K Number
K241506

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-06-27

(30 days)

Product Code
Regulation Number
884.5160
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Annabella Double Breast Pump is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The Annabella Double Breast Pump is intended for a single user.

Device Description

The Annabella Double Breast Pump is a double, electric breast pump system intended to express and collect milk from the breasts of lactating women. Each pump is comprised of a vacuum unit including tubing (connected to a hybrid cable with a "Y" connector) and a massage unit (tongue mechanism) that mimics the baby's sucking motions. The Y hybrid cable delivers the vacuum from the vacuum unit through the Y splitter to both massage units. The user employs buttons on the vacuum unit to control the vacuum intensity levels, suction rate as well as the massage unit speed and location on the breast. The device is supplied with a reusable and adjustable breast shield, a bottle, bottle lids and a charger. The Annabella is a rechargeable Li-ion battery operated device that contains software. The device cannot be operated while connected to the mains AC Power core.

The battery compartment is at the bottom of the pump motor unit and is covered by a battery door. The runtime of the removable lithium-ion battery is influenced by the number and duration of pumping sessions and lasts usually for one day.

The device is to be used in the home.

AI/ML Overview

The provided text describes the Annabella Double Breast Pump (K241506) and its comparison to a predicate device (Annabella Breast Pump K230672). It outlines performance data from non-clinical tests but does not specify acceptance criteria in a quantitative table or directly present reported device performance against such criteria. Instead, it lists the types of tests performed to demonstrate substantial equivalence to the predicate device.

Given the information, here's an attempt to answer your questions based on the provided text.

1. A table of acceptance criteria and the reported device performance

The document does not provide a specific table of acceptance criteria with quantitative values and corresponding reported device performance. It only states that "Non-clinical tests were conducted to verify that the subject device met all design specifications to be considered substantially equivalent to the predicate device."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The text describes only non-clinical performance testing, which typically involves testing of prototype devices or components rather than human subjects. Therefore, the concepts of "sample size used for the test set" and "data provenance (country of origin, retrospective/prospective)" as they relate to human subjects or clinical data are not applicable here. The tests mentioned include:

  • Electrical Safety Testing
  • Electromagnetic compatibility testing
  • Software Verification
  • Biocompatibility (leveraged from predicate device testing)
  • Vacuum level/suction strength testing
  • Battery performance testing
  • Battery status indicator testing
  • Device use life testing

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This question is not applicable as the study involves hardware and software performance testing of a medical device, not a diagnostic or AI-driven system requiring expert ground truth for interpretation of medical images or data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This question is not applicable for the same reasons as #3.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC study was conducted. The device is a breast pump and does not involve AI assistance for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not directly applicable in the context of a breast pump. However, the "Software Verification" was performed according to IEC 62304 and FDA guidance, indicating an assessment of the software's standalone performance and safety within the device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the non-clinical tests mentioned, the "ground truth" would be established by:

  • Industry standards and regulations: e.g., IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, IEC 62304.
  • Design specifications: The device was tested to verify it met "all design specifications."
  • Predicate device characteristics: The subject device's performance (e.g., suction levels, cycle speeds) was compared to that of the predicate device.

8. The sample size for the training set

This question is not applicable as this is not an AI/ML-driven diagnostic device that requires a training set. The software verification focuses on life cycle processes and compliance.

9. How the ground truth for the training set was established

This question is not applicable for the same reasons as #8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

June 27, 2024

Annabella Tech % Dalia Dickman Regulatory Consultant Dalia Dickman Consulting Caesarea ISRAEL

Re: K241506

Trade/Device Name: Annabella Double Breast Pump Regulation Number: 21 CFR 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: Mav 23, 2024 Received: May 28, 2024

Dear Dalia Dickman:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

K241506 - Dalia Dickman

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

{2}------------------------------------------------

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Monica D. Garcia -S

Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K241506

Device Name Annabella Double Breast Pump

Indications for Use (Describe)

The Annabella Double Breast Pump is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The Annabella Double Breast Pump is intended for a single user.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY - K241506

Annabella Tech Ltd's Annabella Double Breast Pump

Submitter

Annabella Tech Ltd. 23/5 Hataas KFAR SABA, 4442525 Israel. +972-9-8845513

Contact Person: Dalia Dickman, PhD. Date Prepared: May 23, 2023

Device Name: Annabella Double Breast Pump

Common or Usual Name: Powered breast pump

Regulation Number: 21 CFR 884.5160

Regulation Name: Powered breast pump

Regulatory Class: II

Product Code: HGX (Powered, Breast, Pump)

Predicate Device

Annabella Breast Pump (K230672)

The predicate device has not been subject to a design-related recall.

Device Description

The Annabella Double Breast Pump is a double, electric breast pump system intended to express and collect milk from the breasts of lactating women. Each pump is comprised of a vacuum unit including tubing (connected to a hybrid cable with a "Y" connector) and a massage unit (tongue mechanism) that mimics the baby's sucking motions. The Y hybrid cable delivers the vacuum from the vacuum unit through the Y splitter to both massage units. The user employs buttons on the vacuum unit to control the vacuum intensity levels, suction rate as well as the massage unit speed and location on the breast. The device is supplied with a reusable and adjustable breast shield, a bottle, bottle lids and a charger. The Annabella is a rechargeable Li-ion battery operated device that contains software. The device cannot be operated while connected to the mains AC Power core.

The battery compartment is at the bottom of the pump motor unit and is covered by a battery door. The runtime of the removable lithium-ion battery is influenced by the number and duration of pumping sessions and lasts usually for one day.

The device is to be used in the home.

{5}------------------------------------------------

Intended Use / Indications for Use

The Annabella Double Breast Pump is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The Annabella Double Breast Pump is intended for a single user.

Predicate Device Comparison

The table below compares the intended use and technological characteristics of the subject and predicate device.

Subject Device:Annabella DoubleBreast PumpPredicate Device:Annabella BreastPump (K230672)Discussion
Intended Use/Indication for useThe Annabella DoubleBreast Pump is apowered breast pumpto be used by lactatingwomen to express andcollect milk from theirbreasts. TheAnnabella is intendedfor a single user.The Annabella is apowered breast pump tobe used by lactatingwomen to express andcollect milk from theirbreasts. The Annabellais intended for a singleuser.Identical
Intended UseExpress and collectmilkExpress and collect milkIdentical
Single User DeviceYesYesIdentical
Environment of UseHomeHomeIdentical
Over the CounterYesYesIdentical
UserInterfaceHardware Interfaces
User ControlOn-off switchVacuum/Cycle-adjustment control.Massage adjustmentcontrol- speed andlocation relative to thebreast center.On-off switchVacuum/Cycle-adjustment control.Massage adjustmentcontrol- speed andlocation relative to thebreast center.Identical
Visual Indicator7 Segment Displayand LED Indications7 Segment Display andLED IndicationsIdentical
Pumping optionsSingle or doubleSingleDifferent – Thesubject devicehas two singleand pumpingoptions. Thepumpingmechanism is thesame for bothdevices.
Media SeparationYesYesIdentical
Specifications
Power SupplyLi-Ion rechargeablebatteryLi-Ion rechargeablebatteryIdentical
Suction LevelsStimulation: 30 - 122Stimulation: 35 - 130Different; there
Subject Device:Annabella Double_Breast PumpPredicate Device:Annabella BreastPump (K230672)Discussion
mmHgmmHgare minor
Expression: 39 -250mmHgExpression: 33 -250mmHgdifferences insuction levels.i
Cycles per SecondStimulation: 1.66- 2.0Expression: 0.7-1.44Stimulation: 1.66- 2.0Expression: 0.8-1.44Different; there are minordifferences incycle speeds.
Maximum vacuum250 mmHg250 mmHgIdentical
Suction setting99Identical
Adjustable SuctionLevelsYesYesIdentical
Let-Down ButtonYesYesIdentical
Massage(TongueMechanism)4 cams rotating at atypical angular speedof 0-60rpm, 9 levelsfor speed.9 levels for location onthe breast (distancefrom the breast center)4 cams rotating at atypical angular speed of0-60rpm, 9 levels forspeed.9 levels for location onthe breast (distancefrom the breast center)Identical
Cycling ControlMechanismMicrocontrollerMicrocontrollerIdentical
Back FlowProtectionYesYesIdentical
2-phase expressionYesYesIdentical

{6}------------------------------------------------

{7}------------------------------------------------

The indications for use of the subject and predicate devices are identical and they have the same intended use (i.e., the collection of breast milk from the breasts of lactating women).

The subject and predicate devices have different technological features, with minor differences in the pumping options, cycle speeds, and vacuum strengths. The different technological characteristics of the subject device, as compared to the predicate device, do not raise different questions of safety and effectiveness.

Performance Data

Non-clinical tests were conducted to verify that the subject device met all design specifications to be considered substantially equivalent to the predicate device.

Electrical Safety and Electromagnetic Compatibility:

  • Electrical Safety Testing was performed in accordance with IEC 60601-1:2005 (3rd Edition)
  • Electrical safety testing for use in home was performed in accordance with IEC 60601-1-11:2015. General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
  • Electromagnetic compatibility testing was performed in accordance with IEC — 60601-1-2:2021. Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.

Software Verification:

Software Verification was performed in accordance with IEC 62304:2015 Ed. 1. 1: Medical Device Software - Software Life Cycle Processes and according to the FDA Guidance document, " Content of Premarket Submissions for Device Software Functions," dated June, 2023.

Biocompatibility:

The materials of all parts that are in contact with the user remain the same for the Annabella Double Breast Pump as was tested for the Annabella Breast Pump, therefore, testing from the predicate device was leveraged to support the biocompatibility of the subject device.

Non-clinical Performance Testing:

  • Vacuum level/suction strength of subject devices was tested for each mode/ cycle.
  • Battery performance testing was conducted to demonstrate that the battery remains functional during its stated use-life.
  • Battery status indicator testing was conducted to demonstrate that the battery status indicator remains functional during its stated battery life.
  • Device use life to demonstrate that the device maintains its specifications throughout its proposed use life.

Conclusions

The results of the performance testing described above demonstrate that the Annabella

{8}------------------------------------------------

Double Breast Pump is as safe and effective as the predicate device and supports a determination of substantial equivalence.

§ 884.5160 Powered breast pump.

(a)
Identification. A powered breast pump in an electrically powered suction device used to express milk from the breast.(b)
Classification. Class II (performance standards).