K Number
K242077

Validate with FDA (Live)

Date Cleared
2024-08-14

(29 days)

Product Code
Regulation Number
862.3100
Age Range
2 - 12
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

PocguideTM Multi-Drug Test Cup OTC is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Butalbital, Oxazepam, Cocaine, 2-ethylidene-1,5-diphenylpyrrolidine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline and Marijuana in human urine at the cutoff concentrations of:

Drug (Identifier) Amphetamine (AMP) Buprenorphine (BUP) Secobarbital (BAR) Oxazepam (BZO) Benzoylecgonine (COC) 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) Methamphetamine (MET) Methylenedioxymethamphetamine (MDMA) Morphine (OPI2000/MOP300) Methadone (MTD) Oxycodone (OXY) Phencyclidine (PCP) Nortriptyline (TCA) Marijuana (THC)
Cut-off level 1000 ng/mL or 500 ng/mL 10 ng/mL 300 ng/mL 300 ng/mL 300 ng/mL or 150 ng/mL 300 ng/mL 1000 ng/mL or 500 ng/mL 500 ng/mL 2000 ng/mL or 300 ng/mL 300 ng/mL 100 ng/mL 25 ng/mL 1000 ng/mL 50 ng/mL

The single or multi-test cups can consist of up to the above listed analytes in any combination. For over-the-counter use.

The tests provide only a preliminary result. A more specific alternative chemical method must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.

PocguideTM Multi-Drug Test Cup is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Butalbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline and Marijuana in human urine at the cutoff concentrations of:

Drug (Identifier)Cut-off level
Amphetamine (AMP)1000 ng/mL or 500 ng/mL
Buprenorphine (BUP)10 ng/mL
Secobarbital (BAR)300 ng/mL
Oxazepam (BZO)300 ng/mL
Benzoylecgonine (COC)300 ng/mL or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300 ng/mL
Methamphetamine (MET)1000 ng/mL or 500 ng/mL
Methylenedioxymethamphetamine (MDMA)500 ng/mL
Morphine (OPI2000/MOP300)2000 ng/mL or 300 ng/mL
Methadone (MTD)300 ng/mL
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Nortriptyline (TCA)1000 ng/mL
Marijuana (THC)50 ng/mL

The single or multi-test cups can consist of up to the above listed analytes in any combination.

The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.

The tests provide only a preliminary result. A more specific alternative chemical method must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.

Device Description

Pocguide™ Multi-Drug Test Cup OTC and Pocguide™ Multi-Drug Test Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine. The devices are a cup format. Each test device is sealed with sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.

AI/ML Overview

The document describes the analytical and user performance of the Pocguide™ Multi-Drug Test Cup OTC and Pocguide™ Multi-Drug Test Cup, which are qualitative lateral flow immunochromatographic assays for detecting various drugs in human urine.

Here's an analysis of the acceptance criteria and the studies performed, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria for the analytical performance (e.g., specific percentages for precision or method comparison). Instead, it presents the results of these studies. For the lay user study, it presents agreement percentages.

However, based on the data presented, the implicit acceptance criteria for the qualitative detection of drugs would be high agreement (ideally 100% at concentrations sufficiently above or below the cutoff) for positive and negative samples, and a certain degree of variability (discordance) near the cutoff, which is expected for qualitative assays.

Here's a summary of the reported performance, primarily from the "Precision/Reproducibility" and "Method Comparison" sections:

Implicit Acceptance Criteria (based on typical assay performance and the provided data)

  • Precision/Reproducibility: Consistent qualitative results (Positive/Negative) across multiple runs and lots, especially for concentrations sufficiently above or below the stated cutoff levels. Some variability (mix of positive and negative results) is expected and acceptable around the cutoff concentration, as this is the region where the device is designed to transition between positive and negative readings.
  • Method Comparison (Accuracy vs. LC-MS/MS):
    • 100% agreement for Drug-Free samples (true negatives).
    • High agreement for "Low Negative" samples (true negatives well below cutoff).
    • High agreement for "High Positive" samples (true positives well above cutoff).
    • Expected discordance (mix of +/- results) for "Near Cutoff Negative" and "Near Cutoff Positive" samples.
  • Lay Person Study: High percentage of agreement on results, demonstrating ease of use and accurate interpretation by lay users. High Flesch-Kincaid reading score/grade level for instructions indicating understandability.

Reported Device Performance

Precision/Reproducibility (See Tables in Section 12.A.a, e.g., Page 8)

  • For concentrations at +100%, +75%, +50%, +25% cutoff: Across all drugs and lots, the device consistently reported 0 negative results and 50 positive results (0-/50+) which indicates 100% agreement for samples well above the cutoff.
  • For concentrations at -25%, -50%, -75%, -100% cutoff: Across all drugs and lots, the device consistently reported 50 negative results and 0 positive results (50-/0+) which indicates 100% agreement for samples well below the cutoff.
  • For the "Cutoff" concentration: As expected for a qualitative test, there was a mix of negative and positive results across all drugs and lots, ranging from 8-/42+ to 17-/33+ (e.g., for AMP 500 Lot 1, 12-/38+ means 12 negative and 38 positive results out of 50 total). This demonstrates appropriate performance around the threshold.

Method Comparison (Accuracy vs. LC-MS/MS) (See Tables in Section 12.B, e.g., Page 19-20)

  • Drug-Free Samples: For all drugs and all three operators, the device consistently reported 0 positive results and showed 100% agreement (e.g., 0+ / 13- for AMP 500).
  • Low Negative Samples (less than -50% of cutoff): For all drugs and all operators, the device consistently reported 0 positive results and showed 100% agreement (e.g., 0+ / 7- for AMP 500).
  • High Positive Samples (greater than +50% of cutoff): For all drugs and all operators, the device consistently reported 100% positive results (e.g., 23+ / 0- for AMP 500).
  • Near Cutoff Negative (Between -50% and cutoff) & Near Cutoff Positive (Between cutoff and +50%): As expected for a qualitative assay, there was a mix of positive and negative results in these ranges, indicating the test's activity around the cutoff. The discordant results table (Pages 22-23) provides specific examples of samples that differ from the LC-MS/MS result near the cutoff. For instance, for AMP 500, samples like AL243 (441.6 ng/mL) were read as Positive by the device, while AL036 (501.2 ng/mL) was read as Negative. This is typical for qualitative tests at the cutoff.

Lay Person Study (See Tables in Section 12.C, e.g., Page 25-27)

  • Agreement Percentages: For concentrations well below (-100%, -75%, -50% cutoff) and well above (+50%, +75% cutoff) the cutoff, the agreement with the expected result was consistently 100% for all drugs and configurations.
  • Near Cutoff: At the -25% and +25% cutoff concentrations, there was slight variability, with agreement mostly at 95.00% (e.g., 19 negative out of 20 at -25% for AMP 500, or 19 positive out of 20 at +25% for AMP 500). This confirms appropriate performance around the cutoff by lay users.
  • Instruction Clarity: "All participants indicated that the device instruction is easy to understand and follow. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7," indicating strong performance against this unstated criterion.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Precision/Reproducibility:
    • Sample Size: For each drug, 9 concentration levels were tested (from -100% cutoff to +100% cutoff, including the cutoff). For each concentration, tests were performed two runs per day for 25 days using three lots of test cups. This means 50 tests per concentration per lot, totaling 450 tests per drug per lot for analytical precision. As there are multiple drugs/cutoffs, the total number of individual tests is substantial. (e.g., for AMP 500, 9 concentrations * 50 tests/concentration * 3 lots = 1350 individual tests for AMP 500 alone).
    • Data Provenance: The document doesn't explicitly state the country of origin. Samples were "prepared by spiking target drug in drug-free urine samples," suggesting controlled laboratory conditions rather than native clinical samples. The study appears to be prospective in nature, as it involves preparing samples and running specific experiments.
  • Method Comparison Study:
    • Sample Size: 80 "unaltered urine clinical samples" were used for each drug (40 negative and 40 positive). These were compared across three operators. Thus, for each drug, 80 unique clinical samples were tested, with data collected from 3 operators, meaning 240 results per drug.
    • Data Provenance: "Unaltered urine clinical samples" suggests real-world urine specimens. The document doesn't specify the country of origin of these samples. It implies a retrospective use of collected samples or a prospective collection for this specific study.
  • Lay Person Study:
    • Sample Size: 280 lay persons participated. Urine samples were prepared at 7 concentration levels per drug (from -100% to +75% cutoff).
    • Data Provenance: The study was likely conducted in a controlled environment as samples were "prepared by spiking drug(s) into drug free-pooled urine specimens." The location (country) of the lay user study is not specified, but it suggests a controlled, prospective study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Ground Truth for Precision/Reproducibility and Lay Person Study: "Each drug concentration was confirmed by LC-MS/MS." and "The concentrations of the samples were confirmed by LC-MS/MS." LC-MS/MS (Liquid Chromatography-Mass Spectrometry/Mass Spectrometry) is a highly accurate analytical chemistry technique. This is an objective, quantitative measurement; therefore, human expert subjective interpretation for ground truth establishment is not typically required.
  • Ground Truth for Method Comparison Study: "The samples were blind labeled and compared to LC-MS/MS results." Again, LC-MS/MS is the ground truth, not human experts.

As such, for an in-vitro diagnostic device of this nature, the "ground truth" is established by highly precise analytical instruments (LC-MS/MS), not by human experts. Therefore, the number and qualifications of human experts establishing ground truth are not applicable in the traditional sense of medical image interpretation or clinical diagnosis.


4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

Not applicable. The ground truth method (LC-MS/MS) for defining the precise concentration of drugs is an objective chemical analysis, not a subjective human interpretation requiring adjudication. For the device's output, it is a qualitative "positive" or "negative" reading based on visual lines, directly observable without complex adjudication. The method comparison study uses results from 3 operators, but their results are compared against the LC-MS/MS, not adjudicated against each other to define a 'truth'.


5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is an in-vitro diagnostic (IVD) device, specifically a point-of-care, qualitative multi-drug test cup, not an AI-assisted diagnostic tool for human image readers (like radiology AI). Therefore, an MRMC study related to AI assistance for human readers is not relevant to this product.


6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical, self-contained immunochromatographic test cup that provides a visual reading (lines appearing/disappearing). It does not involve a software algorithm that performs detection independently. Its performance is the "standalone" performance, as it relies on chemical reactions and visual interpretation.


7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The primary ground truth used for performance evaluation is Gas Chromatography-Mass Spectrometry (GC-MS) or Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions (LC-MS/MS), which are considered the preferred confirmatory methods for drug detection in urine. This is an objective, quantitative chemical analysis.


8. The sample size for the training set

Not applicable. This is a direct chemical/immunological assay device, not a machine learning or AI model. Therefore, there is no "training set" in the context of data-driven model development.


9. How the ground truth for the training set was established

Not applicable, as there is no "training set" for this type of medical device. The device operates based on established chemical and immunological principles, not learning from data.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the Department of Health & Human Services logo. To the right of the HHS logo is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

Aicheck Biotech, Inc. % Jenny Xia Director LSI International Inc 504E Diamond Ave., Suite H Gaithersburg, Maryland 20877

Re: K242077

Trade/Device Name: Pocguide™ Multi-Drug Test Cup OTC, Pocguide™ Multi-Drug Test Cup Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine Test System Regulatory Class: Class II Product Code: NFT Dated: July 12, 2024 Received: July 16, 2024

Dear Jenny Xia:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Joseph A. Digitally signed by Kotarek -S Date: 2024.08.14
10:57:33 -04'00'

Joseph Kotarek, Ph.D. Branch Chief Division of Chemistryand Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K242077 Device Name

PocquideTM Multi-Drug Test Cup OTC; PocquideTM Multi-Drug Test Cup

Indications for Use (Describe)

PocguideTM Multi-Drug Test Cup OTC is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Butalbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dipheny|pyrrolidine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline and Marijuana in human urine at the cutoff concentrations of:

  • Drug (Identifier) Amphetamine (AMP) Buprenorphine (BUP) Secobarbital (BAR) Oxazepam (BZO) Benzoylecgonine (COC) 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) Methamphetamine (MET) Methylenedioxymethamphetamine (MDMA) Morphine (OPI2000/MOP300) Methadone (MTD) Oxycodone (OXY) Phencyclidine (PCP) Nortriptyline (TCA) Marijuana (THC)
    Cut-off level 1000 ng/mL or 500 ng/mL 10 ng/mL 300 ng/mL 300 ng/mL 300 ng/mL or 150 ng/mL 300 ng/mL 1000 ng/mL or 500 ng/mL 500 ng/mL 2000 ng/mL or 300 ng/mL 300 ng/mL 100 ng/mL 25 ng/mL 1000 ng/mL 50 ng/mL

The single or multi-test cups can consist of up to the above listed analytes in any combination. For over-the-counter use.

The tests provide only a preliminary result. A more specific alternative chemical method must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.

PocquideTM Multi-Drug Test Cup is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Butalbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-dipheny|pyrrolidine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline and Marijuana in human urine at the cutoff concentrations of:

Drug (Identifier)Cut-off level
Amphetamine (AMP)1000 ng/mL or 500 ng/mL
Buprenorphine (BUP)10 ng/mL
Secobarbital (BAR)300 ng/mL
Oxazepam (BZO)300 ng/mL

{3}------------------------------------------------

Benzoylecgonine (COC)300 ng/mL or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300 ng/mL
Methamphetamine (MET)1000 ng/mL or 500 ng/mL
Methylenedioxymethamphetamine (MDMA)500 ng/mL
Morphine (OPI2000/MOP300)2000 ng/mL or 300 ng/mL
Methadone (MTD)300 ng/mL
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Nortriptyline (TCA)1000 ng/mL
Marijuana (THC)50 ng/mL

The single or multi-test cups can consist of up to the above listed analytes in any combination.

The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.

The tests provide only a preliminary result. A more specific alternative chemical method must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
--------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY K242077

1.Date:July 11, 2024
2.Submitter:Aicheck Biotech, Inc.17701 Cowan, Ste 230Irvine, CA 92614
3.Contact person:Jenny XiaLSI International Inc.504 East Diamond Ave., Suite HGaithersburg, MD 20877Telephone: 301-525-6856Email: jxia@lsi-consulting.org
4.Device Name:PocguideTM Multi-Drug Test Cup OTCPocguideTM Multi-Drug Test Cup
5.Classification:Class II
Product CodeTarget DrugRegulation SectionPanel
NFTAmphetamine (AMP)862.3100, Amphetamine Test SystemToxicology
PTHSecobarbital (BAR)862.3150, Barbiturate Test SystemToxicology
NGLBuprenorphine (BUP)Morphine (MOP/OPI)Oxycodone (OXY)862.3650, Opiate Test SystemToxicology
NFVOxazepam (BZO)862.3170, Benzodiazepine Test SystemToxicology
NFYCocaine (COC)862.3250, Cocaine and cocainemetabolite test systemToxicology
PTGMethadone (MTD)2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)862.3620, Methadone Test SystemToxicology
NGGMethylenedioxymethamphetamine(MDMA)Methamphetamine (MET)862.3610,Methamphetamine Test SystemToxicology
NGMPhencyclidine (PCP)UnclassifiedToxicology
QAWNortriptyline (TCA)862.3910 Tricyclic antidepressant drugstest systemToxicology
NFW862.3870, Cannabinoids Test SystemToxicology

{5}------------------------------------------------

Cannabinoids (THC)
And Comments of Children Comments of Children Comments of Children

Predicate Devices: 6.

Wondfo T-Cup® Multi-Drug Urine Test Cup (K182701)

7. Intended Use

Pocguide™ Multi-Drug Test Cup OTC is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Butalbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3diphenylpyrrolidine, Methamphetamine, Methylenedioxy-methamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline and Marijuana in human urine at the cutoff concentrations of:

Drug (Identifier)Cut-off level
Amphetamine (AMP)1000 ng/mL or 500 ng/mL
Buprenorphine (BUP)10 ng/mL
Secobarbital (BAR)300 ng/mL
Oxazepam (BZO)300 ng/mL
Benzoylecgonine (COC)300 ng/mL or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine(EDDP)300 ng/mL
Methamphetamine (MET)1000 ng/mL or 500 ng/mL
Methylenedioxymethamphetamine (MDMA)500 ng/mL
Morphine (OPI2000/MOP300)2000 ng/mL or 300 ng/mL
Methadone (MTD)300 ng/mL
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Nortriptyline (TCA)1000 ng/mL
Marijuana (THC)50 ng/mL

The single or multi-test cups can consist of up to the above listed analytes in any combination. For over-the-counter use.

The tests provide only a preliminary result. A more specific alternative chemical method must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem massspectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.

{6}------------------------------------------------

Pocguide™ Multi-Drug Test Cup is competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Butalbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methamphetamine, Methylenedioxy-methamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline and Marijuana in human urine at the cutoff concentrations of:

Drug (Identifier)Cut-off level
Amphetamine (AMP)1000 ng/mL or 500 ng/mL
Buprenorphine (BUP)10 ng/mL
Secobarbital (BAR)300 ng/mL
Oxazepam (BZO)300 ng/mL
Benzoylecgonine (COC)300 ng/mL or 150 ng/mL
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine(EDDP)300 ng/mL
Methamphetamine (MET)1000 ng/mL or 500 ng/mL
Methylenedioxymethamphetamine (MDMA)500 ng/mL
Morphine (OPI2000/MOP300)2000 ng/mL or 300 ng/mL
Methadone (MTD)300 ng/mL
Oxycodone (OXY)100 ng/mL
Phencyclidine (PCP)25 ng/mL
Nortriptyline (TCA)1000 ng/mL
Marijuana (THC)50 ng/mL

The single or multi-test cups can consist of up to the above listed analytes in any combination.

The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.

The tests provide only a preliminary result. A more specific alternative chemical method must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem massspectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.

Device Description 8.

Pocguide™ Multi-Drug Test Cup OTC and Pocguide™ Multi-Drug Test Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.

{7}------------------------------------------------

The devices are a cup format. Each test device is sealed with sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.

ItemProposed DevicePredicate(K182701)
Indication(s) for useFor the qualitative determination of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline and Marijuana in human urine.Same (but the number of drugs detected is different)
MethodologyCompetitive binding, lateral flow immunochromatographic assay based on antigen-antibody reactionSame
Type of TestQualitativeSame
Specimen TypeHuman urineSame
Target Drug andCut Off ValuesTarget DrugCutoff (ng/mL)Same (but the number of drugs detected is different)
Amphetamine (AMP)1000 or 500
Buprenorphine (BUP)10
Secobarbital (BAR)300
Oxazepam (BZO)300
Cocaine (COC)300 or 150
2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)300
Methamphetamine (MET)1000 or 500
Methylenedioxymethamphetamine(MDMA)500
Morphine (MOP 300/OPI 2000)2000 or 300
Methadone (MTD)300
Oxycodone (OXY)100
Phencyclidine (PCP)25
Nortriptyline (TCA)1000
Marijuana (THC 50)50
ConfigurationsTest cupSame
Intended UseFor over-the-counter useSame

Substantial Equivalence Information 9.

10. Standard/Guidance Document Reference (if applicable)

None referenced.

{8}------------------------------------------------

11. Test Principle

Pocguide™ Multi-Drug Test Cup OTC or Pocguide™ Multi-Drug Test Cup is a competitive immunoassay that is used to screen for the presence of drugs of abuse in urine. It is chromatographic absorbent device in which drugs or drug metabolites in a sample competitively combined to a limited number of antibody-dye conjugate binding sites. When the absorbent end of the test device is immersed into the urine sample, the urine is absorbed into the device by capillary action, mixes with the antibody-dye conjugate, and flows across the precoated membrane.

When sample drug levels are at or above the target cutoff (the detection sensitivity of the test), the drug in the sample binds to the antibody-dye conjugate preventing the antibody-dye conjugate from binding to the drug-protein pre-coated in the test region (T). This prevents the development of a distinct colored band in the test region indicating a potentially positive result. When sample drug levels are zero or below the target cutoff, antibody-dye conjugate binds to the drug-protein pre-coated in the test region (T) of the device. This produces a colored test line that, regardless of its intensity, indicates a negative result.

To serve as a procedural control, a colored line will always appear at the control region (C) indicating that proper volume of specimen has been added and membrane wicking has occurred.

12. Performance Characteristics

A. Analytical performance

a. Precision/Reproducibility:

Precision studies were carried out for samples with concentrations of +100% cutoff, +75% cutoff, +50% cutoff, +25% cutoff, cutoff, -25% cutoff, -50% cutoff, -75% cut off and -100% cutoff. Samples with concentration of -100% cutoff were drug-free urine samples. Other samples were prepared by spiking target drug in drug-free urine samples. Each drug concentration was confirmed by LC-MS/MS. For each concentration, tests were performed two runs per day for 25 days using three lots of test cups. The results obtained are summarized in the following tables:

DrugLotNumber+100%cutoff+75%cutoff+50%cutoff+25%cutoffCutoff-25%cutoff-50%cutoff-75%cutoff-100%cut-off
AMP500Lot 10-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
500Lot 20-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+13-/37+50-/0+50-/0+50-/0+50-/0+
AMP1000Lot 10-/50+0-/50+0-/50+0-/50+13-/37+50-/0+50-/0+50-/0+50-/0+
1000Lot 20-/50+0-/50+0-/50+0-/50+13-/37+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+15-/35+50-/0+50-/0+50-/0+50-/0+
BAR300Lot 10-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
300Lot 20-/50+0-/50+0-/50+0-/50+13-/37+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+15-/35+50-/0+50-/0+50-/0+50-/0+
BUP 10Lot 10-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+13-/37+50-/0+50-/0+50-/0+50-/0+
BZOLot 10-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
300Lot 20-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+9-/41+50-/0+50-/0+50-/0+50-/0+
MDMALot 10-/50+0-/50+0-/50+0-/50+14-/36+50-/0+50-/0+50-/0+50-/0+
500Lot 20-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+13-/37+50-/0+50-/0+50-/0+50-/0+
METLot 10-/50+0-/50+0-/50+0-/50+8-/42+50-/0+50-/0+50-/0+50-/0+
500Lot 20-/50+0-/50+0-/50+0-/50+15-/35+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
METLot 10-/50+0-/50+0-/50+0-/50+14-/36+50-/0+50-/0+50-/0+50-/0+
1000Lot 20-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+13-/37+50-/0+50-/0+50-/0+50-/0+
MOPLot 10-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
300Lot 20-/50+0-/50+0-/50+0-/50+15-/35+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+14-/36+50-/0+50-/0+50-/0+50-/0+
OPILot 10-/50+0-/50+0-/50+0-/50+13-/37+50-/0+50-/0+50-/0+50-/0+
2000Lot 20-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+13-/37+50-/0+50-/0+50-/0+50-/0+
MTDLot 10-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
300Lot 20-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
OXYLot 10-/50+0-/50+0-/50+0-/50+9-/41+50-/0+50-/0+50-/0+50-/0+
100Lot 20-/50+0-/50+0-/50+0-/50+8-/42+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
PCP 25Lot 10-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
Lot 20-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
TCALot 10-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
1000Lot 20-/50+0-/50+0-/50+0-/50+12-/38+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
THCLot 10-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
50Lot 20-/50+0-/50+0-/50+0-/50+8-/42+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+9-/41+50-/0+50-/0+50-/0+50-/0+
COCLot 10-/50+0-/50+0-/50+0-/50+16-/34+50-/0+50-/0+50-/0+50-/0+
150Lot 20-/50+0-/50+0-/50+0-/50+17-/33+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+15-/35+50-/0+50-/0+50-/0+50-/0+
COCLot 10-/50+0-/50+0-/50+0-/50+9-/41+50-/0+50-/0+50-/0+50-/0+
300Lot 20-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+13-/37+50-/0+50-/0+50-/0+50-/0+
EDDPLot 10-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
300Lot 20-/50+0-/50+0-/50+0-/50+11-/39+50-/0+50-/0+50-/0+50-/0+
Lot 30-/50+0-/50+0-/50+0-/50+10-/40+50-/0+50-/0+50-/0+50-/0+

{9}------------------------------------------------

{10}------------------------------------------------

b. Linearity/assay reportable range:

Not applicable. This device is intended for qualitative use only.

c. Stability:

The device is stable at 4-30℃ for 24 months based on accelerated stability study.

d. Analytical specificity/Interference:

To test the specificity, drug metabolites and other components that are likely to cross-react in urine samples were spiked into drug-free urine. These urine samples were tested using three lots of the device.

Percent cross-reactivity, provided in the below table, was calculated as the cutoff concentration divided by the concentration of analyte tested that yielded a positive result, multiplied by 100.

Drug/CutoffCompoundMinimumconcentrationrequired to obtaina positive result(ng/mL)% Cross-Reactivity
d-Amphetamine500100%
l-Amphetamine150003.3%
dl- Amphetamine150033.3%
(+/-) 3,4-Methylenedioxyamphetamine(MDA)500010%
AMP 500Phentermine150033.3%
Hydroxyamphetamine80006.25%
d-Methamphetamine>100000-
l-Methamphetamine>100000-
(+/-) 3,4-Methylenedioxyethylamphetamine(MDE)>100000-
(+/-)3,4-Methylenedioxymethamphetamine(MDMA)
β-Phenylethylamine1000000.5%
Tyramine1000000.5%
p-Hydroxynorephedrine1000000.5%
Phenylpropanolamine>100000-
(±)Phenylpropanolamine>100000-
p-Hydroxyamphetamine1000000.5%
d/l-Norephedrine1000000.5%
Benzphetamine>100000-
1-Ephedrine>100000-
l-Epinephrine>100000-
d/l-Epinephrine>100000-
d-Amphetamine1000100%
l-Amphetamine300003.3%
dl- Amphetamine300033.3%
(+/-)3,4-Methylenedioxyamphetamine (MDA)500020%
Phentermine300033.3%
Hydroxyamphetamine800012.5%
d-Methamphetamine>100000-
l-Methamphetamine>100000-
AMP 1000(+/-)3,4-Methylenedioxyethylamphetamine(MDEA)>100000-
(+/-)3,4-Methylenedioxymethamphetamine(MDMA)>100000-
β-Phenylethylamine1000001%
Tyramine1000001%
p-Hydroxynorephedrine1000001%
Phenylpropanolamine>100000-
(±)Phenylpropanolamine>100000--
p-Hydroxyamphetamine1000001%
d/l-Norephedrine1000001%
Benzphetamine>100000-
1-Ephedrine>100000-
1-Epinephrine>100000-
d/l-Epinephrine>100000-
Secobarbital300100%
Amobarbital50006%
Alphenol150200%
Aprobarbital200150%
BAR 300Butabarbital150200%
Butethal50600%
Butalbital250012%
Cyclopentobarbital60050%
Pentobarbital200015%
Phenobarbital50006%
Buprenorphine10100%
Buprenorphine -3-D-Glucuronide1566.67%
Norbuprenorphine2050%
BUP 10Norbuprenorphine-3-D-Glucuronide10010%
Glucuronide
Morphine> 100000-
Oxymorphone> 100000-
Hydromorphone> 100000-
Oxazepam300100%
Alprazolam200150%
a-Hydroxyalprazolam100030%
BZO 300Bromazepam50060%
Chlordiazepoxide150020%
Clobazam100300%
Clonazepam300010%
Clorazepate dipotassium200150%
Delorazepam150020%
Desalkylflurazepam40075%
Diazepam200150%
Estazolam50060%
Flunitrazepam100030%
Midazolam50006%
Nitrazepam100030%
Norchlordiazepoxide200150%
Nordiazepam50060%
Temazepam250120%
Triazolam120025%
Demoxepam200015%
Flurazepam50060%
D,L-Lorazepam150020%
Benzoylecgonine150100%
Cocaine50030%
Cocaethylene62502.4%
COC 150Ecgonine160000.94%
Ecgonine methyl ester>100000-
Norcocaine>100000-
Benzoylecgonine300100%
Cocaine100030%
COC 300Cocaethylene150002%
Ecgonine300001%
Ecgonine methyl ester>100000-
Norcocaine>100000-
3,4-Methylenedioxymethamphetamine(MDMA)500100%
MDMA 5003,4-Methylenedioxyamphetamine HCl (MDA)300016.67%
3,4-Methylenedioxyethylamphetamine(MDEA)500100%
d-methamphetamine> 100000-
d-amphetamine> 100000-
l-methamphetamine500001%
l-amphetamine> 100000-
D(+)-Methamphetamine500100%
(+/-)3,4-Methylenedioxy-n-ethylamphetamine(MDEA)200025%5%
D/L-Methamphetamine500100%
p-Hydroxymethamphetamine150003.33%
D-Amphetamine500001%
L-Amphetamine1000000.5%
Chloroquine100005%
(+/-)-Ephedrine250002%
(-)-Methamphetamine125004%
MET 500(+/-)3,4-Methylenedioxyamphetamine(MDA)200025%
(+/-)3,4-Methylenedioxymethamphetamine(MDMA)200025%
β-Phenylethylamine250002%
Trimethobenzamide500010%
d,l-Amphetamine750000.67%
Mephetermine250002%
(1R,2S)-(-)-Ephedrine500001%
l-phenylephrine1000000.5%
L-Methamphetamine100005%
D(+)-Methamphetamine1000100%
MET 1000(+/-)3,4-Methylenedioxy-n-ethylamphetamine (MDEA)200050%
D/L-Methamphetamine1000100%
p-Hydroxymethamphetamine300003.3%
D-Amphetamine> 100000-
L-Amphetamine1000001%
Chloroquine200005%
(+/-)-Ephedrine500002%
(-)-Methamphetamine250004%
(+/-)3,4-Methylenedioxyamphetamine(MDA)250040%
(+/-)3,4-Methylenedioxymethamphetamine(MDMA)400025%
β-Phenylethylamine500002%
Trimethobenzamide1000010%
d,l-Amphetamine1000001%
Mephetermine500002%
(1R,2S)-(-)-Ephedrine> 100000-
l-phenylephrine> 100000-
L-Methamphetamine200005%
Morphine300100%
Normorphine250120%
Codeine300100%
s-Monoacetylmorphine300100%
Ethyl Morphine100300%
Heroin300100%
Hydrocodone50006%
Hydromorphone200015%
MOP 300Morphinie-3-β-d-glucuronide100030%
Oxycodone>100000--
Oxymorphone1000000.3%
Thebaine300010%
Levorphanol50006%
6-Monoacetylmorphine (6-MAM)150200%
Norcodeine65004.6%
Procaine1000000.3%
Morphine2000100%
OPI 2000Normorphine500004%
Codeine2000100%
s-Monoacetylmorphine2000100%
Ethyl Morphine1500133.3%
Heroin2000100%
Hydrocodone1250016%
Hydromorphone350057.1%
Morphinie-3-β-d-glucuronide2000100%
Oxycodone250008%
Oxymorphone250008%
Thebaine500040%
Levorphanol750002.7%
6-Monoacetylmorphine (6-MAM)1500133.3%
Norcodeine1250016%
Procaine1500001.3%
MTD 300Methadone300100%
Doxylamine300010%
EDDP>100000-
EMDP>100000-
LAAM>100000-
Alpha Methadol>100000-
OXY 100Oxycodone100100%
Dihydrocodeine200000.5%
Hydrocodone100001%
Oxymorphone100010%
Codeine1000000.1%
Hydromorphone320000.3%
Morphine>100000-
Acetylmorphine>100000-
Buprenorphine>100000-
Ethylmorphine>100000-
Thebaine>100000-
PCP 25Phencyclidine25100%
4-Hydroxyphencyclidine125000.2%
TCA 1000Nortriptyline1000100%
Nordoxepine1000100%
Trimipramine200050%
Amitriptyline150066.7%
Promazine150066.7%
Desipramine400250%
Imipramine400250%
Clomipramine125008%
Doxepin1000100%
Maprotiline200050%
Promethazine250004%
Cyclobenzaprine150066.7%
Norclomipramine1000010%
THC 5011-nor-Δ9-THC-9-COOH50100%
(-)-11-nor-9-carboxy-Δ9-THC50100%
11-nor-Δ8-THC-9-COOH30166.67%
11-nor-Δ9-THC-carboxyglucuronide10050%
Cannabinol200000.25%
Cannabidiol1000000.05%
Δ8- Tetrahydrocannabinol13003.8%
Δ9- Tetrahydrocannabinol50001%
11-hydroxy-Δ9-Tetrahydrocannabinol50001%
EDDP 3002-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine300100%
Methadone2000000.15%
EMDP3000000.1%
Doxylamine>100000-
Disopyramide>100000-
LAAM (Levo-alpha-acetylmethadol) HCl>100000-
Alpha Methadol>100000-

{11}------------------------------------------------

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

{16}------------------------------------------------

{17}------------------------------------------------

To evaluate potential interference, non-structurally related compounds were added to drug-free urine and to urine samples containing the target drugs at 25% below and 25% above each corresponding cutoff.

{18}------------------------------------------------

Compounds that show no interference at a concentration of 100μg/mL are summarized in the following table.

following table.
(-) CotinineDuloxetineNimodipine
3-HydroxytyramineEffexorNitroglycerin
AcetaminophenEnalapril MaleateNorethindrone
AcetophenetidinErythromycinO-Hydroxyhippuric Acid
Acetylsalicylic AcidEsomeprazole MagnesiumOlanzapine
AcyclovirEthanol (1%)Omeprazole
AfrinFenofibrateOndansetran
Albumin (100mg/dL)FenoprofenOxalic Acid
AminophyllineFentanyl CitrateOxolinic Acid
AminopyrineFluoxetine HydrochlorideOxymetazoline
Amiodarone HydrochlorideFluvoxaminePaliperidone
Amlodipine MesylateFurosemidePantoprazole
AmoxicillinGabapentinPapaverine
AmpicillinGentisic AcidParoxetine Hydrochloride
ApomorphineGlibenclamidePenfluridol
AripiprazoleGliclazidePenicillin-G
AspartameGlipizidePenicillinV Potassium
AtomoxetineGlucosePhenelzine
Atorvastatin CalciumHaloperidolPioglitazone Hydrochloride
AtropineHemoglobinPiracetam
Benzilic AcidHydrochlorothiazidePravastatin Sodium
Benzoic AcidHydrocortisonePrednisone
BilirubinIbuprofenPropylthiouracil
BupropionIsosorbide DinitrateQuetiapine Fumarate
CaptoprilIsoxsuprineQuinine
CarbamazepineKetamineRanitidine
CefradineKetoconazoleRifampicin
CephalexinKetoprofenRisperidone
Chloral HydrateKratom powderSalicylic Acid
ChloramphenicolLabetalolSerotonin
ChlorothiazideLamotrigineSertraline Hydrochloride
chlorpheniramineLevofloxacin HydrochlorideSildenafil Citrate
CholesterolLevonorgestrelSimvastatin
Ciprofloxacin HydrochlorideLevothyroxine SodiumSodium Valproate
CitalopramLidocaine HydrochlorideSpironolactone
ClarithromycinLisinoprilSulfamethazine
ClonidineLithium CarbonateSulindac
Clopidogrel HydrogenSulphateLiveriteTetracycline
ClozapineLoperamideTetrahydrocortisone 3-(β-D glucuronide)
Conjugated EstrogensLoratadineTetrahydrocortsone 3 -acetate
CortisoneMagnesiumTetrahydrozoline
CreatinineMeperidineThiamine
D,L-OctopamineMeprobamateThioridazine
D,L-PropranololMetoprolol TartrateTopiramate
D,L-TyrosineMifepristoneTramadol Hydrochloride
Deoxy- corticosteroneMinocyclineTrazodone Hydrochloride
DextromethorphanMirtazapineTriamterene
DiclofenacMontelukast SodiumTrifluoperazine
DicyclomineMosapride CitrateTrimethoprim
DiflunisalN-Acetylprocain-amideUric Acid
DigoxinNalidixic AcidValproate
DiphenhydramineNaproxenVerapamil
DirithromycinNiacinamideVitamin B2
DomperidoneNifedipineVitamin C
D-PseudoephedrineNikethamideβ-Estradiol

{19}------------------------------------------------

Interference by pH and specific gravity were also evaluated using pooled urine specimens with concentrations at 25% below and 25% above each corresponding cutoff. The results demonstrated that pH levels of 4 to 9 and specific gravity levels of 1.000 to 1.035 do not affect the results of the assays.

B. Method comparison study

The method comparison studies for the device were performed with three operators. Operators ran 80 (40 negative and 40 positive) unaltered urine clinical samples for each drug. The samples were blind labeled and compared to LC-MS/MS results. The results are presented in the table below:

DrugtestTest CupResultDrug-FreeLowNegative byLC-MS/MS(less than -50%)Near CutoffNegative byLC-MS/MS(Between -50% and theCutoff)Near CutoffPositive byLC-MS/MS(Betweenthe cutoffand +50%)High Positiveby LC-MS/MS(greater than+50%)
AMP500OperatorA+0031623
-1371710
OperatorB+0041623
-1371610
OperatorC+0011523
-1371920
Operator+0011525
AMP1000A-7151700
Operator+0011425
B-7151710
Operator+0011425
C-7151710
BAROperator+0011819
A-10151430
Operator+0001819
B-10151530
Operator+0021919
C-10151320
BUPOperator+0032018
A-10151220
Operator+0042018
B-10151120
Operator+0032018
C-10151220
BZOOperator+0021820
A-10151320
Operator+0021820
B-10151320
Operator+0031820
C-10151220
COC150Operator+0002115
A-10151540
Operator+0002215
B-10151530
Operator+0002115
C-10151540
COC300Operator+0001821
A-9161510
Operator+0001721
B-9161520
Operator+0001721
C-9161520
MDMAOperator+0002018
A-10151520
Operator+0022118
B-10151310
Operator+0001918
C-10151530
Operator+0002016
META-8201240
500Operator+0002016
B-8201240
Operator+0012216
C-8201120
METOperator+0031822
1000A-8131600
Operator+0021722
B-8131710
Operator+0021622
C-8131720
MOPOperator+0012315
A-10141520
Operator+0012215
B-10141530
Operator+0032315
C-10141320
OPIOperator+0012118
A-10151410
Operator+0022118
B-10151310
Operator+0022018
C-10151320
MTDOperator+0011424
A-10181120
Operator+0001424
B-10181220
Operator+0011424
C-10181120
OXYOperator+0042515
A-10151100
Operator+0032415
B-10151210
Operator+0022315
C-10151320
PCPOperator+0011918
A-9161430
Operator+0001918
B-9161530
Operator+0021818
C-9161340
TCAOperator+0032018
A-10151220
Operator+0032118
B-10151210
Operator+0032018
C-10151220
THCOperator+0011820
A-10151420
Operator+0001720
B-10151530
Operator+0021820
C-10151320
EDDPOperator+0032020
A-10141300
Operator+0021820
B-10141420
Operator+0031920
C-10141310

{20}------------------------------------------------

{21}------------------------------------------------

{22}------------------------------------------------

Discordant Results are summarized below.

DrugOperatorSample NumberLC/MS/MS Result (ng/mL)Discordant Result
AMP 500Operator BAL243441.6Positive
AMP 500Operator A, B, CAL073443.7Positive
AMP 500Operator A, BAL251457Positive
AMP 500Operator BAL022475.7Positive
AMP 500Operator AAL143489Positive
AMP 500Operator CAL036501.2Negative
AMP 500Operator A, B, CAL212508.9Negative
AMP 1000Operator CAH051910.3Positive
AMP 1000Operator BAH070968.6Positive
AMP 1000Operator AAH066970.1Positive
AMP 1000Operator B, CAH0021101Negative
BAROperator CBR328279.6Positive
BAROperator A, CBR311296Positive
BAROperator A, BBR321302.1Negative
BAROperator A, B, CBR440307.2Negative
BAROperator BBR226319.83Negative
BAROperator A, CBR227323Negative
BUPOperator A, B, CBP4497.73Positive
BUPOperator A, BBP3068.02Positive
BUPOperator A, B, CBP3299.21Positive
BUPOperator B, CBP3309.42Positive
Operator A, B, CBP27710.5Negative
Operator A, B, CBP45011.3Negative
BZOOperator A, B, CBZ201282Positive
Operator A, B, CBZ213291.2Positive
Operator CBZ257298Positive
Operator BBZ204301Negative
Operator CBZ197305Negative
Operator A, BBZ356317.6Negative
Operator A, CBZ200346Negative
COC 150Operator A, BCL534154Negative
Operator A, B, CCL530156.5Negative
Operator CCL537160.1Negative
Operator CCL573161.4Negative
Operator ACL565163.5Negative
Operator A, B, CCL575166.9Negative
COC 300Operator A, B, CCH513338.5Negative
Operator BCH572346.4Negative
Operator CCH553356.8Negative
MDMAOperator BMM238476.1Positive
Operator BMM305482.34Positive
Operator A, CMM303502Negative
Operator AMM278527.89Negative
Operator CMM284536.21Negative
Operator BMM263543.6Negative
Operator CMM266605Negative
MET 500Operator CML648456.2Positive
Operator BML665506Negative
Operator A, B, CML652511Negative
Operator BML657512Negative
Operator A, B, CML671534Negative
Operator AML662543Negative
Operator AML668564.2Negative
MET 1000Operator A, CMH623962.6Positive
Operator A, B, CMH667981.1Positive
Operator A, BMH651995.8Positive
Operator CMH7671009.7Negative
Operator B, CMH6251083.2Negative
MOPOperator CMO484246Positive
Operator CMO488272.3Positive
Operator A, B, CMO431298Positive
Operator A, B, CMO420301Negative
Operator A, B, CMO374301.6Negative
Operator BMO424313.3Negative
OPIOperator A, B, COP3841869.3Positive
Operator B, COP3791984Positive
Operator COP4722028Negative
Operator A, B, COP3732080Negative
MTDOperator A, CMT753278.8Positive
Operator A, B, CMT709326.5Negative
Operator A, B, CMT872329.8Negative
OXYOperator AOX28677Positive
Operator A, BOX29692.3Positive
Operator A, B, COX27295Positive
Operator A, B, COX25396Positive
Operator COX267100Negative
Operator BOX277102Negative
Operator COX263103Negative
PCPOperator A, CPC38022.257Positive
Operator CPC34123.97Positive
Operator A, B, CPC37125.39Negative
Operator A, B, CPC42526.4Negative
Operator A, B, CPC34027.2Negative
Operator CPC37627.5Negative
TCAOperator A, CTA419817Positive
Operator BTA407849Positive
Operator A, B, CTA431892Positive
Operator A, B, CTA379934Positive
Operator A, BTA4291025Negative
Operator CTA4221069Negative
Operator A, CTA3911117Negative
THCOperator A, CTH36047.7Positive
Operator CTH37448.8Positive
Operator BTH37350.35Negative
Operator A, B, CTH33853.46Negative
Operator A, B, CTH33257.1Negative
EDDPOperator A, B, CED725280.8Positive
Operator A, B, CED709281Positive
Operator A, CED694281.4Positive
Operator BED690325Negative
Operator CED724335Negative
Operator BED717335.7Negative

{23}------------------------------------------------

{24}------------------------------------------------

C. Lay person study

A lay user study was performed at three intended user sites with 280 lay persons. 114 males and 166 females tested Pocguide™ Multi-Drug Test Cup. They had diverse educational and professional backgrounds and their ages ranged from 18 to > 50. The 280 lay persons were split

{25}------------------------------------------------

into two groups of 140 whereby group 1 tested configuration 1 of the cup and group 2 tested configuration 2 of the cup. Urine samples were prepared at the following concentrations: -100%, +/-75%, +/-50%, +/-25% of the cutoff by spiking drug(s) into drug free-pooled urine specimens. The concentrations of the samples were confirmed by LC-MS/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert, 1 blind labeled sample, and a device. The results are summarized below.

CutoffResultsConcentration
Drug(ng/mL)-100%-75%-50%-25%+25%+50%+75%
cutoffcutoffcutoffcutoffcutoffcutoffcutoff
AMP500Negative20202019100
Positive0001192020
Total20202020202020
Agreement (%)100%100%100%95.00%95.00%100%100%
Negative20202020200
Positive0000182020
BAR300Total20202020202020
Agreement (%)100%100%100%100%90.00%100%100%
Negative20202020100
BZOPositive0000192020
300Total20202020202020
Agreement (%)100%100%100%100%95.00%100%100%
10Negative20202019100
Positive0001192020
BUPTotal20202020202020
Agreement (%)100%100%100%95.00%95.00%100%100%
150Negative20202020100
Positive0000192020
COCTotal20202020202020
Agreement (%)100%100%100%100%95.00%100%100%
300Negative20202019100
EDDPPositive0001192020
Total20202020202020
Agreement (%)100%100%100%95.00%95.00%100%100%
Negative20202020100
MDMA500Positive0000192020
Total20202020202020
Agreement (%)100%100%100%100%95.00%100%100%
Negative20202019100
MET500Positive0001192020
Total20202020202020
Agreement (%)100%100%100%95.00%95.00%100%100%
MOP300Negative20202019000
Positive0001202020
Total20202020202020
Agreement (%)100%100%100%95.00%100%100%100%
MTD300Negative20202020100
Positive0000192020
Total20202020202020
Agreement (%)100%100%100%95.00%100%100%100%
OXY100Negative20202019100
Positive0001192020
Total20202020202020
Agreement (%)100%100%100%100%95.00%100%100%
PCP25Negative20202019000
Positive0001202020
Total20202020202020
Agreement (%)100%100%100%95.00%95.00%100%100%
TCA1000Negative20202019100
Positive0001192020
Total20202020202020
Agreement (%)100%100%100%95.00%95.00%100%100%
THC50Negative20202020100
Positive0000192020
Total20202020202020
Agreement (%)100%100%100%100%95.00%100%100%

Results of Pocguide™ Multi-Drug Test Cup: Sample Group 1 for Configuration 1 Cup:

{26}------------------------------------------------

Sample Group 2 for Configuration 2 Cup:

DrugCutoff(ng/mL)ResultsConcentration
-100%-75%-50%-25%+25%+50%+75%
cutoffcutoffcutoffcutoffcutoffcutoffcutoff
AMP1000Negative20202019000
Positive0001202020
Total20202020202020
Agreement (%)100%100%100%95.00%100%100%100%
BAR300Negative20202020200
Positive0000182020
Total20202020202020
Agreement (%)100%100%100%100%90.00%100%100%
BZO300Negative20202019100
Positive0001192020
Total20202020202020
Agreement (%)100%100%100%95.00%95.00%100%100%
BUP10Negative20202020100
Positive0000192020
Total20202020202020
Agreement (%)100%100%100%100%95.00%100%100%
COC300Negative20202020200
Positive0000182020
Total20202020202020
Agreement (%)100%100%100%100%90.00%100%100%
EDDP300Negative20202019100
Positive0001192020
Total20202020202020
Agreement (%)100%100%100%95.00%95.00%100%100%
MDMA500Negative20202020100
Positive0000192020
Total20202020202020
Agreement (%)100%100%100%100%95.00%100%100%
Negative20202020100
MET1000Positive0000192020
Total20202020202020
Agreement (%)100%100%100%100%95.00%100%100%
OPI2000Negative20202020200
Positive0000182020
Total20202020202020
Agreement (%)100%100%100%100%90.00%100%100%
300Negative20202020100
MTDPositive0000192020
Total20202020202020
Agreement (%)100%100%100%100%95.00%100%100%
OXY100Negative20202019100
Positive0001192020
Total20202020202020
Agreement (%)100%100%100%95.00%95.00%100%100%
PCP25Negative20202019000
Positive0001202020
Total20202020202020
Agreement (%)100%100%100%95.00%100%100%100%
TCA1000Negative20202019100
Positive0001192020
Total20202020202020
Agreement (%)100%100%100%95.00%95.00%100%100%
THC50Negative20202019100
Positive0001192020
Total20202020202020
Agreement (%)100%100%100%95.00%95.00%100%100%

{27}------------------------------------------------

{28}------------------------------------------------

Participants were given surveys on the ease of understanding the instruction for use. All participants indicated that the device instruction is easy to understand and follow. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.

Clinical Studies: Not applicable.

13. Conclusion

Based on the test principle and performance characteristics of the device including precision, cutoff, interference, specificity, method comparison and lay-user studies of the devices, it's concluded that Pocguide™ Multi-Drug Test Cup OTC and Pocguide™ Multi-Drug Test Cup are substantially equivalent to the predicate device.

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).