K Number
K241523

Validate with FDA (Live)

Date Cleared
2024-08-27

(90 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The injeTAK Adjustable Tip Needles are intended to be used to deliver a variety of legally marketed drugs into tissues or structures during cystoscopic procedures. They are provided sterile for single use only.

Device Description

The injeTAK Adjustable Tip Needles (injeTAK Needles) are intended to be used to deliver a variety of legally marketed drugs into tissues of structures during cystoscopic procedures. They are provided sterile for single use only.
The needles will contact the bladder tissue and urine fluid in the patient. The typical injection procedure lasts 10-20 minutes; each injection spot contact lasts approximately 5 seconds but may vary slightly depending on the skill of surgeon.
The models of the injeTAK needles are used with a rigid or flexible cystoscope during cystoscopic procedures and differ in length of the needle dependent upon the type of procedure. The intended use does not vary based on the model number or difference in measurement specifications.
The injeTAK Needle comprises of a needle cannula, contrast mark, needle sheath, handle, needle length dial, hub/luer connector, lock-pin, and a seal ring located inside the handle.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study proving device performance:

This document is a 510(k) summary for a medical device called "injeTAK Adjustable Tip Needle (DIS199; DIS201)". The FDA's substantial equivalence determination process for 510(k) clearances typically involves demonstrating that a new device is as safe and effective as a legally marketed predicate device. This often involves non-clinical testing to compare the new device's performance against its design specifications and, by extension, against the predicate.

Given this context, the document primarily focuses on non-clinical testing and comparisons to a predicate device, rather than a clinical study evaluating an AI algorithm or human-in-the-loop performance.


Here's the information extracted and organized based on your request:

1. Table of Acceptance Criteria and Reported Device Performance

The document describes various non-clinical tests performed to ensure the device meets its design specifications and performs as intended. Specific, quantifiable acceptance criteria are not explicitly listed in a table format within the provided text, but the general outcome for each test is stated. The "acceptance criteria" are implied by the statement "passed all requirements" or "met the acceptance criteria."

Acceptance Criteria Category (Implied)Reported Device Performance
Biocompatibility"The performed biocompatibility tests show that the injeTAK needles are safe and perform as intended." "The materials of all components... have not changed and stay as exact the same as existing G1 needle that the device biocompatibility will be well maintained for injeTAK G2 versions."
Packaging Integrity"All test results passed and met the acceptance criteria." "The device packaging performance and integrity will be maintained for injeTAK G2."
Sterilization Efficacy"The testing performed confirms that sterilization output requirements were met."
Functional & Performance"The test results passed the acceptance criteria." "Overall performance and results were positive and meet the related specifications."
Accelerated Aging/Shelf Life"The essential functionality and performance of injeTAK G2 version (DIS199 & DIS201) passed all requirements after accelerated aging and are therefore acceptable for the three year shelf life." This implicitly covers flow rate and needle luer leak tests.
Critical Dimensions"The test was intended to verify if the device critical dimensions... meet the design specifications... The test results passed the acceptance criteria."

2. Sample size used for the test set and the data provenance

The document does not specify exact sample sizes for the non-clinical tests conducted (e.g., how many needles were tested for biocompatibility, packaging, or functional performance). It mentions "the product's packaging testing" and "the testing performed," implying a set of samples were used for each test.

Data Provenance: The tests are described as non-clinical and in-house/manufacturer-conducted tests to verify product design specifications. The data is internal to Laborie Medical Technologies Corp.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This device is an invasive medical instrument (needle) and the evaluation presented is a non-clinical device verification and validation summary, not an AI/software product requiring expert ground truth for interpretation of images or other data. The "ground truth" here is the device's adherence to its engineering specifications and relevant ISO standards.

4. Adjudication method for the test set

Not applicable. This is not a study involving human assessment of data that would require an adjudication method. The "adjudication" in this context would be the internal quality and regulatory review process of the manufacturer.

5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. This is not an AI-assisted diagnostic or predictive device, and no MRMC study was conducted or is relevant to this submission.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

No. This is a hardware device (needle), not an algorithm.

7. The type of ground truth used

The "ground truth" for this medical device's performance evaluation is based on:

  • Engineering Design Specifications: The device must meet its predefined critical dimensions, functional parameters (e.g., flow rate, luer leak), and material properties.
  • International Standards: Adherence to relevant ISO standards for biocompatibility (e.g., ISO 10993 series), sterilization (e.g., ISO 11135, ISO 11137), and packaging (e.g., ISO 11607).
  • Predicate Device Performance: Demonstrating that the new device performs at least as well as the legally marketed predicate device.

8. The sample size for the training set

Not applicable. This is a hardware medical device, not a machine learning model, so there is no "training set."

9. How the ground truth for the training set was established

Not applicable. There is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

August 27, 2024

Laborie Medical Technologies Corp. Thomas Hirte Sr. Director Regulatory Affairs 180 International Drive Portsmouth, New Hampshire 03801

Re: K241523 Trade/Device Name: injeTAK Adjustable Tip Needle (DIS199; DIS201) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FBK Received: August 2, 2024

Dear Thomas Hirte:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See

{2}------------------------------------------------

the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/2/Picture/3 description: The image shows the name "Mark R. Kreitz -S" in a bold, sans-serif font. The text is black against a white background. The letters are evenly spaced and clearly legible.

for Mark J. Antonino, M.S Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K241523

Device Name

injeTAK Adjustable Tip Needle (DIS199; DIS201)

Indications for Use (Describe)

The injeTAK Adjustable Tip Needles are intended to be used to deliver a variety of legally marketed drugs into tissues or structures during cystoscopic procedures. They are provided sterile for single use only.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K241523
510(k) #: K241523510(k) Summary Page 1 of 2Prepared on: 2024-08-01
Contact Details21 CFR 807.92(a)(1)
Applicant NameLaborie Medical Technologies Corp.
Applicant Address180 International Drive Portsmouth NH 03801 United States
Applicant Contact Telephone978-277-6432
Applicant ContactMr. Thomas Hirte
Applicant Contact Emailthirte@laborie.com
Device Name21 CFR 807.92(a)(2)
Device Trade NameinjeTAK Adjustable Tip Needle (DIS199; DIS201)
Common NameEndoscope and accessories
Classification NameEndoscopic Injection Needle, Gastroenterology-Urology
Regulation Number876.1500
Product Code(s)FBK, N/A
Legally Marketed Predicate Devices21 CFR 807.92(a)(3)
---------------------------------------------------------
Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K090830injeTAK Adjustable Tip NeedleFBK
Device Description Summary21 CFR 807.92(a)(4)
-------------------------------------------------

The injeTAK Adjustable Tip Needles (injeTAK Needles) are intended to be used to deliver a variety of legally marketed drugs into tissues
of structures during cystoscopic procedures. They are provided sterile for single use only.
The needles will contact the bladder tissue and urine fluid in the patient. The typical injection procedure
lasts 10-20 minutes; each injection spot contact lasts approximately 5 seconds but may vary slightly
depending on the skill of surgeon.
The models of the injeTAK needles are used with a rigid or flexible cystoscope during cystoscopic procedures and differ in length of the
needle dependent upon the type of procedure. The intended use
does not vary based on the model number or difference in measurement specifications.
The injeTAK Needle comprises of a needle cannula, contrast mark, needle sheath, handle, needle length
dial, hub/luer connector, lock-pin, and a seal ring located inside the handle.

Intended Use/Indications for Use21 CFR 807.92(a)(5)
-------------------------------------------------------

The injeTAK Adjustable Tip Needles are intended to be used to deliver a variety of legally marketed drugs into tissues or structures
during cystoscopic procedures. They are provided sterile for single use only.

Indications for Use Comparison21 CFR 807.92(a)(5)
-----------------------------------------------------

The indications for use of the proposed device remain the same as the predicate device

Technological Comparison21 CFR 807.92(a)(6)
-----------------------------------------------

{5}------------------------------------------------

K241523

The basic technology for the subject device and the predicate device (injeTAK Adjustable Tip Needles) is the same. The injeTAK Needle comprises of a needle cannula, contrast mark, needle length dial, hub/luer connector, lock-pin, and a seal ring located inside the handle. The injeTAK Needles do not include software or a dispensing device (i.e., syringe). The sheath is attached to housing and the needle cannula is attached to a needle hub. Five depths are listed on the needle handle (0, 2, 4, and 5), with each number corresponding to the needle tip length in millimeters. For example, when the needle lock-pin is set at the number 3, the maximum needle tip penetration depth is 3mm. There is at the distal end of the plastic sheath which provides a better view of injection spot.

Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

Non-clinical tests were conducted to verify that the subject device met all design specifications. The substantially equivalent to the predicate. No clinical testing was included in this submission.

Biocompatibility:

No additional testing is required since the standards do not affect the results of the Cytotoxicity & Pyrogenic tests, Sensitization tests or Intracutaneous Reactivity tests previously conducted on the inje TAK needles. The performed biocompatibility tests show that the injeTAK needles are safe and perform as intended. The materials of all components in the device have not changed and stay as exact the same as existing G1 needle that the device biocompatibility will be well maintained for injeTAK G2 versions (DIS 199 & DIS 201).

Packaging:

InjeTAK Packaging Design Verification was completed on the packaging change to provide results of the product's packaging testing against released requirements to ensure the integrity of primary packaging mance at the final point of delivery after enduring the worst-case transportation conditions.

lnjeTAK Adjustable Tip Needles meet the requirements of variously corresponding international standards after the Transportation Simulation Test by functional testing, package integrity testing. All test results passed and met the acceptance criteria.

Sterilization:

The packaging change required the sterilization process to be revalidated in order to maintain the sterilization cycle parameters specified in the original cycle parameters/design. The testing performed confirms that sterilization output requirements were met.

Functional & Performance Testing:

The test was intended to verify if the device critical dimensions, function/performance meet the design specifications stated in the design files. The test results passed the acceptance criteria. Overall performance and results were positive and meet the related specifications.

The packaging material and dimension have not changes and stay as version G1. The overall footprint and weight are equivalent. Therefore, it can be concluded that the device packaging performance and integrity will be maintained for injeTAK G2, and revalidation tests are not needed.

Accelerated Aging and Shelf Life:

No additional accelerated aged tests were needed for the G2 subject device versions (DIS199 & DIS201) of the InjeTAK Needles. Since there was no change in the material, biocompatibility, packaging, aging and/or sterilization process for the G2 needles versions, the accelerated test reports conducted on the G predicate versions (DIS198 & DIS201) of the InjeTAK needles are applicable to the G2 versions as well.

The impact of aging (i.e. shelf life) on functionality and performance of the leject device versions (DIS199 & DIS201) remains unchanged. The three main functions – flow rate, needle leur leek test were evaluated. The essential functionality and performance of hjeTAK G2 version (DIS199 & DIS201) passed all requirements after accelerated aging and are therefore acceptable for the three year shelf life

Verification and Validation testing were performed on the InjeTAK Needles will perform as intended. All testing passed. Details are provided in the verification and validation section of this submission

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.