Intended Use

Shoreline ACS Interbody System: The Shoreline ACS Interbody System with NanoMetalene® surface technology are interbody fusion devices intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) for multiple contiguous levels. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. These devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone. When used as a standalone system, the Shoreline ACS Interbody System, which includes the TruProfile plates and Noprofile spacer, is intended to be used as an adjunct to spinal fusion procedures at multiple contiguous levels of the cervical spine (C2-T1) and must be used with bone screw fixation and locking covers. The Shoreline ACS interbody spacers with > 20° lordosis are intended to be used with supplemental fixation. When the Shoreline ACS Interbody (excluding the No-profile spacer) is used with supplemental fixation, such as anterior cervical plates, the Shoreline ACS Interdody System is intended to be used as an adjunct to spinal fusion procedures at multiple contiguous levels of the cervical spine (C2-T1).

Shoreline RT Interbody System: The Shoreline RT Interbody System with NanoMetalene® surface technology are interbody fusion devices intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) for multiple contiguous levels. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. These devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone. When used as a standalone system, the Shoreline RT Interbody System, which includes the TruProfile plates, is intended to be used as an adjunct to spinal fusion procedures at multiple contiguous levels of the cervical spine (C2-T1) and must be used with bone screw fixation and locking covers. The Shoreline RT interbody spacers with ≥ 20° lordosis are intended to be used with supplemental fixation. When the Shoreline RT Interbody System is used with supplemental fixation, such as anterior cervical plates, the Shoreline RT Interbody System is intended to be used as an adjunct to spinal fusion procedures at multiple contiguous levels of the cervical spine (C2-T1).

Regatta Lateral System: Interbody Device (IBD) Implants (i.e., interbody implants used alone): The Regatta Lateral System with NanoMetalene® surface technology is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD, defined as back pain of discogenic origin, with degeneration of the disc confirmed by history and radiographic studies). It is intended for use at ether one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of DDD with up to Grade 1 spondylolisthesis at the involved level(s). The interior of the interbody spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The Regatta Lateral System is intended for use with supplemental fixation. TruProfile Interbody Implants: The Regatta Lateral System assembled with the TruProfile Lateral Plate, when used with Screws, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The Regatta Lateral System assembled with the 1- hole TruProfile Lateral Plate, when used with Screws, is intended for use with supplemental fixation.

Reef L Interbody System: Interbody Device (IBD) Implants (i.e., interbody implants used alone): The Reef L Interbody System with NanoMetalene® surface technology is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD, defined as back pain of discogenic origin, with degeneration of the disc confirmed by history and radiographic studies). It is intended for use at evel or two contiguous levels in the lumbar spine, from L1 to S1, for the treatment of DDD with up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogence bone graft composed of cancellous, cortical, and/or corticocancellous bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The Reef L Interbody System is intended for use as an adjunct to fusion in the thoracolumbar spine from T1 to T12 and at the thoracolumbar junction (T12-L1) for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two contiguous levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenc origin with degeneration of the disc confirmed by history and radiographic studies. The Reef L Interbody System can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. The Reef L Interbody System is intended for use with supplemental fixation. Hyperlordotic sizes (greater than 200) must be used with legally marketed anterior supplemental fixation. TruProfile Interbody Implants (i.e., interbody implants used with TruProfile Plates): The Reef L Interbody System assembled with the TruProfile Lateral Plate, when used with Screws, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L1 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of nonoperative treatment prior to being treated with the device. The interded to be used with autogenous bone graft, allogeneic bone graft composed of cancellous, corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion. The Reef L Interoody System assembled with the TruProfile Lateral Plate, when used with Screws, can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. The Reef L Interbody System assembled with the 1-Hole TruProfile Lateral Plate, when used with a single Screw, is intended for use with additional legally marketed anterior or posterior supplemental fixation.

Reef TO/TA System: When used as an intervertebral body fusion device, the Reef TO/TA System with NanoMetalene® surface technology is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenc origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had six months of non-operative treatment. The device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone and supplemental fixation.

Meridian Interbody System: Interbody Device (IBD) Implants (i.e., interbody implants used alone): The Meridian System with NanoMetalene® surface technology when used with or without a Spin Plate, is indicated for use as an adjunct to fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients should be skeletally mature and have had at least six (6) months of non-operative treatment. The interded to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The Meridian Interbody is intended for use with supplemental fixation. No-Profile Interbody Implants with Screws: The Meridian System 2-Hole No-profile Interbody, when used with Screws and with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenc origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed cancellous, cortical, and/or corticocancellous bone. The Meridian System 2-Hole No-profile Interbody must be used with the maximum number of screws allowed. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation. The Meridian System 4-Hole No-profile Interbody, when used with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenc origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The Meridian System 4-Hole No-profile Interbody may be used as a standalone device only when at least two Screws are inserted with one inferior and one superior screw trajectory. If the surgeon chooses to use the 4-Hole No-profile Interbody with fewer than two Screws inserted with one inferior and one superior screw trajectory, then additional supplemental fixation system must be used. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation. No-profile Interbody Implants with Inline Fixation Anchors: The Meridian System No-profile Interbody, when used with Inline Fixation Anchors and a No-profile Locking Cover, is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interded to be used with autogenous bone graft and/or allogence bone graft composed of cancellous, corticocancellous bone. The Meridian System No-profile Interbody with Inline Fixation Anchors is intended for use with supplemental fixation. TruProfile Interbody Implants: The Meridian System TruProfile Interbody assembled with the Anterior Plate, when used with Screws, an Anterior Plate Locking Cover, and with or without a Spin Plate, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1.

WaveForm C Interbody System: The WaveForm C Interbody System are interbody fusion devices intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) for multiple contiguous levels. DDD is defined as discogenc pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. These devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone. When used as a standalone system, the WaveForm C Interbody System, which includes the 2,3,4-hole TruProfile plates and 2-hole No-profile interfixated spacer, is intended to be used as an adjunct to spinal fusion procedures at multiple contiguous levels of the cervical spine (C2-T1) and must be used with bone screw fixation and locking covers. When the WaveForm C Interbody System (excluding 2-hole No-profile interfixated spacer) is used with supplemental fixation, such as anterior cervical plates, the WaveForm C Interbody System is intended to be used as an adjunct to spinal fusion procedures at multiple contiguous levels of the cervical spine (C2-T1).

WaveForm L Interbody System: Interbody Device (IBD) Implants (i.e., interbody implants used alone): The WaveForm L Interbody System is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD, defined as back pain of discogenic origin, with degeneration of the disc confirmed by history and radiographic studies). It is intended for use at either one level or two contiguous levels in the lumbar spine, from L1 to S1, for the treatment of DDD with up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The WaveForm L Interbody System is intended for use as an adjunct to fusion in the thoracolumbar spine from T1 to T12 and at the thoracolumbar junction (T12-L1) for the treatment of symptomatic disc degenerative spondylolisthesis at one or two contiguous levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The WaveForm L Interbody System can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. The WaveForm L Interbody System is intended for use with supplemental fixation. Hyperlordotic sizes (greater than 20°) must be used with legally marketed anterior supplemental fixation. TruProfile Interbody Implants (i.e., interbody implants used with TruProfile Plates): The WaveForm L Interbody System assembled with the TruProfile Lateral Plate, when used with Screws, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L1 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of nonoperative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The WaveForm L Interbody System assembled with the TruProfile Lateral Plate, when used with Screws, can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. The WaveForm L Interbody System assembled with the 1-Hole TruProfile Lateral Plate, when used with a single Screw, is intended for use with additional legally marketed anterior or posterior supplemental fixation.

WaveForm TO Interbody System: When used as an intervertebral body fusion device, the system is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had six months of non-operative treatment. The device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone and supplemental fixation.

WaveForm TA Interbody System: When used as an intervertebral body fusion device, the system is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had six months of non-operative treatment. The device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone and supplemental fixation.

WaveForm A Interbody System: Interbody Device (IBD) Implants (i.e., interbody implants used alone): The WaveForm A System Interbody, when used with or without a Spin Plate, is indicated for use as an adjunct to fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients should be skeletally mature and have had at least six (6) months of non-operative treatment. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The WaveForm A System Interbody is intended for use with supplemental fixation. No-profile Interbody Implants with Screws: The WaveForm A System 2-Hole No-profile Interbody, when used with Screws and with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed cancellous, cortical, and/or corticocancellous bone. The WaveForm A System 2-Hole and 3-Hole No-profile Interbody must be used with the maximum number of screws allowed. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation. The WaveForm A System 4-Hole No-profile Interbody, when used with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The WaveForm A System 4-Hole No-profile Interbody may be used as a standalone device only when at least two Screws are inserted with one inferior and one superior screw trajectory. If the surgeon chooses to use the 4-Hole No-profile Interbody with fewer than two Screws inserted with one inferior and one superior screw trajectory, then additional supplemental fixation system must be used. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation. No-profile Interbody Implants with Inline Fixation Anchors: The WaveForm A System No-profile Interbody, when used with Inline Fixation Anchors and with a No-profile Locking Cover, is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interded to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The WaveForm A System No-profile Interbody with Inline Fixation Anchors is intended for use with supplemental fixation. TruProfile Interbody Implants: The WaveForm A System TruProfile Interbody assembled with the Anterior Plate, when used with Screws, an Anterior Plate Locking Cover, and with or without a Spin Plate, is a standalone indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of nonoperative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/ or corticocancellous bone. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation.

Device Description

Shoreline ACS Interbody System: The Shoreline ACS Interbody System featuring NanoMetalene surface technology consists of anterior cervical intervertebral body fusion devices manufactured from polyetheretherketone (PEEK) (per ASTM F2026) with tantalum (per ASTM F560) and/or titanium alloy (Ti-6Al-4V EU per ASTM F136) markers for radiographic visualization and NanoMetalene, which is a one-micron thick surface layer of commercially pure titanium (per ASTM F67) that provides a microscopic roughened surface with nano-scale features. The interbody spacers are offered in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy. The spacers, which include a central graft window for receiving autogenous bone graft and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone, are individually packaged and gamma sterilized. The Shoreline ACS Interbody System can be used with supplemental fixation, such as an anterior plate, or as a standalone construct in combination with non-sterile fixation implants (i.e., plates, bone screws, and locking covers) to create the TruProfile configurations. The fixation implants are manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136). The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

Shoreline RT Interbody System: The Shoreline RT Interbody System featuring NanoMetalene surface technology consists of anterior cervical intervertebral body fusion devices manufactured from polyetheretherketone (PEEK) (per ASTM F2026) with tantalum (per ASTM F560) and/or titanium alloy (Ti-6Al-4V ELI per ASTM F136) markers for radiographic visualization and NanoMetalene, which is a one-micron thick surface layer of commercially pure titanium (per ASTM F67) that provides a microscopic roughened surface with nano-scale features. The interbody spacers are offered in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy. The spacers, which include a central graft window for receiving autogenous bone graft and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone, are individually packaged and gamma sterilized. The Shoreline RT Interbody System can be used with supplemental fixation, such as an anterior plate, or as a standalone construct in combination with the Shoreline ACS Interbody System nonsterile fixation implants (i.e., plates, bone screws, and locking covers) to create the TruProfile configuration. The fixation implants are manufactured from titanium alloy (Ti-6AI-4V ELI per ASTM F136). The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

Regatta Lateral System: The Regatta Lateral System featuring NanoMetalene surface technology consists of lateral lumbar intervertebral body fusion devices manufactured from polyetheretherketone (PEEK) (ASTM F2026) with from tantalum (per ASTM F560) radiographic markers and a one-micron thick surface layer of commercially pure titanium (per ASTM F67). The interbody spacers are available in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy and include central graft windows for receiving autogenous bone graft and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The spacers are individually packaged and gamma sterilized. The Regatta Lateral interbody spacers can be used alone with supplemental fixation or in combination with fixation implants (i.e., plates, screws, and locking covers) to create the TruProfile Interbody configuration. The fixation implants are manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136) and provided non-sterile. The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

Reef L Interbody System: The Reef L Interbody System featuring NanoMetalene surface technology consists of lateral lumbar intervertebral body fusion devices manufactured from polyetheretherketone (PEEK) (per ASTM F2026) with tantalum (per ASTM F560) markers for radiographic visualization and NanoMetalene, which is a one-micron thick surface layer of commercially pure titanium (per ASTM F67) that provides a microscopic roughened surface with nano-scale features. The interbody spacers are available in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy and include central graft windows for receiving autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The spacers are individually packaged and gamma sterilized. The Reef L interbody spacers can be used alone with supplemental fixation or in combination with fixation implants (i.e., plates, screws, and locking covers) to create the TruProfile Interbody configuration. The fixation implants are manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136) and provided non-sterile. The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

Reef TO/TA System: The Reef TO/TA Interbody System featuring NanoMetalene surface technology consists of straight, rectangular-shaped (Reef TO) and curved, banana-shaped (Reef TA), posterior lumbar intervertebral body fusion devices manufactured from polyetheretherketone (PEEK) (per ASTM F2026) with tantalum (per ASTM F560) markers for radiographic visualization and NanoMetalene, which is a one-micron thick surface layer of commercially pure titanium (per ASTM F67) that provides a microscopic roughened surface with nano-scale features. The interbody spacers are available in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy and include a central graft window for receiving autogenous bone graft and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The spacers are individually packaged and gamma sterilized. The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

Meridian Interbody System: The Meridian Interbody System featuring NanoMetalene surface technology consists of anterior lumbar intervertebral body fusion devices manufactured from polyetheretherketone (PEEK) (per ASTM F2026) with tantalum (per ASTM F560) and titanium alloy (Ti-6Al-4V ELI per ASTM F136) markers for radiographic visualization and NanoMetalene, which is a one-micron thick surface layer of commercially pure titanium (per ASTM F67). NanoMetalene surface technology provides a microscopic roughened surface with nano-scale features. The interbody spacers are available in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy and include a central graft window for receiving autogenous and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The spacers are individually packaged and gamma sterilized. The Meridian Interbody System spacers can be used alone with supplemental fixation or in combination with fixation implants to create the TruProfile and No-profile Interbody configurations. Fixation implants include Anterior Plates, Screws, Inline Fixation Anchors, Anterior Plate Locking Covers, and No-profile Locking Covers, all of which are manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136) and provided non-sterile. The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

WaveForm C Interbody System: The WaveForm C Interbody System consists of anterior cervical intervertebral body fusion devices additively manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F3001). The interbody spacers are offered in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy. The spacers, which include a central graft window for receiving autogenous bone graft and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone, are individually packaged and gamma sterilized. The WaveForm C Interbody System can be used with supplemental fixation, such as an anterior plate, or as a standalone construct in combination with the Shoreline ACS Interbody System nonsterile fixation implants (i.e., plates, bone screws, and locking covers) to create the TruProfile and No-profile configurations. The fixation implants are manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136). The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

WaveForm L Interbody System: The WaveForm L Interbody System consists of lateral lumbar intervertebral body fusion devices additively manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F3001). The interbody spacers are available in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy and include a central graft window for receiving autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The spacers are individually packaged and gamma sterilized. The WaveForm L interbody spacers can be used alone with supplemental fixation or in combination with fixation implants (i.e., plates, screws, and locking covers) to create the TruProfile Interbody configuration. The fixation implants are manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136) and provided non-sterile. The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

WaveForm TO Interbody System: The WaveForm TO Interbody System consists of posterior lumbar intervertebral body fusion devices additively manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F3001). The straight, rectangular-shaped interbody spacers are available in a variety of footprints, and lordotic options to accommodate variations in pathology and patient anatomy and include a central graft window for receiving autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The spacers are individually packaged and gamma sterilized. The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

WaveForm TA Interbody System: The WaveForm TA Interbody System consists of posterior lumbar intervertebral body fusion devices additively manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F3001). The curved, banana-shaped interbody spacers are available in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy and include a central graft window for receiving autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The spacers are individually packaged and gamma sterilized. The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

WaveForm A Interbody System: The WaveForm A Interbody System consists of anterior lumbar intervertebral body fusion devices additively manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F3001). The interbody spacers are offered in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy as well as with and without a sodium hydroxide (NaOH) surface treatment that provides a microscopic, roughened surface with nanoscale features. The interbody spacers, which include a central graft window for receiving autogenous and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone, are individually packaged and gamma sterilized. The WaveForm A Interbody System spacers can be used alone with supplemental fixation or in combination with fixation implants to create the TruProfile and No-profile Interbody configurations. Fixation implants include Anterior Plates, Screws, Inline Fixation Anchors, Anterior Plate Locking Covers, and No-profile Locking Covers, all of which are manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136) and provided non-sterile. The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

AI/ML Overview

This is a 510(k) premarket notification for a medical device, specifically a family of intervertebral body fusion devices. This type of submission focuses on demonstrating substantial equivalence to legally marketed predicate devices, rather than presenting novel clinical efficacy data that would typically involve acceptance criteria for device performance as an AI/ML algorithm.

Therefore, the requested information regarding acceptance criteria, study details, sample sizes, ground truth, expert adjudication, and MRMC studies is not applicable in the context of this FDA clearance document. This document primarily addresses the mechanical and material characteristics of the devices and their equivalence to existing technologies.

However, I can extract information related to the demonstration of substantial equivalence:

1. A table of acceptance criteria and the reported device performance

Since this is a 510(k) for an interbody fusion device, the acceptance criteria relate to manufacturing, materials, and mechanical performance, rather than algorithm performance. The document states:

  • "The subject and predicate systems are identical in regard to intended use/indications for use and technological characteristics, including operating principle, design and components, manufacturing, biocompatibility, labeling, sterility, and non-clinical testing (i.e., mechanical performance)."
  • "The subject systems differ from the predicate systems in regard to an additional sterile packaging configuration, which does not raise questions of safety and effectiveness as demonstrated through non-clinical testing."

Non-Clinical Testing Acceptance Criteria (Implied by equivalence to predicate devices and additional testing):

Acceptance Criteria CategoryReported Device Performance (Summary from document)
Intended Use/Indications for UseIdentical to cited predicate devices for various spinal fusion procedures (cervical and lumbar) for degenerative disc disease, spondylolisthesis, and scoliosis.
Technological CharacteristicsIdentical to cited predicate devices regarding: operating principle, design and components, manufacturing, biocompatibility, labeling, sterility, and mechanical performance.
MaterialsPEEK (per ASTM F2026) with tantalum (per ASTM F560) and/or titanium alloy (Ti-6Al-4V ELI per ASTM F136) markers for NanoMetalene® surface technology, commercially pure titanium (per ASTM F67) for NanoMetalene surface. WaveForm systems are additively manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F3001). All materials meet specified ASTM standards.
Sterile PackagingDemonstrated not to raise questions of safety and effectiveness through non-clinical testing (cleaning per ISO 19227, sterilization per ISO 11137, and various package testing per ISO 11607).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not applicable. The document describes a medical device, not an AI/ML algorithm. The "testing" refers to mechanical, material, and packaging assessments, not evaluation against a "test set" of patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable. There is no mention of a test set with ground truth established by experts. The determination is based on substantial equivalence to predicate devices, which involves regulatory review of engineering and manufacturing data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable. There is no test set or adjudication method described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. This is not an AI-assisted device, nor does the submission involve a MRMC comparative effectiveness study.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. This is not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

This information is not applicable. No ground truth in the context of an AI/ML algorithm is relevant here. The "ground truth" for these devices would be established engineering and material standards, and the successful performance of predicate devices.

8. The sample size for the training set

This information is not applicable. This is not an AI/ML algorithm that requires a training set.

9. How the ground truth for the training set was established

This information is not applicable. This is not an AI/ML algorithm that requires a training set or its associated ground truth establishment.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 20, 2024

SeaSpine Orthopedics Corporation Jesse Albright Manager, Regulatory Affairs 5770 Armada Drive Carlsbad, California 92008

Re: K241466

Trade/Device Name: Shoreline ACS Interbody System; Shoreline RT Interbody System; Regatta Lateral System: Reef L Interbody System; Reef TO/TA System; Meridian Interbody System; WaveForm C Interbody System; WaveForm L Interbody System; WaveForm TO Interbody System; WaveForm TA Interbody System; WaveForm A Interbody System

Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX, OVD, PHM, OVE, ODP Dated: May 23, 2024 Received: May 23, 2024

Dear Jesse Albright:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Brent Showalter -S

Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices

{2}------------------------------------------------

OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K241466

Device Name

Shoreline ACS Interbody System: Shoreline RT Interbody System: Regatta Lateral System: Reef TO/TA System: Meridian Interbody System: WaveForm C Interbody System: WaveForm L Interbody System; WaveForm TA Interbody System; WaveForm A Interbody System

Indications for Use (Describe)

Shoreline ACS Interbody System

The Shoreline ACS Interbody System with NanoMetalene® surface technology are interbody fusion devices intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) for multiple contiguous levels. DDD is defined as discogenc pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. These devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone.

When used as a standalone system, the Shoreline ACS Interbody System, which includes the TruProfile plates and Noprofile spacer, is intended to be used as an adjunct to spinal fusion procedures at multiple contiguous levels of the cervical spine (C2-T1) and must be used with bone screw fixation and locking covers. The Shoreline ACS interbody spacers with > 20° lordosis are intended to be used with supplemental fixation.

When the Shoreline ACS Interbody (excluding the No-profile spacer) is used with supplemental fixation, such as anterior cervical plates, the Shoreline ACS Interdody System is intended to be used as an adjunct to spinal fusion procedures at multiple contiguous levels of the cervical spine (C2-T1).

Shoreline RT Interbody System

The Shoreline RT Interbody System with NanoMetalene® surface technology are interbody fusion devices intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) for multiple contiguous levels. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. These devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone.

When used as a standalone system, the Shoreline RT Interbody System, which includes the TruProfile plates, is intended to be used as an adjunct to spinal fusion procedures at multiple contiguous levels of the cervical spine (C2-T1) and must be used with bone screw fixation and locking covers. The Shoreline RT interbody spacers with ≥ 20° lordosis are intended to be used with supplemental fixation.

When the Shoreline RT Interbody System is used with supplemental fixation, such as anterior cervical plates, the Shoreline RT Interbody System is intended to be used as an adjunct to spinal fusion procedures at multiple contiguous levels of the cervical spine (C2-T1).

Regatta Lateral System

Interbody Device (IBD) Implants (i.e., interbody implants used alone):

The Regatta Lateral System with NanoMetalene® surface technology is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD, defined as back pain of discogenic origin, with degeneration of the disc confirmed by history and radiographic studies). It is intended for use at ether one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of DDD with up to Grade 1 spondylolisthesis at the involved level(s). The interior of the interbody spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device.

{4}------------------------------------------------

The Regatta Lateral System is intended for use with supplemental fixation.

TruProfile Interbody Implants:

The Regatta Lateral System assembled with the TruProfile Lateral Plate, when used with Screws, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device.

The Regatta Lateral System assembled with the 1- hole TruProfile Lateral Plate, when used with Screws, is intended for use with supplemental fixation.

Reef L Interbody System

Interbody Device (IBD) Implants (i.e., interbody implants used alone):

The Reef L Interbody System with NanoMetalene® surface technology is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD, defined as back pain of discogenic origin, with degeneration of the disc confirmed by history and radiographic studies). It is intended for use at evel or two contiguous levels in the lumbar spine, from L1 to S1, for the treatment of DDD with up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogence bone graft composed of cancellous, cortical, and/or corticocancellous bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion.

The Reef L Interbody System is intended for use as an adjunct to fusion in the thoracolumbar spine from T1 to T12 and at the thoracolumbar junction (T12-L1) for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two contiguous levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenc origin with degeneration of the disc confirmed by history and radiographic studies. The Reef L Interbody System can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

The Reef L Interbody System is intended for use with supplemental fixation. Hyperlordotic sizes (greater than 200) must be used with legally marketed anterior supplemental fixation.

TruProfile Interbody Implants (i.e., interbody implants used with TruProfile Plates):

The Reef L Interbody System assembled with the TruProfile Lateral Plate, when used with Screws, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L1 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of nonoperative treatment prior to being treated with the device. The interded to be used with autogenous bone graft, allogeneic bone graft composed of cancellous, corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion. The Reef L Interoody System assembled with the TruProfile Lateral Plate, when used with Screws, can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

The Reef L Interbody System assembled with the 1-Hole TruProfile Lateral Plate, when used with a single Screw, is intended for use with additional legally marketed anterior or posterior supplemental fixation.

Reef TO/TA System

When used as an intervertebral body fusion device, the Reef TO/TA System with NanoMetalene® surface technology is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with

{5}------------------------------------------------

degenerative disc disease (DDD). DDD is defined as back pain of discogenc origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had six months of non-operative treatment. The device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, corticocancellous bone and supplemental fixation.

Meridian Interbody System

Interbody Device (IBD) Implants (i.e., interbody implants used alone):

The Meridian System with NanoMetalene® surface technology when used with or without a Spin Plate, is indicated for use as an adjunct to fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients should be skeletally mature and have had at least six (6) months of non-operative treatment. The interded to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The Meridian Interbody is intended for use with supplemental fixation.

No-Profile Interbody Implants with Screws:

The Meridian System 2-Hole No-profile Interbody, when used with Screws and with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenc origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed cancellous, cortical, and/or corticocancellous bone. The Meridian System 2-Hole No-profile Interbody must be used with the maximum number of screws allowed. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation.

The Meridian System 4-Hole No-profile Interbody, when used with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenc origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The Meridian System 4-Hole No-profile Interbody may be used as a standalone device only when at least two Screws are inserted with one inferior and one superior screw trajectory. If the surgeon chooses to use the 4-Hole No-profile Interbody with fewer than two Screws inserted with one inferior and one superior screw trajectory, then additional supplemental fixation system must be used. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation.

No-Profile Interbody Implants with Inline Fixation Anchors:

The Meridian System No-profile Interbody, when used with Inline Fixation Anchors and a No-profile Locking Cover, is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogence bone graft composed of cancellous, corticocancellous bone. The Meridian System No-profile Interbody with Inline Fixation Anchors is intended for use with supplemental fixation.

TruProfile Interbody Implants:

The Meridian System TruProfile Interbody assembled with the Anterior Plate, when used with Screws, an Anterior Plate Locking Cover, and with or without a Spin Plate, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1.

{6}------------------------------------------------

DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation.

WaveForm C Interbody System

The WaveForm C Interbody System are interbody fusion devices intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) for multiple contiguous levels. DDD is defined as discogenc pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. These devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone.

When used as a standalone system, the WaveForm C Interbody System, which includes the 2,3,4-hole TruProfile plates and 2-hole No-profile interfixated spacer, is intended to be used as an adjunct to spinal fusion procedures at multiple contiguous levels of the cervical spine (C2-T1) and must be used with bone screw fixation and locking covers.

When the WaveForm C Interbody System (excluding 2-hole No-profile interfixated spacer) is used with supplemental fixation, such as anterior cervical plates, the WaveForm C Interbody System is intended to be used as an adjunct to spinal fusion procedures at multiple contiguous levels of the cervical spine (C2-T1).

WaveForm L Interbody System

Interbody Device (IBD) Implants (i.e., interbody implants used alone):

The WaveForm L Interbody System is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD, defined as back pain of discogenic origin, with degeneration of the disc confirmed by history and radiographic studies). It is intended for use at either one level or two contiguous levels in the lumbar spine, from L1 to S1, for the treatment of DDD with up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion.

The WaveForm L Interbody System is intended for use as an adjunct to fusion in the thoracolumbar spine from T1 to T12 and at the thoracolumbar junction (T12-L1) for the treatment of symptomatic disc degenerative spondylolisthesis at one or two contiguous levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The WaveForm L Interbody System can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

The WaveForm L Interbody System is intended for use with supplemental fixation. Hyperlordotic sizes (greater than 20°) must be used with legally marketed anterior supplemental fixation.

TruProfile Interbody Implants (i.e., interbody implants used with TruProfile Plates):

The WaveForm L Interbody System assembled with the TruProfile Lateral Plate, when used with Screws, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L1 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of nonoperative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The WaveForm L Interbody System assembled with the TruProfile Lateral Plate, when used with Screws, can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

{7}------------------------------------------------

The WaveForm L Interbody System assembled with the 1-Hole TruProfile Lateral Plate, when used with a single Screw, is intended for use with additional legally marketed anterior or posterior supplemental fixation.

WaveForm TO Interbody System

When used as an intervertebral body fusion device, the system is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had six months of non-operative treatment. The device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone and supplemental fixation.

WaveForm TA Interbody System

When used as an intervertebral body fusion device, the system is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had six months of non-operative treatment. The device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone and supplemental fixation.

WaveForm A Interbody System

Interbody Device (IBD) Implants (i.e., interbody implants used alone):

The WaveForm A System Interbody, when used with or without a Spin Plate, is indicated for use as an adjunct to fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients should be skeletally mature and have had at least six (6) months of non-operative treatment. The interded to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The WaveForm A System Interbody is intended for use with supplemental fixation.

No-profile Interbody Implants with Screws:

The WaveForm A System 2-Hole No-profile Interbody, when used with Screws and with or without a Noprofile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed cancellous, cortical, and/or corticocancellous bone. The WaveForm A System 2-Hole No-profile Interbody must be used with the maximum number of screws allowed. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation.

The WaveForm A System 4-Hole No-profile Interbody, when used with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The WaveForm A System 4-Hole No-profile Interbody may be used as a standalone device only when at least two Screws are inserted with one inferior and one superior screw trajectory. If the surgeon chooses to use the 4-Hole No-profile Interbody with fewer than two Screws inserted with one inferior and one superior screw trajectory, then additional supplemental fixation system must be used. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation.

{8}------------------------------------------------

No-profile Interbody Implants with Inline Fixation Anchors:

The WaveForm A System No-profile Interbody, when used with Inline Fixation Anchors and with a No-profile Locking Cover, is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interded to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The WaveForm A System No-profile Interbody with Inline Fixation Anchors is intended for use with supplemental fixation.

TruProfile Interbody Implants:

The WaveForm A System TruProfile Interbody assembled with the Anterior Plate, when used with Screws, an Anterior Plate Locking Cover, and with or without a Spin Plate, is a standalone indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/ or corticocancellous bone. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{9}------------------------------------------------

510(k) Summary

Contact Details
Applicant Name:SeaSpine Orthopedics Corporation
Address:Phone number:Fax number:5770 Armada Drive, Carlsbad CA(760) 216-5176(760) 683-6874
Contact person (primary):Jesse Albright, Manager, Regulatory Affairs
Date Prepared:May 23, 2024
Device Name(s)Trade Name(s):Shoreline ACS Interbody SystemShoreline RT Interbody SystemRegatta Lateral SystemReef L Interbody SystemReef TO/TA SystemMeridian Interbody SystemWaveForm C Interbody SystemWaveForm L Interbody SystemWaveForm TO Interbody SystemWaveForm TA Interbody SystemWaveForm A Interbody System
Common Name(s):Intervertebral Body Fusion Device
Classification Name(s):Intervertebral Fusion Device with Bone Graft, Lumbar (21 CFR888.3080), Intervertebral Fusion Device With Integrated Fixation,Lumbar (21 CFR 888.3080), Intervertebral Fusion Device WithIntegrated Fixation, Cervical (21 CFR 888.3080), IntervertebralFusion Device With Bone Graft, Cervical (21 CFR 888.3080),Intervertebral Fusion Device With Bone Graft, Thoracic (21 CFR888.3080)
Class:2
Product Code(s):MAX, OVD, PHM, OVE, ODR

{10}------------------------------------------------

Legally Marketed Predicate Devices

510(k) NumberProduct CodeTrade NameManufacturer
Primary Predicate Device
K210497MAX, OVDRegatta Lateral System,Reef TO/TA SystemSeaSpine OrthopedicsCorporation
Additional Predicate Device(s)
K240566MAX, OVD, PHMReef L Interbody System,WaveForm L Interbody SystemSeaSpine OrthopedicsCorporation
K233694MAX, OVDMeridian Interbody System,WaveForm A Interbody SystemSeaSpine OrthopedicsCorporation
K233414OVE, ODPShoreline ACS Interbody System,Shoreline RT Interbody SystemSeaSpine OrthopedicsCorporation
K213359OVE, ODPWaveForm C Interbody SystemSeaSpine OrthopedicsCorporation
K213420MAX, PHMWaveForm TO Interbody System,WaveForm TA Interbody SystemSeaSpine OrthopedicsCorporation

Device Description

Shoreline ACS Interbody System

The Shoreline ACS Interbody System featuring NanoMetalene surface technology consists of anterior cervical intervertebral body fusion devices manufactured from polyetheretherketone (PEEK) (per ASTM F2026) with tantalum (per ASTM F560) and/or titanium alloy (Ti-6Al-4V EU per ASTM F136) markers for radiographic visualization and NanoMetalene, which is a one-micron thick surface layer of commercially pure titanium (per ASTM F67) that provides a microscopic roughened surface with nano-scale features. The interbody spacers are offered in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy. The spacers, which include a central graft window for receiving autogenous bone graft and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone, are individually packaged and gamma sterilized.

The Shoreline ACS Interbody System can be used with supplemental fixation, such as an anterior plate, or as a standalone construct in combination with non-sterile fixation implants (i.e., plates, bone screws, and locking covers) to create the TruProfile configurations. The fixation implants are manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136).

The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

{11}------------------------------------------------

Shoreline RT Interbody System

The Shoreline RT Interbody System featuring NanoMetalene surface technology consists of anterior cervical intervertebral body fusion devices manufactured from polyetheretherketone (PEEK) (per ASTM F2026) with tantalum (per ASTM F560) and/or titanium alloy (Ti-6Al-4V ELI per ASTM F136) markers for radiographic visualization and NanoMetalene, which is a one-micron thick surface layer of commercially pure titanium (per ASTM F67) that provides a microscopic roughened surface with nano-scale features. The interbody spacers are offered in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy. The spacers, which include a central graft window for receiving autogenous bone graft and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone, are individually packaged and gamma sterilized.

The Shoreline RT Interbody System can be used with supplemental fixation, such as an anterior plate, or as a standalone construct in combination with the Shoreline ACS Interbody System nonsterile fixation implants (i.e., plates, bone screws, and locking covers) to create the TruProfile configuration. The fixation implants are manufactured from titanium alloy (Ti-6AI-4V ELI per ASTM F136).

The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

Regatta Lateral System

The Regatta Lateral System featuring NanoMetalene surface technology consists of lateral lumbar intervertebral body fusion devices manufactured from polyetheretherketone (PEEK) (ASTM F2026) with from tantalum (per ASTM F560) radiographic markers and a one-micron thick surface layer of commercially pure titanium (per ASTM F67). The interbody spacers are available in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy and include central graft windows for receiving autogenous bone graft and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The spacers are individually packaged and gamma sterilized.

The Regatta Lateral interbody spacers can be used alone with supplemental fixation or in combination with fixation implants (i.e., plates, screws, and locking covers) to create the TruProfile Interbody configuration. The fixation implants are manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136) and provided non-sterile.

The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

{12}------------------------------------------------

Reef L Interbody System

The Reef L Interbody System featuring NanoMetalene surface technology consists of lateral lumbar intervertebral body fusion devices manufactured from polyetheretherketone (PEEK) (per ASTM F2026) with tantalum (per ASTM F560) markers for radiographic visualization and NanoMetalene, which is a one-micron thick surface layer of commercially pure titanium (per ASTM F67) that provides a microscopic roughened surface with nano-scale features. The interbody spacers are available in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy and include central graft windows for receiving autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The spacers are individually packaged and gamma sterilized.

The Reef L interbody spacers can be used alone with supplemental fixation or in combination with fixation implants (i.e., plates, screws, and locking covers) to create the TruProfile Interbody configuration. The fixation implants are manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136) and provided non-sterile.

The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

Reef TO/TA System

The Reef TO/TA Interbody System featuring NanoMetalene surface technology consists of straight, rectangular-shaped (Reef TO) and curved, banana-shaped (Reef TA), posterior lumbar intervertebral body fusion devices manufactured from polyetheretherketone (PEEK) (per ASTM F2026) with tantalum (per ASTM F560) markers for radiographic visualization and NanoMetalene, which is a one-micron thick surface layer of commercially pure titanium (per ASTM F67) that provides a microscopic roughened surface with nano-scale features. The interbody spacers are available in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy and include a central graft window for receiving autogenous bone graft and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The spacers are individually packaged and gamma sterilized.

The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

Meridian Interbody System

The Meridian Interbody System featuring NanoMetalene surface technology consists of anterior lumbar intervertebral body fusion devices manufactured from polyetheretherketone (PEEK) (per ASTM F2026) with tantalum (per ASTM F560) and titanium alloy (Ti-6Al-4V ELI per ASTM F136) markers for radiographic visualization and NanoMetalene, which is a one-micron thick surface

{13}------------------------------------------------

layer of commercially pure titanium (per ASTM F67). NanoMetalene surface technology provides a microscopic roughened surface with nano-scale features. The interbody spacers are available in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy and include a central graft window for receiving autogenous and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The spacers are individually packaged and gamma sterilized.

The Meridian Interbody System spacers can be used alone with supplemental fixation or in combination with fixation implants to create the TruProfile and No-profile Interbody configurations. Fixation implants include Anterior Plates, Screws, Inline Fixation Anchors, Anterior Plate Locking Covers, and No-profile Locking Covers, all of which are manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136) and provided non-sterile.

The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

WaveForm C Interbody System

The WaveForm C Interbody System consists of anterior cervical intervertebral body fusion devices additively manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F3001). The interbody spacers are offered in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy. The spacers, which include a central graft window for receiving autogenous bone graft and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone, are individually packaged and gamma sterilized.

The WaveForm C Interbody System can be used with supplemental fixation, such as an anterior plate, or as a standalone construct in combination with the Shoreline ACS Interbody System nonsterile fixation implants (i.e., plates, bone screws, and locking covers) to create the TruProfile and No-profile configurations. The fixation implants are manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136).

The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

WaveForm L Interbody System

The WaveForm L Interbody System consists of lateral lumbar intervertebral body fusion devices additively manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F3001). The interbody spacers are available in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy and include a central graft window for receiving autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or

{14}------------------------------------------------

corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The spacers are individually packaged and gamma sterilized.

The WaveForm L interbody spacers can be used alone with supplemental fixation or in combination with fixation implants (i.e., plates, screws, and locking covers) to create the TruProfile Interbody configuration. The fixation implants are manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136) and provided non-sterile.

The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

WaveForm TO Interbody System

The WaveForm TO Interbody System consists of posterior lumbar intervertebral body fusion devices additively manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F3001). The straight, rectangular-shaped interbody spacers are available in a variety of footprints, and lordotic options to accommodate variations in pathology and patient anatomy and include a central graft window for receiving autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The spacers are individually packaged and gamma sterilized.

The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

WaveForm TA Interbody System

The WaveForm TA Interbody System consists of posterior lumbar intervertebral body fusion devices additively manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F3001). The curved, banana-shaped interbody spacers are available in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy and include a central graft window for receiving autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The spacers are individually packaged and gamma sterilized.

The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

WaveForm A Interbody System

The WaveForm A Interbody System consists of anterior lumbar intervertebral body fusion devices additively manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F3001). The interbody

{15}------------------------------------------------

spacers are offered in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy as well as with and without a sodium hydroxide (NaOH) surface treatment that provides a microscopic, roughened surface with nanoscale features. The interbody spacers, which include a central graft window for receiving autogenous and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone, are individually packaged and gamma sterilized.

The WaveForm A Interbody System spacers can be used alone with supplemental fixation or in combination with fixation implants to create the TruProfile and No-profile Interbody configurations. Fixation implants include Anterior Plates, Screws, Inline Fixation Anchors, Anterior Plate Locking Covers, and No-profile Locking Covers, all of which are manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136) and provided non-sterile.

The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

Intended Use/Indications for Use

Shoreline ACS Interbody System

The Shoreline ACS Interbody System with NanoMetalene® surface technology are interbody fusion devices intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) for multiple contiguous levels. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. These devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone.

When used as a standalone system, the Shoreline ACS Interbody System, which includes the TruProfile plates and No-profile spacer, is intended to be used as an adjunct to spinal fusion procedures at multiple contiguous levels of the cervical spine (C2-T1) and must be used with bone screw fixation and locking covers. The Shoreline ACS interbody spacers with ≥ 20° lordosis are intended to be used with supplemental fixation.

When the Shoreline ACS Interbody (excluding the No-profile spacer) is used with supplemental fixation, such as anterior cervical plates, the Shoreline ACS Interbody System is intended to be used as an adjunct to spinal fusion procedures at multiple contiguous levels of the cervical spine (C2-T1).

Shoreline RT Interbody System

The Shoreline RT Interbody System with NanoMetalene® surface technology are interbody fusion devices intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) for multiple contiguous levels. DDD is defined as discogenic pain with

{16}------------------------------------------------

degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. These devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone.

When used as a standalone system, the Shoreline RT Interbody System, which includes the TruProfile plates, is intended to be used as an adjunct to spinal fusion procedures at multiple contiguous levels of the cervical spine (C2-T1) and must be used with bone screw fixation and locking covers. The Shoreline RT interbody spacers with ≥ 20° lordosis are intended to be used with supplemental fixation.

When the Shoreline RT Interbody System is used with supplemental fixation, such as anterior cervical plates, the Shoreline RT Interbody System is intended to be used as an adjunct to spinal fusion procedures at multiple contiguous levels of the cervical spine (C2-T1).

Regatta Lateral System

Interbody Device (IBD) Implants (i.e., interbody implants used alone):

The Regatta Lateral System with NanoMetalene® surface technology is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD, defined as back pain of discogenic origin, with degeneration of the disc confirmed by history and radiographic studies). It is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of DDD with up to Grade 1 spondylolisthesis at the involved level(s). The interior of the interbody spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device.

The Regatta Lateral System is intended for use with supplemental fixation.

TruProfile Interbody Implants:

The Regatta Lateral System assembled with the TruProfile Lateral Plate, when used with Screws, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device.

The Regatta Lateral System assembled with the 1- hole TruProfile Lateral Plate, when used with Screws, is intended for use with supplemental fixation.

{17}------------------------------------------------

Reef L Interbody System

Interbody Device (IBD) Implants (i.e., interbody implants used alone):

The Reef L Interbody System with NanoMetalene® surface technology is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD, defined as back pain of discogenic origin, with degeneration of the disc confirmed by history and radiographic studies). It is intended for use at either one level or two contiguous levels in the lumbar spine, from L1 to S1, for the treatment of DDD with up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of nonoperative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion.

The Reef L Interbody System is intended for use as an adjunct to fusion in the thoracolumbar spine from T1 to T12 and at the thoracolumbar junction (T12-L1) for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two contiguous levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The Reef L Interbody System can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

The Reef L Interbody System is intended for use with supplemental fixation. Hyperlordotic sizes (greater than 20°) must be used with legally marketed anterior supplemental fixation.

TruProfile Interbody Implants (i.e., interbody implants used with TruProfile Plates):

The Reef L Interbody System assembled with the TruProfile Lateral Plate, when used with Screws, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L1 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of nonoperative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft, allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The Reef L Interbody System assembled with the TruProfile Lateral Plate, when used with Screws, can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

The Reef L Interbody System assembled with the 1-Hole TruProfile Lateral Plate, when used with a single Screw, is intended for use with additional legally marketed anterior or posterior supplemental fixation.

{18}------------------------------------------------

Reef TO/TA System

When used as an intervertebral body fusion device, the Reef TO/TA System with NanoMetalene® surface technology is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had six months of non-operative treatment. The device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone and supplemental fixation.

Meridian Interbody System

Interbody Device (IBD) Implants (i.e., interbody implants used alone):

The Meridian System with NanoMetalene® surface technology Interbody, when used with or without a Spin Plate, is indicated for use as an adjunct to fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients should be skeletally mature and have had at least six (6) months of non-operative treatment. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The Meridian Interbody is intended for use with supplemental fixation.

No-Profile Interbody Implants with Screws:

The Meridian System 2-Hole No-profile Interbody, when used with Screws and with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed cancellous, cortical, and/or corticocancellous bone. The Meridian System 2-Hole and 3-Hole No-profile Interbody must be used with the maximum number of screws allowed. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation.

The Meridian System 4-Hole No-profile Interbody, when used with Screws and with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the

{19}------------------------------------------------

device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The Meridian System 4-Hole No-profile Interbody may be used as a standalone device only when at least two Screws are inserted with one inferior and one superior screw trajectory. If the surgeon chooses to use the 4-Hole No-profile Interbody with fewer than two Screws inserted with one inferior and one superior screw trajectory, then additional supplemental fixation system must be used. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation.

No-Profile Interbody Implants with Inline Fixation Anchors:

The Meridian System No-profile Interbody, when used with Inline Fixation Anchors and a Noprofile Locking Cover, is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of nonoperative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The Meridian System No-profile Interbody with Inline Fixation Anchors is intended for use with supplemental fixation.

TruProfile Interbody Implants:

The Meridian System TruProfile Interbody assembled with the Anterior Plate, when used with Screws, an Anterior Plate Locking Cover, and with or without a Spin Plate, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of nonoperative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation.

WaveForm C Interbody System

The WaveForm C Interbody System are interbody fusion devices intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) for multiple contiguous levels. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. These devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone.

When used as a standalone system, the WaveForm C Interbody System, which includes the 2,3,4hole TruProfile plates and 2-hole No-profile interfixated spacer, is intended to be used as an

{20}------------------------------------------------

adjunct to spinal fusion procedures at multiple contiguous levels of the cervical spine (C2-T1) and must be used with bone screw fixation and locking covers.

When the WaveForm C Interbody System (excluding 2-hole No-profile interfixated spacer) is used with supplemental fixation, such as anterior cervical plates, the WaveForm C Interbody System is intended to be used as an adjunct to spinal fusion procedures at multiple contiguous levels of the cervical spine (C2-T1).

WaveForm L Interbody System

Interbody Device (IBD) Implants (i.e., interbody implants used alone):

The WaveForm L Interbody System is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD, defined as back pain of discogenic origin, with degeneration of the disc confirmed by history and radiographic studies). It is intended for use at either one level or two contiguous levels in the lumbar spine, from L1 to S1, for the treatment of DDD with up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion.

The WaveForm L Interbody System is intended for use as an adjunct to fusion in the thoracolumbar spine from T1 to T12 and at the thoracolumbar junction (T12-L1) for the treatment of symptomatic disc degeneration (DDD) or degenerative spondylolisthesis at one or two contiguous levels, including thoracic disc herniation (with myelopathy and/or radiculopathy with or without axial pain). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The WaveForm L Interbody System can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

The WaveForm L Interbody System is intended for use with supplemental fixation. Hyperlordotic sizes (greater than 20°) must be used with legally marketed anterior supplemental fixation.

TruProfile Interbody Implants (i.e., interbody implants used with TruProfile Plates):

The WaveForm L Interbody System assembled with the TruProfile Lateral Plate, when used with Screws, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L1 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The WaveForm L Interbody System assembled with the TruProfile

{21}------------------------------------------------

Lateral Plate, when used with Screws, can also be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.

The WaveForm L Interbody System assembled with the 1-Hole TruProfile Lateral Plate, when used with a single Screw, is intended for use with additional legally marketed anterior or posterior supplemental fixation.

WaveForm TO Interbody System

When used as an intervertebral body fusion device, the system is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had six months of non-operative treatment. The device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone and supplemental fixation.

WaveForm TA Interbody System

When used as an intervertebral body fusion device, the system is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had six months of non-operative treatment. The device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone and supplemental fixation.

WaveForm A Interbody System

Interbody Device (IBD) Implants (i.e., interbody implants used alone):

The WaveForm A System Interbody, when used with or without a Spin Plate, is indicated for use as an adjunct to fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients should be skeletally mature and have had at least six (6) months of non-operative treatment. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The WaveForm A System Interbody is intended for use with supplemental fixation.

{22}------------------------------------------------

No-profile Interbody Implants with Screws:

The WaveForm A System 2-Hole No-profile Interbody, when used with Screws and with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed cancellous, cortical, and/or corticocancellous bone. The WaveForm A System 2-Hole and 3-Hole No-profile Interbody must be used with the maximum number of screws allowed. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation.

The WaveForm A System 4-Hole No-profile Interbody, when used with Screws and with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The WaveForm A System 4-Hole No-profile Interbody may be used as a standalone device only when at least two Screws are inserted with one inferior and one superior screw trajectory. If the surgeon chooses to use the 4-Hole No-profile Interbody with fewer than two Screws inserted with one inferior and one superior screw trajectory, then additional supplemental fixation system must be used. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation.

No-profile Interbody Implants with Inline Fixation Anchors:

The WaveForm A System No-profile Interbody, when used with Inline Fixation Anchors and with a No-profile Locking Cover, is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of nonoperative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The WaveForm A System No-profile Interbody with Inline Fixation Anchors is intended for use with supplemental fixation.

TruProfile Interbody Implants:

The WaveForm A System TruProfile Interbody assembled with the Anterior Plate, when used with Screws, an Anterior Plate Locking Cover, and with or without a Spin Plate, is a standalone

{23}------------------------------------------------

interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of nonoperative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/ or corticocancellous bone. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation.

Summary of Technological Characteristics

The Shoreline ACS Interbody System, Shoreline RT Interbody System, Regatta Lateral System, Reef L Interbody System, Reef TO/TA System, Meridian Interbody System, WaveForm C Interbody System, WaveForm L Interbody System, WaveForm TO Interbody System, WaveForm TA Interbody System, and WaveForm A Interbody System technological characteristics are substantially equivalent to the cited predicate devices. The subject and predicate systems are identical in regard to intended use/indications for use and technological characteristics, including operating principle, design and components, manufacturing, biocompatibility, labeling, sterility, and non-clinical testing (i.e., mechanical performance). The subject systems differ from the predicate systems in regard to an additional sterile packaging configuration, which does not raise questions of safety and effectiveness as demonstrated through non-clinical testing.

Non-Clinical Testing

Testing performed for the additional sterile packaging configuration for the subject interbody spacer systems included cleaning (per ISO 19227), sterilization (per ISO 11137), and various package testing (per ISO 11607).

Clinical Testing

Not applicable. The determination of substantial equivalence is not based on an assessment of clinical performance data.

Conclusions

The submitted data demonstrates that the Shoreline ACS Interbody System, Shoreline RT Interbody System, Regatta Lateral System, Reef L Interbody System, Reef TO/TA System, Meridian Interbody System, WaveForm C Interbody System, WaveForm L Interbody System, WaveForm TO Interbody System, WaveForm TA Interbody System, and WaveForm A Interbody System are substantially equivalent to the cited legally marketed predicates.

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.