(99 days)
The power wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
The D25 power wheelchair is a four wheeled personal mobile device with a complementary chair support system that is powered by two motors. The traveling speed is controlled by the motor, and the traveling direction is controlled by the passenger. This product is a device suitable for disabled people with mobility difficulties and elderly people and it is intended to provide mobility to a disabled or elderly person limited to a seated position. The power wheelchair can be travelled on flat and obstacle ground surface, and direction and speed of the wheelchair can be controlled by the passenger's hand with the help of the joystick. The device can be used to provide indoor and outdoor mobility at a certain distance but not allowed to be travelled on the road or highway. The wheelchair (Model D25) has a base with aluminum alloy frame, two front wheels, two rear wheels, a seat, an adjustable steering column, a tiller console, two electric motors, an electromagnetic brake, 1 rechargeable Lithium-lon Battery with an off-board charger. The movement of the wheelchair is controlled by the rider who operates the throttle lever, speed control dial and handle on the tiller console. The device is installed with an electromagnetic brake that will engage automatically when the wheelchair is not in use and the brake cannot be used manually.
The provided document is a 510(k) premarket notification for a Power Wheelchair (Model D25). This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving a device meets a specific set of acceptance criteria through a clinical study or performance goals with associated metrics and thresholds.
Therefore, the document does not contain the information requested in your prompt regarding acceptance criteria, study design for proving device performance against these criteria, sample sizes, expert involvement, ground truth establishment, or comparative effectiveness studies (MRMC).
The document primarily focuses on:
- Comparison to a predicate device (K113463): It lists similarities and differences in technical characteristics.
- Compliance with consensus standards: It states that the device complies with numerous ISO and IEC standards related to wheelchairs, general medical device safety, and biocompatibility. These standards contain test methods and passing criteria, which implicitly serve as "acceptance criteria" for demonstrating safety and performance in the context of a 510(k) submission for this type of device.
However, the document does not explicitly present:
- A table of "acceptance criteria" with specific quantitative thresholds (like accuracy, sensitivity, specificity, or AUC) that the device itself must meet, alongside its "reported device performance" against those thresholds.
- Details of a standalone algorithm performance study or a multi-reader, multi-case (MRMC) comparative effectiveness study, as these are typically relevant for AI/ML-enabled software devices that perform diagnostic or interpretative functions. This power wheelchair is a physical medical device for mobility.
To answer your request based on the provided text, I must state that the information is not present in the format you've asked for because the submission type (510(k) for a power wheelchair) does not typically require that level of performance study detail as would an AI/ML diagnostic device.
The "study" that proves the device meets the "acceptance criteria" (which are implicitly the requirements of the cited standards) is the non-clinical performance testing summarized in Section VIII.
Here's a breakdown of what is available vs. what is not:
1. A table of acceptance criteria and the reported device performance:
- Not explicitly provided in the requested format with quantitative metrics like accuracy, sensitivity, etc.
- Implicit Acceptance Criteria: Compliance with the listed ISO and IEC standards (e.g., ISO 7176 series for wheelchairs, ISO 10993 series for biocompatibility, IEC 60601-1-2 for EMC). The "reported device performance" is that the device "complied with the requirements of" these standards. For example, for "Minimum braking distance from maximum speed" the subjective device reports "Forward: 0.9 m" while the predicate reports "Forward: 1.5 m". The discussion notes that the subject device has a shorter braking distance, which is considered a positive difference, ensuring safety.
- The "Discussion/Conclusion" column in the comparison table indirectly serves as reported performance against implicitly accepted characteristics.
2. Sample sized used for the test set and the data provenance:
- For physical device testing against standards, the "sample size" is typically a limited number of test units of the device itself (e.g., one or more new units manufactured according to specifications). This is not a "data set" in the way AI/ML models use it.
- Data provenance: Testing was performed by Anhui JBH Medical Apparatus Co., Ltd. in China, as per the manufacturer's location. The document does not specify if the testing was retrospective or prospective, but performance testing of a physical product is inherently prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable/Not provided. This type of information (expert review for ground truth) is for diagnostic or interpretative devices, not a power wheelchair. The "ground truth" for a power wheelchair's performance is determined by physical measurements, engineering tests, and adherence to established safety and performance standards by qualified testing personnel and engineers, not by medical experts establishing diagnostic ground truth.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable/Not provided. This is relevant for image interpretation or diagnostic studies where human agreement on a diagnosis or finding is needed. For a power wheelchair, adherence to a standard is binary (pass/fail) for each test.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, not done. This is explicitly for AI/ML-driven diagnostic aids. This document is for a physical mobility device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This refers to AI algorithm performance. While the device has an "Electronic controller" (software), its performance is integrated into the physical device's function and verified through the ISO 7176-14 standard and general software validation, not a standalone algorithmic performance study as understood in AI/ML.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Engineering specifications and compliance with international consensus standards. For example, braking distance is measured against a standard, not determined by expert consensus on a diagnosis. Stability is tested against physical requirements. Biocompatibility is tested against established ISO standards for material safety.
8. The sample size for the training set:
- Not applicable. This term is exclusively for machine learning models. The power wheelchair is a physical product, not a software algorithm that undergoes machine learning training.
9. How the ground truth for the training set was established:
- Not applicable. (See answer to #8).
In summary, the document describes a traditional 510(k) submission for a physical medical device. The "proof" of meeting "acceptance criteria" comes from demonstrating compliance with established international performance and safety standards through bench testing, rather than an AI/ML-style clinical study with human readers and ground truth data sets.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
October 4, 2024
Anhui JBH Medical Apparatus Co., Ltd. % Ivv Wang Technical Manager Shanghai SUNGO Management Consulting Co., Ltd. Room 1 401, Dongfang Building, 1500# Century Ave. Shanghai. 200122 China
Re: K241854
Trade/Device Name: Power Wheelchair Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: September 6, 2024 Received: September 6, 2024
Dear Ivy Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tushar Bansal -S
for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Rehabilitation Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
Device Name
Power Wheelchair
Indications for Use (Describe)
The power wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510(k) Summary
l. SUBMITTER
Name: Anhui JBH Medical Apparatus Co., Ltd.
Address: No. 116 Qicang Road, Industrial District, Ming Guang City, Chuzhou, Anhui, China
Name of contact person: Wei Yuanwei
Telephone: +86-0550-8108866
Fax: +86-0550-8106688
Email: +86-0550-8106688
Date prepared: 2024-6-26
II. Device
Device trade name: Power wheelchair Model: D25 Classification name: Powered wheelchair Regulation class: 2 Regulation number: 21CFR 890.3860 Panel: Physical Medicine Product code: ITI
III. Predicate device
K113463 Power Wheelchair, PL001
SUZHOU KD Medical Appliance Co. Ltd.
IV. Device description
The D25 power wheelchair is a four wheeled personal mobile device with a complementary chair support system that is powered by two motors. The traveling speed is controlled by the motor, and the traveling direction is controlled by the passenger. This product is a device suitable for disabled people with mobility difficulties and elderly people and it is intended to provide mobility to a disabled or elderly person limited to a seated position. The power wheelchair can be travelled on flat and obstacle ground surface, and direction and speed of
{5}------------------------------------------------
the wheelchair can be controlled by the passenger's hand with the help of the joystick. The device can be used to provide indoor and outdoor mobility at a certain distance but not allowed to be travelled on the road or highway.
The wheelchair (Model D25) has a base with aluminum alloy frame, two front wheels, two rear wheels, a seat, an adjustable steering column, a tiller console, two electric motors, an electromagnetic brake, 1 rechargeable Lithium-lon Battery with an off-board charger. The movement of the wheelchair is controlled by the rider who operates the throttle lever, speed control dial and handle on the tiller console. The device is installed with an electromagnetic brake that will engage automatically when the wheelchair is not in use and the brake cannot be used manually.
V. Indication for use
The power wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
| Attribute | Subject device | Predicate device | Discussion/Conclusion |
|---|---|---|---|
| Manufacturer | Anhui JBH MedicalApparatus Co., Ltd. | SUZHOU KD MedicalAppliance Co. Ltd. | / |
| Proprietary name, model | Power Wheelchair, D25 | power wheelchair, PL001 | / |
| 510(k) number | K241854 | K113463 | / |
| Device classification name | Class II | Class II | Same |
| Classification regulations | 21 CFR 890.3860 | 21 CFR 890.3860 | Same |
| Product code | ITI | ITI | Same |
| Similarities | |||
| Indication for use | The Power Wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. | They are motor driven indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. | Same |
| Intended | disabled or elderly person | disabled or elderly person | Same |
| Attribute | Subject device | Predicate device | Discussion/Conclusion |
| user | limited to a seated position | limited to a seated position | |
| Use condition | indoor and outdoor use | indoor and outdoor use | Same |
| Number of wheels | 4, including two front wheels and two rear wheels | 4, including two pivoting casters and two rear drive wheels | Same |
| Function of wheels | Front wheels: driven wheels suitable for rotation, acceleration, retrogradeRear wheels: driving wheels to control the speed and direction | two pivoting casters: driven wheels suitable for rotation, acceleration, retrogradetwo rear drive wheels: driving wheels to control the speed and direction | Same |
| Movement control method | By Joystick control | By Joystick control | Same |
| Driving system | Direct drive on the rear wheels | Direct drive on the rear wheels | Same |
| Brake system | Intelligent electromagnetic brake system | Intelligent regenerative Electromagnetic brake | Same |
| Max speed forward | Up to 6 km/h (3.75 mph), continuously adjustable | Up to 6 km/h (3.75 mph), variable | Same |
| Maximum distance of travel on the fully charged battery | 20 km | 20 km | Same |
| Main frame material | aluminum alloy, front and rear folding | aluminum alloy, front and rear folding | Same |
| Armrest cushion | Polyurethane (PU) | PU | Same |
| Differences | |||
| Overall dimensions (LxWxH) | 1030 × 610 × 1010 mm | 880 mm×570 mm×890 mm | larger size is designed for bearing more loading weight. All safety and performance have been validated with the maximum rated weight dummy. |
| Folded dimensions (LxWxH) | 610 × 380 × 840 mm | 720 mm×570 mm×400 mm | safety and performance have been validated with the maximum rated weight dummy. |
| Ground clearance | 120 mm | 80 mm | larger ground clearance provides |
| Attribute | Subject device | Predicate device | Discussion/Conclusion |
| Front wheel size/type | 7.9"x2" /PU Solid tire | 6"x 2"/PU Solid tire | larger size of driven wheel will provide better safety performance. |
| Rear wheel size/type | 12.4"×2.4" /PU Solid tire | 8" x 2.4"/PU Solid tire | better stability and safety performance |
| Battery | li-ion battery pack; rechargeable, 24 VDC 11.6Ah | Li-ion, Rechargeable; 24 VDC 20Ah | Same rated voltage with different power capacity will not affect the safety and performance of the subject device. |
| Max Speed backward | 0.8 m/s (2.88km/h) | 2.4 mph (3.84 km/h) | minor difference on max. backward speed will not cause different performance. lower speed will be more safety. |
| Minimum braking distance from maximum speed | Forward: 0.9 m | Forward:1.5 m (59") | subject device has shorter braking distance than the predicate device, such differences will not affect safety and performance of the device. |
| Minimum braking time from maximum speed | 0.6 second | Not available | The minimum braking time for our device is 0.6 second, which will ensure the device can be stopped on time as required, and such differences will not affect safety and performance of the device. |
| Max loading weight | 120kg (264 lbs) | 114 kg (251 lbs) | minor difference on loading weight will not cause different performance. more loading weight provide more convenient and stable performance for |
| Attribute | Subject device | Predicate device | Discussion/Conclusion |
| Maximumsafeoperationalinclinedegree | 8 ° | 9 ° | minor difference onsafe operationalincline degree will notcause new safety andeffectivenessconcerns. Both thestatic and dynamicstability under specificinclining degree havebeen evaluatedaccording to standardISO 7176 series. |
| Batterycharger | Off-board chargerInput: 100-240V, 50/60Hz,1.5AOutput: 24 Vdc, 2A | Off-board, AutomaticTypeInput: 110-220 V / 50-60Hz,Output: 24 Vdc, 2A; | More wide range ofinput voltage in thedevice which will notcause new safety andeffectivenessconcerns raised. |
| Motor | Brush motor; 24VDC;200W; 2pcs | Brushless DC motor; 24VDC; 180 W; 2 pcs | minor difference onmotor power will notcause differentperformance. largerpower will providemore driving force, nosafety andeffectivenessconcerns raised. |
| Electroniccontroller | brush power wheelchaircontroller | Brushless dual-driverocker controller | Both of the controlsystems are evaluatedaccording to standardISO 7176-14 andsoftware validationrequirement and thereare no new safety andeffectivenessconcerns due to thedifference. |
| TurningRadius | 650 mm (25.5") | 31.5" (800 mm) | The minor differencein the turning radiuswill not raise any newsafety andeffectiveness |
| Attribute | Subject device | Predicate device | Discussion/Conclusion |
| Maximum obstacle climbing | 20 mm | 1.2" (30 mm) | Minor difference in the obstacle climbing will not impact the safety and effectiveness of the subject device. |
| seat cushion | Sponge foam covered Denim fabric | PU foam covered by nylon fabric cloth | different material used for parts in contact |
| back cushion | Sponge foam covered Denim fabric | PU foam covered by nylon fabric cloth | with user, which such differences will not impact the safety and effectiveness of the subject device as biocompatibility tests are carried out according to ISO 10993 series. |
VI. Comparison of technological characteristics with the predicate device
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
VII. Summary of substantial equivalence discussion
The power wheelchair complied with the requirements of ISO 7176-1:2014, ISO 7176-2:2017, ISO 7176-3:2012, ISO 7176-4:2008, ISO 7176-5:2008, ISO 7176-6:2018, ISO 7176-7:1998, ISO 7176-8:2014, ISO 7176-9:2009, ISO 7176-10:2008, ISO 7176-11:2012, ISO 7176-13:1989, ISO 7176-14:2022, ISO 7176-15:1996, ISO 16840-10:2021, ISO 7176-21:2009, ISO 7176-22:2014, ISO10993-5:2009, ISO 10993-10:2021, ISO 10993-23: 2021, IEC60601-1-2: 2020.
The indications for use for both devices are the same. Mainframes of two devices are and the frame materials are the same aluminum alloy. The design principles of the controller and Driving system are the same, and both meet the requirements of the ISO 7176-14. Software validation is carried out on both control systems. Brake system and speed control are designed in the same way as well, and both meet the requirements of the ISO 7176-3. Turning radius, Maximum obstacle climbing and Maximum safe operational incline are slightly different while such differences will not impact the safety and effectiveness of the subject device or raise new safety and effectiveness concerns as well as both meet the requirements of the ISO 7176-2 and ISO 7176-10. The biocompatibility of the subject device is evaluated according to
{10}------------------------------------------------
standard ISO 10993-1 and meet the requirements accordingly.
The flame-retardant test of the seat cushion/ backrest of both subject device and predicate device is carried out according to the ISO 16840-10 test. Therefore, both devices are assured to be under the same safety level.
In conclusion, the technological characteristics, features, specifications, materials, mode of operation, and intended use of the device substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.
VIII. Summary of non-clinical testing
A Performance testing-bench
The following performance data were provided to verify that the subject device met all design specifications and provided support of the substantial equivalence determination.
- Risk Analysis developed in accordance with ISO 14971: 2019.
- Software validation
- . ISO 7176-1:2014 Wheelchairs - Part 1: Determination of static stability
- ISO 7176-2:2017 Wheelchairs - Part 2: Determination of dynamic stability of electric wheelchairs
- . ISO 7176-3:2012 Wheelchairs - Part 3: Determination of effectiveness of brakes
- . ISO 7176-4:2008 Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range
- ISO 7176-5:2008 Wheelchairs - Part 5: Determination of dimensions, mass and maneuvering space
- . ISO 7176-6:2018 Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of electric wheelchairs
- . ISO 7176-7:1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions
- ISO 7176-8:2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strength
- . ISO 7176-9:2009 Wheelchairs - Part 9: Climatic tests for electric wheelchairs
{11}------------------------------------------------
- . ISO 7176-10:2008 Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs
- . ISO 7176-11:2012 Wheelchairs -- Part 11: Test dummies
- ISO 7176-13:1989 Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces.
- ISO 7176-14:2022 Wheelchairs -- Part 14: Power and control systems for electrically powered wheelchairs and scooters -- Requirements and test methods
- . ISO 7176-15:1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling.
- . ISO 16840-10:2021 Wheelchair seating - Part 10: Resistance to ignition of postural support devices - Requirements and test method.
- . ISO 7176-21:2009 Wheelchairs - Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters
- . Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2:2020.
A Biocompatibility of patient-contacting material
Biocompatibility Tests are carried out in accordance with ISO 10993-1: 2018, including cytotoxicity (ISO 10993-5:2009), sensitization (ISO 10993-10:2021) and irritation (ISO 10993-23: 2021). Details of the test summary see biocompatibility summary.
IX. Summary of clinical testing
No animal study and clinical studies are available for our device. Clinical testing was not required to demonstrate the substantial equivalence of the power wheelchair to its predicate device.
X. Conclusions
The conclusion drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed predicate device K113463.
N/A