K Number
K241557

Validate with FDA (Live)

Device Name
Cercon® yo ML
Manufacturer
Date Cleared
2024-08-28

(89 days)

Product Code
Regulation Number
872.6660
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Cercon® yo ML is indicated in the anterior and posterior segments for:

  • Crowns
  • Multi-unit bridges (with a maximum of two pontics between abutment crowns)
  • Inlays, onlays and veneers
Device Description

The proposed Cercon® yo ML is a partially sintered ceramic blank composed of yttrium oxide-(yttria-) stabilized zirconium oxide (zirconia) (Y-TZP). Cercon® yo ML is a multilayer disk built out of different shade layers and yttrium oxide contents which create a natural gradient of the restoration after sintering. The proposed device is supplied to dental professionals as a blank and then processed by machining and subsequent sintering to fabricate dental ceramic restorations.

AI/ML Overview

The provided text describes acceptance criteria and the study conducted for the Cercon® yo ML device, specifically focusing on non-clinical performance data.

Here's a breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance

TestTest Method/Applicable StandardAcceptance CriteriaResults
Material Performance (ISO 6872:2015 Amd. 1. 2018)
Flexural StrengthISO 6872:2015 Amd 1. 2018 Dentistry Ceramic Materials$\geq$ 800 MPaPass
CTE (Coefficient of Thermal Expansion)ISO 6872:2015 Amd 1. 2018 Dentistry Ceramic MaterialsThe CTE of the ceramics shall not deviate by more than 0.5 x 10 $^{-6}$ K $^{-1}$ from the value stated by the manufacturer.Pass
SolubilityISO 6872:2015 Amd 1. 2018 Dentistry Ceramic Materials<100µg/cm² for each tested shadePass
RadioactivityISO 6872:2015 Amd 1. 2018 Dentistry Ceramic Materials<1.0 Bq/g of $^{238}$ UPass
Shrinkage factorClass 5 Type II materialAbsolute Accuracy ± 0.002Pass
UniformityISO 6872:2015 Amd 1. 2018 Dentistry Ceramic MaterialsPigments shall be uniformly dispersedPass
Freedom from extraneous materialsISO 6872:2015 Amd 1. 2018 Dentistry Ceramic MaterialsFree from extraneous materials when assessed by visual inspection and radioactivity measurementPass
Compatibility Performance (ISO 9693:2019)
De-bonding / crack-initiationISO 9693:2019 Dentistry - Compatibility testing>20 MPa zirconia material with at least one specified dental veneering ceramicPass
Thermal shock resistanceISO 9693:2019 Dentistry - Compatibility testingMeet critical quenching temperature of at least 120 °CPass
Biocompatibility Performance(Various ISO 10993 standards and USP <151>)Meets ISO 10993 requirements (implied by passing tests)Pass

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not explicitly state the sample sizes for the bench tests. It refers to general standards (ISO 6872, ISO 9693, ISO 10993 series) which typically define the number of samples required for each test. The data provenance is not mentioned, but these are non-clinical engineering and biological tests, not patient-specific data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This section is not applicable. The studies described are non-clinical bench tests and biocompatibility assessments, which rely on established scientific and engineering standards and laboratory measurements, not expert consensus for ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This section is not applicable. Adjudication methods are typically used in clinical studies involving interpretation of data by multiple experts. The described studies are non-clinical bench tests.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There is no mention of an MRMC study or AI assistance in the provided text. The device, Cercon® yo ML, is a partially sintered ceramic blank for dental restorations, not an AI-powered diagnostic or assistive device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This section is not applicable. The device is a material for dental restorations, not a software algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the non-clinical performance data:

  • Material properties (Flexural Strength, CTE, Solubility, Radioactivity, Shrinkage factor, Uniformity, Freedom from extraneous materials): The ground truth is based on the specified physical, chemical, and mechanical properties as defined by the ISO 6872:2015 Amd 1. 2018 standard. The "ground truth" is the established limit or range defined by the standard.
  • Compatibility (De-bonding / crack-initiation, Thermal shock resistance): The ground truth is based on the specified performance criteria defined by the ISO 9693:2019 standard.
  • Biocompatibility: The ground truth is established by the requirements of the ISO 10993 series of standards and USP <151>, which define acceptable biological responses and toxicity levels.

8. The sample size for the training set

This section is not applicable as the studies described are non-clinical bench tests and biocompatibility assessments, not machine learning model training.

9. How the ground truth for the training set was established

This section is not applicable as there is no training set mentioned or implied for this type of device and testing.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the agency's name on the right. The symbol on the left is a stylized depiction of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. The words "FOOD & DRUG" are larger than the other words.

August 28, 2024

Dentsply Sirona Rebecca Sporer Principal Regulatory Affairs Specialist 221 West Philadelphia Street Suite 60W York, Pennsylvania 17401

Re: K241557

Trade/Device Name: Cercon® yo ML Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: May 31, 2024 Received: May 31, 2024

Dear Rebecca Sporer:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bobak Shirmohammadi -S

For Michael E. Adjodha, M.ChE., RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices

{2}------------------------------------------------

OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K241557

Device Name

Cercon® yo ML

Indications for Use (Describe)

Cercon® yo ML is indicated in the anterior and posterior segments for:

  • Crowns

  • Multi-unit bridges (with a maximum of two pontics between abutment crowns)

  • Inlays, onlays and veneers

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Dentsply Sirona 221 West Philadelphia Street Suite 60W York, PA 17401

Image /page/4/Picture/1 description: The image shows the logo for Dentsply Sirona. The logo consists of a blue abstract shape on the left and the words "Dentsply Sirona" in blue on the right. The abstract shape is made up of two curved lines that form an S shape. The words "Dentsply Sirona" are stacked on top of each other.

510(k) SUMMARY for Cercon® yo ML

K241557

1. Submitter Information: Dentsply Sirona 221 West Philadelphia Street Suite 60W York, PA 17404

Contact Person: Rebecca Sporer Telephone Number: 717-849-4793 Email address: Rebecca.sporer@dentsplysirona.com

Date Prepared: 31 May 2024

    1. Device Name:
Proprietary Name:Cercon® yo ML
Classification Name:Porcelain powder for clinical use
CFR Number:21 CFR 872.6660
Device Class:2
Product Code:EIH
    1. Predicate Device:
Primary Predicate Device Name510(k)Company Name
CerconK162888Dentsply Sirona
Predicate Device Names510(k)Company Name
CELTRA PressK161269Dentsply Sirona
Cercon htK112152Dentsply Sirona
Reference Device NameTechnological Characteristics andFormulation510(k)Company Name
CEREC Cercon 4D™ Abutment Blocks,CEREC Cercon 4D™ Abutment SystemK234018Dentsply Sirona

{5}------------------------------------------------

4. Description of Device:

The proposed Cercon® yo ML is a partially sintered ceramic blank composed of yttrium oxide-(yttria-) stabilized zirconium oxide (zirconia) (Y-TZP). Cercon® yo ML is a multilayer disk built out of different shade layers and yttrium oxide contents which create a natural gradient of the restoration after sintering. The proposed device is supplied to dental professionals as a blank and then processed by machining and subsequent sintering to fabricate dental ceramic restorations.

5. Indications for Use:

Cercon® yo ML is indicated in the anterior and posterior segments for:

  • Crowns
  • · Multi-unit bridges (with a maximum of two pontics between abutment crowns)
  • Inlays, onlays and veneers
    1. Comparison of Intended Use and Indications for Use and Technological Characteristics:

The Indications for Use of the proposed device do not constitute a new intended use. The proposed Cercon® yo ML has the same intended use and similar Indications for Use compared to the primary predicate and predicate devices in that they are a ceramic intended for fixed dental prosthetic restorations. Table 6.1 compares the Intended Use and Indications for Use when cared to the predicate devices.

{6}------------------------------------------------

Table 6.1:Comparison of Intended Use and Indications for Use of the proposed and predicate devices
Proposed DevicePrimary PredicatePredicate DevicesDiscussion
Cercon® yo ML(K241557)DeviceCercon®(K162888)CELTRA Press(K161269)Cercon® ht(K112152)
Intended UseCercon® yo ML blanks are aceramic intended for fixeddental prosthetic restorations.Cercon® are a ceramicintended for fixed dentalprosthetic restorations.CELTRA Press are aceramic intended forfixed dental prostheticrestorations.Cercon® ht are aceramic intended forfixed dental prostheticrestorations.Same
Indications forUseCercon® yo ML is indicatedfor in the anterior andposterior segments for:• Crowns• Multi-unit bridges (with amaximum of two ponticsbetween abutment crowns)• Inlays, onlays and veneersCercon® is indicated forall ceramic restorationsfor anterior and posteriorlocations, includingcrowns and 3-unitbridges in anterior andposterior regions.Indications for Use:CELTRA Press is an allceramic system for thecreation ofOcclusal veneersThin veneersVeneersInlaysOnlaysCrowns in the anteriorand posterior region3-unit bridges in theanteriorand posterior region3-unit bridges in thepremolar region up tothe second premolar asthe terminal abutmentCrown, splinted crownor 3 unit bridge up tothe second premolarplaced on top of animplant abutment.Cercon® ht isindicated in theanterior and posteriorsegments for:• Crowns• telescopic primarycrowns• multi-unit bridges(with no more thetwo pontics betweenabutment crowns)The proposed, primarypredicate, and predicatedevices share the indicationfor crowns and bridges withapplications in the anteriorand posterior locations.The proposed, primarypredicate and predicatedevices share an indicationfor multi-unit bridges (e.g. 3-unit bridges).The proposed and predicate(Celtra Press [K161269])devices are both indicated forinlays, onlays and veneers.The proposed device does notinclude an indication fortelescopic primary crowns.

{7}------------------------------------------------

7. Comparison of Technological Characteristics:

The proposed Cercon® yo ML is similar in design, principles of operation and meet the requirements of ISO 6872 when compared to the primary predicate device, Cercon (K16288), and the predicate devices, CELTRA Press (K161269) and Cercon ht (K112152). When comparing the material, the proposed, primary predicate (Cercon ht [K112152]) and reference devices are all zirconia material whereas the predicate (Celtra Press [K161269]) device is a lithium silicate glass material. The proposed Crcon® yo ML is similar in composition to the primary predicate (Cercon [K162888]), the predicate (Cercon ht [K112152]) and the reference (CEREC Cercon 4D™ Abutment Blocks, CEREC Cercon 4D™ Abutment System [K234018]) devices with the main difference in pigment used for esthetics (shade preference). Table 7.1 compares the technological characteristics of the proposed, Cercon® yo ML to the predicates and reference devices.

Table 7.1:Technological Characteristics Comparison the proposed Cercon® yo ML, the primary predicate device Cercon® (K162888), predicates CELTRA Press (K161269) and Cercon® ht (K112152), and the reference device CEREC Cercon 4D™ Abutnent System (K234018)

Item ofProposedPrimaryPredicate DevicesReference DeviceSimilarities and Differences
ComparisonCercon® yo ML(K241557)PredicateCercon®(K162888)CELTRA Press(K161269)Cercon® ht(K112152)CEREC Cercon4DTM AbutmentBlocks, CERECCercon 4DTMAbutment System(K234018)
Product CodeEIHEIHEIHEIHNHA, PNPSimilar. The reference device,CEREC Cercon 4D AbutmentBlocks, CEREC Cercon 4DTMAbutment System (K234018) iscleared under a different productcode as it is marketed as asystem. The block materialwithin the system is processedand compatible with implantsystems and TiBases.
ManufacturerDentsply SironaDentsply SironaDentsply SironaDentsply SironaDentsply SironaSame
RestorationSingle and multi-unitSingle and multi-unitSingle and multi-unitSingle and multi-unitSingle unitSimilar. The reference device,CEREC Cercon 4D AbutmentBlocks, CEREC Cercon 4DTMAbutment System (K234018) isonly used for single restorations.
Item ofProposedPrimaryPredicate DevicesReference DeviceSimilarities and Differences
ComparisonCercon® yo ML(K241557)PredicateCercon®(K162888)CELTRA Press(K161269)Cercon® ht(K112152)CEREC Cercon4DTM AbutmentBlocks, CERECCercon 4DTMAbutment System(K234018)
MaterialZirconiaZirconiaLithium silicateglassZirconiaZirconiaSame except when compared tothe predicate Celtra Press(K161269) which is a lithiumsilicate glass.
SterilityNon-SterileNon-SterileNon-SterileNon-SterileNon-SterileSame
BiocompatibilityMeets ISO 10993requirementsMeets ISO 10993requirementsMeets ISO10993requirementsMeets ISO 10993requirementsMeets ISO 10993requirementsSame
Flexural Strength>800 MPa perISO 6872:2015Type II Class 5>500MPa perISO 6872:2015Type II Class 4>500 MPa perISO 6872:2015Type II Class 4>800MPa per ISO6872:2008 Type IIClass 6>800 MPa per ISO6872:2015 Type IIClass 5The proposed device has ahigher flexural strength whencompared to the primarypredicate and the predicatedevice, Celtra Press (K161269)per the type and class defined inISO 6872:2015. The proposeddevice is comparable in strengthto the predicate, Cercon ht(K112152), and the referencedevice, CEREC Cercon 4DAbutment Blocks, CERECCercon 4DTM Abutment System(K234018).
DesignDiskDiskIngotDiskBlockThe proposed, primary predicate,and the predicate [Cercon ht(K112152)] devices are designedas a disk. The predicate, Celtra
Item ofProposedCercon® yo ML(K241557)PrimaryPredicateCercon®(K162888)Predicate DevicesReference DeviceSimilarities and Differences
ComparisonImage: Cercon yo MLImage: Cercon xtCELTRA Press(K161269)Image: CELTRA PressCercon® ht(K112152)Image: Cercon htCEREC Cercon4D™ AbutmentBlocks, CERECCercon 4D™Abutment System(K234018)Image: CEREC Cercon 4D Abutment Blockspress (K161269) is designed asan ingot and the refence device,CEREC Cercon 4D AbutmentBlocks, CEREC Cercon 4D™Abutment System (K234018) isa block.

{8}------------------------------------------------

Table 7.1.Technological Characteristics Comparison between the primary predicate device Cercon® (K162888), predicates CELTRA Press (K161269) and Cercon® ht (K112152), and the reference device CEREC Cercon 4D™ Abutnent System (K234018)

{9}------------------------------------------------

Table 7.1:Technological Characteristics Comparison between the primary predicate device Cercon® (K162888), predicates CELTRA Press (K161269) and Cercon® ht (K112152), and the reference device CEREC Cercon 4D™ Abutnent System (K234018)

{10}------------------------------------------------

8. Non-Clinical Performance Data:

Performance Bench Testing:

The proposed device was tested and conforms to:

  • ISO 6872:2015 Amd 1. 2018 Dentistry Ceramic Materials O
  • 0 ISO 9693:2019 Dentistry Compatibility testing for metal-ceramic-ceramic-ceramic-ceramic systems

Table 8.1 summarizes the bench testing conducted on the proposed device according to ISO 6872:2015 Amd 1. 2018. Table 8.2 summarizes the bench testing conducted on the proposed device according to ISO 9693:2019.

Table 8.1: Performance Bench Testing according to ISO 6872:2015 Amd. 1. 2018TestMethod/ApplicableStandardAcceptance CriteriaResults
Flexural StrengthISO 6872:2015 Amd 1.2018 Dentistry CeramicMaterials$\geq$ 800 MPaPass
CTE (Coefficient ofThermal Expansion)The CTE of the ceramics shall notdeviate by more than 0.5 x 10 $^{-6}$ K $^{-1}$ fromthe value stated by the manufacturer.Pass
Solubility<100µg/cm² for each tested shadePass
Radioactivity<1.0 Bq/g of $^{238}$ UPass
Shrinkage factorClass 5 Type II materialAbsolute Accuracy ± 0.002Pass
UniformityPigments shall be uniformly dispersedPass
Freedom from extraneousmaterialsFree from extraneous materials whenassessed by visual inspection andradioactivity measurementPass
Table 8.2: Performance Bench Testing according to ISO 9693:2019
Test PerformedTestMethod/ApplicableStandardAcceptance CriteriaResults
De-bonding / crack-initiationISO 9693:2019 Dentistry- Compatibility testing>20 MPa zirconia material with at leastone specified dental veneering ceramicPass
Thermal shock resistancefor metal-ceramic andceramic-ceramic systemsMeet critical quenching temperature ofat least 120 °CPass

{11}------------------------------------------------

Biocompatibility Testing:

A biological risk assessment was conducted on the proposed device, Cercon® yo ML. Review of available information on raw materials, manufacturing processes, chemical characterization tests and existing preclinical biological testing data concludes that the test results meet the requirements of the following ISO 10993 standard series.

The following biological tests were conducted:

  • o ISO 7405: 2018 Dentistry-Evaluation of biocompatibility of medical device used in dentistry
  • o ISO 10993-1:2018 Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process
  • o ISO 10993-5:2009 Biological Evaluation of Medical Devices-Part 5: Tests for in vitro cytotoxicity
  • o ISO 10993-10:2021 Biological Evaluation of Medical Devices-Part 10: Test for Skin Sensitization
  • o ISO 10993-18:2020 Biological Evaluation of Medical Devices-Part 18: Chemical Characterization of Medical Device Within a Risk Management Process
  • o ISO 10993-23:2021 Biological Evaluation of Medical Device-Part 23: Test for Irritation
  • o USP <151> Pvrogen Test (USP Rabbit Test)

Conclusion:

Minor differences in the technological characteristics between the proposed and primary predicate devices were evaluated through performance bench testing. Minor differences in formulation between the proposed, primary predicate, predicate [Cercon ht (K112152)] and reference devices was evaluated through biocompatibility testing. Testing results demonstrated that the proposed device does not raise new questions regarding safety and effectiveness. Therefore, the nonclinical testing data supports the conclusion that the proposed device performs as well as the primary predicates and reference devices.

    1. Clinical Performance Data:
      No data from human clinical studies has been included to support the substantial equivalence of the proposed Cercon® yo ML.

10. Conclusion Regarding Substantial Equivalence:

Cercon® yo ML has the same intended use and similar Indications for Use as the primary predicate (Cercon [K16288]) and the predicates (Celtra Press [K161269] and Cercon ht [K112152]). Cercon® yo ML incorporates the same fundamental technology as the primary predicate (Cercon [K162888]) and the predicate (Cercon ht [K112152]) and the reference device (CEREC Cercon 4D™ Abutment Blocks, CEREC Cercon 4D™ Abutment System [K234018]). The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well or better than the legally marketed primary predicate, predicates and reference devices.

§ 872.6660 Porcelain powder for clinical use.

(a)
Identification. Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.(b)
Classification. Class II.