K Number
K241532

Validate with FDA (Live)

Date Cleared
2024-10-22

(145 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Flexible Video-Choledo-Cysto-Ureteroscope System is indicated for endoscopic examination in the urinary tract and can be used percutaneously to examine the interior of the kidney, and using additional accessories, to perform various diagnostic and therapeutic procedures. The Flexible Video-Choledo-Cysto-Ureteroscope System is also indicated for the examination of bile ducts surgically, and using additional accessories, to perform various diagnostic and therapeutic procedures during cholecystectomy.

Device Description

Imaging Processor System (Including Light Source) is composed of lighting system, image processing board. The lighting system provides the light source for the endoscope probe at the back end. The image processing board receives electronic signals from the front-end camera module and processes them, and finally transmits them to the display through the video interface. Flexible Video-Choledo-Cysto-Ureteroscope is a kind of medical electronic optical instrument, also known as optical camera, which can enter into the human bladder, ureter, biliary and pancreatic duct for observation and diagnosis. The operator delivers the optical camera system to the site of diagnosis and treatment by means of a mechanical part with a flexible insertion tube and a system of bends. The product is equipped with tiny size digital imaging parts -- photoelectric sensors "CMOS", on which the objects in human cavity will be transferred though lens optical system, and converts light signals into electrical signals. The electrical signal will be transferred to Imaging Processor System (Including Light Source) and display images on it's monitor output for doctor observation and diagnosis.

AI/ML Overview

This document is a 510(k) premarket notification for a new medical device, the Flexible Video-Choledo-Cysto-Ureteroscope System (PL-2100). This type of submission relies on demonstrating "substantial equivalence" to a predicate device, meaning it's as safe and effective as a device already legally marketed. Therefore, the "study" referred to is primarily a non-clinical performance evaluation comparing the proposed device to a predicate device, rather than a clinical trial or AI-specific validation study typically associated with AI/ML devices.

Here's an analysis of the provided information concerning acceptance criteria and the "study":

1. Table of Acceptance Criteria and Reported Device Performance:

The document provides a comparative table between the proposed device (PL-2100) and its primary predicate device (Flexible Video-Choledo-Cysto-Ureteroscope System, K211686, Model: PL-1000). The "acceptance criteria" are implied by the predicate device's characteristics, and the "reported device performance" are the proposed device's characteristics. The goal is to show they are "Same" or "Similar" in ways that don't raise new questions of safety or effectiveness.

ItemAcceptance Criteria (Predicate Device)Reported Device Performance (Proposed Device)Comment
Imaging Processor SystemModel: PL-1000Model: PL-2100
Indication for UseSignals from the endoscope are converted to the image and displayed on an inspection monitor.Signals from the endoscope are converted to the image and displayed on an inspection monitor.Same
Target patientsPatients who are considered suitable for the application of this product by the physicianPatients who are considered suitable for the application of this product by the physicianSame
User qualificationsPhysicians (experts approved by endoscopic medical safety administrator at each medical facility, or as defined by government/academic society). Specific training is not required.Physicians (experts approved by endoscopic medical safety administrator at each medical facility, or as defined by government/academic society). Specific training is not required.Same
Place of useMedical facilityMedical facilitySame
Power input100-240V~50/60Hz45VA100-240V~50/60Hz48VASimilar
Imaging outputHDMICVBS&DVI&S-VIDEODifferent
Light SourceLED 2WLED 2WSame
Lamp Life30,000 hrs30,000 hrsSame
ScreenN/A (implied external display)10.1 inchesDifferent
USB Flash MemoryUSB 2.0 (Max:128G; FAT32)USB 2.0 (Max:128G; FAT32)Same
illumination≥2000LUX≥2000LUXSame
Resolutiond = 10mm, MTF50 value of SFR at center not lower than 200lw/pH, MTF50 value at 70% FOV not lower than 80% of center position.d = 10mm, MTF50 value of SFR at center not lower than 200lw/pH, MTF50 value at 70% FOV not lower than 80% of center position.Same
Field of View110°±10%110°±10%Same
Observe the depth of field5-50mm, MTF50 ≥ 200 (LW/PH)5-50mm, MTF50 ≥ 200 (LW/PH)Same
Direction of View0°±2°0°±2°Same
TV distortion< 15%< 15%Same
SNR and Dynamic RangeSNR≥10, DR≥90SNR≥10, DR≥90Same
Color reproducibility testΔE≤23, ΔC≤23ΔE≤23, ΔC≤23Same
ShadingLightness non-uniformity DL≥70, Luminance non-uniformity Dγ≥95Lightness non-uniformity DL≥70, Luminance non-uniformity Dγ≥95Same

2. Sample Size Used for the Test Set and Data Provenance:

The document explicitly states: "No Clinical Study is included in this submission." This means there isn't a "test set" in the sense of patient data used for algorithm validation. The evaluation is primarily based on non-clinical engineering and performance testing.

3. Number of Experts Used to Establish Ground Truth and Qualifications:

Not applicable, as no clinical study or expert-adjudicated ground truth dataset was used for device efficacy or performance evaluation in this 510(k) submission.

4. Adjudication Method for the Test Set:

Not applicable, as no clinical study or expert adjudication process for a test set was performed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

Not applicable. This device is an endoscope system, not an AI/ML-assisted diagnostic device that would typically undergo an MRMC study to evaluate human reader improvement with AI assistance.

6. Standalone Algorithm Performance:

Not applicable. The device is a hardware system (endoscope and imaging processor), not a standalone algorithm.

7. Type of Ground Truth Used:

For the non-clinical performance and safety evaluations, the "ground truth" is established by international and national standards (e.g., IEC 60601-1, IEC 60601-1-2, ISO 10993) and the specified performance characteristics of the predicate device. The proposed device's performance is measured against these established engineering and safety benchmarks.

8. Sample Size for the Training Set:

Not applicable. This is not an AI/ML device, so there is no "training set."

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no training set for an AI/ML algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

October 22, 2024

Shanghai SeeGen Photoelectric Technology Co., Ltd. Yihua Ma RA Supervisor 3 Floor, Building No.1, 4299 JinDu Road, Minhang District Shanghai. 201108 CHINA

Re: K241532

Trade/Device Name: Flexible Video-Choledo-Cysto-Ureteroscope System (PL-2100) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FET, KQM Received: September 11, 2024

Dear Yihua Ma:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

{2}------------------------------------------------

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark R. Kreitz -S

for Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K241532

Device Name

Flexible Video-Choledo-Cysto-Ureteroscope System (PL-2100)

Indications for Use (Describe)

The Flexible Video-Choledo-Cysto-Ureteroscope System is indicated for endoscopic examination in the urinary tract and can be used percutaneously to examine the interior of the kidney, and using additional accessories, to perform various diagnostic and therapeutic procedures. The Flexible Video-Choledo-Cysto-Ureteroscope System is also indicated for the examination of bile ducts surgically, and using additional accessories, to perform various diagnostic and therapeutic procedures during cholecystectomy.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

5.1 Submitter

Submitted by:Shanghai SeeGen Photoelectric Technology Co., Ltd.
Address: 3 Floor, Building No.1, 4299 JinDu Road, MinhangDistrict, Shanghai, China
ContactPerson:Yihua MaRA Supervisor
Shanghai SeeGen Photoelectric Technology Co., Ltd.Address: 3 Floor, Building No.1, 4299 JinDu Road, MinhangDistrict, Shanghai, ChinaPhone: 0086-18616909737Email: mayihua@seegen.com.cn
DatePrepared:May 30, 2024

5.2 Device

Device Name:Flexible Video-Choledo-Cysto-UreteroscopeSystem
Common Name:Choledochoscope and Accessories, Flexible/Rigid;Camera, Surgical and Accessories
Regulatory Class:Class II
Regulation Number:21 CFR 876.1500
Classification Name:Endoscopic Video Imaging System/Component, GU
Product Code:KQM,FET

5.3 Predicate Device

Device Name:Flexible Video-Choledo-Cysto-UreteroscopeSystem, K211686
Common Name:Choledochoscope and Accessories, Flexible/Rigid;Camera, Surgical and Accessories;LED light Source
Regulatory Class:Class II
Regulation Number:21 CFR 876.1500
Regulation Name:Endoscope and Accessories
Product Code:FGB, FBN, FET, FAJ, FGA

{5}------------------------------------------------

5.4 Device Description

Imaging Processor System (Including Light Source) is composed of lighting system, image processing board. The lighting system provides the light source for the endoscope probe at the back end. The image processing board receives electronic signals from the front-end camera module and processes them, and finally transmits them to the display through the video interface. Flexible Video-Choledo-Cysto-Ureteroscope is a kind of medical electronic optical instrument, also known as optical camera, which can enter into the human bladder, ureter, biliary and pancreatic duct for observation and diagnosis. The operator delivers the optical camera system to the site of diagnosis and treatment by means of a mechanical part with a flexible insertion tube and a system of bends. The product is equipped with tiny size digital imaging parts -- photoelectric sensors "CMOS", on which the objects in human cavity will be transferred though lens optical system, and converts light signals into electrical signals. The electrical signal will be transferred to Imaging Processor System (Including Light Source) and display images on it's monitor output for doctor observation and diagnosis.

5.5 Indication for Use:

Flexible Video-Choledo-Cysto-Ureteroscope System is indicated for The endoscopic examination in the urinary tract and can be used percutaneously to examine the interior of the kidney, and using additional accessories, to perform various diagnostic and therapeutic procedures. The Flexible Video-Choledo-Cysto-Ureteroscope System is also indicated for the examination of bile ducts, and using additional accessories, to perform various diagnostic and therapeutic procedures during cholecystectomy.

5.6 Substantial Equivalence and Technological Characteristics

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for SEEGEN. The logo is red and black, with the word "SEEGEN" in large, bold letters. Below the word "SEEGEN" is the phrase "vision borderless" in smaller letters. To the right of the word "SEEGEN" are some Chinese characters.

ItemImaging Processor System(Including Light Source)(Proposed Device),Model:PL-2100Imaging Processor System(Including Light Source),K211686(Primary Predicate Device), Model:PL-1000Comment
Indication for UseSignals from the endoscope are converted to the image anddisplayed on an inspection monitor.Signals from the endoscope are converted to the imageand displayed on an inspection monitor.Same
Target patients:Patients who are considered suitable for the application ofthis product by the physicianPatients who are considered suitable for the application ofthis product by the physicianSame
UserqualificationsPhysicians (experts who have been approved by theendoscopic medical safety administrator at each medicalfacility. If the eligibility requirements are defined by anofficial body such as a government entity and/or anacademic society, follow such requirements.) Specifictraining to use this product is not required.Physicians (experts who have been approved by theendoscopic medical safety administrator at each medicalfacility. If the eligibility requirements are defined by anofficial body such as a government entity and/or anacademic society, follow such requirements.) Specifictraining to use this product is not required.Same
Place of useMedical facilityMedical facilitySame
Power input100-240V~50/60Hz48VA100-240V~50/60Hz45VASimilar
Imaging outputCVBS&DVI&S-VIDEOHDMIDifferent
Light SourceLED 2WLED 2WSame
Lamp Life30,000 hrs30,000 hrsSame
Screen10.1 inchesN/ADifferent
USB FlashMemoryUSB 2.0(Max:128G;FAT32)USB 2.0(Max:128G;FAT32)Same

{7}------------------------------------------------

K241532
Page 4 of 5

Image: SEEEEN 视介光电
vision borderless
illumination≥2000LUX≥2000LUXSame
Resolutiond = 10mm, the MTF50 value of the SFR at the center part shall not be lower than 200lw / pH, and the MTF50 value at 70% of the field of view position shall not be lower than the MTF50 value at 80% of the center position.d = 10mm, the MTF50 value of the SFR at the center part shall not be lower than 200lw / pH, and the MTF50 value at 70% of the field of view position shall not be lower than the MTF50 value at 80% of the center position.Same
Filed ofView110°±10%110°±10%Same
Observethe depthof field5-50mmMTF50 ≥ 200 (LW / PH)5-50mmMTF50 ≥ 200 (LW / PH)Same
Direction of View0°±2°0°±2°Same
TVdistortion< 15%< 15%Same
SNR andDynamicRangeSNR≥10DR≥90SNR≥10DR≥90Same
Colorreducibilityy testΔE≤23ΔC≤23ΔE≤23ΔC≤23Same
ShadingLightness non-uniformity DL≥70Luminance non-uniformity Dγ≥95Lightness non-uniformity DL≥70Luminance non-uniformity Dγ≥95Same

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for SEEGEN, a company specializing in vision technology. The logo features the company name in a stylized, bold font, with the tagline "vision borderless" underneath. To the right of the company name are Chinese characters, likely representing the company's name or a related concept in Chinese. The overall design is modern and sleek, reflecting the company's focus on innovation and cutting-edge technology in the field of vision.

5.7 Substantial Equivalence

The new image processor,(PL-2100), are used as predicate device compared to proposed device Flexible Video-Choledochoscope System manufactured by Shanghai SeeGen Photoelectric Technology Co., Ltd.

5.8 Non-clinical Performance Data

The Flexible Video-Choledo-Cysto-Ureteroscope System has been successfully tested for its functions. Safety testing was performed including electrical safety IEC 60601-1 and IEC 60601-2-18, electromagnetic compatibility per IEC 60601-1-2 and biocompatibility of the patient contact materials per ISO 10993. Additional validations were conducted for the sterilization process, EO residual, transportation and Photobiological safety, optical and colour performance test.

5.9 Clinical Test Data

No Clinical Study is included in this submission.

5.10 Conclusion

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, Based on the information provided in this premarket notification, Shanghai SeeGen Photoelectric Technology Co., Ltd. has demonstrated that proposed device Flexible Video-Choledo-Cysto-Ureteroscope System is substantially equivalent to Boston Scientific Corporation's currently marketed Flexible Video-Choledo-Cysto-Ureteroscope System, K211686.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.