K Number
K242029

Validate with FDA (Live)

Device Name
TRAP EASY
Date Cleared
2024-10-09

(90 days)

Product Code
Regulation Number
876.1075
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is used for drawing osteomedullary substance and/or for bone marrow explantation.

Device Description

Trap Easy is a bone marrow biopsy needle, consisting of a cannula with cylindrical tip and ergonomic handle, a removable pyramidal stylet, a sample trapping device integrated into the cannula and an extraction wire.
The device is used for drawing osteomedullary substance and/or for bone marrow explantation. This device is intended to be used only by experienced and properly trained physicians and is intended to be used exclusively for bone marrow tissue biopsies.
The device is available in different models:

  • 8G x 100 mm
  • 8G x 150 mm
  • 11G x 100 mm
  • 11G x 150 mm
    Trap Easy is supplied in sterile packaging and is designed for single use, non-reusable. At the end of its use, the device should be disposed of as infectious hospital waste.
AI/ML Overview

This looks like a 510(k) summary, which typically does not include raw data or detailed study protocols for performance testing, but rather summarizes the results and demonstrates substantial equivalence to a predicate device. The information provided is sparse regarding the detailed breakdown of the acceptance criteria and study particulars for Trap Easy.

However, based on the provided text, here’s an attempt to extract and infer the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The document mentions acceptance criteria but does not explicitly list them with thresholds. It states that "The average lengths and diameters were compliant with acceptance criteria." and "Performed tests show that all the 50 samples have a good integrity and there were no samples ruined or divided in small parts."

Acceptance Criteria CategoryReported Device Performance
Sample IntegrityAll 50 samples had good integrity; no samples ruined or divided into small parts.
Sample Length of Single BiopsyCompliant with acceptance criteria (specific values not provided in document).
Sample DiameterCompliant with acceptance criteria (specific values not provided in document).
Sample RemovalEasy for all samples.
Mechanical FunctioningNeedles showed no damages to the internal mechanism after all biopsies.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: 50 samples were tested for integrity, length, and diameter.
  • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). The studies appear to be non-clinical performance tests conducted by the manufacturer, HS Hospital Service SPA (located in Italy).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This appears to be a study of the physical characteristics and performance of a device, not an interpretative AI algorithm. Therefore, the concept of "experts to establish ground truth" as it pertains to clinical interpretation is not applicable here. The "ground truth" for the physical performance metrics (integrity, length, diameter, removal ease, mechanical function) would be established by direct measurement and observation during the testing process. The document does not specify the number or qualifications of personnel who performed these tests or evaluated the results.

4. Adjudication Method for the Test Set

Not applicable as this is a device performance test, not a clinical interpretation task requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No. This document describes the performance testing of a physical biopsy device, not a software or AI algorithm requiring an MRMC study.

6. If a Standalone Study (i.e., algorithm only without human-in-the-loop performance) was Done

Not applicable. This is a physical device, not an algorithm.

7. The Type of Ground Truth Used

  • For Sample Integrity, Length, Diameter, and Removal: The ground truth was established through direct observation and quantitative measurements during the performance tests.
  • For Mechanical Functioning: The ground truth was established by direct observation and inspection for damages after the biopsies.

8. The Sample Size for the Training Set

Not applicable. This is a physical device, not an AI model requiring a training set.

9. How the Ground Truth for the Training Set was Established

Not applicable. This is a physical device, not an AI model.


Summary of the Study that Proves the Device Meets Acceptance Criteria:

The provided text summarizes performance tests conducted on the "Trap Easy" device to demonstrate its performance characteristics. The tests evaluated:

  • Sample integrity after biopsy: All 50 tested samples reportedly had "good integrity" with no samples being "ruined or divided in small parts."
  • Sample length of single biopsy: The average lengths were reported as "compliant with acceptance criteria," although the specific criteria or measured values are not detailed.
  • Sample diameter: The average diameters were also reported as "compliant with acceptance criteria," without specific values provided.
  • Sample removal: It was noted that "sample removal was easy for all samples."
  • Mechanical functioning of the needle: The tests concluded that "the needles didn't show any damages to the internal mechanism after all biopsies."

The studies were conducted in a manner "as similar as possible to how the biopsy device is used in a medical procedure." The overall conclusion from these tests is that "Obtained results demonstrated that the device's performances are maintained for the intended use of the device."

It appears these were non-clinical, bench-top performance tests, rather than clinical trials with human subjects. The details regarding the exact methodology, measurement tools, and statistical analysis are not available in this summary document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The logo is simple and professional, and it is easily recognizable.

October 9, 2024

HS Hospital Service SPA % Guido Bonapace Consultant Isemed Srl Via P. Togliatti, 19/X Imola (BO), 40026 Italy

Re: K242029

Trade/Device Name: Trap Easy Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: Class II Product Code: KNW Dated: July 3, 2024 Received: July 11, 2024

Dear Guido Bonapace:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Colin K. Chen -S Digitally signed by Colin K. Chen -S Date: 2024.10.09 14:10:58 -04'00'

Jessica Carr Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

for

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K242029

Device Name

TRAP EASY

Indications for Use (Describe)

The device is used for drawing osteomedullary substance and/or for bone marrow explantation.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

General Information

Applicant:

HS HOSPITAL SERVICE SPA Via Zosimo 13, Rome IT 00178 - Italy Phone: +39 06 9201961 Establishment Registration Number: 8010312 Contact: Mr. Massimiliano Improta

Correspondent/Consultant:

Guido Bonapace Isemed Srl Via P. Togliatti. 19/X, Imola, BO 40026 ITALY Mob. +39 335 537 8686 Tel. +39 0542 683803 Fax +39 0542 698456 Email: regulatory@isemed.eu

Date prepared: 9th October, 2024

Device Name

Device Trade Name: TRAP EASY Common Name: Gastroenterology-urology biopsy instrument Classification name: Instrument, Biopsy Regulation Number: 876.1075 Product Code: KNW

Legally Marketed Predicate Devices

Predicate #Predicate Trade NameProduct code
K033616TRAPMATIC SYSTEM SETKNW

Device Description Summary

Trap Easy is a bone marrow biopsy needle, consisting of a cannula with cylindrical tip and ergonomic handle, a removable pyramidal stylet, a sample trapping device integrated into the cannula and an extraction wire.

The device is used for drawing osteomedullary substance and/or for bone marrow explantation. This device is intended to be used only by experienced and properly trained physicians and is intended to be used exclusively for bone marrow tissue biopsies.

The device is available in different models:

  • 8G x 100 mm
  • 8G x 150 mm
  • 11G x 100 mm
  • 11G x 150 mm

{5}------------------------------------------------

HS HOSPITAL SERVICE SPA K242029

Trap Easy is supplied in sterile packaging and is designed for single use, non-reusable. At the end of its use, the device should be disposed of as infectious hospital waste.

Intended use / Indications for Use

The device is used for drawing osteomedullary substance and/or for bone marrow explantation.

Indications for Use Comparison

Indications for use of subject device are the same of primary predicate devices are intended to be used in the procedure for harvesting bone marrow specimens.

Both subject device and predicate devices are intended to be used by a physician.

Technological Comparison

The Trap Easy medical device has the same technological characteristics, design, material and principle of operation as predicate device Trapmatic System set (K033616). As a matter of facts, both the Trap Easy medical device and the Trapmatic System set consist of an outer cannula with handle and trapping element inside, and an extraction wire. As accessories, both devices possess a safety and luer lock cap.

Trap Easy medical device is characterized by the same principle of operation of primary predicate: both devices trap the sample inside and both devices are characterized by a safety and an activation mechanism of the trap. For the Trap Easy medical device, the safety is represented by the white button, while in the Trapmatic System, this safety feature is the white drawer. Then, to activate the mechanism, subject device is equipped with a lever on the side, while in the Trapmatic device, you need to push down the luer connector. Therefore, the only difference with the primary predicate consists in the safety feature that is present for the activation of the mechanism of action: in the Trap Easy this safety feature is represented by a white button, while in the Trapmatic System, it consists in a white drawer. The devices can therefore be considered substantially equivalent.

Subject device is available in sizes 8G and 11G, which fall within those available for the primary predicate device. In addition, the devices are made of the same materials: both consists of AISI 304 stainless steel cannula and ABS handle.

Both devices are manual instrument, provided for single-use and sterilized with Eto, therefore being substantially equivalent.

The devices can be considered substantially equivalent as the differences do not raise concerns about safety or effectiveness.

Non- Clinical and/or Clinical Tests Summary & Conclusions

Performance of subject device have been evaluated by testing sample integrity after biopsy, sample length of single biopsy, sample diameter, sample removal and by testing mechanical functioning of the needle.

The tests have been conducted in a manner as similar as possible to how the biopsy device is used in a medical procedure.

Performed tests show that all the 50 samples have a good integrity and there were no samples ruined or divided in small parts. Moreover, the sample removal was easy for all samples and the needles didn't show any damages to the internal mechanism after all biopsies. The average lengths and diameters were compliant with acceptance criteria.

{6}------------------------------------------------

HS HOSPITAL SERVICE SPA K242029

Obtained results demonstrated that the device's performances are maintained for the intended use of the device.

In addition, in order to guarantee performance and compliance, devices are subjected to inspections after the manufacturing process.

Nonclinical testing data demonstrate the technological characteristics of subject device Trap Easy and that the medical device perform as intended for its use.

The medical device Trap Easy is sterilized using Ethylene Oxide (EtO), and the sterilization process is validated in accordance with the requirements of ISO 11135:2014/A1:2018. The manufacturer, HS Hospital Service Spa, produces several sterile medical devices, all of which fall under the same process validation. These devices are categorized into four different families based on product type. The sterilization process has been validated and is periodically requalified to ensure the continued ability to effectively sterilize the products.

Additionally, the limits for Ethylene Oxide (EtO) residues and its by-products, Ethylene Glycol (EG) and Ethylene Chlorohydrin (ECH), are defined in accordance with ISO 10993-7, taking into account the limited exposure of the medical device.

Moreover, the Trap Easy medical device is available in two different packaging configurations that serve as the sterile barrier system for the product:

  • A pouch made of Tyvek® 2FS and an OPA/PE film (16x44 cm) ●
  • A blister made of Coated Tyvek® CR27 2FS and PET+PE600 (12x22.5 cm) .

Both packaging configurations have been validated in compliance with the requirements of ISO 11607-1:2019.

Regarding the biocompatibility of the medical device, based on its characteristics, Trap Easy is classified according to Annex A of ISO 10993-1 as an externally communicating medical device with limited (< 24 hours) contact with bone/tissue. The material in direct contact is AISI 304 stainless steel, which is used for both the needle (cannula) and the mandrel (stylet). AISI 304 stainless steel is a well-established material, historically used in the manufacturing of this type of medical device due to its proven safety and performance.

Additionally, Trap Easy in its final finished form is identical to the Trapmatic System set in terms of formulation, processing, and sterilization, and no other chemical have been added.

§ 876.1075 Gastroenterology-urology biopsy instrument.

(a)
Identification. A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification. (1) Class II (performance standards).(2) Class I for the biopsy forceps cover and the non-electric biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.