K Number
K241577

Validate with FDA (Live)

Device Name
Iconix® Anchor
Manufacturer
Date Cleared
2024-07-03

(30 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Iconix() Anchors are intended to be used for soft-tissue to bone fixation in the foot, ankle, knee, hip, hand, wrist, elbow and shoulder. Specific indications are listed below.

Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction

Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Repair

Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/ Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all digits, Digital Tendon Repair

Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valqus Reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction

Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis, Medial Patellofemoral Ligament (MPFL) Repair/ Reconstruction, Quadriceps Tendon Repair

Hip: Capsular Repair, Acetabular Labral Repair, Gluteal Tendon Repair, Proximal Hamstring Repair

Device Description

The Iconix® Anchors are soft-tissue to bone fixation devices, provided preloaded on a disposable inserter. The device is composed of a braided polyester anchor body that contains one or more working sutures. Sutures supplied meet United States Pharmacopeia (USP) requirements for nonabsorbable suture except for diameter. Suture dyes are FDA approved (K100006 and K190817). The standard and the self-punching inserters are comprised of metallic shaft with over molded handle. The device is sterilized by ethylene oxide gas and is provided sterile for single use. The Iconix® Anchors size is 2.3mm. The anchors will be sold sterile for single use with no components or accessories. The product is intended to be used with the same accessories used to implant the predicate device. The device is intended for use in a hospital/clinic/surgical setting, and it is intended for adults only.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device called "Iconix® Anchor." This document is a regulatory submission to the FDA demonstrating that the new device is substantially equivalent to a previously cleared predicate device.

It is crucial to understand that a 510(k) submission for a non-AI/ML device like the Iconix® Anchor typically does not involve the types of studies, acceptance criteria, multi-reader multi-case (MRMC) studies, or ground truth establishment that are relevant for AI/ML-driven diagnostic devices. The questions posed in your prompt are highly specific to AI/ML software as a medical device (SaMD) or AI-augmented medical devices.

The Iconix® Anchor is a physical medical device (a soft-tissue to bone fixation fastener), not an AI/ML diagnostic or therapeutic system. Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are centered around mechanical performance, biocompatibility, sterilization, and substantial equivalence to a predicate device, rather than diagnostic accuracy, reader performance, or AI model metrics.

Based on the provided document, here's how the information relates to your questions, highlighting the differences for a physical device:


1. A table of acceptance criteria and the reported device performance

The document does not present a formal "table of acceptance criteria" with reported performance metrics in the way one would for an AI/ML algorithm (e.g., sensitivity, specificity thresholds). Instead, the acceptance criteria are implicit in meeting the performance requirements for a bone anchor, and the reported performance is qualitative, stating that the device "met all requirements for its intended use" and "performed comparably to the predicate device."

Acceptance Criteria Category (Implicit)Reported Device Performance (from text)
Mechanical Performance"Iconix® Anchor met all requirements for its intended use." "performed comparably to the predicate device in insertion, cyclic and pullout testing."
Biocompatibility"biocompatibility testing per ISO10993-1:2018... validated."
Sterilization"sterilization adoption validation" (device is "sterilized by ethylene oxide gas and is provided sterile for single use").
Shelf-life/Packaging Stability"shelf-life testing" and "stability testing on the product packaging per ISO 11607-1:2006... validated."
Usability"usability engineering validation with simulated use in a cadaveric models performed per EN62366: 2015- Medical devices - Application of usability engineering to medical devices."
Suture Tensile Strength (USP)"The UHMWPE sutures are tested per USP performance requirements for tensile strength." (Stated that sutures meet USP requirements except for diameter).
Substantial Equivalence"The Iconix® Anchor is substantially equivalent to the previously cleared All-Suture Anchor per K231078 'predicate device'."

Regarding AI-specific questions (2-9), the nature of this submission means most are not applicable or the information is not provided because it's not a software/AI device:

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not applicable in the AI/ML sense. The "test set" for a physical device refers to the physical samples tested for mechanical properties, biocompatibility, etc. The document does not specify the number of individual anchors tested, only the types of tests performed (insertion, cyclic, pullout testing) and that cadaveric models were used for usability.
  • Data provenance (country of origin, retrospective/prospective) is not relevant for physical device testing in this context.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable in the AI/ML sense. "Ground truth" for this device is established by physical measurements and engineering validation against established standards (e.g., ISO, USP) and comparison to a predicate device's performance, not by expert consensus on image interpretation. For usability, medical professionals might have been involved in the cadaveric testing, but their number and specific qualifications are not detailed.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. Adjudication methods are for resolving discrepancies in expert interpretations of data, typically in diagnostic studies. This is not relevant for mechanical and biological testing of a physical device.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. MRMC studies are specific to evaluating human reader performance with and without AI assistance for diagnostic tasks. This device is not an AI diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This question pertains to the performance of an AI algorithm independent of human interaction. The Iconix® Anchor is a physical implant, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • For the physical device: The "ground truth" is based on engineering standards, validated test methods (e.g., ISO, USP), and direct physical measurements of properties like tensile strength, insertion force, pullout strength, and biocompatibility. Usability was evaluated using simulated use in cadaveric models.

8. The sample size for the training set

  • Not applicable. This question is for AI/ML algorithms. The Iconix® Anchor does not have a "training set" in the machine learning sense. Any design iterations or preliminary testing during development would be more akin to "training," but not in the data science context.

9. How the ground truth for the training set was established

  • Not applicable. As there is no AI/ML training set, this question is not relevant.

In summary: The provided document is a 510(k) for a physical medical device (a bone anchor), not an AI/ML-driven diagnostic or therapeutic device. The "acceptance criteria" and "study" described are focused on the mechanical, biological, and sterility performance of the physical product, and its substantial equivalence to a predicate device, rather than the performance metrics typically associated with AI/ML systems. Therefore, most of your specific questions related to AI/ML study design are not addressed in this type of regulatory submission.

{0}------------------------------------------------

July 3, 2024

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Riverpoint Medical Bianca Silva de Sousa Regulatory Associate II 825 NE 25th Ave Portland, Oregon 97232

Re: K241577

Trade/Device Name: Iconix® Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: May 31, 2024 Received: June 3, 2024

Dear Mrs. Silva de Sousa:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Robert M. Stefani -S 2024.07.03 09:14:23 -04'00'

For: Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K241577

Device Name

Iconix(R) Anchor

Indications for Use (Describe)

The Iconix() Anchors are intended to be used for soft-tissue to bone fixation in the foot, ankle, knee, hip, hand, wrist, elbow and shoulder. Specific indications are listed below.

Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction

Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Repair

Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/ Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all digits, Digital Tendon Repair

Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valqus Reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction

Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis, Medial Patellofemoral Ligament (MPFL) Repair/ Reconstruction, Quadriceps Tendon Repair

Hip: Capsular Repair, Acetabular Labral Repair, Gluteal Tendon Repair, Proximal Hamstring Repair Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Riverpoint. The logo consists of the letters "RP" in white on a green rounded square. To the right of the square is the word "RIVERPOINT" in green sans-serif font.

510(k) SUMMARY

Iconix® Anchor

Submitter Information

Submitter's Name:Riverpoint Medical
Address:825 NE 25th Ave.Portland, OR 97232
Phone Number:(503) 517-8001
Fax Number:(503) 517-8002
Registration Number:3006981798
Contact Person:Bianca Silva de Sousa(503) 517-8001
Date of Preparation:July 2, 2024

Device Name

Trade Name:Iconix® Anchor
Common or Usual Names:Fastener, Fixation, Nondegradable, Soft Tissue
Classification Name:21 CFR 888.3040: Smooth or Threaded Metallic BoneFixation Fastener

Device Classification

FDA Class:II
Product Classification:888.3040: Smooth or Threaded Metallic Bone FixationFastener
Classification Code:MBI
Review Panel:Orthopedic
Premarket ReviewCenter for Device and Radiological HealthOffice of Health Technology (OHT6: Orthopedic Devices)Division of Health Technology 6C Restorative, Repair andTrauma Devices

Predicate Device

K231078 – All-Suture Anchor

Reference Devices

K233468 - Iconix Anchor

K240213 - Iconix Knotless Anchor

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for Riverpoint. The logo consists of the letters "RP" inside of a rounded square. To the right of the square is the word "RIVERPOINT" in all capital letters. The text and the square are green.

Device Description

The Iconix® Anchors are soft-tissue to bone fixation devices, provided preloaded on a disposable inserter. The device is composed of a braided polyester anchor body that contains one or more working sutures. Sutures supplied meet United States Pharmacopeia (USP) requirements for nonabsorbable suture except for diameter. Suture dyes are FDA approved (K100006 and K190817). The standard and the self-punching inserters are comprised of metallic shaft with over molded handle. The device is sterilized by ethylene oxide gas and is provided sterile for single use. The Iconix® Anchors size is 2.3mm. The anchors will be sold sterile for single use with no components or accessories. The product is intended to be used with the same accessories used to implant the predicate device. The device is intended for use in a hospital/clinic/surgical setting, and it is intended for adults only.

The classification for the Iconix® Anchor is FDA Class II device with product classification 21 CFR &888.3040: Smooth or threaded metallic bone fixation fastener, Product Code MBI.

Intended Use and Indications for Use

The Iconix® Anchors are intended to be used for soft-tissue to bone fixation in the foot, ankle, knee, hip, hand, wrist, elbow and shoulder. Specific indications are listed below.

Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction

Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Repair

Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all digits, Digital Tendon Repair

Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus Reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction

Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis, Medial Patellofemoral Ligament (MPFL) Repair/ Reconstruction, Quadriceps Tendon Repair

Hip: Capsular Repair, Acetabular Labral Repair, Gluteal Tendon Repair, Proximal Hamstring Repair

Performance Data

The sutures used to construct the Iconix® Anchor meet requirements established by the United

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Riverpoint. The logo consists of the letters "RP" in a green rounded square, followed by the word "RIVERPOINT" in green sans-serif font. The logo is simple and modern, and the green color suggests growth and stability.

States Pharmacopeia (USP), except for diameter. The UHMWPE sutures are tested per USP performance requirements for tensile strength.

FDA Guidance "Bone Anchors - Premarket Notification (510(k)) Submissions Guidance for Industry and Food and Drug Administration Staff" and FDA Guidance "Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA" were followed during the preparation of this submission.

Non-clinical performance testing for the Iconix® Anchor included shelf-life testing, sterilization adoption validation, biocompatibility testing per ISO10993- 1:2018 - Biological Evaluation of Medical Devices, stability testing on the product packaging per ISO 11607-1:2006 - Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems, usability engineering validation with simulated use in a cadaveric models performed per EN62366: 2015- Medical devices - Application of usability engineering to medical devices.

Non-clinical mechanical testing was performed to verify the fixation strength of the Iconix® Anchor using insertion, cyclic and pullout testing as compared to the predicate device. Results of performance testing for the Iconix® Anchor device concluded that the device performed comparably to the predicate device in insertion, cyclic and pullout testing and the validations performed demonstrated that the Iconix® Anchor met all requirements for its intended use.

Substantial Equivalence and Comparison of Technical Characteristics

The Iconix® Anchor is substantially equivalent to the previously cleared All-Suture Anchor per K231078 "predicate device." The Iconix® Anchor has the same intended use, same principles of operation, and similar technological characteristics as the predicate device. Both the Iconix® Anchor and the predicate device have the same anchor geometry and method of fixation of suture to anchor. They're also packaged using the same packaging materials and sterilized using the same processes. The only difference between Iconix® Anchor subject device and the predicate device is in technological characteristics, specifically in (i) range of suture size – the subject device contains sutures size 1.2mm and 2.2mm, while the predicate device has 1.2mm and #2. (ii) anchor materials - the anchor of the subject device does not contain a coating. However, the technological characteristics are within the range of currently marketed devices, as seen in the reference devices Iconix Knotless Anchor (K240213) and Iconix Anchors (K233468), and they have been assessed through risk analysis, not raising any new questions of effectiveness. Therefore, the Iconix® Anchor "subject device" is substantially equivalent to the "predicate device" and does not raise any new issues of safety or effectiveness.

Conclusion

The information provided in this Special 510(k) demonstrates that the Iconix® Anchor is substantially equivalent to the predicate device.

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.