/mL) hCG urine samples.
Effects of Specific GravityNo interference in hCG detection across a range of urine specific gravities.Urine samples with a specific gravity between 1.000-1.035 do not interfere with results for either positive (25 mIU/mL) or negative (0 mIU/mL) hCG urine samples.
Hook EffectNo false negative results at very high hCG concentrations.No hook effect observed at hCG concentrations ranging from 2325 to 1,860,000 mIU/mL; all tested concentrations gave a positive result.
Method Comparison with Predicate DeviceSubstantial agreement (e.g., high positive and negative agreement) with the legally marketed predicate device.All formats (Midstream-Dip, Midstream-Midstream, Cassette, Strip) showed 100% positive agreement and 100% negative agreement with the predicate device (HIGHTOP Pregnancy Rapid Test).- Midstream (Dip): 38 Positive, 23 Negative (61 total) - 100% agreement.- Midstream (Midstream): 40 Positive, 20 Negative (60 total) - 100% agreement.- Cassette: 60 Positive, 49 Negative (109 total) - 100% agreement.- Strip: 54 Positive, 49 Negative (103 total) - 100% agreement.
Lay-User Comprehension & PerformanceUsers can adequately understand and use the instructions for use, perform the test, and interpret results correctly.333 subjects participated across three sites. The results from the lay-user study for each device format mirrored the 100% agreement found in the method comparison study.Questionnaire results indicated consumers found the test easy to use, and had no trouble understanding labeling or interpreting results.

2. Sample sizes used for the test set and the data provenance

  • Analytical Performance (Sensitivity/Reproducibility):

    • Sample Size: For each HCG concentration level (0, 12.5, 18.75, 22.5, 25, 50, 100, 200 mIU/mL), and for each of the three device formats (Dip, Midstream, Cassette, Strip), a total of 450 replicates were tested. This translates to (3 lots * 3 operators * 10 runs/day * 5 days = 450 replicates per concentration per format).
    • Data Provenance: The document does not explicitly state the country of origin for the samples used in the analytical performance studies. It mentions "Negative urine collected from females" and "Negative urine samples were spiked with varying hCG concentrations." This suggests controlled laboratory-prepared samples. It's an analytical study, not a clinical trial with patient samples. The manufacturer is based in China.
    • Retrospective/Prospective: These analytical studies appear to be prospective, controlled laboratory experiments designed specifically for device validation.
  • Method Comparison Study & Lay-User Study (combined in the document):

    • Sample Size: Urine samples were collected from 333 women aged 18-49.
      • Synthgene Home Test HCG Test Midstream: 121 samples (61 dip, 60 midstream)
      • Synthgene Home Test HCG Test Cassette: 109 samples
      • Synthgene Home Test HCG Test Strip: 103 samples
      • Note: The sum of samples for each format (121+109+103 = 333) matches the total number of women, suggesting that each woman typically provided a sample tested by one specific format, or multiple samples which were divided among the different test formats. The samples were randomly collected.
    • Data Provenance: "Urine samples were collected from 333 women aged 18 - 49 who presented for pregnancy testing at 3 sites." The document does not specify the country of origin for these clinical samples or the sites. Given the manufacturer's location, China is a possibility, but not explicitly stated.
    • Retrospective/Prospective: The samples were "collected at various times throughout the day and blinded," which suggests a prospective clinical performance study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • For the analytical performance studies, the ground truth was established by the known concentrations of spiked hCG in urine. This falls under controlled laboratory conditions, not requiring expert readers for ground truth.
  • For the method comparison study, the predicate device (HIGHTOP Pregnancy Rapid Test) serves as the comparator, essentially establishing a "ground truth" relative to a legally marketed device. The document states "tested by professionals using the candidate devices and the predicate device." It does not specify the number or qualifications of these "professionals."
  • For the lay-user study, the "ground truth" (actual pregnancy status, i.e., presence or absence of hCG) was likely based on the results from the predicate device that the professionals used, or potentially a reference lab method. The study's primary purpose was to assess the lay user's ability to use and interpret the device, rather than to establish a new ground truth.

4. Adjudication method for the test set

  • Analytical Performance: Not applicable, as the ground truth is defined by the known spiked concentration.
  • Method Comparison and Lay-User Studies: The document does not describe an explicit adjudication method (e.g., 2+1, 3+1). The comparison is direct, stating that samples were "blinded before being tested by professionals using the candidate devices and the predicate device." The results show 100% agreement, suggesting no need for further adjudication or dispute resolution between readers/methods.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This question is not applicable as the device (HCG Test Strip/Cassette/Midstream) is a rapid diagnostic test and not an AI/ML-powered medical device. It does not involve human readers interpreting images with or without AI assistance. The "reader" in this context is the lay user or a professional reading a visual line on a test strip.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • This question is not applicable as the device is not an AI/ML algorithm. Its performance is inherent to its biochemical reaction and visual readout, which is then interpreted by a human user.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

  • Analytical Performance: Ground truth was established by known, spiked concentrations of hCG in urine, controlled in a laboratory setting.
  • Method Comparison and Lay-User Studies: The ground truth for presence/absence of hCG was established by comparison to a legally marketed predicate device (HIGHTOP Pregnancy Rapid Test), used by professionals, and inferred from the women "who presented for pregnancy testing."

8. The sample size for the training set

  • This document describes performance studies for market clearance (510(k)), not the development or training of an AI/ML model. Therefore, there is no "training set" in the context of machine learning. The studies described are validation and verification studies of a finished product.

9. How the ground truth for the training set was established

  • This question is not applicable as there is no training set for an AI/ML model in this context.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 31, 2025

Nanjing Synthgene Medical Technology Co., Ltd. Alvin Oian Registration Manager Building B6-2, No. 9, Weidi Road, Xianlin University Town Xianlin Subdistrict, Qixia District Nanjing, Jiangsu 210000 China

Re: K242135

Trade/Device Name: Synthgene Home Test HCG Test Strip: Synthgene Home Test HCG Test Cassette; Synthgene Home Test HCG Test Midstream Regulation Number: 21 CFR 862.1155 Regulation Name: Human Chorionic Gonadotropin (HCG) Test System Regulatory Class: Class II Product Code: LCX Dated: December 20, 2024 Received: December 20, 2024

Dear Alvin Qian:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

{2}------------------------------------------------

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Joseph A.
Kotarek -S
Digitally signed by
Joseph A. Kotarek -S
Date: 2025.01.31 08:35:49
-05'00'
Joseph Kotarek, Phd.D.
Branch Chief

Branch Chief Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

Device Name

Synthgene Home Test HCG Test Strip; Synthgene Home Test HCG Test Cassette; Synthgene Home Test HCG Test Midstream

Indications for Use (Describe)

Synthgene Home Test HCG Test Midstream is used for the qualitative detection of human chorionic gonadotropin in urine to aid in the early detection of pregnancy.

Synthqene Home Test HCG Test Cassette is used for the qualitative detection of human chorionic gonadotropin in urine to aid in the early detection of pregnancy.

Synthqene Home Test HCG Test Strip is used for the qualitative detection of human chorionic gonadotropin in urine to aid in the early detection of pregnancy.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K242135

1. Date Prepared

January 24, 2025

2. Submitter

Company Name: Nanjing Synthgene Medical Technology Co., Ltd.

Address: Building B6-2, No. 9, Weidi Road, Xianlin University Town, Xianlin

Subdistrict, Qixia District, Nanjing City, Nanjing, Jiangsu, 210000, CHINA

Establishment Registration Number: 3016902172

Contact Person: ALVIN QIAN

Position: Registration Manager

Tel.: +86 15755073877

E-mail: alvin(@syngenebio.com

3. Proposed Device

Trade Name of Device: Synthgene Home Test HCG Test Strip

Synthgene Home Test HCG Test Cassette

Synthgene Home Test HCG Test Midstream

Classification Name: Kit, Test, Pregnancy, Hcg, Over The Counter

Classification: II

Product Code: LCX

Regulation Number: 21 CFR 862.1155

Review Panel: Clinical Chemistry

4. Predicate Devices

510(k) Number: K192123

Trade/Product Name: HIGHTOP Pregnancy Rapid Test Cassette

{5}------------------------------------------------

510(k) summary HIGHTOP Pregnancy Rapid Test Strip

HIGHTOP Pregnancy Rapid Test Midstream

Regulation Name: Human chorionic gonadotropin (HCG) test system Classification name: kit, test, pregnancy, hcg, over the counter Regulation Number: 21 CFR 862.1155 Device Class: Class II Product Code: LCX Manufacturer: Qingdao Hightop Biotech Co., Ltd.

5. Device description

Synthgene Home Test HCG Test is used for the qualitative detection of human chorionic gonadotropin in urine to aid in the early detection of pregnancy. The test devices are in three different formats: Strip, Cassette and Midstream. Each test kit contains the individually packaged test pen/strip/cassette, medical waste bag, urine cup and instructions for Use. The cassette format also contains a dropper. The device is in a readyto-use format and does not require assembly before use.

6. Indication for use

Synthgene Home Test HCG Test Midstream is used for the qualitative detection of human chorionic gonadotropin in urine to aid in the early detection of pregnancy.

Synthgene Home Test HCG Test Cassette is used for the qualitative detection of human chorionic gonadotropin in urine to aid in the early detection of pregnancy.

Synthgene Home Test HCG Test Strip is used for the qualitative detection of human chorionic gonadotropin in urine to aid in the early detection of pregnancy.

7. Comparison of technological characteristics with the predicate devices

The comparison and discussion between the candidate device and the predicate devices are listed in table below.

Table 1 Comparison of Candidate Devices with the Predicate Devices

{6}------------------------------------------------

Similarities
ItemCandidate devicePredicate device (K192123)
Product nameSynthgene Home Test HCG Test StripSynthgene Home Test HCG Test CassetteSynthgene Home Test HCG Test MidstreamHIGHTOP Pregnancy Rapid Test CassetteHIGHTOP Pregnancy Rapid Test StripHIGHTOP Pregnancy Rapid Test Midstream
Intended UseEarly detection of pregnancy.Same
Specimen TypeUrineSame
Assay methodologyImmunochromatographic assaySame
Sensitivity25 mIU/mLSame
ResultsQualitativeSame
Intended userOver the counter useSame
Device formatStrip/Cassette/MidstreamSame
Differences
Reading Time5-20 minutes5 minutes

8. Test Principle

The product adopts the principle of double antibody sandwich. When human chorionic gonadotropin is present in the sample, the HCG antigen in the sample reacts with the colloidal gold-labeled antibody (HCG mAb1) on the conjugate pad to form the labeled antibody-antigen complex. The complex is chromatographed upward by capillary action and captured by the detection line (T line) antibody (HCG mAb2) coated on the nitrocellulose membrane, appearing as a red band. The complex continues to move chromatographically upward, and is captured by the control line (C line) antibody (goat anti-mouse IgG polyclonal antibody) coated on the nitrocellulose membrane, and a red band appears. When the concentration of HCG in the sample is lower than the sensitivity, the detection line (T line) does not develop color.

9. Performance Characteristics

{7}------------------------------------------------

9.1 Analytical performance

Precision/Reproducibility/Sensitivity a

Negative urine collected from females was spiked with hCG standard to hCG concentrations of 0, 12.5, 18.75, 22.5, 25, 50, 100 and 200 mIU/mL. The spiked samples were measured using 3 different lots for each format. Tests were performed by three different operators for each sample concentration in 10 runs per day per lot for 5 days. For each format of the device, a total of 450 replicates were tested for each HCG concentration. Based on the test results, the sensitivity of Synthgene Home Test HCG Test is demonstrated to be 25 mIU/mL. There were no significant differences in performance between lots, between operators, between sites, between time of day and between days. The results are summarized in the table below:

Dip method
SampleNo.HCGconcentration(mIU/mL)Lot 1Lot 2Lot 3TotalresultsPosNeg
-+-+-+-+
S1015001500150045000%100%
S212.515001500150045000%100%
S318.7515001500150045000%100%
S422.52013020130171335739387.3%12.7%
S5250150015001500450100%0%
S6500150015001500450100%0%
S71000150015001500450100%0%
S82000150015001500450100%0%

Midstream Format

Midstream method

SampleNo.HCGconcentration(mIU/mL)Lot 1Lot 2Lot 3TotalresultsPosNeg
-+-+-+-+
S1015001500150045000%100%
S212.515001500150045000%100%
S318.7515001500150045000%100%
S422.52312724126211296838284.9%15.1%
S5250150015001500450100%0%
S6500150015001500450100%0%
S71000150015001500450100%0%
S82000150015001500450100%0%

{8}------------------------------------------------

Cassette Format

SampleNo.HCGconcentration(mIU/mL)Lot 1Lot 2Lot 3TotalresultsPosNeg
-+-+-+-+
S1015001500150045000%100%
S212.515001500150045000%100%
S318.7515001500150045000%100%
S422.51913124126191316238886.2%13.8%
S5250150015001500450100%0%
S6500150015001500450100%0%
S71000150015001500450100%0%
S82000150015001500450100%0%

Strip Format

SampleNo.HCGconcentration(mIU/mL)Lot 1Lot 2Lot 3TotalresultsPosNeg
-+-+-+-+
S1015001500150045000%100%
S212.515001500150045000%100%
S318.7515001500150045000%100%
S422.52312723127211296738385.11%14.89%
S5250150015001500450100%0%
S6500150015001500450100%0%
S71000150015001500450100%0%
S82000150015001500450100%0%

Linearity/assay reportable range: b

Linearity is not applicable since this is a qualitative test.

Traceability, Stability, Expected values (controls, calibrators, or methods) Synthgene Home Test HCG Test is calibrated against reference material traceable to WHO International Standard 6th edition (NIBSC code: 18/244).

A shelf-life stability test of the devices was performed by Real-time Stability Study. The results showed that the devices were stable for 24 months when stored at 2~30℃ in the sealed foil pouch.

Cross reactivity C

To evaluate specificity and cross-reactivity, negative and positive urine samples (0

{9}------------------------------------------------

mIU/mL and 25 mIU/mL) were spiked with various concentrations of glycoprotein hormones LH, FSH, TSH and hCG ß-core fragment. The results showed that there is no interference at 500 mIU/mL LH, 1000 mIU/mL FSH, 1000 mIU/L TSH and 2,000,000 pmol/L hCG ß-core fragment for both negative and positive urine samples.

d Interfering substance

A study was performed to evaluate the potential for interference by certain exogenous and endogenous compounds. Each substance was prepared by diluting stock interference material to the desired concentration. Urine specimens containing 0 and 25 mIU/ml HCG were spiked with the potential interferents to obtain the desired test concentration. The results show that no interferences were observed from substance at the following concentrations for both negative and positive urine samples.

Interfering substanceSubstances concentration
Acetaminophen0.2 mg/mL
Aspirin0.2 mg/mL
Vitamin C0.2 mg/mL
Atropine0.2 mg/mL
Ampicillin0.2 mg/mL
Tetracycline0.2 mg/mL
2,5-Dihydroxybenzoic acid0.2 mg/mL
Caffeine0.2 mg/mL
Phenothiazine0.2 mg/mL
Ethanol1%
Human Albumin20 mg/mL
Triglyceride8 mg/mL
Hemoglobin5 mg/mL
Bilirubin0.02 mg/mL
Glucose20 mg/mL
Acetoacetic Acid20 mg/mL
Asorbic Acid0.2 mg/mL
B-hydroxybutyrate20 mg/mL
Ephedrine0.2 mg/mL
Gentisic Acid0.2 mg/mL
Phenylpropanolamine0.2 mg/mL
Salicylic Acid0.2 mg/mL
EDTA0.8 mg/mL
Benzoylecgonine0.1 mg/mL
Cannabinol0.1 mg/mL
Codeine0.06 µg/mL
Methanol10%
Estriol-17-beta14 mg/mL
Pregnanediol15 µg/mL
Thiophene0.2 mg/mL
Ketone0.2 mg/mL

Effects of Urine pH e

{10}------------------------------------------------

To evaluate potential interference from changes in pH, urine samples containing 0 mIU/mL and 25 mIU/mL hCG were tested at pH values of 4, 5, 6, 7, 8 and 9. The results indicated that changes in pH range of 4 - 9 do not interfere in the results that were either positive or negative for hCG.

f Effects of Specific Gravity

To evaluate potential interference from changes in specific gravity, urine samples containing 0 mIU/mL and 25 mIU/mL hCG were tested at specific gravity values at 1.000, 1.002, 1.009, 1.012, 1.018, 1.028, 1.032 and 1.035. The results demonstrated that urine samples with a specific gravity between 1.000-1.035 do not interfere in the results that were either positive or negative for hCG.

Hook effect test g

Negative urine samples were spiked with varying hCG concentrations (2325, 4650, 9300, 11,625, 58,125, 116,250, 232,500, 465,000, 930,000 and 1,860,000 mIU/mL). All tested concentrations gave a positive result. The results demonstrated that no hook effect was observed at hCG concentrations ranging from 2325 to 1,860,000 mIU/ml.

9.2 Method comparison study

Method comparison with predicate device a

Urine samples were collected from 333 women aged 18 - 49 who presented for pregnancy testing at 3 sites. Samples were randomly collected at various times throughout the day and blinded before being tested by professionals using the candidate devices and the predicate device. A total of 121 samples tested with the Synthgene Home Test HCG Test Midstream, 109 samples tested with the Synthgene Home Test HCG Test Cassette, and 103 samples tested with the Synthgene Home Test HCG Test Strip. Of the 121 samples tested using the midstream format, 61 were tested in dip sampling mode and 60 were tested in midstream sampling mode. The results of the study are summarized in the tables below:

Midstream Format - Dip application method:
----------------------------------------------
Synthgene Home TestHIGHTOP Pregnancy Rapid TestMidstreamTotal
-----------------------------------------------------------------------

{11}------------------------------------------------

HCG Test MidstreamPositiveNegative
Positive38038
Negative02323
Total382361

Midstream Format - Midstream application method :

Synthgene Home TestHCG Test MidstreamHIGHTOP Pregnancy Rapid TestMidstreamTotal
PositiveNegative
Positive40040
Negative02020
Total402060

For Synthgene Home Test HCG Test Cassette

Synthgene Home Test HCG TestCassetteHIGHTOP Pregnancy Rapid TestCassetteTotal
PositiveNegative
Positive60060
Negative04949
Total6049109

For Synthgene Home Test HCG Test Strip

Synthgene Home TestHCG Test StripHIGHTOP Pregnancy Rapid Test StripTotal
PositiveNegative
Positive54054
Negative04949
Total5449103

b Matrix comparison

Not applicable. The device is for use with urine samples only.

9.3 Clinical studies

Clinical Sensitivity a

Not applicable.

Clinical specificity b

Not applicable.

{12}------------------------------------------------

Other clinical supportive data (when a. and b. are not applicable): C

Lay-User Study

A lay user study was conducted at three sites with female subjects who were able to adequately understand and use the instructions for use to conduct testing. These subjects had diverse educational and professional backgrounds and their age range was between 18 and 49 years old. A total of 333 subjects participated in the study, included 121 layusers using the Synthgene Home Test HCG Test Midstream, 109 lay-users using the Synthgene Home Test HCG Test Cassette, and 103 lay-users using the Synthgene Home Test HCG Test Strip.

The test results are summarized below:

Synthgene Home TestHCG Test MidstreamHIGHTOP Pregnancy Rapid TestMidstreamTotal
PositiveNegative
Positive38038
Negative02323
Total382361

Midstream Format - Dip application method

Midstream Format - Midstream application method:

Synthgene Home TestHCG Test MidstreamHIGHTOP Pregnancy Rapid TestMidstreamTotal
PositiveNegative
Positive40040
Negative02020
Total402060

For Synthgene Home Test HCG Test Cassette

Synthgene Home Test HCG TestCassetteHIGHTOP Pregnancy Rapid TestCassetteTotal
PositiveNegative
Positive60060
Negative04949
Total6049109

For Synthgene Home Test HCG Test Strip

Synthgene Home TestHCG Test StripHIGHTOP Pregnancy Rapid Test StripTotal
PositiveNegative

{13}------------------------------------------------

Positive54054
Negative04949
Total5449103

Each lay user was given a questionnaire to assess the readability of the labeling. The results of the questionnaire reflected that the consumers found the test easy to use and that they did not have trouble understanding the labeling and interpreting the results.

Clinical cut-off: d

Not applicable

Expected values/Reference range: e

Not applicable.

10. Conclusion

Based on the test principle and acceptable performance characteristics including precision, interference, specificity, method comparison, and lay-user studies of the devices, it's concluded that the Synthgene Home Test HCG Test Strip, Synthgene Home Test HCG Test Cassette, and Synthgene Home Test HCG Test Midstream are substantially equivalent to the predicate.

§ 862.1155 Human chorionic gonadotropin (HCG) test system.

(a)
Human chorionic gonadotropin (HCG) test system intended for the early detection of pregnancy —(1)Identification. A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class II.(b)
Human chorionic gonadotropin (HCG) test system intended for any uses other than early detection of pregnancy —(1)Identification. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class III.(3)
Date PMA or notice of completion of a PDP is required. As of the enactment date of the amendments, May 28, 1976, an approval under section 515 of the act is required before the device described in paragraph (b)(1) may be commercially distributed. See § 862.3.