K Number
K241641

Validate with FDA (Live)

Date Cleared
2025-01-11

(218 days)

Product Code
Regulation Number
890.3860
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Electric wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

Device Description

The Electric wheelchair, YE200, is a powered wheelchair that is designed to fold when not in use so it can be more easily transported in a car trunk or similar space. It uses rechargeable lithium batteries, and is controlled by a joystick and a few other simple controls which are located on the end of the armrest. The joystick assembly may be installed on either side of the chair to accommodate both right-handed and left-handed users. The Electric wheelchair consists of two armrests, a back rest, a seat cushion, a foldable frame, two rear drive wheels with hub motor and electromagnetic brake assemblies, two pivoting casters, two Li-ion battery pack, an off-board battery charger, a control panel (joystick) with connect cables and an electric motor controller. The device is powered by Li-ion battery pack, which can be recharged by an off-board battery charger that can be plugged into an AC socket outlet when the device is not in use. The user can activate the controller handle (joystick) to control the speed and direction of the wheelchair movement. In addition, when the patient releases the joystick, the joystick will return back to the central position and the wheelchair will be automatically stopped soon due to automatic electromagnetic brake system. Once the joystick is activated again move to other position, the wheelchair will be re-energized.

AI/ML Overview

The provided text is a 510(k) summary for an electric wheelchair (Model YE200). It details the device, its intended use, a comparison to a predicate device, and a summary of non-clinical testing performed to demonstrate substantial equivalence.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided document:

1. A table of acceptance criteria and the reported device performance

The document does not present a formal "acceptance criteria" table with specific pass/fail values. Instead, it provides a comparison table between the subject device (YE200) and the predicate device (DC01), highlighting similarities and differences. The "Discussion/Conclusion" column in this table serves as the performance assessment against the predicate. The underlying "acceptance criteria" appear to be the successful execution and passing of specific ISO standards relevant to electric wheelchairs.

Here's a re-imagined table focusing on key performance attributes where differences exist, and the stated conclusion regarding compliance with standards. It's important to note that the document states "all passed" or "will not affect safety/performance," implying that the measured performance met the unseen criteria of the referenced standards.

AttributePredicate Device (DC01) PerformanceSubject Device (YE200) PerformanceAcceptance Criteria (Implied by Standards & Conclusion)Reported Device Performance (as per Discussion/Conclusion)
Max speed forward6 km/h5.7 km/hPerformance within range of predicate and/or within ISO 7176-6 standards.Minor difference, will not cause different performance.
Minimum braking distance from maximum speed (Forward)0.5m0.75mPerformance within range of predicate and/or within ISO 7176-3 standards.Minor difference, all relevant tests performed according to ISO 7176-3, no safety and effectiveness concerns.
Maximum safe operational incline degree8 °6 °Performance within range of predicate and/or within ISO 7176 series standards for static and dynamic stability.Minor difference, no new safety and effectiveness concerns as static and dynamic stability evaluated per ISO 7176 series.
Maximum distance of travel on fully charged battery20 km16 kmPerformance within range of predicate and/or adequate for intended use.Minor difference, will not cause different performance. No new safety/effectiveness concerns.
Turning Radius800 mm875 mmPerformance adequate for intended use; ergonomic considerations.Minor difference, may cause slight inconvenience in narrow spaces, but no new safety/effectiveness concerns.
Maximum obstacle climbing20 mm15 mmPerformance adequate for intended use and/or meets ISO 7176-10 standards.Minor difference, will not impact safety and effectiveness.
Biocompatibility of patient-contacting materialMet ISO 10993 seriesMet ISO 10993 seriesCompliance with ISO 10993-5, -10, -23.Biocompatibility tests carried out and met requirements.
Frame MaterialCarbon fiberAluminum alloyMaterial change must not affect safety/performance as demonstrated by testing.Different material, but all safety and performance tests conducted and passed. No effect on safety/performance.
Overall Dimensions1110mm x 700mm x 980mm1107mm x 620mm x 950mmDimensions must allow for safe and effective use.Minor difference, will not cause different performance. Validated with maximum rated weight dummy.
Folded Dimensions810mm x 700mm x 400mm420mm x 620mm x 800mmDimensions must allow for intended transportability.No specific conclusion beyond the data presented.
Ground Clearance160mm100mmSufficient ground clearance for safe operation.Minor difference, will not impact safety and effectiveness.
Front Wheel Size/Type182mm*45mm (7" x 1.75") / PU Solid tire182mm*55mm (7" x 2.1")/ PU Solid tireAppropriate wheel characteristics for safe and effective mobility.Minor difference on dimension, will not cause different performance.
Rear Wheel Size/Type203mm*50mm (8"x1.95") / PU solid tires350mm*61mm (13.7" x 2.4")/ PU solid tireAppropriate wheel characteristics for safe and effective mobility.Minor difference on dimension, will not cause different performance.
Battery Charger Input100-240V, 50/60Hz, 1.5A100-240V, 50/60Hz, 1.1AInput current differences must not raise new safety or effectiveness concerns.Minor input current difference, no new safety/effectiveness concerns.
Battery (Capacity)Lithium-ion, 24V6Ah, 6Ah x 24VDCLi-ion battery, Rechargeable; 24VDC 7.8Ah*2Battery capacity suitable for stated travel distance; no new safety/effectiveness concerns.Minor difference on capacity, no new safety/effectiveness concerns.
Back and Seat Cushion MaterialLinen cloth filled with PU foamPolyester Mesh fabric filled with spongeMaterial change must not impact safety/effectiveness, especially regarding biocompatibility.Different material, but no impact on safety/effectiveness as biocompatibility tests performed.

Study Proving Device Meets Acceptance Criteria:

The study to prove the device meets the acceptance criteria is primarily a non-clinical performance evaluation through testing against recognized international standards (ISO and IEC), and a comparison to a legally marketed predicate device.

2. Sample sized used for the test set and the data provenance:

  • Sample Size: The document does not specify a numerical "sample size" in terms of number of devices tested. It refers to "the subject device" (singular), implying that at least one physical unit of the YE200 electric wheelchair was subjected to the various performance, strength, climatic, and EMC tests. For biocompatibility, individual material samples would have been tested.
  • Data Provenance: The document does not explicitly state the country of origin of the data collection or whether it was retrospective or prospective. However, given that Shenzhen Yattll Industry Co., Ltd is based in China and the submission is to the FDA, it is highly likely that the testing was performed by a certified lab, possibly in China or a mutually recognized testing facility. Since it's for a 510(k) submission for a new device model, the testing data would be prospective relative to the submission date, generated specifically for this application.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable in the context of this 510(k) summary. The "ground truth" for an electric wheelchair's performance is established by objective, measurable outputs against defined engineering and safety standards (e.g., maximum speed, braking distance, stability angles, EMC limits, biocompatibility results), not by expert consensus or interpretation of complex medical images. The testing labs would have qualified technicians and engineers performing these measurements according to ISO/IEC protocols.

4. Adjudication method for the test set:

This is not applicable. Adjudication methods (like 2+1 or 3+1) are typically used in medical imaging studies where there's subjective interpretation by multiple readers/experts that needs to be reconciled to establish a "ground truth" for diagnosis. For engineering performance tests like those for an electric wheelchair, the tests yield objective, quantifiable results, not subjective interpretations requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This is not applicable. This document is for an electric wheelchair, which is a physical mobility device, not an AI-powered diagnostic or assistive tool for human readers/clinicians interpreting data. Therefore, an MRMC study is not relevant to demonstrating its safety and effectiveness.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

This is not applicable. This device is an electric wheelchair, which does not have an AI algorithm that functions in a standalone capacity. It is a user-controlled device with embedded electronic controls. While software validation was performed, it's for the control system of the wheelchair, not for an AI algorithm for diagnostic purposes.

7. The type of ground truth used:

The "ground truth" for the performance of the electric wheelchair is objective, measurable performance criteria defined by internationally recognized standards (ISO 7176 series, ISO 10993 series, IEC 60601-1-2) and engineering specifications. For example:

  • Static stability: Measured in degrees.
  • Braking distance: Measured in meters.
  • EMC: Measured against specific radiated/conducted emission and immunity limits.
  • Biocompatibility: Results of specific biological tests (cytotoxicity, sensitization, irritation) against defined pass/fail criteria.
  • Strength/Fatigue: Ability to withstand specified forces for a given number of cycles.

The predicate device's performance also serves as a crucial point of comparison, establishing substantial equivalence rather than a medical "ground truth" in the diagnostic sense.

8. The sample size for the training set:

This is not applicable. This is not an AI/machine learning device that requires a "training set" of data. The device's functionality is based on established engineering principles and pre-programmed controls, not learned patterns from data.

9. How the ground truth for the training set was established:

This is not applicable as there is no "training set" for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 11, 2025

Shenzhen Yattll Industry Co., Ltd % Ivy Wang Technical Manager Shanghai Sungo Management Consulting Co. Ltd. 14th Floor, 1500# Century Avenue Shanghai, Shanghai 200122 China

Re: K241641

Trade/Device Name: Electric Wheelchair (YE200) Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: December 12, 2024 Received: December 12, 2024

Dear Ivy Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Heather L. Dean -S

Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K241641

Device Name

Electric Wheelchair (YE200)

Indications for Use (Describe)

The Electric wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

K241641

l. SUBMITTER

Name: SHENZHEN YATTLL INDUSTRY CO., LTD

Address: Room 907, A&I International Center, No. 8 Gaoerfu blvd, Longhua Dist, Shenzhen

City, Guangdong Province, China

Name of contact person: Kevin Shih

Telephone: +86-13712089970

Email: yattll06@yattll.com

Date prepared: 2024-12-12

II. Device

Device trade name: Electric Wheelchair

Model: YE200

Classification name: Powered wheelchair

Regulation class: 2

Regulation number: 21CFR 890.3860

Panel: Physical Medicine

Product code: ITI

III. Predicate device

K212092

Electric Wheelchair, DC01

Anhui JBH Medical Apparatus Co., Ltd.

IV. Device description

The Electric wheelchair, YE200, is a powered wheelchair that is designed to fold when not in use so it can be more easily transported in a car trunk or similar space. It uses rechargeable lithium batteries, and is controlled by a joystick and a few other simple controls which are located on the end of the armrest. The joystick assembly may be installed on either side of the chair to accommodate both right-handed and left-handed users.

{5}------------------------------------------------

The Electric wheelchair consists of two armrests, a back rest, a seat cushion, a foldable frame, two rear drive wheels with hub motor and electromagnetic brake assemblies, two pivoting casters, two Li-ion battery pack, an off-board battery charger, a control panel (joystick) with connect cables and an electric motor controller. The device is powered by Li-ion battery pack, which can be recharged by an off-board battery charger that can be plugged into an AC socket outlet when the device is not in use.

The user can activate the controller handle (joystick) to control the speed and direction of the wheelchair movement. In addition, when the patient releases the joystick, the joystick will return back to the central position and the wheelchair will be automatically stopped soon due to automatic electromagnetic brake system. Once the joystick is activated again move to other position, the wheelchair will be re-energized.

V. Indication for use

The Electric wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. It is suitable for disabled people with mobility difficulties and elderly people.

AttributePredicate deviceSubject deviceDiscussion/Conclusion
ProprietarynameElectric wheelchairElectric wheelchair/
ModelDC01YE200/
510(k)numberK212092K241641/
DeviceclassificationnameClass IIClass IISame
Classificationregulations21 CFR 890.386021 CFR 890.3860Same
Product codeITIITISame
Similarities
Indication foruseThe wheelchair (Model: DC01) is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.The Electric wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.Same
AttributePredicate deviceSubject deviceDiscussion/Conclusion
Intended userdisabled people and elderly peopledisabled people and elderly peopleSame
Use conditionindoor and outdoor useindoor and outdoor useSame
Number of wheels4, including two front wheels and two rear wheels4, including two pivoting casters and two rear drive wheelsSame
Function of wheelsFront wheels: driven wheels suitable for rotation, acceleration, retrogradeRear wheels: driving wheels to control the speed and directiontwo front casters: driven wheels suitable for rotation, acceleration, retrogradetwo rear wheels: driving wheels to control the speed and directionSame
Folding mechanismfront and rear foldingfront and rear foldingSame.
Movement control methodBy Joystick controlBy Joystick controlSame
Driving systemDirect drive on the rear wheelsDirect drive on the rear wheelsSame
Brake systemSmart Electromagnetic brakeAutomatic electromagnetic brakeSame
Electronic controllerDual Drive Controller for BrushlessDual Drive Controller for BrushlessSame
MotorBrushless DC motor 180W x 24 VDC x 2 pcsBrushless DC motor; 24 VDC; 180 W; 2 pcsSame
Max loading weight120 kg (264 lbs)120 kg (264 lbs)Same
ArmrestPU (polyurethane)PU (polyurethane)Same
DifferencesFrame design and materialThe frame of the wheelchair is type capable of front and rear close. The main part of the frame can be folded for saving space and convenient storage and transportation. The main frame is made of carbon fiber material.The frame of the wheelchair is type of aluminum frame with front and rear folding structure and flip-up armrest. The main frame is made of Aluminum alloy material.Although the different frame material used on the subject device compared to the predicate device, while all safety and performance tests are conducted on the subject device and all passed. Such difference will not affect the safety and performance of the subject device.
AttributePredicate deviceSubject deviceDiscussion/Conclusion
Overalldimensions(LxWxH)1110 mm x 700 mm x 980 mm1107mmX620mmX950mmMinor difference on wheelchairdimensions will not cause differentperformance. All safety andperformance have been validated withthe maximum rated weight dummy.
Foldeddimensions(LxWxH)810 mm x 700 mm x 400 mm420mmX620mmX800mm
Groundclearance160 mm100mmMinor difference on ground clearancewill not impact the safety andeffectiveness of the subject device.
Front wheelsize/type182mm*45mm (7" x 1.75") /PUSolid tire182mm*55mm (7" x 2.1")/PU Solid tireMinor difference on dimension ofwheels will not cause differentperformance.
Rear wheelsize/type203mm*50mm (8"x1.95") /PUsolid tires350mm*61mm (13.7" x 2.4")/PU solid tire
Max speedforward6 km/h5.7km/hMinor difference on max. forwardingspeed will not cause differentperformance.
MinimumbrakingdistancefrommaximumspeedForward: 0.5mForward: 0.75 mMinor difference on braking distancebetween subject device andpredicate device, while all relevanttests are performed according to ISO7176-3, no safety and effectivenessconcerns will be affected.
Maximumsafeoperationalinclinedegree8 °6 °Minor difference on safe operationalincline degree will not cause newsafety and effectiveness concerns areraised as both the static and dynamicstability under specific incliningdegree have been evaluatedaccording to standard ISO 7176series.
BatterychargerOff-board chargerInput: 100-240V, 50/60Hz,1.5A,Output: 24 Vdc, 2A;Off-board chargerInput: 100-240V, 50/60Hz,1.1A,Output: 24 Vdc, 2A;Minor input current difference will notcause new safety and effectivenessconcerns raised.
BatteryLithium-ion, 24V6Ah, 6Ah x 24VDCLi-ion battery, Rechargeable;24 VDC 7.8Ah*2Same rated voltage, minor differenceon battery capacity will not causedifferent performance. This differencewill not raise any new safety andeffectiveness concerns.
AttributePredicate deviceSubject deviceDiscussion/Conclusion
Maximum distance of travel on the fully charged battery20 km16 kmMinor difference on travel distance will not cause different performance. This difference will not raise any new safety and effectiveness concerns.
Turning Radius800 mm875 mmThe minor difference in the turning radius is caused by different size of wheelchair and may cause a little bit inconvenience when it turns in a narrow space while the minor difference will not raise any new safety and effectiveness concerns.
Maximum obstacle climbing20 mm15 mmMinor difference in the obstacle climbing will not impact the safety and effectiveness of the subject device.
back and sea cushionlinen cloth filled with PU foamPolyester Mesh fabric filled with spongeDifferent material used for parts in contact with user, which such differences will not impact the safety and effectiveness of the subject device as biocompatibility tests are carried out according to ISO 10993 series.

VI. Comparison of technological characteristics with the predicate device

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

VII. Summary of substantial equivalence discussion

The subject electric wheelchair complied with the requirements of ISO 7176-1:2014, ISO 7176-2:2017, ISO 7176-3:2012, ISO 7176-4:2008, ISO 7176-5:2008, ISO 7176-6:2018, ISO 7176-7:1998, ISO 7176-8:2014, ISO 7176-9:2009, ISO 7176-10:2008, ISO 7176-11:2008, ISO 7176-13:1989, ISO 7176-14:2008, ISO 7176-15:1996, ISO 16840-10: 2021, ISO 7176-21:2009, ISO 7176-22:2014, ISO10993-5:2009, ISO 10993-10:2021, ISO 10993-23: 2021, IEC 60601-1-2: 2020.

The intended uses for both devices are the same. Both mainframes of the two devices are folded by way of front and rear close. As for different frame materials used and minor difference on safe operational incline degree for both devices, considering all safety and performance tests are carried out on the subject device with favorable result, such difference will not affect the safety and performance of the subject device. The design principles of the

{9}------------------------------------------------

controller and Driving system are the same, and both meet the requirements of the ISO 7176-14:2008. Software validation is carried out on both control systems. Brake system and speed control are designed in the same way as well, and both meet the requirements of the ISO 7176-3:2012. Turning radius, Maximum obstacle climbing and ground clearance are slightly different while such differences will not impact the safety and effectiveness of the subject device or raise new safety and effectiveness concerns as well as both meet the requirements of the ISO 7176-2:2017, ISO 7176-10:2008.

The biocompatibility of the Predicate device and Subject device meet the requirements of the ISO 10993-5:2009 & ISO 10993-10:2021 & ISO 10993-23: 2021.

All seat cushion/back cushion and armrest are made of flame retardant material for both devices. Therefore, both devices are assured to be under the same safety level.

In conclusion, the technological characteristics, features, specifications, materials, mode of operation, and intended use of the device substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.

VIII. Summary of non-clinical testing

> Performance testing-bench

The following performance data were provided to verify that the subject device met all design specifications and provided support of the substantial equivalence determination.

  • Risk Analysis developed in accordance with ISO 14971: 2019.
  • Software validation
  • ISO 7176-1:2014 Wheelchairs - Part 1: Determination of static stability
  • . ISO 7176-2:2017 Wheelchairs - Part 2: Determination of dynamic stability of electric wheelchairs
  • ISO 7176-3:2012 Wheelchairs - Part 3: Determination of effectiveness of brakes
  • ISO 7176-4:2008 Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range
  • . ISO 7176-5:2008 Wheelchairs - Part 5: Determination of dimensions, mass and maneuvering space

{10}------------------------------------------------

  • . ISO 7176-6:2018 Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of electric wheelchairs
  • . ISO 7176-7:1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions
  • . ISO 7176-8:2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strength
  • . ISO 7176-9:2009 Wheelchairs - Part 9: Climatic tests for electric wheelchairs
  • . ISO 7176-10:2008 Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs
  • ISO 7176-11:2012 Wheelchairs -- Part 11: Test dummies
  • ISO 7176-13:1989 Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces.
  • ISO 7176-14:2022 Wheelchairs -- Part 14: Power and control systems for electrically ● powered wheelchairs and scooters -- Requirements and test methods
  • . ISO 7176-15:1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling.
  • . ISO 16840-10:2021 Wheelchair seating - Part 10: Resistance to ignition of postural support devices - Requirements and test method.
  • ISO 7176-21:2009 Wheelchairs - Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters
  • . ISO 7176-22: 2014 Wheelchairs - Part 22: Set-up procedures
  • . Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2:2020.

A Biocompatibility of patient-contacting material

Biocompatibility Tests are carried out in accordance with ISO10993-1: 2018, including cytotoxicity (ISO 10993-5:2009), sensitization (ISO 10993-10:2021) and irritation (ISO 10993-23: 2021). Details of the test summary see biocompatibility summary.

IX. Summary of clinical testing

No animal study and clinical studies are available for our device. Clinical testing was not required to demonstrate the substantial equivalence of the power wheelchair to its predicate

{11}------------------------------------------------

device.

Conclusions x.

The conclusion drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed predicate device K212092.

§ 890.3860 Powered wheelchair.

(a)
Identification. A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.(b)
Classification. Class II (performance standards).