K Number
K241414

Validate with FDA (Live)

Date Cleared
2024-09-26

(132 days)

Product Code
Regulation Number
870.5175
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CrossWise™ RF Transseptal Cannula and accessories are used to create an atrial septal defect in the heart. Secondary indications include infusing solutions including heparinized saline and mixtures of 50% contrast media and 50% saline.

Device Description

The CrossWise RF Transseptal Access System is used to puncture the Fossa Ovalis (FO) to establish transcatheter access from the right atrium to the left atrium. Monopolar radiofrequency (RF) energy is delivered between the CrossWise focal force electrode and a patient return electrode. The unique design of the focal force electrode minimizes trauma to cardiac tissue unless RF energy is applied. A colored sleeve is provided on the CrossWise RF Transseptal Cannula handle to indicate to the user when the tip of the catheter is still within the dilator.

The CrossWise RF Transseptal Cannula is supplied with a compatible dilator and a Super Stiff 0.032" PTFE-Coated Fixed Core J-Tip guidewire for vascular introduction using an over-thewire technique. The RF Transseptal Cannula is connected to a ValleyLabs Force2 Electrosurgical Generator (Medtronic, Inc) via the CrossWise RF Adapter Cable (packaged separately - Model CW-1002) and a commercially available Electrosurgical Pencil.

The CrossWise RF Transseptal Cannula is designed to facilitate injection of heparinized saline and/or contrast solution. The dimensions for the CrossWise RF Transseptal Cannula can be found on the device label. The CrossWise RF Transseptal Access System offers four configurations compatible with various commercially available guide sheaths (Table 5). All four configurations require use of a single model (CW-1002) CrossWise RF Adapter Cable Model which is packaged separately.

AI/ML Overview

This document does not describe a study involving device performance metrics relevant to AI/ML, such as sensitivity, specificity, or F1 score, or any human reader studies using AI assistance. The provided text is a 510(k) summary for a medical device (CrossWise RF Transseptal Access System) to demonstrate substantial equivalence to predicate devices, focusing on non-clinical performance data like biocompatibility, sterilization, electrical safety, and bench testing.

Therefore, many of the requested categories related to acceptance criteria, device performance, ground truth, and human reader studies cannot be populated from the given text.

Here's a breakdown of the information that can be extracted:

1. A table of Acceptance Criteria and the Reported Device Performance:

The document doesn't provide specific numerical acceptance criteria for performance metrics in a table format. Instead, it states that the device was tested to conform with various international standards and that "The nonclinical bench data support the safety of the device and demonstrate that the CrossWise RF Transseptal System performs as intended in the specified use conditions."

The types of tests performed and the general findings are:

Test CategoryAcceptance Criteria (Implied by standard conformance)Reported Device Performance (Summary)
BiocompatibilityConformance to ISO 10993-1Meets requirements for intended use.
SterilizationConformance to ISO 11135, SAL of 10⁻⁶Achieves a sterility assurance level (SAL) of 10⁻⁶.
Electrical Safety & EMCConformance to IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2Complies with applicable sections of these standards.
Bench Testing(Implied by specific performance tests)Performs as intended in specified use conditions.
  - Packaging Integrity(Implied)Passed (implied by overall conclusion).
  - Label Integrity(Implied)Passed (implied by overall conclusion).
  - Visual & Dimensional(Implied)Passed (implied by overall conclusion).
  - Introducer Set Compatibility(Implied)Passed (implied by overall conclusion).
  - Electrical Functionality/Compatibility(Implied)Passed (implied by overall conclusion).
  - Electrical Safety(Implied)Passed (implied by overall conclusion).
  - Mechanical Functionality(Implied)Passed (implied by overall conclusion).
  - Mechanical Durability(Implied)Passed (implied by overall conclusion).
  - Puncture Performance(Implied)Passed (implied by overall conclusion).
  - Particulate Generation(Implied)Passed (implied by overall conclusion).
  - Radiopacity(Implied)Passed (implied by overall conclusion).
  - Corrosion Resistance(Implied)Passed (implied by overall conclusion).

2. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: Not specified in terms of number of devices or number of tests for each type, other than "Devices were subjected to 2X sterilization and distribution simulation". The document states "Design verification testing was performed on the CrossWise RF Transseptal Access System with CrossWise RF Adapter Cable at two time points: immediately after manufacturing (T=0) and after six months of accelerated aging (T=6M AA)."
  • Data Provenance: Not explicitly stated, however, the testing appears to be internal validation performed by the manufacturer, Circa Scientific, Inc., or a contract lab on their behalf. It is non-clinical (bench).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. This is a non-AI/ML medical device submission based on physical and electrical performance. Ground truth would be based on instrument readings, standardized methods, and expert evaluation of physical properties, not clinical expert consensus.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable for this type of device and testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. This is not an AI/ML device.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • No. This is not an AI/ML device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • For the non-clinical performance data, the "ground truth" is based on:
    • Validated test methods outlined in international standards (e.g., ISO 10993 for biocompatibility, IEC 60601 for electrical safety, ISO 11135 for sterilization).
    • Objective measurements from laboratory equipment and physical inspection.
    • Functional verification against design specifications.

8. The sample size for the training set:

  • Not applicable. This is not an AI/ML device.

9. How the ground truth for the training set was established:

  • Not applicable. This is not an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 26, 2024

Circa Scientific, Inc. % Jennifer Willner President JW Regulatory Consulting, LLC 406 Wacouta St. Suite 417 Saint Paul, Minnesota 55101

Re: K241414

Trade/Device Name: CrossWise RF Transseptal Access System (Models: CW-1085S, CW-1085A, CW-1012W, CW-1012C); CrossWise RF Adapter Cable (Model CW-1002) Regulation Number: 21 CFR 870.5175 Regulation Name: Septostomy Catheter Regulatory Class: Class II Product Code: DXF Dated: May 17, 2024 Received: May 17, 2024

Dear Jennifer Willner:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Alexandra K. Manaras -S

For Kalkidan Molla Acting Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K241414

Device Name

CrossWise RF Transseptal Access System (Models: CW-1085A, CW-1085A, CW-1012W, CW-1012C); CrossWise RF Adapter Cable (Model CW-1002)

Indications for Use (Describe)

The CrossWise™ RF Transseptal Cannula and accessories are used to create an atrial septal defect in the heart. Secondary indications include infusing solutions including heparinized saline and mixtures of 50% contrast media and 50% saline.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801)
-------------------------------------------------------------------------------------------------------------------------------------------------------------

| X | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) SUMMARY

510(k) Number: K241414

Date Prepared: September 26, 2024

Table 1: Submitter Information

Manufacturer:Manufacturer's Contact Person:
Circa Scientific, Inc.Jennifer Willner, President
14 Inverness Dr E, Suite H-136JW Regulatory Consulting LLC
Englewood, CO 80112Phone: (612) 240 - 8904
US FDA ERN: 3009437315Email: Jennifer@JWRegulatoryConsulting.com

Table 2: Device Information

Trade NameCrossWise™ RF Transseptal Access SystemCrossWise™ RF Adapter Cable
Common NameTransseptal Access System
Classification NameCatheter, Septostomy
Regulation21CFR 870.5175
Product CodeDXF
Regulatory Classification:Class II
Device Panel:Cardiovascular

The Circa Scientific CrossWise RF Transseptal Access System with CrossWise RF Adapter Cable is substantially equivalent to the previously cleared predicate Cross Vascular RF Transseptal Needle with Connection Cable Predicate Device (Table 3) and has aspects similar to those in the AcQCross Qx Integrated Transseptal Dilator/Needle Reference Device (Table 4).

Table 3: Predicate Devices

Predicate DeviceManufacturerFDA 510(k)
Cross Vascular RF Transseptal NeedleCross Vascular, Inc.K232852

Table 4: Reference Device

Reference DeviceManufacturerFDA 510(k)
AcQCross Qx Transseptal Dilator/NeedleAcutus MedicalK220047

Device Description

The CrossWise RF Transseptal Access System is used to puncture the Fossa Ovalis (FO) to establish transcatheter access from the right atrium to the left atrium. Monopolar radiofrequency (RF) energy is delivered between the CrossWise focal force electrode and a patient return electrode. The unique design of the focal force electrode minimizes trauma to cardiac tissue unless RF energy is applied. A colored sleeve is provided on the CrossWise RF Transseptal Cannula handle to indicate to the user when the tip of the catheter is still within the dilator.

The CrossWise RF Transseptal Cannula is supplied with a compatible dilator and a Super Stiff 0.032" PTFE-Coated Fixed Core J-Tip guidewire for vascular introduction using an over-thewire technique. The RF Transseptal Cannula is connected to a ValleyLabs Force2

{5}------------------------------------------------

Electrosurgical Generator (Medtronic, Inc) via the CrossWise RF Adapter Cable (packaged separately - Model CW-1002) and a commercially available Electrosurgical Pencil.

The CrossWise RF Transseptal Cannula is designed to facilitate injection of heparinized saline and/or contrast solution. The dimensions for the CrossWise RF Transseptal Cannula can be found on the device label. The CrossWise RF Transseptal Access System offers four configurations compatible with various commercially available guide sheaths (Table 5). All four configurations require use of a single model (CW-1002) CrossWise RF Adapter Cable Model which is packaged separately.

CrossWise Transseptal Access SystemCompatible Commercial Introducer Set
DescriptionREFDescriptionREF
CrossWise 8.5Fr, 71cmC1 SwartzCW-1085SSwartz Braided Transseptal GuidingIntroducer 8.5Fr, 63cm, SL0 and SL1[St Jude Medical / Abbott]407451407453
CrossWise 8.5Fr, 97cmC1 AgilisCW-1085AAgilis NxT Steerable Introducer 8.5Fr,71cm, Small, Medium, Large Curl[St Jude Medical / Abbott]408309408310G408324
CrossWise 12Fr, 88cm C1WatchmanCW-1012WWatchman[Boston Scientific]M635TU70010
CrossWise 12Fr, 88cm C1FlexCathCW-1012CFlexCath Advance Steerable Sheath 12F[Medtronic]4FC12

Table 5: Compatible Introducer Sets

Indications for Use

The CrossWise™ RF Transseptal Cannula and accessories are used to create an atrial septal defect in the heart. Secondary indications include infusing solutions including heparinized saline and mixtures of 50% contrast media and 50% saline.

Comparison of Technological Characteristics with the Predicate and Reference Device

The Subject and Predicate (PD) and Reference (RD) Devices are based on the same technological elements of delivering RF power in a monopolar mode between their distal tip electrode and commercially available Patient Return Electrode. Both devices are also loaded through commercially available transseptal introducer sets and have connections at their proximal end to a RF generator. The Subject and PD/RD devices both contain holes at the distal end to facilitate injection of contrast solution.

The following technological differences exist between the Subject and PD/RD device(s):

  • Materials of construction
  • Inner lumen dielectric coating .
  • Tip electrode geometry
  • . Tip protrusion indicator sleeve

These differences do not raise new questions regarding safety or effectiveness and do not impact the intended use of the device.

{6}------------------------------------------------

FeatureSubject DevicePredicate Device 1 (K232852)Reference Device 2 (K220047)Analysis of Differences
ProductNameCrossWise RF TransseptalAccess SystemCross Vascular RF TransseptalNeedleAcQCross Qx IntegratedTransseptal Dilator/Needle
ManufacturerCirca Scientific, Inc.Cross Vascular, Inc.Acutus Medical, Inc.
Product Code/ RegulationDXF / 21CFR 870.5175DXF / 21CFR 870.5175DYB, DRE / 21CFR 870.1340Identical to PD
Indicationsfor UseThe CrossWise RF TransseptalSystem is used to create anatrial septal defect in the heart.Secondary indications includeinfusing solutions includingheparinized saline and mixturesof 50% contrast media and 50%saline.The Cross Vascular RFTransseptal Needle is used tocreate an atrial septal defect inthe heart. Secondary indicationsinclude infusing solutionsincluding heparinized saline andmixtures of 50% contrast mediaand 50% saline.The AcQCross™ Qx IntegratedTransseptal Dilator/Needle isindicated to puncture the interatrialseptum to gain access to the leftside of the heart whereby variouscardiovascular catheters areintroduced.Identical to PD; Same as RD withaddition of infusing solutions
EnergyDeliveryRF (monopolar mode)RF (monopolar mode)RF (monopolar mode)Identical to PD & RD
Key DeviceComponentsCannula Handle with electricalconnector and luer connectorNeedle Handle with electricalconnector and luer connectorIntegrated Dilator/Needle Handlewith electrical connector and luerconnectorEquivalent to PD & RD
Inner and outer polymer coatedtubular stainless steel shaftOuter polymer coated SST ShaftElongated shaft with tapered tipand central lumen to track overguidewire.Hollow stainless steel transseptalneedle. Shaft and needle connectedto proximal handle.Similar to PD & RD; minor designdifferences do not raise newquestions of safety or effectiveness
Radiopaque ShaftRadiopaque Shaft/TipRadiopaque Shaft/TipIdentical to PD & RD
RF Adapter CableConnection CableAdapter Cables – EGM & ESIdentical to PD & RD
DilatorIntegrated dilator/needleSimilar – PD relies on dilator fromIntroducer Set while RD has
FeatureSubject DevicePredicate Device 1 (K232852)Reference Device 2 (K220047)Analysis of Differences
integrated dilator/needle; Minordesign differences do not raise newquestions of safety or effectiveness
GuidewireGuidewireIdentical to RD
Sizes /Dimensions8.5F, 71cm, C1 Curve8.5F, 97cm, C1 Curve12F, 88cm, C1 Curve8.5F, 71cm, C0 Curve8.5F, 71cm, C1 Curve8.5F, 98cm, C0 Curve8.5F, 98cm, C1 Curve8.5F, 61cm8.5F, 71cm8.5F, 63cm12F, 65cm12F, 70cmSimilar to PD & RD; Subject Devicenot available in 61-65cm lengths butdoes not raise new questions ofsafety or effectiveness
GuidewireCompatibility0.032"0.032"0.032"Identical to PD & RD
IntroducerCompatibilityAgilis, SwartzWatchmanFlexCathAgilis, SwartzHeartspanWatchmanFlexCathAgilis, SwartzVizigoSimilar; Subject Device notcompatible with all of RDcompatible devices but does not raisenew questions of safety oreffectiveness
DeviceMaterials(PatientContacting)Handle: Polypropylene, WhitecolorantShaft: Stainless Steel, Adcoat41-3500, White colorantDistal Tip: Stainless SteelGuidewire: Stainless Steel,PTFEDilator Shaft: High DensityPolyethylene, Barium Sulfate,Blue ColorantHandle: Polypropylene, bluecolorantShaft: 304 Stainless Steel,Adcoat 41-3500, blue colorantDistal Tip: GoldShaft: Polyethylene HexeneCopolymer; ethylenehomopolymer; barium sulfate withblue colorantNeedle: 304 Stainless steelHypotube: 304 stainless steelLuer fitting: polycarbonateSimilar; minor design differences donot raise new questions of safety oreffectiveness
PackagingSterile Pouch with IFU in ShelfCartonSterile Pouch with IFU in ShelfCartonSterile Pouch with IFU in ShelfCartonIdentical to PD & RD
FeatureSubject DevicePredicate Device 1 (K232852)Reference Device 2 (K220047)Analysis of Differences
SterilizationMethod /SALEO / 10-6EO / 10-6EO / 10-6Identical to PD & RD
Single-UseDevices?YesYesYesIdentical to PD & RD
Inserted Overa Guidewire?YesNoYesIdentical to RD
Shelf-Life6M, however intend to extendup to 2 years upon successfulcompletion of testing toidentical protocol6M, however intend to extend upto 2 years upon successfulcompletion of testing to identicalprotocol12 MonthsIdentical to PD
Non-pyrogenic?YesYesYesIdentical to PD & RD
ConnectionCable?YesYesYesIdentical to PD & RD
Single Use orMulti-UseConnectionCable?Single UseSingle UseSingle UseIdentical to PD & RD
CompatibleRF GeneratorValleyLabs Force 2Electrosurgical GeneratorCross Vascular RF GeneratorRFG-01-00Valleylabs Force 2 ElectrosurgicalGeneratorIdentical to RD

Table 6: Substantial Equivalence Comparison Table

Circa Scientific Inc.

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

Performance Standards

The CrossWise RF Transseptal Access System has been developed in conformance with the following standards:

  • ISO 10993-1:2018, Biological evaluation of medical devices Part 1: Evaluation and ● testing within a risk management process
  • FDA Guidance: Use of International Standard ISO 10993-1, "Biological evaluation of ● medical devices - Part 1: Evaluation and testing within a risk management process", September 2023
  • . IEC 60601-1:2020, Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance
  • . IEC 60601-1-2:2014+A1:2020, Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests
  • IEC 60601-2-2:2023, Medical Electrical Equipment Part 2-2: Particular Requirements . For The Basic Safety And Essential Performance Of High Frequency Surgical Equipment And High Frequency Surgical Accessories
  • ISO 11135:2014, Sterilization of health-care products Ethylene oxide -. Requirements for the development, validation and routine control of a sterilization process for medical devices
  • ISO 10993-7:2008, Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals
  • AAMI TIR42:2021 Evaluation of Particulate Associated with Vascular Medical Devices
  • ISO 11607-1:2019, Packaging for terminally sterilized medical devices Part 1: . Requirements for materials, sterile barrier systems and packaging systems
  • ASTM D4169:2022, Standard Practice for Performance Testing of Shipping Containers . and Systems
  • ISO 15223-1:2021, Medical devices Symbols to be used with medical device labels, ● labeling and information to be supplied
  • ISO 20417:2021, Medical devices Information to be supplied by the manufacturer ●
  • ISO 11070-1:2014, Sterile single-use intravascular introducers, dilators and guidewires ●
  • ISO 80369-7:2021. Small-bore connectors for liquids and gases in healthcare . applications - Part 7: Connectors for intravascular or hypodermic applications

{10}------------------------------------------------

Nonclinical Performance Data

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility

Biocompatibility testing was performed on the CrossWise RF Transseptal Access System in accordance with ISO 10993-1, Biological evaluation of medical devices - Part I: Evaluation and testing within a risk management process. The panel of biocompatibility testing included the following recommended tests:

  • Cytotoxicity ●
  • Irritation/Intracutaneous Reactivity ●
  • Acute Systemic Toxicity ●
  • Sensitization ●
  • Hemocompatibility: ●
    • o Hemolysis, Direct and Indirect Methods
    • Complement Activation o
    • o Partial Thromboplastin Time (PTT)
    • Heparinized Platelet and Leukocyte (PL&L) Count Assay with Comparison Article O
    • In Vivo Thrombogenicity in Canine o
  • Material Mediated Pyrogenicity ●

The results demonstrate that the CrossWise RF Transseptal Access System meets the requirements of ISO 10993-1 and is biocompatible for its intended use.

Sterilization

Sterilization validation adoption was performed on the CrossWise RF Transseptal Access System with CrossWise RF Adapter Cable in accordance with ISO 11135:2014 Sterilization of health-care products- Ethylene Oxide- Requirements for development, validation and routine control of a sterilization process for medical devices. The CrossWise RF Transseptal Access System with CrossWise RF Adapter Cable is subjected to a similar ethylene oxide (EO) sterilization process as the Predicate Devices to meet a sterility assurance level (SAL) of 10°.

Electrical Safety and Electromagnetic Compatibility (EMC)

Electrical safety and EMC testing were conducted on the CrossWise RF Transseptal Access System with CrossWise RF Adapter Cable. The testing complies with the applicable sections of IEC 60601-1:2020, Medical electrical equipment – Part 1: General requirements for basic safety and essential performance, IEC 60601-1-2: 2014 +A1: 2020, Medical electrical equipment -Part 1-2: General requirements for the basic safety and essential performance – Collateral standard: Electromagnetic compatibility – Requirements and tests, and IEC 60601-2-2:2023, Medical electrical equipment – Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical

Circa Scientific Inc.

{11}------------------------------------------------

accessories.

This testing is consistent with that conducted by the Predicate Devices.

Bench Testing

Design verification testing was performed on the CrossWise RF Transseptal Access System with CrossWise RF Adapter Cable at two time points: immediately after manufacturing (T=0) and after six months of accelerated aging (T=6M AA). Devices were subjected to 2X sterilization and distribution simulation before the following types of testing was conducted:

CrossWise RF Transseptal Access System:

  • Packaging Integrity
  • Label Integrity ●
  • . Visual & Dimensional
  • Introducer Set Compatibility
  • Electrical Functionality/Compatibility ●
  • Electrical Safety
  • Mechanical Functionality
  • Mechanical Durability ●
  • Puncture Performance
  • Particulate Generation
  • Radiopacity ●
  • . Corrosion Resistance

CrossWise RF Adapter Cable:

  • Packaging Integrity ●
  • Label Integrity ●
  • Visual & Dimensional ●
  • Electrical Functionality and Compatibility ●
  • Electrical Safety ●
  • Mechanical Functionality ●
  • Mechanical Durability

Conclusions

The CrossWise RF Transseptal Access System with CrossWise RF Adapter Cable is of similar materials and design of the PD/RD devices and has similar technical requirements. The nonclinical bench data support the safety of the device and demonstrate that the CrossWise RF Transseptal System performs as intended in the specified use conditions.

The CrossWise RF Transseptal Access System with CrossWise RF Adapter Cable is substantially equivalent to the Predicate (Table 3) and Reference (Table 4) Devices.

§ 870.5175 Septostomy catheter.

(a)
Identification. A septostomy catheter is a special balloon catheter that is used to create or enlarge the atrial septal defect found in the heart of certain infants.(b)
Classification. Class II (performance standards).