AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SAFElife™ Fentanyl Urine Home Test (Cassette) is a competitive binding, lateral flow immunochromatographic assay for qualitative detection of Fentanyl in human urine at cutoff concentration of 1 ng/mL.

For in vitro diagnostic use. For Over The Counter (OTC) use.

The test provides only preliminary test results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC/MS-MS) is the recommended confirmatory method.

The SAFElife™ Fentanyl (FTY) Urine Test Cassette is a competitive binding, lateral flow immunochromatoqraphic assay for qualitative detection of Fentanyl (FTY) in human urine at cutoff concentration of 1 ng/mL.

For in vitro diagnostic use.

lt is not intended to distinquish between prescription drug or abuse of the drug. Clinical consideration and professional judgment should be applied to the drug of abuse test result, particularly in evaluating a preliminary positive result.

The test provides only preliminary test results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC/MS-MS) is the recommended confirmatory method.

The SAFElife™ T-Dip Fentanyl Urine Home Test (Dip Card) is a competitive binding, lateral flow immunochromatographic assay for qualitative detection of Fentanyl in human urine at cutoff concentration of 1 ng/mL.

For in vitro diagnostic use. For Over The Counter (OTC) use.

The test provides only preliminary test results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC/MS-MS) is the recommended confirmatory method.

The SAFElife™ T-Dip Fentanyl (FTY) Urine Test Panel is a competitive binding, lateral flow immunochromatographic assay for qualitative detection of Fentanyl (FTY) in human urine at cutoff concentration of 1 ng/mL.

For in vitro diagnostic use.

It is not intended to distinguish between prescription drug or abuse of the drug. Clinical consideration and professional judgment should be applied to the drug of abuse test result, particularly in evaluating a preliminary positive result. The test provides only preliminary test results. To obtain a confirmed analytical result, a more specific alternate chemical must be used. Chromatography/Mass Spectrometry (GC/ MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC/MS-MS) is the recommended confirmatory method.

Device Description

The Wondfo SAFElife™ Fentanyl Tests are immunoassays intended for the qualitative detection of fentanyl in human urine. Each Wondfo SAFElife™ Fentanyl Test consists of a Test Device in format of Cassette or Dip Card, and a package insert. Each Test Device is sealed with sachets of desiccant in an aluminum pouch.

AI/ML Overview

Acceptance Criteria & Study Analysis for SAFElife™ Fentanyl Urine Home Test

The SAFElife™ Fentanyl Urine Home Test (and its variants) are competitive binding, lateral flow immunochromatographic assays for qualitative detection of Fentanyl in human urine at a cutoff concentration of 1 ng/mL. The device provides preliminary test results, with confirmation requiring a more specific alternate chemical method like GC/MS or LC/MS-MS.

1. Acceptance Criteria and Reported Device Performance

The provided documentation does not explicitly state formal acceptance criteria in the typical "pass/fail" format for each performance characteristic. However, the study results implicitly demonstrate the device's acceptable performance based on standard expectations for qualitative drug screening tests.

Given the information, we can infer the acceptance criteria and compare them to the reported performance:

Acceptance Criteria (Inferred)Reported Device Performance
Precision:Cassette Precision:
Consistent results for samples significantly below and above the cutoff concentration.-100% to -50% cutoff: All 50 tests across 3 lots were negative (50-/0+).
Acceptable performance around the cutoff, demonstrating the ability to differentiate positive and negative samples.-25% cutoff: 46-48 negative, 2-4 positive across 3 lots.
Cutoff (1 ng/mL): 23-28 positive, 22-27 negative across 3 lots.
+25% to +100% cutoff: All 50 tests across 3 lots were positive (50+/0-).
Dip Card Precision:
-100% to -50% cutoff: All 50 tests across 3 lots were negative (50-/0+).
-25% cutoff: 47-48 negative, 2-3 positive across 3 lots.
Cutoff (1 ng/mL): 23-28 positive, 22-27 negative across 3 lots.
+25% to +100% cutoff: All 50 tests across 3 lots were positive (50+/0-).
Stability:The devices are stable at 36-86°F for 24 months based on accelerated stability study.
Interference:Over 100 common substances (medications, endogenous compounds) showed no interference at specified concentrations (typically 100 µg/mL or physiological/pathological levels) when spiked into drug-free urine and fentanyl-positive urine (at -50% and +50% cutoff). This indicates a high level of specificity against a wide range of potential interferents.
No significant interference from common substances at physiological or therapeutic concentrations.
Specificity (Cross-Reactivity):While some fentanyl analogs showed cross-reactivity (e.g., Acrylfentanyl, Isobutyryl fentanyl, Furanyl fentanyl at 100%; Butyryl fentanyl, Carfentanil at 50%), this is expected for drugs belonging to the same class or with similar chemical structures. Many other opioids (e.g., Morphine, Codeine, Buprenorphine, Oxycodone) and diverse compounds showed no cross-reactivity at 100 µg/mL, indicating good specificity against a broad range of other substances.
Acceptable cross-reactivity profile, with minimal false positives from unrelated compounds, and expected reactivity to close analogs.
Effect of Urine Specific Gravity & pH:No differences in test results were observed across specific gravity ranges of 1.000 to 1.035 and pH ranges of 4 to 9. All samples at or above +50% Cut-Off were positive, and all samples at or below -50% Cut-Off were negative. This demonstrates robust performance across varying urine physiological conditions.
Robust performance across a physiologically relevant range of urine specific gravity and pH.
Method Comparison (Clinical Sample Performance):Cassette: For 84 unaltered clinical samples (41 negative, 43 positive per LC/MS), typically 0 false positives for negative samples, 0-2 false negatives for high positive samples (with 1 false negative at 1.058 ng/mL, just over cutoff). A few "Near Cutoff Negative" (LC/MS between -50% and cutoff, i.e., 0.5-1 ng/mL) were reported positive by the rapid test, and some "Near Cutoff Positive" (LC/MS between cutoff and +50%, i.e., 1-1.5 ng/mL) were reported negative. The discordant results (e.g., sample 39 at 0.825 ng/mL, sample 41 at 0.914 ng/mL being positive) indicate reasonable performance near the cutoff.
High concordance with a confirmatory method (LC/MS) for both negative and positive samples, especially for samples well away from the cutoff.Dip Card: Similar performance to the cassette. For 84 samples, typically 0 false positives for negative samples. A few false negatives for samples just above cutoff (e.g., 1.038 ng/mL, 1.015 ng/mL, 1.077 ng/mL). Discordant results (e.g., samples at 0.885, 0.914, 0.804, 0.825 ng/mL reported positive) demonstrate appropriate sensitivity around the cutoff. Overall, the method comparison shows acceptable agreement with LC/MS, particularly for samples clearly above or below the cutoff. The expected variation occurs in samples very close to the 1 ng/mL cutoff.
Lay-user Study:Cassette & Dip Card: 100% correct results for samples well below (-100% to -50% cutoff) and well above (+25% to +75% cutoff). 95% correct results for -25% cutoff (0.7 ng/mL), with 1 out of 20 samples incorrectly read as positive. All lay users found instructions easy to follow, with a Flesch-Kincaid reading level below 7. This demonstrates the device's suitability for OTC (home) use.
High percentage of correct results by lay users, demonstrating ease of use and interpretation in a home setting.

2. Sample Size Used for the Test Set and Data Provenance

  • Precision Studies Test Set:

    • For each of the 9 concentration levels, 50 tests were performed for each of the 3 device lots.
    • Total tests for Precision (Cassette): 9 concentrations * 50 tests/concentration * 3 lots = 1350 tests.
    • Total tests for Precision (Dip Card): 9 concentrations * 50 tests/concentration * 3 lots = 1350 tests.
    • Data Provenance: The samples were "prepared by spiking fentanyl in negative samples." The document does not specify the country of origin but implies laboratory-prepared controlled samples. The study is prospective in nature as samples were prepared for the purpose of the study.
  • Interference Test Set: Not explicitly stated as a single "test set" size. "These urine samples were tested using three batches of each device." The number of samples for each interferent tested at specific concentrations (drug-free and fentanyl-spiked) is not quantified.

    • Data Provenance: Laboratory-prepared controlled samples with spiked substances. Most likely retrospective analysis of prepared samples.
  • Specificity (Cross-Reactivity) Test Set: Similar to interference, not a single "test set" size. "drug metabolites and other components that are likely to interfere in urine samples were tested using three batches of device." The number of individual compounds tested is large (dozens of fentanyl analogs and many other opioids/non-opioids).

    • Data Provenance: Laboratory-prepared controlled samples with spiked substances. Most likely retrospective analysis of prepared samples.
  • Effect of Urine Specific Gravity and pH Test Set: Not explicitly stated as a single "test set" size. "These samples were tested using three lots of device." Samples were prepared by spiking target fentanyl at -50% and +50% Cut-Off levels across specific gravity and pH ranges.

    • Data Provenance: Laboratory-prepared controlled samples. Most likely retrospective analysis of prepared samples.
  • Method Comparison Studies Test Set:

    • Cassette: 84 "unaltered clinical samples."
    • Dip Card: 84 "unaltered clinical samples." (It's unclear if these are the same 84 samples for both formats or separate sets of 84).
    • Data Provenance: "unaltered clinical samples." The country of origin is not specified, but the clinical nature suggests they were collected from human subjects. The study appears to be retrospective as these samples were "blind labeled" and then tested.
  • Lay-user Study Test Set:

    • For each device format (Cassette and Dip Card), 7 different concentration levels were tested, with 20 samples per concentration.
    • Total samples per device format: 7 concentrations * 20 samples/concentration = 140 samples.
    • Total samples overall: 140 samples (Cassette) + 140 samples (Dip Card) = 280 samples.
    • Data Provenance: Samples were "prepared at the following concentrations by spiking fentanyl into drug free-pooled urine specimens." The samples were then blind-labeled and randomized. The study is prospective since samples were prepared for the study.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Precision, Interference, Specificity, Effect of SG/pH, Lay-user Studies: The "ground truth" (actual fentanyl concentration or absence) was established by LC/MS (Liquid Chromatography/Mass Spectrometry) for all spiked samples. This is a highly accurate analytical method, and human experts are not directly establishing the ground truth in terms of visual interpretation for these studies. The LC/MS results are considered the definitive quantitative measurement.
  • Method Comparison Studies: The ground truth for the 84 "unaltered clinical samples" was established by LC/MS. The document states, "The samples were blind labeled and compared to LC/MS results." No human experts (e.g., pathologists, radiologists) were involved in establishing the ground truth for these quantitative drug levels.

4. Adjudication Method for the Test Set

The concept of "adjudication" (e.g., 2+1, 3+1) typically applies to situations where multiple human readers are interpreting results and their agreement/disagreement needs to be resolved. In this case, the device output is a qualitative (positive/negative) result based on a visual line, and the ground truth is established by a quantitative analytical method (LC/MS).

  • For the Precision studies, the results are quantitative counts of "positive" or "negative" for each lot and concentration, implying a direct read of the test.
  • For the Method Comparison Studies, the rapid test results from the operators were directly compared to the LC/MS results. Discordant results are simply listed. No adjudication process is described for resolving discrepancies in the rapid test readings themselves among different operators, but rather the discrepancy is noted between the rapid test result and the LC/MS ground truth.
  • For the Lay-user study, each participant (lay person) used one device and provided their result. The comparison was then made between the lay person's result and the LC/MS confirmed concentration. The study report only mentions "Lay person results" (No. of Positive / No. of Negative) implying their individual readings were collected and analyzed for correctness.

Therefore, an adjudication method in the traditional sense (e.g., expert panel review to resolve conflicting interpretations) was not applied, as the output is a qualitative visual reading compared to an analytical gold standard.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done.

An MRMC study typically compares the performance of multiple human readers on multiple cases, often with and without AI assistance, to determine if the AI improves human reader performance (e.g., sensitivity, specificity, efficiency). The studies presented here focus on the standalone performance of the device itself (precision, analytical performance, and comparison to LC/MS) and a lay-user study for ease of use. There is no mention of human readers being compared with and without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, the primary studies presented are essentially standalone performance studies for the device.

The SAFElife™ Fentanyl Urine Home Test is a qualitative lateral flow immunoassay. Its "algorithm" is the biochemical reaction that produces a visual line. The precision studies, interference studies, specificity studies, effect of SG/pH studies, and method comparison studies all evaluate the device's ability to correctly classify samples as positive or negative based on its inherent properties, without human interpretation being the primary variable of interest for method comparison against LC/MS.

While the "reading" of the line is ultimately done by a human (either a trained operator in the method comparison or a lay-user in that specific study), the performance characteristics described (e.g., sensitivity at cutoff, cross-reactivity) are intrinsic to the device's chemical and physical design, functioning as an "algorithm-only" or "device-only" performance evaluation against a gold standard. The lay-user study specifically tests the human-in-the-loop aspect for OTC use, but the core analytical performance is evaluated as standalone.

7. The Type of Ground Truth Used

The primary ground truth used across all analytical performance studies (Precision, Interference, Specificity, Effect of SG/pH, and Method Comparison) and the Lay-user study was Liquid Chromatography/Mass Spectrometry (LC/MS).

LC/MS is a highly sensitive and specific analytical technique used to identify and quantify substances in complex mixtures, making it a robust "gold standard" for determining precise drug concentrations in urine samples.

8. The Sample Size for the Training Set

The document does not describe a "training set" in the context of an algorithm that learns from data. This device is a biochemical immunoassay, not a software-based AI algorithm that requires a training phase. Its performance is determined by its physical and chemical design, not by learning from a dataset.

Therefore, the concept of a "training set" is not applicable to this device.

9. How the Ground Truth for the Training Set Was Established

As noted above, a "training set" is not applicable because this is a biochemical immunoassay, not a machine learning algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Guangzhou Wondfo Biotech Co., Ltd. % Joe Shia Director LSI International Inc 504E Diamond Ave., Suite H Gaithersburg, Maryland 20877

Re: K241741

Trade/Device Name: SAFElife™ Fentanyl Urine Home Test (Cassette); SAFElife™ Fentany) (FTY) Urine Test Cassette; SAFElife™ T-Dip Fentanyl Urine Home Test (Dip Card); SAFElife™ T-Dip Fentanyl (FTY) Urine Test Panel Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate Test System Regulatory Class: Class II Product Code: NGL Dated: June 13, 2024 Received: June 17, 2024

Dear Joe Shia:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Digitally signed by
Joseph A. Kotarek Joseph A. Kotarek -S Date: 2024.07.16
11:44:38 -04'00' Joseph Kotarek, Ph.D. Branch Chief Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K241741 Device Name

SAFElife™ Fentanyl Urine Home Test (Cassette); SAFElife™ Fentanyl (FTY) Urine Test Cassette; SAFElife™ T-Dip Fentanyl Urine Home Test (Dip Card); SAFElife™ T-Dip Fentanyl (FTY) Urine Test Panel

Indications for Use (Describe)

The SAFElife™ Fentanyl Urine Home Test (Cassette) is a competitive binding, lateral flow immunochromatographic assay for qualitative detection of Fentanyl in human urine at cutoff concentration of 1 ng/mL.

For in vitro diagnostic use. For Over The Counter (OTC) use.

The test provides only preliminary test results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC/MS-MS) is the recommended confirmatory method.

The SAFElife™ Fentanyl (FTY) Urine Test Cassette is a competitive binding, lateral flow immunochromatoqraphic assay for qualitative detection of Fentanyl (FTY) in human urine at cutoff concentration of 1 ng/mL.

For in vitro diagnostic use.

lt is not intended to distinquish between prescription drug or abuse of the drug. Clinical consideration and professional judgment should be applied to the drug of abuse test result, particularly in evaluating a preliminary positive result.

The test provides only preliminary test results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC/MS-MS) is the recommended confirmatory method.

The SAFElife™ T-Dip Fentanyl Urine Home Test (Dip Card) is a competitive binding, lateral flow immunochromatographic assay for qualitative detection of Fentanyl in human urine at cutoff concentration of 1 ng/mL.

For in vitro diagnostic use. For Over The Counter (OTC) use.

The test provides only preliminary test results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC/MS-MS) is the recommended confirmatory method.

The SAFElife™ T-Dip Fentanyl (FTY) Urine Test Panel is a competitive binding, lateral flow immunochromatographic assay for qualitative detection of Fentanyl (FTY) in human urine at cutoff concentration of 1 ng/mL.

For in vitro diagnostic use.

It is not intended to distinguish between prescription drug or abuse of the drug. Clinical consideration and professional judgment should be applied to the drug of abuse test result, particularly in evaluating a preliminary positive result. The test provides only preliminary test results. To obtain a confirmed analytical result, a more specific alternate chemical must be used. Chromatography/Mass Spectrometry (GC/ MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC/MS-MS) is the recommended confirmatory method.

{3}------------------------------------------------

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY K241741

1. Date:July 8, 2024
2. Submitter:Guangzhou Wondfo Biotech Co., Ltd.No.8 Lizhishan Road, Science City, Huangpu DistrictGuangzhou, China
3. Contact person:Joe ShiaLSI International Inc.504E Diamond Ave., Suite HGaithersburg, MD 20877Telephone: 240-505-7880Email: shiajl@yahoo.com
4. Device Names:SAFElife™ Fentanyl Urine Home Test (Cassette)SAFElife™ Fentanyl (FTY) Urine Test CassetteSAFElife™ T-Dip Fentanyl Urine Home Test (Dip Card)SAFElife™ T-Dip Fentanyl (FTY) Urine Test Panel
Product CodeClassificationRegulation SectionPanel
NGLII21 CFR § 862.3650Opiate Test SystemToxicology (91)
    1. Predicate Devices:
      AllTest Fentanyl Urine Test Cassette (K233417)
    1. Indications for Use
      The SAFElife™ Fentanyl Urine Home Test (Cassette) is a competitive binding, lateral flow immunochromatographic assay for qualitative detection of Fentany] in human urine at cutoff concentration of 1 ng/mL.

For in vitro diagnostic use. For Over The Counter (OTC) use.

The test provides only preliminary test results. To obtain a confirmed analytical result, a more specific alternate chemical must be used. Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC/MS-MS) is the recommended confirmatory method.

The SAFElife™ Fentanyl (FTY) Urine Test Cassette is a competitive binding, lateral flow immunochromatographic assay for qualitative detection of Fentanyl (FTY) in human urine at cutoff concentration of 1 ng/mL.

For in vitro diagnostic use.

It is not intended to distinguish between prescription drug or abuse of the drug. Clinical consideration and professional judgment should be applied to the drug of abuse test result, particularly in evaluating a preliminary positive result.

The test provides only preliminary test results. To obtain a confirmed analytical result, a more specific alternate chemical must be used. Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC/MS-MS) is the recommended confirmatory method.

{5}------------------------------------------------

The SAFElife™ T-Dip Fentanyl Urine Home Test (Dip Card) is a competitive binding, lateral flow immunochromatographic assay for qualitative detection of Fentany] in human urine at cutoff concentration of 1 ng/mL.

For in vitro diagnostic use. For Over The Counter (OTC) use.

The test provides only preliminary test results. To obtain a confirmed analytical result, a more specific alternate chemical must be used. Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC/MS-MS) is the recommended confirmatory method.

The SAFElife™ T-Dip Fentanyl (FTY) Urine Test Panel is a competitive binding, lateral flow immunochromatographic assay for qualitative detection of Fentanyl (FTY) in human urine at cutoff concentration of 1 ng/mL.

For in vitro diagnostic use.

It is not intended to distinguish between prescription drug or abuse of the drug. Clinical consideration and professional judgment should be applied to the drug of abuse test result, particularly in evaluating a preliminary positive result. The test provides only preliminary test results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC/MS-MS) is the recommended confirmatory method.

    1. Device Description
      The Wondfo SAFElife™ Fentanyl Tests are immunoassays intended for the qualitative detection of fentanyl in human urine. Each Wondfo SAFElife™ Fentanyl Test consists of a Test Device in format of Cassette or Dip Card, and a package insert. Each Test Device is sealed with sachets of desiccant in an aluminum pouch.
    1. Substantial Equivalence Information
      A summary comparison of features of the Wondfo SAFElife™ Fentanyl Test and the predicate devices is provided in following table.
ItemDevicePredicate – K233417
Indication(s)for UseFor the qualitative determination offentanyl in human urine.Same
Calibrator and Cut-OffValuesFentanyl (FTY)1 ng/mlSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays basedon the principle of antigen antibodyimmunochemistry.Same
Type of TestQualitativeSame
Specimen TypeHuman UrineSame
Intended UseFor OTC useSame
ConfigurationsCassette, Dip CardCassette
Storage4-30°CSame

Table 1: Features Comparison of Wondfo SAFElife™ Fentanyl Test and the Predicate Device

{6}------------------------------------------------

9. Test Principle

The Wondfo SAFElife™ Fentanyl Tests are immunoassays based on the principle of competitive binding.

During testing, a urine specimen migrates upward by capillary action. Fentanyl, if present in the urine specimen below 1 ng/mL, will not saturate the binding sites of antibody-coated particles in the test device. The antibody-coated particles will then be captured by immobilized fentanyl conjugate and a visible colored line will show up in the test line region. The colored line will not form in the test line region if the fentanyl level exceeds 1 ng/mL because it will saturate all the binding sites of anti-fentanyl antibodies.

To serve as a procedural control, a colored line will always appear at the control line region indicating that proper volume of specimen has been added and membrane wicking has occurred.

10. Performance Characteristics

    1. Analytical Performance
    • a. Precision

Precision studies were carried out for samples with concentrations of -100% cut off, -75% cut off, -50% cut off, -25% cut off, cut off, +25% cut off, +50% cut off, +75% cut off and +100% cut off. These samples were prepared by spiking fentanyl in negative samples. Each fentanyl concentration was confirmed by LC/MS. All sample aliquots were blindly labeled by the person who prepared the samples and didn't take part in the sample testing. For each concentration, tests were performed two tests per day for 25 days per device lot in a randomized order.

Cassette

Lot Number-100% cut off-75% cut off-50% cut off-25% cutoffcut off+25% cut off+50% cut off+75% cut off+100% cut off
Lot 150-/0+50-/0+50-/0+48-/2+28+/22-50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+46-/4+27+/23-50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+48-/2+25+/25-50+/0-50+/0-50+/0-50+/0-

Dip Card

LotNumber-100%cut off-75%cut off-50%cut off-25%cutoffcut off+25%cut off+50%cut off+75%cut off+100%cut off
Lot 150-/0+50-/0+50-/0+47-/3+24+/26-50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+47-/3+28+/22-50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+48-/2+23+/27-50+/0-50+/0-50+/0-50+/0-

c. Stability

The devices are stable at 36-86F for 24 months based on the accelerated stability study.

d. Interference

Potential interfering substances found in human urine of physiological or pathological conditions were added to drug-free urine and target drug fentanyl urine with concentrations at 50% below and 50% above Cut-Off levels. These urine samples were tested using three batches of each device. Compounds that showed no interference at a concentration of 100µg/mL or specified concentrations are summarized in the following tables.

AcetaminophenDoxepinNortriptyline
Acetone (1000 mg/dL)Ecgonine methyl esterNoscapine
AcetophenetidinEphedrineO-Hydroxyhippuric acid
Acetylsalicylic acidErythromycinOctopamine
Albumin (100 mg/dL)Ethanol (1%)Oxalic acid (100 mg/dL)
AlbuterolFenoprofenOxazepam
AminopyrineFluphenazineOxolinic acid
AmitriptylineFurosemideOxymetazoline
AmobarbitalGalactose (10 mg/dL)Papaverine
AmoxicillinGamma globulin (500 mg/dL)Penicillin G
AmpicillinGentisic acidPerphenazine
ApomorphineGlucose (3000 mg/dL)Phencyclidine
Ascorbic AcidHemoglobinPhenelzine
AspartameHydralazinePhenobarbital
AtropineHydrochlorothiazidePrednisone
Benzilic acidHydrocortisonePropoxyphene (50 µg/mL)
Benzoic acidHydroxytyraminePropranolol
BenzoylecgonineIbuprofenPseudoephedrine
BilirubinImipramineQuinine
Boric acid (1% w/v)IsoproterenolRanitidine
BupropionIsoxsuprineRiboflavin (10 mg/dL)
CaffeineKetamineSalicylic acid
CarbamazepineKetoprofenSecobarbital
Chloral hydrateLabetalolSerotonin (5-hydroxytyramine)
ChloramphenicolLidocaineSulfamethazine
ChlorothiazideLoperamideSulindac
ChlorpromazineMaprotilineTetrahydrocortisone 3-(β-Dglucuronide)
CholesterolMeperidine (50 µg/mL)Tetrahydrocortisone 3-acetate
ClomipramineMeprobamateTetrahydrozoline
ClonidineMethapyrileneThiamine
CortisoneMethaqualoneThioridazine
CotinineMethoxyphenamineTriamterene
CreatinineMetronidazole (300 µg/mL)Trifluoperazine
CyclobenzaprineN-AcetylprocainamideTrimethoprim
DeoxycorticosteroneNaCl (4000 mg/dL)Tyramine
DesipramineNalidixic acidUrea (2000 mg/dL)
DextromethorphanNaloxoneUric acid
DiclofenacNaltrexoneValproic acid (250 µg/mL)
DiflunisalNaproxenVenlafaxine
DigoxinNiacinamideVerapamil
DiphenhydramineNicotineZomepirac
DL-TryptophanNifedipineβ-Estradiol
DL-TyrosineNorethindrone

{7}------------------------------------------------

e. Specificity

To test specificity, drug metabolites and other components that are likely to interfere in urine samples were tested using three batches of device. The lowest concentration that caused a positive result for each compound are listed below. There are no differences in the test results for different device format.

{8}------------------------------------------------

Fentanyl (Cutoff=1ng/mL)Minimumconcentration requiredto obtain a positiveresult (ng/mL)% Cross-Reactivity
Acetyl fentanyl166.25%
Acrylfentanyl1100.00%
ω-1-Hydroxyfentanyl20,0000.005%
Isobutyryl fentanyl1100.00%
Ocfentanil2.343.48%
Butyryl fentanyl250.00%
Furanyl fentanyl1100.00%
Valeryl fentanyl2.540.00%
(±) β-hydroxythiofentanyl2.540.00%
4-Fluoro-isobutyrylfentanyl333.33%
Para-fluorobutyryl fentanyl425.00%
Para-fluoro fentanyl2.540.00%
(+)-3-cis-methyl fentanyl502.00%
Carfentanil250.00%
Sufentanil156.67%
Alfentanil75000.01%
Despropionyl fentanyl (4-ANPP)2,0000.05%
Remifentanil>100 µg/mL< 0.001%
Norfentanyl>100 µg/mL< 0.001%
Acetyl norfentanyl>100 µg/mL< 0.001%
Norcarfentanil>100 µg/mL< 0.001%
Trazodone250000.004%

The following opioids compounds were tested at a concentration of 100ug/mL. Negative results were obtained for all these compounds. There is no cross-reactivity for these compounds using the Wondfo SAFElife™ Fentanyl Tests of all three device formats.

6-Acetyl morphineNaloxone
AmphetamineNaltrexone
BuprenorphineNorbuprenorphine
BuprenorphineglucuronideNorcodeine
CodeineNorketamine
DextromethorphanNormeperidine
DihydrocodeineNormorphine
EDDPNoroxycodone
EMDPOxycodone
FluoxetineOxymorphone
HeroinPentazocine (Talwin)
HydrocodonePipamperone
HydromorphoneRisperidone
KetamineTapentadol
LevorphanolThioridazine
MeperidineTilidine
MethadoneTramadol
MorphineTramadol-O-Desmethyl
Morphine-3-glucuronideTramadol-N-Desmethyl

{9}------------------------------------------------

f. Effect of Urine Specific Gravity and Urine pH

To investigate the effect of urine specific gravity and urine pH, urine samples, with 1.000 to 1.035 specific gravity or urine samples with pH 4 to 9 were spiked with target fentanyl at 50% below and 50% above Cut-Off levels. These samples were tested using three lots of device. Results were all positive for samples at and above +50% Cut-Off and all negative for samples at and below -50% Cut-Off. There are no differences in the test results for different device format.

    1. Comparison Studies
      Method comparison studies for the Wondfo SAFElife™ Fentanyl Test were performed using three different lots of the device. Operators ran 84 (41 negative and 43 positive) unaltered clinical samples. The samples were blind labeled and compared to LC/MS results are presented in the tables below.

Cassette

NegativeLowNegative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
Operator1Positive0022220
Negative8131810
Operator2Positive0022320
Negative8131800
Operator3Positive0022320
Negative8131800
Discordant Results
----------------------
OperatorSample IDLC/MS Result (ng/mL)Rapid Test Result
Operator 1, 2390.825Positive
Operator 1, 2, 3410.914Positive
Operator 3400.885Positive
Operator 1431.058Negative

Dip Card

NegativeLow Negative byLC/MS(less than -50%)Near Cutoff Negative byLC/MS(Between -50% andcutoff)Near Cutoff Positive byLC/MS(Between thecutoff and +50%)High Positiveby LC/MS(greater than+50%)
Operator1Positive0022220
Negative8131810
Operator2Positive0032120
Negative8131720

{10}------------------------------------------------

OperatorPositive0012120
3Negative8131920

Discordant Results

OperatorSample IDLC/MS Result (ng/mL)Rapid Test Result
Operator 1, 2, 3400.885Positive
Operator 1410.914Positive
Operator 2370.804Positive
Operator 2390.825Positive
Operator 1441.038Negative
Operator 2, 3421.015Negative
Operator 2, 3451.077Negative

3. Lay-user study

A lay user study was performed at three intended user sites with 140 lay persons for each device format. They had diverse educational and professional backgrounds and ranged in age from 18 to >50 years. Urine samples were prepared at the following concentrations; -100%, +/-75%, +/-50%, +/-25% of the cut-off by spiking fentanyl into drug free-pooled urine specimens. The concentrations of the samples were confirmed by LC/MS. Each sample was aliquoted into individual containers, blind-labeled and randomized. Each participant was provided with the package insert, 1 blind labeled sample and a device. The results are summarized below:

NumberofsamplesFentanyl Concentrationby LC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
% of CutoffNo. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff200.2020100
-50% Cutoff200.5020100
-25% Cutoff200.711995
+25% Cutoff201.2200100
+50% Cutoff201.4200100
+75% Cutoff201.7200100

Cassette

Dip Card

% of CutoffNumberofsamplesFentanyl Concentrationby LC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff200.2020100
-50% Cutoff200.5020100
-25% Cutoff200.711995
+25% Cutoff201.2200100
+50% Cutoff201.4200100
+75% Cutoff201.7200100

Lay-users were also given surveys on the ease of understanding the package insert instructions. All lay users indicated that the device instructions can be easily followed. A Flesch-Kincaid

{11}------------------------------------------------

reading analysis was performed on the package insert and the score revealed a reading grade level of less than 7.

4. Clinical Studies

Not applicable.

11. Conclusion

Based on the test principle and acceptable performance characteristics including precision, cut-off, interference, specificity, method comparison and Lay-user studies of the devices, it's concluded a substantial equivalence decision.

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).