K Number
K241394

Validate with FDA (Live)

Date Cleared
2024-12-20

(218 days)

Product Code
Regulation Number
862.1155
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The iHealth® Early Pregnancy Test is an OTC in vitro diagnostic test used to assess early pregnancy by detecting the presence of the hCG (human chorionic gonadotropin) hormone in urine. The test device is intended for use as an aid in early detection of pregnancy, in some cases as early as six (6) days before the missed period, i.e., as early as five (5) days before the day of the expected period.

The iHealth® Early Pregnancy Test Strip is an OTC in vitro diagnostic test used to assess early pregnancy by detecting the presence of the hCG (human chorionic gonadotropin) hormone in urine. The test device is intended for use as an aid in early detection of pregnancy, in some cases as early as six (6) days before the missed period, i.e., as early as five (5) days before the day of the expected period.

Device Description

The iHealth® Early Pregnancy Test is a lateral flow immunoassay technique device designed for the qualitative determination of human chorionic gonadotropin (hCG) concentration in human urine samples, and is indicated for use as an aid in early detection of pregnancy. The iHealth® Early Pregnancy Test is designed to detect as early as six (6) days before the day of the missed period, i.e., as early as five (5) days before the day of the expected period. Test result is displayed to the user as two lines for a 'Pregnant' result and one line for a 'Not Pregnant' result.

The iHealth® Early Pregnancy Test are in two different formats: Stick and Strip format. Two test formats use identical strips and each test strip in the device consists of: Sample pad, colloidal gold binding pad, nitrocellulose membrane, absorbent paper and PVC board, colloidal gold binding pad coated with hCG monoclonal antibody 1, test line(T Line) coated with hCG-monoclonal antibody 2 and control line(C Line) coated with goat anti-mouse IgG antibody.

Two formats have the same performance specifications. The difference is that Stick format is designed to be tested in dip or midstream method, and Strip format is tested in dip method only.

AI/ML Overview

The provided document describes the iHealth® Early Pregnancy Test and iHealth® Early Pregnancy Test Strip, which are over-the-counter in vitro diagnostic devices used to detect human chorionic gonadotropin (hCG) in urine for early pregnancy detection.

Here's a breakdown of the acceptance criteria and the studies performed:

1. A table of acceptance criteria and the reported device performance:

The document doesn't explicitly state "acceptance criteria" in a separate table. However, based on the performance characteristics presented, the key acceptance criteria for this type of device would typically revolve around its analytical sensitivity (detection limit) and its ability to accurately classify pregnant and non-pregnant states. The sensitivity of hCG detection is a critical performance metric for early pregnancy tests.

Table 1: Key Performance Metrics for iHealth® Early Pregnancy Tests

Performance MetricAcceptance Criteria (Implied)Reported Device Performance
Analytical Sensitivity (hCG detection)Detect 10 mIU/mL hCG with 100% positivity.100% positive at 10 mIU/mL hCG for all formats and lots tested.
Specificity (False Positive Rate)No false positives in non-pregnant individuals.0% false positives in 300 urine samples from non-pregnant premenopausal, peri-menopausal, and postmenopausal women.
Hook EffectNo hook effect at high hCG concentrations.No hook effect observed up to 1,000,000 mIU/mL.
Shelf-Life Stability24 months at 2°C - 30°C.24 months met at 36°F - 86°F (2°C - 30°C).
Cross-Reactivity (hLH, hFSH, hTSH)No cross-reactivity at specified concentrations.No cross-reactivity observed with 500 mIU/mL hLH, 1000 mIU/mL hFSH, 1000 µIU/mL hTSH.
Effect of hCG β-core fragmentNo effect on performance at specified concentrations.No effect observed up to 1,000,000 pmol/L.
Interfering SubstancesNo interference from common substances.No interference effect observed at tested concentrations of numerous substances listed (e.g., Acetaminophen, Glucose, Hemoglobin).
Urine pH effectNo effect on performance across physiological pH range.No effect observed for pH ranging from 4.0 to 9.0.
Urine Specific Gravity effectNo effect on performance across physiological specific gravity range.No effect observed for specific gravity ranging from 1.000 to 1.035.
Method Comparison (Conformity with Predicate)100% conformity with predicate device.100% conformity between candidate and predicate for both positive and negative samples.
Lay User Study (Conformity with Professional)100% conformity with professional results.100% positive and 100% negative conformity observed.
Lay User Spiked Standard (10 mIU/mL)100% positive detection by lay users.100% positive for 10 mIU/mL hCG.
Lay User Spiked Standard (0, 2, 3 mIU/mL)0% positive detection by lay users.0% positive for 0, 2, 3 mIU/mL hCG.

2. Sample sizes used for the test set and the data provenance:

  • Analytical Sensitivity (Precision):

    • Test Set Sample Size: 405 individual tests for each hCG concentration (0, 3, 5, 7.5, 10, 12.5, 15, 25 mIU/mL) for each device format (Strip, Stick-dip, Stick-midstream), totaling 8 concentrations x 405 tests = 3240 tests per device type. The total number of individual readings is significantly higher due to multiple lots and operators. For the 5 mIU/mL the total samples are 170 negative and 235 positive in the Strip format.
    • Data Provenance: Not explicitly stated, but the study was performed "at three sites" by "nine operators," suggesting a multi-site internal study. The samples were "Not-pregnant female urine was spiked with hCG traceable to the 5th WHO international standard".
  • Specificity Study:

    • Test Set Sample Size: 300 urine samples (100 from premenopausal, 100 from peri-menopausal, 100 from postmenopausal women).
    • Data Provenance: Not explicitly stated, but implies collected urine samples.
  • Hook Effect Test:

    • Test Set Sample Size: Not explicitly quantified, but performed on urine samples spiked with hCG at 0, 10, 1,000, 10,000, 100,000, and 1,000,000 mIU/mL.
  • Cross-Reactivity & Interfering Substances:

    • Test Set Sample Size: Not explicitly quantified. Performed on negative and 10 mIU/mL hCG positive urine samples spiked with various substances.
  • Method Comparison Study:

    • Test Set Sample Size: 204 pregnant and non-pregnant subjects. 109 positive samples and 95 negative samples in total across both formats. iHealth® Early Pregnancy Test Strip: 103 samples (54 positive, 49 negative). iHealth® Early Pregnancy Test Stick: 101 samples (55 positive, 46 negative).
    • Data Provenance: Not explicitly stated, but implies collected samples from subjects.
  • Lay User Study (Conformity with Professional):

    • Test Set Sample Size: 204 pregnant and not pregnant women subjects. iHealth® Early Pregnancy Test Strip: 103 subjects (54 positive, 49 negative). iHealth® Early Pregnancy Test Stick: 101 subjects (55 positive, 46 negative).
    • Data Provenance: Women subjects with diverse educational and professional backgrounds and ages between 18 and 55 years old who tested their own urine samples.
  • Lay User Spiked Standard Study:

    • Test Set Sample Size: 100 women.
    • Data Provenance: Women with diverse educational and professional backgrounds (18-55 years old) who performed testing using spiked samples.
  • Detection of hCG in Early Pregnancy Clinical Samples:

    • Test Set Sample Size: 83 subjects for each day relative to expected missed period (-8 to +1).
    • Data Provenance: Early pregnancy urine samples collected from subjects.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Analytical Sensitivity (Precision): The results were generated by "nine operators." Their specific qualifications are not detailed, but their involvement across three sites suggests trained personnel in a laboratory or clinical setting.
  • Method Comparison and Lay User Studies: The "professional testing" serves as the reference standard against which the lay user and candidate devices were compared. The number and qualifications of these "professionals" are not specified in the document.
  • Detection of hCG in Early Pregnancy Clinical Samples: The "early pregnancy urine samples" themselves imply a clinical diagnosis of pregnancy, which would typically be established by medical professionals. However, the number and qualifications of these professionals are not explicitly stated.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

The document does not describe an explicit adjudication method for reconciling conflicting results, such as 2+1 or 3+1. For the analytical studies, the results are presented as counts of positive/negative outcomes, implying a direct outcome rather than expert adjudication. For the comparison studies, results are compared to a "professional" result, which acts as a gold standard, so an adjudication process for discrepancies would implicitly fall under how the "professional" result was determined.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This is not applicable to the device described. The iHealth® Early Pregnancy Test is an in vitro diagnostic (IVD) device that provides a visual result (lines on a stick/strip) for human interpretation. It is not an AI-assisted diagnostic tool for human readers interpreting complex medical images or data. Therefore, an MRMC study assessing AI assistance is not relevant here.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This is not applicable. The device provides a visual output that is then interpreted by the user (human-in-the-loop). Even the "professional testing" involves human interpretation of the device's output. The device itself is not a standalone algorithm that provides a diagnosis without human interaction/interpretation of the visual signal.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Analytical Sensitivity (Precision): The ground truth was established by spiking not-pregnant female urine with known concentrations of hCG traceable to the 5th WHO international standard. This is a scientific, controlled ground truth.
  • Specificity: The ground truth was based on urine samples from normal, non-pregnant women. Clinical status (non-pregnant) served as the ground truth.
  • Method Comparison and Lay User Studies: The ground truth for these studies was the result obtained by professional testing using the candidate device or the predicate device. This implies a reference method result, likely also interpreted by professionals.
  • Detection of hCG in Early Pregnancy Clinical Samples: The ground truth for "early pregnancy urine samples" implies a clinical diagnosis of pregnancy based on medical assessment, which is outcomes data (known pregnancy status) rather than just a subjective expert consensus on the device's result.

8. The sample size for the training set:

The document describes performance studies (analytical and clinical comparisons) but does not mention a training set sample size. This is expected as the iHealth® Early Pregnancy Test is a lateral flow immunoassay, which is a chemical/biological test, not a machine learning model that requires a "training set."

9. How the ground truth for the training set was established:

Since there is no "training set" in the context of a machine learning model for this device, this question is not applicable. The device's performance is intrinsically linked to the chemical properties of the antibodies and reagents used, validated through the performance studies described.

{0}------------------------------------------------

December 20, 2024

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Andon Health Co., Ltd. Liu Yi Official Correspondent No 3. Jinping Street, Ya An Road, Nankai District Tianiin. China

Re: K241394

Trade/Device Name: iHealth® Early Pregnancy Test; iHealth® Early Pregnancy Test Strip Regulation Number: 21 CFR 862.1155 Regulation Name: Human Chorionic Gonadotropin (HCG) Test System Regulatory Class: Class II Product Code: LCX Dated: November 15, 2024 Received: November 15, 2024

Dear Liu Yi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Joseph A. Digitally signed by
Joseph A. Joseph A. Kotarek -S Kotarek -S Date: 2024.12.20
12:59:38 -05'00' Joseph Kotarek, Ph.D. Branch Chief Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K241394

Device Name iHealth® Early Pregnancy Test iHealth® Early Pregnancy Test Strip

Indications for Use (Describe)

The iHealth® Early Pregnancy Test is an OTC in vitro diagnostic test used to assess early pregnancy by detecting the presence of the hCG (human chorionic gonadotropin) hormone in urine. The test device is intended for use as an aid in early detection of pregnancy, in some cases as early as six (6) days before the missed period, i.e., as early as five (5) days before the day of the expected period.

The iHealth® Early Pregnancy Test Strip is an OTC in vitro diagnostic test used to assess early pregnancy by detecting the presence of the hCG (human chorionic gonadotropin) hormone in urine. The test device is intended for use as an aid in early detection of pregnancy, in some cases as early as six (6) days before the missed period, i.e., as early as five (5) days before the day of the expected period.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

1. Submitter's Information

Name:Andon Health Co., Ltd.
Address:No 3, Jinping Street, Ya An Road, Nankai District, Tianjin, P.R. China Tianjin China
Phone Number:86-22-87611660
Contact:Mr. Liu Yi
Email:fda@jiuan.com
2. Device Information
Trade Name:iHealth® Early Pregnancy TestiHealth® Early Pregnancy Test Strip
Model:IHC-3000
Common Name:Human chorionic gonadotropin (HCG) test system
510(k) NumberK241394
Device class:Class II
Classification Name:Kit, Test, Pregnancy, Hcg, Over The Counter
Regulation Number:21 CFR 862.1155
Product Code(s):LCX
Panel:Clinical Chemistry (75)
3. Predicate Device
Manufacturer:SPD Swiss Precision Diagnostics GmbH
Device Name:Clearblue® Early Pregnancy Test
510(k) Number:K213379
ClassificationII
  • LCX Product Code :

{5}------------------------------------------------

4. Device Description

The iHealth® Early Pregnancy Test is a lateral flow immunoassay technique device designed for the qualitative determination of human chorionic gonadotropin (hCG) concentration in human urine samples, and is indicated for use as an aid in early detection of pregnancy. The iHealth® Early Pregnancy Test is designed to detect as early as six (6) days before the day of the missed period, i.e., as early as five (5) days before the day of the expected period. Test result is displayed to the user as two lines for a 'Pregnant' result and one line for a 'Not Pregnant' result.

The iHealth® Early Pregnancy Test are in two different formats: Stick and Strip format. Two test formats use identical strips and each test strip in the device consists of: Sample pad, colloidal gold binding pad, nitrocellulose membrane, absorbent paper and PVC board, colloidal gold binding pad coated with hCG monoclonal antibody 1, test line(T Line) coated with hCG-monoclonal antibody 2 and control line(C Line) coated with goat anti-mouse IgG antibody.

Two formats have the same performance specifications. The difference is that Stick format is designed to be tested in dip or midstream method, and Strip format is tested in dip method only.

5. Intended Use

The iHealth® Early Pregnancy Test is an OTC in vitro diagnostic test used to assess early pregnancy by detecting the presence of the hCG (human chorionic gonadotropin) hormone in urine. The test device is intended for use as an aid in early detection of pregnancy, in some cases as early as six (6) days before the day of the missed period, i.e., as early as five (5) days before the day of the expected period.

The iHealth® Early Pregnancy Test Strip is an OTC in vitro diagnostic test used to assess early pregnancy by detecting the presence of the hCG (human chorionic gonadotropin) hormone in urine. The test device is intended for use as an aid in early detection of pregnancy, in some cases as early as six (6) days before the day of the missed period, i.e., as early as five (5) days before the day of the expected period.

6. Comparison to Predicate Device

The following table is the comparison summary of the proposed subject device and predicate device.

ItemSubject DevicePredicate Device (K213379)
iHealth® Early Pregnancy TestClearblue® Early Pregnancy Test

{6}------------------------------------------------

iHealth® Early Pregnancy TestStrip
Similarities
Intended UseQualitative detection of humanhCG for an aid in early detectionof pregnancy.Same
'Early Test'ClaimThe test device is intended for use asan aid in early detection ofpregnancy, in some cases as early assix (6) days before the day of themissed period, i.e. as early as five (5)days before the day of the expectedperiod.Same
Target UserOver-The-Counter useSame
Sample MatrixUrineSame
AnalytehCGSame
Read Time3-5 minutesSame
hCGSensitivity10mIU/mLSame
Test PrincipleLateral flow sandwich immunechromatographic assay with visualresult displaySame
Assay MobilePhaseGold conjugateSame
ControlMechanismVisualSame
ResultsDisplayVisual Parallel Line2 Lines = Pregnant1 Line = Not PregnantSame
Test TypeQualitativeSame
Differences
Device FormatTwo formats: Strip and StickOne format
SampleapplicationStrip format: dipStick format: midstream and dipIn-stream and dip
TraceabilityWHO 5th International Standard forhCGWHO 4th International Standard forhCG
AnalyteDetectionRecognizes intact hCG and hCG β-subunitDetects intact hCG.Scavenger system to removeintact hCG in the presence of

{7}------------------------------------------------

FSH.

7. Performance characteristics

A. Analytical performance

a. Precision/Sensitivity

Not-pregnant female urine was spiked with hCG traceable to the 5th WHO international standard with hCG concentrations of 0, 3, 5, 7.5, 10, 12.5, 15 and 25mIU/mL. Each standard was tested with devices from three different lots for each format. The study was performed by nine operators over five days at three sites.

The results are summarized in the tables below:

hCGconcentration(mIU/mL)Lot 1Lot 2Lot 3Total result
Negative(n)Positive(n)Positive(%)Negative(n)Positive(n)Positive(%)Negative(n)Positive(n)Positive(%)Negative(n)Positive(n)Positive(%)
013500135001350040500
313500135001350040500
5587757.0%538260.7%597656.3%17023558.0%
7.5419469.6%439268.1%3510074.1%11928670.6%
100135100%0135100%0135100%0405100%
12.50135100%0135100%0135100%0405100%
150135100%0135100%0135100%0405100%
250135100%0135100%0135100%0405100%

iHealth® Early Pregnancy Test Strip format:

iHealth® Early Pregnancy Test Stick format (dip method)

hCGconcentration(mIU/mL)Lot 1Lot 2Lot 3Total result
Negative(n)Positive(n)Positive(%)Negative(n)Positive(n)Positive(%)Negative(n)Positive(n)Positive(%)Negative(n)Positive(n)Positive(%)
013500135001350040500
313500135001350040500

{8}------------------------------------------------

iHealth® Early Pregnancy Test FDA 510(k) Files

5627354.1%587757.0%686749.6%18821753.6%
7.5488764.4%379872.6%439268.1%12827768.4%
100135100%0135100%0135100%0405100%
12.50135100%0135100%0135100%0405100%
150135100%0135100%0135100%0405100%
250135100%0135100%0135100%0405100%

iHealth® Early Pregnancy Test Stick format (midstream method)

hCGLot 1Lot 2Lot 3Total result
concentration( mIU/mL )Negative(n)Positive(n)Positive(%)Negative(n)Positive(n)Positive(%)Negative(n)Positive(n)Positive(%)Negative(n)Positive(n)Positive(%)
013500135001350040500
313500135001350040500
5676850.4%637253.3%666951.1%19620951.6%
7.5449167.4%488764.6%498663.7%14126465.2%
100135100%0135100%0135100%0405100%
12.50135100%0135100%0135100%0405100%
150135100%0135100%0135100%0405100%
250135100%0135100%0135100%0405100%

Based on the above results, the sensitivity of iHealth® Early Pregnancy Test and iHealth® Early Pregnancy Test Strip is demonstrated to be 10 mIU/mL.

b. Linearity/Assay measuring range

N/A. This is a qualitative detection device.

c. Hook effect test

Not-pregnant female urine was spiked with hCG to concentrations of 0 mIU/mL, 10mIU/mL, 1,000 mIU/mL, 10,000 mIU/mL, 100,000 mIU/mL and 1,000,000 mIU/mL. The results demonstrated that no hook effect was observed at hCG concentration up to 1,000,000 mIU/mL.

d. Stability

{9}------------------------------------------------

The iHealth® Early Pregnancy Test and iHealth® Early Pregnancy Test Strip have been verified for shelf-life stability for 24 months at both 36°F - 86°F (2°C - 30°C).

e. Specificity and cross reactivity

Specificity study to determine false-positive result rate

To evaluate specificity, 300 urine samples were collected from normal, non-pregnant premenopausal (ages 1840 years old), peri-menopausal (4155 years old) and postmenopausal (>55 years old). 100 people for each age group. Each urine sample was tested with iHealth® Early Pregnancy Test Strip format, iHealth® Early Pregnancy Test Stick format (dip method and midstream method). No false positive results were observed for any of the age groups.

Effect of cross reactivity

To evaluate cross-reactivity, negative and positive urine samples (0 and 10 mIU/mL hCG) were spiked with potential cross reactants (500 mIU/mL hLH, 1000 mIU/mL hFSH, 1000 µIU/mL hTSH). No cross-reactivity was observed at tested concentration.

Effect of hCG β -core fragment (hCG β cf)

To evaluate the effect of the hCG ß-core fragment, negative and positive urine samples (0 and 10 mIU/mL hCG) were spiked with hCG ß-core fragment at concentrations of 0 pmol/L, 250,000 pmol/L, 500,000 pmol/L and 1,000,000 pmol/L. The performance of iHealth® Early Pregnancy Test was not affected by hCG ß-core fragment concentrations up to 1,000,000 pmol/L.

f. Interfering substance

Interference from exogenous and endogenous substances

To evaluate the effect of the potential interfering substance, negative and positive urine samples (0 and 10 mIU/mL hCG) were spiked with the interfering substance to obtain the certain desired test concentration. No interference effect was observed at the tested concentration shown in table below:

SubstancesConcentration
Acetaminophen20mg/dL
Acetoacetic Acid500mg/dL
Acetylsalicylic Acid20mg/dL
Albumin2000mg/dL
Ampicillin20mg/dL
Ascorbic Acid20mg/dL
Atropine20mg/dL
Benzoylecgonine10mg/dL

{10}------------------------------------------------

B-hydroxybutyrate1000mg/dL
Bilirubin2mg/dL
Caffeine20mg/dL
Cannabinol10mg/dL
Codeine6μg/dL
EDTA80mg/dL
Estriol-17-beta1400μg/dL
Ethanol1.00%
Gentisic Acid20mg/dL
Glucose2000mg/dL
Hemoglobin500mg/dL
Ketone20mg/dL
Methanol10%
Phenothiazine20mg/dL
Phenylpropanolamine20mg/dL
Pregnanediol1500μg/dL
Salicylic Acid20mg/dL
Tetracycline20mg/dL
Thiophene20mg/dL

Effect of urine pH

To evaluate the effect of urine pH, negative and positive urine samples (0 and 10 mIU/mL hCG) were tested at pH values of 4, 5, 6, 7, 8 and 9. The results demonstrate that pH ranging from 4.0~9.0 in urine samples does not affect the performance of iHealth® Early Pregnancy Test and iHealth® Early Pregnancy Test Strip.

Effect of urine specific gravity

To evaluate the effect of urine specific gravity, negative and positive urine samples (0 and 10 mIU/mL hCG) were tested at urine specific gravity of 1.000, 1.005, 1.015, 1.030 and 1.035. The results demonstrate that urine with a relative specific gravity ranging from 1.000 to 1.035 does not affect the performance of iHealth® Early Pregnancy Test and iHealth® Early Pregnancy Test Strip.

B. Comparison study

a. Method comparison study

The performance of the new device was compared to the predicate device. A total of 204 pregnant and non-pregnant subjects collected samples and completed the study to obtain

{11}------------------------------------------------

a valid result. 109 were positive samples and 95 were negative samples. All samples were tested with candidate and predicate devices by professionals.

The results are summarized in the table below:

Predicate device
Candidate DevicePositiveNegativetotal
iHealth®EarlyPregnancyTest StripPositive54054
Negative04949
total5449103

iHealth® Early Pregnancy Test Strip format (dip method)

iHealth® Early Pregnancy Test Stick format (dip and midstream methods)

Candidate DevicePredicate device
PositiveNegativetotal
iHealth®EarlyPregnancyTestPositive55055
Negative04646
total5546101

The conformity between the candidate device (iHealth® Early Pregnancy Test and iHealth® Early Pregnancy Test Strip) and the predicate device is 100%.

b. Lay user study

Lay User Study

A total of 204 pregnant and not pregnant women subjects with diverse educational and professional backgrounds and ages between 18 and 55 years old participated in the Lay User Study. Each subject tested her own urine sample using iHealth® Early Pregnancy Test and iHealth® Early Pregnancy Test Strip according to the labeling and provided a sample for professional testing.

The results are summarized in the table below:

iHealth® Early Pregnancy Test Strip format (dip method)

Candidate DeviceProfessional
PositiveNegativetotal
Lay userPositive54054
Negative04949
total5449103

{12}------------------------------------------------

iHealth® Early Pregnancy Test Stick format (dip and midstream methods)
Lay userCandidate DeviceProfessional
PositivePositiveNegativetotal
55055
Negative04646
total5546101
iHealth® Early Pregnancy Test Stick format (dip and midstream methods)

From the above tables, the lay person results showed 100% positive and 100% negative conformity with the professional results.

Each lay user was given a questionnaire to assess the comprehensibility of the labeling. The results of the questionnaire reflected that the consumers found the test easy to use and that they did not have trouble understanding the labeling and interpreting the results.

Lay User Spiked Standard

The second lay user study was conducted with a total of 100 women with diverse educational and professional backgrounds between the ages of 18 years to 55. Lay users performed the testing using samples spiked with 0, 2, 3, 5, 10 and 15 mIU/mL hCG using iHealth® Early Pregnancy Test and iHealth® Early Pregnancy Test Strip. Results were summarized below.

SamplenumberhCGconcentrationLay users Results% positive
Number ofpositiveNumber ofnegative
B0 mIU/mL01000%
A2 mIU/mL01000%
E3 mIU/mL01000%
D5 mIU/mL465446%
F10 mIU/mL1000100%
C15 mIU/mL1000100%

iHealth® Early Pregnancy Test Strip format

iHealth® Early Pregnancy Test Stick format (dip method)
SamplenumberhCGconcentrationLay users Results% positive
Number ofpositiveNumber ofnegative
B0 mIU/mL01000%
A2 mIU/mL01000%
E3 mIU/mL01000%

{13}------------------------------------------------

D5 mIU/mL435743%
F10 mIU/mL1000100%
C15 mIU/mL1000100%

iHealth® Early Pregnancy Test Stick format (simulated midstream method)

SamplenumberhCGconcentrationLay users Results% positive
B0 mIU/mL01000%
A2 mIU/mL01000%
E3 mIU/mL01000%
D5 mIU/mL445644%
F10 mIU/mL1000100%
C15 mIU/mL1000100%

C. Other clinical study

Detection of hCG in Early Pregnancy Clinical Samples

Early pregnancy urine samples from day -8 to day +1 of their expected period were collected. The early pregnancy urine samples of each day from 83 subjects were tested using iHealth® Early Pregnancy Test Strip format and iHealth® Early Pregnancy Test Stick format (dip method and midstream method).

The results are summarized in the table below:

Dayrelative toEMPDay relative toMissed PeriodPercentPregnantResult(%)NumberPregnantResultNumber NotPregnant ResultTotalSamples
-8-95%47983
-7-813%117283
-6-728%236083
-5-678%651883
-4-593%77683
-3-499%82183
-2-399%82183
-1-2100%83083

iHealth® Early Pregnancy Test Strip format

{14}------------------------------------------------

0-1100%83083
+10100%83083
Dayrelative toEMPDay relative toMissed PeriodPercentPregnantResult(%)NumberPregnantResultNumber NotPregnant ResultTotalSamples
-8-95%47983
-7-812%107383
-6-727%226183
-5-678%651883
-4-593%77683
-3-499%82183
-2-399%82183
-1-2100%83083
0-1100%83083
+10100%83083

iHealth® Early Pregnancy Test Stick format (dip method)

iHealth® Early Pregnancy Test Stick format (midstream method)

Dayrelative toEMPDay relative toMissed PeriodPercentPregnantResult(%)NumberPregnantResultNumber NotPregnant ResultTotalSamples
-8-96%57883
-7-813%117283
-6-727%226183
-5-678%651883
-4-593%77683
-3-499%82183
-2-399%82183
-1-2100%83083
0-1100%83083
+10100%83083

8. Conclusion

Based on the test principle and performance characteristics of the device, it's concluded that iHealth® Early Pregnancy Test and iHealth® Early Pregnancy Test Strip are substantially equivalent to the predicate device.

{15}------------------------------------------------

§ 862.1155 Human chorionic gonadotropin (HCG) test system.

(a)
Human chorionic gonadotropin (HCG) test system intended for the early detection of pregnancy —(1)Identification. A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class II.(b)
Human chorionic gonadotropin (HCG) test system intended for any uses other than early detection of pregnancy —(1)Identification. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class III.(3)
Date PMA or notice of completion of a PDP is required. As of the enactment date of the amendments, May 28, 1976, an approval under section 515 of the act is required before the device described in paragraph (b)(1) may be commercially distributed. See § 862.3.